Pengevaluasian ketaatan pedoman proteksi radiasiyang dilakukan pada semua FKG di IIran
Intraoral
cross sectionaL ; Quesioner
Kurangnya kontrol kualitas secara rutin dan kurangnya jaminan
kesehatan dari fasilitas proteksi radiasi tersebut
EkstraoralKontrol
Kualitas
Perlunya pertimbangan lebih lanjut tentang prinsip-prinsip proteksi radiasi di FKG terutama pada
pengendalian mutu dan jaminan kualitas
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 3/26
Latar Belakang
• Keputusan medis → 30-50%
dental radiologi• Radiasi berbahaya → ALARA
dan kontrol kualitasditingkatkan
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 4/26
Tujuan
• Untuk mengawasi ketaatan dalam
peaksanaan prinsip-prinsip proteksi
radiasi semua FKG di Iran• Untuk memperhatikan dan
menentukan prinsip-prinsip proteksi
IR, ER dan implementasi kontrolkualitas
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 5/26
Bahan dan Metode
• Penelitian dilakukan dengan
metode potong melintang →
september 2008 dan februari 2009→ 18 FKG di Iran
• Kuisioner : → IR → ER → kontrol
kualitas
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 6/26
Intraoral radiography N (%)
Film speed D-Speed
0 ( 0 )
D&E Speed
3 (16.7)
E-Speed
15 (83.3)
F-Speed
0 (0)
PID type Round
16 (88.9)
Rectangular
0 (0)
Round or rectangular
2 (11.1)
Cone
0 (0)
PID length < 8 inch
1 (5.5)
8 inch
7 (38.9)
12 inch
7(38.9)
≤ 16 inch
3 (16.7)
Protective
covers
Lead apron only
11 (61.1)
Thyroid collar only
0 ( 0 )
Both
7 (38.9)
None
0 (0)
Protective
covers usage
Pregnant women
11 (61.1)
Pregnant women&childs
2 (11.1)
Pregnant women,childs
& Full mouth series
5 (27.8)
All patients
0 (0)
Performing the
radiography
Students only
4 (22.2)
Radiology technicians
only 1 (5.6)
Dentists only
2 (11.1)
Occasionally
11 (61.1)
Holding disable
patients
Dentist
0 (0)
Patient's attendances
18 (100)
Technician of
radiology0 ( 0 )
nurses
0 (0)
kVp setting < 60 KVp
0
60 KVp
2 (11.1)
61-70 KVp
12 (66.7)
> 70 kVp
4 (22.2)
Tube filtration 1 mmAl ≤
0 ( 0 )
1.5 1 mmAl
8 (44.4)
2 1 mmAl
10 (55.6)
> 2 mmAl
0 (0)
exposure room
protection
leaded wall
16 (88.9)
Distance – direction
law
0(0)
Adequate thickness of
gypsum or concrete
2 (11.1)
lead partitions
0 (0)
Table 1. Status of dose saving practices in intraoral radiography (IR) practices in Iranian dental schools.
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 7/26
• Intraoral
•
PID = bulat dan dengan panjang 8 inch dan12 inch dari sumber film.
•
14 fkg –
fasilitas radiograpy digital- tidak rutindigunakan,hanya pada saat penelitian.
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 8/26
Table 2. Status of dose saving practices in extraoral radiography (ER) practices in Iranian dental schools.
Extraoral radiography N (%)
Film-screen
Speed
100 ≤
0 (0)
200-300
0 (0)
400 ≥
18 (100)
Digital radiography
0 (0)
Protective
Covers
Lead apron only
15 (83.3)
Thyroid collar only
0 (0)
Both
3 (16.7)
None
0 (0)
Protectivecovers usage
Pregnant women12 (66.6)
All patients4 (22.2)
Pregnant women &Childs 1 (5.6)
None1(5.6)
Performing the
Radiography
Students only
2 (11.1)
Dentists Only
2 (11.1)
Technician of
radiology only
11 (61.1)
Occasionally
3 (16.7)
Holding disable
Patients
Dentist
0 (0)
Patient's attendances
18 (100)
Technician of
radiology 0 (0)
nurses
0 (0)
kVp setting < 60 KVp
0
60 KVp
0 (0)
61-70 KVp
9 (50.0)
> 70 KVp
9(50.0)
Tube
Filtration
1 mmAl ≤
0 (0)
1.5 1 mmAl
0 (0)
2 1 mmAl
8 (44.4)
> 2 mmAl
10 (55.6)
Exposure room
protection
Leaded wall
18 (100)
lead partitions
0 (0)
Adequate thickness
of wall 0 (0)
Distance – direction
Law 0(0)
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 9/26
• Ekstraoral
• Semua fkg menggunakan intensifying screens
(100%).
• Leaded rubber aprons – apron wanita hamil -
baik IR maupun ER – semua fkg kecuali satu.
• Paling sering – proteksi dari dinding ruangan
yang berlapis timbal(Pb) di IR dan ER.
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 10/26
Table 3. Status of dental school regarding to quality control periods.
