Modeling reactive multiphase flow and transport of concentrated solutions
Estimation of the left ventricular ejection fraction using a novel multiphase, dark-blood,...
Transcript of Estimation of the left ventricular ejection fraction using a novel multiphase, dark-blood,...
AJR:169, July 1997 101
Estimation of the Left VentricularEjection Fraction Using a NovelMultiphase, Dark-Blood,Breath-Hold MR ImagingTechnique
Shantanu Sinha1Richard Mather1Usha Sinha1Jonathan Goldin1Gregory Fona row2Hyo-Chun Yoon1
OBJECTIVE. In this paper, we evaluate a recently pro�x)sed dual-phase dark-blood MR se-
quence for estimating the left ventricular ejection fraction, compare Simpson’s method of estinla-
tion of ejection fraction with a model based on the biplane method. assess the reproducibility of
both nlethods, and fiiially. test a semiautomated niethod for contouring the endocardial border.
SUBJECTS AND METHODS. An MR pulse sequence was implen�ented to acquire cardiac
images itt Ifotli diastolic ;uld systolic phases within a single breath-hold. A special magnetization
prepa�ation scheille rendered blood dark while a segmciited acquisition allowed breath-hold scantimes. Five healthy volunteers �md five patients with cardiac disease were imaged. Ejection fractions
were estimated using ( 1 ) long-axis and four-chaniber biplane views with :ui ellipsoid mxlel �uxl (2)
a series of short-axis views in conthination with Simpson’s model. These values of ejection frac-
tions were then compared with values obtained tiom echocardiography.
RESULTS. Estiniates of ejection llactiotls obtained using biplane ellipsoid volume and Simp-
SOIl’S rule methods varied by l4�/ in healthy volunteers. However, for patients with severe car-
dion�yopathy, differences between the values of ejection traction obtained with the two methods
varied by as much as 1 5O�k . Ejection fraction estimates obtained from MR inutges with the hi-
plane ellipsoid ulethod at�l fIonl echocardiography varied by approximately l4�X for all sub-jects. Ejection fraction estiit�ttes obtaitied with the seiniautoinated algoritlini agreed well with
estinlates obtained with ii�anual contours made by experienced radiologists. lntraohserver van-
ability was low for both the short-axis (3(/ ) and biplane (4(/ ) methods. 1-lowever, interobsenver
variability of the biplane n�ethod ( I 2�/ ) exceeded that of the shont-axis method (4�% ). lntenex-
;ti�ination vaniahility (9Y ) was the largest litctor in detenuining the reproducibility of the ejec-
tion fraction estiniates.
CONCLUSION. Breath-hold dark-blood MR inlaging technique with sinlultaneous acqui-
sition of a series of short-axis views during systolic and diastolic Phases pemlits rapid and accu-
rate estimates Of ejection fnactions in healthy subjects and in p�ttients. Model-based biplane MR
iniaging methods are less reliable in patients with global cardion�yopathy. The estinlation of
ejection fractions can he automated using the proposed contouring algonithu� aixl the (lark-blood
short-axis views.
Received May 14, 1996; accepted after revisionJanuary22, 1997.
Presented at the annual meeting of the AmericanRoentgen Ray Society, San Diego, May 1996.
1 Department of Radiological Sciences, UCLA School of
Medicine, Los Angeles, CA 90095-1721. Addresscorrespondence to S. Sinha, Department of RadiologicalSciences, 82-220 CHS, UCLA Medical Center,10833 Le Conte Ave., Los Angeles, CA 90095-1721.
2Department of Cardiology, UCLA School of Medicine,
Los Angeles, CA 90095-1721.
AJR 1997;169:101-1 12
0361-803X/97/1 691- 101
© American Roentgen Ray Society
T he inherent soil-tissue contrast of
MR iniaging conuhined with necent
improvements in hardware and soft-
ware capabilities has significantly increased its
potential as an iniaging technique for non-
invasive assessn�ent of cartliac function. This
increased potential includes detennining left
ventricular v(.)lulfles, ejection fraction, mass,
wall i�otion ahnoniialities, �uid functional mt or-
nlation such as i11yoc�u’dial perfusion in patients
with congenital or acquired heart disease 1 1-41.
Geometry-based methods of estintation of the
left ventricular volume and mass, such as the
single-plane and biphule ellipsoidal niethods.
have leen conventionally used in ech(x.ardiog-
mph)’ (5 . These geometry-based metliids have
also been applied to the assessment of ventnicu-
lat’ volumes 110111 MR imaging in � el’firt to t’e-
duce scan times 161. Though these methods
have the adv;ultage of’ reduced imaging tinles.
they are based 011 the ellipsoidal shell t1�xlel of
the left ventricle, which requires axial symnle-
try ol’ the left ventricle atxut its long axis
throughout the c�trdiac cycle. ;ui assunlptmon of-
ten invalid in patients with ii�yocardial disease.
In contrast, volunle estimates lr’oiii a series of’
short-axis views using Simpson’s rule do not
use any geolnetnc asswnptions :111(1yield accu-
Note-Echo. = echocardiography, N/A � not available.
Sinha et al.
102 AJR:169, July 1997
rate volumes. However, the disadvantage of the
latter technique lies in the relatively longer ac-
quisition times. Each short-axis view requires a
separate acquisition and furthermore, in the esti-
mation of ejection fractions and wall motion ab-
normalities, each view needs to be acquired in
different cardiac phases.
In estimating global cardiac parameters,
contouring of the epicardial and endocardial
borders is an important step. Automated meth-
ods of segmenting the left ventricle and myo-
cardium are desirable not only to reduce the
time and effort that would otherwise be in-
volved in contouring the many images but also
to minimize interobserver and intraobserver
variabilities. Semiautomated methods of myo-
cardial segmentation have been attempted by
various groups, generally using bright-blood
images that depict blood as brighter than
myocardium. The efficacy of segmentation de-
pends essentially on two factors: good blood-
myocardium contrast and uniform blood signal
intensity. Obtaining bright-blood breath-hold
images is relatively easy using a segmented
fast gradient-echo technique, but these images
have variable blood signal intensities that are
dependent on the cardiac phase. Also, blood-
myocardium contrast may be suboptimal in
some images: these images, therefore, are not
always amenable to an automatic contouring
method [4]. Thus, it is desirable to suppress the
blood signal intensity to obtain both high myo-
cardium-blood contrast and uniformity of
blood signal intensity to facilitate automation
of endocardial border detection.
We recently proposed a new scheme to si-
multaneously acquire diastolic and systolic
images of the heart in the same R-R interval
in breath-hold times [7]. This study had four
objectives.
The first objective was to evaluate the ul-
trafast sequence for estimation of the left yen-
tricular ejection fraction using a series of
short-axis slices covering the heart from the
apex to the base (Simpson’s rule).
The second objective was to compare this
geometry-independent method of estimating
ejection fraction with that obtained from a
model-based ellipsoid biplane method. The
latter method involves acquisition of only two
planes so that if this biplane method can be
shown to have accuracy and reproducibility
comparable to those of the Simpson’s rule
method. using the biplane method will repre-
sent a considerable savings in scan time.
