Post-stall airfoil performance and vertical-axis wind turbines
Transcript of Post-stall airfoil performance and vertical-axis wind turbines
Post-stall airfoil performance and vertical-axis
wind turbines
John Rainbird∗, Joaquim Peiro† and J. Michael Graham‡
Imperial College, London, SW7 2AZ, United Kingdom
Abstract
Sensitivity of VAWT start-up to aerofoil aerodynamic performance char-
acteristics is analyzed. Existing post-stall aerofoil data and the corrections
used in acquiring it are critiqued. Uncertainties in the existing data large
enough to impact on VAWT start-up modeling are established, with wind
tunnel blockage identified as a potential cause of the discrepancies. Using
a conventional closed-section wind tunnel with blockage corrections applied
to data and a passive, tolerant test section design, tests are conducted on
five NACA 0015 aerofoils with a range of chord lengths. The passive tunnel
was found to be better at minimizing the effects of blockage and it makes
no use of correction formulae which are not strictly applicable when flow is
separated since they are based on potential flow theory. Preliminary results
are presented from a study varying Reynolds number showing little varia-
tion in post-stall performance for the NACA 0015 in the range of Reynolds
numbers from 6× 104 to 2.5× 105.
∗Graduate Student, Department of Aeronautics; [email protected]. AIAA Student
Member.†Senior Lecturer, Department of Aeronautics, [email protected].‡Professor of Unsteady Aerodynamics, Department of Aeronautics,
[email protected]. AIAA Member.
1
Nomenclature
c chord, m
g gap between slat aerofoils in tolerant tunnel, m
p plenum depth, m
s spacing between slat aerofoils, m
t aerofoil thickness, m
Ae model equivalent area, m2
Cd drag coefficient
Cl lift coefficient
Cm moment coefficient about the quarter chord
DS double slatted wall tolerant test section
H tunnel test section height, m
HAWT horizontal axis wind turbine
L tunnel test section length, m
OAR open area ratio
Re Reynolds number
SS single slatted wall tolerant test section
VAWT vertical axis wind turbine
Subscripts
u uncorrected measurement
Symbols
α angle of incidence, degrees
εsc solid blockage factor
εwc wake blockage factor
λ tip-speed ratio
σ standard deviation
A aspect ratio
1 Introduction
When measuring the forces generated by an aerofoil in a closed test-section wind
tunnel experiment, readings are influenced by by factors such as the turbulence
intensity of flow in the wind tunnel, the aspect ratio, A, of the wing under test
(this can impact even two-dimensional tests), the chord-to-height ratio, c/H, of the
2
aerofoil and tunnel, and the length-to-height ratio, L/H, of the tunnel test section.
Through good experiment design and by application of blockage corrections to raw
experimental measurements, the effects of these factors can be minimized, allowing
data close to that expected in equivalent free air conditions to be achieved.
Wind tunnel testing of aerofoils does not usually extend much beyond the
range of incidences −20° ≤ α ≤ +20°, since in most applications aerofoils are
not expected to encounter stall, which occurs at incidences of around 10° - 15°.The blades of wind turbines are a notable exception during the start-up phase of
operation, with the inner regions of horizontal axis wind turbine (HAWT) blades
and the full length of vertical axis (VAWT) blades experiencing apparent incidences
well beyond these limits at low tip-speed ratios, λ. VAWT performance can be
sensitive to minor changes to blades and VAWT modeling techniques are similarly
sensitive to small differences in input blade force data, making accurate post-stall
data critical for the analysis of these machines.
In the range of incidences covered in conventional wind tunnel testing, aerofoils
are near-parallel to flow, so blockage of the tunnel flow is minimal. Flow is for
the most part attached, allowing blockage corrections, based partly on potential
flow theory, to be applied with confidence [1]. Once testing extends past stall, flow
detaches, reducing the applicability of these corrections, with blockage higher due
to the greater area of aerofoil presented normal to the flow.
There is a lack of aerofoil data covering a full range of incidences (0° to 180°for symmetrical aerofoils, 0° to 360° for others), particularly at the low Reynolds
numbers relevant to the modeling of start-up of a VAWT. Differences between
the data sets found are significant to the outcomes of such modeling. This paper
documents an attempt to achieve the best possible post-stall data at low Reynolds
numbers using conventional wind tunnels and a tolerant tunnel test section, applied
to deeply stalled aerofoils for the first time.
In this paper, the term “post-stall” is used to refer to aerodynamic character-
istics in the range of incidences between stall angles, approximately ±10° ≤ α ≤±170°. The term “deep stall” refers to the range of incidences ±20° ≤ α ≤ ±160°and “the immediate vicinity of stall” to 10° < α < 20°.
3
(a) Aerofoil coefficients [2] (b) Turbine CQ envelope
(c) Turbine start-up behaviour
Figure 1: Effects of small changes in post-stall lift on VAWT performance.
2 Sensitivity of VAWT start-up behavior to aero-
foil data
Using Strickland’s multiple-streamtube BEM VAWT model [3] with aerofoil coef-
ficients based on Sheldahl and Klimas’s widely-used dataset [2] the sensitivity of
VAWTs to uncertainty in post-stall performance can be demonstrated. Fig. 1(a)
shows lift and drag coefficients, Cl and Cd, against α for the NACA 0018 aerofoil,
where Cl type A is per the reference, but type B has been artificially modified to
exhibit 10% lower post-stall peaks in Cl (at α ≈ 45°, 135°). Fig. 1(b) shows the
resulting plots of turbine torque coefficient CQ against λ from the BEM code using
Cl types A and B (the same Cd data, per the reference, was used for both runs).
4
Fig. 1(c) contains results from a time-dependent extension to the code showing
development of λ with time from a standing start. The change in post-stall be-
havior leads to a prediction of negative torque production at 0.5 < λ < 1.7. The
turbine is unable to self-start to full rotational velocities, reaching only this first
λ = 0.5 limit of positive torque, in spite of a peak in positive torque production
identical to the unaltered case, at λ = 3.5, and the same upper limit in positive
torque at λ = 6.7. Hill et al. [4] demonstrated similar large impacts on modelled
turbine performance with small changes to lift and drag coefficient input.
In this paper, self-starting ability is defined after Bianchini et al. [5] as when a
turbine accelerates through its entire power curve to its fastest equilibrium state
unaided. Even though the turbine using Cl type B rotates under its own power,
it does not self-start by this definition.
