A reverse-flow injection analysis method for the determination of dissolved oxygen in fresh and...
-
Upload
independent -
Category
Documents
-
view
2 -
download
0
Transcript of A reverse-flow injection analysis method for the determination of dissolved oxygen in fresh and...
A reverse-flow injection analysis method for the determinationof dissolved oxygen in fresh and marine waters
Saowapha Muangkaew a, Ian D. McKelvie b,*, Michael R. Grace b,Mongkon Rayanakorn a, Kate Grudpan a, Jaroon Jakmunee a,
Duangjai Nacapricha c
a Department of Chemistry, Faculty of Science, Chiang Mai University, Chiang Mai 50200, Thailandb Water Studies Centre, School of Chemistry, PO Box 23, Monash University, Vic. 3800, Australia
c Department of Chemistry, Faculty of Science, Mahidol University, Bangkok 10240, Thailand
Received 4 March 2002; received in revised form 12 March 2002; accepted 13 March 2002
Abstract
A reverse flow injection method (rFIA) based on the Winkler titration chemistry, is reported for the determination of
dissolved oxygen (DO) in natural waters. Manganese(II) sulfate is injected into a continuously flowing stream of sample
and subsequently merges with a reagent stream of sodium hydroxide and sodium iodide. Manganese(II) hydroxide that
is formed reacts with DO in the sample to form an oxidized manganese hydroxyoxide floc. Addition of 10% sulfuric
acid dissolves this floc, and under acidic conditions, the triiodide ion formed is detected by photometry in a flow
through cell at a wavelength of 440 nm. The method is rapid (48 measurements per h), repeatable (R.S.D. ca. 3%, n�/
3), and has a calculated detection limit of 0.25 mg l�1 (P�/0.001). No interference from nitrite or ferrous ions was
observed at concentrations typically found in natural waters. The method has been successfully applied to on-line
measurement of DO in sediment respiration reactors.
# 2002 Elsevier Science B.V. All rights reserved.
Keywords: Reverse-FIA; Winkler; Iodine; Photometric detection
1. Introduction
Natural waters in equilibrium with the atmo-
sphere typically contain dissolved oxygen (DO)
concentrations in the range from 5 to 15 mg l�1 O2
depending on the water temperature, salinity and
altitude [1]. The DO concentration present in
water reflects atmospheric dissolution, as well as
autotrophic and heterotrophic processes that pro-
duce and consume oxygen, respectively.
Analysis of DO is extremely important in
determining water quality. It provides information
on the biological and biochemical reactions occur-
ring in a water body, and is, therefore, an
important indicator of stream metabolism [2]. In
situ measurements of this parameter can be used as
a primary indicator of water quality, and regula-
* Corresponding author. Tel./fax: �/61-3-990-54558
E-mail address: [email protected] (I.D.
McKelvie).
Talanta 58 (2002) 1285�/1291
www.elsevier.com/locate/talanta
0039-9140/02/$ - see front matter # 2002 Elsevier Science B.V. All rights reserved.
PII: S 0 0 3 9 - 9 1 4 0 ( 0 2 ) 0 0 4 1 4 - 9
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
tory agencies recommend a minimum DO require-ment for maintenance of fish populations, e.g.
DO]/6 mg l�1 or 80�/90% saturation measured
over at least one diurnal cycle [3]. Measurement of
DO is also the basis of many respirometric tests,
including the 5-day Biochemical Oxygen Demand
(BOD5) test.
The classical method for DO determination is
the Winkler titration. In this method, DO oxidizesmanganese(II) hydroxide floc produced in situ, to
form a brown colored, oxidized manganese hydro-
xyoxide [4]. Addition of sulfuric acid in the
presence of excess iodide dissolves this hydroxy-
oxide floc, producing triiodide that is then titrated
with standardized thiosulfate using a starch in-
dicator.