The third objective was to assess the repro-
ducibility of left ventricular ejection fraction
measurements obtained using Simpson’s rule
method and the biplane ellipsoid method. The
calculation of the left ventricular ejection
fraction from the short-axis images requires
either manual or computer-assisted outlining
of the endocardial borders. Similarly. the esti-
mation of left ventricular ejection fraction
from the biplane views requires a manual
measurement of certain left ventricular di-
mensions. Both the manual outlining and the
manual measurements are prone to inter- and
intraobserver variabilities.
The fourth objective was to test an en-
docardial border detection algorithm by com-
parison of the ejection fractions derived by the
semiautomated method with those from the
manual contouring by experienced radiolo-
gists. The manual method of contouring the
endocardial borders is both time-consuming
and prone to operator variabilities. The semi-
automated algorithm for contouring the en-
docardial borders was developed in-house
(Sinha 5, unpublished data).
Evaluation of the two methods of ejection
fraction estimation involved comparison with
the clinical technique used at our site for global
cardiac function evaluation-that is, echocardi-
ography. Echocardiography was not taken as a
gold standard because the potential problems
that limit accuracy of ejection fraction estima-
tion using this technique are well known and
are discussed later. Rather. such comparisons
were made essentially because echocardiog-
raphy is the routine clinical method for evaluat-
ing cardiac parameters.
Subjects and Methods
Study Design
The dark-blood dual-phase sequence was used to
acquire cardiac images in five healthy volunteers
(subjects I. 2, 3. 5, and 9) with no history of cardiac
disease. and five cardiac patients (subjects 4, 6. 7. 8,
and 10) from whom informed consent was obtained
(Table I ). The average age for the healthy subjects
was 33 years old. Patients 4 and 10 were suffering
from severe global cardiomyopathy. and patients 6,
7. and 8 had a history of myocardial infarction.
Echocardiography scans were obtained for all
healthy volunteers by an experienced technologist,
and patient echocardiography data were obtained
from the patient’s most recent study. To assess the
reproducibility of the measurements, healthy volun-
teers 1 and 3 were scanned in the MR scanner onthree different occasions.
MR Imaging
ECG-gated cardiac MR images of all subjects
were acquired on a I .5-T MR scanner (Magnetom
Vision; Siemens Medical Systems, Iselin, NJ). Sub-
jects were scanned lying supine with ECG electrodes
on the anterior chest wall. A circularly polarized.
phased array body coil. provided commercially by
Siemens. yielded excellent image quality uniformly
throughout the chest region.Dark-bloodturbo fast low-angle shot (turboFLASH) cardiac
images acquired in breath-hold times has been pro-
posed earlier 18. 9J. However. these sequences per-
mitted acquisition only during diastole and could
not be used for estimating ejection fractions be-
cause such estimations require determination of left
ventricular volumes both in end diastole and in end
systole. We designed a magnetization-prepared seg-
mented turboFLASH sequence to simultaneously
acquire images at these two cardiac phases (end di-
astole and end systole). The details of the sequence
are given in the study by Sinha and Sinha 171 and
are summarized in the following text. The magneti-
zation preparation period consists of two 180#{176}RF
pulses designed to selectively suppress blood signal
(Fig. 1). The first l80� pulse is non-slice-selective
and inverts spins from the entire volume. The see-
ond 180#{176}RF pulse is slice-selective and thus re-
stores the spins in the slice of interest to
equilibrium. Magnetization of the static spins in the
slice is recovered, whereas that of blood flowing
into the slice remains inverted. In the single phase
acquisition during diastole. the signal from blood is
minimized by positioning the acquisitioll window
near the null point (T��ii) of blood (8. 91. In the
present scheme of dual-phase images. however. the
two acquisitions are placed symmetrically about
Tnuti. We have shown earlier. from both theoretic
considerations and experimental observations. that
for typical choice of sequence parameters. blood is
suppressed by a factor of 0.3 of its maximum value
at either acquisition [7]. Furthermore. the sequence
is run in a segmented mode that implies the collec-
tion of several lines of K-space in one acquisition
window. Artifacts from the increased motion of the
heart at systole were niinimized by optimizing the
T��ii i
T�,j
(1)(2)
(3)
Fig. 1.-Sequence diagram for dual-phase dark-blood segmented K-space acquisition (five, seven, nine, 11, or13 lines per segment) shows two acquisitions positioned at end-systolic and end-diastolic cardiac phases withtimings set using equations 1-3. Blood signal is suppressed by positioning two acquisitions symmetrically aroundzero-point crossing of inverted blood magnetization, Tnull’ Pressure wave was not measured but is shown to helpidentify relative time points of acquisition windows. The it pulses refer to 180#{176}RF pulses required for magnetiza-tion preparation, whereas a pulses are RF pulses for segmented gradient echoes. Td = delay of preparation pulsesfrom R-wave, T1= delay between preparation pulses and start offirst acquisition, 2T8� total time of segmentedacquisition, I� = time interval between end of first acquisition and beginning of second. Tes time to end systole,
Ted time to end diastole.
�l 4�
�e 20
�e -20
�-40
I�
slrww._2� �
�- --- �--
�
:��i
VT� or T�
4-912Taq
Time
Fig. 2.-Schematic of phase reordering shows relative position of end-diastolic and end-systolic time points (154phase lines, 11 lines per segment). During each R�R interval, phase-encoding levels shown in one segment (Seg. 1,Seg. 2) are traversed once each in end-diastolic (Ted) and end-systolic (Tes) time points. Note that end-diastolic andend-systolic time points are centered in acquisition windows (2Taq) where low K-space lines are acquired first.
MR Imaging of Left Ventricular Ejection Fraction
AJR:169, July 1997 103
length of the acquisition window (which for a fixed
TR ofthe turboFLASH sequence increases with the
number of K-space lines in the segment). The tim-ings Ii,, the time interval between the end of the
first acquisition and the beginning ofthe second; T�,the interval between the preparation pulses and the
beginning of the first acquisition; and Td’ �e inter-val between the R-wave and the preparation pulses
(Fig. 1) of the sequence, are adjusted by the opera-
tor so that the acquisitions, each of duration 2Taq’
are centered at end systole, I�, and at end diastole,T�, around TraIl’ the time at which the magnetiza-
tion of blood spins goes through zero, according to
the equations:
I� = (Ted _T��) -
1 = Tnull - 2Taq _(I�/ 2),
Td(TesTaq)h1.
