ON THE IMPACT OF THE PRE-IRRADIATION TLDs READOUT ON THEIR POSTIRRADIATION
Transcript of ON THE IMPACT OF THE PRE-IRRADIATION TLDs READOUT ON THEIR POSTIRRADIATION
International Journal of Physics
and Research (IJPR)
ISSN 2250-0030
Vol. 3, Issue 2, Jun 2013, 63-76
© TJPRC Pvt. Ltd.
ON THE IMPACT OF THE PRE-IRRADIATION TLDs READOUT ON THEIR POST-
IRRADIATION GLOW CURVES
ARAFA ABD EL-HAFEZ1, NABIL EL-FARAMAWY
2, JEEHAN FAHAD AL-SHAREEF
3 & SOAD EL-FIKI
4
1Ionizing Radiation Metrology Laboratory, National Institute for Standards (NIS), Giza, Egypt
2,4Department of Physics, Faculty of Science, Ain Shams University, Abbassia, Cairo, Egypt
3Kingdom of Saudi Arabia, Riyadh
ABSTRACT
In this work the impact of reading out of the un-irradiated LiF TLDs based on the post irradiation readout was
investigated. Four types of LiF detectors were used. The first three TLDs were MTS-N, MTS-6 and MTS-7 which differ
only in the percentage abundance of Li6 and Li
7 isotopes. The present results were discussed and compared with similar
LiF (TLD-100, TLD-600 and TLD-700) fabricated by another company. MCP-N LiF dosimeters were also studied and
compared with similar detectors of different only in the doping materials. The current results indicated that all of the LiF
doped with Mg and Ti detectors were affected dramatically by the pre-irradiation readout process regardless of the
fabrication companies. Otherwise, the impact of different doping materials was contradictious. These dramatic changes
could be attributed to the occurrence of the nonradiative recombination through the irradiation process which yield high
temperature peaks in the glow curve of the investigated LiF: Mg,Ti in contrary to LiF: Mg,Cu,P detectors.
KEYWORDS: Zero Dose Readout, Pre-Irradiation Heat Treatment, Thermoluminescence, LiF:Mg,Ti, LiF:Mg,Cu,P
INTRODUCTION
There are several techniques for radiation dosimetry, among all these techniques, thermoluminescence dosimetry
(TLD) is now the most widely used technique in many fields like personal and environmental radiation exposure (Gilvin
and Perks, 2010, Sandouqa et al., 2011), geochronology and space dosimetry. In addition, TLDs are extensively used in
both of diagnostic (Shousha et. al., 2011) and therapeutic medical applications (Abd El-Hafez et. al., 1997a and b, Czopyk
et al., 2007, Sharma et al., 2011), and in other health related aspects as well.
Thermoluminescence phenomena results in forming the characteristic glow curves for the phosphor material. The
glow curves obtained for each material are different, and each glow peak is ascribed to the recombination centers and is
related to traps (Furetta, 2011). One of the major problems that afflict TLD in general is the complexity of the glow curve
obtained with many TLD materials. Hence, a glow curve analytical tool kit (GCAFIT) was used for assessing individual
behavior of each peak in terms of both intensity and position (Abd El-Hafez et al., 2011a).
Improper use of a thermoluminescent material in a dosimetric application may lead to numerous errors in
determined doses (Moscovitch and Horowitz, 2006). To assess the radiation doses precisely, the sources of dose errors
should be identified and minimized (Robertson1981). Reproducibility and overall accuracy of dose estimations with TLD
was assessed for standard procedures used in a variety of areas, such as radiation therapy, diagnostic radiology, mailing
dosimetry service for radiotherapy (ASTM, 1998; Izewska et al., 2007), and space missions (Jarvis et al., 2011). A
background signal of a TL dosimeter is due to several factors, such as tribo-and chemi-luminescence, stimulation of the TL
phosphor by UV and visible light, IR emission of the heating element and its surroundings, dark current fluctuations of the
PM tube, and residual signals of the TL phosphor from its previous irradiations. All those effects can be reduced or
64 Arafa Abd El-Hafez, Nabil El-Faramawy, Jehan Fahad Al-Shareef & Soad El-Fiki
eliminated if appropriate procedures are followed (Furetta, 2010). Typically, standard practice procedures for TL dosimetry
involve pre-irradiation background (also known as ‗‗zero dose‘‘) readings of TLDs especially at low dose levels, to be
subtracted from readings of irradiated dosimeters (ASTM, 1998; Izewska et al., 2007). However, some experimentalists do
skip this procedure at different circumstances: (1)if dosimeters are being irradiated to high doses compared to which
background values could be neglected, (2)when high level of dosimetry with minimized uncertainties is not within the
scope,(3) when dealing with large number of dosimeters, it was thought to be enough to select randomly few dosimeters
for pre-irradiation background measurements instead of reading all dosimeters for saving time, and (4) when the glow
curves are a subject of an analytical study regardless of doses delivered (Ixquiac-Cabrera, et al., 2011, Harvey, and
Kearfott, K. 2011).
