Comparison of VMAT and IMRT strategies for cervical cancer patients using automated planning

7
Original article Comparison of VMAT and IMRT strategies for cervical cancer patients using automated planning Abdul Wahab M. Sharfo , Peter W.J. Voet, Sebastiaan Breedveld, Jan Willem M. Mens, Mischa S. Hoogeman, Ben J.M. Heijmen Department of Radiation Oncology, Erasmus MC-Cancer Institute, Rotterdam, The Netherlands article info Article history: Received 5 September 2014 Received in revised form 4 February 2015 Accepted 9 February 2015 Available online xxxx Keywords: VMAT IMRT Locally advanced cervical cancer Automated knowledge-based planning abstract Background and purpose: In a published study on cervical cancer, 5-beam IMRT was inferior to single arc VMAT. Here we compare 9, 12, and 20 beam IMRT with single and dual arc VMAT. Material and methods: For each of 10 patients, automated plan generation with the in-house Erasmus- iCycle optimizer was used to assist an expert planner in generating the five plans with the clinical TPS. Results: For each patient, all plans were clinically acceptable with a high and similar PTV coverage. OAR sparing increased when going from 9 to 12 to 20 IMRT beams, and from single to dual arc VMAT. For all patients, 12 and 20 beam IMRT were superior to single and dual arc VMAT, with substantial variations in gain among the study patients. As expected, delivery of VMAT plans was significantly faster than delivery of IMRT plans. Conclusions: Often reported increased plan quality for VMAT compared to IMRT has not been observed for cervical cancer. Twenty and 12 beam IMRT plans had a higher quality than single and dual arc VMAT. For individual patients, the optimal delivery technique depends on a complex trade-off between plan quality and treatment time that may change with introduction of faster delivery systems. Ó 2015 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology xxx (2015) xxx–xxx Introduction Volumetric modulated arc therapy (VMAT) and intensity modulated radiotherapy with static gantry angles (IMRT) have been compared for various tumour sites, both regarding plan qual- ity and treatment time [1]. In most studies, VMAT was a new tech- nology and plans were compared to previously delivered clinical IMRT plans. Systematic comparisons with variations in the number of IMRT beams and VMAT arcs are scarce [2–4]. Several studies on advanced cervical cancer have demonstrated reduced OAR doses with IMRT compared to 3D conformal radiotherapy [5–9]. To the best of our knowledge, only Cozzi et al. [10] have compared VMAT and IMRT for cervical cancer patients. Recently, we have developed Erasmus-iCycle, an optimizer for automated, multi-criterial beam profile optimization and beam angle selection for coplanar and non-coplanar IMRT [11–17]. In Erasmus-iCycle, the common manual, trial-and-error tweaking of plan parameters by dosimetrists is replaced by a fully automated procedure, based on lexicographic optimization. The automation allows objective comparison of treatment strategies, e.g. with vari- able numbers of IMRT beams, or considering non-coplanar vs. coplanar IMRT [14–16]. Currently, Erasmus-iCycle optimizes beam fluences. For generation of clinical plans, Erasmus-iCycle plans are reconstructed and segmented in the clinical treatment planning system (TPS). In a recent prospective study, we have demonstrated superior plan quality of Erasmus-iCycle plans that were ‘manually’ reconstructed in the clinical TPS (semi-automatic plan generation), compared to the clinical routine of trial-and-error treatment plan- ning. In 97% of cases, the treating physician selected the semi-au- tomatically generated plan for treatment, rather than the IMRT plan generated manually by a dosimetrist without prior knowledge of the Erasmus-iCycle result [17]. Apart from the improved plan quality, the semi-automated procedure also reduced the planning hands-on time by 50% compared to manual planning. Cozzi et al. [10] have retrospectively compared single arc VMAT plans with previously delivered 5-field equi-angular IMRT plans for treatment of cervical cancer patients. In this study, we have sys- tematically compared single and dual arc VMAT plans with IMRT plans with 9, 12 and 20 beams. All these plans were generated semi-automatically (above). Apart from the mutual comparisons, the semi-automatic plans were also compared to clinically delivered 9-beam IMRT plans. http://dx.doi.org/10.1016/j.radonc.2015.02.006 0167-8140/Ó 2015 Elsevier Ireland Ltd. All rights reserved. Corresponding author. E-mail address: [email protected] (A.W.M. Sharfo). Radiotherapy and Oncology xxx (2015) xxx–xxx Contents lists available at ScienceDirect Radiotherapy and Oncology journal homepage: www.thegreenjournal.com Please cite this article in press as: Sharfo AWM et al. Comparison of VMAT and IMRT strategies for cervical cancer patients using automated planning. Radiother Oncol (2015), http://dx.doi.org/10.1016/j.radonc.2015.02.006

Transcript of Comparison of VMAT and IMRT strategies for cervical cancer patients using automated planning

Radiotherapy and Oncology xxx (2015) xxx–xxx

Contents lists available at ScienceDirect

Radiotherapy and Oncology

journal homepage: www.thegreenjournal .com

Original article

Comparison of VMAT and IMRT strategies for cervical cancer patientsusing automated planning

http://dx.doi.org/10.1016/j.radonc.2015.02.0060167-8140/� 2015 Elsevier Ireland Ltd. All rights reserved.

