OPTIBAR PSM 2010 - Downloads

40
Electronic pressure switch with flush welded diaphragm OPTIBAR PSM 2010 OPTIBAR PSM 2010 OPTIBAR PSM 2010 OPTIBAR PSM 2010 Handbook Handbook Handbook Handbook © KROHNE 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

Transcript of OPTIBAR PSM 2010 - Downloads

Electronic pressure switch with flush welded diaphragm

OPTIBAR PSM 2010OPTIBAR PSM 2010OPTIBAR PSM 2010OPTIBAR PSM 2010 HandbookHandbookHandbookHandbook

© KROHNE 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without

the prior written authorisation of KROHNE Messtechnik GmbH.

Subject to change without notice.

2 www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

Copyright 2020 by

KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany)

: IMPRINT :::::::::::::::::::::::::::::::::::::::

CONTENTS

3www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

OPTIBAR PSM 2010

1 Safety instructions 5

1.1 Intended use ..................................................................................................................... 51.2 Technical limits ................................................................................................................ 51.3 Certification ...................................................................................................................... 51.4 Safety instructions from the manufacturer ..................................................................... 6

1.4.1 Copyright and data protection ................................................................................................ 61.4.2 Disclaimer ............................................................................................................................... 61.4.3 Product liability and warranty ................................................................................................ 71.4.4 Information concerning the documentation........................................................................... 71.4.5 Warnings and symbols used................................................................................................... 8

1.5 Safety instructions for the operator................................................................................. 8

2 Device description 9

2.1 Scope of delivery............................................................................................................... 92.2 Nameplate ...................................................................................................................... 10

3 Installation 11

3.1 General notes on installation ......................................................................................... 113.2 Installation specifications .............................................................................................. 113.3 Mounting ......................................................................................................................... 12

3.3.1 Installation steps for connections acc. to DIN 3852............................................................. 133.3.2 Installation steps for connections acc. to EN 837 ................................................................ 133.3.3 Installation steps for NPT connections ................................................................................ 133.3.4 Installation steps for clamp and Varivent connections........................................................ 143.3.5 Installation steps for flange connections ............................................................................. 14

4 Electrical connections 15

4.1 Safety instructions.......................................................................................................... 154.2 Terminal assignment ..................................................................................................... 154.3 Electrical connection diagram ....................................................................................... 164.4 Switching output signals ................................................................................................ 17

5 Start-up 20

5.1 Commissioning............................................................................................................... 205.2 Keypad functions ............................................................................................................ 205.3 Menu structure ............................................................................................................... 21

6 Service 23

6.1 Maintenance ................................................................................................................... 236.2 Recalibration .................................................................................................................. 236.3 Spare parts availability................................................................................................... 236.4 Availability of services .................................................................................................... 23

CONTENTS

4 www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

OPTIBAR PSM 2010

6.5 Repairs............................................................................................................................ 236.6 Returning the device to the manufacturer..................................................................... 24

6.6.1 General information.............................................................................................................. 246.6.2 Form (for copying) to accompany a returned device............................................................ 25

6.7 Disposal .......................................................................................................................... 25

7 Technical data 26

7.1 Technical data................................................................................................................. 267.2 Dimensions ..................................................................................................................... 297.3 Measuring ranges........................................................................................................... 32

8 Description IO-LINK interface 33

8.1 General description ........................................................................................................ 338.2 SIO-mode (standard IO mode)........................................................................................ 338.3 IO-Link mode (communication mode) ........................................................................... 338.4 Process data ................................................................................................................... 338.5 Error message................................................................................................................ 348.6 Event codes..................................................................................................................... 348.7 Parameter data............................................................................................................... 35

9 Notes 38

SAFETY INSTRUCTIONS 1

5

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

1.1 Intended use

The OPTIBAR PSM 2010OPTIBAR PSM 2010OPTIBAR PSM 2010OPTIBAR PSM 2010 pressure transmitter is designed to measure the absolute pressure and gauge pressure in gases and liquids.

1.2 Technical limits

The device was constructed solely for use within the technical limits indicated on the nameplate and in the technical data. Applications outside of these limits are not permitted and could lead to significant risk of accident. For this reason, observe the following limits:

• Do not exceed the maximum working pressure (MWP).• Do not exceed the indicated permissible operating temperature range.• The permissible ambient temperatures given may not be exceeded or undershot.• Check the materials used for the wetted parts (e.g. gasket, process connection, separating

diaphragm etc.) for suitability as regards process compatibility.

1.3 Certification

CE markingThe device fulfils the statutory requirements of the following EU directives:

• EMC Directive 2014/30/EU• EMC specification acc. to EN 61326-1:2013• RoHS Directive 2011/65/EU

The manufacturer certifies successful testing of the product by applying the CE marking.

Pressure equipment directive (PED)Devices with a permissible pressure PS ≤ 200 bar (20 MPa) comply with Pressure equipment directive (PED) 2014/68/EU Article 4 Section (3) and are not subject to a conformity assessment. These devices were designed and manufactured in accordance with sound engineering practice (SEP).

The CE marking on the device does not apply to the Pressure Equipment Directive.

CAUTION!Responsibility for the use of the measuring devices with regard to suitability, intended use and corrosion resistance of the used materials against the measured fluid lies solely with the operator.