An initial cine scout scan was used to detetminethe temporal location of the end-diastolic (T�) and
end-systolic (Tes) �&dftit fr�tfl�S. ThC timings are
referenced to the start of the R-wave trigger. Two ac-
quisition windows, each of a duration of 70 msec
centered around the end-systolic and the end-dia-
stolic time points, serve to acquire the images at the
two phases. Both nequisition windows are within the
same R-R interval (Fig. 1). In each such acquisitionsegment� 11 phase-encodinglthes are acquired. Thesephase tines are sequentially reordered with the cen-
tral K-space lines in the middle of the acquisitionwindow (Fig. 2). The end-diastolic and end-systolic
tune points were centered on the respective acquisi-
tion windows (i.e., on the central K-space lines).The schemejust described works well for spatial
locations from the apex tojust beyond the mid yen-
tricle level. Cardiac motion is more pronounced to-
ward the base of the heart than at the apex [10].This severe motion of the more basal slices results
in loss of myocardia.l signal intensity in the systolic
phase because of saturated spins outside the slice ofinterest moving into the slice. The problem was cir-
cumvented by positioning the preparatory inversionpulses at systole and acquiring the slice in the fol-
lowing R-R interval at the same cardiac phase
during systole. The end-diastolic acquisition isunchanged. The excellent image quality obtainedusing this variation of the pulse sequence is shownin the Results section. This latter scheme, which
stretches each segmented acquisition to two R-R in-
tervals, is used only for the basal slices.The imaging protocol consisted of first acquiring
scout scans to locate the long-axis, short-axis, and
four-chamber views (the long-axis view is equivalentto the vertical long-axis view, and the four-chamberview is equivalent to the horizontal long-axis view).This step was followed by a cine scan in the short-
axis view at the level of the mid ventricle to accu-
rately determine � and I�,j. A direct examination ofthe ECG signal should be sufficient to locate the sys-
tolic and diastolic time points, but our experienceshowed that the cine scans were able to locate more
accurately and consistently the required cardiac
phases. The cine sequence (TR � 12 msec), permit-
(4)
ted acquisitions of 18-22 temporal slices (depending
on the R-R interval of the subject). These multiphase
cine images of the same slice were then examined vi-
sually, and the end-systolic and end-diastolic frames
were identified by the smallest and largest endocar-
dial areas. respectively, of the left ventricular cham-ber. The dual-phase dark-blood images were then
acquired under breath-hold conditions with the two
acquisition window locations determined from the
cine scan and other sequence timings determined
from equations 1-3. Typical values of these timings
are as follows: R-R, 800 msec; T�, 6&) msec; T�,
335 msec; I�, 240 msec; I�, 275 msec; and Td’
I 7 msec. I� and Td were adjusted for each subject toobtain the best blood suppression. The image planes
acquired with the dark-blood sequence were a long-
axis view, a four-chamber view, and a series of con-
tiguous short-axis views from apex to base. Each
breath-hold acquisition yielded images of the sameslice at the two points in the cardiac cycle in l6to apeproximately 21 sec. A pause of 6 to approximately
10 sec was allowed between scans. Optimization of
sequence parameters led to the following choices for
the dual-phase cardiac acquisitions reported here:(I) TRItE of the tUTbOFLASH acquisition, 6.3/3.2
msec; field ofview, 45#{176};(2) number ofK-space linesand segments, 11 (resulting in an acquisition window
of 70 msec); and (3) geometry parameters: slicethickness, 10 mm; field of view, 230-250 mm; spac-
ing, contiguous.
For the purposes of comparison with the tech-
nique most commonly used at present, we also ac-quired bright-blood breath-hold images for onevolunteer with scan parameters similar to those used
for the dark-blood images. Because these imageswere acquired to facilitate a direct comparison withdark- and bright-blood images obtained under simi-
lar conditions and in the same scan time, we used a
segmented breath-hold turboFLASH cine scan.
Echocardiogrophy
Echocardiography was performed on all healthyvolunteers to obtain long-axis and four-chamber
views during end systole and end diastole. All im-
aging was performed on an AlL Ultramark 9 (Ad-
vanced Techniques Laboratory, Bothell, WA) with
a 2.25-MHz phased array transducer. All patients
had echocardiography within I month before theMR imaging study. No patient had therapeutic in-
tervention between the time of echocardiographyand the MR imaging study. In addition, histories
obtained from all patients revealed no significantchange in symptomatology related to their cardio-myopathy between studies.
Left Ventricular Volume Estimation
From MR short-avis images-Simpson’s rulewas used to directly measure the left ventricular vol-
umes from the set of endocardial contours of theshort-axis images. The end-diastolic and end-sys-
tolic chamber volume of the left ventricle were oh-tamed by integrating the areas bound by the
endocardial contours (in the two cardiac phases) and
multiplying by the section thickness. The cardiac
(5)
Sinha et al.
104 AJR:169, July 1997
contours were outlined manually, once for all the
subjects, by two experienced cardiovascular radiolo-gists. In addition, the intraobserver variability was
estimated by each radiologist repeating the manualcontouring of particular studies (three times for each
study). These studies were the repeated MR exami-nations on subjects I and 3. Finally, the contours
were also determined using a semiautomated algo-
rithm (briefly described in the following text).
From the dual-plane MR images-The ventricu-
lar volumes (V) from the long-axis and four-cham-
her images were calculated using the equation [1 11:
(irL� (4A� (4A2V=I- lxi Xl
�6) tjrDi) �jtD2
where L = long-axis length of the ventricle, Al =
cross-sectional area at the mid papillary muscle level
on the long-axis view, DI = diameter at the mid pap-
illary muscle level on the long-axis view, A2 =
cross-sectional area at the mid papillary muscle levelon the four-chamber view, D2 = diameter at the mid
papillary muscle level on the four-chamber view.
Given the geometric constraint, A = ir(D I 2)2, thevolume reduces to the formula:
v= DIXD2X(�
The measurements of Dl , D2, and L for each sub-ject were again performed by the two radiologists
with repeated measurements on the same scans asfor the short-axis views.
Front echocardiography.-Ventricular volumes
were calculated using biplane apical views with a
modified Simpson’s rule approach [1 1]:
V = � � D.D x4s..l �20
where the length L of the ventricle is divided into
20 disks (i = 1 to i = 20) from base to apex with thediameters, Dl and D2, of each disk determined in
two apical views, four-chamber and long-axis.
Morphologic Measurements from MR Images
Manual techniques-Images were transferred
from the MR host computer to a SPARC5 work-station (Sun Microsystems, Mountain View, CA)
for image analysis. Images were displayed en-larged by a factor of 2 to facilitate manual mea-
surement.s. Endocardial borders were traced using
the mouse, and the algorithm calculated the num-
ber of pixels inside the trace. Distance measure-
ments were made subsequent to the user’s clickingat appropriate start and end points.
Semiautomated image pux-essing.--Image pro-cessing was performed on a SPARC5 workstationwith image processing tools (median filter, erosion
and dilation morphologic operators) based on the ap-plications software SunVision (an image processing
applications package) (Sun Microsystems). The algo-rithm locates both the endocardial and epicardial bor-
ders but, for the sake of brevity, only the stage
corresponding to the endocardial border detection isdescribed in the following text. The process is similar
to that for detecting the epicardial border. The userinitiates the program forthe isolation ofthe left ventri-
dc by identifying the approximate center of the leftventricle and a region of interest encompassing the
left ventricle. The histogram of the region of interestis displayed on a panel that essentially consists ofthree peaks corresponding to the background (and
blood), myocardium, and fat. On an intensity scale of0-255, typical blood signal intensity is less than 2&,
myocardium signal intensity ranges from 28 to 75,
and fat signal intensity is greater than 75. The user
sets the lower threshold interactively on the histo-
gram. The image is first changed to black and white
so that only pixels with values equal to or less than thelower threshold are set to white and all others are set
to black. The next step is to isolate the region spatially
connected to the center of the left ventricle (the user-defined center). The size of the isolated region is
checked against preset values. and the object is
eroded until the criterion is met. The size criterion
serves to break false connections to large isointense
objects such as the chest wall and liver. Each erosion
is followed by a connectivity operation to isolate the
new area. The connectivity algorithm isolates thestructure in which all pixels are connected to the seed
pixel through their four nearest neighbors. Once the
size criterion is fulfilled, the image is dilated the samenumber oftimes that it was eroded. The dilation is fol-
lowed by a sequential contouring algorithm that iden-tifies the endocardial border pixel that can then be
used for subsequent calculations of ejection fraction.