⇑ Corresponding author.E-mail address: [email protected] (A.W.M. Sharfo).

Please cite this article in press as: Sharfo AWM et al. Comparison of VMAT and IMRT strategies for cervical cancer patients using automated plRadiother Oncol (2015), http://dx.doi.org/10.1016/j.radonc.2015.02.006

Abdul Wahab M. Sharfo ⇑, Peter W.J. Voet, Sebastiaan Breedveld, Jan Willem M. Mens,Mischa S. Hoogeman, Ben J.M. HeijmenDepartment of Radiation Oncology, Erasmus MC-Cancer Institute, Rotterdam, The Netherlands

a r t i c l e i n f o a b s t r a c t

Article history:Received 5 September 2014Received in revised form 4 February 2015Accepted 9 February 2015Available online xxxx

Keywords:VMATIMRTLocally advanced cervical cancerAutomated knowledge-based planning

Background and purpose: In a published study on cervical cancer, 5-beam IMRT was inferior to single arcVMAT. Here we compare 9, 12, and 20 beam IMRT with single and dual arc VMAT.Material and methods: For each of 10 patients, automated plan generation with the in-house Erasmus-iCycle optimizer was used to assist an expert planner in generating the five plans with the clinical TPS.Results: For each patient, all plans were clinically acceptable with a high and similar PTV coverage. OARsparing increased when going from 9 to 12 to 20 IMRT beams, and from single to dual arc VMAT. For allpatients, 12 and 20 beam IMRT were superior to single and dual arc VMAT, with substantial variations ingain among the study patients. As expected, delivery of VMAT plans was significantly faster than deliveryof IMRT plans.Conclusions: Often reported increased plan quality for VMAT compared to IMRT has not been observedfor cervical cancer. Twenty and 12 beam IMRT plans had a higher quality than single and dual arc VMAT.For individual patients, the optimal delivery technique depends on a complex trade-off between planquality and treatment time that may change with introduction of faster delivery systems.

� 2015 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology xxx (2015) xxx–xxx

Introduction

Volumetric modulated arc therapy (VMAT) and intensitymodulated radiotherapy with static gantry angles (IMRT) havebeen compared for various tumour sites, both regarding plan qual-ity and treatment time [1]. In most studies, VMAT was a new tech-nology and plans were compared to previously delivered clinicalIMRT plans. Systematic comparisons with variations in the numberof IMRT beams and VMAT arcs are scarce [2–4]. Several studies onadvanced cervical cancer have demonstrated reduced OAR doseswith IMRT compared to 3D conformal radiotherapy [5–9]. To thebest of our knowledge, only Cozzi et al. [10] have compared VMATand IMRT for cervical cancer patients.

Recently, we have developed Erasmus-iCycle, an optimizer forautomated, multi-criterial beam profile optimization and beamangle selection for coplanar and non-coplanar IMRT [11–17]. InErasmus-iCycle, the common manual, trial-and-error tweaking ofplan parameters by dosimetrists is replaced by a fully automatedprocedure, based on lexicographic optimization. The automation

allows objective comparison of treatment strategies, e.g. with vari-able numbers of IMRT beams, or considering non-coplanar vs.coplanar IMRT [14–16]. Currently, Erasmus-iCycle optimizes beamfluences. For generation of clinical plans, Erasmus-iCycle plans arereconstructed and segmented in the clinical treatment planningsystem (TPS). In a recent prospective study, we have demonstratedsuperior plan quality of Erasmus-iCycle plans that were ‘manually’reconstructed in the clinical TPS (semi-automatic plan generation),compared to the clinical routine of trial-and-error treatment plan-ning. In 97% of cases, the treating physician selected the semi-au-tomatically generated plan for treatment, rather than the IMRTplan generated manually by a dosimetrist without prior knowledgeof the Erasmus-iCycle result [17]. Apart from the improved planquality, the semi-automated procedure also reduced the planninghands-on time by 50% compared to manual planning.

Cozzi et al. [10] have retrospectively compared single arc VMATplans with previously delivered 5-field equi-angular IMRT plans fortreatment of cervical cancer patients. In this study, we have sys-tematically compared single and dual arc VMAT plans with IMRTplans with 9, 12 and 20 beams. All these plans were generatedsemi-automatically (above). Apart from the mutual comparisons,the semi-automatic plans were also compared to clinicallydelivered 9-beam IMRT plans.

anning.

DJ
高亮
权衡
DJ
高亮
系统比较固定野调强射野数目和VMAT弧个数这种文献相对较少
DJ
高亮
DJ
高亮
减少计划制作时间?

