Installation & Operation Manual Continuous Level Indicator
-
Upload
independent -
Category
Documents
-
view
1 -
download
0
Transcript of Installation & Operation Manual Continuous Level Indicator
Installation & Operation Manual
Continuous Level Indicator
NIVOMETER MLC 830/831/832
Manufactured by
Nivo Controls Pvt. Ltd. 104 – 115 Electronic Complex
Indore 452003. INDIA
i
MLC 830
Safety Guidelines All the precautions and warning notices as mentioned in the manual
must be strictly followed to ensure personal safety and to protect the product and the other connected devices.
Qualified Personnel
This instrument should be operated and used strictly in accordance with this manual. Only a technically qualified person with basic knowledge regarding electrical practices is authorized to install and operate the instrument.
Disclaimer In order to continuously improve quality, Nivo Controls Pvt. Ltd. reserves the right to modify the technical data without prior notice. Responsibility as to the suitability and intended use of the instrument rests solely with the operator. Improper installation or operation of the instrument may result in loss of warranty.
MLC 830
ii
Contents 1 Application 1
2 Symbols 1
3 System Description 1
4 Installation Guidelines 5
5 Wiring 6
6 Calibration 7
7 Maintenance 9
8 Troubleshooting 9
9 Technical Data 11
10 Operating Principle 12
11 Appendix 13
12 Warranty 18
09/MKT/217/V0
MLC 830
1
1 Application These capacitance instruments are ideally suited for continuous
measurement of liquids and slurries of homogeneous compositions having a Stable dielectric constant. The performance of these instruments is not affected by the properties of the material such as conductivity, corrosiveness and viscosity.
2
Symbols The following symbols may have been used on the instrument or in the documentation. They should be interpreted as described below.
Symbol Description
Indicates Direct Current
Indicates Alternating Current
Earth (Ground) Terminal
Protective Conductor Terminal This terminal should be connected to ground.
Indicates an action or procedure which, if not performed correctly can result in either personal injury or incorrect operation of the device. Comply strictly with these instructions.
Risk of electric shock
3
System Description The Probe
The complete instrument consists of a Sensor and a Transmitter. 1. The Probe 2. The Preamplifier S-11 3. The Nivometer MLC
.
The probe is the active element of the system that goes within the container from top and comes in contact with the material whose level is to be measured.
The filling material may or may not be electrically conductive .It is important to select the correct probe for material in each case, keeping into consideration the size, aggressiveness, abrasion, buildup tendencies and other characteristics of the material. Fully insulated probe (rod or rope) is used for continuous level measurement. For mounting these probes (vertically) bush or flange construction is given .A cast aluminium housing (head) is provided at the top of the probe for mounting the Preamplifier S-11.
2
MLC 830
The Preamplifier S-11:
The Preamplifier circuit converts the change in capacitance, which results due to change in material level inside the container, into level proportional DC signal. The level proportional signal is transmitted to Nivometer MLC for further processing. The Preamplifier circuit is mounted on a glass epoxy board. The glass epoxy board is then encased in plastic housing and is filled with synthetic potting resin. Potting this circuit protects It against the influence of dust, moisture and vibration.
The Nivometer MLC:
The level proportional signal from Preamplifier S-11 is evaluated in the
Nivometer. The Nivometer also provided the DC supply voltage for the Preamplifier S-11.it house the transformer, DC power supply and the control unit.
Standardized signal of 4-20mA is available at the output of the Nivometer for
remote display. The material level is also indicated by the built in digital display. Potentiometers for adjusting 0% and the 100% level are accessible on removing the front panel.
Nivometer available in following two versions.
NIVOMETER MLC
Nivometer MLC 830/831/832 with Rod Probe
Probe Rod ~Ø12
Preamplifier Housing
Probe Rod ~Ø12Cable for SensorConnection
Preamplifier Housing
Pre-Amplifier1 2 3 4 5 6
Probe Hsg. Green Wire
MLC 830
S11Sr. No. O/F No.
NC
+
NC
InputInput
LEDNIVOMETER
%
7 Segement Display
CurrentOutput4-20 mA
Power SupplyRated Voltage 230V AC ORRated Voltage 24V DC
`` Remote- In Remote Version electronics and probe are interconnected using a cable supplied with the instrument.
