Multichannel Bipotentiostat Integrated With a Microfluidic Platform for Electrochemical Real-Time...
-
Upload
independent -
Category
Documents
-
view
2 -
download
0
Transcript of Multichannel Bipotentiostat Integrated With a Microfluidic Platform for Electrochemical Real-Time...
>TBCAS-2011-Sep-0136-Reg.R1<
1
Abstract— An electrochemical detection system specifically
designed for multi-parameter real-time monitoring of stem cell
culturing/differentiation in a microfluidic system is presented. It
is composed of a very compact 24-channel electronic board,
compatible with arrays of microelectrodes and coupled to a
microfluidic cell culture system. A versatile data acquisition
software enables performing amperometry, cyclic voltammetry
and impedance spectroscopy in each of the 12 independent
chambers over a 100kHz bandwidth with current resolution down
to 5pA for 100ms measuring time. The design of the platform, its
realization and experimental characterization are reported, with
emphasis on the analysis of impact of input capacitance (i.e.
microelectrode size) and microfluidic pump operation on current
noise. Programmable sequences of successive injections of
analytes (ferricyanide and dopamine) and rinsing buffer solution
as well as the impedimetric continuous tracking for 7 days of the
proliferation of a colony of PC12 cells are successfully
demonstrated.
Index Terms - electrochemical measurements, multichannel
potentiostat, microfluidics, stem cell monitoring, impedance
spectroscopy.
I. INTRODUCTION
ellular dynamics and the complex pathways that regulate
the differentiation of stem cells in vitro are currently of
primary interest in biology and medicine [1]. In
particular, the use of pluripotent stem cells shows great
promise for the treatment of neurodegenerative disorders [2].
A feasible approach to investigate stem cell differentiation
towards adult neural cells is the combination of several
complementary detection techniques. Due to the specificity of
fluorescent markers, optical microscopy is by far the preferred
tool in biological investigation. However, when studying
electrogenic cells, the microscope is naturally replaced by
Manuscript received September, 2011. This work was supported by the
EU FP7 EXCELL project under grant NMP4-SL-2008-214706.
M. Vergani, M. Carminati, G. Ferrari and M. Sampietro are with
Dipartimento di Elettronica e Informazione, Politecnico di Milano, Milano,
Italy (phone: +39.02.2399.3773, fax: +39.02.2399.3574, email:
[email protected], [email protected]).
C. Caviglia, A. Heiskanen, K. Zor, D. Sabourin, M. Dufva, M. Dimaki and
J. Emnéus are with the Department of Micro- and Nanotechnology, Technical
University of Denmark, Lyngby, Denmark.
E. Landini and R. Raiteri are with the Department of Biophysical and
Electronic Engineering, Università di Genova, Genova, Italy.
C. Comminges and U. Wollenberger are with the Institute for
Biochemistry and Biology, University Potsdam, Golm, Germany.
instrumentation with a direct electrical interface with cells,
such as patch-clamp [3]-[4] to record the activity of ion
channels (current detection) and multi-electrode arrays to map
the propagation of extracellular potentials (voltage detection)
in cultured neural networks [5]. Even in the case of non-
electrogenic cells, electrochemistry enables the detection of
molecules produced by cellular metabolism with suitable
(micro- and nano-molar) sensitivity [6]. The powerful
combination of electrochemical detection techniques with the
availability of planar microelectrodes, enabled by
microelectronic fabrication technologies, provides sub-cellular
spatial resolution [7]-[8].
The aim of this work is the realization of a versatile
platform that allows automatic monitoring of several cell
culture chambers and capturing detailed information of the
population status by combining optical and multiple electrical
detection techniques. For this purpose, a compact multichannel
bipotentiostat has been specifically designed to be integrated
in a microfluidic culture system having an array of planar
microelectrodes that allow for parallel screening of several cell
populations. This integrated solution is intended to bridge the
gap between microfluidic bioreactors with electrodes but
operated with bulky bench-top electrochemical instruments [9]
and sophisticated ad-hoc systems oriented to chemo-physical
monitoring of stem cell cultures [10] that are not designed to
encompass all electrochemical techniques such as
amperometry, voltammety and impedance spectroscopy.
The instrument has been designed to perform: (i) automatic
tracking of impedance (both at a single frequency and in the
full frequency range from 10mHz to 100kHz) in order to
monitor over time the adhesion and proliferation of a cell
population that affects the electrode-solution interface
impedance [11], and (ii) selective electrochemical detection of
electro-active species, in particular catecholamines (a family of
neurotransmitters, such as dopamine, released by cells), which
is carried out using cyclic voltammetry or amperometry [12].
The release of these molecules, contained in vesicles that are
fused with the cellular membrane, is called exocytosis. An
exocytotic event can be electrically detected using planar gold
microelectrode as reported in [13]. The continuous, spatially
resolved and label-free monitoring of exocytosis is believed to
represent an early indicator of the differentiation into a neural
lineage. Furthermore, preliminary results [14]-[15], indicate
that also electrochemical impedance spectroscopy (EIS)
Multichannel Bipotentiostat Integrated with a
Microfluidic Platform for Electrochemical Real-
Time Monitoring of Cell Cultures Marco Vergani, Marco Carminati, Member, IEEE, Giorgio Ferrari, Member, IEEE, Ettore Landini,
Claudia Caviglia, Arto Heiskanen, Clément Comminges, Kinga Zor, David Sabourin, Martin Dufva,
Maria Dimaki, Roberto Raiteri, Ulla Wollenberger, Jenny Emnéus and Marco Sampietro
C
>TBCAS-2011-Sep-0136-Reg.R1<
2
represents a promising label-free tool to detect morphological
changes during the stem cell differentiation process. Specific
requirements for stem cell monitoring include the capability to
apply a sinusoid of tens of µV (for EIS) and record the current
with a resolution of a few pA (for amperometry).
