Schedule-induced drinking as a function of percentage reinforcement1
VISCOSIMETRY -APPLICATION Laboratory excersise 1: Determination of percentage of 1,2-diol bonds in...
Transcript of VISCOSIMETRY -APPLICATION Laboratory excersise 1: Determination of percentage of 1,2-diol bonds in...
1
VISCOSIMETRY - APPLICATION
Laboratory excersise 1: Determination of percentage of 1,2-diol bonds in
polyvinyl alcohol from limiting viscosity number1
Introduction
While the basic chemical structure of a synthetic polymer is usually well understood, many physical
properties depend on such characteristics as chain length, degree of chain branching, and molar
mass, which are not easy to specify exactly in terms of a molecular formula. Moreover the
macromolecules in a given sample are seldom uniform in chain length or molar mass (which for a
linear polymer is proportional to chain length); thus the nature of the distribution of molar masses is
another important characteristic. A polymer whose molecules are all of the same molar mass is said
to be monodisperse; a polymer in which the molar masses vary from molecule to molecule is said to
be polydisperse. Specimens that are approximately monodisperse can be prepared in some cases by
fractionating a polydisperse polymer; this fractionation is frequently done on the basis of solubility in
various solvent mixtures.
This experiment is concerned with the linear polymer polyvinyl alcohol 2CH CHOHn
(PVOH), which is prepared by methanolysis of the polyvinyl acetate (PVAc) obtained from the direct
polymerization of the monomer vinyl acetate 3 2CH COO CH=CH . As ordinarily prepared,
polyvinyl alcohol shows a negligible amount of branching of the chains. It is somewhat unusual
among synthetic polymers in that it is soluble in water. This makes polyvinyl alcohol commercially
important as a thickener, as a component in gums, and as a foaming agent in detergents.
A characteristic of interest in connection with PVOH and PVAc is the consistency of orientation of
monomer units along the chain. It is assumed in the formula 2CH CHXn
, where X denotes –
OH nebo –OOCCH3 group that all monomer units go together "head to tail", it means the groups
2CH a CHX are bonded. skupina. However, occasionally a monomer unit will join onto the
chain in a "head-to-head" fashion, yielding a chain of the form
2 2 2 2CH CHX CHX CH CHX CH CHX CHm m
The frequency of head-to-head linkage depends on the relative rates of the normal growth-step reaction α
k
R M R M (1.1)
and the abnormal reaction ,
k
R M R M (1.2)
where R is the growing polymer radical, the arrow representing the predominant monometer
orientation). The rates in tum must depend on the activation energies:
1 Shoemaker, Garland, Nibler, "Experiments in Physical Chemistry," McGraw-Hill Publishing Company, Toronto
using different experimental equipment; The book is at disposal in the Department of physiacal chemistry library
2
exp
exp
exp
a
a
a
EA
k RT ES
Ek RTA
RT
(1.3)
where a a aE E E is the additional thermal activation energy needed to produce abnormal
addition and S is the steric factor representing the ratio of the probabilities that the abnormally and
normally approaching monomer will not be prevented by steric or geometric obstruction from being
in a position to form a bond. If the (1.1) reaction is faster than the reaction (1.2), it is probable that
the activation energy of the first reaction aE is lower than the activation energy of the second
reaction aE . This hypothesis was confirmed by Flory and Leutner, when they found out that the
frequency of head-to-head linkages CHX CHX in PVAc increase with increasing
polymerization temperature. The values of steric factor S and difference of activation energies
a a aE E E determine wheter the formation of the “head-to-head linkages“ CHX CHX is
reduced by the steric effect or by thermal activation barrier.
In this experiment, the method of Flory and Leutner will be used to determine the fraction of head-
to-head attachments in a single sample of PVOH. The method depends on the fact that in PVOH a
head-to-head linkage is a 1,2-glycol structure, and 1,2-glycols can bespecifically and quantitatively
cleaved by periodic acid or periodate ion. Treatment of PVOH with periodate should therefore break
the chain into a number of fragments, bringing about a corresponding decrease in the effective
molar mass. All that is required is a measurement of the molar mass of a specimen of PVOH before
and after treatment with periodate.
