Irradiation Damage Tests on Backside-Illuminated CMOS APS Prototypes for the Extreme Ultraviolet...
Transcript of Irradiation Damage Tests on Backside-Illuminated CMOS APS Prototypes for the Extreme Ultraviolet...
> REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) <
1
Abstract— Complementary metal oxide semiconductor
(CMOS) active pixel sensor (APS) prototypes made using 0.18
µm technology process have been developed for the Extreme
Ultraviolet Imager of the Solar Orbiter mission. Backside
illuminated CMOS APS devices i.e., 256 x 256 pixels area, 10 µm
pixel pitch with different pixel designs have been fully
characterized in the visible and in EUV wavelengths. A set of
irradiation tests were carried out to investigate the degradation
of the devices expected in the space environment conditions of
Solar Orbiter. Total ionizing dose effects from grounded
measurements are presented up to 150 krad[SiO2]. The prototype
sensors show the immunity to single-event latch up at linear
energy transfer’s of 67.7 MeV cm2/mg but were observed to
suffer from strong degradations after proton irradiation test
(with a cumulated fluence up to 4x1011 protons/cm2) and from
single event functional interrupt.
Index Terms— Image sensors, Radiation effects, Space
applications.
I. INTRODUCTION
he Extreme Ultraviolet Imager (EUI) onboard Solar
Orbiter consists of a suite of two high-resolution imagers
centered at wavelengths of 17.4 nm (HRI17.4) and 121.6 nm
(HRILy) and one dual-band full Sun imager (FSI) telescope at
17.4 nm and 30.4 nm [1], [2].
For the EUI instruments, the APSOLUTE project for “APS
Optimized for Low-noise and Ultraviolet Tests and
Manuscript received November 15, 2012. The APSOLUTE development
project was supported by the Belgian Federal Science Policy Office
(BELSPO), which is part of the EUI project (ref: EUI PEA C90343).
A. BenMoussa, B. Giordanengo and S. Gissot are with the Solar Terrestrial Center of Excellence (STCE), Royal Observatory of Belgium, Circular 3, B-
1180 Brussels, Belgium (corresponding author: 32 2 373 02 76; fax: 32 2 374
98 22; e-mail: [email protected]). G. Meynants, X. Wang, B. Wolfs and J. Bogaerts are with the CMOSIS nv,
Coveliersstraat 15, 2600 Antwerpen, Belgium.
U. Schühle is with the Max-Planck-Institut für Sonnensystemforschung, 37191, Katlenburg-Lindau, Germany.
G. Berger is with the Catholic University of Louvain-la-Neuve, Chemin du
cyclotron 2, B-1348 Louvain la Neuve, Belgium. A. Gottwald, C. Laubis, U. Kroth and F. Scholze are with the Physikalisch-
Technische Bundesanstalt (PTB), Abbestr. 2-12, D-10587 Berlin, Germany.
A. Soltani is with the Institut d'Electronique, de Microélectronique et de Nanotechnologie (IEMN) F-59652, Villeneuve d'Ascq, France.
T. Saito is with the department of Environment & Energy, Tohoku Institute
of Technology, Sendai, Miyagi 982-8577, Japan.
Experiments” has been initiated to develop a complementary
metal oxide semiconductor (CMOS) active pixel sensor (APS)
with 10 μm pixel pitch. To be sensitive to the extreme
ultraviolet (EUV) range i.e., for the HRI17.4 and FSI
telescopes, a backside illumination (BSI) approach is proposed
[3], [4].
In this paper, we report on the irradiation damage tests
performed on the BSI-APSOLUTE prototypes. We
characterize their dark current and responsivity in the visible
and EUV i.e., at around 17 nm spectral range, before and after
irradiation tests, conducted with:
- gamma rays to simulate a total ionization dose (TID), using
Cobalt-60 (60
Co) sources up to 150 krad[SiO2],
- protons for displacement damages (DD) with a cumulated
fluence up to 4x1011
particles/cm2,
- heavy ions for a single event effect (SEE) with ions at linear
energy transfer (LET) values up to 67.7 MeV cm2/mg.
We conclude with a summary of the various results and
offer recommendations to reduce expected degradation for the
fabrication of the EUI flight model (FM) detectors.
II. EXPERIMENTAL DETAILS
A. APSOLUTE Structure
APSOLUTE, 256 x 256 pixels sensors (rolling shutter)
containing 16 pixel designs, organized in blocks of 64 x 64
pixels, have been fabricated in a 0.18 µm CMOS process
technology (see Fig. 1(a)). Each pixel structure is based on the
pinned photodiode (PPD) where their in-pixel transistors differ
in shape, size and number. One of the main advantages of the
PPD is that the thin (70 nm) and heavily doped i.e., 2x1018
cm-
3 analyzed by secondary ion mass spectrometry, p
+ layer can
significantly reduce the dark current.
A particularity of the APSOLUTE sensor is that it applies a
so-called ‘dual-transfer’ scheme to achieve a high dynamic
range and minimize the noise contribution from the analog
chain. The approach reads out individual pixels through a high
gain (HG) and low gain (LG) path. The associated floating
diffusion nodes (FDHG=2.7 fF and FDLG=10 fF) determine the
sensitivity of the pixel path. For efficient charge transfer from
the PPD to the FD node, the transfer gate (TX) transistor
shape and size were optimized by adapting the gate width (W).
As an example, pixel designs #9 to #12 contain 6 transistors
Irradiation damage tests on backside-illuminated
CMOS APS prototypes for the Extreme
Ultraviolet Imager on-board Solar Orbiter
A. BenMoussa, S. Gissot, B. Giordanengo, G. Meynants, X. Wang, B. Wolfs, J. Bogaerts, U. Schühle
G. Berger, A. Gottwald, C. Laubis, U. Kroth , F. Scholze, A. Soltani, and T. Saito
T
> REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) <
2
(T) per pixel with 2 transfer gates (TXHG and TXLG) as shown
in Fig. 1(b). Pixel design #9 and #12 have the same regular
TXHG (WHG=2.2 µm) but with an extra p+ implant for the latter
on the FDHG that increases the full well capacity (FWC). The
goal of this implant is to reduce leakage between the PPD and
the FD during the exposure time. Pixel design #10 has a
smaller TXHG (WHG=1.2 µm) and pixel design #11 has a ‘U-
shaped trench-gated’ [5] TXHG (WHG=2.2 µm). For pixel
design #9 to #12, TXLG has the same design i.e., WLG=2.7 µm
and LLG=0.6 µm and the source follower (SF) transistors width
and length are identical i.e., 0.8 µm.