Measurement of DO for BOD, respirometrictests, and in-field studies is more commonly and
conveniently performed using an electrometric
technique, based on either galvanic or voltam-
metric membrane sensors. However, these probes
must be frequently calibrated, often against the
Winkler titration [5], may be insensitive to changes
in oxygen concentrations of less than 1 mg l�1 [2],
and require a sample volume of several millilitersand active sample transport across the gas diffu-
sion membrane for successful measurement. For
these reasons measurement by non-electrometric
techniques may be preferred in some instances to
the use of the DO electrode.
While the Winkler titration, and its various
modifications, offers a robust means of determin-
ing DO [1], the precision of the classical titrationmay be inadequate for the purposes of accurate
respirometric measurements, especially for low
concentrations. In an attempt to address this,
Hartwig and Michael [6] used photometric detec-
tion at 350 nm of the triiodide complex formed in
the Winkler reaction to perform photometric
titrations and achieved precision of better than
0.4% R.S.D. for shipboard measurements.Various suggestions have been made for auto-
mating either the total determination of DO, or
detection of the iodine formed by the Winkler
reaction, based on flow injection analysis. For
example, Novic [5] used a flow injection system
with amperometric detection, and Sakai et al. [7]
used the fluorescence quenching of 2-thionaphthol
to determine I2 produced by the Winkler reaction.In both these methods, the formation of iodine
was performed in batch, off-line, thus avoiding
likely problems associated with the formation of
manganese(II) hydroxide precipitate in the flow
lines of an FIA system. However, because of this
off-line step, neither method would be suitable for
the fully automated determination of DO. Few
FIA methods have been described that measureDO directly. Of these, two involve the reaction of
molecular oxygen with the redox indicator, leuco-
methylene blue and photometric detection of the
product species [8,9], while another is based on the
quenching of the room temperature phosphores-
cence of a sol�/gel metal�/chelate by oxygen [10].
In this paper, we describe the development of a
flow injection system for DO measurement basedon the spectrophotometric detection of triiodide
ion from the Winkler reaction. To avoid the
problem of flow tube blockage by manganese(II)
hydroxide that occurs if Mn2� is continually
merged with sodium hydroxide, we have adopted
a reverse flow (reagent) injection approach in
which Mn2� is injected into a continuous flowing
stream of sample before merging with a mixedreagent stream of NaOH and KI. The oxidized
manganese hydroxide floc that forms is readily
transported through the flow tubing, and is
completely dissolved after merging with the sulfu-
ric acid stream. Triiodide produced is measured
photometrically at 440 nm. The method is both
rapid and reproducible, can be performed using
any simple laboratory or field FIA system withphotometric detection capability, and is eminently
suitable for on-line monitoring applications.
2. Experimental
2.1. Chemicals
All reagents were of analytical-reagent grade.The 5% (w/v) solution of manganese(II) sulfate
was prepared by dissolving 36.2 g of mangane-
se(II) sulfate-4-hydrate in 500 ml ultra pure water
(Continental Water systems). Sulfuric acid (10% v/
v) was prepared by dissolving 56 ml of 96%
sulfuric acid in 500-ml ultra pure water. Sodium
S. Muangkaew et al. / Talanta 58 (2002) 1285�/12911286
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
hydroxide (0.5% (w/v)) and potassium iodide(0.2% (w/v)) were prepared by dissolving 2.5 g
sodium hydroxide and 1.0 g of potassium iodide in
500 ml ultra pure water. An acidic iodide flushing
solution was prepared by mixing 10% (v/v) sulfuric
acid solution and the combined sodium hydroxide
and potassium iodide solution in a 3:1 ratio.
