Design, Performance, and Applications of a Hybrid X-Ray/MR System for Interventional Guidance
-
Upload
independent -
Category
Documents
-
view
0 -
download
0
Transcript of Design, Performance, and Applications of a Hybrid X-Ray/MR System for Interventional Guidance
INV ITEDP A P E R
Design, Performance, andApplications of a HybridX-Ray/MR System forInterventional GuidanceDigital flat-panel detectors make it possible to view combined x-ray and magnetic
resonance images to more-accurately guide medical diagnosis and treatment.
By Rebecca Fahrig, Arundhuti Ganguly, Prasheel Lillaney, John Bracken,
John A. Rowlands, Zhifei Wen, Huanzhou Yu, Viola Rieke, Juan M. Santos,
Kim Butts Pauly, Daniel Y. Sze, Joan K. Frisoli, Bruce L. Daniel, and Norbert J. Pelc
ABSTRACT | Image-guided minimally invasive procedures
have made a substantial impact in improving patient manage-
ment, reducing the cost, morbidity, and mortality of treatments
and making therapies available to patients who would other-
wise have no option. X-ray fluoroscopy and magnetic reso-
nance imaging (MRI) are two powerful tools for guiding
interventional procedures but with very different strengths
and weaknesses. X-ray fluoroscopy offers very high spatial and
temporal resolution and is excellent for guiding and deploying
devices. MRI offers tomographic imaging with complete
freedom of plane orientation, outstanding soft tissue discrim-
ination, and the ability to portray physiological responses
during treatment. We have shown that it is feasible to fully
integrate an X-ray fluoroscopy system into the bore of an
interventional MR scanner to provide a single congruent field
of view, with integration requiring minor modifications to the
flat-panel digital detector, and using a static-anode X-ray tube.
Given the limited availability of the MR scanner platform (0.5T
GE Signa SP magnet), and the X-ray fluence limitations of the
static-anode X-ray tube, we are now investigating the technol-
ogy developments required to place a rotating-anode digital
flat-panel X-ray system immediately adjacent to a closed-bore
MRI system. These types of hybrid systems could have
enormous impact in the diagnosis and treatment of oncologic,
cardiovascular, and other disorders.
KEYWORDS | Hybrid imaging; interventional image guidance;
magnetic resonance imaging; X-ray fluoroscopy
I . INTRODUCTION
Combining X-ray fluoroscopy and magnetic resonance
imaging (MRI) systems for guidance of interventional
procedures has recently becomemore available, with several
vendors offering dual-modality suites [1]–[8]. In most cases,
these suites require a long travel distance between the two
imaging modalities, on the order of several tens of feet. Thisdesign strategy was driven by the use of relatively standard
X-ray components (e.g., image intensifiers and rotating
anode X-ray tubes) and by the need to ensure minimal
impact of one system on the other. As a result, switching
back and forth between the two modalities is time
consuming and awkward, leading to potential risk and loss
of image registration. It would also invariably lead to less
Manuscript received July 17, 2007; revised xxxx. This work was supported in part by the
National Institutes of Health under Grants NIBIB R01 EB000198 and RR09784, GE
Healthcare, NSERC, and the Lucas Foundation.
R. Fahrig, A. Ganguly, V. Rieke, K. B. Pauly, D. Y. Sze, J. K. Frisoli, B. L. Daniel, andN. J. Pelc are with the Department of Radiology, School of Medicine, Stanford
University, Stanford, CA 94301 USA (e-mail: [email protected]).
P. Lillaney is with the Departments of Radiology and Bioengineering, Stanford
University, Stanford, CA 94301 USA.
J. Bracken and J. A. Rowlands are with the Department of Medical Biophysics,
University of Toronto and Sunnybrook Health Sciences Center, Toronto,
ON M4N 3M5, Canada.
Z. Wen was with the Departments of Radiology and Physics, Stanford University,
Stanford, CA 94301 USA. He is now with the Department of Medical Physics,
University of Wisconsin-Madison, Madison, WI 53706 USA.
H. Yu was with the Departments of Radiology and Electrical Engineering, Stanford
University, Stanford, CA 94301 USA. He is now with GE Healthcare, Menlo Park,
CA 94086 USA.
