Analytical Methods ValidationM. Hatta PrabowoDepartment of Pharmaceuticals ChemistryUniversitas Islam Indonesia
Chemical analysisThe bioanalytical part of bioequivalence trials should be conducted according to the applicable principle of Good Laboratory Practice (GLP)
Good Laboratory Practice (GLP) Test plan (Analytical protocol) Sample traceability Documentation, possible to reconstruct
the study Analytical method validation report Analytical report signed by responsible
investigator
Pre-study phase The method used must be well
characterised Stability Specificity Accuracy Precision Limit of quantitation Response function
Validation objectiveTo demonstrate that the analytical procedure is suitable for its intended purpose
Analytical method validation
Analytical Procedure
Selectivity
Accuracy
Precision- within-run- between-run
Recovery
Limit of Quantitation LOQ
Calibration curve
Robustness
Validation
Stability
Specificity (selectivity) Ability of an analytical method to
measure only what it is intended to measure
Blank samples from six different subjects Will other drugs, metabolites or
endogenous components interfere in the measurements?
AccuracyThe closeness of mean test results obtained by the analytical method to the true value (concentration) of the analyte.
Accuracy Accuracy should be measured at
minimum 3 levels At least 5 determinations per
concentration The mean value should be within 15% of
the actual value At the lower limit of quantitation level
within 20% is acceptedxx xx xx xxxx
xx
PrecisionThe closeness of individual measurements of an analyte when the procedure is applied repeatedly to multiple aliquotes of a single homogenous volume of biological matrix
Precision Precision should be measured at
minimum 3 levels At least 5 determinations per
concentration The calculated CV should not exceed 15% At the lower limit of quantitation level, CV
should not exceed 20% Subdivided into within-run and between-
run
Precision and Accuracy
Conc.nmol/l
Accuracy(%)
Precision
% %
n
0.76 0.6 5.1 6.1 1823 3.6 1.7 1.8 18122 3.9 1.3 1.3 18
Within-run Between-run
Lower limit of quantitation The lowest standard on the calibration curve should be accepted as the lower limit of quantitation (LLOQ)
if
Lower limit of quantification The analyte responce at LLOQ is at least
5 times the blank response The peak should be identifiable and
discrete Precision within 20% CV Accuracy of 80-120%
LLOQ (1.50 nmol/l) for morphine
Sample No. nmol/l Accuracy LLOQ 1 1.58 5.3% LLOQ 2 1.61 7.3% LLOQ 3 1.46 -2.6% LLOQ 4 1.44 -4.0% LLOQ 5 1.50 0.0% LLOQ 6 1.49 -0.7% Mean: 1.51 0.9%
SD: 0.067 CV%: 4.5
Calibration/Standard curve
A calibration curve is the relation between instrument response and known concentrations of the analyte
Should be prepared in the same biological matrix as the samples
Should consist of 6-8 samples covering the expected range Should include LLOQ and a blank sample Should include a zero sample (with internal standard) Same curve fitting, weighting in prestudy and study Any changes should be documented
Sample dilution Any required sample dilutions should use
like matrix Dilution QC sample should be used
85
90
95
100
105
110
115
0 10 20 30 40 50 60 70 80 90 100 110
Sample No.
Foun
d co
ncen
trat
ion
%
Robustness
Stability of your substance
In the automatic injector
In plasma during storage
In room temperature (4 h)
In Freeze/Thaw tests
In stock solutions
Analytical method validation
Analytical Procedure
Selectivity
Accuracy
Precision- within-run- between-run
Recovery
Limit of Quantitation LOQ
Calibration curve
Robustness
Validation
Stability
References1. Guidance for Industry
Bioanalytical Method Validation FDA, May 2001
2. Workshop Report: Shah, V.P. Et al., Pharmaceutical Research: 1992; 9:588-592.
3. Workshop Report: Shah, V.P. et al., Pharmaceutical Research: 2000; 17:1551-1557
Costs Validation = 130-180.000 SEK Stability = 15-20.000 SEK for each time
point QA = 11.000 SEK/study
The study phase (1) ...in which the validated bioanalytical
method is applied to the actual analysis of samples from the biostudy mainly in order to confirm the stability, accuracy and precision.
The study phase (2) Calibration curve in each run Six Quality Control samples in each run Pre-stablished SOPs for procedures (method) Acceptance criteria for a run
- accuracy and precision of the calibration curve- accuracy and precision of the QC samples- repeat analysis
It is preferable to analyse all study samples from a subject in a single run
The study phase (3) The QC samples should be used to
accept or reject the run (2 samples at 3 levels)
Four QC samples out of six should be within 15% of their nominal value
Two QC samples can be outside ±15% but not both at the same concentration
System suitability test
Signal to noise ratio is above 5 for the substance.
The peak shape is acceptable after visual inspection of the chromatogram
The retention times are within 10% of the previous run.
The lowest calibration sample is injected before each run.
The system is accepted if:
The analytical report should include
Results for all calibration curves Results for all quality control samples Representative number of chromatograms Should include data from subjects who
eventually dropped-out Reanalysed samples and the reason for
reanalyses The analytical validation report The responsible investigator(s) should sign
for their respective section of the report
Chiral active substances The bio-analytical method should be
enantiomeric Unless
Both products contain the same stable singel enantiomer
Both products contain the racemate and both enantiomers show linear pharmacokinetics
Top Related