The zero dose reading is very important when the dosimeters are used to measure low doses. As the measured
doses increase, the background doses and their variations become less important and can be entirely neglected at high
delivered doses. Quantitatively, a zero dose reading is a combination of two main components: a reading without a
dosimeter, Lo (dark current) and a reading of the unexposed dosimeter, Lu (Furetta, 2010). The mean value of the zero dose
reading LBG is given by
LBG= Lo +Lu (1)
In our previous work (Abd El-Hafez and Maghraby 2011, Abd El-Hafez et al., 2011b), we investigated
qualitatively: what extent pre-irradiation background measurements affect the final readout of the dosimeters after
irradiation. We selected commonly used lithium fluoride (LiF) from the numerous available thermoluminescent materials
(Horowitz, 1984; Horowitz and Yossian, 1995; Ranogajec-Komor, 2003). LiF as a TL dosimeter is known for its high
sensitivity, stability, and approximate tissue equivalency (Jung H. et al., 2003). Since the discovery of its useful TLD
properties at the University of Wisconsin (Daniel et al., 1953), LiF was extensively studied (e.g., Buenfil and Brandan,
1997; Yang et al., 2006; Fung, 2004; Kim et al., 2008; Harvey et al., 2010) and is now arguably the most extensively used
reference thermoluminescent material in a wide spectrum of applications. Selecting specifically LiF dosimeters to evaluate
the effect of pre-irradiation background measurements on the post-irradiation dosimeter readout values has made the
results interesting to a wide audience.
The readout of the un-irradiated thermoluminescence detectors (TLDs) plays important role when measuring low
doses in the range of personal and diagnostic dosimetry. Recently, Abd El-Hafez and Maghraby (2011), as a defence of the
pre-irradiation readouts (zero dose), noticed that readouts of LiF:Mg,Ti Harshow fabrication named TLD-100, TLD-600
and TLD-700 dosimeters background before irradiation (regardless of values of background) can enhance TL output and
may cause changes to the glow curves both qualitatively in terms of peaks positions and peaks relative intensities, and
quantitatively in terms of areas under the curve and the maxima of the peaks.
It may be concluded that pre-irradiation background measurements for the studied TLDs is not only important for
evaluation of residual doses when considering high level of accuracy, but its most importance (as a form of heat treatment)
in the enhancement of their sensitivities and adjustment of glow curves peaks distribution and relative ratios. Impacts on
glow curves enhancement were dramatically in cases of sensitized TLD-700 and sensitized TLD-600, and were minor in
case of CaF2:Tm TLD-300 dosimeters (Abd El-Hafez et al., 2011b). It is highly recommended to perform background
measurements before irradiation regardless how high the radiation dose is, and even the background value will be
neglected either the thermoluminescent dosimeters were sensitized or un-sensitized and or the Li isotopes are (enriched 6Li,
7Li or natural). The condition of better enhancement is that the TL material should have the critical mechanism is that of
On The Impact of the Pre-Irradiation TLDs Readout on Their Post-Irradiation Glow Curves 65
competition during the stage of readout. In contrary to CaF:Tm (TLD-300) which clearly shows that the critical mechanism
is that of competition during the stage of absorption of radiation.
Hence, it is of some importance to investigate if this enhancement in glow curves is observable in another
fabricated LiF such as MTS-6, MTS-7 and MTS-N; and to what extent the doping impurities such as Mg, Cu, P (MCP-N)
play a role, if it has the same enhancement associated with the pre-irradiation readouts.
INSTRUMENTS AND METHODS
Radiation Source and Dose Determination
The Cesium-137 gamma source model is GB-150, was manufactured by Atomic Energy of Canada Limited on
April 1970 with original activity of 1000 Ci. (3.7 x 1010
kBq) and was used for irradiation. Air kerma (Kair) was determined
according to the International Atomic Energy Agency (IAEA) code of practice TRS-(381) (IAEA, 1997). Kair
determination was performed using the secondary standard dosimetry system of the National Institute of Standards (NIS) –
Egypt, which is composed of the NPL-2560 electrometer (UK) and the NE2561 ionization chamber (UK), the secondary
standard calibration system was calibrated at the Bureau International des Poids et Mesures (BIPM).
The expanded uncertainty associated to Kair determination was about 0.9 % at 95% level of confidence (coverage
factor = 2).
TLD Reader
The used TLD reader is Harshaw/Bicron 4500 TLD Reader which is equipped with two photomultiplier tubes that
can read independently; the Reader operates on WinREMS Software, which runs under Windows on a separate computer.
Thermoluminescent Dosimeters
Four different LiF-based TLDs have been used in this study: Three types of LiF:Mg,Ti phosphors with
commercial names MTS-7 (enriched with 7Li), MTS-6 (enriched with
6Li), and MTS-N (with natural Li) were used beside
another type of LiF:Mg,Cu,P phosphors named as MCP-N manufactured by Institute of Nuclear Physics in Karkow,
Poland (IFJ). The dimensions of all the dosimeters were 3.2 x 3.2 x .9 mm3. The annealing were performed using an
electric furnace model (NEY VulcanTM
3-550) manufactured in (U.S.A).
Experimental Method
LiF dosimeters (all families) were annealed, and kept in dark. Two cases were investigated, in the first case:
background was readout, while the second case was irradiated without reading of the background. Each set is composed of
not less than four dosimeters.
All the TLDs were grouped together and exposed to a well-defined radiation dose from Cs-137 gamma source.
Dosimeters readings were acquired using linear heating rate 3°C/s over the range (100 – 400 C°). This low heating rate was
used in order to avoid possible shifts, and hence extreme overlapping in glow peaks (Abd El-Hafez et al., 1999; Yazici,
2004).
Thermoluminescent crystals are inherently re-usable. The crystals require special thermal treatments (annealing)
prior to re-use. The annealing was carried out in a separate annealing oven.