Experimental determination of molar mass of polymers
For the determination of very high molar masses, freezing-point depressions, boilingpoint elevations,
and vapor-pressure lowerings are too small for accurate measurement. Osmotic pressures are of a
convenient order of magnitude, but measurements are timeconsuming. The technique to be used in
this experiment depends on the determination of the intrinsic viscosity of the polymer. However,
molar-mass determinations from osmotic pressures are valuable in calibrating the viscosity method.
Two other methods, light scattering and gel permeation chromatography (GPC), could also be used
as alternatives or in addition to the viscosity technique. Light scattering is a classical method for
determining molar masses, and if scattering as a function of angle is measured, it is also possible to
learn something of the polymer shape, i.e., to distinguish rodlike from spherical forms. In the case of
gel permeation chromatography, molecules of different sizes are carried by a liquid solvent over a
substrate gel network of uniformly sized pores, which can be chosen in a range from 10 to 100,000
nm. Smaller solute molecules diffuse into and out of the pores and hence exhibit a longer retention
time than larger solute molecules, which are not retained and travel through the column with the
solvent. This method is simple, but to obtain accurate molar masses of a given polymer type from the
retention times it is necessary to reproduce the flow rates carefully and to calibrate the column with
polymers of known molar mass and similar chemical composition.
As it was mentioned above, molar mass will be determined by viscosity measurement in this work.
Viscosity will be mesured by the Stabinger SVM 3000 instrument from Anton Paar company, which is
3
a special rotation viscometer. The rotation viscometer is, generally, based on the fact that there is a
function dependence of torque, which develops at constant rotation of a body in liquid, on viscosity
of this liquid. Stabinger SVM3000 differs from standard rotational viscometers by a patented feature:
hollow cylinder freely floats in liquid and is centred in the cell during rotation of outer jacket by
centrifugal forces. This arrangment eliminates unwanted influence of mechanical friction between
mechanical parts of normal rotational viscometers, where rotating body is connected directly to the
motor.
Theory of evaluation of molar mass from viscosity measurement
Einstein showed that viscosity of a fluid in which small rigid spheres are present in dilute and
uniform suspension is related to the viscosity0 of the pure fluid (solvent) by the expression
0
51
2
v
V
(1.4)
where v is the volume occupied by all the spheres and V is the total volume. The left side of the
equation (1.4) is denoted as a increment of relative viscosity i (specific viscosity sp ). For
nonspherical particles the numerical coefficient on the right side of equation (1.4) is higher than 5 2 ,
but should be still a constant for any given shape provided the rates of shear are sufficiently low to
avoid preferential orientation of the particles.
Limiting viscosity number (formerly intrinsic viscosity) is defined as the ratio of the specific
viscosity to the weight concentration of solute, in the limit of zero concentration
2 20 0
2 2 0
1lim lim lni
w ww w
(1.5)
where 2w is conventionally defined as the concentration in grams of solute per 100 mL of
solution/suspension. Both 2i w and 2 01 lnw show a reasonably linear concentration dependence at
low concentrations 2w . The dependence 2 01 lnw vs.
2w usually has small negative slope, while
2i w vs. 2w plot has larger positive slope. Therefore both plots can be used for the extrapolation to
obtain , although 2 01 lnw for a single dilute solution will give a fair approximation to .
If the internal density of the spherical particles (polymer molecules) is independent of their size (i.e.,
the volume of the molecule is proportional to its molar mass), the intrinsic viscosity should be
independent of the size of the particles and hence of no value in indicating the molar mass. This,
however, is not the case; to see why we must look into the nature of a polymer macromolecule as it
exists in solution.