Table 1Applied wish-list for all study patients.

Constraints

Volume Type Limit

PTV Max 105% ofDp

PTV shell1.5 cm

Max 75% ofDp

PTV shell2.5 cm

Max 65% ofDp

PTV shell4 cm

Max 50% ofDp

UnspecifiedTissues

Max 105% ofDp

Objectives

Level Priority Volume Type Goal Parameters

1 PTV ; LTCP 1 Dp = 46 Gy,a = 0.75

2 Skin ring 2 cm ; Max 23 Gy

3 Small bowel ; Mean 40 Gy1 4 Rectum ; Mean 40 Gy

5 Bladder ; Mean 40 Gy

6 Small bowel ; Mean 20 Gy2 7 Rectum ; Mean 20 Gy

8 Bladder ; Mean 20 Gy

9 Small bowel ; Mean 10 Gy3 10 Rectum ; Mean 10 Gy

11 Bladder ; Mean 10 Gy

12 Unspecifiedtissues

; Mean –

2 Comparison of VMAT and IMRT for cervical cancer

Materials and methods

Patients and clinical treatment plans

Planning CT-scans of 10 randomly selected, previously treatedcervical cancer patients were included in this study. All scanshad delineated small bowel (SB), rectum, bladder, and a planningtarget volume (PTV). To establish the PTV, an internal target vol-ume (ITV) was first constructed using a pretreatment establishedmotion model from a full and an empty bladder CT-scan, whichwas then uniformly expanded by a 1 cm margin [18]. Patients weretreated in prone position lying on a small bowel displacement sys-tem (belly board), and using step-and-shoot 9 field IMRT with 10-MV photons at a Synergy linear accelerator (Elekta AB, Sweden).The linacs were equipped with the MLCi2 multi-leaf collimator(MLC) and the dose rate was 600 MU/min. The prescription dosewas 46 Gy delivered in 23 fractions. As defined by the clinical pro-tocol, all plans were created to deliver P95% of the prescribed doseto a minimum of 99.5% of the PTV, and no more than 0.2% of thePTV should receive P110%. Cervical cancer target volumes are gen-erally large. For this reason we did not use commonly applied D98%

and D2% parameters for these patients to evaluate low and high PTVdoses, respectively. The main planning goal was to create highlyconformal plans providing the best sparing of the delineated OARswith a higher concern for the SB which has the most severe com-plications [19]. All plans were generated with our clinical TPS(Monaco version 3.3, Elekta AB, Sweden).

Systematic comparisons of planning strategies

For each patient, 5 plans were semi-automatically generated ina two-step process. In the first step, Erasmus-iCycle ([13], andbelow) was used for fully automatic generation of a high quality20-beam equi-angular IMRT plan. Using the Erasmus-iCycle resultsfor guidance, an expert planner (Abdul Wahab M. Sharfo) thenmanually designed 5 plans with the clinical TPS; single arc VMAT(VMAT), dual arc VMAT (2VMAT), and IMRT plans with 9 (9DMLC),12 (12DMLC) and 20 (20DMLC) equi-angular beams, deliveredwith dynamic MLC. To this purpose, the Erasmus-iCycle outputwas used to generate a patient specific plan prescription for Mona-co [20]. For all plans, the goal was to maximally approach (orsupersede) the corresponding Erasmus-iCycle plan results, consid-ering the established hard constraints and priorities for the variousobjectives (details below). In order to avoid bias in planningresults, plan generations for individual patients were spread outin time. Moreover, for each patient, the five plans were generatedin an arbitrary order, and consecutive generation of plans for a sin-gle patient was avoided. The expert stopped working on the plansif plan quality levelled off and no significant further improvementwas expected; there was no time limit. As in clinical practice, semi-automatic plans were generated for 10-MV beams, the MLCi2 mul-ti-leaf collimator, and a maximum dose rate of 600 MU/min.

Generated plans were visually inspected by the clinician par-ticipating in this study (Jan Willem M. Mens) to assess accept-ability for clinical use. For small bowel, the volumes receiving>15 Gy and >45 Gy, V15Gy and V45Gy [21], and the mean dose, Dmean,were used for quantitative plan comparisons. For rectum and blad-der, Dmean was used. Paired two-sided Wilcoxon signed-rank testswere performed to compare the different planning strategies. Dif-ferences were considered statistically significant for p < 0.05.

For a subset of 5 patients, all 5 semi-automatically generatedplans were actually delivered at one of our treatment units in theabsence of the patient to record treatment delivery times.

Automatic plan generation with Erasmus-iCycle

Erasmus-iCycle has been described in detail in [13]. In thisstudy, the optimizer was used for fluence profile optimization,

Please cite this article in press as: Sharfo AWM et al. Comparison of VMAT anRadiother Oncol (2015), http://dx.doi.org/10.1016/j.radonc.2015.02.006

using a fixed, equi-angular 20-beam setup (above). Plan optimiza-tion with Erasmus-iCycle is based on an a priori defined ‘wish-list’with hard constraints, and prioritized objectives with goal values[11,12]. Each treatment site has a fixed wish-list. In several studies,we have observed superior plan quality with this approach com-pared to ‘manual’ trial-and-error tweaking of TPS parameters bydosimetrists in a clinical setting [14,16].