3 Existing Aerofoil data
Few studies have been found that extend over a full range of incidence angles
for any aerofoil profile. All of those that have been found for the NACA 0015,
namely those of Pope [6] and Sheldahl and Klimas [2] are reproduced in Figs 2
and 3, which show Cl and Cd against α respectively. Alongside them are some
results from the current study, for an aerofoil with a c/H of 0.1, and the output of
NASA’s AERODAS model [7] for an infinite aspect ratio, 15% thick aerofoil. Thin
aerofoil theory lift is included in Fig. 2. Pope’s study has c/H = 0.17 and Sheldahl
and Klimas’s is 0.07, so they are similarly blocked to the results of the the current
study included in the figures. All three studies have used techniques to minimize
the impacts of lift interference, solid blockage and wake blockage. The current
study follows the correction methodology explained in Section 4 and Sheldahl and
Klimas use similar methods. Pope used a tunnel equipped with a “breather” to
atmosphere that reduced solid and wake blockage, so the data was corrected for
lift interference only.
The current study was taken at a Reynolds number, Re, of 1.5× 105, Sheldahl
and Klimas’s at Re = 3.6× 105 and Pope’s at Re = 1.23× 106. Reynolds number
dependency is evident in the pre-stall results, with higher maximum Cl reached at a
higher incidence the higher the Reynolds number, with Pope reaching Clmax = 1.20
at α = 15°, Sheldahl and Klimas Clmax = 1.07 at α = 13.7° and the current study
5
Figure 2: Cl vs α for selected post-stall studies.
Clmax = 0.93 at α = 12°. After stall, agreement between Sheldahl and Kilmas and
the c/H = 0.1 is strong, other than for Cd in the range 80° < α < 110°, where
Sheldahl and Kilmas’s results are less smooth. Post-stall lift and drag peaks are
lower for the Pope study, with the lift peaks occurring at approximately 50° and
145° rather than 45° and 140° as for the other studies. These differences are likely
due to the differences in methods used for blockage minimization. A Reynolds
number dependency is hard to discern in the post stall region from this limited
sample size.
Before stall, the AERODAS model replicates performance of chosen exper-
imental data, here that of Sheldahl and Klimas has been used. After stall, a
universal model is applied based on thickness ratio, t/c, andA, with lift and drag
represented as hyperbolic distributions around calculated peaks:
Clmax = 1.190(1.0− (t/c)2
) (0.65 + 0.35e(−(9.0/A)2.3)
)Cdmax = 2.3e(−(0.65t/c)0.9)
(0.52 + 0.48e(−(6.5/A)1.1)
) (1)
The model is based on an assumption that post-stall aerofoil performance is inde-
pendent of Reynolds number (note that there is no input based on flow conditions),
6
Figure 3: Cd vs α for selected post-stall studies.
a view that is supported by Sheldahl and Klimas’s study, which features runs at
Re = 5× 105 and Re = 6.8 × 105 as well as the Re = 3.6 × 105 reproduced here,
with no significant differences between the runs noted in the deep stall region.
Given the limited amount of NACA 0015 data available, NACA 0012 studies
covering a full range of incidences have also been reviewed. The similarity of the
two profiles means that performance in deep stall should also be similar. Fig. 4
shows a composite plot of the three studies included in Figs 2 and 3, alongside
those of Bergeles et al. [8], Critzos et al. [9], Massini et al. [10] and Sheldahl and
Klimas [2] for the NACA 0012. AERODAS output for a 15% thick aerofoil and
thin aerofoil theory Cl are again included. Further details of the studies included in
the figure, along with the rest of the data from the current study and an additional
study of Jacobs [11] referred later on in this paper, are summarized in Table 1.
Note that the study of Critzos et al. features two runs using the same apparatus,
at Re = 5 × 105 and Re = 1.36 × 106, with higher post-stall lift and drag peaks
exhibited for the higher Reynolds number.
Pre-stall agreement of the studies is strong, with lift at low incidences in all
7
Figure 4: Cl and Cd vs α for the c/H = 0.1 aerofoil from this study, alongside
the studies of Bergeles et al [8], Critzos et al [9], Massini et al [10], Sheldahl and
Klimas [2] and Pope [6]
cases comparable to thin aerofoil theory, but post-stall agreement is generally poor.
At the second lift peak at an approximate incidence of 45°, values of Cl are spread
between extremes of 0.98 and 1.18, a difference of 20%. At peak drag at around
90°, Cd is between 1.81 and 2.08, a 15% variation. AERODAS output is at the
upper end of the spread of data. The wide variety of Reynolds numbers at which
the studies were taken make comparisons in the vicinity of stall unwise, given
the sensitivity of performance to this variable here. The modeling presented in
Section 2 shows that a 10% difference in post-stall lift peaks is sufficient to alter
the output of a BEM code from prediction of self-start to non-prediction, below
the level of uncertainty between data sets reproduced here. These uncertainties are
perhaps the reason that numerical models of VAWTs can struggle to predict the
successful self-starting of straight-bladed VAWTs with symmetrical blades without
some additional modification to input aerofoil data [12, 5], in spite of experimental
evidence that self-starting does occur [4, 13, 14].
8
Source Post-stall peak Re c/H A
Cl Cd (×106)
NACA 0012 studies
Bergeles et al. [8] 1.08 1.90 0.76 0.14 1.70
Critzos et al. LTPT [9] 1.13 2.08 1.80 0.07 6.00
Critzos et al. LTPT [9] 1.15 1.96 0.50 0.07 6.00
Critzos et al. Langley 7 × 10
[9]1.14 2.07 1.36 0.10 7.00
Massini et al. [10] 1.18 1.99 0.96 0.13 2.00
Sheldahl and Klimas [2] 1.10 1.83 0.36 0.07 6.00
NACA 0015 studies
c/H = 0.10 (this study) 1.07 1.85 0.06 0.10 10.00
c/H = 0.10 (this study) 1.06 1.86 0.15 0.10 10.00
c/H = 0.10 (this study) 1.06 1.82 0.25 0.10 10.00
c/H = 0.15 (this study) 1.02 1.84 0.15 0.15 6.67
c/H = 0.20 (this study) 1.04 1.83 0.15 0.20 5.00
c/H = 0.25 (this study) 1.09 1.83 0.15 0.25 4.00
c/H = 0.30 (this study) 1.18 1.82 0.15 0.30 3.33
Jacobs [11] - - 0.17 0.081 6.001
Pope [6] 0.98 1.81 1.23 0.17 1.67
Sheldahl and Klimas [2] 1.05 1.86 0.36 0.07 6.00
Table 1: Summary of studies referenced in this paper.