II. SYSTEM OVERVIEW
A. Architecture
As illustrated in Fig. 1, the complete experimental setup is
composed of a compact microfluidics/potentiostat assembly
that can be placed under a standard upright optical
microscope. The electronic board is tightly assembled with the
microfluidic system, while the peristaltic pumps [16] are on
the side. Stringent volume constraints have consequently
guided the board layout, in particular in proximity to the
microscope objective. Fig. 2 shows the vertical cross-section
of the complete platform. The potentiostat is fabricated on a
0.8mm thick multilayer PCB (Fig. 3) assembled in contact
with the microfluidic platform used to culture the cells on a
microelectrode array (MEA). The sealing between the rigid
PMMA substrate of the microfluidic system and the MEA chip
is provided by a biocompatible silicon adhesive tape. The
contact pads of the MEA chip are vertically connected to the
potentiostat using an array of 48 spring-loaded pins for easy
and fast “plug and play” replacement of the disposable MEA
chip. Beyond the highly increased practical convenience for
the user, the miniaturization achieved by this mounting
(avoiding connection cables between the electrodes and the
potentiostat) makes it possible to minimize the impact of input
parasitic capacitance, and consequently keep the signal
resolution as good as possible.
The MEA, placed at the bottom of the cell culture chambers
(see inset 2 of Fig. 3), consists of a single silicon chip hosting
an array of 12 independent measuring sites for cell based
assays. Each measuring site is composed of an individually
addressable microfluidic cell culture chamber containing one
counter (CE), reference (RE) and a pair of interdigitated
working electrodes (WE) (see inset 1 of Fig. 3). The two WEs
(WEa and WEb) can be addressed independently and each is
composed of 12 fingers (10µm width and spacing, 500µm
length) covering an area of 6·104µm
2. All the 48 planar
electrodes of the array are made of gold, deposited on a silicon
dioxide substrate, patterned with a lift-off process and
passivated with a silicon nitride layer (500nm thick). The REs
have been fabricated using gold instead of standard materials
as Ag/AgCl because of its simpler fabrication process. The use
of pseudo-reference gold electrodes results in a reference
potential shift and in a less controlled drift, that is acceptable
during short amperometric measurements (max. ~10 min) [9].
A custom software, run by a laptop computer, controls the
instrument through a USB interface by means of a portable
commercial data acquisition unit (NI USB-6259 by National
Instrument). (Fig. 1) The potentiostat board is connected to
this acquisition unit through an intermediate interconnection
unit that allows matching the single cable (68 pins for power
supply, analog and digital signals) from the potentiostat to the
two cables required by the USB-6259, thus simplifying the
handling of the board mounted with the pumps on the
microscope stage.
B. Operating Modes
The system can work in two different operating modes: (i)
two-electrode mode, where the instrument applies a voltage
signal to the CE and reads the current signal from the WE, and
(ii) three-electrode mode, where the voltage signal is applied
to the electrode-electrolyte interface in a closed-loop manner,
by driving the CE and monitoring the potential at the RE, still
reading the current in WE. The first operating mode is mainly
used when performing impedance spectroscopy measurements,
whereas the second one is typically used when performing
amperometric and voltammetric recordings to avoid the
distortion due to the non linear voltage drop at the CE-
electrolyte interface. Several electrochemical measurements
can thus be performed on the array by changing the operating
mode and the voltage stimulation waveform.
For maximum versatility, impedance spectroscopy can be
performed in vertical or coplanar mode. In the first case, the
sinusoidal signal is applied to the CE and the current is read at
one WE, thus probing a single WE-solution interface for
standard Electrical Cell Impedance Sensing (ECIS) detection
[17]. In the second case, the signal is applied to one set of
fingers (WEb) of the interdigitated couple, thus measuring the
impedance between the two sets (a versus b) [18].
In addition to these operating modes, WEa and WEb of
each pair of interdigitated electrodes can be biased at two
independent potentials (e.g. the oxidation and the reduction
potential of a specific target molecule). This feature facilitates
redox cycling experiments, where oxidation of a chemical
species occurs at one WE, following reduction of the oxidized
form at the other WE of the interdigitated structure. The
product of the reaction at one electrode becomes the source of
the reaction at the other electrode, and the target molecule
contributes to the current response with several electrons. This
has the overall effect of improving the current response of the
detection technique through signal amplification [19].
III. MULTICHANNEL POTENTIOSTAT HARDWARE DESIGN
The potentiostat electronic board (see Fig. 4) is composed
of two fundamental blocks: the voltage generation circuit
(often referred to as the potentiostatic circuit) and the current
reading amplifiers. This architecture in which the voltage is
forced (and current is sensed) is the natural choice to perform
all the measurements with the same analog front-end.
A. Potentiostatic Circuit
The voltage generation circuit is schematically shown in
Fig. 5. The voltage of the REs is read by 12 buffers
implemented using the OPA4141 JFET input operational
amplifiers and they are used to protect the small RE from non
idealities, such as current leakage from the analog
multiplexers. Two analog multiplexers (DG406) select the
active RE and CE in the MEA. The desired voltage signal is
>TBCAS-2011-Sep-0136-Reg.R1<
3
applied to the non-inverting input of an operational amplifier
(OPA124) that drives the voltage of the active CE. Due to the
negative feedback, the same voltage is applied also to the non
inverting input of the operational amplifier, which has the
same potential as the active RE. The voltage generation circuit
has ±2V voltage range to match the potential windows of
standard redox reactions and has 3µV resolution. To ensure
these performances, the outputs of two DACs (±10V range, 16
bit resolution) have been summed with a different attenuation,
respectively by a factor of 5 and 100. The OPA4141 buffers
have a 10MHz closed loop bandwidth, while the OPA124 has
been chosen for its 1.5MHz gain-bandwidth product and 65°
phase margin. This choice maintains the stability of the
potentiostatic loop under the expected operating conditions.
B. Current Reading Amplifiers
The electronic board (see Fig. 4) implements four
independent current sensing channels and analog multiplexers
to address the 24 working electrodes. With respect to using a
dedicated transimpedance amplifier for every WE, our choice
reduces size so to accommodate the integrated system on the
stage of an optical microscope without paying for the added
load of the multiplexer ADG1234, that has been chosen for its
low stray capacitance to ground, not influencing the resolution.