The MTS-N, MTS-6 and MTS-7 were annealed at 400 °C for one hour followed by fast cooling to 100°C and
maintained for two hour. Each dosimeter passes two cases; the first case (Case A) represents the readout after gamma
irradiation for dosimeters experienced pre-irradiation background measurements, while the second case (Case B)
66 Arafa Abd El-Hafez, Nabil El-Faramawy, Jehan Fahad Al-Shareef & Soad El-Fiki
represents the readout of dosimeters irradiated without experiencing pre-irradiation background measurements. After
annealing the TL detectors, the TL reader was switched on and left for about half an hour to reach the stability of the
electronic system. Moreover, the TLD readout was always after 24 hours post irradiation; in order to get rid of the short
lived TL peaks that occur at low temperatures.
RESULTS
A sensitivity correction factor is determined for each used detector by exposing all the dosimeters to the same
dose and repeating that four times. The final reading for TLDs is obtained by correcting each dosimeter‘s reading with its
sensitivity correction factor.
LiF:Mg,Ti (MTS-6)
Figure (1a) illustrates the glow curves of MTS-6 in the temperature range 100–400 oC with readout heating rate 3
oC/s and irradiated with 0.2 Gy from
137Cs gamma source, the dashed curve represents the case (A) without pre-irradiation
readout and the solid one represents the case (B) with pre-irradiation readout.
From the figure, it is so easy to notice that the change in sensitivity is not uniform over the heating range; there
are several obviously occurred changes. A remarkable enhancement of the glow curve as a whole is observed as the first
effect of pre-irradiation readout for the integral value of the glow curve and the maximum of the dosimetric peak 5 (P5).
The entire low temperature glow curve (P2:P5) is observed to be increased as a second effect of the pre-irradiation
readout. In contrary, the high temperature peaks (P7:P8) tends to be constant with a little decrease is noticed. Peak 3 (P3)
which is not apparent in the case without pre-irradiation readout is very clear and higher resolution is induced for P3 and
P5.The same behavior is noticed for MTS-6 irradiated with the other examined dose 0.4 Gy as in Figure (2a).
Figures (1b) and (2b) illustrate the ratio between the two glow curves point by point with and without pre-
irradiation readouts for the two examined doses 0.2 and 0.4 Gy respectively.
This ratio is not constant for all points, it has a value of one or less in the high temperature region from about
260:400 oC and has values more than 2 up to 6 at least at low temperatures from 100 : 260
oC. Several peaks in the ratio
curve with maxima correspond to the real positions of glow peaks which are not the apparent ones and represented by the
dashed vertical lines.
The reason of that behavior is that the temperature of the real peak stands few degrees of temperature less than the
apparent peak because of the contribution of the adjacent higher temperature one.
The lower values between the peaks of the ratio curve are represented by the dotted vertical lines. Differentiating
between the thermoluminescence intensity values of the pre-irradiation glow curve with respect to time of heating or
temperature, since the heating rate is constant and representing it as upper curve (c) in figures (1) and (2). From curves (b)
and (c), it can be predicted the existence of peaks 4 (not apparent in this case) and peak 6 (may appears only when analysis
the glow curve with computerized glow curve deconvolution).
The small peak at about 190 oC on the ratio curve (b) and on the differential curve (c) represents the position of
peak 4 of the thermoluminescent glow curve.
A small shift in the peak positions (the temperature at which the maxima of the peaks is occurred) is observed as a
result of pre-irradiation readout towards high temperatures for most peaks except for P3 shifted towards lower
temperatures.
On The Impact of the Pre-Irradiation TLDs Readout on Their Post-Irradiation Glow Curves 67
Figure 1: (a) Illustrates the Glow Curves of MTS-6 in the Temperature Range 100–400 oC with Readout Heating
Rate 3 oC/s and Irradiated with 0.2 Gy from
137Cs Gamma Source, the Dashed Curve Represents the Case
(A) without Pre-Irradiation Readout and the Solid one Represents the Case (B) with Pre-Irradiation Readout.
(b) Illustrate the Ratio between the Two Glow Curves Point by Point with and without Pre-Irradiation Readouts.
(c) the Differential of the Thermoluminescence Intensity Values of the Pre-Irradiation Readout Glow Curve with
Respect to time of Heating or Temperature Since the Heating Rate is Constant
Figure 2: (a) Illustrates the Glow Curves of MTS-6 in the Temperature Range 100–400 O
c with Readout Heating
Rate 3 O
c/S and Irradiated with 0.4 Gy from 137
Cs Gamma Source, the Dashed Curve Represents the Case
(A) without Pre-Irradiation Readout and the Solid One Represents the Case (B) with Pre-Irradiation Readout.
(b) Illustrate the Ratio between the Two Glow Curves Point by Point with and Without Pre-Irradiation Readouts.
(c) the Differential of the Thermoluminescence Intensity Values of the Pre-Irradiation Readout Glow Curve with
Respect to Time of Heating or Temperature Since the Heating Rate is Constant
LiF:Mg,Ti (MTS-7)
Figures (3a) and (4a) which represent the intensity of the glow curve of MTS-7 irradiated with the same dose 0.2
Gy and 0.4 Gy respectively without pre-irradiation (dashed curve) and with pre-irradiation (solid curve) readout. The same
behavior is recorded for MTS-7 as that for MTS-6 represented in figures (1a) and (2a). The ratio between the two
investigated case (with and without pre-irradiation readouts) for all points of the two glow peaks are plotted versus
68 Arafa Abd El-Hafez, Nabil El-Faramawy, Jehan Fahad Al-Shareef & Soad El-Fiki
temperature as shown in figures (3b:4b). A small shift in the peak positions is observed as a result of pre-irradiation
readout towards high temperatures for most peaks except for P3 shifted towards lower temperatures.
