PERBEDAAN KADAR TOTAL PROTEIN DALAM …repository.unimus.ac.id/1183/1/ABSTRAK.pdf · Laboratorium...
Transcript of PERBEDAAN KADAR TOTAL PROTEIN DALAM …repository.unimus.ac.id/1183/1/ABSTRAK.pdf · Laboratorium...
v
PERBEDAAN KADAR TOTAL PROTEIN DALAM SERUM MENGGUNAKAN
REAGEN BIURET YANG DILETAKKAN DALAM ALAT KIMIA ANALYSER
SEGERA, 24 JAM, 48 JAM
DAN 72 JAM
Setyo Budi Raharjo 1, Budi Santosa 2, Harun Nurrachmat3
1. Program Study DIV Analis Kesehatan Fakultas Ilmu Keperawatan dan Kesehatan
Universitas Muhammadiyah Semarang
2. Laboratorium Patologi Klinik Fakultas Ilmu Keperawatan dan Kesehatan Universitas
Muhammadiyah Semarang.
3. Instalasi Laboratorium Klinik RSUD Tugurejo Semarang
ABSTRAK
Pemeriksaan kadar total protein serum menggunakan alat kimia analyser. Stabilitas reagen biuret
selama 1 tahun pada suhu 2 - 25° C, 6 bulan pada suhu ruang ( 25 - 35° C ) atau sebelum masa
kadaluwarsa yang tertera pada label etiket. Reagen Biuret jika sudah diletakkan pada alat kimia
analyser, setelah ± 2 hari ( ± 48 jam ) digunakan untuk running pemantapan mutu internal, hasil
pemeriksaan keluar dari nilai rentang target. Tujuan Penelitian untuk mengetahui perbedaan kadar
total protein dalam serum yang menggunakan reagen biuret yang diletakkan dalam kimia analyser
segera, 24 jam, 48 dan 72 jam. Jenis penelitian adalah penelitian analitik kuantitatif. Sampel
diambil secara random berjumlah 10 sampel, kemudian diperiksa menggunakan reagen yang
diletakkan dalam alat kimia analyser selama segera, 24 jam, 48 jam dan 72 jam. Hasil pemeriksaan
menunjukan rerata hasil kadar protein serum menggunakan reagen yang diletakkan pada alat kimia
analyser segera: 6.8 gr/dl, selama 24 jam : 6.6 gr/dl, selama 48 jam : 6.3 gr/dl dan selama 72 jam :
5.9 gr/dl. Hal ini menunjukkan kecenderungan menurun antara kadar total protein dalam serum
menggunakan reagen biuret yang diletakkan dalam alat kimia analyser segera, selama 24 jam, 48
jam dan 72 jam. Uji statistik Kruskal Wallis menunjukkan nilai kemaknaan 0.000 dengan taraf
kemaknaan 0.05 yaitu 0.000 < 0.005 sehingga dapat disimpulkan bahwa ada perbedaan yang
bermakna antara kadar total protein yang diperiksa menggunakan reagen biuret yang diletakkan
pada alat kimia analyser selama segera, 24 jam, 48 jam dan 72 jam.
Kata kunci : Kadar total protein, lama penyimpanan reagen biuret, alat kimia analyser.
http://repository.unimus.ac.id
vi
DIFFERENCES OF TOTAL PROTEIN CONTENT IN SERUM USING
REAGEN BIURETS IMPLEMENTED IN CHEMICAL ANALYSER
IMPORTANT IMMEDIATELY, 24 HOURS, 48 HOURS
AND 72 HOURS
Setyo Budi Raharjo 1, Budi Santosa 2, Harun Nurrachmat3
1. Four Years Diploma Study Program Health Analyst Faculty of Nursing and Health University
of Muhammadiyah Semarang
2. Clinical Pathology Laboratory Faculty of Nursing and Health Sciences University of
Muhammadiyah Semarang
3. Installation of Clinical Laboratory of Tugurejo Hospital Semarang
ABSTRACT
Examination of total serum protein content using a chemical analyzer. Stability of the biuret
reagent for a year at a temperature of 2 - 25 ° C, 6 months at room temperature (25 - 35 ° C) or
before the expiration date indicated on the label of etiquette. Biuret reagents if already placed on
the chemical analyzer, after ± 2 days (± 48 hours) is used for running quality control out of the
results of the target range. The objective of the study was to investigate the difference in total
serum protein levels using biochemical reagents placed in chemical analyzers immediately, 24
hours, 48 and 72 hours. The type of research is quantity analytical research. Samples were collected
by random samples of 10 samples, then examined using reagents placed in the chemical analyzer
for immediate, 24 hours, 48 hours and 72 hours. The results showed that serum protein
concentrations were performed with reagents placed on the chemical analyzer immediately: 6.8 g /
dl for 24 hours: 6.6 g / dl, for 48 hours: 6.3 g / dl and for 72 hours: 5.9 g / dl. This shows a
declining trend between total serum protein levels using biuretic reagents placed in the chemical
analyzer immediately, for 24 hours, 48 hours and 72 hours. Kruskal Wallis statistic test showed the
significance value of 0.000 with significance level of 0.05 0.000 < 0.005 so it can be concluded
that there is a significant difference between the total protein content examined using biuret reagent
placed on chemical analyzer immediate, for 24 hours, 48 hours, and 72 hours. .
Keywords: Total protein content, long storage of biuret reagent, chemical analyzer.
http://repository.unimus.ac.id