INFORMATION!The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose.

DANGER!For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.

1 SAFETY INSTRUCTIONS

6

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

1.4 Safety instructions from the manufacturer

1.4.1 Copyright and data protection

The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.

The contents and works in this document are subject to copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.

The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.

The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.

We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.

We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.

1.4.2 Disclaimer

The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect or incidental and consequential damages.

This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.

Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.

The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.

SAFETY INSTRUCTIONS 1

7

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

1.4.3 Product liability and warranty

The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation or operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.

1.4.4 Information concerning the documentation

To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.

If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.

This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of icons as shown below.

1 SAFETY INSTRUCTIONS

8

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

1.4.5 Warnings and symbols used

Safety warnings are indicated by the following symbols.

• HANDLINGHANDLINGHANDLINGHANDLINGThis symbol designates all instructions for actions to be carried out by the operator in the specified sequence.

i RESULTRESULTRESULTRESULTThis symbol refers to all important consequences of the previous actions.

1.5 Safety instructions for the operator

DANGER!This warning refers to the immediate danger when working with electricity.

DANGER!This warning refers to the immediate danger of burns caused by heat or hot surfaces.

DANGER!This warning refers to the immediate danger when using this device in a hazardous atmosphere.

DANGER!These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.

WARNING!Disregarding this safety warning, even if only in part, poses the risk of serious health problems. There is also the risk of damaging the device or parts of the operator's plant.

CAUTION!Disregarding these instructions can result in damage to the device or to parts of the operator's plant.

INFORMATION!These instructions contain important information for the handling of the device.

LEGAL NOTICE!This note contains information on statutory directives and standards.

WARNING!In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device.

DEVICE DESCRIPTION 2

9

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

2.1 Scope of delivery

The following items are supplied with the device:• Measuring device in ordered version• For mechanical connections DIN 3852: O-ring (pre-assembled)• Product documentation

INFORMATION!Inspect the packaging carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.

INFORMATION!Do a check of the packing list to make sure that you have all the elements given in the order.

INFORMATION!Look at the device nameplate to ensure that the device is delivered according to your order.Check for the correct supply voltage printed on the nameplate.

INFORMATION!Assembly materials and tools are not part of the delivery. Use the assembly materials and tools in compliance with the applicable occupational health and safety directives.

2 DEVICE DESCRIPTION

10

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

2.2 Nameplate

The device can be identified by its nameplate. It provides the most important data.

INFORMATION!Look at the device nameplate to ensure that the device is delivered according to your order.Check for the correct supply voltage printed on the nameplate.

Figure 2-1: Example for a nameplate

1 Manufacturer logo and address2 CE marking and other approvals (e.g. ATEX, PED,...)3 Serial number4 Type code5 Barcode6 Electrical connection diagram7 Specifications for process conditions (measuring range, MWP (= Maximum Working Pressure) and electrical data (sig-

nal output and power supply))8 Product name

VGKY4RA0S30V00SM1000OPTIBAR PSM 2010

Ub+ : Ub- : SIO / IO-Link: S+ : PE : connector

1

2

34

INSTALLATION 3

11

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

3.1 General notes on installation

3.2 Installation specifications

INFORMATION!Inspect the packaging carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.

INFORMATION!Do a check of the packing list to make sure that you have all the elements given in the order.

INFORMATION!Look at the device nameplate to ensure that the device is delivered according to your order.Check for the correct supply voltage printed on the nameplate.

WARNING!Install the device only when depressurised and without power!

DANGER!For installation the respective regulations for explosion protection have to be fulfilled.

INFORMATION!For installations outdoor and in damp areas, the following points must be observed:• To ensure that no moisture can get into the connector, the device should be connected

electrically immediately after installation. Otherwise a moisture admission has to be prevented e.g. by using a suitable protection cap.

• Select an installation if possible, where a mounting position allows draining of spray and condensate. Sealing surfaces should not be submerged!

• When using devices with cable glands or output, the cable should be looped facing down so that any liquid that collects on the cable can drip off.

• Install the device so it is protected from direct sunlight. In the worst case scenario, the permissible operating temperature will be exceeded in the presence of direct sunlight. This can negatively affect or damage the functionality of the device. In addition, it can lead to temporary measuring errors if the internal pressure of the device increases due to the sunlight.

• When installing outside where the risk of lightning or overvoltage may exist and damage the device, we recommend installing suitable overvoltage protection between the supply device or control cabinet and the device.

3 INSTALLATION

12

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

3.3 Mounting

INFORMATION!• Handle this highly sensitive electronic measuring device with care, both in and out of the

packaging!• Only remove the packaging and any protection cap from the device immediately before

installing to prevent damage to the diaphragm! Keep the supplied protection cap! Remove the protection cap slowly and carefully to avoid any negative pressure on the diaphragm.

• Handle the unprotected diaphragm with extreme care; it is very easily damaged.• A device with a gauge reference in the housing (small hole next to the electrical connection)

must be installed so that the gauge reference necessary for measurement is protected from dirt and moisture. Should the pressure transmitter be exposed to fluid admission, the air pressure compensation is blocked by the gauge reference. Accurate measurement in this state is not possible. It can also result in damage to the pressure transmitter.