J. M. Santos is with the Department of Electrical Engineering, Stanford University,
Stanford, CA 94301 USA.
Digital Object Identifier: 10.1109/JPROC.2007.913506
468 Proceedings of the IEEE | Vol. 96, No. 3, March 2008 0018-9219/$25.00 �2008 IEEE
than optimal use of the modalities since one may bereluctant to move the patient repeatedly. In a recent paper
by Dick [9] it was emphasized that during interventional
procedures in such suites Bmultiple external devices and
connections required careful supervision and several minutesof preparation when moving in either direction.[
The advent of digital flat-panel detectors that are more
resistant than X-ray image intensifiers to stray magnetic
fields opened the door to alternative hybrid imaging systemsthat place both modalities in close proximity. This allows
faster, easier switching between modalities, maintaining
multimodality registration and providing a truly hybrid
imaging system. This technology enabled the development
of our static-anode X-ray fluoroscopy system integrated
into the bore of a vertical gap interventional MRI scanner
(Signa SP, GE Milwaukee, WI)Vthe SP-XMR. Here we
describe the technical challenges that were overcome toallow close integration of the two imaging systems
including the following critical issues: 1) X-ray tube and
detector operation within the magnetic field, 2) MR field
homogeneity and noise properties with the X-ray system in
place, and 3) registration and integration approaches.
Our experience with the SP-XMR system is guiding
development of our next generation of truly hybrid XMR
systems. Since the vertical gap MRI system is no longercommercially available and was limited in field to 0.5 T,
attention is turning to short closed-bore systems, such as
the Espree (Siemens Medical Solutions, Erlangen,
Germany; 1.2 m long, 75 cm inner bore diameter). This
and/or similar systems from other vendors may become
the most prevalent interventional MRI platform, and we
therefore propose to widen the availability of truly hybrid
XMR systems by developing the technology for placementof an X-ray system directly adjacent to a high-field closed-
bore MRI system while maintaining the shortest possible
distance between the two systems.
II . THE SP-XMR SYSTEM
A. System ConfigurationThe current implementation of the SP-XMR consists of
1) the GE BInova[ fluoroscopy capable flat-panel detector
with a 20� 20 cm2 area and 0.2� 0.2 mm2 pixels slightly
modified for operation at high magnetic fields, placed at a
field of 0.5 T; 2) the OEC 9800 workstation, which
controls X-ray exposure, image acquisition, processing and
display, data communication, and archiving; and 3) the
fixed-anode X-ray generator and mainframe from an OEC8800 mobile C-arm system modified to allow separation of
the high-voltage generator from the tube (distance 10 m)
and with the X-ray tube mounted in a small housing. Both
the X-ray tube and housing were modified to remove all
magnetic components. The X-ray tube is in a field of
strength�0.2 T when placed in the overhead bracket, with
the external magnetic field aligned with the internally
applied electric field between the anode and cathode. Thehigh-voltage cables are passed through waveguides in the
penetration panel, with radio-frequency (RF) shielding
between the penetration panel and generator to eliminate
RF leakage into the room. The X-ray tube is installed in the
bracket between the two donut-shaped superconducting
magnets, allowing a true anterior-posterior (AP) view of a
patient supported axially in the magnet. A homemade
collimator with remote controls has been built and installed,allowing intraprocedural modification of the X-ray field of
view (FOV). Construction of an integrated cradle-handling
platform for side-dock procedures allows positioning flexi-
bility of the detector. The overall system is shown in Fig. 1.
We have found that automatic exposure control, online
image subtraction [e.g., for digital subtraction angiography
(DSA) studies], continuous image storage, and image trans-
mission via Digital Imaging and Communications inMedicine (DICOM) format to other systems (e.g., the MR
real-time imaging workstation) are critical during clinical
interventions. Amore detailed description of systemupgrades
and modifications has been provided previously [10].