The circuit details of each current reading channel are
shown in Fig. 6. Every transimpedance amplifier biases the
active WE at the desired voltage, while simultaneously reading
the current flowing through it. The inactive WEs are biased
through the ADG1234 at the same potential as the active one,
allowing the maintenance of the electrochemical steady state
conditions at the WE-liquid interface and leading to a fast
response when they are selected. The ADA4817 JFET input
operational amplifier has been chosen to implement the
transimpedance stage due to its low noise and dynamic
performance (equivalent noise sources of 4nV/√Hz and
2.5fA/√Hz, 410MHz gain bandwidth product). The feedback
resistor and capacitor have been set to 100kΩ and 10pF,
respectively, to ensure a 100kHz bandwidth and a 400fA/√Hz
input referred current noise. The current amplifier has been
optimized for a maximum capacitive input load of 30nF, five
times the double layer capacitance of our WEs estimated on a
typical capacitance of 0.1pF/µm2 [12] of a metal immersed in
phosphate buffered saline (PBS). A second amplification stage
(based on the instrumentation amplifier INA128 with a gain of
10) and a second-order filter complete the current
amplification chain. The 100kHz bandwidth, consistent with
the performance offered by commercial systems (ECIS Z by
Applied Biophysics and exCELLigence by Roche) is sufficient
to reach the resistive plateau given by the solution resistance
as for the fabricated electrodes in PBS the corner frequency
between the double layer capacitance (6nF) and the solution
resistance (~1kΩ) is at 24kHz. Beyond allowing the use of a
USB acquisition board, the choice to limit the bandwidth of
the current sensing amplifier enables the increase of the
feedback resistor up to 100kΩ, reducing the input referred
current noise at low frequencies down to 400fA/√Hz. The low-
pass (10kHz) filtered output (LPF) is used in amperometric
and voltammetric recordings to relax the sampling and
processing requirements for the acquisition board and PC.
C. Summary of Performance
The electronic performance of the multichannel potentiostat
is summarized in Table 1. The details of the experimental
characterization are reported in sec. V: here we anticipate that
all the design specifications have been achieved. For
voltammetry mode, the maximum scan rate (SR) has been
determined from the bandwidth of the current sensing
amplifier using the relation BW = 40V-1SR [20] that relates
the delay of a single pole to a maximum potential error of
4mV.
The current resolution of 5pA has been measured using a
proper digital filter to limit the bandwidth of the current
sensing circuit at 5Hz. This performance translates into the
ability to perform amperometric detection of neurotransmitter
exocytosis from a cell population corresponding to current
pulses with amplitudes in the 800pA range and durations of
several seconds [21]-[22].
IV. DATA ACQUISITION AND ANALYSIS SOFTWARE
In order to control and record data from the electronic
board, as well as to display and process the data in real-time,
with a single user interface, a custom software has been
developed in LabVIEW environment (National Instruments,
Austin TX). The software has a modular structure in order to
allow easy addition of new measurement techniques/protocols
as well as new analysis features. The modules have been
designed to be compatible with all DAQmx National
Instruments boards, requiring 8 analog acquisition channels, 4
analog generation channels, and 14 digital lines. The chosen
NI USB-6259 acquisition unit (see Fig. 1) allows a
multiplexed 1MSample/s acquisition-generation rate and is
connected to a laptop computer via USB in order to maintain
the “plug-and-play” architecture of the entire system.
The software is composed of different acquisition modules
which facilitate electrochemical measurements using four main
techniques: amperometry, cyclic voltammetry (both in single
mode and redox cycling mode), impedance measurements,
both in ECIS and coplanar configuration and both at a single
frequency (impedance tracking) or in a selectable frequency
TABLE 1
SUMMARY OF SYSTEM PERFORMANCE
Generation DC voltage range ±2V
Generation DC voltage resolution 3µV
Min. generation AC amplitude 30µV
Transimpedance gain 1MΩ
Max. input current ±10µA
Current resolution B~5Hz 5pA
Current resolution B~100kHz 2.3nA
Operating frequency range (EIS) 10mHz - 100kHz
Max. scan rate (CV) 2.5kV/s
Max. input capacitance 30nF
Power dissipation (supply voltage) 4W (±15V)
>TBCAS-2011-Sep-0136-Reg.R1<
4
range (impedance spectroscopy and tracking of the spectra
over time). Impedance is extracted from the acquired data
through a digital implementation of a phase-sensitive lock-in
detection algorithm [23], taking advantage of the synchronous
generation and acquisition of analog samples enabled by the
board. Measurements using any of the techniques can be
conducted in an arbitrary sequence of electrodes in the array at
a freely defined time interval in order to automate time laps
experiments on cells in culture.
Since a calibration procedure is necessary to ensure high
accuracy in impedance spectroscopy by compensating
magnitude and phase errors (due to stray input impedance and
to the finite bandwidth of the conditioning circuits), an
automatic calibration circuit has been implemented. Whenever
necessary, a relay connects each input to previously
characterized impedance (a RC cell) mounted on the board,
then a full spectrum is recorded and the correction factors,
calculated for each frequency from discrepancy with respect to
the ideal spectrum, are stored.
All recorded data are displayed during the measurement and
some data analysis can be performed on-line in order to help
the user to monitor the behavior of the cultured cells in real-
time. For example, the user can select an equivalent circuit
from a set of pre-programmed equivalent electrical circuits in
order to fit automatically each of the recorded impedance
spectra. The fitting algorithms, implemented as compiled C#
libraries in the LabVIEW code for maximum speed of
execution, provide the characteristic electrical parameters of
the electrode-cell-solution interfaces.
V. EXPERIMENTAL RESULTS
A. Functional Characterization
The electronic instrumentation has been fully tested in terms
of bandwidth, gain, linearity and noise. Fig. 7 shows the
transfer functions of the four independent Current Reading
Amplifiers, measured with an HP4195A network/spectrum
analyzer. When operated in full bandwidth, they show a
constant gain (1MΩ) up to a frequency bandwidth above
100kHz whereas when the LPF output of Fig. 6 is used the
bandwidth is properly limited to about 10kHz. This second
option is important in increasing the resolution of the system
when speed is not a concern as in cyclic voltammetry. This can
be better understood by considering the input referred current
noise of the electronic platform for different capacitances of
the sensing electrode, measured with the HP4195A, reported
in Fig. 8. When loading the Current Reading Amplifier with
minimum capacitance (Cin = 56pF), we obtain the intrinsic
noise floor of the electronic board which is flat and below
500fA/Hz. The curve of Cin = 5.6nF represents the typical
operative situation when the WE of the array chip is connected
to the board. The scatter points in the graph are the values
expected from theoretical simulations in P-Spice based on the
voltage noise of the amplifier and on Cin. Because of the high
input capacitance that increases the noise spectral density at
high frequencies, the bandwidth limiter at 10kHz is very
effective in reducing the rms noise of the platform to a factor
of almost 30.