2.2. Preparation of calibration standards of varying
dissolved oxygen concentrations
Solutions containing various DO concentrations
within the range of 0�/14 mg l�1 O2 were prepared
in sealed 1 l bottles, the caps of which were fitted
with a DO probe connected to a YSI Model 55
handheld DO meter, a thermometer, and a tubefor air or nitrogen gas sparging and a sampling
line. Using these bottles, 1 l volumes of ultra pure
water at temperatures from 0 to 50 8C were
sparged with either air or nitrogen gas to obtain
a ranges of different concentrations of DO, as
indicated by the in situ oxygen electrode. Samples
of each water prepared in this way were withdrawn
and the DO concentration determined by Winklertitration [1], which was used as the principal
reference method.
Ideally the FIA DO method should give a linear
response over a wide range of oxygen concentra-
tion values. As this was the case, a simple two
point calibration based on the difference in the
theoretical oxygen saturation values at different
salinities and temperatures would be feasible, inmuch the same manner that theoretical oxygen
saturation values at specified temperatures are
used to calibrate DO electrodes [1]. This would
avoid the need to calibrate against the Winkler
titration value.
2.3. Flow injection system
The flow injection system used for development
of the DO method is shown in Fig. 1. The FIA
system was equipped with two peristaltic pumps
(Ismatec CA4E, 20 rpm) and an electrically
actuated rotary six-port valve (Rheodyne, Model
5301). A three way solenoid switching valve (SV)
(LEE, Model No LFAA 1201618H) was used to
introduce an acidic iodide wash solution between
samples. Valve switching and data acquisition was
performed using Flow Control Software (A-Chem
Technologies, Melbourne) and was run on a 486
computer. A variable wavelength spectrophoto-
metric (Applied Biosystems, Model 759A) with a 6
ml flow cell was used at 440 nm. PTFE (0.8-mm
i.d.) tubing was used for knitted reaction coils
(M1, 2 and 3) and all flow tubing. All reagents
were pumped at a flow rate of 1.4-ml min�1.
3. Results and discussion
Two criteria were applied in determining the
optimum conditions for the flow injection mani-
fold, viz; highest sensitivity and shortest practic-
able analysis time. However, because the method
also involves the formation of an hydroxide floc,
there is the potential for blockage problems within
the flow lines, and this becomes a practical
constraint in the choice of reagents, injection
volume and flow rates.
3.1. Wavelength selection
The maximum absorption wavelength of aqu-
eous iodine solution was measured at 380 nm. At
this wavelength the absorbance of a sample with
8.5 mg l�1 O2 was approximately 1.9, and as this
was near the upper end of the absorbance range of
the detector used (0�/2), it represents the practical
upper limit for determination of DO. Since the
desirable linear detection range for O2 in natural
waters is 5�/15 mg l�1 O2, the sensitivity at 380 nm
was deemed to be too great. Therefore, a wave-
length of 440 nm was selected because under this
condition of reduced sensitivity, measurement of
waters with concentrations as high as 15 mg l�1
O2 could be accommodated. However, to max-
imize the sensitivity and resolution of waters with
low DO concentrations, a wavelength of 380 nm
should be used.
S. Muangkaew et al. / Talanta 58 (2002) 1285�/1291 1287
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
3.2. Effect of injection volume of manganese(II)
sulfate solution on peak height and shape
Injection volumes from 25 to 350 ml were used to
determine the optimum value, i.e. the volume that
gave the largest peak response and a single peak.
Fig. 2 shows that the greatest peak height was
obtained at an injection volume of 50 ml. Doublet
peaks were observed for all injection volumes
larger than this, suggesting that for all but the
smallest injections (25 and 50 ml) of manganese(II)
sulfate, the reaction between DO in the sample
stream and the bolus of manganese(II) hydroxide
that forms, was complete at the leading and
trailing edges, and incomplete in the mid portion
of the injected zone because of a paucity of DO in
this region. While it may be possible to overcome
doublet peak formation by more efficient mixing,
say by the use of a longer mixing coil at M1, this
would be counterproductive because the increased
dispersion would offset any gains in sensitivity.
For these reasons, a 50-ml injection volume was
chosen for all subsequent measurements.