B. Performance of the X-Ray System in theMagnetic Field
Conventional fluoroscopic X-ray systems present, inprinciple, several problems that prevent operation in a
magnetic field. X-ray image intensifiers (XRIIs) and X-ray
tubes both contain electrons moving in vacuum through
relatively large distances (20–60 cm and 1–2 cm, respec-
tively). The deflection of these moving electrons by a
magnetic field can be considerable, making both devices
difficult to operate in the stray field of a powerful
electromagnet. Fortunately, the invention and developmentof the solid state flat-panel fluoroscopic imager to replace the
XRII has reduced the problem of using a detector in a
magnetic field to manageable proportions. There is,
unfortunately, no such simple replacement possible for the
X-ray tube, for which a vacuum approach is essential. For
this case, a more incremental approach is necessary and is
discussed below. In addition, we need to understand the
operation of the induction motor used, ingeniously to permitthe rotation of the anode inside the vacuum of the tube,
without the need for a rotating vacuum seal. Finding a
replacement for this mechanism is challenging, and so
means for extending the utility of a conventional tube in a
magnetic field are carefully considered.
DetectorVInvestigation of Modulation Transfer Function(MTF) and Detective Quantum Efficiency (DQE): Flat-paneldetectors can be either direct conversion (X-ray transduced
directly into a charge image by a photoconductor such as
amorphous selenium and stored on individual pixel
capacitors) or indirect conversion (X-ray transduced to light
in a phosphor, often cesium iodide, and this light is absorbed
and converted into a charge image by individual photodiodes
at each pixel). In both conversion methods, a very large
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
Vol. 96, No. 3, March 2008 | Proceedings of the IEEE 469
integrated circuit built on a monolithic glass substrate
coordinates the acquisition, temporary storage, and readoutof the latent charge image. This large integrated circuit is
called an active matrix and uses control (gate) lines and data
(readout) lines formed lithographically to sequence out the
latent image charge stored at the pixels (see Fig. 2). The core
technology enabling such large integrated circuits is
hydrogenated amorphous silicon (a-Si:H), which is the
same technology used in liquid crystal displays for laptop and
desktop monitors [11].There is no need for any ferromagnetic components on
the active matrix array itself, and in practice, the
manufacturing conditions exclude them. This is fortunate,
as it would be very difficult and expensive to change these
processes. In contrast, however, the auxiliary electronics
needed to drive the active matrix (gate drivers, data line
amplifiers and multiplexers, power supplies, transformers,
and timing circuits) built on a printed circuit board as wellas the physical housing are generally made without any
consideration of their magnetic compatibility, and there-
fore it is to be expected that multiple magnetically
sensitive and ferromagnetic components will be present.
However, since these are built individually from compo-
nents and not mass-produced in a clean room as is the
active matrix, there is ample opportunity to modify and
eliminate troublesome parts.The only active components on an active matrix array are
thin-film transistors [(TFTs), metal–oxide–semiconductor
field-effect transistors], which are used as switches to
control the flow of charge from the pixels to the data lines.When the TFT is off, there is no current passing through the
a-Si:H channel and therefore no opportunity for the
magnetic field to affect it in any way; when it is on, a very
Fig. 2. Schematic diagram of themain components of an activematrix
array. The only active components are the thin-film transistors,
which have currents that are too small to be affected by external
magnetic fields.
Fig. 1. Installation and components of the SP-XMR hybrid system. (a) X-ray system installed in the bore of the SP magnet with patient shown
in ‘‘side dock’’ geometry. (b) High-voltage generator and mainframe in the equipment room of the magnet. (c) Digital flat-panel detector.
(d)Static-anodeX-ray tube insertwithmagnetic components removed. (e)Closeupofcollimatorcontrols,hiddenbehindX-ray imagedisplay in (a).
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
470 Proceedings of the IEEE | Vol. 96, No. 3, March 2008
tiny current flows in the channel of the TFT. However, thedrift velocity of the carriers is so low that the deflection of
the electrons is negligible; thus the TFT is also immune to
magnetic fields when on. In addition, the very small size of
the current also ensures that there is no magnetic field to
speak of generated by the TFT. Flat-panel detectors also
contain photodiodes: X-ray photodiodes in direct conversion
systems and optical photodiodes in indirect conversion
systems. Here again the intrinsically low drift velocities ofcarriers and the tiny quantities of generated charge cannot
either be deflected significantly by a magnetic field or
generate any external field themselves.