Fig. 1. (a) Pixel arrangement of the 16 pixel designs (256x256 pixels format),
organized in blocks of 64x64 pixels. (b) Schematic representation of the dual TX 6-T PPD pixel design (BSI approach). It should be noted that pixel
designs #8 and #13-16 using a dual source follower (SFHG, SFLG) transistors
could not meet the desired specification showing a limited FWC (<30 ke-) and were not fully characterized.
The EUV sensitivity of the APSOLUTE prototypes is
achieved with BSI on a silicon-on-insulator (SOI) material (p-
type, Si epilayer resistivity of 30 Ω cm). The epitaxial Si layer
is thinned down to either 2.75 or 2.6 µm depending on the
prototypes to gain further sensitivity in the EUV range.
Fig. 2 shows the basic analog output sensor architecture. Each
pixel contains 512 analog data samples, two for each pixel due
to the dual-transfer scheme. 512 column gain stages (up to
16x) are implemented. The output stage converts the signal to
a fully differential signal, which is sent to an off-chip analog-
to-digital converter. The APSOLUTE sensor architecture
details and operation are reported elsewhere [3], [4].
Fig. 2. Analog output sensor architecture of APSOLUTE (256x256 pixel format). The top right panel shows the simplified pixel analog path
architecture.
B. Irradiation Conditions
Three APSOLUTE prototypes containing all pixel designs
structures (denoted APT-BSI-642-0xx in the following) were
characterized before, during, and after each irradiation test and
annealing step. Although the bias applied to the test devices
shall be worst-case conditions to produce the greatest
radiation-induced damage [6], the prototypes were grounded
during 60
Co -rays and proton irradiation tests and during the
post-annealing. However it should be noted that during the
Solar Orbiter mission phase, the duty cycle of EUI will be less
than 20% with long interruptions between solar observation
periods. Moreover during the cruise phase of Solar Orbiter
(approximately 3 years), the EUI detectors will remain almost
all the time unbiased.
The 60
Co irradiations were carried out at the Cyclotron
Research Center (CRC) facility at Louvain-La-Neuve
(Belgium) [7]. Three devices were irradiated with a dose level
up to 150 krad[SiO2] corresponding to the TID in SiO2
expected over the Solar Orbiter mission lifetime (under the
worst case of 1 mm thick aluminum shielding) [8]. In space,
detectors are usually exposed to a low and continuous dose
rate. However, for this accelerated test, 1 krad[SiO2]/h was
recommended by the Evaluation Testing Method (ESCC
22900) [6].
The proton irradiations were performed at the Institut de
Physique Nucléaire et Atomique et de Spectroscopie at the
University of Liège (Belgium). Three APSOLUTE detectors
were tested in a vacuum chamber at the available energies of
10 and 15 MeV, with three different fluence levels up to
4x1011
p/cm2.
The single event effect (SEE) tests were carried out at the
Heavy Ion Facility (HIF) CYCLONE of the CRC. The
radiation sources were high-LET i.e., low penetration
cocktails of ions capable of delivering the required fluence of
107 ions/cm
2 at 5 different LETs up to 67.7 MeVcm
2/mg (cf.
Table I). The flux was around 104 ions/cm
2/s (dosimetry error
bars are within 10%) and was incident normal to the detector
surface.
1(5T, single TX)
2(5T, single TX)
3(5T, single TX)
4(5T, single TX)
5(5T, single TX)
6(5T, single TX)
7(5T, single TX)
8(7T, single TX,
dual SF)
9(6T, dual TX)TXHG (regular)W/L=2.2/0.6
10(6T, dual TX)TXHG (regular)W/L=1.2/0.6
11(6T, dual TX)
TXHG (U-shaped):W/L=2.2/0.6
12(6T, dual TX)TXHG (regular)W/L=2.2/0.6
(FDHG p+ implant)
13(8T, dual TXand dual SF)
14(8T, dual TXand dual SF)
15(8T, dual TXand dual SF)
16(8T, dual TXand dual SF)
64 pixels
64
pix
els
a)
native SiO2
p-Si (epitaxial 2.6 or 2.75 µm)
Backside illumination
Shallow TrenchIsolation (STI)
Si handle wafer (750 µm)
pinned photodiodeFDHG (n+)
SCR
SiO2-SiO2 bonding
TXHG Vout
Ibias
Columni
SFrow select
reset
Vdd (3.3V)
p-well
b)
Field-free region
S (switch)
FDLG
TXLG
p+
(n)
Pixel Array
256 x 256
512 Column Gain Stage
Column Multiplexer (512:1)
Output Stage
SP
I r
eg
iste
rP
ixel C
on
tro
l
Column
Stage Control
MUX Clock
and Sync.
Analog Data Output
External
Pixel Control
Sensor Settings
> REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) <
3
TABLE I PARAMETERS OF THE IRRADIATION TEST CAMPAIGN
Gamma TID krad[SiO2]
1.173 & 1.332 MeV up to 150
Proton Fluence (cm-2) DD (TeV/g)
10 MeV 1x1011 915 60 2x1011 1830 120
4x1011 3660 240
15 MeV 5x1010 390 20 1x1011 780 40
4x1011 3120 170
Heavy Ions LET (MeV.cm2/mg)
N 3.3 0.5 Ne 6.4 1.0
Ar 15.9 2.5
Kr 40.4 6.5 Xe 67.7 10
C. Optical Setup
The optical measurements in the 450-1000 nm spectral
range were carried out at the Detector Measurement
Laboratory (DeMeLab, STCE) using a collimated and tunable
monochromatic light beam emitted by a 100 W tungsten light
source combined with a monochromator and integrating
sphere. The distance between the integrating sphere and the
detectors was kept at a large enough distance for the beam to
be considered approximately normal to the detector surface.