The studies reproduced in Fig. 4 are, to the best of the authors’ knowledge,
all of those available on symmetrical aerofoils for a full range of incidences, other
than the very low Reynolds number work of Zhou [15] and Poisson-Quiton and
de Sievers’s study [16] for which no information about experimental configuration
could be found. Both use the NACA 0012. See Lindenberg [17] for a more complete
1Studies were conducted in a cylindrical test section with a finite wing
9
review of post-stall research, including cambered profiles.
Reynolds number effects alone cannot be responsible for the variation in post-
stall performance seen between the referenced studies, given the non-linearity of
the relationship between it and lift and drag peaks Table 1. For example, the
study of Massini et al.[10] has the highest lift peak, but two Critzos et al. studies
were taken at higher Reynolds numbers. Note though that if the two studies of
Critzos et al. (denoted LTPT in Table 1) and that of Sheldahl and Klimas for the
NACA 0012 are isolated, the impact of other variables can be ignored, since they
share common c/H,A and aerofoil profile. These do seem to suggest that higher
post-stall peaks occur with higher Reynolds number
As mentioned before, the influence of secondary effects on experiments can
be minimized through sensible apparatus design. Introduction of suitable screens
and honeycomb inserts in a wind tunnel with a suitably large contraction ratio
can minimize turbulence in the flow. Ensuring a test section is long enough to
isolate the model under test from flow features caused by the contraction limits
any problems associated with the length-to-height ratio of the tunnel working
section.
End wall effects can introduce some three-dimensionality to separated flow
of an otherwise 2D experiment, leading to the formation of stall cells. These
occur as the stall develops and generate some downwash effects. Aspect ratio of
the aerofoil under test can therefore impact on the lift coefficient around stall.
End wall effects can be limited by using a high A aerofoil, or by utilizing end-
plates to isolate aerofoils from the wind tunnel boundary layer. If these are not
practical, consideration of how stall cells will form on the aerofoil when designing
an experiment (by applying the findings of Weihs and Katz [18], for example) can
also help.
It is not always practical to limit blockage through similar means. Large wind
tunnels are costly to build and operate, while small aerofoils are difficult to man-
ufacture accurately and require faster flows than larger ones to attain the same
Reynolds numbers. Blockage corrections are instead used to process blockage ef-
fects out of results taken in closed jet wind tunnels. Less conventionally, similar
corrections can be used with results taken in open-jet tunnels, or blockage tolerant
test sections can be used. The c/H ratio is an indicator of blockage. The results
summarized in Table 1 suggest that blockage may be affecting them even after the
10
application of corrections.
4 Blockage corrections
Results for this study taken in conventional closed test sections have been corrected
for lift interference, solid blockage and wake blockage using the formulas given in
ESDU 76028[1], modified for use at extreme incidence angles as follows.
The solid blockage factor, εsc is prescribed in the reference[1], as
εsc =πAe
6H2
[1 + 1.2
(t
c
)] [1 + 1.1
(ct
)αu
2]
(2)
where Ae is the equivalent area of the model (the area of the two-dimensional aero-
foil section) and t is its thickness, and αu is uncorrected incidence in radians. The
“incidence effect factor” contained in the second set of square brackets quantifies
the impact of model incidence on blockage. As the impact reduces with increasing
incidence at α > π/2, it has been modified at α > π/2 to[1 + 1.1
(ct
)(π − αu)2
](3)
Total blockage (εsc + εwc, where εwc is the wake blockage factor) is assumed by
ESDU to be small enough to neglect terms of its square. These have been restored
for the current study as the assumption does not hold post-stall. Compressibility
factors have been omitted throughout due to the low Mach numbers of experiments
(< 0.1).
5 Limitations of corrections
As mentioned, corrections are based on potential flow theory and so are limited
by the assumptions of the theory. Flow is assumed to be inviscid, suitable for
the Reynolds numbers of the studies in Table 1 away from boundary layers, and
incompressible, suitable as Mach numbers are below 0.3. The aerofoil is taken to
be small in relation to the tunnel, since it is modeled by a combination of a single
source, vortex and doublet in the derivation of the corrections. This does not
hold true for larger c/H ratios, with ratios that would be acceptable for pre-stall
aerofoil testing being less so at extreme incidences due to the larger area presented
11
normal to flow. Use of potential flow theory also requires an assumption that flow
is attached, which is not the case beyond stall.
Use is still made of corrections where separation exists - the ESDU corrections
used here are partly based on Maskell’s empirical corrections for bluff bodies and
stalled wings [19] so they are, in part at least, suitable for such applications.
ESDU state that use has been made of the corrections for flows with “some degree
of separation”, but that “they should clearly be used with caution” [1] in these
conditions. They also state that use is commonly made of the corrections with c/H
ratios of up to 0.35 [1] (the largest in this study is 0.3) but it should be noted that
this almost certainly applies to experiments covering a narrow range of incidences.
6 The tolerant tunnel
An alternative to blockage corrections which is more compatible with separated
flows and larger blockage is the use of tolerant wind tunnel test sections. ESDU
provide corrections for open jet wind tunnels [1] which are of an opposite sign to
those for closed jets. It stands to reason therefore that by using test sections with
semi-permeable walls blockage constraint can be minimized.
The tolerant test section used in this study is based on the design of Parkinson
[20]. Certain sections of tunnel wall are replaced with aerofoil shaped slats run-
ning perpendicular to flow, with slots between them through which air can enter
and exit the main test section, see Fig. 5, avoiding blockage caused by the model
under test. The slats are aerofoil shaped to limit the risk of flow separation around
them which would impede readings taken from the model. Plenums of stagnant
air enclose the permeable walls to ensure conservation of mass flow is maintained.
The bi-dimensionality of the design inhibits the development of significant trans-
verse flows, which can impede the quality of readings taken from a experiment
attempting to reproduce 2-D flows.
The design is versatile, being capable of minimizing blockage for a variety of
model sizes once a suitable open area ratio (OAR, a measure of open to slatted wall
areas, defined as g/s using the notation in Fig. 5) has been settled on [21]. The
best OAR may be specific to a tunnel and test model paring, and is settled on by
testing a range of models of different sizes but with the same shape in the tunnel
and comparing results. This makes experimentation using this tunnel design more
12
costly and time consuming than conventional solid walled testing; even ignoring
the cost of initial fabrication of the tolerant test section, experiments to select
the best OAR must be completed before main experiments, requiring significantly
longer tunnel time and fabrication of several models of different sizes.