• Ensure that no mechanical stress is applied to the pressure port during installation as this may result in a shift in the characteristic curve. This applies in particular to very small pressure ranges as well as to devices with plastic pressure ports.

• With hydraulic systems, arrange the device so that the pressure port faces up (venting).

WARNING!Do not screw in using the housing! Tightening this way can cause damage to the rotary mechanism on the housing.

CAUTION!• Prior to installing the transmitter, it is essential to verify whether the version of the device on

hand completely fulfils the technical and safety requirements of the measuring point. This applies in particular to the measuring range, overpressure resistance, temperature, explosion protection and operating voltage.

• Check the materials used for the wetted parts (e.g. gasket, process connection, separating diaphragm etc.) for suitability as regards process compatibility.

• The device must not be heated by radiated heat (e.g. exposure to the sun) to an electronics housing surface temperature above the maximum permissible ambient temperature. If it is necessary to prevent damage from heat sources, a heat protection (e.g. sun shade) has to be installed.

INSTALLATION 3

13

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

3.3.1 Installation steps for connections acc. to DIN 3852

• Make sure that the O-ring fits properly in the intended groove.• Ensure that the raised face of the receiving part has a smooth surface.• Screw the device into the thread by hand.• Devices with wrench flats must be tightened with a wrench. For information on tightening

torque refer to the following table.

3.3.2 Installation steps for connections acc. to EN 837

• When sealing, use a suitable gasket that corresponds to the product and pressure to be measured (e.g. a copper gasket).

• Ensure that the raised face of the receiving part has a smooth surface.• Screw the device into the thread by hand.• Then tighten the device with the wrench.

3.3.3 Installation steps for NPT connections

• When sealing, use a suitable sealing agent that is compatible with the product (e.g. PTFE tape).

• Screw the device into the thread by hand.• Then tighten the device with the wrench.

Wrench size Tightening torque [Nm]

Wrench size of steelWrench size of steelWrench size of steelWrench size of steel

G1/4 approx. 5

G1/2 approx. 10

G3/4 approx. 15

G1 approx. 20

G1 1/2 approx. 25

Wrench size of plasticWrench size of plasticWrench size of plasticWrench size of plastic

All sizes max. 3

Wrench size Tightening torque [Nm]

G1/4 approx. 20

G1/2 approx. 50

Wrench size Tightening torque [Nm]

1/4 NPT approx. 30

1/2 NPT approx. 70

3 INSTALLATION

14

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

3.3.4 Installation steps for clamp and Varivent connections

Notes for versions acc. to 3A standard

The device has to be installed in the way that drainability (E1.8.1) and demands acc. to L1.1 are met. The device must be installed with the pressure port downwards or to the side.

The user is responsible for:• using a suitable sealing material• defining adequate service intervals• using an elastomer sealing material according to 3A standard 18-03 and 21CFR177.2600• the right dimension of the gasket

Make sure that welding sockets are mounted flush inside the tank.

General procedure• When sealing, use a suitable gasket that corresponds to the product and pressure to be

measured.• Place the gasket on the corresponding fitting.• Centre the clamp or Varivent connection over the corresponding fitting and gasket.• Then use a suitable connecting element (e.g. half ring or clamp ring connection) to attach the

device according to the manufacturer's specifications.

3.3.5 Installation steps for flange connections

• When sealing, use a suitable gasket that corresponds to the product and pressure to be measured (e.g. a fibre gasket)

• Put the gasket between connecting flange and counter flange.• Then fasten the device using 4 or 8 screws (depending on flange version) to the counter flange.

INFORMATION!The device must be assembled according to 3A standard 74-06.

ELECTRICAL CONNECTIONS 4

15

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

4.1 Safety instructions

4.2 Terminal assignment

DANGER!All work on the electrical connections may only be carried out with the power disconnected.Take note of the voltage data on the nameplate!

DANGER!Observe the national regulations for electrical installations!

DANGER!For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.

WARNING!Observe without fail the local occupational health and safety regulations.Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.

INFORMATION!Look at the device nameplate to ensure that the device is delivered according to your order.Check for the correct supply voltage printed on the nameplate.

WARNING!Install the device only when depressurised and without power!

Electrical connections Description M12x1 (4-pin)

Supply + 1

Supply - 3

Signal 1 IO-Link / PNP / NPNswitchable (SIO)

4

Signal 2 PNP / NPN / mA / Vswitchable

2

Shielding Pressure-port

4 ELECTRICAL CONNECTIONS

16

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

4.3 Electrical connection diagram

Figure 4-1: Electrical connection diagram 2-wire 4...20 mA

1 Supply +2 Supply - 3 Signal 2 4 Power supply5 IO-Link Master

I / U

ELECTRICAL CONNECTIONS 4

17

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

4.4 Switching output signals

The output for the switching signals can be configured with two different function, both for normally open and normally closed actuators.