A 3 x 3 rectangular structuring element is used for the
morphologic operations of both erosion and dilation.
Erosion diminishes the isolated volume, removes sub-volumes that do not fill out the structuring element,
(6) and forces spatial isolation by breaking false connec-
tions. Dilation enlarges the volume and fills holes and
indentations that are smaller than or equal to the struc-turing element. The epicardial surface was also con-
toured by a similar procedure but was not used in anyof our calculations and so is not described here.
Contour refinement-A further refinement to
the contour found in the previous step was incorpo-
rated using the concept of active contours, or“snakes,” which have the ability to conform to dif-
ferent contours within an image [ 12]. We describethe refinement procedure briefly in the following
text. The snake algorithm requires an initial active
contour (a rough approximation to the desired con-tour in the image) to be specified, which in our al-gorithm is the output of the previous stage. The true
border between the myocardium and the left ventri-
cle is automatically estimated by the snake algo-
rithm on the basis of iteratively minimizing the
energy of the active contour. The energy of the con-tour consists of different components such as the in-temal energy, image gradient energy, and externalenergy. The internal energy is defined from the
smoothness and compactness of the contour, the gra-
I0
MR Imaging of Left Ventricular Ejection Fraction
AJR:169, July 1997 105
dient energy is the negative of the sum of the gradi-
ent strengths at cacti contour point. and the external
energy consists �)f any knowledge-based constraints
the user can add to the equation.
Data Analysis
(.Ofll/)(IILsOfl I)(’t1i�e(’!1 (1(1/k- (111(1 hiight-blood tnt-
tiges.-The contours obtained from the semiauto-
niated algorithm followed by the refinement were
analyied as follows: The center of mass was esti-
mated using the border pixels. and intensity gradients
along the radial direction were calculated. Three pixel
points on either side of each border pixel were exani-
ned for the one with the maximum gradient value.
(The border pixel may not be the maxirnuni gradient
pixel because the snake algorithm chooses the con-
tour pixel on the basis of the maxiniurn gradient and
other cotisiderations. I Comparisons were then made
using the histograni distribution of pixel gradient val-
ues for all the border pixels.
Co,tzjaii.oni 1)t’titt’t’!i ili//t’re’tit t(’(/nnques.-Per-centage ofejection fractions (EF) was calculated as:
. EDt: -ESVLF=-- xl()0.
EDV
where EDV is the end-diastolic volume and ESV is
the end-systolic volume of the left ventricle. The
ejection fractions deteriiiined front MR iiaaging.
using the short-;ods images and the biplane method
for all the subjects. were compared with those front
echocaidiography. A sepalate coniparison was also
niade of the ejection fractions detenitined from
echocardiogranis and those determined by the semi-
autontated algorithm front the short-axis MR im-
ages. (‘oniparisons were expiessed is linear
regression plots I.)i the ejection fractions from the
two techniques. A slope close to one for the linear
regression line indicates that the ejection fractions
froii� the tVv() methods coiiipaied well. vhereas a re-
gression coefficient of I indicates a high degree of
correlation between the two data sets.
Reproducibility
Variations in ejection fiactioti estiniations front
MR inlages could potentially arise froiu three dif-
ferent factois: ( I ) interexaiilination variaL)ility and
( 2) intra- and ( 3 ) interobserver s ariation in deter-
mining the left ventricular endocardial llx)rder
( Simpson ‘s rule ) and left ventricular diiiiensions
( biplane aethod). We investigated the contrihu-
tions of these three factors to the reproducibilit) of
ejeCtiOil fraction estintates. 1 ftc f irst factor. inter-
(7 1 exaniination variability . v�as deterntined by scaii-
ning two subjects on three separate occasions s�ith
identical MR protocols. The interexamination
variahilit� was expressed in terms of the average
coefficient of variation of the ejection fraction es-
tintated front the three epetilions. lfiiee uianual
contouring trials sverc lx�rf�i tied on each of tss o
Patients at three different exaiiiiihttions (a total ofsix studies) to evaluate tfte iniraofserver variation
of the left s entricuLir est iilhtt Oil . I�iL ft contouring
trial was separated fiuni the pies bus OOC h� 7 I .4
days to decrease any remnant meiiior� ef fects. In
addition. the iadiologist [)erfornhing the contouring
WdS kept unav�arc of patient deniograpliics. 1 he
interobserver s ariahilit� s� a- deternniicd from the
ejection fraction estinlaies h� the iv�o observers oii
all the studies. The iiti aohsers er ‘. ailat (lit iii theestiniation of ejection fractions using tfie seiiiiau-
tomated algorithiii ssas estiin�iied front three sepa-
rate trials on each of six studies (the sante as the
Oiies used iii estiin�itii1g utti �iohsers er s �ti i�ihifities
by niaiiual contouring ntethotls . li�.aclt stiirf� svas
contoured three tinics using the ‘eniiatitoinated al-
goritliiii b� ()itC ohsei s er s� ith �i stiff iciiiitl� long
tinie inters at betsvccii Lritfs. �\ll cocf ficieiuts of
\�iriiItiO1t 55Ci� expiesseii in teiins of peicenLige of
SD of the mean.
Results
A typical set of four-cltanihcr �uid long-axis
inlages (in end-diastolic �utd end-systolic
phases) using the d;uk-hkxxl sequence is shown
in Figure 3. The distance measurements I) I.
Fig. 3.-Typical dark-blood gated MR images of leftventricle in 40-year-old healthy male volunteer.A. Vertical long-axis view of heart at end diastole. Notetwo ellipsoidal axis lengths, Dl and L, required for cal-
culating ventricular volume according to equation 4.
B, Similar vertical long-axis view at end systole showssystolic thickening of myocardium.C, Horizontal long-axis (or four-chamber) view of leftventricle at end diastole. Lengths L and 02 are shown,
the latter being last of three measurements of axis
lengths required in equation 4.D, Horizontal long-axis view at end systole similar tothat in C.
Sinha et aI.
106 AJR:169, July 1997
Fig. 4.-Short-axis MR images ofheart at different spatial and tempo-ral locations in 37-year-old healthymale volunteer using dark-blood se-quence. Signal-to-noise ratio and ex-tent of blood suppression seen hereare representative of most imagesacquired in 10 subjects.A and B, Images acquired at end dias-tole (A), with most apical image onleft, may be compared with corre-sponding spatial images acquired inend systole (B).C and D, More basal slices are shownat end diastole (C), with correspond-ing images at end systole (D). Mostbasal slice is on right Note clear de-lineation of left ventricle (LV, C) andright ventricle (RV, C) in these basalslices, which undergo large degree ofdistortion during cardiac cycle.