H
gs
c
p
Figure 5: Parkinson’s slatted ceiling tunnel configuration for aerofoil testing.
Parkinson’s design was originally developed as it is shown in Fig. 5 for the
testing of aerofoils before stall. Only the wall opposite the aerofoil’s suction surface
was made semi-permeable, since this surface is the more strongly affected by wall
constraints, and it was thought that the aerofoil under test would draw air from a
lower plenum into the main test section ahead of the model mounting, degrading
the flow past the experiment [22]. In this configuration, an OAR of 0.6 was found
to give the best results for the pre-stall range of incidences, while 0.7 was preferred
for preliminary work on stalled aerofoils at α = 20° [23].
Later work adapted the tunnel design by replacing both the floor and ceiling
with slats and a plenum chamber following numerical modeling which suggested
this would provide better results than the single slatted wall [24]. Both slatted
areas were set to an OAR of 0.59 for this work. Limited experimental results were
presented to support the numerical modeling, with no comparisons made to the
earlier, single-walled work. The double slatted wall design was later applier to the
testing of bluff bodies [21]. A range sizes of cylinders, cylinders with splitter plates
and flat plates were accommodated using similar OARs of around 0.6.
13
7 Experimental apparatus
Experiments were conducted in Imperial College’s 3’ × 3’ low turbulence wind
tunnel. The tunnel test section has a 915mm × 915mm cross section and a length
of 2390mm. The turbulence intensity of this tunnel is less than 0.1% when empty,
making it ideal for low Reynolds number aerofoil testing [25]. A Parkinson-style
test section was fabricated for the tunnel, with a plenum depth of 350mm and
a length of 2178mm. Slats have a NACA 0015 profile and a chord of 90mm.
These dimensions are similar to those of Parkinson’s original tunnel, which had
an identical tunnel height, 92mm chord NACA 0015 slats and a plenum of length
2440mm and depth of 300mm
The test section can be configured with a single slatted wall, double slatted
walls or as a conventional closed jet section. Slats are mounted between aluminum
runners and can be slid along the test section and added or removed at will,
enabling OAR to be adjusted.
7.1 Aerofoils
Five NACA 0015 aerofoils were tested, this profile was chosen as it is frequently
used in experimental and numerical VAWT research, and there are some existing
post-stall studies for it against which comparisons can be made (see Figs 2 and 3).
All five aerofoils have a span of 915mm and chords such that their c/H ratios are
0.1, 0.15, 0.2, 0.25 and 0.3.
The aerofoils were manufactured using Nylon selective laser sintering (SLS) 3D
printing technology. Each consists of 4 separate printed parts with a 228.75mm
span (the print chamber of the printer limits the size of parts), with a core of two
915mm long steel rods. Printed Nylon parts can shrink and deform as they cool -
steps were taken to minimize the risk of this happening by limiting wall thickness in
the parts to 3mm, ensuring heat dissipates quickly. Ribbing and cross-struts were
printed into the parts to give them sufficient stiffness. The Nylon sections were
assembled around the steel cores using epoxy resin, with a final smooth surface and
accurate profile achieved using filler, sand paper and paints. While sanding and
filling the aerofoils, profile accuracy was checked against female profile templates,
produced using a laser cutter with a rated accuracy of < 100µm which compares
favorably to the typical accuracy of Nylon SLS printers, around 0.35mm for parts
14
of the size used here.
7.2 Model mounting and data measurement
The aerofoils were mounted vertically in the wind tunnel at the half chord, between
two end plates that sit flush with the tunnel floor and ceiling, see Fig. 6. They
are mounted at the half chord to avoid uneven blocking of the tunnel at extreme
incidences. Each end plate is attached to an adapter plate which in turn attaches
to an ATI Industrial Automation Gamma IP65 6-component force and torque
transducer. The upper transducer is mounted to the the framework of the tunnel
via a bearing unit, see Fig. 7, while the bottom is mounted to a stepper motor
that controls aerofoil incidence, again via a baring, see Fig. 8.
It was decided not to isolate the aerofoils from the wind tunnel boundary layer
due to their insignificant displacement thickness at the model mounting point.
Results presented later in Figs 13 and 15 show little impact of aerofoil aspect ratio
on readings around stall, suggesting any end wall effects caused by the boundary
layers are minimal.
Since the transducers rotate with the aerofoil, forces normal and tangential
to the aerofoil chord are measured, as is moment about the half chord. These
are post-processed into lift, drag and moment about the quarter chord. Drag
is further processed by subtracting the tare drag force associated with the end
plates, obtained by measuring the forces on them when the tunnel is run without
an aerofoil present. For lightly loaded aerofoils, this tare force is likely to be
overestimated at high incidences, where the separated flow of the aerofoil reduces
the velocity over a large portion of the end plates, reducing the friction drag on
them. When aerofoils are loaded highly enough to induce bending in them, the
upstream edges of the endplates encroach into the flow slightly, so the tare drag
is likely to be underestimated. Tare drag is an insignificant proportion of total
drag post-stall, with these under- or overestimates representing an even smaller
proportion of the total, so no adjustment has been made for them.
7.3 Data acquisition and experiment control
Dynamic pressure difference across the tunnel contraction is measured using a mi-
cromanometer, feeding into a PC via a serial port. This is processed into flow
15
Figure 6: The c/H = 0.2 aerofoil mounted in the wind tunnel with double slatted
walls, OAR = 0.55.
velocity with a calibration obtained using a pitot-static tube mounted in place of
a test aerofoil at the tunnel midspan. Flow and aerofoil incidence are controlled by
the PC via the outputs on a National Instruments NI USB-6229 data acquisition
board. The incidence is altered using the stepper motor, with a tracking of its
rotation stored at each data acquisition point. In addition incidence is also mea-
sured by an optical encoder attached to the motor shaft. The tracked incidence
and encoder output are checked after each movement to ensure the motor has
16
Figure 7: The top assembly of the apparatus, with tunnel floor removed, showing
end plate, adapter, force transducer, bearing mount and attachments to the timber
framework of the test section.
not slipped at any point. The output of the force transducers is digitized using a
pair of NI PCI-6220 data acquisition boards, with simultaneous acquisition from
both. If any of of the components of the load cell are saturated during the first 10
samples taken, raw data is saved for further analysis. An average of the readings
is saved for all data points. The transducers are sampled at a rate of 1000 Hz (a
much higher rate than the highest frequencies of the main vibrations induced in
the rig) for 120 seconds for incidences in the range 45° ≤ |α| ≤ 135° and 30 sec-
onds for other incidences, long enough to capture numerous periods of the lowest
frequencies of force fluctuations.