To assess the accuracy of the electronic section of the
platform and its bandwidth, an EIS measurement has been
performed on gold interdigitated electrodes immersed in PBS.
The impedance of the same sample by using our multichannel
platform has been compared to a stand-alone commercial
instrument (Agilent E4980A Precision LCR meter) as a
reference. Two impedance measurement modes have been
tested: the vertical (impedance of a single WE with respect to
the CE) and the coplanar (impedance of WEa with respect to
WEb) techniques. The comparison of experimental results by
the two instruments is presented in Fig. 9. In the frequency
range where both instruments can work, the same impedance
values have been recorded, thus proving the correct behavior
of the proposed multichannel potentiostat.
To further assess the capability of the instrument to address
each single electrochemical cell of an array and to handle
cyclic voltammetry measurements, the electrochemistry of the
interdigitated gold electrodes with standard redox mediators
has been sequentially tested. A self-assembled monolayer of
cysteamine (a short thiol molecule with an amine endgroup)
was deposited on the gold electrodes in order to get a
controlled and defined electrode surface, and then cyclic
voltammograms were recorded from all the 24 WEs versus the
on-chip gold RE when both immersed in a solution containing
10mM ferricyanide (K3Fe(CN)6) and ferrocyanide
(K4Fe(CN)6) in PBS. Fig. 10 depicts the voltammograms,
which indicate a very good reproducibility between all the
cysteamine modified electrodes.
B. Dynamic Electrochemical Measurements
The key feature of the platform to have the electronics
tailored onto a microfluidic network, thus resulting in an
integrated system for programmable on-line multi-
electrochemical measurements, has been investigated with a
sequence of amperometric tests. These comprised multiple
injections of a 62.5µM solution of ferricyanide in PBS into the
measurement chambers separated by injection of rinsing buffer
solution. During the experiment, the redox reaction was
continuously monitored. To this aim the cysteamine modified
WE were kept at -350mV with respect to the gold RE using
the three electrodes configuration. The amperometric current-
time trace generated by the system can be seen in Fig. 11. The
baseline was recorded while flushing PBS with 34µl/min flow
rate. Then the pump started to inject the ferricyanide solution
from a reservoir at the same flow rate. It takes about 12s for
the solution to reach the inlet of the microfluidic chamber.
This causes a delay before the current response (a negative
current signal due to reduction of ferricyanide) is observed.
The transient time, before reaching the saturation signal, is due
to the diffusion of the analyte in the chamber previously filled
with PBS buffer. After stopping the flow, the current response
significantly decreases due to purely diffusion-based mass
transfer. After starting a flow of PBS, the current response
reaches the previous value until the PBS completely rinse the
chamber removing the ferricyanide. As a result, the current
>TBCAS-2011-Sep-0136-Reg.R1<
5
response returns to the initial baseline. By repeating the same
experiment with different ferricyanide concentrations and
evaluating the magnitude of the current response with respect
to the PBS baseline, we have obtained the calibration curve of
Fig. 12, showing linear response with respect to concentration.
For each of the concentration values, the analyte was injected
three times in order to assess the repeatability of the
measurement, obtaining a deviation smaller than 1%. The
responses from 2 different WEs in the same chamber have
been recorded, giving very similar response with a deviation of
only 0.5%. The rms input noise in this experiment is
dominated by interferences caused by the peristaltic pumps
(see next section for further details) and can be quantified in
105pArms. Under these dynamic experimental conditions, the
detection limit is consequently of 1.6µM.
C. Analysis of the Noise Effects of Peristaltic Pumps
The sequence of amperometric tests has also allowed to
carefully explore the interplay between microfluidic operation
and electronic signal detection. The disturbing effect of the
peristaltic pumping on amperometric recordings is well known
and has previously been reported in the literature [9]. To
highlight the effect of our peristaltic pumps, Fig. 13(a) depicts
a current signal resulting from the injection of 10mM
potassium ferricyanide in one microfluidic chamber. The
signal has been acquired with 10kHz bandwidth (from LPF,
see Fig. 6) at 40kHz sampling frequency to avoid aliasing
effects. Observing the black line signal of Fig. 13(a), the effect
of peristaltic pumping is evident: when the pumps are on
spikes are presents, when the pumps are off fluctuations
drastically reduce. Also by disconnecting the microfluidics
from the motors but still operating the motors, the fluctuations
reduce drastically as well, thus verifying that electromagnetic
interferences are not significantly contributing to the noise.
Instead, we attribute the main noise contribution to the
pulsating nature of the flow that produce a movement of
charged ions perturbing the electrode-electrolyte interface and
thus resulting in spikes of ionic current.
To investigate the effectiveness of heavy filtering in
reducing these spikes, a bandwidth reduction down to 5Hz
(using a simple mobile average digital filter with 100ms
period) has been applied. A value lower than 5Hz would have
affected the tracking of relevant biochemical signals. The
positive effect of the 5Hz filter in smoothing the curve can be
seen (gray curve) in Fig. 13(a). Nevertheless, the effect of the
pumps has not been eliminated as the residual rms noise can be
quantified in 105pArms. As a comparison, Fig. 13(b) shows the
system response to a 100pA input step after the same 5Hz
filtration when the signal is applied to the current sensing
amplifier through a 4.7nF input capacitance and the flow is not
playing a role. From Fig 13(b) a 5pArms noise level can be
quantified as the intrinsic current resolution of the system. The
residual noise contribution of the peristaltic pumps can be thus
considered the limiting noise level of our system when
operating dynamic measurements.