LiF:Mg,Ti (MTS-N)
The same behavior is recorded for MTS-N as shown in Fig. (5a) which represent the intensity of the glow curve of
MTS-N irradiated with the same dose 0.2 Gy without pre-irradiation (dashed curve) and with pre-irradiation (solid curve)
readout. The same behavior is noticed for MTS-N irradiated with the other examined dose 0.4 Gy as in Fig. (6a). Peaks 3
and 4 (P3 and P4) which are not apparent in the case without pre-irradiation readout are obvious and higher resolution is
induced for P3 and P5. Peak 4 appear as a hump in the case A (with pre-irradiation readout). The same behavior is noticed
for MTS-N irradiated with the other examined dose 0.4 Gy as in Fig. (6a). Noticeable decrease in peaks P7 and P8 of MTS-
N higher than that of MTS-6 and MTS-7 with corresponding smaller increase of peaks P2:P5 (low temperature peaks) than
that of MTS-6 and MTS-7. The ratio between the two investigated case (with and without pre-irradiation readouts) for all
points of the two glow peaks are plotted versus temperature as shown in figures (5b and 6b). The same small shift in the
peak positions is observed as a result of pre-irradiation readout towards high temperatures for most peaks except for P3
shifted towards lower temperatures.
LiF:Mg,Ti (MCP-N)
Representing the TL intensity glow curves of LiF:M,Cu,P (MCP-N) on the linear scale, it yields one peak only in
the two cases A and B (with and without pre-irradiation readouts). This single peak did not interpret the ratio curve (b) nor
the differential curve (c) as the discussed cases of LiF:Mg,Ti (MTS-6, MTS-7 and MTS-N). So, representing it on the log
scale as in Figures 7a and 8a yields two peaks at low temperature region which are responsible for observed high ratio in
that region on the ratio curve (b) beside the main peak at about 235oC. In contrary to the LiF:Mg,Ti dosimeters, little
effects of the pre-irradiation readout are noticed in the integral value of the glow curve (area under the glow curve) and in
the height of the main dosimetric peak.
Figure 3: (a) Illustrates the Glow Curves of MTS-7 in the Temperature Range 100–400 O
c with Readout Heating
Rate 3 O
c/S and Irradiated with 0.2 Gy from 137
Cs Gamma Source, the Dashed Curve Represents the Case
(A) without Pre-Irradiation Readout and the Solid One Represents the Case (B) with Pre-Irradiation Readout.
(b) Illustrate the Ratio between the Two Glow Curves Point by Point with and without Pre-Irradiation Readouts.
(c) the Differential of the Thermoluminescence Intensity Values of the Pre-Irradiation Readout Glow Curve with
Respect to Time of Heating or Temperature Since the Heating Rate is Constant
On The Impact of the Pre-Irradiation TLDs Readout on Their Post-Irradiation Glow Curves 69
Figure 4: (a) Illustrates the Glow Curves of MTS-7 in the Temperature Range 100–400 O
c with Readout Heating
Rate 3 O
c/S and Irradiated with 0.4 Gy from 137
Cs Gamma Source, the Dashed Curve Represents the Case
(A) without Pre-Irradiation Readout and the Solid One Represents the Case (B) with Pre-Irradiation Readout.
(b) Illustrate the Ratio between the Two Glow Curves Point by Point with and without Pre-Irradiation Readouts.
(c) The Differential of the Thermoluminescence Intensity Values of the Pre-Irradiation Readout Glow Curve with
Respect to Time of Heating or Temperature Since the Heating Rate is Constant
Figure 5: (a) Illustrates the Glow Curves of MTS-N in the Temperature Range 100–400 oC with Readout Heating
Rate 3 oC/s and Irradiated with 0.2 Gy from
137Cs Gamma Source, the Dashed Curve Represents the Case
(A) without Pre-Irradiation Readout and the Solid One Represents the Case (B) with Pre-Irradiation Readout.
(b) Illustrate the Ratio between the Two Glow Curves Point by Point with and without Pre-Irradiation Readouts.
(c) The Differential of the Thermoluminescence Intensity Values of the Pre-Irradiation Readout Glow Curve with
Respect to Time of Heating or Temperature Since the Heating Rate is Constant
70 Arafa Abd El-Hafez, Nabil El-Faramawy, Jehan Fahad Al-Shareef & Soad El-Fiki
Figure 6: (a) Illustrates the Glow Curves of MTS-N in the Temperature Range 100–400 oC with Readout Heating
Rate 3 oC/s and Irradiated with 0.4 Gy from
137Cs Gamma Source, the Dashed Curve Represents the Case
(A) without Pre-Irradiation Readout and the Solid One Represents the Case (B) with Pre-Irradiation Readout.
(b) Illustrate the Ratio between the Two Glow Curves Point by Point with and without Pre-Irradiation Readouts.
(c) The Differential of the Thermoluminescence Intensity Values of the Pre-Irradiation Readout Glow Curve with
Respect To Time of Heating or Temperature Since the Heating Rate is Constant
DISCUSSIONS
For quantitative investigation of the impact of pre-irradiation readout on the post-irradiation readout, the values of
the ratio (R) which is defined as the intensity of the thermoluminescence signals with pre-irradiation readout to that
without pre-irradiation treatment. Table (1) represents this impact as the ratio between the total areas in the range 100-
400oC in both cases which increase with R between 1.83 and 2.69 for all LiF:Mg,Ti dosimeters and higher values of R for
the low temperature area ranged from 2.5 up to 4.13 with a lowest value for MTS-N.