Thus in conclusion, flat-panel detectors (both direct and
indirect conversion) contain no essential intrinsic compo-
nents that are magnetic or cause magnetic disturbances, but
do typically contain extrinsic components or housings that
must be modified or replaced. We found that simplemodifications to the onboard power conditioning circuit
(replacement of inductive filters and extension of cables so
that power supplies experience fields below 0.015 T) of the
GE Inova flat panel was sufficient to ensure robust operation
of the detector when the panel was placed in fields as high as
0.5 T. To verify this, we carried out two experiments to
compare the resolution and noise properties of the detector
at fields of 0 and 0.5 T using standard techniques to measureMTF using a slanted edge and DQE [12]. We found that
inherent detector resolution was not affected by the
magnetic field, although overall system resolution measured
at the location of the patient showed a decrease of�10% for
spatial frequencies of 0.5 cycles/mm and above due to a
change in the focal spot size (see below). The noise in the
X-ray imaging system was identical inside and out of the
magnetic field.
X-Ray TubeVInvestigation of Focal Spot Size andDistribution With E and B Parallel: In our SP-XMR system,
the strength of the magnetic field at the location of the
tube is �0.2 T. Interaction between this strong magnetic
field and the electron beam can lead to deflection and/or
defocusing of the beam. The relatively simple design of the
system allowed us to minimize the deflection of theaccelerating electrons by aligning the anode-cathode
direction with the main magnetic field that, by symmetry,
is perpendicular to the plane midway between the two
donuts (see Fig. 1). Each component of the X-ray tube was
investigated in detail, including the effects of the ac
filament current causing mechanical oscillation of the
filament (shown to be negligible) [10] and the impact of
field strength on focal spot distribution, described here.A finite-element (FE) program (OPERA-3d; Vector
Fields, U.K.) was used to simulate the electron trajectories;
care was taken to match the modeled geometry to our fixed-
anode X-ray tube [13]. Thermionic emission was modeled
using Child’s law current limit model (temperature: 2500 K;
work function: 4.5 eV; total tube current: 4 mA). The
electrostatic Poisson’s equation was solved numerically to
obtain the electrostatic field, including the effects caused byspace charge in the electron beam. The space-charge density
was found by calculating the trajectories of electrons emitted
from the filament under the influence of the electrostatic
field and an external uniform magnetic field. Focal spot
Bimages[ were obtained by plotting and projecting the
electron current density map on the target through a
modeled pinhole to an image plane.
Experimental verification of themodel was carried out byobtaining pinhole images of the focal spot (30 �m tantalum
pinhole 12.4 cm from the X-ray tube, with an overall
magnification of 8.9) with tube and detector mounted on an
MR-compatible stand. A Tesla meter (F.W. Bell 5080; Sypris
Test & Measurement, Orlando, FL) was used to define the
point midway between the two donuts where the magnetic
field strength was 0.208 TVa field strength equivalent to
that at the focal spot location of the installed system. Imageswere acquired at 65 kVp and 4 mA at 0.208 T, and for
comparison in the anteroom of the MR suite at a residual
field of less than 0.001 T using the same stand and geometry.
Results for fields of 0 and 0.208 T are shown in Fig. 3.
The increase in focal spot size shown here accounts for the
change in system resolution (MTF) when the imaging
system operates within the magnetic field. These results
show that even low field strengths affect the size and currentdensity distribution of the focal spot of the X-ray tube.
C. Performance of the MR System in the Presence ofthe X-Ray System
Three main concerns arise when placing X-ray
components close to theMR FOV: 1) magnetic components
Fig. 3.Magnetic field dependent focal spot size at two field strengths:
top ¼ 0 mT, bottom ¼ 208 mT showing FE simulations of the electron
trajectories, the resulting simulated focal spots, and measured focal
spots obtained by placing the nonmagnetic X-ray tube at the fields
strengths indicated.
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
Vol. 96, No. 3, March 2008 | Proceedings of the IEEE 471
could cause B0 field inhomogeneity leading to imagedistortion and signal loss; 2) eddy currents could cause
artifacts; and 3) RF noise could cause both loss of signal-to-
noise ratio (SNR) and zipper artifact. As a first step, we
reduced, as far as possible, magnetic components in both
tube and detector. In the X-ray tube, the nickel focusing
cup and all screws, etc., were replaced with nonmagnetic
stainless steel, and the tube housing was constructed out of
brass. For the detector, the housing was redesigned inaluminumwith a carbon-fiber front face, all housing screws
were replaced with nonmagnetic stainless steel, and nickel-
plated connectors were replaced with plastic equivalents.