D2. and L required for the volume estimation ments from the four-chamber and long-axis
from the biplane method are also indicated images (Fig. 3). Short-axis images of the
(Fig. 3). The long-axis length of the ventricle, sanse subject covering the entire heart were
L, is taken to be the average of the measure- acquired with the dark-blood dual-phase se-
quence (Fig. 4). Images at each spatial loca-
tion were obtained at end diastole and end
systole in a single breath-hold. The time re-
quired to image the heart froni apex to base
A
AJR:169, July 1997 107
Fig. 5.-Short-axis MR images of leftventricle are shown with endocardialand epicardial contours from semiau-tomated algorithm superimposed aswhite pixels.A and B, End-diastolic images (A) cor-respond to end-systolic images (B).These imageswere acquired atthreespatial locations near mid ventricle
level. Relatively high degree of suc-cess of automatic contouring is facili-tated to large extent by good contrastbetween myocardium and ventricularspace in these dark-blood images.
was 8-l() ruin, including preparation and re-
construction time.
The endocardial and epicardial borders of
the heart, as deteniiined from the semiauto-
mated algorithm. are shown superposed on a
set ofdark-blood cardiac images (Fig. 5). The
semiautoniated algorithni took approximately
6 mm for a series of eight images (eight spa-
tial locations at one cardiac phase). Additional
time (3-5 mm) was required for nianual inter-
ventions for the proper identification of the
epicardial border, which was often contiguous
with isointense liver tissue or muscle of the
chest wall. The endocardial borders (lid not
require any manual intervention because the
myocardium-blood contrast was high and
hence tuore amenable to automated detection.
However. soiite of the ntost apical slices did
register blood as bright: manual editing was
required for these images.
Brmght-hkxxl imiiages of the heart showed
nonunifon�iity of blood signal imitemisity at
5(.)I11C locations and cardiac phases (Fig. 6). An
additional drasvhack was that in the basal
Fig. 6.-Bright-blood MR images nearbase of ventricle.A and B, End-diastolic (A) and end-sys-tolic (B) phases are shown at approxi-mately same location as dark-bloodimages in Figure 4B. Note thatwhereasoverall contrast between myocardiumand blood is high, blood signal intensityis nonuniform. In certain regions, par-ticularfy in posterior wall in systolicphase (arrow, B), distinguishing yen-tricular space from myocardium isdifficult, which leads to errors in en-docardial border detection.
I
V0
20
10
10 20 30 40 50 60 70 80 90 100
Intensity Gradient Strength (arbitrarily assigned units)
I
.1:
EF from MR Short-Axis Imiges EF from MR Images
Using Biplane Method (%)
Sinha et al.
108 AJR:169, July 1997
Fig. 7.-Histogram of pixel intensity gradient values at endocardial border in dark-blood (solid line) and bright-blood (dottedline) images. Histogram shown here is for typicalslice (dark and bright blood) in short-axis view. Gradientvalue atany pixel is maximum absolute value ofgradient along a radial line within three pixels of border pixel identifiedby snakeS algorithm.
slices, the signal intensity of blood was isoin-
tense to that of myocardium, particularly near
the posterior wall (Fig. 6). The spatial location
of these slices corresponds roughly to that of
the dark-blood images in Figure 4B. The histo-
grams of gradient intensity distribution of bor-
der pixels indicate that in typical dark-blood
images, the average gradient values are larger
than those in bright-blood images (Fig. 7).
The average ejection fraction of the five
volunteers was 55% (Table 1). A wide range
of ejection fractions was found for the patients
(most lower than the ejection fraction range in
healthy volunteers) and particularly for the
two patients with severe global cardiomyopa-
thy (subjects 4 and 10 in Table 1). Measure-
ments of left ventricular ejection fraction
estimates from MR short-axis images (manual
contouring) and from echocardiographic im-
ages are compared in Figure 8. A significant
correlation was observed between the two
techniques, although echocardiographic val-
ues were underestimated compared with val-
ues from MR short-axis images (slope of
regression line = 0.68, 2 correlation coeffi-
cient = .88). A similar comparison of ejection
Fig. 8-Ejection fractions (EF) estimated from manualcontouring (by observer 1) of short-axis MR images ofheart using Simpson’s rule compared with estimatesfrom echocardiography. Patient data are shown byfilled circles and volunteer data by open circles. Equa-tion to regression line is given in form of y= (slope) x x
+ (intercept), and coefficient of regression as R2. Slopeof regression line was 0.68 (compared with 1.0 for per-fect agreement shown as dashed line) together withvalue of.88forR2. Inthis graph, y=0.68x+ 18.
Fig. a--Ejection fractions (Efl estimated from biplanemethod, using ellipsoid biplane model (with leftventricularmeasurements by observer 1) applied to cardiac MR im-ages compared with estimates from echocardiography.Patient data are shown by solid circles and volunteer databy open ones. Equation to regression line is given as y =
(slope)xx+(intercept)and coefficientofregression as R2.Value of 0.99 for slope of regression line (compared with1.Ofor line ofidentity [dashedlinej) revealsthattwo meth-
ods agree reasonably well. Significant degree of correla-don was obtained (R = .82). Inthis graph, y=O.93x-2.8.
‘S..
0
aa
ci‘0m�aB.
.� z
B0
EF from MR Short-Axis Images EF from Manual Contouring of
MR Short-Axis Images
mnterexamination Va,IabIIIty of Ejection Fraction Contoured by Observer I
.
MR Imaging ExaminationSubject 1 Subject 2
Biplane (%) Short Axis (%) Biplane (%) Short Axis (%)
Numberl
Number2
Number3
%SD
62
67
66
3
66
61
55
10
67
57
52
13
58
50
49
9
lntraobserver Varlablllty of EjectIon Fraction for Observer I
Exam. Subj.
Trial 1 (%) Trial 2 (%) Trial 3 (%) SD (%)
BiplaneShortAxis
.
BiplaneShortAxis
.
BiplaneShortAxis
.
BiplaneShortAxis
1
2
3
4
5
6
1
1
1
3
3
3
62
67
66
67
57
52
66
61
55
58
50
49
59
68
69
61
54
53
62
58
55
56
51
48
67
70
73
64
60
50
63
63
53
56
49
48
6
3
5
5
5
3
3
4
2
2
2
2
MR Imaging of Left Ventricular Ejection Fraction
AJR:169, July 1997 109
Fig. 10.-Ejection fraction (EF) estimates from corn-puter-generated contours of short-axis MR imagescompared with those from echocardiography. Patientdata are shown by filled circles and volunteer data byopen circles. Equation to regression line is given as y= (slope) x x + (intercept), and coefficient of regres-sion as R2. Value of 0.65 for slope of regression line(compared with 1.0 for line of identity shown asdashedline), together with value of .85 for R 2, revealsresults closely parallel to those shown in Figure 9. Inthis graph, y= 0.65x + 18.8.