3.3. Effect of mixing coil lengths
Using a fixed injection volume of 50 ml, theinfluence of independently varying the mixing coils
M1, 2 and 3 (Fig. 1) on final signal response and
analysis time was investigated. Increasing the
length of the mixing coil, M1, from 0�/200 cm,
caused further dispersion of the injected volume of
manganese(II) sulfate resulting in a decreasing
peak height (Fig. 3).
Not surprisingly, increasing the length of M2,immediately after the addition of the sodium
Fig. 1. Reverse FIA manifold for DO determination. P1, P2: peristaltic pumps, M1, M2 and M3: knitted reaction coils of specified
lengths, V: injection valve, SV: 3-way solenoid switching value and D; flow-through photometric detector.
Fig. 2. Peak heights obtained from the determination of DO in
air-saturated ultra pure water at 22 8C (8.5 mg l�1 O2) by the
proposed rFIA method at various injection volumes from 25 to
350 ml. Error bars are 9/sn�1 for n�/3.
Fig. 3. Peak heights obtained from the determination of DO in
air-saturated ultra pure water at 22 8C by the proposed rFIA
method with various mixing coil lengths for M1(j), M2 (")
and M3 (').
S. Muangkaew et al. / Talanta 58 (2002) 1285�/12911288
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
hydroxide/iodide reagent stream results in anincreased peak response because additional mixing
promotes the reaction between DO and the
manganese(II) hydroxide.
The dissolution of the oxidized manganese
hydroxide floc following the addition of sulfuric
acid is rapid, and increasing the length of mixing
coil M3 (Fig. 1) had no effect other than to
decrease the peak response and increase the peakwidth (Fig. 3).
On the basis of these observations, mixing coils
M1 and 3 were omitted entirely from the final
version of the flow injection manifold. A length of
100 cm was chosen for M2, and this represents a
compromise between maximizing the sensitivity
and maintaining an acceptable sample throughput.
One additional modification was made to themanifold. Oxidized manganese hydroxyxide floc is
formed on-line after the merging of the sodium
hydroxide/iodide reagent stream with the injected
Mn(II) and sample stream. While the bulk of the
oxidized manganese hydroxyoxide floc is trans-
ported through the manifold tubing, and subse-
quently dissolved by the addition of sulfuric acid,
it was observed that a small residual amount ofbrown precipitate remained in mixing coil M2, and
that this gradually accumulated over time. While
this process had no effect on the reproducibility of
the analyses, if ignored, gradual accumulation
would result in the complete occlusion of the
flow tubing completely. This potential problem
was overcome by incorporating a solenoid SV, in
the NaOH/KI reagent line (Fig. 1), and using thisto introduce an acidic iodide wash solution after
every third injection. This washing process was
controlled by the instrument software, and pro-
grammed such that it occurred during the aspira-
tion of a new sample into the FIA system. An
improvement on this configuration would be to
move the SV to the sample line, which would mean
that sample switching could be achieved withoutthe need to stop the pump, while manifold washing
was occurring.
3.4. Analytical figures of merit
The DO calibration data obtained using the
proposed rFIA system for a DO concentration
range of approximately 2�/13 mg l�1 O2 fitteda linear plot, according to the equation:
Peak Height (Abs)
�0:0414[DO]�0:4784;
for which r2�0:9945 (n�6)
Linearity over the calibration range means that
a two point calibration method can be used, i.e.
using high and low DO standards, whose concen-
trations are defined by the saturation concentra-
tion at particular temperature and salinity values.
This offers the most feasible option for calibration
in the field.It should be noted that there is a substantial
blank signal that originates from DO in the
reagents, and while this could be reduced by
sparging the reagents with nitrogen before use,
this is generally not practicable.
The calculated detection limit was 0.25 mg l�1
O2 (P�/0.001). The relative standard deviation for
the proposed rFIA method was typically approxi-mately 3% for triplicate injections, and a
sample throughput of 48 injections per h was
possible.