D. Impedimetric Tracking of Cell Proliferation
Dopaminergic PC12 cells have been used as a robust model
cell line for preliminary cell-based assays with the proposed
platform. The proliferation of PC12 has been continuously
monitored using automatic EIS tracking. To further
demonstrate the versatility of the instrument, commercially
available electrodes (8W2x1E array by Applied Biophysics)
have been employed. These arrays contain 16 gold disk
electrodes with 250µm nominal diameter, corresponding to a
WE area of 4.9·104µm
2, i.e. 5nF, compatible with the
instrument capability to handle a large input capacitance.
Before starting the experiment, the sterilized chambers were
coated with laminin to promote cellular adhesion. Dulbecco’s
Modified Eagle’s medium (F12/Glutamax) supplemented with
15% horse serum, 2.5% fetal bovine serum and 1%
penicillin/streptomycin was used as culture medium. About
2.5·105 cells have been initially seeded into each well and they
were cultured for 7 days (at 37° C, in a humidified atmosphere
with 5% CO2). During the first 3 hours after cell seeding,
impedance spectra were recorded every 13 minutes to observe
in more detail the cell settling and further on cell adhesion and
spreading. Then spectra were acquired every hour, while
taking a picture of the electrodes with a bright-field
microscope every 24 hours in correspondence with the change
of culture medium. 30 points were acquired in the frequency
range 100Hz-100kHz, each with an averaging time of 2s that
provided a sufficient noise reduction.
The amplitude of the applied sinusoidal signal was set to
only 215µV in order to limit the current flowing through the
cells below a safety threshold value (250nA for PC12, 1A
for HeLa) without the inconvenience of an external series
resistor, commonly adopted [17]. Fig. 14(a) reports the
normalized impedance spectra taken during the experiment.
For each time instant, the normalized impedance is defined as
Znorm(t,f) = (|Z(t,f)| - |Z(0,f)|)/ |Z(0,f)|; where |Z(t,f)| is the
magnitude of the impedance spectrum acquired at the given
time instant. This method is commonly used to highlight
relative variations of the impedance magnitude with respect to
the impedance magnitude measured at the beginning of the
experiment (|Z(0,f)|). Consistently with what was reported in
[18], it has been possible to determine a peak frequency
(69kHz, approximately at the corner between the double layer
capacitive slope and the solution resistance plateau), where
impedance shows the highest sensitivity to the population
growth. The initial decrease of the impedance magnitude is
most probably due to the thermal transient occurred to the
system after it has been placed inside the incubator. The
increase in impedance magnitude at high frequency is
correlated with the electrode coverage fraction, as confirmed
by optical images shown in Fig. 14(b). It reaches a maximum
of 35% at confluence (t = 125 hour) and then slightly
decreases when some cells start to detach.
E. Dopamine Detection
In parallel to the demonstration of the platform capability to
host PC12 cultures for several days and monitor their growth
in a real-time and label-free manner, preliminary tests with
>TBCAS-2011-Sep-0136-Reg.R1<
6
cell-related analytes have been performed. In particular,
among various neurotransmitters secreted by cells, dopamine
is of major interest being a clear and early indicator of the
differentiation of stem cells into adult neural cells. As shown
in Fig. 15(a), the presence of dopamine can be detected by
means of impedance spectroscopy (VDC = 75mV). In this case,
performing a full spectrum from 10mHz to 100kHz of a
diluted solution of dopamine in PBS (5µM), a charge transfer
resistance of 140MΩ and a mass-transport Warburg term
become apparent at low frequency.
As impedance spectroscopy is not specific to the redox
reaction, being measured at a constant VDC potential, selective
detection of catecholamines can be performed by means of
cyclic voltammetry. In Fig. 15(b), the same concentration of
dopamine is detected: the oxidation and reduction peaks
(ΔI~2nA) are visible (at a scan rate of 100mV/s), with respect
to a background signal measured with buffer only (PBS). In
this measurement, the redox cycling mechanism achieved with
the pair of interdigitated working electrodes is also shown. The
bipotentiostat allows performing cyclic voltammetry both in
single mode and in cyclic mode. In the latter, WEb is scanned
while WEa is biased at the reduction potential (0V). A current
increase of a factor 1.5 is observed, supported by an extra
cycling current of -2nA measured at WEa. Further biological
experiments performed by means of the proposed platform,
such as the amperometric detection of the dopamine exocytosis
from a population of PC12 cells, are reported in [22].
VI. CONCLUSION
A novel and compact 24-channel instrument tailored for an
array of planar microelectrodes for multi-parametric
quantitative electrochemical measurements has been presented.
The bipotentiostat is originally integrated with a microfludic
platform to allow automatic monitoring of cell cultures.
Though designed for a specific combination of ad-hoc
microfrabicated electrodes and microfludics, meant for opto-
electrical monitoring of neural stem cell cultures, it is suitable
for routinary use in biology investigations, provinding a user-
friedly software and being compatile with a wide range of
commercial arrays of electrodes. The experimental
characterization of the portable system confirmed the
achievement of the design specifications (100kHz bandwidth,
few pA resolution). Continuous operation during 7 days of
impedimetric tracking of PC12 proliferation and dopamine
detection in the µM range have been sucessfully demostrated.
REFERENCES
[1] F.H. Gage, “Mammalian neural stem cells,” Science, vol. 287, pp. 1433-
1438, 2000.
[2] O. Lindvall, Z. Kokaia and A. Martinez-Serrano, “Stem cell therapy for
human neurodegenerative disorders- how to make it work,” Nat. Med.,
vol. 10, pp. S42-S50, 2004.
[3] F.J. Sigworth and K.G. Klemic, “Microchip Technology in Ion-Channel
Research,” IEEE Trans. Nanobiosci., vol. 4, no. 1, pp. 121-127, 2005.
[4] P. Weerakoon, E. Culurciello, K. G. Klemic, and F. J. Sigworth, “An
Integrated Patch-Clamp Potentiostat With Electrode Compensation,”
IEEE Trans. Biomed. Circuits Syst., vol. 3, pp. 117-125, 2009.
[5] L. Berdondini, K. Imfeld, A. Maccione, M. Tedesco, S. Neukom, M.