The process is fully automated for a given set of incidences, the PC control
system running from a Matlab script. It rotates the aerofoil to each incidence
prescribed in turn and at each one it checks the tunnel dynamic pressure, adjusting
if necessary, then records this along with temperature, absolute pressure, incidence
according to both motor and encoder and forces and torques from the transducers.
17
Figure 8: The bottom assembly of the apparatus, with tunnel floor removed,
showing end plate, adapter, force transducer, bearing mount and the cooling unit
of the stepper motor.
7.4 Accuracy of measurements
The transducers used have been factory-calibrated within the last two years where
they were rated to a 95% confidence level to within 1% for force measurements
and 1.5% for torque measurements. Recent checks performed on them found no
significant deterioration in this performance.
A small amount of flexibility was introduced to the system to prevent the
transducers from maxing out under moment loads induced by the rig itself (under
no aerodynamic loads). This was achieved by inserting rubber O-rings between
the adapter plates and transducers. The adapters are held to the transducers with
countersunk headed bolts, passing through the O-rings. This flexibility allowed
some “play” in the incidence of the aerofoil, as did a small amount of looseness in
the stepper motor holding torque. Total play in incidence was ≈ ±0.5° when forced
by hand, with the tapered shape of the countersunk heads and the magnetic forces
of the motor ensuring the aerofoil returned to the intended incidence on release.
18
During runs of the experiment, no significant incidence play was noted other than
extreme incidences (in the range 45° ≤ |α| ≤ 135°). The impact on forces of an
incorrect incidence reading here would be less than 0.005 for Cd and 0.02 for Cl
due to the low rate of change of forces with incidence in this region, but given the
design ensures incidence would oscillate about the angle intended, errors induced
by rig flexibility are likely to be far lower.
The largest source of error comes from creep in the force transducers. An offset
is taken at the beginning and end of each run of the experiment with no flow in
the tunnel and these are compared, with any runs where the differences between
the two are overly large discarded. As a measure of the impact of zero movement,
the end offset is processed into force coefficients using the start offset and the flow
conditions for zero incidence for each run. For Reynolds numbers of 1.5×105 , the
maximum values of Cl and Cd calculated using this method are less than 0.02 (or
around 2% at stall Cl) and Cm are less than 0.012.
The maximum errors in the system accounting for all of the above are esti-
mated at approximately 3%, this is consistent with maximum differences between
repeated runs for the same aerofoil and tunnel configuration.
8 Results
Measurements of normal force, tangential force and moment about the half chord
were taken for the five aerofoils for −2° < α < 184°, the test section configured
with solid walls (with blockage corrections applied to data), with a single Parkin-
son slatted wall (SS), with OARs of 0.71, 0.79, 0.83 and 0.88, and with double
Parkinson slatted walls (DS), with OARs of 0.50, 0.55, 0.59, 0.63 and 0.71.
8.1 Selection of data for inclusion in this report
Results are presented for the solid walled tunnel raw and post-processed with data
corrections as described in Section 4. Results are also presented for the SS-0.88
and DS-0.71 tunnel configurations. Solid walled data allows easier comparison to
existing post-stall studies, none of which have utilized Parkinson’s tunnel design,
and also allows assessment of the capabilities of blockage corrections at the ex-
treme incidences and blockages used in this study. The SS-0.88 and and DS-0.71
19
were chosen as the best OARs of their respective tunnel configurations, based on
experimental results, see Section 9 for the methodology used in this judgement.
8.2 Comparisons to existing data
Fig. 9 shows results for the c/H = 0.1 aerofoil taken in the solid walled tunnel,
corrected for blockage. Also included are the data of Sheldahl and Klimas [2] (as
before, taken atRe = 3.6×105) and a study by Jacobs[11], taken atRe = 1.66×105.
All studies are included in Table 1. Thin aerofoil theory and performance for the
NACA 0015 modelled using Xfoil with matched Reynolds number are also included.
Figure 9: Cl and Cd vs α for selected low Reynolds number studies.
Before stall, this study’s results exhibit a gradient greater than 2π up to α = 4°,then a shallower gradient between there and stall, which occurs at 12°, with a Clmax
of 0.93. The steep gradient at α < 4° is caused by a laminar separation bubble near
the trailing edge at these low incidences, which moves towards the leading edge
with increasing incidence. This is a feature of the NACA 0015’s performance at low
20
Reynolds numbers, hence the less pronounced impact on Sheldahl and Klimas’s
data, which features a more mild change in gradient at α = 6.5°. Though the Xfoil
output does not agree to the experimental data well, it features the same change
in gradient at α = 4°, and evidence of the laminar separation bubble is visible in
pressure distributions. See Fig. 10 for the pressure distribution for α = 3°, where
the bubble can be seen on the suction surface at 50-60% of the chord.
Figure 10: Xfoil pressure distribution at α = 3°
Jacobs’s study was conducted with a finite wing ofA = 6. The data has been
corrected to an infinite aspect ratio using methods based on lifting line theory.
Stall point should be reasonably accurately rendered, this occurs at α = 11.34°with Clmax = 0.92, values close to those of the current study, as is expected
given the similar Reynolds numbers of the experiments. The characteristic kink in
the polars of the infinite wings and Xfoil simulation is not present in Jacobs’s lift
curve, as the laminar separation bubble has less of an impact on the performance of
wings since it does not extend across the whole span. Jacobs’s study exhibits a far
smoother lift curve up to its stall point, close to a straight line from α = 0°. Lifting
line theory is not valid beyond stall; though it is assumed Jacobs’s corrections were
applied beyond stall, they have done little to correct behavior here to something
expected of an aerofoil - the study shows the gentler stall characteristics expected
of a finite wing.
The current study and that of Sheldahl and Klimas extend beyond this plot, see
Figs 2 and 3 for plots that continue to 180°. As discussed in Section 3, agreement
21
between both is excellent at 20° < α < 180°, the post-stall drag peaks of the two
studies are identical, at 1.86, and lift peaks less than 1% different. Given the
similar blockage of the two studies and their very close agreement, along with the
findings of Sheldahl and Klimas for Reynolds numbers of 3.6 × 105, 5 × 105 and
6.8× 105, where again little variation in post-stall performance was noted, it can
be concluded that post-stall performance for the NACA 0015 is independent of
Reynolds number, at least in the range of Reynolds numbers 1.5 × 105 < Re <
6.8× 105.