Gauss (statistical) polymer coil
A polymer such as PVOH contains many single bonds, around which rotation is possible. If the
configurations around successive carbon atoms are independent and unrelated, it will be seen that
two parts of the polymer chain more than a few carbon atoms apart are essentially uncorrelated in
regard to direction in space. The molecule is then "statistically coiled" and resembles a loose tangle
of yarn (see Fig.1):
4
Figure 1. – a scheme of Gauss (statistical) polymer coil
Simple statistical treatments show that the mean distance between the two ends of the chain, and
indeed also the effective mean diameter d of the coiled molecule regarded as a rough sphere, should
be proportional to the square root of the chain length and thus to the square root of the molar mass:
1 2d M (1.6)
The volume mv occupied by a molecule should then vary as 3 2M . The number of molecules in a
given weight of polymer varies inversely with the molar mass M; hence the total volumeof the spheres is
3 2 1 2cVv M cVM
M (1.7)
Two equations (1.5) and (1.7) give
1 2KM (1.8)
where K is a constant. This treatment is much simplified; it ignores among other things the problem
of "excluded volume"; that is, the fact that the chain cannot coil altogether randomly because it is
subject to the restriction that no two parts of the chain may be at the same point in space at the
same time. This restriction becomes more and more important the higher the molar mass. Even
more serious is the effect of solvent; the above treatment tacitly assumes a "poor solvent," which
would barely get the polymer into solution. A "good solvent," by solvating the polymer, makes the
size of the statistical coil increase faster with chain length than it otherwise would (equation (1.7)),
owing to enhancement of the excluded volume effect. Accordingly, instead of equation (1.8) we
might write
aKM (1.9)
where K and a are empirical parameters characteristic both of the polymer itself and of the
solvent. The exponent a varies from about 0.5, for well-coiled polymer molecules in a poor solvent,
to as much as 1.7 for a rigidly extended "rodlike" polymer molecule.
Flory and Leutner,l working with monodisperse specimens of PVOH differing from one another in
molar mass over a wide range (obtained by fractionating polydisperse commercial PVOH),
established a correlation between the molar mass, as determined from osmotic pressure
measurements, and the increment of relative viscosity i . They found that for PVOH in aqueous
solution at 25°C the values of K a a parameters from equation (1.9) are
4 0,762,0 10 M (1.10)
and if molar mass is expressed then
5
1,3247,6 10M (1.11)
where molar mass M is in the g/mol unit. molární hmotnost je v g/mol. The equation (1.11) also
holds for a polydisperse sample of PVOH, but the'molar mass in this case is M , the viscosity-
average molar mass defined below.
Number-Average Molar MassNM and Viscosity-Average Molar Mass M
For a polydisperse polymer, any determination of molar mass must yield an average of some sort.
When a colligative property such as osmotic pressure is used, the average is a number average NM 2
0
0
N
MP M dM
M
P M dM
(1.12)
where P M is the molar-mass distribution function; that is, P M dM is proportional to the
number of molecules with molar masses M and M dM . Viscosity-Average Molar Mass M
obtained from the value of limiting viscosity number is given by the equation3
1
0
0
aM P M dM
M
MP M dM
(1.13)
and its value is different from the NM value in the case of polydisperse polymer (certainly except for
the case of monodisperse polymer, where values of molar mass determined by osmotic pressure and
viscosity measurement, respectivally, are the same ). The ratio NM M is a constant given by a
form of the distribution function P M and by value of the parameter a . A commonly encountered
distribution function, and one that is likely to be valid for PVOH, is one that arises if the probability of
a chain-termination reaction during the polymerization is constant with time and independent of the
chain length already achieved. It is also the most likely function for the product resulting from
cleavage with periodate if the head-to-head structures CHX CHX can be assumed to be
randomly distributed along the PVOH chains. This distribution function is
2 If there is a countable number of monodisperse fractions, then N i i
i
M x M , where ix je molar fraction
of i-th fraction and iM is its molar mass
3 If there is a countable number of monodisperse fractions, then
1 a
a
i i
i
M W M
, where iW is mass
fraction of i-th fraction and iM is its molar mass
6
1
NM M
N
P M eM
(1.14)
If this distribution function is used in the equations (1.12) a (1.13), the folowing relation can be
derived:
1
1 1a
NM M a a (1.15)
where is the gamma function (Euler integrál of the second kind). For 0,76a (the value given