The cervical cancer wish-list as used in this study is presentedin Table 1. The constraints for shells around the PTV at 1.5, 2.5and 4 cm were defined to ensure a steep dose fall-off outside thePTV, as well as a conformal dose distribution. Among the objec-tives, PTV coverage had the highest priority. To achieve adequatecoverage, the Logarithmic Tumour Control Probability (LTCP) wasused as objective function [22]. A skin ring of 2 cm wide, extendingfrom the body contour towards the PTV, was defined to control theentrance dose (priority 2). In line with current clinical practice,sparing of SB had a higher priority than reducing delivered doseto the rectum and bladder. Reduction of the mean SB, rectumand bladder doses was performed using a multi-level approach[13], i.e., by repeated use of the objective functions with decreasingpriorities and goal values (Table 1).

Results

Semi-automatic plan generation

All generated plans were clinically acceptable. Based onachieved plan parameters of an Erasmus-iCycle 20-beam IMRTplan, the planner succeeded in all cases to generate a clinicallydeliverable 20DMLC Monaco plan with highly similar plan quality.As in the clinical plans, in all semi-automatically generated plans,at least 99.5% of the PTV received 95% or more of the prescribeddose, while the volume receiving >110% was below 0.2%. For thewhole population and all plans, the minimum dose in the PTV-1 cc was 92.2% ± 1.1% of the prescribed dose. For each patient, dif-ferences in PTV coverage and PTV mean and minimum doses

d IMRT strategies for cervical cancer patients using automated planning.

DJ
高亮
DJ
高亮
DJ
高亮
DJ
高亮
DJ
高亮
靶区外扩不同距离限值剂量分布

A.W.M. Sharfo et al. / Radiotherapy and Oncology xxx (2015) xxx–xxx 3

among the various plans were insignificant. Because of the highsimilarities in PTV doses in all plans for each of the patients, inthe remainder of this section, plan quality comparisons are basedon differences in OAR dose delivery. Fig. 1 shows for the includedpatient 8, comparisons between 20DMLC, 2VMAT and clinical dosedistributions in axial planes. Conformality is clearly highest for20DMLC and lowest for the clinical plan.

Semi-automatic vs. clinical plan quality

The last row of Table 2 demonstrates that for all semi-automaticIMRT and VMAT procedures (20DMLC, 12DMLC, 9DMLC, 2VMAT,VMAT), the 5 OAR plan parameters were on average lower thanthe corresponding parameters in the clinical plans. For the 3 SBplan parameters, reductions in the semi-automatic procedureswere always statistically significant (last column of Table 2). Forthe (lower priority) rectum and bladder mean doses, reductionswere statistically significant for 20DMLC, 12DMLC, 9DMLC, and2VMAT, but not for VMAT. The last column in Table 2 shows thatfor each semi-automatic procedure, the majority of patients hadlower OAR parameter values than in the clinical plan. Superiorityof the semi-automatically generated plans over the clinicallyapplied plans is confirmed in Fig. 2.

Semi-automatic dual arc vs. single arc VMAT plan quality

On average, semi-automatic dual arc VMAT planning resulted inreduced OAR dose delivery compared to single arc (Fig. 2 andTable 2). Each of the 5 evaluated OAR plan parameters was lowestwith 2VMAT for 50% or more of patients (Table 2). For SB V45Gy andSB Dmean, the differences between 2VMAT and VMAT were statisti-cally significant (Table 2). While both for 2VMAT and VMAT, SBV45Gy, SB V15Gy, and SB Dmean were statistically significantly lowerthan in the clinical plan, significant reductions for rectum andbladder Dmean were only observed for dual arc planning (Table 2).

Semi-automatic 20DMLC vs. 12DMLC vs. 9DMLC plan quality

Fig. 2 and Table 2 show that all mean OAR parameter valuesreduce when going from 9 to 12 beams and from 12 to 20. Bothbeam additions result in a benefit for the majority of patients(Table 2).

Semi-automatic IMRT vs. VMAT plan quality

Fig. 2 and Table 2 demonstrate that all semi-automatic IMRTstrategies (20DMLC, 12DMLC, and 9DMLC) have better OAR spar-ing than single arc VMAT plans. The differences are also mostly sta-tistically significant (Table 2). For all OAR parameters, the majorityof patients also benefitted from 20DMLC and 12DMLC compared todual arc VMAT (2VMAT) (Table 2). Figs. 3 and 4 provide for eachpatient separately, detailed comparisons between 20DMLC and2VMAT, and the clinical plan (Fig. 3 only). Both figures underlinethe overall superiority of semi-automatic 20DMLC plan qualitycompared to semi-automatic 2VMAT. There are however substan-tial variations among patients in the differences between the twostrategies. For patients 2, 8 and 9 these differences are relativelysmall, while patients 3–6 and 10 show higher gains of 20DMLC.Treatment times of 20DMLC plans are clearly enhanced comparedto 2VMAT plans. In the Discussion section, improved plan quality isdiscussed in the context of treatment time enhancement.