Koudelka-Hep and S. Martinoia, “Active pixel sensor array for high
spatio-temporal resolution electrophysiological recordings from single
cell to large scale neuronal networks,” Lab Chip, vol. 9, pp. 2644-2651,
2009.
[6] R.M. Wightman, “Probing cellular chemistry in biological systems with
microelectrodes,” Science, vol. 311, pp. 1570-1574, 2006.
[7] I. L. Jones, P. Livi, M. K. Lewandowska, M. Fiscella, B. Roscic, A.
Hierlemann, “The potential of microelectrode arrays and
microelectronics for biomedical research and diagnostics,” Anal.
Bioanal. Chem., vol. 399, pp. 2313-2329, 2011.
[8] L. Li, X. Liu, W. A. Qureshi, and A. J. Mason, “CMOS Amperometric
Instrumentation and Packaging for Biosensor Array Applications,”
IEEE Trans. Biomed. Circuits Syst., vol. 5, pp. 439–448, 2011.
[9] J. Gottschamel, L. Richter, A. Mak, C. Jungreuthmayer, G. Birnbaumer,
M. Milnera, H. Brückl, and P. Ertl, “Development of a disposable
microfluidic biochip for multiparameter cell population measurements,”
Anal. Chem., vol. 81, pp. 8503-8512, 2009.
[10] X. Yue, E.M. Drakakis, M. Lim, A. Radomska, H. Ye, A. Mantalaris,
N. Panoskaltsis, and A. Cass, “A real-time multi-channel monitoring
system for stem cell culture process,” IEEE Trans. Biomed. Circuits
Syst., vol. 2, pp. 66-77, 2008.
[11] C. Spegel, A. Heiskanen, L.H.D. Skjolding, and J. Emnéus, “Chip based
electroanalytical systems for cell analysis,” Electroanalysis, vol. 20, pp.
680-702, 2008.
[12] A.J. Bard and L.R. Faulkner, Electrochemical methods: fundamentals
and application, 2001.
[13] C. Spégel, A. Heiskanen, J. Acklid, A. Wolff, R. Taboryski, J. Emnéus,
and T. Ruzgas, “On-Chip determination of dopamine exocytosis using
mercaptopropionic acid modified microelectrodes,” Electroanalysis,
vol. 19, pp. 263-271, 2007.
[14] P.O. Bagnaninchi and N. Drummond, “Real-time label-free monitoring
of adipose-derived stem cell differentiation with electric cell-substrate
impedance sensing,” PNAS, vol. 108, pp. 6462-6467, 2011.
[15] H.E. Park, D. Kim, H.S. Koh, S. Cho, J.-S. Sung, and J.Y. Kim, “Real-
time monitoring of neural differentiation of human mesenchymal stem
cells by electric cell-substrate impedance sensing,” J. Biomed. Biotech.,
p. 485173-485191, 2011.
[16] P. Skafte-Pedersen, D. Sabourin, M. Dufva, and D. Snakenborg, “Multi-
channel peristaltic pump for microfluidic applications featuring
monolithic PDMS inlay,” Lab Chip, vol. 9, pp. 3003-3006, 2009.
[17] I. Giaever and C.R. Keese, “Monitoring fibroblast behavior in tissue
culture with an applied electric field,” PNAS, vol. 81, pp. 3761-3764,
1984.
[18] L. Wang, L. Wang, H. Yin, W. Xing, Z. Yu, M. Guo, and J. Cheng,
“Real-time, label-free monitoring of the cell cycle with a cellular
impedance sensing chip,” Biosens. Bioelectron., vol. 25, pp. 990-995,
2010.
[19] M. Paeschke, U. Wollenberger, C. Köhler, T. Lisec, U. Schnakenberg
and R. Hintsche, “Properties of interdigital electrode arrays with
different geometries”, Anal. Chim. Acta, vol. 305, pp. 126-136, 1995
[20] R.M. Wightman and D.O. Wipf, “High-speed cyclic voltammetry,”
Accounts Chem. Res., vol. 23, pp. 64-70, 1990.
[21] A. Heiskanen, C. Spégel, J. Tønnesen, Z. Fohlerova, L. Goulart, J.
Hansen, M. Kokaia, M. Dufva, and J. Emnéus, “Development of a
microfluidic on-line culture system for combined electrochemical and
optical real-time detection of cellular processes,” Proc. of µTAS 2008,
San Diego (USA), vol. 2, pp. 1168-1171, 2008.
[22] K. Zór, M. Vergani, A. Heiskanen, E. Landini, M. Carminati, V.
Coman, I. Vedarethinam, P. Skafte-Pedersen, M. Skolimowski, A.
Martinez Serrano, M. Kokaia, T. Ramos Moreno, A. Ghio, W.E.
Svendsen, M. Dimaki, Zs. Kerestes, M. Adamovski, U. Wollenberger,
D. Sabourin, G. Ferrari, R. Raiteri, M. Sampietro, M. Dufva, J. Emnéus,
"Real-time monitoring of cellular dynamics using a microfluidic cell
culture system with integrated electrode array and potentiostat.," Proc.
of µTAS 2011, Seattle (USA), vol. 1, pp. 1532-1535, 2011.
[23] M. Carminati, G. Ferrari, and M. Sampietro, “Attofarad resolution
potentiostat for electrochemical measurements on nanoscale
biomolecular interfacial systems.,” Rev. Sci. Instrum., vol. 80, pp.
124701:1-10, 2009.
>TBCAS-2011-Sep-0136-Reg.R1<
7
Marco Vergani, born in 1985, received the Bachelor
degree cum laude from Politecnico di Milano (Italy) in
2007 and a Double Master of Science cum laude from
the same university and Politecnico di Torino in 2009.
In January 2010 he started a PhD program in
Information Engineering at Politecnico di Milano. He
is currently working on low-noise instrumentation for
electrical measurements on biological samples.
Marco Carminati, born in 1981, received B.Sc. and
M.Sc. in Electronic Engineering, both cum laude from
the Politecnico di Milano (Italy), in 2003 and 2005
respectively. In 2008 he spent a semester at MIT
working on BioMEMS and microfluidics. In 2009 he
completed his PhD in Electronics and Information
Science at DEI, Politecnico di Milano where he is
currently a post-doctoral fellow working in the field of
bio-chemical sensors and instrumentation.