above for polyvinyl alcohol) we obtain
1,89NM M (1.16)
For the sake of completeness it can be added that so called mass-average molar mass WM can be
defined WM 4
2
0
0
W
M P M dM
M
MP M dM
(1.17)
Series of different molar mass values for one sample is as follows:
pro 0,5 1
pro 1
N W
N W
M M M a
M M M a
(1.18)
Determination of Frequency of Head-to-Head CHX CHX Bonds in a PVOH Sample
This determination is based on the assumption that degradation arises exclusively from cleavage of
1,2-glycol structures and that all such structures are cleaved. The relative frequency of head-to-
head CHX CHX bonds in a PVOH sample is equal to the increase in the number of molecules
present in the system, divided by the total number of monomer units represented by all molecules in
the system:
0
0 0
M C C
CM CM
N N N
N N
(1.19)
MN is a number of reversly head-to-head oriented monomer units, 0
CMN is a total number of all
monomer units in the whole system, 0
CN is a total number of all polymer particles in an original
4 If there is a countable number of monodisperse fractions, then W i i
i
M W M , where iW is mass fraction
of i-th fraction and iM is its molar mass
7
sample and CN is a total number of all polymer particles in the sample after cleaving of the 1,2-
glycol structures. Because numbers of particles in the relation (1.19) are in inverse proportion to the
respective molar masses, it is possible to write
0
1 1
1
N NM M
M
(1.20)
where aN NM M jsou number average molar masses before and after degradation (cleaving of the
1,2-glycol bonds), respectivally, and 0M is molar mass of the monomer unit, so 44 g/mol in the case
of PVOH. Therefore
44 1 1N NM M (1.21)
and using the equation (1.16) we get the relation with viscosity-average molar masses
83 1 1M M (1.22)
Experimental setup
1. Preparation of the stock solutions. The preparation of PVOH solutions is time-consuming,
because it is necessary to dissolve PVOH in hot water and to stir gently to avoid a formation of
foam. Therefore it is better to prepare one stock solution with the highest concentration and
other solutions with lower concentrations will be prepared by a dilution of this stock solution.
The stock solution of the original sample of PVOH is prepared by dissolution of ca. 4.2 g of the
polymer (the mass is determined using an analytical balance with accuracy 0.1 mg) in ca. 200 ml
of hot distilled water using gentle stiring to minimize amount of foam. After dissolving and cooling
down the solution is poured quantitatively into 250 ml volumetric flask (it is necessary to wash
out the rest of the solution from the original bottle into the volumetric flask by pure distilled
water). Then pure distilled water is added up to the mark of the volumetric flask and the solution
is slowly mixed. The stock solution of the degradated PVOH is prepared in a similar way. Again ca.
4.2 g of the original polymer (the mass is determined using an analytical balance with accuracy
0.1 mg) is dissolved in ca. 200 ml of hot distilled water using gentle stiring to minimize amount of
foam. After dissolving and cooling down, 1.25 g of solid KIO4 is added into the solution and it is
warmed up to 70°C and stirred until the periodate is dissolved. Then the solution is cooled down
and is poured quantitatively into 250 ml volumetric flask (again, it is necessary to wash out the
rest of the solution from the original bottle into the volumetric flask by pure distilled water). Then
pure distilled water is added up to the mark of the volumetric flask and the solution is slowly
mixed. The prepared stock solutions of the original and degradated PVOH, resp. have mass
concentration around 16 g/l (the accurate value of concentration is calculated).
2. Preparation of diluted solutions. 50 and 25 ml of the stock solutions are pipetted into 100 ml
volumetric flasks and refilled by pure distiled water up to the mark. Particular mass
concentrations 2w (unit g/l) are to be calculated.
3. Measurement of viscosity. Measurement of dynamic viscosity is performed on Stabinger SVM
3000 at temperature 25°C according to following procedure
8
a) First, the button MODE on the front panel of the apparatus (see fig. 2) is pressed and the
procedure M6-REPEAT S is chosen by arrows buttons (see fig. 2 SET/MODE). The choice is
confirmed by ENTER. The chosen procedure is based on a performance of three subsequent
measurements, when an operator is asked automatically to refill the cells by new amount of
the sample. The final result is an arithmetic average of these three values. The apparatus
automatically checks, if a value of standard deviation is within a defined range. If the value of
standard deviation is out of the range, the operator is asked to repeat measurement. This
procedure (MODE) minimizes problems with insifficiently washed cells.