Planning times

Fully automated plan generation with Erasmus-iCycle togenerate input for the manual planning in the clinical TPS took

Please cite this article in press as: Sharfo AWM et al. Comparison of VMAT anRadiother Oncol (2015), http://dx.doi.org/10.1016/j.radonc.2015.02.006

on average 4.5 h per patient and was done overnight (no workload).The average hands-on planning time for generating a plan in theclinical TPS, based on Erasmus-iCycle output, was 1.5 h with aminimum of 0.5 h and a maximum of 3 h; VMAT hands-on plan-ning times were almost a factor of 2 larger than that for DMLC.The average hands-on planning time spent on generation of clinicalplans was 3 h with a range between 2 and 5 h, depending on thecomplexity of the case. The mentioned times for plan generationin the clinical TPS include times for Monte Carlo dose calculation.

Treatment delivery times

For the studied subset of 5 patients, the recorded mean treat-ment delivery times for VMAT, 2VMAT, 9DMLC, 12DMLC and20DMLC were 3.0 [2.9,3.2], 4.1 [3.8,4.5], 10.5 [9.1,12], 12.8[10.3,15.2], and 15.6 min. [14,17.6], respectively.

Discussion

VMAT and IMRT plans have been compared in many studies andfor various tumour sites [1]. In most studies, VMAT was a noveltreatment technique, compared to previously delivered IMRT.Often, VMAT and IMRT plans were generated by different planners,and possibly under different conditions (e.g. available planningtime). In this study, we have systematically compared single anddual arc VMAT, and IMRT with 9, 12, and 20 beams for treatmentof locally advanced cervical cancer patients. All plans were gener-ated in a semi-automatic procedure. For each patient, the basis forthe production of the five semi-automatic plans was a 20-beamequi-angular IMRT plan, automatically generated with the in-house Erasmus-iCycle. The output (i.e. achieved plan parameters)of this Erasmus-iCycle plan was used as a guide for manual con-struction of the two VMAT and three IMRT plans in the clinicalTPS. To the best of our knowledge this is the first comprehensivestudy presenting for a single treatment site a comparison ofclinically deliverable plans (i.e. generated with a clinical TPS) withvariation of both the number of IMRT beams and the number ofVMAT arcs. Previous studies by Bortfeld [2], and Quan et al. [3]were based on plans generated with non-clinical optimizationalgorithms. Guckenberger et al. [4] investigated VMAT with multi-ple arcs for complex targets compared to one clinical IMRT plan.

Instead of generating a 20-beam equi-angular IMRT plan, Eras-mus-iCycle could also have been used to generate for each patienta 20-beam plan with an individualized beam setup. However, astudy for a subset of patients did not show significant planimprovements for individualized beam arrangements, while calcu-lation times went up. The same study showed no significantimprovements for plans with more than 20 beams. The presentedhands-on planning times in the Results section demonstrate thatconstruction of IMRT and VMAT plans in the clinical TPS, basedon an automatically generated 20-beam Erasmus-iCycle plan, is atime-efficient method that sets the workload to minimal by theautomatic generation of Monaco cost functions.

In a previous prospective clinical study on head-and-neck can-cer we have demonstrated clear superiority of the semi-automaticprocedure over the general trial-and-error treatment planning bydosimetrists [16]. In this study, we have confirmed the superiorityof semi-automatic planning for cervical cancer patients in a retro-spective comparison; for each patient, all five semi-automaticallygenerated plans had better OAR sparing than the clinically applied9-beam IMRT plan. Plan quality increased when the number ofbeams (IMRT) or arcs (VMAT) went up. This can be attributed toincreases in the degrees of freedom for generating high qualityplans. Interestingly, 12DMLC and 20DMLC plans had better OARsparing than dual and single arc VMAT. Detailed analyses demon-

d IMRT strategies for cervical cancer patients using automated planning.

Table 2Pair-wise comparisons of planning strategies.

In each comparison (table cell), going from top to bottom, data refer to (a) SB V45Gy, (b) SB V15Gy, (c) SB Dmean, (d) Bladder Dmean, and (e) Rectum Dmean. Below the tablediagonal, the ‘‘A/B’’ in the cells refer to A: plan parameter value for the strategy along the vertical axis minus the parameter value for the strategy along the horizontal axis,averaged over the 10 patients in the study, B: average of the patient-mean OAR parameter values in the two compared strategies. Cells above the diagonal show p values and‘‘(n|m)’’: for n patients, the strategy indicated at the horizontal axis has lowest OAR dose, while for m patients the strategy mentioned at the vertical axis is superior. NS: nostatistically significant difference.