MLC 830
3
Mounting Bush
Cable Gland
Insulation
Current Output4-20 mA
Power SupplyRated Voltage 230V AC ORRated Voltage 24V DC
Probe Rod ~Ø12
NIVOMETER
%
Housing Top View
LED
7 Segement Display
Electronics(Nivometer+Preamplifier)
NIVOMETER MLC
Nivometer MLC 830/831/832 with Rod Probe
Compact- In compact Version electronics and probe form a single mechanical entity.
4
MLC 830
The Nivometer MLC Available Models
1. Nivometer MLC 830 - Indicator:
Indicates the material level by built in 7 Segment Digital Display and gives 4-20mA output.
2. Nivometer MLC 831 - One Set Point: Functions same as MLC 830-Indicator and gives alarm if material level
above or below the selected alarm level depending upon selection of fail safe mode Maximum or Minimum respectively.
3. Nivometer MLC 832 - Pump Control: Functions same as MLC 830-Indicator along with an additional facility of
maintaining material level between two levels, low and high (Pump Control). It indicates level alarm condition only for one level, either low or high depending on selection of fail safe mode Maximum or Minimum.
Fail Safe Mode
Depending on the process requirement, Maximum fail safe mode or Minimum fail safe mode can be selected in Nivometer.
1. Maximum fail safe mode means that the relay de-energizes when the level exceeds the desired level (Probe Covered) or when mains supply fails.
2. Minimum fail safe means that the relay de-energizes when the level
drops below the desired level (Probe Uncovered) or mains supply fails. As relay de-energized condition is used for alarm condition, so apart from level alarm condition, even in case of instrument failure or power failure, the operator gets an alarm. This results in better reliability of operation.
Factors Influence Measurement Accuracy
1. The distance between the probe and the container wall should be
constant throughout the probe length. 2. High viscous material will have a tendency to stick to the probe. This will
influence to some extent to the 0% level setting.
MLC 830
5
3 Installation
Ensure that the mains supply is switched off during installation
Please refer figures in Appendix for some installation tips.
• Fully insulated Probe (rod or rope type) with mounting bush or flange is mounted in a container vertically from top.
• Probe should be firmly fixed on a vibration free foundation. • During installation rotate the mounting bush of the instrument not the
housing. • Select the mounting location such that during filling of the container
the material does not fall directly on the probe. Otherwise use protection shown in Appendix.
• For container having irregular shape or which are filled with the liquid having low dielectric constant, or strong agitator foam formation, rod probe with ground tube should be used or probe should be mounted on a standpipe.
• When probe is mounted in a concrete container, the metal rod grid in the RCC should be electrically connected to the ground terminal at the probe head.
• For non-conductive container, a counter electrode (300mm wide strip) made from conductive material should be fixed.
• With high bulk temperatures, there should be a heat shield between the vessel and the housing of the instrument.
Preamplifier S-11
The preamplifier is mounted in protective cast aluminium housing on the top of the probe by means of a single screw. The preamplifier consists of a semi-conductor dvice, which limits the working temperature at the Probe head about80˚C. In case, the temperature at the Probe head exceeds the above limit, thePreamplifier is mounted in a separate weather proof housing and connected tothe Probe
Nivometer MLC
Nivometer in Cast aluminium housing can be directly installed in the field. While installing the cable ducts provided with rubber seal should be always point downwards to prevent dust and moisture entry into the instrument. Also take care to protect the instrument from direct sunlight by providing sunshield. Special attention should be paid not to install Nivometer on support and walls, which are being continuously or periodically exposed to heavy vibrations. If this is unavoidable, install the instrument on the rubber cushioning to protect it from direct surges.
6
MLC 830
4 Wiring The following procedure should be carried out for wiring the instrument.
Rotate the top cover of the housing to loosen and remove it. Hold housing base while opening and closing the cover.
Unscrew the flap and move aside. Remove the display module Connect the instrument as per the wiring diagram give below.