Window functionWhen the window function is activated, the output signal is set when the applied pressure is within the upper (FH) and lower (FL) pressure threshold:

Figure 4-2: Switching output signals (window function)

1 FH2 Pressure range3 FL4 Hno (normally open)5 Hnc (normally closed)

01

01

P

t

t

t

P

P

4 ELECTRICAL CONNECTIONS

18

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

Hysteresis-functionWhen the hysteresis-function is activated, the output signal is activated when the pressure reaches the switching pressure (SP) and deactivated when the release pressure (rP) is applied:

Figure 4-3: Switching output signals (hysteresis-function)

1 SP2 Hysteresis3 rP4 Hno (normally open)5 Hnc (normally closed)

0

0

1

1

P

t

t

t

ELECTRICAL CONNECTIONS 4

19

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

Hysteresis-functionTo eliminate an unintended switching signal caused by pressure spikes, a damping-constant (dr, dS) can be configured for both switching-functions:

Figure 4-4: Switching output signals (hysteresis-function with damping)

1 SP2 Hysteresis3 rP4 Hnc (normally closed)5 dS6 dr7 Hno (normally open)

01

01

P

t

t

t

5 START-UP

20

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

5.1 Commissioning

The signal converter may only be started up after it has been completely installed and checked by appropriately qualified personnel. Switch on the operating voltage for start-up.

Prior to applying the operating voltage check that • the pressure transmitter is completely installed• the process connection fits properly• the signal and, if necessary, supply lines are properly connected• the impulse lines are completely filled with the process medium

5.2 Keypad functions

The display and adjustment module is used for indication of measuring values and adjustment.

The device is operated via the two keys 2 on the display and adjustment module. The LED-Display 1 indicates the individual menu items.

Figure 5-1: Keypad

1 LED Display2 Function buttons

Key Description Function

Short press Skip through menu 1 – 5

Long press Count up parameter values quickly

Short press Select option within one menu

Long press Save adjusted value and jump pack to current menu

+ Press both simultaneously Jump back to indication

START-UP 5

21

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

5.3 Menu structure

Function Description

Operating mode

SP1 | FH1 Adjustment of switch-on point 1Set value from where switching point 1 should be activated (SP1). If the window function is activated, the value is the upper pressure limit of the window (FH1).

rP1 | FL1 Adjustment of switch-off point 1Set value from where switching point 1 should be deactivated (rP1). If the window function is activated, the value is the lower pressure limit of the window (FL1).

SP2 | FH2 Adjustment of switch-on point 2Set value from where switching point 2 should be activated (SP2). If the window function is activated, the value is the upper pressure limit of the window (FH2).

rP2 | FL2 Adjustment of switch-off point 2Set value from where switching point 2 should be deactivated (rP2). If the window function is activated, the value is the lower pressure limit of the window (FL2).

Additional menuASt2 | AEn2(Only if Output-Signal 2is activated)

Adjustment of output-signal 2Analog-output 2. Change possibilities: 5% at zero (ASt2), 90-100% at span (AEn2)

EF - (Extended function)

rES Factory-reset

dS1 Adjustment of switch-on delay 1Set value for switch-on delay after reaching switch-on point 1(Adjustable from 0.0...50.0 s)

dr1 Adjustment of switch-off delay 1Set value for switch-off delay after reaching switch-off point 1(Adjustable from 0.0...50.0 s)

dS2 Adjustment of switch-on delay 2Set value for switch-on delay after reaching switch-on point 2(Adjustable from 0.0...50.0 s)

dr2 Adjustment of switch-off delay 2Set value for switch-off delay after reaching switch-off point 2(Adjustable from 0.0...50.0 s)

ou1 Adjustment of switch-output 1Switching functions for switch-output:- Hno: Hysteresis function, normally open- Hnc: Hysteresis function, normally closed- Fno: Window function, normally open- Fnc: Window function, normally closed

ou2 Adjustment of switch-output 2See "ou1"

Uni Change UnitsSelect unit for indicated and adjusted pressure valuebAr = barnnBa = mbarPsi = PSIΠPA = MPa

5 START-UP

22

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

EF - (Extended function)

FLIP Rotate indication by 180°

Lo Min-ValueShow minimal applied pressure during measurement. In case of interruption of power supply the value is lost

Hi Max-ValueShow maximal applied pressure during measurement. In case of interruption of power supply the value is lost

---- Reset "Min-Value" and "Max-Value"

SEt0 Adjustment of Zero-PointAdjustment/Correction of zero-point of indicated value and analog-output signal by up to 3% of nominal pressure

dAP Damping of measurementSet value of damping(0...1000 ms in 10 ms steps)

codE Access protectionSet password for access protection0000 = no password (deactivated)1000...9999 adjustable (activated)

o1 Output-Signal 1Switch between PNP- and NPN-functions

o2 Output-Signal 2Switch between PNP- and NPN-functions, 4...20 mA and 0...10 V

hent Operating time in [h]

Pent Number of pressure spikes

Function Description

SERVICE 6

23

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

6.1 Maintenance

In principal, the device is maintenance free. If necessary, clean the device housing when switched off using a damp cloth and a non-aggressive cleaning solution.

Depending on the product, deposits or contamination can still occur on the diaphragm. If the product has a known affinity, the operator must determine the cleaning intervals accordingly. Once the device has been properly taken out of commission, the diaphragm can generally be carefully cleaned using a non-aggressive cleaning solution and a soft brush or sponge. If the diaphragm is calcified, decalcification by the manufacturer is recommended.