Fig.2: Keyboard on the front panel of the SVM3000. MENU – main menu; PRINT-
saving or printing of results; SET- set of temperature value and sample info; MODE- choice of the apparatus mode – it determines experimental procedure; START – button ti start a measurement (repeated, too), or button to switch on/off a motor durinf filling of the cells; PUMP – air pump for drying of the cell to be switched on/off; INFO- opens further screans with further information
b) The menu for writing of sample name and defining temperature of measurement is brought
about by pressing of the button SET (fig.2). The external keybord or the keybord from the
front panel (see fig. 2) is used for listing of menu items. The chosen item is confirmed by key
ENTER.
c) Make sure that the motor of the outer rotor is switched off (if it is not true, it is necessary to
switch the rotor off by pressing of the button START (see fig. 2). The liquid sample is slowly
sucked into 5 ml syrringe without bubbles (if there are bubbles in the syrringe, they should
be removed by turning the surringe over and pressing the piston). After that the syrringe is
put into the port D1 (see fig.3) and at least 2 ml of the liquid sample is inserted into the
apparatus by gentle pressing5 of the syrringe piston. Then the motor of the rotor is switched
5 The power for pushing should not excess ca. 20 N. It corresponds to weight of the 2 kg body.
9
on for ca. 10 seconds by pressing of the button START on the front panel (see fig. 2) to
ensure optimal filling of the cells. After ca. 10 seconds the motor is switched off by pressing
START button and further 1 ml of the sample is inserted into the apparatus (liquid should
appeared in the waste tube – see fig. 3). Then measurement can be started.
Fig.3: Scheme of the inject part of the Stabinger SVM3000.
d) The real measurement is started by further pressing of the START button and it is waited
until the apparatus obtains the first value of viscosity, switches off the motor and asks the
operator to push further 1 ml of liquid. After pushing of the sample the START button is
pressed to start the second measurement and it is. Then the third measurement is
performed in the same way as previous measurements. If there is no problem, the apparatus
calculates the final value as an arithmetic average from the three values and the operator
saves the result by the button PRINT (fig.2). If the standard deviation is out of the defined
range, measurement has to be repeated after pushing further amount of the sample into the
apparatus.
4. Data treatment. The values of limiting viscosity numbers both for original and degradated
polymer are obtained by extrapolation of the 2i w vs. 2w dependencies. After that the viscosity
average molar masses aM M
are obtained
from
values according to the equation (1.11).
Finally the value of the relative frequency of head-to-head CHX CHX bonds in a PVOH
sample is calculated from the equation (1.22) and the relation this value to the rate constants
values k a k in the equation (1.3) is discussed.
10
Laboratory excersise 2: Finding of numerical values of parameters in equation describing correlation between molar mass and limiting viscosity number (intrisnic viscosity) in the case of polyvinyl alcohol Introduction
While the basic chemical structure of a synthetic polymer is usually well understood, many physical
properties depend on such characteristics as chain length, degree of chain branching, and molar
mass, which are not easy to specify exactly in terms of a molecular formula. Moreover the
macromolecules in a given sample are seldom uniform in chain length or molar mass (which for a
linear polymer is proportional to chain length); thus the nature of the distribution of molar masses is
another important characteristic. A polymer whose molecules are all of the same molar mass is said
to be monodisperse; a polymer in which the molar masses vary from molecule to molecule is said to
be polydisperse. Specimens that are approximately monodisperse can be prepared in some cases by
fractionating a polydisperse polymer; this fractionation is frequently done on the basis of solubility in
various solvent mixtures.
This experiment is concerned with the linear polymer polyvinyl alcohol 2CH CHOHn
(PVOH), which is prepared by methanolysis of the polyvinyl acetate (PVAc) obtained from the direct
polymerization of the monomer vinyl acetate 3 2CH COO CH=CH . As ordinarily prepared,
polyvinyl alcohol shows a negligible amount of branching of the chains. It is somewhat unusual
among synthetic polymers in that it is soluble in water. This makes polyvinyl alcohol commercially
important as a thickener, as a component in gums, and as a foaming agent in detergents.