The impact on the high temperature area is noticed as a small decrease with average value equals 0.785 for MTS-
6 and MTS-7. This decrease is enlarged for MTS-N to a value of 0.60. There is a little impact of 1.12 for MCP-N which
has different impurities than MTS dosimeters. As TLD models can be classified as models in which the critical processes
occur during the absorption of radiation, and models in which the critical processes occur during heating (i.e. during TL
readout).
The evidence, at least for LiF:Mg,Ti materials, whether the Lithium is 6Li or
7Li or natural Li, clearly shows that
the critical mechanism is that of competition during the stage of TL readout. For LiF:Mg,Cu,P (MCP-N), during heating,
electrons released from traps, may either recombine with trapped holes to produce TL or be retrapped in deeper traps,
which act as competing centers. The competition–during heating process was first suggested by Rodine and Land (1971).
This explains the different behavior between LiF:Mg,Ti and LiF:Mg,Cu,P.
On The Impact of the Pre-Irradiation TLDs Readout on Their Post-Irradiation Glow Curves 71
Figure 7: (a) Illustrates the Glow Curves of MCP-N in the Temperature Range 100–400 oC with Readout Heating
Rate 3 oC/s and Irradiated with 0.2 Gy from
137Cs Gamma Source, the Dashed Curve Represents the Case
(A) without Pre-Irradiation Readout and the Solid One Represents the Case (B) with Pre-Irradiation Readout.
(b) Illustrate the Ratio between the Two Glow Curves Point by Point with and without Pre-Irradiation Readouts.
(c) The Differential of the Thermolumine-Scence Intensity Values of the Pre-Irradiation Readout Glow Curve with
Respect To Time of Heating or Temperature Since the Heating Rate is Constant
Figure 8: (a) Illustrates the Glow Curves of MCP-N in the Temperature Range 100–400 oC with Readout Heating
Rate 3 oC/s and Irradiated with 0.4 Gy from
137Cs Gamma Source, the Dashed Curve Represents the Case (A)
without Pre-Irradiation Readout and the Solid One Represents the Case (B) with Pre-Irradiation Readout. (b)
Illustrate the Ratio between the Two Glow Curves Point by Point with and without Pre-Irradiation Readouts. (c)
The Differential of the Thermoluminescence Intensity Values of the Pre-Irradiation Readout Glow Curve with
Respect to Time of Heating or Temperature Since the Heating Rate is Constant
Table (1), also represents the ratio for the maxima of the apparent peaks for all the investigated dosimeters. The
lowest increase for the impact is observed for P2, the highest for P3 followed by P5 and the decrease for LiF:Mg,Ti is
observed for P7 and P8. It is well known that, especially, pre-irradiation heat treatments between 100 and 200 ºC cause a
72 Arafa Abd El-Hafez, Nabil El-Faramawy, Jehan Fahad Al-Shareef & Soad El-Fiki
series dramatic on the TL properties of TLD-100 as well as on the trapping parameters of the individual TL glow peaks
(Bos et al., 1992: Piter et al, 1996: Yazici, 1998).
Kitis and Otto (2000) for TLD-700 glow curves, observed a glow peak between P2 and P3 with a Tm very close to
the Tm of P2, and they observed a glow peak between P2 and P3 but with a Tm very close to the Tm of P3. It is first observed
by Johnson (Johnson, 1974) and labeled as P3a. A few other authors have reported this observation (Fairchild et al., 1978:
Perks and Marshall, 1991). It is not clear however, what experimentally conditions are necessary for the creation of this
peak (Horowitz and Yossian, 1995). It was reported (Yuan and McKeever, 1988 and Piters et al., 1996) that P2 and P3 have
been associated with with magnesium cation vacancy dipoles MgV and P4 and P5 with trimmers or higher order cluster of
(MgV)3. The basic processes in which MgV dipoles and clusters of dipoles (dimers, trimmers and so on) are involved can
be schematically presented by:
MgV (MgV)2 (MgV)3 precipitates (2)
Where, ‗precipitates‘ stand for higher order clusters. A high temperature has the effect of pushing the defect
equilibrium to the left and at a low temperature the defect equilibrium is forced to the right of equation (2).
Table 1: The Ratio between the Thermoluminescence Glow Curves (Peaks) with Pre-Irradiation
Readout to without Pre-Irradiation Readout for MTS-6, MTS-7, MTS-N and MCP-N
Irradiated with 0.2 and 0.4 Gy from 137
Cs γ-Source
Ratio (Case A/Case B)
MTS-6 MTS-7 MTS-N MCP-N MTS-6 MTS-7 MTS-N MCP-N
0.2 Gy 0.4 Gy
Total Area (100:400) 2.38 2.43 1.83 1.12 2.69 2.77 1.83 1.12
Area P2:P5 (100-250) 3.46 3.42 2.5 3.69 4.13 2.55
Area P6:P8 (250-400) 0.76 0.77 0.6 0.81 0.79 0.61
Max. of P2 3.1 2.7 2.3 3.60 3.4 3.2 2.3 3.05
Max. of P3 5.89 5.1 4.8 5.8 5.3 4.06
Max. of P4 2.8 3.02
Max. of P5 3.6 3.6 2.1 1.11 3.6 3.2 2.18 1.11
Max. of P7 0.78 .79 0.8 0.78 0. 9 0.59
Max. of P8 0.85 0.8 0.5 1.0 1.0 0.51
It was observed that the high temperature side in the case without applying the pre-irradiation read out process is
about 1.5 or 2 times greater than that in the case of applying it. The contribution of the low temperature side in the case of
applying the pre-irradiation read out process is greater than that in the case without applying it. Of course, one can ask;
when shall we apply the pre-irradiation read out process? In fact, this is depending on the natural type and the aim of the
experiment or the study. Many studies (Buenfil and Brandan, 1997: Berger and Hajek, 2008: Pardhan et al., 2009: Datz et
al., 2010) are based on using or studying the high temperature glow peaks.