*

* **

0

5

10

15

20

25

30

35

40

Perc

enta

ge D

iffer

ence

(%)

Planning Technique

SB V15GySB V45Gy SB Dmean

Bladder Dmean Rectum Dmean

20DMLC 12DMLC 9DMLC 2VMAT VMAT ClinicalTAMVCLMD

20DMLC 12DMLC 9DMLCDMLC

2VMAT VMATVMAT

Fig. 2. For the semi-automatic planning strategies 12DMLC, 9DMLC, 2VMAT and VMAT and the clinical plans, percentage increases of mean OAR plan parameter valuesrelative to the semi-automatic 20DMLC. Average absolute values for 20DMLC: SB V45Gy = 102 cc, SB V15Gy = 345 cc, SB Dmean = 20.6 Gy, Rectum Dmean = 42 Gy and BladderDmean = 42.7 Gy. ⁄Difference from 20DMLC not statistically significant.

4 Comparison of VMAT and IMRT for cervical cancer

strated that there were substantial variations among patients inthe OAR sparing advantage of 20DMLC compared to 2VMAT. Possi-bly, the lower number of directions with IMRT compared to VMAT

Please cite this article in press as: Sharfo AWM et al. Comparison of VMAT anRadiother Oncol (2015), http://dx.doi.org/10.1016/j.radonc.2015.02.006

is compensated by an increase in beam modulation. To our bestknowledge, only Cozzi et al. [10] have previously compared VMATand IMRT plans for treatment of cervical cancer. The involved

d IMRT strategies for cervical cancer patients using automated planning.

0

20

40

60

80

100

0 20 40 60 80 100 120

Volu

me

(%)

Dose (%)

0

20

40

60

80

100

0 20 40 60 80 100 120

Volu

me

(%)

Dose (%)

0

20

40

60

80

100

0 20 40 60 80 100 120

Volu

me

(%)

Dose (%)

0

20

40

60

80

100

0 20 40 60 80 100 120

Volu

me

(%)

Dose (%)

0

20

40

60

80

100

0 20 40 60 80 100 120

Volu

me

(%)

Dose (%)

0

20

40

60

80

100

0 20 40 60 80 100 120

Volu

me

(%)

Dose (%)

0

20

40

60

80

100

0 20 40 60 80 100 120Vo

lum

e (%

)Dose (%)

0

20

40

60

80

100

0 20 40 60 80 100 120

Volu

me

(%)

Dose (%)

0

20

40

60

80

100

0 20 40 60 80 100 120

Volu

me

(%)

Dose (%)

0

20

40

60

80

100

0 20 40 60 80 100 120

Volu

me

(%)

Dose (%)

Clinical 20DMLC 2VMATPTV Small Bowel Bladder Rectum

Fig. 3. DVHs for PTV (red), SB (blue), bladder (green) and rectum (orange) for all study patients for clinical (solid), 20DMLC (dashed), and 2VMAT (dotted) plans.

A.W.M. Sharfo et al. / Radiotherapy and Oncology xxx (2015) xxx–xxx 5

clinically delivered IMRT plans had 5 equi-angularly spaced beams.For VMAT, only single arc dose delivery was considered. Theyfound superior plan quality with VMAT compared to IMRT. VMATimproved dose homogeneity and sparing of OARs in the mediumto high dose region. For a subset of patients, we also generated

Please cite this article in press as: Sharfo AWM et al. Comparison of VMAT anRadiother Oncol (2015), http://dx.doi.org/10.1016/j.radonc.2015.02.006

5-beam IMRT plans and also observed clearly inferior plan qualitycompared to both single and dual arc VMAT (data not provided),confirming the findings of Cozzi et al. [10].

In a theoretical study, Bortfeld and Webb [23] have discussedplan quality limitations of single arc VMAT with very short delivery

d IMRT strategies for cervical cancer patients using automated planning.

0

20

40

60

80

100

V45Gy

V15Gy

SB Dmean

Bladder Dmean

Rectum Dmean

P1 P2 P3 P4 P5 P6 P7 P8 P9 P10

Rectum Dmean

Bladder Dmean

SB Dmean

V15Gy

V45Gy

(2VMAT) (20DMLC)

OA

Rs P

lan

Para

met

er (G

y, %

)

Study Patient

Fig. 4. For each of the 10 patients in this study (P1-P10 along the horizontal axis), OAR plan parameters are presented for dual-arc VMAT (2VMAT) and 20-beam IMRT(20DMLC).

6 Comparison of VMAT and IMRT for cervical cancer

times. The current study also finds limitations of VMAT, not onlyfor single arc VMAT, but also for dual arc. Moreover, the evidenceprovided in our paper is all obtained from clinically deliverabledose distributions, approved by a physician. In contrast, publishedplanning studies often report better or similar plan quality for sin-gle or dual arc VMAT compared to IMRT, or advantages for PTV anddisadvantages for OARs, or vice versa [1].