Connection Diagram for Separate Version
Sensor S-11
56 4 3 2 1
7643 521+ - E
Mains Relay
119 108
- +
Sensor
Power Supply24V DC ±10%(24V DC Rated Voltage)
C NO NC 20mA
4mA
CurrentO/P
Sensor S-11
56 4 3 2 1
7643 521L N E
Mains Relay
119 108
- +
Sensor
Power Supply85 to 265V AC, 50 Hz(Rated Voltage 230V AC)
C NO NC 20mA
4mA
CurrentO/P
NIVOMETER MLC Terminals NIVOMETER MLC Terminals
Connection Diagram for Compact Version
7643 521L N E
Mains Relay
119 108
- +
Sensor
Power Supply85 to 265V AC, 50 Hz(Rated Voltage 230V AC)
C NO NC 20mA
4mA
CurrentO/P
Internally Connected(Factory Setting)
NIVOMETER MLC Terminals
7643 521+ - E
Mains Relay
119 108
- +
Sensor
Power Supply24V DC ±10%(Rated Voltage 24V DC )
C NO NC 20mA
4mA
CurrentO/P
Internally Connected(Factory Setting)
NIVOMETER MLC Terminals
Note- • In Compact Version Sensor S-11 internally connected to Nivometer. • Relay output is available only for Nivometer MLC-831/832.
MLC 830
7
5 Calibration
Ci C
Close
Open
Field SettingDIP Switch (For Range Selection C)Closing Switch : Range Closing Switch : Range Closing Switch : Range *(Closing Switch one at a time)
*
Field SettingDIP Switch (For Selecting Intial Capacitance C )Open Switch : Range Closing Switch : Range
i
Nivometer MLC 830 To calibrate the MLC 830 following procedure should be followed:
a) Connect the Preamplifier S-11 to the Nivometer terminals as per the
connection diagram b) Connect the mains power supply. c) Fill the vessel upto 0% level (i.e. the minimum level from which material is
to be measured.) d) Select the initial capacitance range І (Ci) by opening the DIP Switch І. e) Adjust 0% on display by turning the 0% potentiometer. f) If not adjustable, then select next initial capacitance range П by closing the
DIP Switch І. g) Repeat step (e). h) Fill the vessel upto 100% level (i.e. the maximum level upto which material
is to be measured.) i) Select the range І (∆C) by closing the DIP Switch П. . j) Adjust 100% on display by turning the 100% potentiometer. k) If 100% not adjusted then first open the DIP Switch П, then close DIP
Switch Ш for selecting range П, repeat step (j). Again if 100% not adjustable then first open DIP Switch Ш, then close DIP Switch ІV for selecting capacitance range ІV, repeat step (j)
l) Check the calibration once again by emptying the tank to 0% level and filling upto 100% level.
m) The output will vary corresponding to 0 %( 4mA) and 100% (20mA) level of the material being measured.
Nivometer MLC 831
To calibrate the MLC 831 following procedure should be followed:
a) Connect the Preamplifier S-11 to the Nivometer terminals as per the connection diagram.
b) Connect the mains supply. c) Select the required fail safe mode using jumper JP1. d) Keep the switch SW1 at “ALARM” position. Set the required alarm level
with the help of HL set POT, at this instance the display shows the alarm set value.
e) Fill the vessel upto 0% level (i.e. the minimum level from which the material is to be measured.)
f) Select the initial capacitance range І (Ci) by opening the DIP Switch І. g) Adjust 0% on display by turning 0% potentiometer. h) If 0% not adjustable then select the initial capacitance П (Ci) by closing
DIP Switch І. i) Repeat step (g.) j) Fill the vessel upto 100% level (i.e. the maximum level upto which the
material is to be measured.) k) Select the range І (∆C) by closing the DIP Switch П.
Close one DIP Switch at a time.
8
MLC 830
l) Adjust 100% on display by turning 100% potentiometer. m) If 100% not adjusted then first open the DIP Switch П, then close
DIP Switch Ш for selecting range П, repeat step (l). Again if 100% not adjustable then first open DIP Switch Ш, then close DIP Switch ІV for selecting capacitance range Ш, repeat step (l)
n) Check the calibration by once again emptying the tank to 0% and filling upto 100% level.
o) The output will vary corresponding to 0 %( 4mA) and 100% (20mA) level of the material being measured. Check alarm set value.
Close one DIP Switch at a time.
Switch SW1 should be at Normal position for normal Operation.