Therefore, in such studies, it is not recommended to apply the pre-irradiation readout process especially with this
described heat treatment. We observe that the percentage contribution of P5 has increased by about three times in the case
of applying the pre-irradiation read out process. Therefore, in the studies which are based on using or studying P5 [e.g.
Chen and Hag-Yahya, 1996; Weizman et al, 2000; Weiss et al., 2008; Yazici, 2004) it is recommended to apply the pre-
irradiation read out process.
If we suggest that P4 is due to electron traps and P5 is due to hole traps, we can use Lawless et al. model (Lawless
et al., 2009) to explain the complexity structure of the glow P4 and P5. Their mode has one electron trap, one hole trap, and
one recombination center from which radiation is emitted when an electron recombines at the center. Under some
On The Impact of the Pre-Irradiation TLDs Readout on Their Post-Irradiation Glow Curves 73
circumstances, this model shows two peaks. The first peak is due to thermal stimulation of electrons from the electron trap.
These recombine with holes in the center until those holes are depleted.
At this point, the TL emission due to electron recombination is stopped, and due to the lack of holes in center,
electrons emitted from the electron trap are re-trapped. This continues until free holes start to be created from thermal
stimulation of the hole trap. This replenishes holes in the center and allows electron recombination to continue. This results
in a second TL peak, in which the second peak has the initial rise and peak shape associated with the hole trap, not the
electron trap. If the time between the electron re-trapping in the hole centers and their recombination at the luminescence
center is too short, this will result in a complex overlapping of P4 with P5.
The shift in the peak quality, i.e., peak position towards higher temperature as an impact of pre-irradiation readout
was held responsible for the increase in the contribution of non-radiative transitions due to the presence of the competing
traps as previously concluded in our previous work (Abd El-Hafez et al., 2011b). It was inferred that the glow peaks
occurring at higher temperatures must exhibits higher thermal quenching. This is attributed to the energy dissipated in an
indirect transition, however, which is much less than the band-gap energy and may thus be dissipated either radiatively (via
photons) and non-radiatively (via phonons). Non-radiative capture of free carriers takes place because the lattice vibrations
cause the energy level to change its position in the forbidden gap.
For large enough vibrations, the energy level crosses into the conduction band and captures a free electron. The
lattice relaxation which follows the capture lowers the position of the level back into the energy gap, the excess energy
being propagated as lattice phonons (McKeever, 1985). However, the glow curve is controlled by the release of the charge
carriers from traps and not by the properties of the luminescence centers (Horowitz, 1984).
CONCLUSIONS
The pre-irradiation read out process causes dramatic effects on the behavior and shape, of the glow peaks in the
glow curves of the LiF:Mg,Ti (MTS-6, MTS-7 and MTS-N ) as well as (TLD-100, TLD-600, and TLD-700) dosimeters
manufactured by different companies with the same impurities.
The pre-irradiation readout process causes little impact only on the height of the glow peaks or area for the
LiF:Mg,Cu,P (MCP-N) as well as CaF2:Tm (TLD-300) which have no high temperature peaks and the critical mechanism
is that of competition-during the stage of absorption of radiation.Pre-irradiation readout causes an increasing of the main
dosimetric peak P5 in the glow curve by about three times double, but at the same time it causes a significant decreasing in
the area contribution of the high temperature glow peaks (P7, and P8) by about 0.6:0.8. Thus, according to the type and the
aim of the experiment, one can decide whatever to apply the pre-irradiation process or not.
In the studies which are based on using or studying the high temperature glow peaks, it is not recommended to
apply the pre-irradiation readout process on the dosimeters. While, in the studies which are based on using or studying the
main dosimetric glow P5, it is recommended to apply the pre-irradiation readout process on the dosimeters. Further studies
should be performed soon regarding behavior of each resolved peak as a function of radiation quality and dose level
effects.