In line with overwhelming evidence in the literature, we havealso observed substantial treatment delivery time reductions withVMAT compared to IMRT. Compared to 20DMLC and 12DMLC, dosedelivery times with 2VMAT reduced on average by 11.5 and8.7 min, respectively. These studies were performed with theMLCi2 multi-leaf collimator. Recently, the Agility MLC (Elekta AB)with 160 leaves with very fast motion has shown to reduce VMATdelivery times by 40% when compared to MLCi2 with equivalentplan quality and delivery accuracy [24]. For a subset of 5 patientswe observed mean reductions in the predicted treatment deliverytimes when using the Agility MLC instead of MLCi2 of 1.5 min/42.9% for single arc VMAT and 2.1 min/42% for dual arc VMAT,and 6.5 min/42.5%, 5.7 min/43.9%, and 7.2 min/45%, for IMRT with9, 12, and 20 beams, respectively. Treatment time reductions havea potential clinical impact in terms of reducing the risk of intra-fraction motion and enhancing patient comfort during treatment.Moreover, short treatment times can increase patient throughputin the clinic and eventually reduce waiting lists and costs. On theother hand, in this paper we have observed for some cervical can-cer patients clearly enhanced plan quality for IMRT compared toVMAT, especially when 20 IMRT beams are applied. Based on theresults of this study, the clinical decision has been made to starttreating cervical cancer patients with dual arc VMAT. This decisionwill be re-considered when the fast Agility MLCs will become avail-able in our department. We are currently developing a system forfully automatic generation of a set of IMRT and VMAT plans foreach individual patient to evaluate on a per patient basis plan qual-ity differences versus delivery time differences. These studies willinclude modelling of intra-fraction motion.

In this study, we have investigated IMRT delivered with dynam-ic multi-leaf collimation. From a theoretical point of view, thisapproach could result in a higher plan quality than step-and-shootIMRT because of the increased degree of freedom. It could howeveralso result in somewhat longer treatment times. In the end this all

Please cite this article in press as: Sharfo AWM et al. Comparison of VMAT anRadiother Oncol (2015), http://dx.doi.org/10.1016/j.radonc.2015.02.006

depends on implementation in the (commercial) TPS software.Here we have chosen for the dynamic IMRT technique as we main-ly focused on plan quality.

A potential limitation of this study is that plans were generatedsemi-automatically, possibly resulting in some ‘noise’ in ourresults. In a recent study, we have developed a system for fullyautomatic VMAT plan generation for prostate cancer patients,based on an Erasmus-iCycle pre-optimization [20]. Currently,efforts are on-going to also realize this for cervical cancer patientsand to extend the method for static gantry IMRT.

Conclusions

For locally advanced cervical cancer patients we have system-atically compared single and dual arc VMAT with 9, 12, and 20beam IMRT. IMRT plan quality increased when going from 9 to12 to 20 beams, and dual arc VMAT was superior to single arc.For all patients, the 20-beam IMRT plan had the overall highestquality. On the other hand, the VMAT plans had lowest treatmentdelivery times. For individual patients, the optimal delivery tech-nique is dependent on a complex trade-off between plan qualityand treatment time.

Conflict of interest

This work was in part funded by a research grant of Elekta AB,Stockholm, Sweden. Erasmus MC Cancer Institute also has aresearch collaboration with Accuray Inc., Sunnyvale, USA.

Appendix A. Supplementary data

Supplementary data associated with this article can be found, inthe online version, at http://dx.doi.org/10.1016/j.radonc.2015.02.006.

References

[1] Teoh M, Clark CH, Wood K, Whitaker S, Nisbet A. Volumetric modulated arctherapy: a review of current literature and clinical use in practice. Br J Radiol2011;84:967–96.

d IMRT strategies for cervical cancer patients using automated planning.

A.W.M. Sharfo et al. / Radiotherapy and Oncology xxx (2015) xxx–xxx 7

[2] Bortfeld T. The number of beams in IMRT-theoretical investigations andimplications for single-arc IMRT. Phys Med Biol 2010;55:83–97.

[3] Quan EM, Li X, Li Y, et al. A comprehensive comparison of IMRT and VMAT planquality for prostate cancer treatment. Int J Radiat Oncol Biol Phys2012;83:1169–78.

[4] Guckenberger M, Richter A, Krieger T, Wilbert J, Baier K, Flentje M. Is a singlearc sufficient in volumetric-modulated arc therapy (VMAT) for complex-shaped target volumes? Radiother Oncol 2009;93:259–65.

[5] Ahamad A, D’Souza W, Salehpour M, et al. Intensity-modulated radiationtherapy after hysterectomy: comparison with conventional treatment andsensitivity of the normal-tissue-sparing effect to margin size. Int J Radiat OncolBiol Phys 2005;62:1117–24.