3.5. Effect of salinity on the determination of
dissolved oxygen concentration by the proposed
rFIA method
As salinity increases, there is a decrease in the
oxygen solubility. Increasing ionic strength causes
no interference in the classical Winkler titration,
but could conceivably have a kinetic effect on thereactions in a flow injection method. To investi-
gate whether the proposed method is tolerant of
changes in salinity, air-saturated ultra pure water
at 20�/21 8C with a range of different salinities
was analyzed by the proposed method, and
compared with measured values using a DO meter
and the theoretical saturation value for the same
conditions of temperature and salinity. The results(Table 1) show excellent agreement between the
rFIA results, those from the DO meter
(salinity corrected readings) and theoretical
values [1], and indicate that this method is suitable
for DO measurements over a wide range of
salinities.
S. Muangkaew et al. / Talanta 58 (2002) 1285�/1291 1289
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
3.6. Effect of potential interferents on the
determination of dissolved oxygen concentration by
the proposed rFIA method
Nitrite at concentrations greater than 50 mg l�1
NO2�1 is a potential interferent in the Winkler
titration because in acid medium it reacts with
excess iodide to produce iodine and N2O2. Further
exposure to air will produce more nitrite and hence
more iodine [2], and this cyclic process can lead to
significant overestimation of the true DO concen-
tration. In natural waters the nitrite concentration
would seldom be higher than 200 mg l�1. Potential
interference of nitrite on the rFIA DO determina-
tion was studied by adding nitrite in the concen-
tration range 0�/600 mg l�1 NO2� to samples of
oxgenated water. Table 2 shows that the proposed
method is tolerant to nitrite up to a concentration
of 600 mg l�1. This tolerance for relatively high
concentrations of nitrite is not surprising given
that the reactions are all performed on-line in a
closed system that offers no opportunity for
reoxygenation of the sample after final acidifica-tion and subsequent production of triiodide.
Iron(II) at concentrations greater than 1 mg l�1
is also known to interfere in the DO determina-
tion[11], and while the Fe(II) concentration in
most natural waters is generally lower than 1 mg
l�1, we chose to test for interference from this
species over the concentration range of 0�/1.5 mg
l�1. Table 2 shows that DO concentrationsdetermined using the proposed rFIA method
were unaffected by increasing Fe(II) concentra-
tions, demonstrating that there was no interfering
effect of Fe(II) for DO measurements of natural
waters.
3.7. Application to real samples
A method comparison between this proposed
rFIA method and Winkler titration values was
performed on six natural water samples, and astrong correlation of the developed rFIA and
Winkler titration methods was obtained, with a
linear regression equation of:
DOrFIA�0:9136[DO]winkler�0:5849;
r2�0:9926 (n�6)
To evaluate the longer term reliability of the
rFIA method developed, the sample intake line of
the FIA system was connected directly to a stirred
reactor containing tap water that had been deox-ygenated by nitrogen sparging. A DO probe and
the proposed rFIA method were used to determine
DO in the water sample simultaneously over a 7 h
period. After initial deoxygenation, the water in
the reactor was allowed to reoxygenate up until
1500 h, and, thereafter, the temperature of the
Table 1
DO concentrations of air-saturated ultra pure water at 20�/
21 8C with various salinities (0�/35 ppt)
Salinity (ppt) Dissolved oxygen concentration (mg l�1)
Theoretical RFIA (%R.S.D.) DO meter
0 9.04 8.90 (2.2) 8.94
5 8.83 8.63 (1.2) 8.39
10 8.57 8.25 (2.4) 8.18
20 7.93 7.87 (1.2) 7.74
30 7.48 7.16 (1.4) 7.02
35 7.26 7.16 (4.2) 7.04
Table 2
Effect of added nitrite and Fe(II) on DO measurements (mean9/S.D., n�/3)
Nitrite (mg l�1) DO concentration (mg l�1) Fe(II) (mg l�1) DO concentration (mg l�1)
0 8.719/0.31 0.0 7.899/0.27
100 8.669/0.70 0.2 7.859/0.34
200 8.549/0.48 0.4 7.599/0.20
300 8.529/0.24 0.8 7.919/0.46
400 8.649/0.28 1.0 7.699/0.32
500 8.729/0.29 1.2 7.879/0.17
600 8.639/0.26 1.5 8.059/0.05
S. Muangkaew et al. / Talanta 58 (2002) 1285�/12911290
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
reactor was manipulated to obtain a range of DOvalues. Fig. 4 shows that DO concentrations
obtained from the proposed rFIA method closely
as followed the theoretical values over the latter 4
h period of time (DOrFIA�/0.998DOTheoretical�/
0.195, r2�/0.9642, n�/3). However, the DO con-
centrations obtained from the DO probe were
lower than the theoretical values, and essentially
constant during the last 3 h 30 min, an effect that isprobably due to inadequate liquid transport across
the surface of the DO electrode membrane.