REFERENCES
1. Abd El-Hafez, A.I., Eissa, H.M., Eman, S.A., & Fadel, M.A. (1997a). Measurement of Dose Enhancement by
Neutron Capture of Boron in Cf-252 Interstitial Brachytherapy. Radiat. Meas, 28, 441-444
74 Arafa Abd El-Hafez, Nabil El-Faramawy, Jehan Fahad Al-Shareef & Soad El-Fiki
2. Abd El-Hafez A.I., Eissa, H.M., Eman, S.A., & Fadel, M.A. (1997b). Intercomparative Studies of Brachytherapy
Techniques Combined with Cf-252 Boron Neutron Capture Therapy. Radiat. Meas., 28, 435-440
3. Abd El-Hafez, A.I., (1999). The readout-heating rate effects on the performance of LiF (TLD-600 and TLD-700)
thermoluminescence detectors and determination of the trap parameters. Egyptian Journal of Solids, 22, 149-162
4. Abd El-Hafez, A.I., Yasin, M.N., & Sadek, A.M. (2011a). GCAFIT- New Tool for Glow Curve Analysis in
Thermoluminescence nanodosimetry. Nucl. Inst. Meth. (A), 637, 158-163
5. Abd El-Hafez, A.I., Maghraby, A., & El-Faramawy, N.A. (2011b). In Defence of Thermoluminescence Dosimeter
Zero Dose Readouts. Journal of American Science, 7, 796-803
6. Abd El-Hafez, A.I. & Maghraby, A. (2011). Impacts of Pre-Irradiation Background Reading On
Thermoluminescence Dosimeter Glow Curve. Appl. Radiat. & Isot. 69, 1533–1539
7. ASTM. (1998). Standard practice for use of thermoluminescence-dosimetry (TLD) systems for radiation
processing. pp.1956-1998
8. Berger, T. & Hajek, M., (2008). On the Linearity of the High-Temperature Emission from 7LiF:Mg,Ti (TLD-
700)‖. Radiat. Meas., 43, 1467-1473
9. Bos, A.J.J., Vijverberg, R.N.M., Piters, T.M. & MecKeever, S.W.S., (1992). Effect of Cooling and Heating Rate
on Trapping Parameters in LiF:Mg,Ti Crystals. J. Phys., 25, 1249-1257
10. Buenfil, A.E., & Brandan, M.E., (1997). Influence of Repeated Use and Annealing Atmosphere on LiF TLD-100
Response to α-Particles‖, Appl. Radiat. Isot., 48, 1193-1196
11. Chen, R., & Hag-Yahya, A., (1997). A New Possible Interpretation of the Anomalous Fading in
Thermoluminscent Materials as Normal Fading in Disguise. Radiat. Meas., 27, 205-210
12. Czopyk, L.K.M.. Olko, P., Swakon, J., Waligorski, M.P.R., Kajdrowicz, T., Cuttone, G., Cirrone, G.A.P., & Di
Rosa, F., (2007). Two-dimensional dosimetry of radiotherapeutical proton beams using thermoluminescence foils.
Radiat. Prot. Dosim., 126, 185–189
13. Daniel, F., Boyd, C.A., & Sanders, D.F., (1953). Thermoluminescence as a research tool. Science 117, 343–349
14. Datz, H., Horowitz., Y.S., Oster, L., & Margaliot, M., (2010). Characteristics of the High Temperature
Thermoluminescence in LiF:Mg,Ti (TLD-100): the Effect of Batch History. Radiat. Meas., 45, 710-712
15. Fairchild, R.G., Mattern, P.L., Lengweilerk, K., & Levy, P.W., (1978). Thermoluminescence of LiF (TLD-100):
Glow Curve Kinetics. ,J. Appl. Phys., 49, 4523-4533
16. Fung, K.K.L., (2004). Investigation of dosimetric characteristics of the high sensitivity LiF:Mg,Cu,P
thermoluminescent dosimeter and its applications in diagnostic radiology: A review. Radiography 10, 145–150
17. Furetta, C., Guzmán, S., Ruiz, B., & Cruz-Zaragoza, E., (2011). The initial rise method extended to multiple
trapping levels in thermoluminescent materials. Appl. Radiat. Isot. 69, 346–349
18. Furetta, C., (2010). Handbook of Thermoluminescence second ed
. World Scientific, Singapore.
19. Gilvin, P.J., & Perks C.A., (2011). Practical low dose limits for passive personal dosimeters and the implications
for uncertainties close to the limit of detection. Radiat. Prot. Dosim., 144 102–106
On The Impact of the Pre-Irradiation TLDs Readout on Their Post-Irradiation Glow Curves 75
20. Jung, H., Lee, K.J., Kim, J.L., (2003). A personal thermoluminescence dosimeter using LiF:Mg,Cu,Na,Si
detectors for photon fields. Appl. Radiat. Isot. 59, 87–93
21. Harvey, J., & Kearfott, K., (2011). Reproducibility of glow peak fading characteristics of thermoluminescent
dosimeters. Radiat. Meas., 46, 319-322
22. Harvey, J., Haverland, N., & Kearfott, K., (2010). Characterization of the glow-peak fading properties of six
common thermoluminescent materials. Appl. Radiat. Isot. 68, 1988–2000
23. Horowitz, Y.S., Ben Shachar, B., & Yossian, D., (1993). Study of the long-term stability of peaks 4 and 5 in TLD-
100: correlation with isothermal decay measurements at elevated temperatures. J. Phys., D: Appl. Phys., 26, 1475
1481
24. Horowitz, Y.S., (1984). Thermoluminescence and Thermoluminescent Dosimetry. CRC press, Inc, Boca Raton,
Florida,
25. Horowitz, Y.S., & Yossian, D., (1995). Computerized glow curve deconvolution applications to thermolumine-
scence dosimetry. Radiat. Prot. Dosim. 60, 3-3
26. IAEA, (1997). The Use of Plane Parallel Chambers in High-Energy Electron and Photon Beams: An International
Code of Practice. Technical Report Series No. 381
27. Ixquiac-Cabrera, J.M., Brandan, A., Martínez- Dávalos, M.E., Rodríguez-Villafuerte, M., Ruiz-Trejo, C., &
Gamboa-deBuen, I. (2011). Effect of spectral shape in the relative efficiency of LiF:Mg,Ti exposed to 20 keV
effective energy X-rays. Radiat. Meas., 46, 389-395
28. Izewska, J., Dietmar, G., Bera, P., Thwaites, D., Arib, M., Saravi, M., Sergieva, K., Li, K., Garcia, F., Kumar, A.,
& Bulski, M., (2007). A methodology for TLD postal dosimetry audit of high-energy radiotherapy photon beams