6.2 Recalibration

During the life cycle of the device, the offset or full-scale of the device may shift. If this occurs, note that the signal value output will deviate based on the set start or end value of the measuring range. If one of these phenomena does occur following prolonged use, recalibration is recommended to ensure continued high accuracy.

6.3 Spare parts availability

The manufacturer adheres to the basic principle that functionally adequate spare parts for each device or each important accessory part will be kept available for a period of 3 years after delivery of the last production run for the device.

This regulation only applies to spare parts which are subject to wear and tear under normal operating conditions.

6.4 Availability of services

The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, maintenance, technical support and training.

6.5 Repairs

Repairs must be carried out exclusively by the manufacturer or the manufacturer authorised specialist companies.

INFORMATION!Improper cleaning can result in irreparable damage to the measuring cell. For this reason, never use sharp objects or compressed air to clean the diaphragm.

INFORMATION!For more precise information, please contact your local sales office.

6 SERVICE

24

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

6.6 Returning the device to the manufacturer

6.6.1 General information

This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems.

WARNING!Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:• Due to statutory regulations on environmental protection and safeguarding the health and

safety of the personnel, the manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.

• This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle.

WARNING!If the device has been operated with toxic, caustic, radioactive, flammable or water-endangering products, you are kindly requested:• to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from

such dangerous substances,• to enclose a certificate with the device confirming that it is safe to handle and stating the

product used.

SERVICE 6

25

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

6.6.2 Form (for copying) to accompany a returned device

6.7 Disposal

CAUTION!To avoid any risk for our service personnel, this form has to be accessible from outside of the packaging with the returned device.

Company: Address:

Department: Name:

Telephone number:

Fax number:

Email address:

Manufacturer order number or serial number:

The device has been operated with the following medium:

This medium is: radioactive

water-hazardous

toxic

caustic

flammable

We checked that all cavities in the device are free from such substances.

We have flushed out and neutralized all cavities in the device.

We hereby confirm that there is no risk to persons or the environment caused by any residual media contained in this device when it is returned.

Date: Signature:

Stamp:

LEGAL NOTICE!Disposal must be carried out in accordance with legislation applicable in your country.

Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:

According to the directive 2012/19/EU, the monitoring and control instruments marked with the WEEE symbol and reaching their end-of-life must not be disposed of with other wastemust not be disposed of with other wastemust not be disposed of with other wastemust not be disposed of with other waste.The user must dispose of the WEEE to a designated collection point for the recycling of WEEE or send them back to our local organisation or authorised representative.

7 TECHNICAL DATA

26

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

7.1 Technical data

INFORMATION!• The following data is provided for general applications. If you require data that is more

relevant to your specific application, please contact us or your local sales office.• Additional information (certificates, special tools, software,...) and complete product

documentation can be downloaded free of charge from the website (Downloadcenter).

Measuring systemMeasuring principle Piezoresistive measuring cell

Application range Measurement of gauge and absolute pressure in gases and liquids

Measuring range 0.1...40 bar / 1.5...580 psi; refer also to chapter "Measuring ranges"Adjustment (in relation to the nominal range):Zero: ± 5%Span: 90...100%

Display and user interfaceDisplay on signal converter 4-digit, 7-segment LED-Display 22.5 x 10.5 mm / 0.89 x 0.41"

4 LED's for indication of unit (bar, mbar, PSI, Mpa)Status LED for IO-Link and switching-outputsDisplay infinitely rotatable up to -210° and +100°Ambient temperatures below -20°C may affect the readability of the display

Display function Display of measured valueAll parameters are accessible via the operating menu

Operating Local operation via 2 softkeys on the display and adjustment module

Remote control PACTware™ via IODD and USB IO-Link interface

Measuring accuracyReference conditions Medium: air

Temperature: ambient temperature

Ambient pressure: 1013 mbar / 14.7 psi

Nominal position: vertical, pressure port down

Power supply: 24 VDC

Pressure type Gauge pressure / absolute pressure

Measuring accuracy according to IEC 60770 (terminal based)(Hysteresis, non-linearity, non-repeatability)

Nominal pressure (PN) < 0.4 bar / 5.8 psi: ≤± 0.5% of URL

Nominal pressure (PN) ≥ 0.4 bar / 5.8 psi: ≤± 0.35% of URL

(URL = Upper Range Limit)

Ambient temperature effect on zero and span

≤± 0.3% of URL per 10K in compensated range of -25...85°C / -13...185°F

Long-term stability ≤±0.3% of URL within one year under reference conditions

Step response time ≤ 12 ms (T90)

Vacuum resistance PN ≥ 1 bar / 14.5 psi: vacuum resistantPN < 1 bar / 14.5 psi: on request

TECHNICAL DATA 7

27

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

Operating conditionsTemperatureTemperatureTemperatureTemperature

Nominal temperature -20...+85°C / -4...+185°F

Ambient temperature -40…+85°C / -40...+185°F

Process temperature Silicone oil: -40...+125°C / -40...+257°F

Food grade oil: -10...+125°C / +14...+257°F

With cooling fins (optional):

Silicone oil:PN > 0 barg: -40...+300°C / -40...+572°F;PN < 0 barg: -40...+150°C / -40...+302°F

Food grade oil:PN > 0 barg: -10..+250°C / +14...+482°F;PN < 0 barg: -10...+150°C / +14...+301°F

Installation conditionsMounting position Any - factory calibration carried out with pressure port down.