Nivometer MLC 832
To calibrate the MLC 832 following procedure should be followed:
a) Connect the Preamplifier S-11 to the Nivometer terminals as per the connection diagram.
b) Connect the mains supply. c) Select the required fail safe mode using jumper JP1. d) Rotate the LL potentiometer fully anticlockwise direction. e) Keep the Toggle Switch SW1 at “ALARM” position and set switch
SW2 to HL position .set the required high level alarm with the help of HL: set POT, at this instance the display shows high level alarm value.
f) Now toggle Switch SW2 to LL position and set the low level alarm by LL set POT, at this instance display shows low level alarm value.
g) Now toggle Switch SW1 to normal position and SW2 to HL position.
h) Fill the vessel upto 0% level (i.e. the minimum level from which material is to be measured.)
i) Select the initial capacitance range І (Ci) by opening the DIP Switch І.
j) Adjust 0% on display by turning the 0% potentiometer. k) If not adjustable, then select next initial capacitance range П by
closing the DIP Switch І. l) Repeat step (j). m) Fill the vessel upto 100% level (i.e. the maximum level upto which
material is to be measured.) n) Select the range І (∆C) by opening the DIP Switch П. o) Adjust 100% on display by turning the 100% potentiometer. p) If 100% not adjusted then first open the DIP Switch П, then close
DIP Switch Ш for selecting range П, repeat step (o). Again if 100% not adjustable then first open DIP Switch Ш, and then close DIP Switch ІV for selecting capacitance range Ш, repeat step (o).
Close one DIP Switch at a time.
q) Check the calibration by once again emptying the tank upto 0% level and filling upto 100% level. Check alarm set values.
r) The output will vary corresponding to 0% (4mA) and 100% (20mA) level of the material being measured..
Switch SW1 should be at Normal position for normal Operation.
The Instrument will show negative display if the level is below 0% or the level is above 199%.
MLC 830
9
6
Maintenance
Never attempt to loosen or remove the process connections or instrument housing while the process is under pressure.
7
The electronic part of the instrument needs no maintenance, as there are no moving parts. In case, the sensor requires cleaning due to deposited material, periodically clean the sensor taking care not to scratch or press the surface of the transducer. Ensure that the cable glands and the housing lid are sealed to prevent the ingress of dust and moisture.
Troubleshooting General Checks: • If the system does not function properly, then check that Nivometer
is properly connected. • There is true earth connection to the container and Preamplifier • Preamplifier is properly tightened in probe housing and free from any
deposition.
Checking of Probe
The Preamplifier must be removed from the Probe Head. When the Probe is uncovered an infinitely high ohmic resistance must be present between the earth of Probe head and Probe connection. Résistance value which can be measured indicate a faulty Probe.
Remove Preamplifier from the Probe head. Keep it connected electrically with Nivometer. Instead of a Probe, Connect a capacitance decade (10-4500 pF) between Probe connection and earth terminal (6) to stimulate material level. Operating Voltage
18V DC between Preamplifier terminal 3 (+) & 4(-) operating voltage too small or absent indicates
1) Nivometer Defective or 2) Current consumption of Preamplifier is
high
Checking of Preamplifier S-11
Current Consumption 5mA between Preamplifier terminal 3 and Nivometer Terminal 11.
Sensor S-11
56 4 3 2 1
7
NIVOMETER MLC Terminals
643 521L N E
Mains Relay
119 108
- +
Sensor
Power Supply85 to 265V AC, 50 Hz(Rated Voltage 230V AC)
C NO NC 20mA
4mA
CurrentO/P
10
MLC 830
Sensor S-11
56 4 3 2 1
7
NIVOMETER MLC Terminals
643 521+ - E
Mains Relay
119 108
- +
Sensor
Power Supply24V DC ±10%(24V DC Rated Voltage)
C NO NC 20mA
4mA
CurrentO/P
Hf Voltage
11Vpp /33 KHz between Probe connection point and ground terminal (6) of Preamplifier Hf voltage to small or absent indicates either Preamplifier is defective or Probe is defective.
Signal Current
Measure between Preamplifier Terminal (1) and Nivometer terminal (9) depend on capacitance. Signal current should vary linearly with change in capacitance value when capacitance changed between probe connection point and earth terminal (6) of Preamplifier
Checking of Nivometer
Measure the following voltages on Analog PCB. +18V±5% = At IC Z3’s Pin 8 W.r.t. Pin4.