in non-reference conditions. Radioth. Oncol. 84, 67–74
29. Jarvis, A., Caffrey, J.A., & Hamby, D.M., (2011). A review of instruments and methods for dosimetry in space.
Adv. Space Res. 47, 563–574
30. Johnson, T.L., (1974). Quantitative Analysis of the Growth and Decay of the TL Peaks in LiF (TLD-700)‖, In:
Proc. 4th
Int. Conf. on Luminescence Dosimetry, Krakow Inst., 197-217
31. Jung, H., Jai L., Kun, K., Jang-Lyul, L., & Sang, Y., (2004). Development of a personal dosimeter badge system
using sintered LiF:Mg, Cu, Na, Si TL detectors for photon fields. Radiat. Meas. 38, 71–80
32. Kim, J.L., Lee, J.I., Pradhan, A.S., Kim, B.H., & Kim, J.S., (2008). Further studies on the dosimetric
characteristics of LiF:Mg,Cu,Si—a high sensitivity thermoluminescence dosimeter (TLD). Radiat. Meas. 43, 446–
449
33. Kitis, G., & Otto, T., (2000). Peculiarities of the Glow-Peak 5a of LiF:Mg,Ti , Nucl. Instr. Meth. in Phys. Res. B
160, 262-273
34. Lawless J.L., Chen R., & Pagonis, V., (2009). On the theoretical basis for the duplicitous thermoluminescence
peak. J. Phys. D: Appl. Phys. 42, 1-8
35. McKeever, S.W.S., (1985). Thermoluminescence of Solids, Cambridge University Press
76 Arafa Abd El-Hafez, Nabil El-Faramawy, Jehan Fahad Al-Shareef & Soad El-Fiki
36. Moscovitch, M., & Horowitz, Y., (2006). Thermoluminescent materials for medical applications: LIF:Mg,Ti and
LiF:Mg,Cu,P. Radiat. Meas. 41, S71–S77
37. Perks, C.A. & Marshall, M., (1991). Technique for Thermoluminescence Glow Curve Analysis. Radiat. Prot.
Dosim., 38, 261-269
38. Piters, T.M., Bos, A.J.J., & Van der Burg, B., (1996). Effects of Annealing on Glow Peak Parameters of
LiF:Mg,Ti (TLD-100) Dosimetry Material. Radiat. Prot. Dosim. 65, 203-206
39. Pradhan, A.S., Lee, J.I., & Kim, J.L., (2009). Further Studies on Higher Temperature TL Glow Peaks of
7LiF:Mg,Ti. Appl. Radiat. Isot., 67, 1078-1083
40. Ranogajec-Komor, M., (2003). Thermoluminescence dosimetry-application in environmental monitoring. Radiat.
Saf. Manage. 2, 2–16
41. Robertson, M.E.A., (1981). Identification and Reduction of Errors in Thermoluminescence Dosimetry Systems
Ph.D. Thesis, Department of Physics, University of Surrey, Pitman Ltd.
42. Rodine, E.T., & Land, P.L., (1971). Electronic Defect Structure of Single- Crystal ThO2 by thermoluminescence.
Phs. Rev. B4, 2701-2724
43. Sandouqa, A.S., Haddadin, I.M., & Abu-Khaled, Y.S., (2011). Hand equivalent doses of nuclear medicine staff in
Jordan: Preliminary experimental studies. Radiat. Meas. 46, 250-253
44. Sharma, P.K., Jamema, S.V., Kaushik, K., Budrukkar, A., Jalali, R., Deshpande, D.D., Tambe, C.M., Sarin, R., &
Munshi, A., (2011). Electron Arc Therapy for Bilateral Chest Wall Irradiation: Treatment Planning and
Dosimetric Study. Clinical Oncology 23, 216-222
45. Shousha, H.A., Abd El-Hafez, A.I., & Fawzia, A., (2011). Dosimetric study of the effective doses resulting during
dental X-ray and panoramic radiotherapy. Radiat. Eff. & Def. in Solids. 1, 67-73
46. Weiss, D., Horowitz, Y.S. and Oster, L.,(2008) ―Delocalized Recombination Kinetics Modeling of LiF:Mg,Ti
(TLD-100) Glow Peak 5 TL System‖, Radiat. Meas., 43, 254-258
47. Weizman, Y., Horowitz, Y.S., & Oster, L., (2000). Optically Induced Conversion of Glow Peaks 4 and 5 in
LiF:Mg,Ti due to Localized and de-localized Recombination Mechanisms‖, Journal of Luminescence, 87-89, 552-
554
48. Yang, B., Gao, H., & Townsend, P.D., (2006). Low temperature thermoluminescence of annealed LiF:Mg, Cu, P.
Nucl. Instrum. Methods Phys. Res. B 247, 324–330
49. Yazici, A.N., (1998). The Effect of Pre-Irradiation Heat Treatment on the Evaluated Trapping Parameters of
LiF:Mg,Ti (TLD-100). Radiat. Prot. Dosim., 80, 379-348
50. Yazici. A. N. 2004, The influence of heating rate on the TL response of the main glow peaks 5 and 4+ 5 of
sensitized TLD-100 treated by two different annealing protocols. Nucl. Instr. Meth. in Phys. Res. B 215, 174–
180.
51. Yuan, X.L., & McKeever, S.W.S., (1988). Impurity clustering effects in magnesium-doped lithium fluoride‖,
Phys. Stat. Sol. (a), 108, 545-551