Experimental determination of molar mass of polymers
For the determination of very high molar masses, freezing-point depressions, boilingpoint elevations,
and vapor-pressure lowerings are too small for accurate measurement. Osmotic pressures are of a
convenient order of magnitude, but measurements are timeconsuming. The technique to be used in
this experiment depends on the determination of the intrinsic viscosity of the polymer. However,
molar-mass determinations from osmotic pressures are valuable in calibrating the viscosity method.
Two other methods, light scattering and gel permeation chromatography (GPC), could also be used
as alternatives or in addition to the viscosity technique. Light scattering is a classical method for
determining molar masses, and if scattering as a function of angle is measured, it is also possible to
learn something of the polymer shape, i.e., to distinguish rodlike from spherical forms. In the case of
gel permeation chromatography, molecules of different sizes are carried by a liquid solvent over a
substrate gel network of uniformly sized pores, which can be chosen in a range from 10 to 100,000
nm. Smaller solute molecules diffuse into and out of the pores and hence exhibit a longer retention
time than larger solute molecules, which are not retained and travel through the column with the
solvent. This method is simple, but to obtain accurate molar masses of a given polymer type from the
retention times it is necessary to reproduce the flow rates carefully and to calibrate the column with
polymers of known molar mass and similar chemical composition.
As it was mentioned above, molar mass will be determined by viscosity measurement in this work.
Viscosity will be mesured by the Stabinger SVM 3000 instrument from Anton Paar company, which is
11
a special rotation viscometer. The rotation viscometer is, generally, based on the fact that there is a
function dependence of torque, which develops at constant rotation of a body in liquid, on viscosity
of this liquid. Stabinger SVM3000 differs from standard rotational viscometers by a patented feature:
hollow cylinder freely floats in liquid and is centred in the cell during rotation of outer jacket by
centrifugal forces. This arrangment eliminates unwanted influence of mechanical friction between
mechanical parts of normal rotational viscometers, where rotating body is connected directly to the
motor.
Theory of evaluation of molar mass from viscosity measurement
Einstein showed that viscosity of a fluid in which small rigid spheres are present in dilute and
uniform suspension is related to the viscosity0 of the pure fluid (solvent) by the expression
0
51
2
v
V
(2.1)
where v is the volume occupied by all the spheres and V is the total volume. The left side of the
equation (2.1) is denoted as a increment of relative viscosity i (specific viscosity sp ). For
nonspherical particles the numerical coefficient on the right side of equation (2.1) is higher than 5 2 ,
but should be still a constant for any given shape provided the rates of shear are sufficiently low to
avoid preferential orientation of the particles.
Limiting viscosity number (formerly intrinsic viscosity) is defined as the ratio of the specific
viscosity to the weight concentration of solute, in the limit of zero concentration
2 20 0
2 2 0
1lim lim lni
w ww w
(2.2)
where 2w is conventionally defined as the concentration in grams of solute per 100 mL of
solution/suspension. Both 2i w and 2 01 lnw show a reasonably linear concentration dependence at
low concentrations 2w . The dependence 2 01 lnw vs.
2w usually has small negative slope, while
2i w vs. 2w plot has larger positive slope. Therefore both plots can be used for the extrapolation to
obtain , although 2 01 lnw for a single dilute solution will give a fair approximation to .
If the internal density of the spherical particles (polymer molecules) is independent of their size (i.e.,
the volume of the molecule is proportional to its molar mass), the intrinsic viscosity should be
independent of the size of the particles and hence of no value in indicating the molar mass. This,
however, is not the case; to see why we must look into the nature of a polymer macromolecule as it
exists in solution.