Giorgio Ferrari, born in 1973, received the Master
degree and the Ph.D. degree from Politecnico di
Milano in electronics engineering in 1999 and 2003,
respectively. Since 2005, he has been an Assistant
Professor of electronics at Politecnico di Milano. The
focus of his research activities is on the development of
novel instrumentation to probe electrical properties of
materials and biosamples at the nanoscale. He has
coauthored more than 30 peer-reviewed publications
and three patents.
Ettore Landini, born in 1984, received his Master
Degree in Bioengineering in 2008 from the University
of Genova He is now taking his PhD in
Nanotechnologies at the Department of Biophysical
and Electronic Engineering (University of Genova). He
works mainly with the Atomic Force Microscope to
investigate the changes in the mechanical properties of
articular cartilage due to degenerative pathologies like
osteoarthritis.
Claudia Caviglia, born in 1985, received her Bachelor
degree in Biomedical Engineering from University of
Genova in 2008 and her Master degree cum laude in
Bioengineering from the same university in 2010. In
May 2011, she started a PhD program at DTU,
Nanotech. She is currently working on development of
safer non-viral gene transfection vectors.
Arto Heiskanen, born in 1965, received B.Sc in
biochemistry in 1987 at Åbo Akademi, Finland and
B.Sc. in chemistry in 2001 at University of the
Philippines. In 2004, he received M.Sc. and in 2009
PhD at the Department of Analytical Chemistry of
Lund University, on electrochemical monitoring of
cells. Currently, he is a Assistant Professor at DTU
Nanotech, focusing on the development of microfluidic
systems. He has co-authored 20 peer-reviewed
publications.
Clément Comminges, born in 1981, received his
Master degree in electrochemistry in 2004 from the
University Pierre et Marie Curie, Paris (France) and
Ph.D in electrochemistry in 2007 from the University
Paris 12 (France). Since 2010, he has a Postdoctoral
position in Potsdam University (Germany) in the group
of molecular enzymology. His research interests are the
sensitive electrochemical detection of biological
relevant compounds and biosensors.
Kinga Zór born in 1980, received B.Sc. from Babeş-
Bolyai University (Romania) in 2003. Currently she is
a PhD student at the Department of Biotechnology,
Lund University (Sweden). Her research is focused on
electrochemical (bio)sensors in miniaturized systems
for bioanalysis and on application of microfluidic on-
line cell culture systems for real-time monitoring of
cellular dynamics. During 2011 She spent a six months
period at DTU Nanotech.
David Sabourin, born in 1975, received both B.Sc. in
Biochemistry and B.A.Sc. in Chemical Engineering
from the University of Ottawa (Canada) in 1999. From
1999 to 2007 he held a position at i-STAT, a
manufacturer of microfluidic diagnostic devices. From
2007 to 2010 he completed his PhD at DTU Nanotech
developing a microfluidic platform with innovative
scalable interconnection and pumping solutions.
Martin Dufva, born in 1967, received master degree
in molecular biology in 1998 and PhD degree at the
medical faculty of University of Göteborg (Sweden) in
2001. He worked with DTU Nanotech since 2001 and
since 2005 holds a position as Associate Professor. His
research interests are microsystems for cell and
molecular biology research and actuation and
technologies for cancer and genetic diagnostics. He has
co-authored >50 publications.
Maria Dimaki received an M.Eng. from the National
Technical University of Athens, Greece, in 2000 and
an M.Sc. from the Imperial College, London in 2001.
She received her Ph.D. in 2005 from the Technical
University of Denmark. She is currently an associate
professor with the Nano-Bio Integrated Systems group
(NaBIS) at DTU Nanotech. Her research interests
involve the development of micro- and nanoelectrodes
in microfluidic systems for electrophysiological and
electrochemical measurements on neuron cultures.
Roberto Raiteri got his Master Degree in Electronic
Engineering in 1993 from the University of Genova
and his PhD in electronics in 1997 from University of
Trento. Since 2001 he is Assistant Professor in
Bioengineering at the University of Genova. His main
research interests deal with the use of Scanning Probe
Microscopy-based methods to characterize
biomolecules and biological tissues at the nanometer
scale. He is co-author of 32 peer-reviewed international
publications.
Ulla Wollenberger, received her PhD in Enzymology
from the Academy of Sciences of the GDR in 1984. In
1994 she joined the University Potsdam (Germany).
Currently she is group leader in the Department of
Molecular Enzymology and is working in the field of
bioelectrochemistry, biosensors and bioelectronic. She
published more than 120 peer reviewed papers and a
textbook.
Jenny Emnéus, born in 1961, received M.Sc. in 1986
and PhD in 1992 at Lund University, where she also
held the position of Associate Professor until 2007. In
2007, she became Professor at the Department of
Micro- and Nanotechnology, DTU, to lead research
focusing on development of microfluidic bioassay
systems. She has coordinated 6 European projects.
During her research career, she published over 100
peer-reviewed publications.
>TBCAS-2011-Sep-0136-Reg.R1<
8
Marco Sampietro, born in 1957, received his Master
Degree in Nuclear Engineering in 1982 from
Politecnico di Milano where he is now full professor of
Electronic Circuits and Devices. He is responsible for
the activities in high-sensitivity instrumentation for the
nanoscience. He is co-author of more than 150 peer-
reviewed international publications. He has been vice-
dean of the Faculty and the coordinator of many
national and international research projects and
scientific partner in 4 European projects.
Microfluidics
Multichannel Potentiostat
Acquisition unit
Array of cell culture chambers
USB
Control and analysis software
Fig. 1 Schematic view of the whole system architecture showing the
electronic board tailored onto the cell culture chambers and coupled to
microfluidics in a compact assembly placed under an upright microscope.
Spring
pins
Cells
Microscope
PMMA
Board
Silicon tape
Liquid inlet
MicroelectrodesChip
Screws
Contact pads
2mm
0.8mm
Fig. 2 Cross-section of the electronic board at the cell culture site integrated
with the array of planar microelectrodes and the microfluidic culture system.