Fig. 11.-Ejection fraction (EF) estimates obtained from semiautomated contouring of MR short-axis images com-pared with those from manually contoured (observer 1) short-axis images. Patient data are shown by filled circlesand volunteer data by open circles. Line ofidentity is indicated by dashed line. Equation to re9ression line is given asy= (slope) x X+ (intercept), and coefficient of regression as R2. Hi9h degree of correlation (R’ = .98) and near perfectagreement(slope of regression line = 1.07) indicate good reliability of semiautomated contouring process comparedwith manual contouring by experienced radiologists. In this graph, y= 1.Olx- 2.07.
Note-Exam. = examination, Subj. = subject.
fraction estimates from the biplane MR imag-
ing method with those from echocardio-
graphic images reveal both a significant
correlation and a closer agreement of the esti-
mates (Fig. 9) (slope of regression line = 0.93,
2 = .82). Results of the regression analysis
of ejection fraction estimates obtained from
MR short-axis images (semiautomated con-
touring) compared with those from echocar-
diographic images closely paralleled those
obtained for the manually contoured data (Fig.
10) (slope of regression line = 0.65, 2 .85).
A direct comparison of the ejection fraction
estimates from short-axis MR images pro-
cessed by the manual versus the semlauto-
mated method revealed a high degree of
correlation and agreement (Fig. 1 1) (slope of
regression line = I .07, 2 .98).
An examination of the estimates of ejection
fraction from three independent scans each on
two of the subjects using the biplane and Simp-
son’s rule methods revealed that the average
value of the percentage of SD caused by inter-
examination variability was about 9% for both
methods of ejection fraction estimation (Ta-
ble 2). The results of the contouring by ob-
server 1 are given in Table 2; similar results
were obtained by observer 2. The intraob-
server variability (for observer 1) was esti-
mated as the percentage of SD ofthe mean left
ventricular ejection fraction obtained from
three trials each on six scans (Table 3). Again,
as in the case of the interexarnination variabil-
ity, the results of observer 2 closely tracked
those of observer 1 and therefore are not listed
here. The average percentages of SD of ejec-
tion fraction estimates caused by intraobserver
variability (averaged for both observers) were
4.5% for the biplane method and 2.5% for the
short-axis images. Figures 12A and l2B are
comparisons of the ejection fractions obtained
by the two observers for all the subjects. The
I..Ii
0a
I
II
0aC
aC
Ejection Fraction, Observer 2 (%)
A
Ejection Fraction, Observer 2 (%)
B
Sinha et al.
110 AJR:169, July 1997
Fig. 12.-lnterobserver variability in usingshort-axis and biplane methods to estimateejection fractions. Patient data are shownby filled circles and volunteer data by opencircles. Une of identity is indicated bydashed line. Equation to regression line isgiven as y= (slope) x x+ (intercept), and co-efficient of regression as R2.A, Ejection fractions obtained bytwo observ-ers using same short-axis images. Note highdegree of correlation (slope = 1.0, R2 = .96)between estimates obtained by two observ-ers, indicating low interobserver variabilitywhen using short-axis images. In this graph,y= 1.Ox+ 0.18.B, Ejection fraction obtained bytwo observ-ers using biplane method. Note lack ofagreement and degree of scatter, indicatedby slope of 0.63 and R2 of .46 between val-ues of ejection fraction estimates. In thisgraph. y=0.63x+21.
slope of the regression lines and the coeffi-
cients of correlation are 1 .0 and .96, respec-
tively, for the short-axis images, whereas for
the biplane method these values are 0.63 and
.46, respectively. The interobserver variabili-
ties are 12% and 4% for the biplane and Simp-
son’s rule methods, respectively. The
intraobserver variability in the estimates of
ejection fraction using the semiautomated at-
gorithm, averaged over all six studies (three
trails each), was approximately 3%.
Discussion
Cine MR imaging provides reliable and
reproducible measurements of left ventricu-
lar ejection fraction and left ventricular myo-
cardial mass [1-4]. However, cine images
acquired with conventional gradient-echo se-
quences require scan times of 10-20 mm.
Recently, a breath-hold segmented cine se-
quence yielding bright-blood images has
been proposed and evaluated for reproduc-
ibility of left ventricular volume estimations
[4, 13]. However, the bright-blood tech-
niques suffer from variable blood signal in-
tensity through the cardiac cycle as seen in
Figure 6 (e.g., image at the right in Fig. 6A
compared with the corresponding dark-blood
images in Fig. 4). Such nonuniformity of
blood signal intensity can lead to problems
when using semiautomated algorithms to
outline the endocardial border [4]. The reason
for the variable signal intensity of the left
ventricle is that blood registers bright on cine
images because it is flowing at varying rates
within the ventricle compared with a relatively
static myocardium. However, slow-flowing
blood near the myocardiat wall (especially in
the diastolic phase) may have a reduced sig-
nat intensity and may often be difficult to
distinguish from the myocardium in bright-
blood images because it is subject to the
same saturation effects that render the myo-
cardium hypointense compared with the rap-
idly flowing blood. The new sequence
proposed here overcomes this problem by
suppressing the left ventricular blood signal
by a preparation scheme. Furthermore, the
interval between the preparation pulses and
the collection of data allows enough time
even for the slow-moving blood present
within the slice to move out of the slice and
be replaced by dark blood. In conjunction
with a phased array surface coil, the sequence
yields images with good blood-to-myocar-
dium contrast and high signal-to-noise ratio.
The extent of blood suppression varied
among the subjects, and in some slices the
intensity of the suppressed blood appeared
nonuniform (Fig. 4). Nonuniformity of blood
signal intensity may arise from the fact that
not all the blood in the slice of interest has
been replaced by inverted blood from out-
side the imaging plane. The extent of this re-
placement depends on several factors such as
the thickness of the reinverted slice, velocity
of blood, cardiac phase, and spatial location.
However, even in the worst case, the blood
suppression was sufficient to permit unam-
biguous detection of the myocardial border.
Contrast between the blood and the myo-
cardium can be quantified in terms of the in-
tensity gradient at the endocardial border
pixels. The semiautomated algorithm used
here (and in general most such algorithms)
uses the detection of intensity discontinuities
between homogenous regions to identify the
edges. Larger values of the intensity gradient
will thus result in a greater reliability of seg-
menting the endocardial border using auto-
mated techniques, which was the motivation
behind comparing the gradients of pixels on
the endocardial border for dark- and bright-
blood images. The histogram in Figure 7
shows the average gradient strengths of the
border pixels to be higher for the dark-blood
images than for the bright-blood images and
quantitatively indicates that dark-blood im-
ages have an increased blood-to-myocardium
contrast over bright-blood images. A draw-
back of imaging the heart at only two phases,
such as at end diastole and end systole as in
our scheme, may be an inability to assess par-
ticular cardiac conditions such as diastolic
dysfunction or asynergic contraction that ap-
pear normal at end systole.
The superiority of Simpson’s technique
over the biplane technique is well established,
but our rationale for investigating the model-
based biplane method is as follows. The se-
quence discussed in this paper does permit left
ventricular function estimations in relatively
short scan times using the more accurate Simp-
son’s technique. To cover the entire heart,
eight to 10 breath-holds are necessary-one
breath-hold for each spatial location. Al-
though this pace is not demanding for volun-
teers, conceivably some patients will not be
able to perform these eight to 10 breath-holds.