Fig. 11 shows some preliminary results from an extension to the current study,
where Reynolds number was varied. Included is lift and drag for the c/H = 0.1
aerofoil at Re = 1.5 × 105 as in earlier plots in this paper, alongside runs at
Re = 6 × 104 and Re = 2.5 × 105. There is very strong agreement between the
post-stall regions of the lift and drag curves, with results near-identical for all
three Reynolds numbers. Post-stall lift peaks are within 0.5% of each other and
drag peaks within 1.4%. This adds further support to the hypothesis that post-
stall performance of aerofoils is independent of Reynolds number, at least in the
ranges covered by this and Sheldahl and Klimas’s study. Note that the run at
Re = 2.5 × 105 was aborted at α = 100° due to excessive blade loads saturating
the sensors.
8.3 Solid walled tunnel
Fig. 12 shows results from the solid walled tunnel for the c/H = 0.1, 0.15, 0.2,
0.25 and 0.3 aerofoils at Re = 1.5× 105, as taken from the force transducers with
no blockage corrections applied. Fig. 13 shows the same data after application
of corrections. In both figures, the results for the c/H = 0.1 aerofoil have been
emboldened, in part to allow easier comparison to earlier figures which all contain
data from this aerofoil, and also because as the lowest blocked of all the aerofoils,
data from it should be the most reliable.
Comparing the two figures shows that the adjustments made by the corrections
are not trivial, even for the c/H = 0.1 aerofoil the adjustment to lift at the
post-stall peak is more than 5.5%, and more than 12.5% at the drag peak. The
corrections do a good job of collapsing the wide spread of raw data down to a
reasonable level of agreement.
22
Figure 11: Cl and Cd vs α for the c/H = 0.1 aerofoil at Reynolds numbers of
6× 104, 1.5× 105 and 2.5× 105 taken in the solid walled tunnel and corrected for
blockage
Pre-stall agreement between the corrected aerofoils is strong, as is to be ex-
pected given the applicability of potential flow theory where flow is attached and
the low blockage at low incidences. Post-stall agreement is less impressive. While
most of the lift data does cluster, the highest lift peaks are significantly higher
than the others. These are for the c/H = 0.3 aerofoil. The drag of the same aero-
foil is higher than the others at incidences of 40° < α < 60° and 110° < α < 140°,and the lowest of them all at 85° < α < 110°. These results suggest that the
blockage of the largest aerofoil is beyond the capabilities of the corrections used,
with drag data appearing to be over-corrected at the post-stall peak. The results
of the c/H = 0.25 aerofoil are affected in a similar fashion to a less limited extent,
suggesting it too may be too large in comparison to the tunnel for the corrections
to handle.
The c/H = 0.2 aerofoil stalls slightly ahead of the others, at 11° rather than
12°, and lift is lower than for the others in the range of incidences between stall
and 35°; this aerofoil has a slight camber (< 0.5%) on one of its sections due to
warping of the Nylon SLS material which could not be corrected with filling and
sanding. Some anomalous results were noted for the c/H = 0.2 aerofoil for several
23
Figure 12: Cl and Cd vs α for c/H = 0.1, 0.15, 0.2, 0.25 and 0.3 aerofoils taken in
the solid walled tunnel (raw data). c/H = 0.1 results emboldened.
tunnel configurations, it is thought that this slight camber is to blame.
8.4 Tolerant tunnel
Fig. 14 shows results for the SS-0.88 configured tolerant tunnel, where again the
results for the c/H = 0.1 aerofoil have been emboldened. Also included are the
corrected results from the solid walled tunnel for the same aerofoil to allow com-
parison with earlier figures. Pre-stall agreement is reasonable for both lift and
drag for the five aerofoils. Stall points are on the whole lower than for the solid,
corrected data. The c/H = 0.3 aerofoil stalls at 12° as it does in the sold tunnel,
but the c/H = 0.2 stalls at 10° and the remaining studies at 11°.After stall, the collapse of lift is better than for the solid corrected tunnel,
though the tolerant tunnel has similar problems to the correction regime with
accommodating the c/H = 0.3 aerofoil; its lift is larger that of the other aerofoils
in the range of incidences 55° < α < 85° and 105° < α < 145°. Drag collapse is
24
Figure 13: Cl and Cd vs α for c/H = 0.1, 0.15, 0.2, 0.25 and 0.3 aerofoils taken in
the solid walled tunnel and corrected for blockage. c/H = 0.1 results emboldened.
worse at the peak, with Cd of the c/H = 0.3 aerofoil far higher than that of the
others. Parkinson’s student Hameury [21] found that a 33% blocked normal flat
plate was too large for the tolerant tunnel to handle, since it diverted flow to such
an extent that it caused some of the slat aerofoils to stall, it is likely the largest
aerofoil behaves similarly when it is near normal to the flow.
Both lift and drag results for the c/H = 0.1 aerofoil are generally larger in
the post-stall region of incidences than the corrected, solid walled data. At the
post-stall peak at α = 90°, Cdmax = 1.98, 6.2% higher than for the corrected data.
The positive lift peak at α = 40° is 1.12, 5.5% higher.
With an OAR of 0.88, the slatted wall contained three aerofoils. Limited testing
was carried out with no aerofoils in place for an OAR of 1, for 85 < α < 95°, with
results still showing wider spread than the corrected, solid walled data. Though
Parkinson’s preliminary work on stalled aerofoils used a single slatted wall[23], this
research moved on to his double slatted wall bluff body design in search of a better
25
Figure 14: Cl and Cd vs α for c/H = 0.1, 0.15, 0.2, 0.25 and 0.3 aerofoils taken in
the SS-0.88 tunnel. c/H = 0.1 results emboldened. Corrected solid walled data
for the c/H = 0.1 aerofoil also included.
collapse of data.
Fig. 15 shows results for the DS-0.71 tunnel configuration. c/H = 0.1 results
have been emboldened and corrected data from the solid walled tunnel for the same
aerofoil included as before. Pre-stall agreement of the studies is strong, with stall
points at identical incidences to the solid, corrected data (11° for the c/H = 0.2
aerofoil, 12° for all others).
After stall, the collapse of data is excellent, better than for the corrected, sold
wall tunnel or the C-0.88 configuration. Other than a region of over-correcting
of drag, for the c/H = 0.3 aerofoil at incidences of 75° < α < 130° and for the
c/H = 0.25 at incidences of 85° < α < 110°, lift and drag results for all aerofoils
are similar over the whole range of incidences. The largest aerofoil likely forces
slat aerofoils into stall, as in the SS-0.88 tunnel.
