RP-HPLC-ESI-QTOF-MS Qualitative Profiling, Antioxidant, Anti ...
DEVELOPMENT AND VALIDATION OF ZOLEDRONIC ACID BY USING RP-HPLC Method. Shanmugasundaram.P*.,...
Transcript of DEVELOPMENT AND VALIDATION OF ZOLEDRONIC ACID BY USING RP-HPLC Method. Shanmugasundaram.P*.,...
January-March 2013 Volume 1 Issue 1 Page 62
INTERNATIONAL JOURNAL OF FRONTIERS IN SCIENCE AND TECHNOLOGY
Research Article
ISSN 2321 – 0494
www.ijfstonline.org Received on : 25.02.2013,Revised and Accepted on : 20.03.2013
DEVELOPMENT AND VALIDATION OF ZOLEDRONIC ACID BY USING RP-HPLC
METHOD
Shanmugasundaram.P*., Rasagna.,Viswanath.V., Ravichandiran V,M.Sumithra.
School of Pharmaceutical Sciences, VISTAS, Vels University, Pallavaram, Chennai –
600117.
Abstract
The present work deals with the development of a precise, accurate, simple, specific,
reliable and less time consuming RP-HPLC method for the estimation of zoledronicacid.A
newer RP-HPLC method was developed for bulk drug and formulations. The proposed method
gives reliable assay results with short analysis time using mobile phase of triethylamine (pH3) :
acetonitrile. The proposed method was developed by HPLC Shimadzu Separation Module LC-
20AT Prominence Liquid Chromatograph using Hypersil silica, C18 250× 4.6mm, 5µ column
column with an injection volume of 20 l was injected and eluted with a mobile phase
composition of Triethylamine buffer PH
3 : Acetonitrile (50:50), which is pumped at a flow rate
of 1.1ml/min and detected by uv detector at 215nm. The retention time was found to be 3.649
min thus the method is not time consuming and can be used in laboratories for the routine
analysis of formulation. The method was validated in accordance with ICH parameters. The
method was also specific and robust.
KEYWORDS: ICH parameters, RP-HPLC, zoledronicacid
Corresponding author: Dr.P.Shanmugasundaram, E.Mail:[email protected]
INTRODUCTION
Zoledronic acid is an Antihypocalcemic, bone density conservation agent chemically
known as (1-hydroxy-2 (1h-imidazol- 1-yl)-1-phasphonoethyl)phosphoric acid. It is almost
white powder and freely soluble in water. It is mainly used to treat high blood calcium levels
namely hypercalcaemia and also used in cancer chemotherapy to treat bone cancer problems
that may occur with multiple myeloma. Figure 1 gives the structure of zoledronicacid
January-March 2013 Volume 1 Issue 1 Page 63
shanmugasundaram.et.al., www.ijfstonline.org
MATERIALS & METHODS:
Equipment and Apparatus Used:
Single Pan balance (Metler Toledo)
Control Dynamics pH meter (Metler Toledo)
HPLC Shimadzu Separation ModuleLC-20AT Prominence Liquid Chromatograph
UV Visible Detector (SPD 20A Prominence)
Chromatographic data Software: Spinchrom
Hypersil silica, C18 250× 4.6mm., 5µ column.
Vacuum filter pump,(NUPORE filtrartion systems, 0.45 µ, 47mm)
Mobile phase reservoir
Ultrasonicator
Reagents used:
Acetonitrile HPLC grade
triethylamine,laboratory reagent
Water (HCL)
Orthophosphoric acid
METHOD DEVELOPMENT
Chromatographic parameters:
Instrument : HPLC Shimadzu Separation Module LC-20AT Prominence
Liquid Chromatograph
Detector : U.V – Visible detector
Column : Hypersil silica, C18 250× 4.6mm, 5µ column
Temperature : 30oC
Flow rate : 1.1ml/min
Wave length : 215 nm
Runtime : 10 min
Injection volume : 20 l
Mobile Phase: : Triethylamine buffer PH
3Acetonitrile(50:50)
Diluent : Mobile phase
Retention time : 3.649 min
January-March 2013 Volume 1 Issue 1 Page 64
shanmugasundaram.et.al., www.ijfstonline.org
Selection of Mobile phase:
A number of trials were made to find out the ideal solvent system (mobile phase) for
eluting the drug. The mobile phase containing Methanol: Water (50:50), Methanol: Phosphate
buffer PH
3 (50:50) , Acetonitrile : Triethylamine buffer PH
3 (50:50) was tried. Better peak
resolution and adequate retention time were obtained with the ratio of Triethylamine buffer PH
3: Acetonitrile (HPLC grade) (50:50)
Preparation of Mobile phase:
The mobile phase was prepared by mixing Triethylamine : Acetonitrile: (pH adjusted 3.0
with orthophosphroic acid) in the ratio of 50:50. The mobile phase is then sonicated using UV-
Sonicator to remove the impurities and dissolved gases, as they may lead to unwanted peaks in
the chromatogram.