The responsivity in the EUV spectral range was measured
in the laboratory of the Physikalisch-Technische
Bundesanstalt (PTB) at the electron storage ring BESSY II
(Berliner Elektronenspeichering Gesellschaft für
Synchrotronstrahlung) of the Helmhotz-Centre Berlin. The
detector calibration chamber was under ultra-high vacuum
conditions, and manipulation stages allowed us to scan the
sample area and to toggle between test and calibrated
reference detectors [9]. The detector characterizations before
irradiation are reported elsewhere [4].
III. RESULTS AND DISCUSSION
A. Dark current
The dark current (DC), expressed in e-/s/pix, is the signal
measured in the absence of incident photons and is a limiting
factor for long-time exposures, where DC shot noise forms a
dominant source of noise.
The DC measurements between two irradiation doses were
performed in air inside an optical black box. The detector was
held in place by a holder, which was electrically isolated from
but thermally connected to a thermoelectric cooling system. A
calibrated resistance thermometer (Pt100) was added close to
the detector surface to better estimate the absolute temperature
error, which is estimated to be around 1K. The variation of the
DC (cf. inset of Fig. 3b) was measured over the temperature
range with the detector mounted inside a stainless steel
vacuum chamber. The refrigerated/heating circulator system
(model Huber Unistat 405) was used to thermalize the detector
through a copper cold finger. This setup allows us to stabilize
the temperature with ± 0.05 K.
The mean DC is derived from the linear fit of the integrated
signal versus the integration time (IT) from 0% up to 50% of
the saturation level [4]. In case of a non-linearity is observed
(typically at low IT during the irradiation tests), the best-fit
straight line is performed on the slope region up to 80% of the
saturation level. Note that during integration time, the TX low
level is 0V (high level is 3.3 V), which still provides anti-
blooming during exposure.
To minimize the influence of the temporal noise,
measurements were performed by taking 100 frames for each
IT steps. To convert the signal from digital number (DN) to e-,
two different methods were used to extract the detector
conversion gain ( ): the Photon Transfer Curve [10] and the
mean-variance analysis [11]. For both methods, the
measurements were performed with fixed and high-intensity
light sources and short exposure times to negate the DC
contribution [4].
1) Total Ionization Dose-induced DC
Fig. 3(a) shows the mean DC (and distribution) measured at
room temperature as a function of the TID. The DC shows a
relatively small increase i.e., 1 to 2 e-/s/pix/krad up to 110
krad[SiO2]. The variation of the mean DC below 110
krad[SiO2] is not fully understood and could be related to the
non-stabilization of the detector temperature. Note that the
maximum duration for the characterization of the three
APSOLUTE detectors between two -ray doses was only two
hours to avoid any annealing effect [6].
Different mechanisms could contribute to the DC increase
related to the ionization effect in oxides (e.g., SiO2) [12]-[16].
Intensive studies have recently been performed to understand
and mitigate the TID-induced effect on 4T PPD image sensors
[17]-[19]. They all report a linear DC increase as a function of
-rays exposure level in contrast to our measurements. Thus
the small TID-induced DC below 110krad[SiO2] might be
related to a combination of factors, such as the irradiation
condition (e.g., dose rate [20], applied bias condition [21]) and
the pixel design (e.g., geometry, doping concentration, BSI
approach).
Since the scaling of the process technology (here 0.18 m) has
already demonstrated the hardness of MOSFET gate oxides up
to 1 Mrad[SiO2] [22], major sources of leakage current are
due to radiation-induced defects located in the Shallow Trench
Isolation (STI) [23]-[25]. As in any PPD CMOS image sensor
process, the STIs are protected by a p-well, which has proved
to be robust against STI-induced leakage [26]. For
APSOLUTE, the space charge region (SCR) ends inside the p-
well region and depending on the concentrations, the SCR is
about 30 to 50 nm away from the STI (cf. Fig. 1b).
Above 110 krad[SiO2], the DC sharply increases up to at least
a factor two at 150 krad[SiO2]. The sudden increase is not yet
understood. During the TID irradiation, we could observed a
decrease in the FWCLG as reported in [18],[27] but after
annealing at room temperature, the FWCLG increases back to
its pre-irradiation level (within 2% error bar). For APSOLUTE
prototypes, the FWC is dependent on the TX size and shape,
and it is limited by the TX sub-threshold current and not only
> REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) <
4
by the size (area) of the photodiode [4], [28]. Ongoing
investigations have already shown that the TX leakage could
be reduced significantly with an additional implant in the FD
region. This approach strongly reduces the dependence of the
FWC as a function of the photon flux and is ready to be
implemented for the EUI FM detectors.
Fig. 3. (a) Mean DC at room temperature as a function of the TID (pixel
design #12) and (b) after irradiation, as a function of elapsed time during
annealing (in hours). The inset of a) shows the DC distribution as a function of the TID and the inset of b) shows the mean DC as a function of 1000/T for
APT-BSI-642-019 (irradiated up to 150 krad[SiO2]). The detectors were
grounded during the irradiation and post annealing.
After irradiation (see Fig. 3(b)), the DC continues to increase
during the first 12 hours showing a reverse annealing which is
generally attributed to an increase of new interface states with
time created after irradiation in the MOS oxides (see page 176
in [12]). After 24 hours the DC recovers at an exponential
decay with time during two weeks at room temperature. After
a two-week period, the DC levels off and remains almost
constant, however it did not completely return to its pre-
irradiated value (~ +40%).
The inset of Fig. 3(b) shows the mean DC Arrhenius plots
after the 60
Co -rays irradiation tests and annealing. At low
temperature and depending of the pixel design, the extracted
activation energy (Ea) is very small (e.g., 0.05 eV for pixel
design #12). Before irradiation, the same DC behavior has
been observed [4], which does not correspond to the
Shockley-Read-Hall (SRH) generation mechanism. This
leakage current, which is dominant at low temperature, is still
unclear and our current hypothesis is that process-like direct
(band-to-band) tunneling under the TX transistor (with 7 nm
thick SiO2 gate oxide) could explain this unusual behavior.