+18V±5% = At IC Z5’s Pin 8 W.r.t. Pin4
-12V±20% = At IC Z10’s Pin 2 W.r.t. TP4
Measures the following Test point voltages with respect to GND (Test Point TP4) on Analog PCB.
+5V±5% = At TP1
-5V±5% = At TP2
+15V±20% = At TP3
+1±0.005V = At TP5 (when level is 100%)
0 ± 0.005V = At TP5 (when level is 00%)
Check level proportional current output at terminal 7 and terminal 8 of Analog PCB. When level is 100% current output is 20mA and when level is 00% current output is 4mA. Note- See Appendix for PCB Layouts.
MLC 830
11
8
Technical Data
Initial Capacitance
1) 30 to 65 pF 2) 60 to 280 pF
Capacitance Variation
1) 10 to 100 pF 2) 70 to 500 pF 3) 450 to 4500pF
Indication
1) 0 to 100% (On a 2-½ digit 7 segments LED Display.) 2) Bicolour LED
• Red LED for Alarm. (MLC-831/832) • Green LED for Normal. (MLC-831/832)
Output
1) 4-20 mA (Optically Isolated) with Max. Load resistance 600E. 2) One set of potential free changeover contact rated 3A at 230V AC for non inductive load. (Only for MLC-831/832)
System Accuracy ±2 %
Instrument Accuracy ±0.5 %
Relay Status 1) De-energized in Power Fail or Alarm Condition. 2) Energized in Normal Condition.
Fail Safe Mode
Maximum/Minimum (Field Adjustable) (Selectable for MLC-831/832)
Mains Supply
• 85-265VAC, 50Hz(Rated Voltage 230V AC) or • 24V DC ± 10% (Rated Voltage 24V DC)
Cable Entry M16 Cable Glands (3 Nos.) suitable for Dia. 4.5
mm to 10 mm. Option- ½ ” NPT 2Nos.
Power 7W (Max.)
Housing Material: Cast Aluminium Paint: Polyurethane 2K
Protection
IP67
Operating Temperature
0-60 °C
12
MLC 830
9 Operating Principle
The Capacitance measuring system works on the principle of change in capacitance (∆C). The capacitance of parallel plate capacitor is..
1) Directly proportional to the cross sectional area of plate ‘A’. 2) Inversely proportional to the distance ‘D’ between them. 3) Directly proportional to the dielectric constant of insulating material
Єr.
In Actual practice, Probe (Rod or Rope) and container wall form two plates of the capacitor. When the container is empty, the dielectric medium separating the two plates is air with dielectric constant approximately equal to 1. when the container is filled with the material , air displaced by the material having dielectric constant greater than 1,which thereby results in a increase of the capacitance. This increase of capacitance is directly proportional to change in material level provided:
1. Material’s dielectric constant remains unchanged. 2. Distance between Probe and container wall is same over the entire
measuring range. therefore, for a continuous level measurement, it is very important that
dielectric constant of the filling material does not change with time. Otherwise, error in measurement will take place.
Dielectric constant of the material can change, when there is a change in chemical composition of the liquid whose level is to be measured.
MLC 830
13
10 Appendix Probe Installation
Drawbacks 1. Probe mounted close to container wall. 2. Subject to sunlight.
Drawbacks: 1. Probe to close to container wall. 2. Subject to sunlight. 3. Material inflow directed on rope Probe.
• Weight will touch container wall. • Probe and insulation will get damaged soon.
14
MLC 830
Dimensional Drawing Remote Version of Nivometer MLC-830/831/832
Power SupplyRated Voltage 230V ACORRated Voltage 24V DC
Flange (optional)
Nivometer MLC 830/831/832 with Rod Probe
Ø84
112" BSP
93
Preamplifier Housing
Cable Gland
Mounting Bush
Insulation
Min
. 300
Max
. 350
0
±10 ±0
.5
±0.5
ControllerHousing
Cable Gland
Mounting BracketCarbon Steel
LEDNIVOMETER
%
7 Segement Display
155x155
116.
5
Probe
Ø6.5x4 Nos.