Preparation of Triethylamine buffer:
Dissolve 4.0 ml of Triethylamine in 100oml of water to produce 0.02% buffer pH is
adjusted to 3.0 with 10% v/v Orthophosphoric acid.
Preparation of standard stock solution:
Stock solution of zolidronicacid(1mg/ml) was prepared by dissolving 25mg of
Zoledronicacid in 25 ml of volumetric flask containing mobile phase. The solution was
sonicated for about 10 min and then made up to volume with mobile phase. Daily working
standard solutions of Zolidronicacid was prepared by suitable dilution of the stock solution with
appropriate mobile phase.
Preparation of sample solution:
5 vials of zoledronicacid injection 25ml was transferred into volumetric flask, 70 ml of diluent
was added to it and was shaken by mechanical stirrer and sonicated for about 10 minutes by
shaking at intervals of five minutes each and was diluted up to the mark with diluent and the
solution was filtered through 0.45 m filter before injecting into the HPLC system.
Procedure:
Separately inject the blank, triplicate injections of standard and duplicate injections of
sample preparations to the HPLC system and record the chromatograms. The results for sample
were shown in table.
METHOD VALIDATION
1 LINEARITY:
The linearity of the method was demonstrated over the concentration range of 60-140µg
/ ml. Aliquots of 80-120 µg / ml were prepared from stock solution and labeled as solution 1, 2,
January-March 2013 Volume 1 Issue 1 Page 65
3, 4,5 respectively. The solutions were injected in to HPLC system as per test procedure. A
calibration curve was plotted for concentration Vs peak area and was given in the Figure1.
shanmugasundaram.et.al., www.ijfstonline.org
Acceptance criteria:
Correlation Coefficient (r2) should be not less than 0.9990.
2 .PRECISION
Method precision:
Five samples of 100 µg/ml solutions were prepared from stock solution and injected into
the HPLC system as per test procedure.
Acceptance criteria:
% Relative standard deviation of results should not be more than 2.
System precision:
Five samples of 100 µg/ml solutions were prepared from stock solution and injected into
the HPLC system as per test procedure.
Acceptance criteria:
% Relative standard deviation of results should not be more than 2.
Intraday Precision:
Intraday precision of test method is demonstrated by 7 injections of the same batch(same conc)
of samples at 0, 1, 2, 3, 4,5 and 6 hrs on same day.
Interday Precision:
Interday precision of test method is demonstrated by 5 injections of the same batch(same conc)
of samples on 5 successive days.
3 . ACCURACY
The accuracy of the method was evaluated by determination of recovery of
zoledronicacid at three levels of concentrations. The sample solutions were spiked with
zoledronicacid standard solutions corresponding to 80, 100, and 120% of nominal analytical
concentrations. (80µg/ml, 100µg/ml and 120µg/ml). The results showed good recovery within
limits (98% – 102%).
4 SPECIFICITY
A) zoledronicacid Identification
Solutions of Standard and Sample were prepared as per test procedure and injected into the
HPLC system.
B) Stress degradation studies by treating with alkali and acid:
Zoledronicacid was treated with 0.1M NaOH and 0.1M HCl. Solutions of sample was
prepared as per test procedure and injected into the HPLC system.
January-March 2013 Volume 1 Issue 1 Page 66
shanmugasundaram.et.al., www.ijfstonline.org
5 ROBUSTNESS:
The robustness of the proposed method was determined by analysis of aliquots from
homogenous lots by differing physical parameters like flow rate and mobile phase composition
which may differ but the responses were still within the specified limits of the assay.
Effect of variation of Mobile Phase Composition
A study was conducted to determine the effect of variation in mobile phase ratio by Changing
the ratio of mobile phase i.e Triethylamine: Acetonitrile: (pH adjusted 3.0withorthophosphroic
acid)in the ratio of 50:50. Standard solution was prepared and injected into the HPLC system
and recorded the chromatograms.
Acceptance criteria
Tailing Factor of zoledronicacid standard should not be more than 2 for Variation in
composition of mobile phase.
The % RSD of asymmetry of zoledronicacid standard should be not more than 2 for
Variation in composition of mobile phase.
Effect of variation in Buffer PH
A study was conducted to determine the effect of variation in Buffer PH. Standard solution
was prepared and injected into the HPLC system by maintaining PH 3.5and 4.5. The effect of
variation in Buffer PHwas evaluated. The results were discussed in the Table: 5
Acceptance criteria
The tailing factor of standard should be not more than 2.0 for Variation in flow.
The % RSD of Asymmetry of zoledronicacid standard should be not more than 2 for variation
in flow.