The dark signal (of individual pixel) shows also temporal
fluctuations after TID irradiation due to the recombination-
generation centers at the Si-SiO2 interfaces [29], [30]. As
shown in Fig. 4, this effect is integration time (IT) and
temperature dependent, and therefore, it is referred as the dark
current random telegraph signal (DC-RTS). The DC-RTS
should not be confused with the MOSFET or SF-RTS [31].
Fig. 4. Dark signal RTS (e-) at 20°C of one pixel (pixel design #9 and #12)
of APT-BSI-642-019 at three different integration time after TID tests and
annealing. The inset shows the RTS of one pixel (pixel design #10) as a function of the temperature between +20°C and +40°C.
2) Displacement Damage-induced DC
For the available proton energies i.e., 10 and 15 MeV,
Coulomb and elastic nuclear interactions are mainly
responsible for displacement damage [32]. The Non-Ionizing
Energy Loss (NIEL, expressed in MeV cm2/g) describes the
rate of the particle energy loss due to atomic displacements
inside the material through nuclear elastic/inelastic reactions
[33]-[36]. The product of the NIEL and the particle fluence
gives the non-ionizing (displacement damage) dose.
As an example, Fig. 5 shows the DC distribution as a function
of the proton fluence (at 15 MeV proton) and after 1 week
post-annealing at room temperature. The mean DC, measured
just after the proton irradiations, increases almost linearly with
the displacement damage (DD) dose. The high DC value
observed during proton irradiation (see histogram at 4
1011
p/cm2
in Fig. 5) is partially attributed to the collisions of
the incident protons with the Si nuclei leading to the creation
of Frenkel defects, such as the creation of vacancy-interstitial
pair atoms [12]. After 1 week post-annealing, the mean DC
decreases by a factor 3.
The inset of Fig. 5 shows the mean DC Arrhenius plots after
the proton irradiation tests and annealing. The extracted Ea is
located close to the Si mid-gap, i.e., 0.70 eV at high
temperature and 0.62 eV at lower temperature, which suggests
0 20 40 60 80 100 120 140 1600.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
103
100
101
102
103
104
pre-irradiation
150 krad[SiO2]
130 krad[SiO2]110 krad[SiO
2]
Dark Current (e-/s/pix)
Pix
el C
ou
nt
60 krad[SiO2]
APT-BSI-642-019
a)
1.5 to 2 e-/s/pix/kradMea
n D
ark
Cu
rren
t (k
e-/s
/pix
)
Pixel design #12
APT-BSI-642-011
APT-BSI-642-012
APT-BSI-642-019
Total Ionisation Dose (krad[SiO2])
TIDthreshold
=110 krad[SiO2]
0 100 200 300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.210
0
101
102
103
104
105
106
107
Ea2
= 0.0125 eV
Ea2
= 0.05 eV
Mea
n D
ark c
urr
ent
(e-/
s/p
ix)
1000/T (K-1)
APT-BSI-642-019
150 krad[SiO2] + annealing
Pixel design #10
Pixel design #12
Ea1
= 1.12 eV
reverse annealing
b)
M
ean
Dar
k C
urr
ent
(ke-
/s/p
ix)
RT: Room Temperature
at 100°C
Annealing Time (h)
RT
0 20 40 60 80 100 120 140 160 180 200
2.0x103
4.0x103
6.0x103
8.0x103
1.0x104
0 20 40 60 80 100 120 140 160 180
1000
1500
2000
+40°C
+30°C
+20°C
Dar
k S
ignal
(a.
u)
Time (min)
one pixel (design #10), IT=0.59s
pixel design #9
pixel design #12
IT=0.59s
IT=1.77s
IT=1.18s
Dar
k S
ign
al (
e-)
at 2
0°C
Time (min)
APT-BSI-642-019
> REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) <
5
that the SRH generation emission is the dominant mechanism
over the whole temperature range i.e., between +100°C and -
35°C. However to learn more about the exact nature of these
defects that exhibit distinctive DC generation rates, specific
measurements, characterization and annealing are required
that fall beyond the scope of this study.
Fig. 5. Dark current distribution at room temperature of APT-BSI-642-017
(pixel design #12) as a function of the proton fluence at 15 MeV. The inset shows the mean DC as a function of 1000/T. Note that the detector was
grounded during the irradiation and post annealing.
Fig. 6 shows the thermal generation rate increase i.e., DC
increase normalized to the pixel depletion volume with the
universal dark current damage factor (Kdark)[37]. The large
deviation between the experimental results and the Kdark curve
(underestimated by a factor 3) cannot be dominantly attributed
only to the ionization damage contribution induced by protons.
Fig. 6. Mean increase of the thermal generation rate of APT-BSI-642-017 as a
function of the displacement damage dose. The red line corresponds to the universal damage factor Kdark at 300K after 1 week post-annealing [37]. The
inset shows a close view of the dark signal RTS (IT=0.345s) at 40°C of one pixel (pixel design #12) after the proton irradiation tests and annealing.
Indeed to allow comparison with the 60
Co -rays DC results,
the proton TID level has been estimated (not shown). As an
example, at 40 krad[SiO2] there is more than two orders of
magnitude between the mean DC of APT-BSI-642-017 i.e.,
41.35 ke-/s/pix estimated from the deviation of the Kdark curve,
and the mean DC measured during the 60
Co -rays irradiation
on APT-BSI-642-019 i.e., 319 e-/s/pix. Note the intra-
columnar recombination [16] is neglected, which is an
important parameter to be accounted for when comparing the
effect of the two different radiation sources.
The inset in Fig. 6, show the DC-RTS of pixel design #9 and
#12. In contrast with the RTS observed after the 60
Co -rays
irradiation tests (cf. Fig. 4), we observed more discrete RTS
levels of the dark signal due to meta-stable generation centers
located in the SCR [29]. The DC-RTS induced by ionization
and by displacement damage processes will be studied in more
detail in future works since it may impact the low-light
sensitivity of the EUI instrument.