Gauss (statistical) polymer coil
A polymer such as PVOH contains many single bonds, around which rotation is possible. If the
configurations around successive carbon atoms are independent and unrelated, it will be seen that
two parts of the polymer chain more than a few carbon atoms apart are essentially uncorrelated in
regard to direction in space. The molecule is then "statistically coiled" and resembles a loose tangle
of yarn (see Fig.1):
12
Figure 1. – a scheme of Gauss (statistical) polymer coil
Simple statistical treatments2 show that the mean distance between the two ends of the chain, and
indeed also the effective mean diameter d of the coiled molecule regarded as a rough sphere, should
be proportional to the square root of the chain length and thus to the square root of the molar mass:
1 2d M (2.3)
The volume mv occupied by a molecule should then vary as 3 2M . The number of molecules in a
given weight of polymer varies inversely with the molar mass M; hence the total volumeof the spheres is
3 2 1 2cVv M cVM
M (2.4)
Two equations (2.2) and (2.4) give
1 2KM (2.5)
where K is a constant. This treatment is much simplified; it ignores among other things the problem
of "excluded volume"; that is, the fact that the chain cannot coil altogether randomly because it is
subject to the restriction that no two parts of the chain may be at the same point in space at the
same time. This restriction becomes more and more important the higher the molar mass. Even
more serious is the effect of solvent; the above treatment tacitly assumes a "poor solvent," which
would barely get the polymer into solution. A "good solvent," by solvating the polymer, makes the
size of the statistical coil increase faster with chain length than it otherwise would (equation (2.4)),
owing to enhancement of the excluded volume effect. Accordingly, instead of equation (2.5) we
might write
aKM (2.6)
where K and a are empirical parameters characteristic both of the polymer itself and of the
solvent. The exponent a varies from about 0.5, for well-coiled polymer molecules in a poor solvent,
to as much as 1.7 for a rigidly extended "rodlike" polymer molecule.
The equaton (2.6) is valid for polydispersed polymers, too. In this case, viscosity-average molar mass
M , whis is defined below, is placed in the equation (2.6).
13
Number-Average Molar MassNM and Viscosity-Average Molar Mass M
For a polydisperse polymer, any determination of molar mass must yield an average of some sort.
When a colligative property such as osmotic pressure is used, the average is a number average NM 6
0
0
N
MP M dM
M
P M dM
(2.7)
where P M is the molar-mass distribution function; that is, P M dM is proportional to the
number of molecules with molar masses M and M dM . Viscosity-Average Molar Mass M
obtained from the value of limiting viscosity number is given by the equation7
1
0
0
aM P M dM
M
MP M dM
(2.8)
and its value is different from the NM value in the case of polydisperse polymer (certainly except for
the case of monodisperse polymer, where values of molar mass determined by osmotic pressure and
viscosity measurement, respectivally, are the same ). The ratio NM M is a constant given by a
form of the distribution function P M and by value of the parameter a . A commonly encountered
distribution function, and one that is likely to be valid for PVOH, is one that arises if the probability of
a chain-termination reaction during the polymerization is constant with time and independent of the
chain length already achieved. It is also the most likely function for the product resulting from
cleavage with periodate if the head-to-head structures CHX CHX can be assumed to be
randomly distributed along the PVOH chains. This distribution function is
1
NM M
N
P M eM
(2.9)
6 If there is a countable number of monodisperse fractions, then N i i
i
M x M , where ix je molar fraction
of i-th fraction and iM is its molar mass
7 If there is a countable number of monodisperse fractions, then
1 a
a
i i
i
M W M
, where iW is mass
fraction of i-th fraction and iM is its molar mass
14
If this distribution function is used in the equations (2.7) and (2.8), the folowing relation can be
derived:
1
1 1a
NM M a a (2.10)
where is the gamma function (Euler integrál of the second kind).
For the sake of completeness it can be added that so called mass-average molar mass WM can be
defined WM 8
2
0
0
W
M P M dM
M
MP M dM
(2.11)
Series of different molar mass values for one sample is as follows:
pro 0,5 1
pro 1
N W
N W
M M M a
M M M a
(2.12)
Experimental setup
1. Preparation of the stock solutions. The preparation of PVOH solutions is time-consuming,
because it is necessary to dissolve PVOH in hot water and to stir gently to avoid a formation of
foam. Therefore it is better to prepare one stock solution with the highest concentration and
other solutions with lower concentrations will be prepared by a dilution of this stock solution.
The stock solution of the each sample of PVOH with known molar mass is prepared by dissolution
of ca. 4.2 g of the polymer (the mass is determined using an analytical balance with accuracy
0.1 mg) in ca. 200 ml of hot distilled water using gentle stiring to minimize amount of foam. After
dissolving and cooling down the solution is poured quantitatively into 250 ml volumetric flask (it is
necessary to wash out the rest of the solution from the original bottle into the volumetric flask by
pure distilled water). Then pure distilled water is added up to the mark of the volumetric flask and
the solution is slowly mixed. The prepared stock solutions of the PVOH have mass concentration
around 16 g/l (the accurate value of concentration is calculated).