Effect of variation of Flow Rate
A study was conducted to determine the effect of variation in flow rate. Standard solution
was prepared and injected into the HPLC system by keeping flow rates 0.8 ml / min and 1.2 ml /
min. The results were discussed in the Table: 6
Acceptance criteria
The tailing factor of standard should be not more than 2 for Variation in flow.
The % RSD of ofzoledronicacid standard should be not more than 2 for variation in
flow.
January-March 2013 Volume 1 Issue 1 Page 67
Effect of variation in Temperature
A study was conducted to determine the effect of variation in temperature. Standard
solution was prepared and injected into the HPLC system by keeping temperature 30˚c, 27˚c
and 25˚c. The effect of variation of temperature was evaluated. The results were discussed in the
Table: 7
shanmugasundaram.et.al., www.ijfstonline.org
Acceptance criteria
The tailing factor of standard should be not more than 2.0 for Variation in flow.
The % RSD of asymmetry of zoledronicacid standard should be not more than 2.0 % for
variation in flow.
6 RUGGEDNESS:
Ruggedness is the degree of reproducibility of the results obtained under a variety of
conditions. It is checked that the results are reproducible under differences in conditions,
analysts and instruments. And hence the proposed method was found to be rugged.
The Chromatograms were shown in Fig:8
7 LIMIT OF DETECTION (LOD)
The parameter LOD was determined on the basis of response and slope of the regression
equation. The LOD for this method was found to be 0.989.
8 LIMIT OF QUANTIFICATION (LOQ)
The parameter LOQ was determined on the basis of response and slope of the regression
equation. The LOQ for this method was found to be 2.1872.
9. SYSTEM SUITABILITY
A Standard solution of zoledronicacid working standard was prepared as per procedure
and was injected six times into the HPLC system. The system suitability parameters were
evaluated from standard Chromatograms obtained by calculating the % RSD of retention times,
tailing factor, theoretical plates and peak areas from six replicate injections.
Acceptance criteria: The % RSD for the retention times of principle peak from 6
replicate injections of each Standard solution should be not more than 2
The number of theoretical plates (N) for the zoledronicacid peaks should be not less than
2000. The Tailing factor (T) for the zoledronicacid peaks should be not more than 2
From the system suitability studies it was observed that all the parameters were within limit.
Hence it was concluded that the Instrument, Reagents and Column were suitable to perform the
assay.
January-March 2013 Volume 1 Issue 1 Page 68
The results were discussed in Table 3
ASSAY
Assay of different formulations available in the market was carried by injecting sample
corresponding to equivalent weight into HPLC system. And percentage purity was found out
by following formulae. Recovery studies were carried out.]
shanmugasundaram.et.al., www.ijfstonline.org
Calculated the percentage purity of zoledronicacid present in tablet using the formula:
Splarea x Stddilx Avgwt x P
Percentage purity = ----------------------------------------
Stdarea x Spldil x L.C x 100
Where,
P = (%) potency of zoledronicacid working standard use
L.C = Label claim
Avgwt = Average weight of tablets
RESULTS AND DISCUSSIONS
LINEARITY: In the new RP-HPLC method for zoledronicacid by using Methanol:
Acetonitrile: Acetate Buffer (40:40:20) as a mobile phase we obtained a linearity curve by
taking the sample at concentrations ranging from 80-120 µg/ml. Retention time – 3.42. From
the peak response obtained correlation coefficient was found to be 0.999.
ACCURACY
The accuracy of the method was evaluated by determination of recovery of
zoledronicacid at three levels of concentrations. The sample solutions were spiked with
zoledronicacid standard solutions corresponding to 80, 100, and 120% of nominal analytical
concentrations. (80µg/ml, 100µg/ml and 120µg/ml). The results showed good recovery within
limits (98% – 102%).
PRECISION
Method precision:
Five samples of 100 µg/ml solutions were prepared from stock solution and injected into
the HPLC system as per test procedure.
The % Relative standard deviation of results should be 0.09.
System precision:
January-March 2013 Volume 1 Issue 1 Page 69
Five samples of 100 µg/ml solutions were prepared from stock solution and injected into
the HPLC system as per test procedure.
The % Relative standard deviation of results should be 0.10.
shanmugasundaram.et.al., www.ijfstonline.org
SYSTEM SUITABILITY
A Standard solution of zoledronicacid working standard was prepared as per procedure
and was injected six times into the HPLC system. The system suitability parameters were
evaluated from standard Chromatograms obtained by calculating the % RSD of retention times,
tailing factor, theoretical plates and peak areas from six replicate injections.
The % RSD for the peak areas of principle peak from 6 replicate injections of each Standard
solution should be 0.07 (limit should be less than-2)
ROBUSTNESS: The studies were carried out by using different chromatographic condition
like follows:
a) Change in flow rate
b) Change in organic phase composition in the mobile phase
c) Change in pH of mobile phase
The % RSD was found to be below 2% i.e. within the limit. Thus the results show that the
method is precise under different a chromatographic condition which is shown in table below.