Subsequent to this evaluation, the proposed
rFIA DO method has been successfully applied
in the measurement of oxygen uptake kinetics in a
series of sediment respiration studies in both
laboratory and field situations.
Enhancement of the sensitivity, and hence theresolution, of DO over a narrower range of values,
e.g. in anaerobic waters, could readily be achieved
by performing measurements at a wavelength
closer to the lmax of 380 nm.
4. Conclusion
A reverse flow injection analysis technique for
determination of DO in natural water has been
successfully developed based on the Winkler titra-
tion method. The proposed reagent injection
method overcomes manifold blockage problems
of precipitated manganese(II) hydroxide thatwould occur if conventional sample injection
FIA were used, and the method is very tolerant
of the common interferents Fe2� and NO2�. The
DO concentrations obtained using this rFIA
method were in close agreement with the theore-
tical saturation and Winkler titration DO values of
a number of real samples. The method has been
successfully deployed in tracking DO variations in
a stirred reactor over a period of several hours,
and is now being routinely used for this purpose in
our laboratory and in the field.
Acknowledgements
Saowapha Muangkaew thanks the Thai Minis-
try of University Affairs and PERCH (Postgrad-
uate Education and Research Program in
Chemistry) for their support of her research inAustralia and Thailand. Technical assistance by
Mr Peter Ellis is gratefully acknowledged.
References
[1] APHA-AWWA-WEF, Standard methods for the Exam-
ination of Water and Wastewater, 20th ed., American
Public Health Association, Washington, DC, 1998.
[2] R.G. Wetzel, G. Likens, Limnological Analysis, second
ed., Springer, New York, 1991, p. 391.
[3] ANZECC, Australian Water Quality Guidelines for Fresh
and Marine Waters, Australian and New Zealand Envir-
onment and Conservation Council, 1992.
[4] M.L. Lopez, J.A. Santaballa, Education in Chemistry,
1999, pp. 162.
[5] M. Novic, B. Pihlar, M. Dular, Fresenius’ Z. Anal. Chem.
332 (1988) 750.
[6] E.O. Hartwig, J.A. Michael, Environ. Sci. Technol 12
(1978) 712.
[7] T. Sakai, H. Takio, N. Teshima, H. Nishikawa, Anal.
Chim. Acta 438 (2001) 117.
[8] A. Sanz-Martinez, A. Rios, M. Valcarcel, Anal. Chim.
Acta 284 (1993) 189.
[9] V.V. Kuznetsov, M.V. Murasheva, J. Anal. Chem. 51
(1996) 993.
[10] J.M. Costa-Fernandez, M.E. Diaz-Garcıa, A. Sanz-Medel,
Anal. Chim. Acta 360 (1998) 17.
[11] H.L. de Medina, Chemical parameters, in: L.M.L. Nollet
(Ed.), Handbook of Water Analysis, Marcel Dekker, New
York, Basel, 2000, p. 921.