Dimensions For detailed information refer to chapter "Dimensions and weights".

Weight Min. 230 g / 0.51 lb (depending on pressure port)

MaterialsHousing Stainless steel 1.4404 / AISI 316L

Fill fluid Silicone oil

Wetted partsWetted partsWetted partsWetted parts

Pressure port Stainless steel 1.4404 / AISI 316L

Separating diaphragm Stainless steel 1.4435 / AISI 316L

Sealing FKM, EPDM (PN ≤ 160 bar)

Process connectionsThread Thread ISO 228 G1/2", DIN 3852 (PN ≥ 1 bar / 14.5 psi),

Thread ISO 228 G3/4", DIN 3852 (PN ≥ 600 mbar / 8.7 psi), Thread ISO 228 G1", DIN 3852; Thread ISO 228 G1", coneThread ISO228 G1/2" hygienic design; DIN3852Thread ISO228 G1" hygienic design; DIN3852

Clamp Clamp DN25 ISO 2852; clamp DN38 ISO 2852; clamp DN51 ISO 2852

Others on request.

Electrical connectionOutput signal 1 IO-Link / SIO (PNP/NPN) switchable

Output signal 2 4...20 mA / 0...10 V (3-wire) / with PNP / NPN switchable

Power supply Ub = 18...30 VDC

Load 4...20m A / 3-wire: Rlmax ≤ (Ub - Ubmin) / 0.036 A [Ohm]0...10 V /3-wire: Rlmin ≥ Ub / 0.0012 A [Ohm]

Short circuit protection Continuously

Reverse polarity protection In the event of reversed connections there is no damage but also no function.

7 TECHNICAL DATA

28

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

Ripple Supply: 0.05% URL Load: ≤ 0.1% URL

Electrical connection Connector M12x1, 4-pin

Properties switching signalSwitch point accuracy ≤ +/- 0.5% URL

Repeatability ≤ +/- 0.1% URL

Max. switching current 150 mA

Switching frequency max. 170 Hz

Delay 0...50 seconds

IO-LinkInterface IO-Link 1.1; Slave

Data transmission 38.4 kbit/s (COM2)

Mode SIO / IO-Link

Standard IEC 61132-2, IEC 61161-9

Approvals and certificatesCE The device fulfils the statutory requirements of the EU directives. The manufacturer

certifies that these requirements have been met by applying the CE marking.

Electromagnetic compatibility (EMC) acc. to EN 61326-1:2013

EMC Directive: 2014/30/EU

For more information consult the relevant declaration of conformity.

Pressure equipment directive 2014/68/EU

Other standards and approvalsOther standards and approvalsOther standards and approvalsOther standards and approvals

Vibration resistance acc. to EN 60068-2-6

10g RMS (25...2000 Hz)

Shock resistant (impact) according to EN 60068-2-27

500g / 11 ms (PN ≥ 60 bar / 870 psi)

Hygienic 3-A approved; FDA approved materials

TECHNICAL DATA 7

29

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

7.2 Dimensions

Figure 7-1: Dimensions

Dimensions

a b c

[mm] ["] [mm] ["] [mm] ["]

26 1.02 35.5 1.40 64 2.52

7 TECHNICAL DATA

30

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

The entire length of the device is made up of the electrical connection (a), the transmitter housing (e) and the process connection (k).With cooling fins (optional) additional 32 mm / 1.26".

Pressure transmitter with threaded connection

Figure 7-2: Dimensions for pressure transmitter with threaded connection

1 G1/2" DIN 3852 flush2 G3/4" DIN 3852 flush3 G1" DIN 3852 flush4 G1/2" DIN 3852 flush hygienic design5 G1" DIN 3852 flush hygienic design6 G1" cone

Dimensions

1 2 3 4 5 6

[mm] ["] [mm] ["] [mm] ["] [mm] ["] [mm] ["] [mm] ["]

e 60 2.36 60 2.36 60 2.36 60 2.36 60 2.36 60 2.36

f Ø26.5 Ø1.04 Ø26.5 Ø1.04 Ø26.5 Ø1.04 Ø26.5 Ø1.04 Ø26.5 Ø1.04 Ø26.5 Ø1.04

g WS27 WS34 WS41 WS27 WS41 WS34

h G½" G¾" G1" G½" G1" G1"

k 18 0.71 20 0.79 22 0.87 23.5 0.93 23.5 0.93 26 1.02

m 15 0.59 17 0.67 19 0.75 20.5 0.81 20.5 0.81 - -

n - - - - - - - - 7.5 0.3 9 0.35

TECHNICAL DATA 7

31

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

The entire length of the device is made up of the electrical connection (a), the transmitter housing (e) and the process connection (k).