Results for the c/H = 0.1 aerofoil are similar to those from the corrected solid
26
Figure 15: Cl and Cd vs α for c/H = 0.1, 0.15, 0.2, 0.25 and 0.3 aerofoils taken in
the DS-0.71 tunnel. c/H = 0.1 results emboldened. Corrected solid walled data
for the c/H = 0.1 aerofoil also included.
walled tunnel, though slightly lower in magnitude in the post-stall region. Post-
stall positive lift peak is 1.03, 3.1% lower, while the drag peak is 1.7% lower at
1.83.
9 Discussion
Parkinson’s double slatted wall bluff-body tolerant tunnel was found to perform
best with an OAR of around 0.6 for flat plates normal to the flow [23]. Experi-
ments were carried out using pressure tapped models with c/H of 0.083, 0.194 and
0.333. The judgement of the best tunnel configuration was made by plotting vari-
ous parameters against OAR. Plots of base pressure coefficients, drag coefficients,
Strouhal number and required blockage corrections were interpreted by seeing at
which OAR they are most similar for the range of model sizes. Standard deviations
of pressure distributions from a reference distribution were judged by assessing at
which OAR they are at a minimum. Pressure distributions for the three models
were also plotted on the same axis for each OAR in turn, with those OARs giving
the best collapse of data judged to be best.
27
Since good quality force coefficients are the aim of the research detailed here,
judgement of the best tunnel configuration has been based on measured forces. In
order to apply the methods used by Hameury[21] to the current study’s results
would require separate analysis of each incidence at which data was taken, since
his flat pate readings were taken for α = 90° so are designed for a singe incidence.
Results for both lift and drag require analysis over a wide range of incidences.
Two metrics have been devised by which to judge the tunnels, the first is a sum
of standard deviations (∑σ). Standard deviations are taken of the data for all
five aerofoils at each incidence at which readings were taken, then summed over
a range of incidences for each tunnel configuration. Fig. 16 shows plots of this
metric for both lift and drag against OAR, for pre-stall ( −2° < α < 10°) and post-
stall (20° < α < 160°) ranges of incidences. The standard deviations have been
calculated using only data from this study, without reference to a benchmark force,
since no reliable benchmark exists given the spread found in results of existing post-
stall studies. The lowest sum of standard deviations should be the best tunnel
configuration, since this indicates the lowest spread of data between the different
sized models.
For the pre-stall performance, the best tunnel overall is the DS-0.71 tolerant
configuration, as it has the lowest∑σ for both lift and drag, with the next best
being the corrected solid-walled tunnel. The best performing single slatted wall
tunnel is the SS-0.71, the nearest OAR tested to 0.6, which Parkinson found to be
the best single slatted configuration [23]. Single slatted OARs below 0.71 were not
tested as the main focus of research is on post-stall aerofoil performance, and post
stall performance of the lower single slatted OARs can be seen to be unimpressive
in Fig. 16(c).
Post-stall, the DS-0.71 is again the best performer for lift with the corrected
solid tunnel in second place. For drag, this order is reversed, mostly due to the
over-correction of the drags of the c/H = 0.25 and c/H = 0.3 aerofoils at extreme
incidences. Omitting the results for these extreme incidences (70° < α < 110°)from the sum of drag standard deviation, the DS-0.71 is judged to be the best
tunnel. The SS-0.88 is the best of the single slatted wall tunnels post-stall.
The second metric is the sum of the differences between the force coefficients of
the smallest and largest aerofoils, Cl c/H=0.3−Cl c/H=0.1 or Cd c/H=0.3−Cd c/H=0.1.
The best performing tunnel should be the one where this is closest to zero. Should
28
(a) Single slatted wall pre-stall performance (b) Double slatted wall pre-stall performance
(c) Single slatted wall post-stall performance (d) Double slatted wall post-stall performance
(e) Key
Figure 16: Summations of standard deviations for lift and drag, plotted against
OAR. Note that while corrected solid walled data has an OAR of 0, it is shown
plotted at 0.9 or 0.75.
the metric be negative, the tunnel over-corrects for blockage. The only configura-
tion to have a negative measurement was the DS-0.71. This suggests the best con-
figuration may lay between FC-0.71 and DS-0.63, which is the third best perform-
ing tunnel overall using the∑σ metric, and has positive
∑Cl c/H=0.3 − Cl c/H=0.1
and∑Cd c/H=0.3 − Cd c/H=0.1. An OAR of 0.67 can be achieved using the 90mm
chord slat aerofoils, but this configuration is still to be tested.
29
Results for the c/H = 0.1 aerofoil, the lowest blocked of the five tested, are
close for both the DS-0.71 tunnel and the corrected, solid-walled tunnel results.
The latter represents current best practice, but makes use of corrections that are
not applicable where flows are separated. The tolerant tunnel does not require
corrections and based on the metrics described above, with the DS-0.71 configura-
tion performs better than current best practice across a range of model sizes, with
potential for the tunnel to perform even better with the as yet untested DS-0.67
configuration
10 Conclusions
Results are presented in a study to establish a reliable set of post-stall aerofoil
data, using conventional, solid walled tunnels and a passive, tolerant design with
slatted walls. Five NACA 0015 aerofoils with c/H ratios of 0.1, 0.15, 0.2, 0.25 and
0.3 were tested over a 0° − 180° of incidence using the conventional test section,
with blockage corrections applied to data, and several configurations of the tolerant
design, where no corrections are required.
The tunnel types were analyzed to assess which test section minimizes the ef-
fects of blockage on the forces measured on the aerofoils. The tolerant tunnel,
configured with two slatted walls with an OAR of 0.71 was found to be the best,
giving the lowest spread between data from the five different sized aerofoils. No-
tably, it reduced blockage effects better than the blockage corrections applied to
solid walled tunnel data, an approach that represents current best practice.
Results for the c/H = 0.1 aerofoil, the smallest and so least blocked of those
tested, are similar for both the DS-0.71 and the corrected, solid walled tunnel.
That two different methods for blockage reduction give comparable data allows
confidence to be placed in it as a close approximation to equivalent free-air values.
The results also strongly agree with the earlier study of Sheldahl and Klimas [2]
in the post-stall range of incidences
Analysis of the tolerant tunnel configurations tested suggest it is possible that
the DS-0.67 configuration, as yet untested, could provide an even better reduction
in blockage effects.