Using the biplane method instead of Simp-
son’s method requires only two breath-holds,
thereby making the procedure less demand-
ing. We investigated the differences in left
ventricular estimation between the Simpson’s
and the biplane methods to quantify the
amount of error in the latter method, using the
proposed sequence. Thus, although accurate
measurements may not be possible using the
biplane method, that method may still enable
MR Imaging of Left Ventricular Ejection Fraction
AJR:169, July 1997 111
quick left ventricular function estimates to be
made in even very sick patients, which until
now was not possible with MR imaging.
However. maximum deviations from true es-
timates of ejection fraction occur in sick pa-
tients because the ellipsoid model is not
strictly valid in such cases. Simpson’s method
will be the protocol of choice in all subjects
capable of eight to 10 breath-holds, and the
biplane method can be used only in patients
who have difficulty holding their breath.
The normalized average difference (in per-
centage) between ejection fraction estimations
from the biplane method and from short-axis
images was about 15% excluding the two pa-
tients with severe cardiomyopathy. The aver-
age value of this quantity for the two patients
was approximately 150%. The large discrep-
ancy for the two patients was due to the fail-
ure of the geometry-based model in those
cases. Thus, it would not be feasible to accu-
rately calculate ejection fractions in patients
with global cardiomyopathy using the biplane
method, but that method could be used in such
patients to obtain an approximate figure.
However, the biplane method can be used to
obtain ejection fractions with sufticient accu-
racy in patients with. for example. history of
myocardial infarction (subjects 6, 7, and 8 in
Table 1 ). The regression plots in Figures 8-10
show a significant correlation between ejec-
tion fractions obtained from MR imaging and
from echocardiography techniques. The better
agreement between ejection fl-actions esti-
mated from echocardiography and from the
MR imaging biplane method is not surprising
because in both cases similar geometric as-
sumptions were involved in the calculation of
the let’t ventricular volumes. A comparison of
ejection fraction estimates from echocardiog-
raphy and front MR imaging using Simpson’s
rule revealed that the estimates from the former
were consistently less than those from MR
imaging for both manual (Fig. 8) and auto-
mated (Fig. 10) contouring. Ejection fraction
estimates from short-axis MR images are
more accurate because the echocardiography
estimate is based on the geometric assump-
tion of elliptical ventricular cross section that
is suboptimal for the precise measurement of
left ventricular dimensions. Potential maccu-
racies do exist in ejection fraction estimations
from short-axis images, arising mainly from
errors in locating the end points of the left
ventricle both at the base and at the apex.
However, because the calculation of the ejec-
tion fraction involves a subtraction of these
measurements made at systole froni those
made at diastole, this particular error is likely
to be smaller than that arising from the geo-
metric assumptions in the biplane method.
Another potential source of error is the tempo-
ral resolution of the images. In gradient-echo
imaging protocols, the temporal resolution
can be as small as 12 msec, but the ultrafast
segmented technique implemented in this
study uses an acquisition window of 70 msec.
Because the motion of the heart is most pro-
nounced at systole, the end-systolic volume
will depend on the duration of the acquisition
window. However, an acquisition window of
about 60-msec duration can still be used for
the accurate depiction of the end-systolic vol-
umes [4]. Our study uses an acquisition win-
dow of 70 msec, but this duration is likely to
be reduced in the future with the use of shorter
gradient-rise times, a hardware improvement
that is being made increasingly more available
in clinical scanners.
The correlation coefficients and slope of the
regression lines were similar for the plots in
Figures 8 and 10. Such similarity of behavior
was again reflected in the excellent correlation
(with a slope of 1.07) between the ejection
fraction values obtained from the manual and
from the semiautomatic methods (Fig. 1 1).
This finding is encouraging because it implies
that the left ventricular contours generated by
the sentiautomated algorithm agree closely
with the contours generated by the expert. An
automated or semiautomated technique is criti-
cal for a clinical implementation of left yen-
tricular function estimates in a time-efficient
manner. As indicated in the introduction. the
success of the semiautomated method depends
on the robustness of the algorithm, the signal
intensity and contrast-to noise ratio of the left
ventricle and myocardium, and the signal ho-
mogeneity of both blood and myocardium.
The comparison of the dark-blood technique
with the equivalent bright-blood scheme
shows an improvement of both contrast be-
tween left ventricle and myocardium and the
homogeneity of the blood signal. The signal-
to-noise ratio of the images was significantly
enhanced by the use of the phased array sur-
face coil compared with the body coil. Cer-
tainly the improved signal-to-noise ratio is not
an advantage of our sequence alone. but the
combination of the improved contrast. homo-
geneity (using dark-blood sequences). and in-
creased signal-to-noise ratio (from the use of
the phased array coil) resulted in an ability to
easily contour the left ventricle both manually
and with the computer.
Potentially, the manual interactions in-
volved in the semiautomatic niethod could
lead to observer-dependent variations in left
ventricular ejection fraction estimation. The
manual interactions involved are identifica-
tion of a seed point within the left ventricle
and tracing a region of interest encompassing
the left ventricle and the ittyocardiunt, and se-
lection of an intensity threshold separating left
ventricle blood from myocardium. We tested
the variation (arising from the manual inter-
ventions) in the estimation of ejection fraction
using the semiautontated algorithm. The in-
traobserver variability of approximately 3%
was small and close to that obtained using
manual contouring by experienced radiolo-
gists. The results showed that the semiauto-
mated algorithm can be used successfully in a
routine clinical setting.
Intraobserver variability was small (4%
and 3% for the biplane and short-axis methods,
respectively). This small variability shows that
reproducibility cannot be improved signifi-
cantly by having the san�e observer make the
nteasurements on two or more occasions and
then take the mean of these nteasurements.
The larger variability in the biplane method is
expected because small errors in the distance
determinations ofthe biplane method can give
rise to large errors in left ventricular volume
calculations. However, the ntain difference in
the two niethods was seen in the interobserver
variabilities (Fig. 1 2). The interobserver van-
ability was high for the biplane method be-
cause each observer had to make subjective
estimations of the locations of various cardiac
landmarks on the four-chamber and long-axis
images (Dl, D2. and L in Fig. 3). Conversely.
the contouring of the endocardial borders on
the short-axis images was much less subjec-
tive, especially with the improved myocar-
dium-to-blood signal intensity contrast of the
present sequence. The large interobserver
variability of the biplane method shows that it
can be used only in clinical situations in
which scan time is at a premium. In such
cases, the measurement reproducibility can
be improved by having five or six experts
perform the measurenients twice and then
taking the mean of all the measurements. Be-
cause only three distance measurements must
be made per examination, such repeated mea-
surements by different experts may be a time-
efficient clinically feasible alternative to the
short-axis method. We did not attempt to
sentiautomate the measurements required to
estimate left ventricular ejection fraction
front the biplane method. It is intrinsically
harder to semiautomate these nteasurements
compared with contouring the endocardial
borders from short-axis images because the
algorithm will have to be designed to recog-
Sinha et al.
112 AJR:169, July 1997
nize landmarks (Fig. 3). Semiautomation may
decrease the interobserver variability (com-
pared with completely manual methods) in es-
timating left ventricular ejection fraction from
biplane methods.