Controller
Ø84
112" BSP
93
Preamplifier Housing
Cable Gland
Mounting Bush
Insulation
Min
. 300
Max
. 350
0
±10 ±0
.5
±0.5
ControllerHousing
Cable Gland
Mounting BracketCarbon Steel
12
Probe Rod ~Ø12
25
Pro
be L
engt
h
Ø125Ø125
Input
105
116.5
CurrentOutput4-20 mA
12
Probe Rod ~Ø12
25
Pro
be L
engt
h
Ø125Ø125
Input
Cable for SensorConnection
105
116.5
155x155
116.
5
Probe
Ø6.5x4 Nos.
Controller
Note: All dimensions are in mm. unless otherwise specified.
Note: All dimensions are in mm. unless otherwise specified.
Ø6.5x4 Nos.
Controller
Ø84
112" BSP
93
Preamplifier Housing
Cable Gland
Mounting Bush
±0.5
±0.5
ControllerHousing
Cable Gland
Mounting BracketCarbon Steel
25
Ø125Ø125
Input
Cable for SensorConnection
105
116.5
155x155
116.
5
Probe
%
7 Segement Display
CurrentOutput4-20 mA
Power SupplyRated Voltage 230V ACORRated Voltage 24V DC
Flange (optional)
Nivometer MLC 830/831/832 with Rope Probe
Pro
be L
engt
h Wire Rope Ø3
Max
. 100
00
Min
. 300
Weight
Ø38
76
+20
25
Ø125Ø125
Input
105
116.5
155x155
116.
5
Probe
Ø6.5x4 Nos.
Controller
Ø84
112" BSP
93
Preamplifier Housing
Cable Gland
Mounting Bush
±0.5
±0.5
ControllerHousing
Cable Gland
Mounting BracketCarbon Steel
LEDNIVOMETER
MLC 830
15
Pro
be L
engt
h
112" BSP
Mounting Bush
Controller Housing25
Cable Gland
Min
. 100
Max
. 350
0
Insulation
155
Ø125
Current Output4-20 mA
Power SupplyRated Voltage 230V ACORRated Voltage 24V DC
±10
180
12
Probe Rod ~Ø12
NIVOMETER
%
Note: All dimensions are in mm. unless otherwise specified.
Housing Top View
Flange (Optional)
Nivometer MLC 830/831/832 with Rod Probe
LED
7 Segement Display
Current Output4-20 mA
Power SupplyRated Voltage 230V ACORRated Voltage 24V DC
NIVOMETER
%
Housing Top View
LED
7 Segement Display
112" BSP
Mounting Bush
Controller Housing
25
Cable Gland
Ø125
180
Pro
be L
engt
h Wire Rope Ø3
Max
. 100
00
Min
. 300
Weight
Ø38
76
+20
Flange (optional)
Note: All dimensions are in mm. unless otherwise specified.
Nivometer MLC 830/831/832 with Rope Probe
155
Dimensional Drawing Compact Version of Nivometer MLC-830/831/832
16
MLC 830
PCB Layouts Component Layout of MLC- 830/831/832
F1
CN1
CN2
Power Supply PCB
OPEN
1 2 3 4 Ci C
MIN MAX
VR7
DIPSW
LED
NORMAL
ALARM
SW1 SW2 HL
LL
VR1VR6
VR2VR5
0%ADJ
HLSET
LLSET100%
ADJ
JP1
*
Display PCB Note-Don’t disturb setting of * marked components.
MLC 830
17
Component Layout of MLC 830 Analog PCB
CN6 CN7
E N L
NCNOCENL +4mA
20mAE-+
CN4
Analog PCB
Z10
VR8
TP4
TP2
VR4
VR3*
*
*
TP3
Z5
Z3
TP1
VR9*TP5
TP6
Note-Don’t disturb setting of * marked components.
18
MLC 830
Warranty Standard guarantee is for twelve months from the date of commissioning or eighteen months from the date of dispatch which ever is earlier. Guarantee term is strictly for any manufacturing fault i.e. any defect in material or workmanship. We undertake to make good the defect at our own expense, provided notice is given to us as soon as it is discovered and the instrument is forwarded to our factory insured and freight paid and with seals unbroken. Our responsibility is limited to the cost of removing the defect in the instrument itself.
Customer Service
Nivo Controls Pvt. Ltd. 104 – 115 Electronic Complex Pardesipura Indore – 452010 India Phone: +91 731 2553248 / 49 Fax: +91 731 2550075 Url: www.nivocontrols.com Email: [email protected]