Pressure transmitter with cooling fins and hygienic process connections

Figure 7-3: Dimensions for pressure transmitter with cooling fins and hygienic process connections

1 Cooling fins (optional)2 Clamp DIN 32678 / ISO 2852, 3-A approved

3 Varivent® DN40/50, 3-A approved

Dimensions

1 2 3

[mm] ["] [mm] ["] [mm] ["]

e 50 1.97 50 1.97 48 1.89

f Ø26.5 Ø1.04 Ø26.5 Ø1.04 Ø26.5 Ø1.04

h - -DN25: 50.5

DN38: 50.5

DN51: 64

DN25: 2.0

DN38: 2.0

DN51: 2.5

Ø84 Ø3.3

k - - 5 0.2 17 0.67

r 32 1.26 - - - -

INFORMATION!Other hygienic connections are available on request.

7 TECHNICAL DATA

32

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

7.3 Measuring ranges

Pressure in bar

Pressure in psi

Nominal pressure (gauge/abs.)

0.10 0.16 0.25 0.40 0.50 0.60 1 1.6

Max. working pressure (MWP)

0.5 1 1 2 2 5 5 10

Burst pressure (OPL) 1.5 1.5 1.5 3 3 7.5 7.5 15

Nominal pressure (gauge/abs.)

2 2.5 4 5 6 10 16 25 40

Max. working pressure (MWP)

10 10 20 20 40 40 80 80 105

Burst pressure (OPL) 15 15 25 25 50 50 120 120 210

Nominal pressure (gauge/abs.)

1.45 2.32 3.63 5.80 7.5 8.70 14.5 23.2

Max. working pressure (MWP)

7.3 14.5 14.5 29 29 72.5 72.5 145

Burst pressure (OPL) 21.8 21.8 21.8 43.5 43.5 108.8 108.8 217.6

Nominal pressure (gauge/abs.)

29 36.3 58.0 72.5 87.0 145 232.1 362.6 580

Max. working pressure (MWP)

145 145 290 290 580 580 1160 1160 1523

Burst pressure (OPL) 217,6 217.6 362.6 362.6 725 725 1740 1740 3046

DESCRIPTION IO-LINK INTERFACE 8

33

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

8.1 General description

The device is equipped with an IO-Link interface as standard in order to exchange process data, diagnostic reports and status messages with a superordinate control level.

IO-Link is the first standardized IO technology worldwide (IEC 61131-9) for the communication with sensors and actuators. The powerful point-to-point communication is based on the long established 3-wire sensor and actuator connection without additional requirements regarding the cable material. IO-Link is no fieldbus but the further development of the existing, tried-and-tested connection technology for sensors and actuators.

8.2 SIO-mode (standard IO mode)

In this mode the sensor operates like a normal pressure sensor with standard output signals. The digital output is always on Pin 4 (output 1) of the M12 connector. Depending on the version, pin 2 (output 2) can be an analog or an additional digital output.

8.3 IO-Link mode (communication mode)

The pressure sensor switches into IO-Link communication mode when operating under an IO-Link master. IO-Link communication is only possible via Pin 4 of the M12 connector.

8.4 Process data

The process data length of the sensor is 16 bits. The switching states (BCD1 and BCD2) as well as the current measured values are transmitted. The 14 bits of the measured value are scaled according to the measuring range.

Baud rate COM 2 (38,4 kBaud)

Input process data length 2 Byte

Minimum cycle time 5 ms

IO-Link version V 1.1 (backward compatible V 1.0)

SIO-Modus Yes

15 bit15 bit15 bit15 bit 14...214...214...214...2 1111 0000

Signed Bit Measurement BDC2 /output 2

BDC1 /output 1

INFORMATION!Please note the bit sequence, otherwise the process value will be misinterpreted.

8 DESCRIPTION IO-LINK INTERFACE

34

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

8.5 Error message

8.6 Event codes

Error CodeError CodeError CodeError Code DescriptionDescriptionDescriptionDescription

0x8011 Index not available

0x8012 Subindex not available

0x8023 Access denied

0x8030 Parameter value out of range

0x8033 Parameter length overrun

0x8034 Parameter length underrun

Event-Codes IOLinkEvent-Codes IOLinkEvent-Codes IOLinkEvent-Codes IOLink1.11.11.11.1

Event-Codes IOLinkEvent-Codes IOLinkEvent-Codes IOLinkEvent-Codes IOLink1.01.01.01.0

Device statusDevice statusDevice statusDevice status TypeTypeTypeType

No malfunction 0x0000 0x0000 0 Notification

General malfunction.Unknown error

0x1000 0x1000 4 Error

Process variable range over-run.Process data uncertain

0x8C10 0x8C10 2 Warning

Process variable range under-run.Process data uncertain

0x8C30 0x8C10 2 Warning

DESCRIPTION IO-LINK INTERFACE 8

35

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

8.7 Parameter data

The parameter data for the pressure sensor correspond to the Smart Sensor profile (V1.0).