B. External Quantum Efficiency (EQE)
The external quantum efficiency (EQE) is defined as the ratio
of the number of detected photons per second, , to
the number of incident photons, , measured by the
calibrated reference detector. is computed by
summing over the beam region in DN and dividing by the
effective conversion gain ( ) as follow:
[ ⁄ ] ∑
with,
[ ⁄ ] [ ⁄ ] [ ⁄ ]
where is the DN value of the ith
pixel corrected from the
detector offset in the region of interest (ROI), is the
conversion gain measured in the visible range and is the
external quantum yield defined, at a given wavelength, as the
ratio of the number of collected photoelectrons to the number
of incident photons. In the EUV range, is higher than
one as shown in the inset of Fig. 7(b). The EQE was measured
by estimating at each wavelength and to reduce the
random error, 100 frames were taken per IT at regular
intervals. The same measurements and analysis were repeated
under ‘dark’ conditions to remove the DC.
1) TID-induced EQE
As an example, Fig. 7(a) shows the EQE of APT-BSI-642-
012 between 450 and 1000 nm. The EQE was measured
several weeks after the 60
Co -ray irradiation and it decreases
by 2% at 500 nm, 24% at 600 nm and around 18% between
650 and 900 nm. A different wavelength dependency with a
stronger decrease towards the UV range has been reported in
similar image sensor technology [18],[26]. However, as shown
in Fig. 7(a), the EQE below 500 nm does not seem to be
affected by the 60
Co photons. If the measurements between
450 and 500 nm are correct, it implies that the TID-induced
defects do not impact the APT-BSI-642-012 EQE.
Note that APSOLUTE detectors are not fully depleted (cf.
Fig. 1(b)). For APT-BSI-642-012, the field-free region
thickness is approximately 950 nm (Siepi = 2.75 m). The inset
of Fig. 7(a) shows APT-BSI-642-019, which has a thinner
102
103
104
105
106
10-3
10-2
10-1
100
101
102
103
2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.210
0
101
102
103
104
105
106
107
Ea2
= 0.62 eV
Ea1
= 0.70 eV
Mea
n D
ark
cu
rren
t (e
-/s/
pix
)
1000/T (K-1)
4x1011
p/cm2 (15 MeV) + annealing
Pixel design #12
pre-irradiation
5x1010
p/cm2
1x1011
p/cm2
4x1011
p/cm2
Norm
aliz
ed P
ixel
Co
un
t
Dark Current (e-/s/pix)
APT-BSI-642-017 (15 MeV)
1 week annealing (RT)
102
103
104
101
102
103
104
105
106
30 40 50 601000
1100
1200
Dar
k S
ign
al (
DN
)
Time (min)
IT=0.345s, +40°C
Ther
mal
Gen
erat
ion R
ate
(e-/
s/
m3)
data (APT-BSI-642-017, Pixel design #12)
Kdark
= 0.19 e-m
-3s
-1(TeV/g)
-1
Non-ionizing absorbed dose (TeV/g)
> REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) <
6
field-free region of 800 nm (Siepi= 2.6 m) and as a
consequence, starts to show a stronger decrease of its EQE
(e.g., 12% at 500 nm) between 450 and 500 nm following the 60
Co -rays irradiation test.
An important remark is that APSOLUTE detectors do not
have a passivation layer, only a SiO2 native oxide layer on top
of the illuminated surface. Moreover in the BSI approach, the
p+ implant layer which could reduce the surface generation
DC and improve the near UV responsivity is at the bottom of
the Si epilayer i.e., facing the SiO2-SiO2 wafer bond and the
handle wafer (see Fig. 1(b)).
Fig. 7(b) shows the absolute EQE of APT-BSI-642-012 in
the EUV range. The EQE decreases by almost 38 % at 17.4
nm after the 60
Co -ray irradiation test. At this wavelength, the
absorption length i.e., the inverse of the absorption coefficient
in Si is about 520 nm [38] and is equal to that of the visible at
around 490 nm.
Fig. 7. EQE of APT-BSI-642-012 (pixel design #10) before and after 60Co γ-
ray irradiation tests a) between 450 and 1000 nm and b) between 15 and 20 nm. The inset of a) shows the APT-BSI-642-019 (pixel design #10) EQE
between 450 and 1000 nm, and inset of b) shows the theoretical and
experimental between 13 and 20 nm.
In such a situation where the absorption length is shallower
than the junction depth, the probability of the surface
recombination is governed by the absorption coefficient of the
material [39] since the density of the photogenerated electrons
and holes becomes higher as the photon penetration depth
becomes shorter. Based on this theory, the stronger decrease
of the EQE in the EUV range cannot be explained by the
surface recombination loss. One of the possible explanations
for the cause of the large change in EQE over the EUV range
may be because of a change in surface layer thickness
(oxidation) or surface contamination since the oxide is
transparent in the visible but absorptive in the EUV and
therefore a slight change in thickness or contamination can
result in a large change in EUV transmittance.
2) Displacement Damage-induced EQE
The EQE was also measured after the proton irradiation tests
(see Fig. 8) where it is found to decrease by approximately
65% at 17.4 nm. Here, the decrease can be explained by the
fact that the proton irradiation has strongly affected the charge
collection within the pixel i.e., SCR due to bulk displacement
damage, increasing the probability of recombination of the
photo-generated carriers. Although precaution was taken, a
surface contamination should not be excluded.
Fig. 8. Absolute EQE of APT-BSI-642-018 (pixel design #12) between 5 and 20 nm wavelength before and after proton tests at 10 MeV (4x1011 p/cm2) and
1 week annealing at RT. The inset shows the normalized FWCLG and GLG of
APT-BSI-642-017 (pixel design #12) as a function of the fluence at 15 MeV proton energy. The filled symbols are the values after 1000 h annealing at RT.