The quality of data from the tolerant tunnel suggests that its use is valid for
stalled aerofoils. Given that no reliance is placed on correction methodologies
30
that are not strictly applicable where flow is separated, and the tunnel is bet-
ter at correcting for blockage than current best practice, it represents a better
way of conducting post-stall experiments, all be it at the expense of greater time
requirements and apparatus costs than conventional solid-walled tunnel testing.
Preliminary results are presented from a study varying Reynolds number using
a conventional, solid-walled tunnel. Experiments were conducted at Re = 6× 104,
1.5×105 and 2.5×105, with little variation in post-stall performance noted. Again,
the study of Sheldahl and Klimas, taken at a Reynolds numbers of 3.6 × 105,
5 × 105 and 6.8 × 105, agrees strongly in the post-stall range of incidences. It
can be concluded that post-stall performance of the NACA 0015 is independent of
Reynolds number effects in the range 6× 104 to 6.8× 105.
Acknowledgments
This work was supported by the Environmental Services Association Education
Trust (ESAET).
31
References
[1] ESDU, “Lift-interference and blockage corrections for two-dimensional sub-
sonic flow in ventilated and closed wind-tunnels,” Tech. Rep. 76028, Engi-
neering Sciences Data Unit, 1978.
[2] Sheldahl, R. and Klimas, P., “Aerodynamic characteristics of seven symmetri-
cal airfoil sections through 180-degree angle of attack for use in aerodynamic
analysis of vertical axis wind turbines,” Tech. Rep. SAND80-2114, Sandia
National Laboratories, 1981.
[3] Strickland, J., “Darrieus turbine: a performance prediction model using mul-
tiple streamtubes,” Tech. Rep. SAND75-0431, Sandia National Laboratories,
1975.
[4] Hill, N., Dominy, R., Ingram, G., and Dominy, J., “Darrieus turbines: the
physics of self-starting,” Proceedings of the Institution of Mechanical Engi-
neers Part A-Journal of Power and Energy , Vol. 223, No. A1, 2009, pp. 21–
29.
[5] Bianchini, A., Ferrari, L., and Magnani, S., “Start-up behavior of a three-
bladed H-Darrieus VAWT: experimental and numerical analysis,” Proceedings
of the ASME Turbo Expo, 2011, pp. 6–10.
[6] Pope, A., “The forces and pressures over a NACA 0015 airfoil through 180
degrees angle of attack,” Tech. Rep. E-102, Georgia Institute of Technology,
1947.
[7] Spera, D., “Models of lift and drag coefficients of stalled and unstalled air-
foils in wind turbines and wind tunnels,” Tech. Rep. NASA/CR-2008-215434,
NASA, 2008.
[8] Bergeles, G., Athanassiadis, N., and Michos, A., “Aerodynamic characteristics
of NACA 0012 aircraft in relation to wind generators,” Wind Engineering ,
Vol. 7, No. 4, 1983, pp. 247–262.
[9] Critzos, C., Heyson, H., and Boswinkle, R., “Aerodynamic characteristics of
NACA 0012 airfoil section at angles of attack from 0 to 180 Degrees,” Tech.
Rep. TN 3361, NACA, 1955.
[10] Massini, G., Rossi, E., and D’Angelo, S., “Wind tunnel measurements of aero-
dynamic coefficients of asymetrical airfoil sections for wind turbine blades
32
extended to high angles of attack,” European Community Wind Energy Con-
ference, Herning, Denmark, June 1988, pp. 241–245.
[11] Jacobs, E. and Sherman, A., “Airfoil section characteristics as affected by
variations of the Reynolds number,” Tech. Rep. TR 586, NACA, 1937.
[12] Rossetti, A. and Pavesi, G., “Comparison of different numerical approaches
to the study of the H-Darrieus turbines start-up,” Renewable Energy , Vol. 50,
No. 2013, 2013, pp. 7–19.
[13] Rainbird, J., The aerodynamic development of a vertical axis wind turbine,
MEng Project Report, University of Durham, UK, 2007.
[14] Dominy, R., Lunt, P., Bickerdyke, A., and Dominy, J., “Self-starting capa-
bility of a Darrieus turbine,” Proceedings of the Institution of Mechanical
Engineers, Part A: Journal of Power and Energy , Vol. 221, No. 1, 2007,
pp. 111–120.
[15] Zhou, Y., Alam, M., Yang, H., Guo, H., and Wood, D., “Fluid forces on a
very low Reynolds number airfoil and their prediction,” International Journal
of Heat and Fluid Flow , Vol. 32, No. 1, 2011, pp. 329–339.
[16] Poisson-Quinton, P. and de Sievers, A., “Etude aerodynamique d’un element
de pale d’helicoptere,” AGARD Conference Proceedings No. 22 , Gottingen,
Germany, December 1967, pp. 4.1–4.35.
[17] Lindenburg, C., “Stall coefficients: aerodynamic airfoil coefficients at large
angles of attack,” IEA Symposium on the Aerodynamics of Wind Turbines ,
National Renewable Energy Laboratory, Colorado, USA, December 2000.
[18] Weihs, D. and Katz, J., “Cellular patterns in post-stall flow over unswept
wings,” AIAA Journal , Vol. 21, No. 12, 1983, pp. 1757–1759.
[19] Maskell, E., “A theory of the blockage effects on bluff bodies and stalled wings
in a closed wind tunnel,” Tech. Rep. R. & M. No. 3400, British ARC, 1963.
[20] Parkinson, G., “The tolerant tunnel: concept and performance,” Canadian
Aeronautics and Space Journal , Vol. 36, No. 3, 1990, pp. 130–134.
[21] Hameury, M., Development of the tolerant wind tunnel for bluff body testing ,
Ph.D. thesis, University of British Columbia, 1987.
33
[22] Williams, C., A new slotted-wall method for producing low boundary correc-
tions in two-dimensional airfoil testing , Ph.D. thesis, University of British
Columbia, 1975.
[23] Parkinson, G., “A tolerant wind tunnel for industrial aerodynamics,” Journal
of Wind Engineering and Industrial Aerodynamics , Vol. 16, No. 23, 1984,
pp. 293–300.
[24] Malek, A., “An investigation of the theoretical and experimental aerodynamic
characteristics of a low-correction wind tunnel wall configuration for airfoil
testing,” 1983.
[25] Selig, M., Deters, R., and Williamson, G., “Wind Tunnel Testing Airfoils at
Low Reynolds Numbers,” 49th AIAA Aerospace, Sciences Meeting, AIAA,
Vol. 875, 2011, pp. 4–7.
34