Fig. 3 Photograph of the complete instrumentation platform. The board is
black to minimize autofluorescence. (Inset 1: enlargement of a sensing site of
the MEA; inset 2: window enabling microscope observations).
CEsREs
WE1-6,a
WE1-6,b
WE7-12,a
WE7-12,b
NI-USB
6259MEA
Current reading
amplifier
Potentiostatic
circuit
RE
WEa CE
Sensing site
ADC
DAC
Current reading
amplifier
Current reading
amplifier
Current reading
amplifier
WEb
Fig. 4 Multichannel potentiostat architecture comprising the potentiostatic
circuit driving the RE/CE and 4 current amplifiers connected to the 24 WEs
hosted in the MEA chip, through low-parasitics analog switches.
+
-To
CE+
-
200 Ω
1 kΩ
20 kΩ
From
RE
+
-
+
-
OPA4141
DG
406
VAC
VDC
100
V-
5
V- ACDC
OPA124
2/3 electrode
selector
DG
406
1.
.
.
.
12
1.
.
.
.
12
THS4032
Adder
Fig. 5 Detailed architecture of the potentiostatic circuit: the DC and AC
signal, generated by two DACs, are summed with different attenuation (5 and
100) to allow high accuracy. The potentiostatic loop is closed by OPA124.
1 kΩ
3.9 nF
10 kΩ
390 pF
100 kΩ
10 pF
+
-ADA4817
+
-
INA128
x 10
Full BW
LPFADG
1234
IinIin
VWE VWE
1.
.
.
.
.
6
Dout
Fig. 6 Detailed architecture of the current reading chain composed by the
input multiplexer, a transimpedance amplifier and INA-based second stage.
>TBCAS-2011-Sep-0136-Reg.R1<
9
100 1k 10k 100k 1M100
1k
10k
100k
1M
10 kHz Channels
Tra
nsim
pedance (
)
Frequency (Hz)
100 kHz Channels
Fig. 7 Transfer function of the four Current Reading Amplifiers operating at
full bandwidth (100kHz) or at increased resolution (10kHz).
100 1k 10k 100k100f
1p
10p
Input N
ois
e (
A/s
qrt
(Hz))
Frequency (Hz)
Cin = 56 pF
Cin = 560 pF
Cin = 5.6 nF
Fig. 8 Spectral density of the input referred current noise measured with
different input capacitances. Scatter points represent the values obtained by
theoretical simulations, while the measured values are plotted with continuous
lines. With a 5.6nF input capacitance, the capacitance resolution estimated
from this spectrum is 2fF (VAC = 10mV at 10kHz, 1s measurement time).
1 10 100 1k 10k 100k 1M100
1k
10k
100k
1M
10M
Agilent LCR:
Coplanar
Vertical
Multichannel Potentiostat:
Coplanar
Vertical
Imp
ed
an
ce
Ma
gn
itu
de
)
Frequency (Hz) Fig. 9 Test result of the impedance measurement. The impedance spectra
recorded with the designed multichannel platform overlap the measurement
performed with a stand-alone commercial instruments (Agilent Precision LCR
Meter E4980A) both for coplanar and vertical measurement technique.
-0.4 -0.2 0.0 0.2 0.4-4.0µ
-3.0µ
-2.0µ
-1.0µ
0.0
1.0µ
2.0µ
3.0µ
4.0µ
Cu
rre
nt
(A)
Voltage (V) Fig. 10 Cyclic voltammograms recorded on a whole array of WEs. A 10mM
solution of ferricyanide (K3Fe(CN)6) and ferrocyanide (K4Fe(CN)6) in PBS
was used as the electrolyte. The recordings were made at the scan rate of
50mV/s vs the on-chip gold RE.
0 20 40 60 80 100-12.5n
-10.0n
-7.5n
-5.0n
-2.5n
0.0
2.5n PBSStopPBS
Cu
rre
nt
(A)
Time (s)
Analyte
Fig. 11 Amperometric recording of the injection of a 62.5µM potassium
ferricyanide solution (flow rate 34µl/min, WE biased at -350mV vs gold, 5Hz
low pass digital filter).
100µ 1m 10m
10n
100n
1µ
WEa
WEb
Linear fit
Cu
rre
nt
(A)
Concentration (M)
Fig. 12 Calibration curve for the amperometric detection of potassium
ferricyanide injected at the flow rate of 34µl/min. Both WEs were biased at -
350mV vs gold RE.
>TBCAS-2011-Sep-0136-Reg.R1<
10
Fig. 14 One week tracking of the proliferation of a population of PC12 cells: (a) plot of normalized impedance spectra recorded over time and (b)
corresponding microscopic images taken at the indicated hour.
0 10 20 30 40-3.5µ
-3.0µ
-2.5µ
-2.0µ
-1.5µ
-1.0µ
-500.0n
0.0
500.0n
1.0µ
1.5µ
(b)
C
urr
ent
(A)
Time (s)
Pumps ON Pumps ON
(a)
0 10 20 30
0
50p
100p
150p
Curr
ent
(A)
Time (s)
Fig.13 (a) Amperometric tracking showing the disturbing effect of peristaltic
pumping (black line is the signal acquired at 10kHz bandwidth whereas the
grey line is the result of a 5Hz filtration). (b) Response of the system to a
100pA input current step. The acquired signal has been digitally filtered with
5Hz bandwidth.
10m 1 100 10k
1k
10k
100k
1M
10M
100M
-0.2 0.0 0.2 0.4
-4
-2
0
2
4
6
8
10Warburg
CPE
Rsol
RCT
Imp
ed
an
ce
Ma
gn
itu
de (
)
Frequency (Hz)
WEb Buffer Only
WEb Single
WEb Cycling
WEa Cycling
Cu
rre
nt
(nA
)
Potential vs. Ag/AgCl (V)
5M DA
(a) (b)
Fig. 15 Detection of 5µM of dopamine (DA) with (a) impedance spectroscopy
(VDC = 75mV, VAC = 10mV) and (b) cyclic voltammetry (100mV/s scan rate).
The redox cycling operation is also demonstrated as the current signal at WEa
is -2nA during the oxidation of dopamine.