Fig. 4. DO concentrations in reactor measured on-line over a 7
h period (1230�/1930) using the rFIA method and a DO
electrode. Theoretical DO saturation values are shown for
comparison.
S. Muangkaew et al. / Talanta 58 (2002) 1285�/1291 1291
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
VITA
Name: Miss Saowapha Muangkaew
Date of Birth: August 31, 1974
Academic status:
- B.Sc. (Education), Second Class Honors, Prince of Songkla University, 1995
- M.S. (Applied Analytical and Inorganic Chemistry), Mahidol University,
1999
- Ph.D. student in analytical chemistry, Chiang Mai University
Practical Experience:
- Visiting Graduate Student of Water Studies Centers, Monash University,
Australia, 2001
Scholarships
The Ministry of University Affairs (PSU, Surathani Campus)
The Postgraduate Education and Research Program in Chemistry (PERCH)
University Mobility in Asia and the Pacific (UMAP)
List of Publications:
International Journals
1. S. Muangkaew, I.D. McKelvie, M.R. Grace, M. Rayanakon, K. Grudpan,
J. Jakmunee and D. Nacapricha, “A reverse-flow injection analysis method
for the determination of dissolved oxygen in fresh and marine waters”
Talanta. 58, 1285-1291, 2002.
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
104
International Conferences
1. S. Muangkaew, I. D. McKelvie, M. R. Grace, M. Rayanakorn, K. Grudpan,
J. Jakmunee, and D. Nacapricha, “The Development of A Reversed FIA
Method for the Determination of Dissolved Oxygen In Fresh and Marine
Water” The 11th International Conference on Flow Injection Analysis and The
38th Semi-annual Meeting of the Japanese Association for Flow Injection
Analysis, Chiang Mai, 2001.
National Conferences
1. K. Grudpan, R. Chantiwas, S. Muangkaew, P. Aumpan, W. Praditwiengcome,
L. Hamer, J. Jakmunee and M. Rayanakorn “Determination of Some
Anions in Water Sample by Flow injection Dialysis-Ion Chromatography”
25th Congress on Science and Technology, Pitsanuloke, 1999.
2. S. Muangkaew, M. Rayanakorn, K. Grudpan, J. Jakmunee, and D. Nacapricha
“Derivatization of Iodide Prior to HPLC Determination” RGJ Seminar
NO.2 Analytical Chemistry and Chemistry In The North, Chiang Mai, 2000.
3. S. Muangkaew, M. Rayanakorn, K. Grudpan, J. Jakmunee, and D. Nacapricha
“Development of High Performance Liquid Chromatography For The Determination
of Iodide at Trace Levels” 26th Congress on Science and Technology, Bangkok, 2000.
4. S. Muangkaew, M. Rayanakorn, K. Grudpan, J. Jakmunee, and D. Nacapricha
“Effect of Ionic Strength on the Determination of Iodide by Flow Injection
Potentiometry” 27th Congress on Science and Technology, Songkla, 2001.
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
105
5. S. Muangkaew, I. D. McKelvie, M. R. Grace, M. Rayanakorn, K. Grudpan,
J. Jakmunee, and D. Nacapricha, “Performance Test of the rFIA System
Developed For Dissolved Oxygen Determination” 28th Congress on
Science and Technology, Bangkok, 2002.
6. S. Muangkaew, J. Jakmunee, S. Lapanantnoppakhun, M. Rayanakorn,
D. Nacapricha, and K. Grudpan, “Determination of Iodide by Flow Injection
In-Valve Derivatization with High Performance Liquid Chromatography”
29th Congress on Science and Technology, Khon Kaen, 2003.
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d
ÅÔ¢ÊÔ·¸Ô ìÁËÒÇÔ·ÂÒÅÑÂàªÕ§ãËÁèCopyright by Chiang Mai UniversityA l l r i g h t s r e s e r v e d