As shown in the inset of Fig. 8, the FWC decreases by
almost 17% at high displacement damage dose, suggesting a
reduction of the doping concentration in the PPD [40] or a
shift in the TX transistor threshold voltage [28] though
ionization effect. After 1000h annealing at room temperature
the FWC increases, but this is still around 10% below the pre-
irradiation level. Note that we did not observe (within the 5%
error bar) any change in the mean value of the detector
conversion gain ( ) after 60
Co -rays (see inset of Fig. 6(b)) or
after the proton irradiation tests (see inset of Fig. 8). The fact
that is unchanged suggests that the FD capacitances and the
CMOS electronics i.e. the column readout circuitry, have not
been affected by the irradiation tests.
500 600 700 800 900 10000.0
0.2
0.4
0.6
0.8
1.0
500 600 700 800 900 10000.0
0.2
0.4
0.6
0.8
1.0APT-BSI-64
2-019 (Si
epi=2.6 m)
pre-irradiation
150 krad[SiO2] + annealing
EQ
E
Wavelength (nm)
a)
APT-BSI-642-012 (Si
epi=2.75 m)
pre-irradiation
150 krad[SiO2] and annealing
EQ
E
Wavelength (nm)
5 10 15 200.0
0.2
0.4
0.6
0.8
1.0
1.2
13 14 15 16 17 18 19 200
10
20
30
Theoretical QYext
(Si)
Qu
antu
m Y
ield
(e-
/ph)
pre-irradiation
after irradiation / annealing
Wavelength (nm)
EQ
E
APT-BSI-642-012 (Si
epi= 2.75 m)
pre-irradiation (EQE17.4nm
=52%)
150 krad[SiO2] and annealing (EQE
17.4nm=32%)
Wavelength (nm)
b)5 10 15 20
0.0
0.2
0.4
0.6
0.8
1.0
1.2
0 1x1011
2x1011
3x1011
4x1011
0.8
0.9
1.0
1.1
1.2
GLG
FWCLG
Fluence (p+/cm
2)
No
rmal
ized
F
WC
, G
APT-BSI-642-017 (Pixel design #12)
FWCLG
1000 h annealing
GLG
1000 h annealing
EQ
E
APT-BSI-642-018 (Si
epi= 2.6 m)
pre-irradiation (EQE17.4nm
= 62 %)
4x1011
p/cm2 (10 MeV) and annealing (EQE
17.4nm= 22 %)
Wavelength (nm)
> REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) <
7
C. Heavy Ion Single Event Effects (SEE)
The SEE is of increasing concern to the space community
due to damage of cosmic rays and solar energetic particles.
During the heavy ion irradiation tests, three APSOLUTE
detectors were exposed and biased with the dedicated
electronic board allowing video recording inside the vacuum
chamber. The image sensor was running on a mains power
supply (VDD) of 3.3V and the pixel array was controlled by
the pixel control block (see in Fig. 2). As a particular case of
SEE, single event latch up (SEL) is a serious occurrence, since
it is long-lived and potentially destructive. During the
irradiation no SEL has been detected on the three APSOLUTE
detectors up to a LET of 67.7 MeVcm2/mg (which is assumed
to be constant within the detector thickness). However single
event failure interrupt (SEFI) and single event transient (SET)
were detected as shown in the inset of Fig. 9 (bottom panel)
but without causing any long-term effect. Indeed, SEFI is a
type of anomaly (soft latch) caused by a single ion strike
leading to a non-functionality of the sensor that fails to
respond to control signals (block condition) and this state may
last as long as the power is maintained [12]. Cycling of the
power supply voltage was sufficient to restore normal
operation. SEFI is most likely to be linked to the serial
peripheral interface (SPI) bus setting being corrupted by heavy
ions and thus different sensor functionality errors may appear.
For instance, gain, offset or biasing settings may be drastically
changed.
The APSOLUTE sensitivity (occurrence yield) is shown in
Fig. 9 where the heavy ions cross section curve is defined as
the number of SEFI events divided by the particle fluence
(ion/cm2). To determine the saturation cross section (σsat), data
are fitted using a Weibull distribution [42].
Fig. 9. Cross section curve of APT-BSI-642-020 (pixel design #9 to #12) for SEFI as a function of the particle LET at +35°C. The inset shows successive
dark frames (rolling shutter) for pixel design #9 to #12 (4x64x64 pixels)
before (top) and after (bottom) a SEFI leading to a block condition (white screen). The ion is Kr with LET = 40.4 MeV.cm2/mg and irradiations were
performed at normal incidence.
The inset of Fig. 9 shows also the presence of hot pixels
created before the irradiation tests [4]. It has been confirmed
that the hot pixels are introduced by metal contamination
during the SOI CMOS processing, whereby the contaminated
particles are stuck in the pixel (around the STI sidewall) and
therefore contribute significantly to the DC. The EUI FM
detectors should not suffer from the hot pixel problem due to
improved solutions including a different pixel design and
CMOS processing which has already been investigated with
promising results [41].
IV. CONCLUSIONS AND RECOMMENDATIONS
Backside illuminated CMOS APS prototypes with different
pixel designs have been fully characterized regarding dark
current and responsivity in the visible and EUV spectral
ranges before and after grounded irradiation tests. The
radiation hardness has been investigated by exposure to 60
Co
-rays, protons and heavy ions.
Even if the physical degradation mechanisms are not yet
fully understood, for EUV applications the minimization of
the detector surface defects and the recombination losses is
crucial. In order to prevent the growth of a native oxide, a
backside surface treatment including a buried backside p+
implant should be applied to the EUI FM detectors to reduce
the DC and improve the detector optical performance. In
addition to minimize the DC, the leakage current observed
before irradiation and probably from tunneling effect should
be cancelled or significantly reduced.
DC-RTS was not investigated in detail in this work but is
strongly recommended for future studies of the EUI FM
detectors. DC-RTS might become a serious limitation during
low-light level solar observations.
For APSOLUTE, the sensors were grounded during the
Co60 -rays and proton irradiation tests because the processing
board was assumed to be not radiation tolerant. However
future irradiation tests and post annealing shall be performed
preferentially under operating conditions i.e., the worst-case
bias condition. In addition, for displacement damage tests,
higher proton energies (e.g., 30 and 60 MeV) shall be
performed to assess also the nuclear interactions contribution.