2. Preparation of diluted solutions. 50 and 25 ml of the stock solutions are pipetted into 100 ml
volumetric flasks and refilled by pure distiled water up to the mark. Particular mass
concentrations 2w (unit g/l) are to be calculated.
8 If there is a countable number of monodisperse fractions, then W i i
i
M W M , where iW is mass fraction
of i-th fraction and iM is its molar mass
15
3. Measurement of viscosity. Measurement of dynamic viscosity is performed on Stabinger SVM
3000 at temperature 25°C according to following procedure
a) First, the button MODE on the front panel of the apparatus (see fig. 2) is pressed and the
procedure M6-REPEAT S is chosen by arrows buttons (see fig. 2 SET/MODE). The choice is
confirmed by ENTER. The chosen procedure is based on a performance of three subsequent
measurements, when an operator is asked automatically to refill the cells by new amount of
the sample. The final result is an arithmetic average of these three values. The apparatus
automatically checks, if a value of standard deviation is within a defined range. If the value of
standard deviation is out of the range, the operator is asked to repeat measurement. This
procedure (MODE) minimizes problems with insifficiently washed cells.
Fig.2: Keyboard on the front panel of the SVM3000. MENU – main menu; PRINT-
saving or printing of results; SET- set of temperature value and sample info; MODE- choice of the apparatus mode – it determines experimental procedure; START – button ti start a measurement (repeated, too), or button to switch on/off a motor durinf filling of the cells; PUMP – air pump for drying of the cell to be switched on/off; INFO- opens further screans with further information
b) The menu for writing of sample name and defining temperature of measurement is brought
about by pressing of the button SET (fig.2). The external keybord or the keybord from the
front panel (see fig. 2) is used for listing of menu items. The chosen item is confirmed by key
ENTER.
c) Make sure that the motor of the outer rotor is switched off (if it is not true, it is necessary to
switch the rotor off by pressing of the button START (see fig. 2). The liquid sample is slowly
sucked into 5 ml syrringe without bubbles (if there are bubbles in the syrringe, they should
be removed by turning the surringe over and pressing the piston). After that the syrringe is
put into the port D1 (see fig.3) and at least 2 ml of the liquid sample is inserted into the
16
apparatus by gentle pressing9 of the syrringe piston. Then the motor of the rotor is switched
on for ca. 10 seconds by pressing of the button START on the front panel (see fig. 2) to
ensure optimal filling of the cells. After ca. 10 seconds the motor is switched off by pressing
START button and further 1 ml of the sample is inserted into the apparatus (liquid should
appeared in the waste tube – see fig. 3). Then measurement can be started.
Fig.3: Scheme of the inject part of the Stabinger SVM3000.
d) The real measurement is started by further pressing of the START button and it is waited
until the apparatus obtains the first value of viscosity, switches off the motor and asks the
operator to push further 1 ml of liquid. After pushing of the sample the START button is
pressed to start the second measurement and it is. Then the third measurement is
performed in the same way as previous measurements. If there is no problem, the apparatus
calculates the final value as an arithmetic average from the three values and the operator
saves the result by the button PRINT (fig.2). If the standard deviation is out of the defined
range, measurement has to be repeated after pushing further amount of the sample into the
apparatus.
4. Data treatment. The values of limiting viscosity numbers are obtained by extrapolation of the
2i w vs. 2w dependencies for particular samples with known molar mass. The obtained
experimental data vs. molar mass is obtained for determination of parameter values of the
equation (2.6). E.g.the equation (2.6) can be rearranged to obtain linear dependence ( ln vs.
ln M ) and then the least squares method can be applied. Finaly the obtained results will be
9 The power for pushing should not excess ca. 20 N. It corresponds to weight of the 2 kg body.
17
compared to literature values of K and a parameters (a=0,76; K = 2,010-4), especially with regard
to prospective polydispersity of samples and the way, in which known values of molar masses
were obtained.
Literature
Shoemaker, Garland, Nibler, "Experiments in Physical Chemistry," McGraw-Hill Publishing Company, Toronto (1989).