Index hex Subindex hex

Object name Single Value Default Comment

0x02 0x00 System Commands

0x81 = delete Min-/Max-Value

The action is executed by writing in the subindex0x82 = res

0xA0 = Set0

0x03 0x00 Data Storage Index

0x01: Upload Start

0x02: Upload End

0x03: Download Start

0x04: Download End

0x05: Data storage Break

0x0C 0x00 Device Access Lock

0x00: Unlocked 0x00: Unlocked

0x01: Parameter access - Lock

0x02: Data Storage - Lock

0x04: Local parameterization - Lock

0x08: Local user interface - Lock

0x03: Parameter access & Data Storage - Lock

0x05: Parameter access & Local parameterization - Lock

0x09: Parameter access & Local user interface - Lock

0x06: Data Storage & Local parameterization - Lock

0x0C 0x00 Device Access Lock

0x0A: Data Storage & Local user interface - Lock

0x07: Data Storage & Parameter access & Local

parameterization - Lock

0x0B: Data Storage & Parameter access & Local user

interface - Lock

0x24 0x00 Device 0x00 Device is operating properly

Status 0x02 Out-of-Specification

0x04 Failure

0x3D 0x01 Switch Point 0x00: Value as specified

Logic 1

8 DESCRIPTION IO-LINK INTERFACE

36

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

0x3D 0x02 Switch Point 0x80: Hysteresis NO

0x82: Window NO

0x80: HNo

Mode 1 0x81: Hysteresis NC

0x83: Window NC

0x3D 0x03 Switch Point 0x0000: No Hysteresis

Hysteresis 1

0x3F 0x01 Switch Point 0x00: Value as specified

Logic 2

0x3F 0x02 Switch Point 0x80: Hysteresis NO

0x82: Window NO

0x80: HNo

Mode 2 0x81: Hysteresis NC

0x83: Window NC

0x3F 0x03 Switch Point 0x0000: No Hysteresis

Hysteresis 2

0x93 0x00 Switch Point 0x01 - NPN Output

Typ 1 0x00 - PNP Output

0x97 0x00 Switch Point 0x01 - NPN Output

0x02 - 0...10 V Output

Typ 2 0x00 - PNP Output

0x03 - 4...20 mA

0xD4 0x00 Unit 0x00 bar 0x00: bar Pressure unit of the display changed, the IO-Link process data are not changed

0x01 mbar

0x02 PSI

0x03 MPa

Index hex Subindex hex

Object name Single Value Default Comment

DESCRIPTION IO-LINK INTERFACE 8

37

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

Index Index Index Index hexhexhexhex

SubindexSubindexSubindexSubindexhexhexhexhex

Object nameObject nameObject nameObject name AccessAccessAccessAccess LengthLengthLengthLength Value Value Value Value RangeRangeRangeRange

GradientGradientGradientGradient UnitUnitUnitUnit DefaultDefaultDefaultDefault

0x3C 0x01 Set Point 1 = SP1 R/W 2 Byte Process Data

100%

0x3C 0x02 Set Point 2 = rP1 R/W 2 Byte Process Data

0%

0x3E 0x01 Set Point 1 = SP2 R/W 2 Byte Process Data

100%

0x3E 0x02 Set Point 2 = rP2 R/W 2 Byte Process Data

0%

0x52 0x00 Temperature R 2 Byte -40...125 1 C 0

0x57 0x00 Operating hours R 4 Byte 0...4294967295

1 H 0

0x60 0x00 Password W 2 Byte 0000...9999 0

0x98 0x00 Pressure peaks R 4 Byte 0...4294967295

1 0

0xD0 0x00 Delay Switching Time 1

R/W 2 Byte 0...500 0.1 Sec 0

0xD1 0x00 Delay Back Switching Time 1

R/W 2 Byte 0...500 0.1 Sec 0

0xD2 0x00 Delay Switching Time 2

R/W 2 Byte 0...500 0.1 Sec 0

0xD3 0x00 Delay Back Switching Time 2

R/W 2 Byte 0...500 0.1 Sec 0

0xD5 0x00 Min Pressure Value

R 2 Byte Process Data

0xD6 0x00 Max Pressure Value

R 2 Byte Process Data

0xD7 0x00 Measure damping R/W 2 Byte 0...1000 in 10 ms steps

1 ms 0

0x0010 0 Get Vendor Name R 64 Byte Process Data

0x0011 0 Get Vendor Text R 64 Byte Process Data

0x0012 0 Get Product Name R 64 Byte Process Data

0x0013 0 Get Product ID R 64 Byte Process Data

0x0014 0 Get Product Text R 64 Byte Process Data

0x0015 0 Get Serial Number R 64 Byte Process Data

0x0016 0 Get Hardware Revision

R 64 Byte Process Data

0x0017 0 Get Software Revision

R 64 Byte Process Data

9 NOTES

38

OPTIBAR PSM 2010

www.krohne.com 03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

NOTES 9

39

OPTIBAR PSM 2010

www.krohne.com03/2020 - 4008116901 - MA OPTIBAR PSM 2010 R01 en

KROHNE – Process instrumentation and measurement solutions

• Flow

• Level

• Temperature

• Pressure

• Process Analysis

• Services

Head Office KROHNE Messtechnik GmbHLudwig-Krohne-Str. 547058 Duisburg (Germany)Tel.: +49 203 301 0Fax: +49 203 301 [email protected]

© K

RO

HN

E 03

/202

0 -

4008

1169

01 -

MA

OP

TIB

AR

PSM

201

0 R

01 e

n -

Subj

ect t

o ch

ange

with

out n

otic

e.

The current list of all KROHNE contacts and addresses can be found at:www.krohne.com