[6] Heron DE, Gerszten K, Selvarai RN, et al. Conventional 3D conformal versusintensity-modulated radiotherapy for the adjuvant treatment of gynecologicmalignancies: a comparative dosimetric study of dose-volume histograms.Gynecol Oncol 2003;91:39–45.

[7] Mundt AJ, Lujan AE, Rotmensch J, et al. Intensity-modulated whole pelvicradiotherapy in women with gynecologic malignancies. Int J Radiat Oncol BiolPhys 2002;52:1330–7.

[8] Portelance L, Chao KS, Grigsby PW, Bennet H, Low D. Intensity-modulatedradiation therapy (IMRT) reduces small bowel, rectum, and bladder doses inpatients with cervical cancer receiving pelvic and para-aortic irradiation. Int JRadiat Oncol Biol Phys 2001;51:261–6.

[9] van de Bunt L, van der Heide UA, Ketelaars M, de Kort GA, Jurgenliemk-SchulzIM. Conventional, conformal, and intensity-modulated radiation therapytreatment planning of external beam radiotherapy for cervical cancer: theimpact of tumor regression. Int J Radiat Oncol Biol Phys 2006;64:189–96.

[10] Cozzi L, Dinshaw KA, Shrivastava SK, et al. A treatment planning studycomparing volumetric arc modulation with RapidArc and fixed field IMRT forcervix uteri radiotherapy. Radiother Oncol 2008;89:180–91.

[11] Breedveld S, Storchi PRM, Keijzer M, Heemink AW, Heijmen BJM. A novelapproach to multi-criteria inverse planning for IMRT. Phys Med Biol2007;52:6339–53.

[12] Breedveld S, Storchi PRM, Heijmen BJM. The equivalence of multi-criteriamethods for radiotherapy plan optimization. Phys Med Biol2009;54:7199–209.

Please cite this article in press as: Sharfo AWM et al. Comparison of VMAT anRadiother Oncol (2015), http://dx.doi.org/10.1016/j.radonc.2015.02.006

[13] Breedveld S, Storchi PRM, Voet PWJ, Heijmen BJM. ICycle: integrated,multicriterial beam angle, and profile optimization for generation ofcoplanar and noncoplanar IMRT plans. Med Phys 2012;39:951–63.

[14] Rossi L, Breedveld S, Heijmen BJM, Voet PWJ, Lanconelli N, Aluwini S. On thebeam direction search space in computerized non-coplanar beam angleoptimization for IMRT-prostate SBRT. Phys Med Biol 2012;57:5441–58.

[15] Voet PWJ, Breedveld S, Dirkx MLP, Levendag PC, Heijmen BJM. Integratedmulticriterial optimization of beam angles and intensity profiles for coplanarand noncoplanar head and neck IMRT and implications for VMAT. Med Phys2012;39:4858–65.

[16] Voet PWJ, Dirkx MLP, Breedveld S, Fransen D, Levendag PC, Heijmen BJM.Toward fully automated multicriterial plan generation: a prospective clinicalstudy. Int J Radiat Oncol Biol Phys 2013;85:866–72.

[17] Voet PWJ, Dirkx MLP, Breedveld S, Heijmen BJM. Automated generation ofIMRT treatment plans for prostate cancer patients with metal hip prostheses:comparison of different planning strategies. Med Phys 2013;40:071704.

[18] Bondar ML, Hoogeman MS, Mens JW, et al. Individualized nonadaptive andonline-adaptive intensity-modulated radiotherapy treatment strategies forcervical cancer patients based on pretreatment acquired variable bladderfilling computed tomography scans. Int J Radiat Oncol Biol 2012;83:1617–23.

[19] Roeske JC, Bonta D, Mell LK, Lujan AE, Mundt AJ. A dosimetric analysis of acutegastrointestinal toxicity in women receiving intensity-modulated whole-pelvic radiation therapy. Radiother Oncol 2003;69:201–7.

[20] Voet PWJ, Dirkx MLP, Breedveld S, Al-Mamgani A, Incrocci L, Heijmen BJM.Fully automated volumetric modulated arc therapy plan generation forprostate cancer patients. Int J Radiat Oncol Biol Phys 2014;88:1175–9.

[21] Kavanagh BD, Pan CC, Dawson LA, et al. Radiation dose-volume effects in thestomach and small bowel. Int J Radiat Oncol Biol Phys 2010;76:S101–7.

[22] Alber M, Reemtsen R. Intensity modulated radiotherapy treatment planning byuse of a barrier-penalty multiplier method. Opt Methods Softw2007;22:391–411.

[23] Bortfeld T, Webb S. Single-Arc IMRT? Phys Med Biol 2009;54:N9–N20.[24] Bedford JL, Thomas MD, Smyth G. Beam modeling and VMAT performance

with the Agility 160-leaf multileaf collimator. J Appl Clin Med Phys2013;14:4136.

d IMRT strategies for cervical cancer patients using automated planning.