LIMIT OF DETECTION (LOD)
The LOD for this method was found to be 0.989
LIMIT OF QUANTIFICATION (LOQ)
The LOQ for this method was found to be 2.1872
All the ICH validation parameters were found to be within the limits
CONCLUSION:
The praposed analytical method is simple, economical
,rapid,sensitive,reproduciable,and accurate for the estimation of zolidronic acid.A newer RP-
January-March 2013 Volume 1 Issue 1 Page 70
HPLC method was developed for bulk drug and formulations. The praposed method gives
reliable assay results with short analysis time using mobile phase of triethylamine (pH3) :
acetonitrlie
The above method does not suffer from any interferance due to common
excipients. Thus it was show that praposed methods could be successfully appiled toestimate
commercial pharmaceutical products contining zolidronic acid .Thus the above findings will
enable the quantification of the drug for future investigation in the field of analytical chemistry.
shanmugasundaram.et.al., www.ijfstonline.org
Figure 1: Zoledronicacid structure
Figure 2: linearity
Figure 3: chromatogram of of zoledronicacid.by RP-HPLC Method
January-March 2013 Volume 1 Issue 1 Page 71
shanmugasundaram.et.al., www.ijfstonline.org
Figure 4: System Suitability of zoledronicacid.
Table 1: Standard Concentration To Peak Response For zoledronicacid
Parameters
Zoledronicacid
215nm
Concentration (µg/ml) 80-120 µg/ml
January-March 2013 Volume 1 Issue 1 Page 72
Correlation 0.999
Slope 22481.8
Y- intercept 17879
Acecptance criteria: The Correlation coefficient (r2) should be more than 0.99
shanmugasundaram.et.al., www.ijfstonline.org
Table 2: specificity of zoledronicacid.
Parameters Zoledronicacid
Area under curve
2106934
20699684
2107438
2341854
2345450
Mean 2676908
SD 25079
%RSD 0.80
Retention time (min) 3. 42
January-March 2013 Volume 1 Issue 1 Page 73
Theoretical plates 5919
Tailing factor 1. 24
shanmugasundaram.et.al., www.ijfstonline.org
Table 3: System suitability for Zoledronicacid by RP-HPLC
Parameter Observed Value
USP tailing 1.24
Retention time 3.43
January-March 2013 Volume 1 Issue 1 Page 74
shanmugasundaram.et.al., www.ijfstonline.org
REFERENCES
[1] Michael , schartz , krull. Analytical method development and validation 2004; 24-46
[2] Sharma bk. Instrument methods of chemical analysis ,Introdiuction analysis chemistry ,goel
publishing house meerut, 23th edition; 2004, p. 223-38
[3] Basic edition in analytical chemistry,17th
edition; 2001, p. 323-42
S.No Area response of Zoledronic acid
1.0 2086027
2.0 2084140
3.0 2087209
4.0 2088028
5.0 2088003
6.0 2087320
Average 2086788
SD 2145
%RSD 0.07
January-March 2013 Volume 1 Issue 1 Page 75
[4] Willard hh, Merritt ll. Instrumental method of analysis, cbs publishers and distributors ,
New delhi, 7th
edition; 1986, p. 256-64
[5] lyles kw et al., zoledronicacid and clinical fractures and mortality after hip fractures. 2007;
sep 26.
[6] Durie bg, katz m, Crowley j. Osteonecrosis of the jaw and bisphosphonates. N engl j med
2005; jul 7:353 (1) : 99-102.
[7] Francois legay Sonia gauron fabienne deckert , ghislaine gusset Ulrike pfaar. Christian
ravera hansjorg wiegand journal of pharmaceutical and biomedical analysis 2002; 30(4):
897-911
[8] Nadan srinivas., journal of liquid chromatography and related technologies , 2011; 30(4):
476-89.
[9] Mallikarjuna rao m. k., srinivasu ch., prathima rani s., shiva kumar p., rajendra kumar k b,
chandrashekar .m., veerenra., Journal of pharmaceutical and biomedical analysis 2005;
15(3-4) : 781-90.
[10] Xiao Keng Liu, Jiang B.Fang, Nina Cauchon Pengzu Zhou, Journal Of Pharmaceutical
And Biomedical Analysis. 2008;46(4): 639-44.
[11] zan xie ye jiang di qun zhanng journal of chromatography A, 173-76.
[12] Rakesh Kumar Singh, RP-HPLC Method Development And Validation For
Simultaneous Estimation Of Ranitidine Hydrochloride And Domperidone In Combined
Tablet Dosage Form. Ijpsr, 2010; 1(8).