The interexamination variability showed
that when a subject is studied on several occa-
sions, with the different examinations read by
the same observer, the estimate of ejection
fractions may vary by approximately 9% for
the healthy heart (for both the biplane and
Simpson’s rule methods). Keeping this factor
in mind is important when attempting to detect
changes in ejection fraction, as in a follow-up
study. The interexamination variability is the
most significant source of variation in deter-
mining left ventricular ejection fractions from
serial short-axis images from the dark-blood
breath-hold sequence and was similar to that
reported earlier for MR short-axis images from
spin-echo and gradient-echo imaging and from
echocardiography [2, 14].
Interexamination variability has contnibu-
tions that are biologic and technical in origin
[2, 1 1, 14]. Physiologic differences depend on
various factors, such as the ventricular after-
load, hydration state, heart rate, and blood
pressure, that cannot be controlled easily.
Technical factors include the data acquisition
method and analysis. In data acquisition, sev-
eral factors may limit reproducibility, such as
the thickness of imaging planes, acquisition
plane positioning, and temporal location and
resolution of the imaging planes. As men-
tioned previously, one of the potential sources
of error in estimating the ejection fractions is
the temporal resolution of the MR images
limited by the acquisition window duration.
Because the present scheme also requires po-
sitioning of the end-systolic and end-diastolic
acquisition windows, we undertook a study
of the interexamination variability. For the
biplane plane method, small errors in slice po-
sitioning required for the generation of the
four-chamber and short-axis images in re-
peated MR examinations may lead to a larger
contribution from technical variability to the
total interexamination variance. It is surpris-
ing that the variability of ejection fractions
determined from the biplane and from the
short-axis methods was in the same range.
This similar variability may reflect the fact
that biologic variability contributes signifi-
cantly to the reproducibility of repeated stud-
ies [14]. All the MR imaging was carried out
with the same protocols for scout images, and
a radiologist prescribed the four-chamber and
short-axis views on the basis of the scout im-
ages. In other studies unrelated to this one [2,
1 1, 14], interexamination variabilities in the
same range of 10% were estimated using both
MR imaging (spin- and gradient-echo se-
quences) and echocardiography. These meth-
ods are significantly different from the present
method (especially echocardiography) so that
common technical contributions to variability
cannot explain the same range of interexami-
nation variabilities found across all these stud-
ies. The biologic factor is common to all these
techniques, which leads us to conclude that it
is the major source of interexamination van-
ability. In fact, the biologic factor has been
identified as a major contributor by other
studies also [2, 14]. However, the contribution
from technical sources, as detailed previously,
cannot be discounted. The determination of
the interexamination variabilities will help set
a range beyond which changes in left ventric-
ular function can be considered significant.
Reproducibility may be improved by in-
corporating a repeated examination and tak-
ing the average of two such examinations
[2]. This procedure was not clinically feasi-
ble when using spin-echo or gradient-echo
sequences. However, with the ultrafast se-
quences, it is clinically possible to actually
have two separate examinations covering
the heart from apex to base. Furthermore,
the use of thinner slices will improve accu-
racy and reproducibility. Again, the use of
ultrafast sequences will permit the acquisi-
tion of a larger number of thinner imaging
sections in reasonable scan times.
In conclusion, this paper reports our pre-
liminary feasibility study on the use of the
dual-phase, dark-blood serial short-axis im-
ages acquired by the present imaging scheme
to yield reproducible estimates of the ejection
fraction in a limited number of both patients
and healthy volunteers. The suppression of
blood in the left ventricle in combination with
the high signal-to-noise ratio obtained with
the phased array surface coil permits auto-
mated contouring of the endocardial borders.
The estimation of ejection fraction by apply-
ing the Simpson’s method to the short-axis
images is more reliable than the biplane
method; the latter is not suitable for patients
with global cardiomyopathy because of the
invalidity of the assumptions in the geometric
model in these cases. The proposed semiauto-
mated algorithm can be successfully imple-
mented to automate the estimation of ejection
fractions from dank-blood dual-phase images.
The speed of the acquisition of the images
coupled with the potential to automate their
segmentation and analysis suggests the possi-
bility of implementing this technique on a
routine clinical basis for evaluating global left
ventricle function.
References
I . Van Rossum AC, Visser FC, Sprenger M, Van Ee-
nige Mi. Valk J, Roos JR Evaluation of magnetic
resonance imaging for determination of left yen-
tricular ejection fraction and comparison with an-
giography. Am J Cardiol 1988:62:628-633
2. Pattynama PMT, Lamb HJ, van der Velde EA, van
der Wall EE, de Roos A. Left ventricular measure-
ments with cine and spin-echo MR imaging: a
study of reproducibility with variance component
analysis. Radiology 1993:187:261-268
3. Pattynama PMT, de Roos A, van der Wall EE, van
Voorthusien AE. Evaluation of cardiac function
with magnetic resonance imaging. Am Heart J
1994:128:595-607
4. Baldy C. Douek P. Croisille P. Magnin IE, Revel D,
Amiel M. Automated myocardial edge detection
from breath-hold cine-MR images: evaluation of
left ventricular volumes and mass. Magn Reson
imaging 1994:12:589-598
5. Wyatt HL. Meerbaum S. Heng MK, et al. Cross
sectional echocardiography. III. Analysis of mathe-
matical models for quantifying volume of symmet-
tic and asymmetric left ventricles. Am Heart J
1980:100:821-8286. Dell’ltalia U, Blackwell GG, Pearce Di, Thom B,
Pohost GM. Assessment of ventricular volumes us-
ing cine magnetic resonance in the intact dog: a
comparison of measurement methods. Itivest Ra-
dial 1994:29:162-167
7. Sinha S. Sinha U. Black blood dual phase Turbo-
FLASH MR imaging of the heart. J Ma,tii Reson
Imaging 1996:6:484-494
8. Edelman RR, Chien D, Kim D. Fast selective black
blood MR imaging. Radiology 1991:181:655-660
9. Liu Y. Riederer SJ. Ehman RL. Magnetization-pre-
pared cardiac imaging using gradient echo acquisi-
tion. Magn Reson Med 1993;30:27l-275
10. Rogers WJ, Shapiro EP, Weiss JL, et al. Quantifica-
tion of and correction for left ventricular systolic
long-axis shortening by magnetic resonance tissue
tagging and slice isolation. Circulation 1991;84:
721-731
I 1. Otto CM, Pearlman AS. Echocardiographic evalu-
ation of left and right ventricular systolic function.
In: Otto CM, Pearlman AS, eds. Textbook of dm1-
dal echocardiograph�� 1st ed. Philadelphia: Sam-ders, 1995:85-115
12. Ranganath S. Contour extraction from cardiac
MRI studies using snakes. IEEE Tratis Med Imag-
ing 1995: 14:328-338
13. Sakuma H, Globits S. Bourne MW, Shimakawa A,
Fm TK, Higgins CB. Improved reproducibility in
measuring left ventricular volumes and mass using
multicoil breath-hold cine MR imaging. J Magn
Reson Imaging 1996:6:124-12714. Gordon EP, Schnittger I, Fitzgerald PJ, Williams P,
Popp RL. Reproducibility of left ventricular vol-
umes by two-dimensional echocardiography. J Am
Coil Cardiol 1983:2:506-513