Although robustness to single event latch-up has been
demonstrated up to 67.7 MeVcm2/mg, SEFI effects must be
properly analyzed. It is recommended to implement a triple
redundancy concept for the EUI FM detectors, which will
largely reduce SPI setting errors caused by single event upset.
In addition, during EUI solar science operations, it is
recommended to frequently update the SPI setting after
acquiring a certain amount of frames to ensure the corrupted
data is removed and corrected from time to time.
Finally the irradiation test campaigns of the APSOLUTE
developments give promising results for EUI detector
fabrication which impose the use of large, thinned arrays for
back-side illumination, namely formats of 2048 x 2048 pixels
for the two high resolution imagers and 3072 x 3072 pixels for
the full Sun imager, all with a 10 µm pixel pitch.
ACKNOWLEDGMENT
We would like to thank the EUI detector working group
members. A special thank is addressed to Dr Matthew West
0 10 20 30 40 50 60 7010
-4
10-3
10-2
#9 #10 #11 #12
#9 #10 #11 #12
LET (MeVcm2/mg)
Cro
ss S
ecti
on (
cm2)
APT-BSI-642-020 (+35°C)
SEFI
fit (sat
= 0.96 mm2, LET
th=15.9 MeV.cm
2/mg)
> REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) <
8
(ROB) for reviewing the paper.
REFERENCES
[1] http://eui.sidc.be/ [2] J.-P. Halain, P. Rochus, T. Appourchaux, D. Berghmans, L. Harra, U.
Schühle, F. Auchère, A. Zhukov, E. Renotte, J-M. Defise, L. Rossi, K.
Fleury-Frenette, L. Jacques, J-F. Hochedez, and A. BenMoussa, "The technical challenges of the Solar-Orbiter EUI instrument", Proc. SPIE
7732, pp. 77320R, 2010.
[3] X. Wang, B. Wolfs, J. Bogaerts, G. Meynants, and A. BenMoussa. “High dynamic range (HDR) backside illuminated (BSI) CMOS image
sensor for extreme UV detection”, Proc. SPIE 8298, pp. 82980B, 2012.
[4] A. BenMoussa, B. Giordanengo, S. Gissot, G. Meynants, X. Wang, B. Wolfs, J. Bogaerts, U. Schühle, G. Berger, A. Gottwald, C. Laubis, U.
Kroth, and F. Scholze.“ Characterization of backside-illuminated CMOS
APS prototypes for the Extreme Ultraviolet Imager on-board Solar Orbiter”, IEEE Trans. Elec. Dev., 60 (5), pp. 1701-1708, May 2013.
[5] S. A. Suliman, O. O. Awadelkarim, S. J. Fonash, R. S. Ridley, G. M.
Dolny, J. Hao, and C. M. Knoedler. “Electron and hole trapping in the
bulk and interface with Si of a thermal oxide grown on the sidewalls and
base of a U-shaped silicon trench”, Solid State Electr. 46, pp. 837-845,
2002. [6] ESCC Radiation Standards. “Total Dose Steady-State Irradiation Test
Method” ESA/ESCC 22900. http://www.ecss.nl/.
[7] http://www.cyc.ucl.ac.be/ [8] J. Sorensen: “Solar Orbiter Environmental Specification – Issue 3.0”,
TEC-EES-03-034/J, 2010.
Available: http://rapidlibrary.com/files/esa-solar-orbiter-environmental-specification-issue-1-3-pdf_ulczccybq9i89on.html.
[9] F. Scholze, B. Beckhoff, G. Brandt, R. Fliegauf, R. Klein, B. Meyer, D.
Rost, D. Schmitz, and M. Veldkamp, “The new PTB-beamlines for high accuracy EUV reflectrometry at BESSY II”, Proc. SPIE 4146, pp. 72-
82, 2000.
[10] J. Janesick, “Photon Transfer”, SPIE Press Book, pp. 35, 2007. [11] R.P. Sperline, A. K. Knight, C. A. Gresham, D. W. Koppenaal, G. M.
Hieftje, and M. B. Denton, “Read-noise characterization of focal plane
array detectors via mean-variance analysis”, Appl. Spectrosc., 59 (11),
pp 1315-1323, 2005.
[12] A. Homes-Siedle, and L. Adams, “Handbook of radiation effects”,
Second edition, oxford university press, 2002. [13] H.J. Barnaby, “Total-ionizing-dose effects in modern CMOS
technologies”, IEEE Trans. Nucl. Sci., 53, n°6, pp. 3103-3121, Dec.
2006. [14] T. P. Ma, and P. V. Dressendorfer, “Ionizing radiation effects in MOS
devices and circuits”, John Wiley and Sons, 1989.
[15] D. M. Fleetwood, P. S. Winokur, R. A. Reber, T. L. Meisenheimer, J. R. Schwank, M. R. Shaneyfelt, and L. C. Riewe, “Effects of oxide traps,
interface traps, and “borders traps” on metal-oxide-semiconductor devices”, J. Appl. Phys., 73, issue 10, pp. 5058, 1993.
[16] T.R. Oldham. “Total ionizing dose effects in MOS oxides and devices”.
IEEE Trans. Nucl. Sci., 50, n°3, pp. 483-499, 2003. [17] V. Goiffon, C. Virmontois, P. Magnan, P. Cervantes, S. Place, M.
Gaillardin, S. Girard, P. Paillet, M. Estribeau, and P. Martin-Gonthier,
“Identification of radiation induced dark current sources in pinned photodiode CMOS image sensors”, IEEE Trans. Nucl. Sci., 59, n°4, pp.
918-926, Aug. 2012.
[18] V. Goiffon, M. Estribeau, O. Marcelot, P. Cervantes, P. Magnan, M. Gaillardin, C. Virmontois, P. Martin-Gonthier, R. Molina, F. Corbiere,
S. Girard, P. Paillet, and C. Marcandella, “Radiation effects in pinned
photodiode CMOS image sensors: Pixel performance degradation due to total ionizing dose”, IEEE Trans. Nucl. Sci., 59, n°6, pp. 2878-2887,
Dec. 2012.