Quality control periods
Monthly monthly - irregulary Not done
or shorter Anually
B+F density of intraoral films 1 (5.6) 0 (0) 6 (33.3) 11 (61.1)
B+F density of extraoral films 5 (27.8) 0 (0) 6 (33.3) 7 (38.9)
Darkroom lighting and safelight condition 2 (11.1) 1 (5.6) 10 (55.5) 5 (27.8)
Output measurements of x-ray sets 0 (0) 2 (11.1) 9 (50.0) 7 (38.9)
Monitoring of received dose by personnel 2 (11.1) 16 (88.9) 0 (0) 0 (0)
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 11/26
• Quality control periods hanya beberapa
sekolah
• Pengukuran dosis radiasi reguler menggunakan
film badges dan prosesor otomatis sangat
efektif
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 12/26
Table 4. Agreements of dental school status with guidelines of EC and NRPB in different aspects.
Agreement N ( %)
Item
IR ER
1- Utilizing the rectangular Collimator 2 (11.1) --------
2- PID length 17 (94.4) --------
3- Existence of protective covers 18 (100) 18 (100)4- Policy for use of protective covers 5 (27.8) 4 (22.2)
5- Responsible person for radiography 12 (66.7) 16 (88.9)
6- Existence of radiation monitoring systems 17 (94.4) 17 (94.4)
7- Responsible person for holding disable patients 18 (100) 18 (100)
8- Regular measurements of films sensitivity 5 (27.8) 1 (5.6)
9- Measurements of darkroom lighting 3 (16.7) 2 (11.1)
and safelight condition
10- personnel dose received monitoring 17 (94.4) 18 (100)
11- Quality control and output measurements 0 (0) 2 (11.1)
of X-ray sets
12- Responsible person about quality control 13 (72.2) 13 (72.2)programs
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 13/26
• Kesepakatan diperoleh dari beberapa
rekomendasi dari komisi Eropa dan NRPB.
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 14/26
Figure 1. Agreements of dental school status with some of radiation protection guidelines
0
10
20
30
40
50
60
70
80
90
100
Item
Tube filtration in ER
Exposure roomprotection in IRTube filtration in ER
Tube filtration in IR
exsistance of Digital
ERuse of Digital ER
exsistance of DigitalIRuse of Digital IR
ER automaticprocessing system
IR automaticprocessing systemtype of Film screen inERFilm-screen speedinERFilm Speed in IR
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 15/26
Pembahasan ( Diskusi )
Dental Radiography →
kualitas gambar dari rahangdan struktur maxillofacial →
sedikit paparan → proteksi
radiasi
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 16/26
Penggunaan E-speed film radiografi 83,3%
Dengan kombinasi D-speed film 16,3%
Kecepatan radiasi bertambah
Sedikit paparan sinar X
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 17/26
Tidak digunakannya bentuk tubehead“rectangular”
Hanya 11,1%
Tidak membatasi area paparan sinar X
Paparan radiasi berlebihan
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 18/26
Penggunaan 70 kVp sebanyak
88,9%
Voltase terlalu tinggi
Tidak sesuai dengan yang telah
ditentukan (65 kVp)
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 19/26
Penggunaan tubehead dengan
panjang 8-12 inci sebanyak 38,9%
Tubehead panjang (long cone)
Mengurangi radiasi sinar X
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 20/26
Penggunaan apron
Mengurangi radiasi sinar X yangterpancar dan diterima pasien
Banyak dilakukan pada semua pasien yang hamil, baik intraoral maupun
extraoral
Berdampak positif
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 21/26
Penggunaan filtrasi dengan tebal minimal
1,5 mm hanya sebanyak 44,4%
Filtrasi dapat mengurangi paparan radiasisinar X
Tetapi, jarang digunakan
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 22/26
Hampir semua FKG menggunakan
peralatan digital
Namun, jarang digunakan
Dengan alasan : permasalahan
mencetak, jarang membuat arsip
gambar, operator yang kurang mahir, pasien yang tidak terbiasa
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 23/26
Semua FKG memiliki prosesor
otomatisSemua klinik gigi memiliki prosesor otomatis
Penatalaksanaan RadiografiIntraoral lebih cepat
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 24/26
Kontrol kualitas tidak dilakukan di kebanyakanFKG dan klinik gigi
Berupa : pemeriksaan sensitivitas film,
kebocoran cahaya untuk kamar gelap, kondisicahaya yang aman dan pengukuran output unitsinar X
Efektivitas pemeriksaan tidak berkualitas
dan tidak terjamin (unsafety)
5/17/2018 PERSENTASI KELOMPOK RKG2 - slidepdf.com
http://slidepdf.com/reader/full/persentasi-kelompok-rkg2 25/26
Kesimpulan
1. Prinsip-prinsip proteksi radiasi harus
dilaksanakan di setiap FKG dan
klinik gigi
2. Kontrol kualitas harus sering
dilakukan di setiap FKG dan klinik
gigi3. Dilaksanakannya program asuransi
bagi pasien untuk indikasi paparan
sinar X yang berlebihan