[19] S. Place, J.P. Carrere, S. Allegret, P. Magnan, V. Goiffon, and F. Roy, “Rad tolerant CMOS image sensor based on hole collection 4T pixel
pinned photodiode”, IEEE Trans. Nucl. Sci., 59, n° 6, pp. 2888-2893,
Dec. 2012. [20] F. Faccio, H.J. Barnaby, X.J. Chen, D.M. Fleetwood, L. Gonella, M.
McLain, and R.D. Schrimpf, “Total ionizing dose effects in shallow
trench isolation oxides”, Microelectron. Reliab., 48, pp. 1000-1007, 2008.
[21] M. Innocent,”A radiation tolerant 4T pixel for space applications”,
presented at the Proc. Int. Image Sensor Workshop (IISW), Bergen, 2009.
[22] M. Gaillardin, V. Goiffon, S. Girard, M. Martinez, P. Magnan, and P.
Paillet,” Enhanced radiation-induced narrow channel effects in commercial 0.18 m bulk technology”, IEEE Trans. Nucl. Sci., 58, n°6,
pp. 2807-2815, 2011.
[23] C. R. Moon, J. J. Jung, D. W. Kwon, J. R. Yoo, D. H. Lee, and K. Kim, “Application of plasma-doping (PLAD) technique to reduce dark current
of CMOS image sensors”, IEEE Elec. Dev. Lett., 28, n°2, pp. 114-116,
2007. [24] H. I. Kwon, I. M. Kang, B.-G. Park, J. D. Lee, and S. S. Park, “The
analysis of dark signals in the CMOS APS imagers from the
characterization of test structures”, IEEE Trans. Elec. Dev., 51, n°2, pp. 178-184, 2004.
[25] V. Goiffon, C. Virmontois, P. Magnan, S. Girard, and P. Paillet,
“Analysis of total dose-induced dark current in CMOS image sensors from interface state and trapped charge density measurements”, IEEE
Trans. Nucl. Sci., 57, n°6, pp. 3087-3094, Dec. 2010.
[26] P. R. Rao, “Charge-Transfer CMOS Image Sensors: Device and Radiation Aspects”, PhD thesis, Deflt University of Technology, The
Netherlands, pp. 102, 2009.
[27] J.P. Carrere, J.P. Oddou, S. Place, C. Richard, D. Benoit, C. Jenny, M. Gatefait, C. Aumont, A. Tournier, and F. Roy, “New mechanism of
plasma induced damage on CMOS image sensor: Analysis and process
optimization”, Sol. St. Elec. 65-66, pp. 51-56, 2011. [28] G. Meynant, “Global shutter pixels with correlated double sampling for
CMOS image sensors”, Adv. Opt. Techn., 2, n° 2, pp. 177-187, 2013.
[29] V. Goiffon, P. Magnan, P. Martin-Gonthier, C. Virmontois, and M. Gaillardin, “Evidence of a novel source of random telegraph signal in
CMOS imager sensors”, IEEE Electr. Dev. Lett., 32, n° 6, pp. 773-775, Jun. 2011.
[30] C. Virmontois, V. Goiffon, P. Magnan, O. Saint-Pé, S. Girard, S. Petit,
G. Rolland, and A. Bardoux, “Total ionizing dose versus displacement damage dose induced dark current random telegraph signals in CMOS
image sensors”, IEEE Trans. Nucl. Sci., 58, n° 6, pp. 3085-3094, Dec.
2011. [31] X. Wang, “Noise in Sub-micron CMOS image sensors”, PhD thesis,
Deflt University of technology, The Netherlands, 2008.
[32] I. Jun, M. A. Xapsos, S. R. Messenger, E. A. Burke, R. J. Walters, G. P. Summers, and T. Jordan, “Proton nonionizing energy loss (NIEL) for
devices applications”, IEEE Trans. Nucl. Sci., 50, pp. 1924-1928, 2003.
[33] E.A. Burke, “Energy dependence of proton-induced displacement damage in silicon”, IEEE Trans. Nucl. Sci., 33, n°6, pp. 1276-1281,
1986.
[34] C.J. Dale, P.W. Marschall, G.P. Summers, E.A. Wolicki, and E.A. Burke, “ Displacement damage equivalent to dose in silicon devices”,
Appl. Phys. Lett., 54, pp. 451-453, 1989.
[35] G. C. Messenger,”A summary review of displacement damage from high energy radiation in silicon semiconductors and semiconductors
devices”, IEEE Trans. Nucl. Sci., 39, n°3, pp. 468-473, 1992.
[36] A. Akkerman, J. Barak, M.B. Chadwick, J. Levinson, M. Murat, and Y. Lifshitz, ”Updated NIEL calculations for estimating the damage induced
by particles and -rays in Si and GaAs”, Rad. Phys. Chem., 62, pp. 301-
310, 2001. [37] J.R. Srour, “Universal damage factor for radiation-induced dark current
in silicon devices”, IEEE Trans. Nucl. Sci., 47, n° 6, pp. 2451–2459,
Dec. 2000. [38] E.D. Palik (Ed.), “Handbook of optical constants of solids”, Academic
Press, New York, pp. 547, 1985.
[39] H. J. Hovel, “ Solar cells in semiconductors and semimetals”, Ed. by R.K. Willardson and A.C. Beer, Academic Press, New York, vol. 11, pp.
24, 1975.
[40] C. Virmontois, V. Goiffon, F. Corbiere, P. Magnan, S. Girard, and A. Bardoux, “Displacement damage effects in pinned photodiode CMOS
image sensors”, IEEE Trans. Nucl. Sci., 59, n° 6, pp. 2872-2877, Dec.
2012. [41] G. Meynants, W. Diels, J. Bogaerts, and W. Ogiers, “Emission
microscopy analysis of hot cluster defects of imagers processed on SOI”,
International Image Sensor Workshop – IISW, Snowbird, Utah, USA, June 12th, 2013.
[42] E.L Petersen, J.P. Pickel, J.H. Adams, and E.C. Smith, "Rate prediction
for single event effects – A critique", IEEE Trans. Nucl. Sci., vol. NS-39, pp. 1577-1599, 1992.