The packaging of some small components in the detector
was ferromagnetic, and nonmagnetic alternatives were not
available. Even though these components were very small,
they still caused some distortion in B. Below we describe a
solution for this problem as well as our investigations ofeddy currents and RF noise.
Magnetic Field Homogeneity: To obtain a map of the
main magnetic field within the imaging volume of the
scanner, complex images of a large doped-water phantom
were collected at two echo times using a two-dimensional
gradient echo (GRE) sequence (34 � 34 cm2 FOV, 66
3 mm slices, (�TE of 2 or 3 ms). From the unwrappedphase difference map of two echo time images, the raw
field map was calculated. The zeroth-order offset and the
first-order linear component were removed since they can
be handled by the scanner center frequency and linear
shims. We assume that the ferromagnetic components
(FCs) inside the detector are located at a few discrete
locations and are small diameter (� a few centimeters)
much smaller than the isocenter-to-detector distance (�30cm) (see Fig. 1). Each FC may therefore be approximated
by a magnetic dipole pointing in the direction of the
external field and located near the FC. This suggests that
appropriate placement of permanent magnets may be used
to eliminate the effect of the magnetic components. For
this to be practical, the intrinsic coercivity of the
permanent magnet must be greater than the external
magnetic field, which ensures that the permanent magnetwill not be demagnetized by the external field. Therefore
the magnetic field disruption caused by the detector can be
canceled by the permanent magnets oriented in a direction
opposite to B.
This passive shimming approach was investigated
using NdFeB magnets (Dexter Magnetic Technologies,
Elk Grove Village, IL) of various sizes and strengths, with
optimum placement determined using a fitting procedure.The field degradation from the detector and the
effectiveness of the passive shimming were quantified
by measuring the peak-to-peak field inhomogeneity (p-p)
and the standard deviation of the field ð�fÞ in a
cylindrical volume ð�28 � 20 cm2Þ. It was also evaluatedqualitatively with a steady-state free precession (SSFP)
sequence (TR/TE¼ 11:64=5:56 ms). The SSFP sequence
was selected for these tests since it is important forinterventional guidance and is very sensitive to magnetic
field inhomogeneity.
The p-p and �f were 5.4 and 0.55 �T, respectively,without the detector present. The inhomogeneity param-
eters were much larger with the detector in place, 12.6
and 1.68 �T, respectively. Nonlinear least squares
fitting, optimizing both strengths and locations of the
dipoles, predicted that two ideal dipoles would reducethe field inhomogeneity with the detector to 5.5 (p-p)
and 0.52 �T ð�fÞ. These two dipoles were physically
realized using two sets of permanent magnets, each
set consisting of a number of identical NdFeB magnets
(0.2 � 0.2 � 0.5 in3Þ matching the ideal dipole
strength. The two sets of magnets were placed just
below the detector in custom-made holders at locations
prescribed by the fitting (relative to the center of thedetector, with the power and other cables pointing in
the �y direction, one at x ¼ �8:5, y ¼ 1:5 cm and the
second at x ¼ 9:9, y ¼ 1:8 cm). The field map of the
detector with the shimming magnets in place showed a
deviation of 5.6 (p-p) and 0.57 �T ð�fÞ, compared to 5.4
and 0.55 �T for the baseline field with no detector in
the magnet. Fig. 4 clearly shows the severe impact of the
inhomogeneity due the presence of the detector and theimprovement achieved from the passive shimming with
high-coercivity magnets.
Dynamic Effects: Wrapping the detector in multiple
layers of aluminum foil (�100 �m in total) provides
sufficient reduction of RF noise such that the X-ray
detector can be left on during MR imaging. This allows the
detector to reach temperature equilibrium, which, bystabilizing the gain, reduces the fixed pattern noise of the
X-ray images. To verify the performance of the shield, MR
images of a water phantom were taken using a GRE
sequence (TR/TE 150/minfull) and a transmit/receive
Fig. 4. SSFP images showing the effects of field homogeneity with
(a) baseline (i.e., no X-ray detector present), (b) with the detector in
the bore of the magnet, and (c) with the detector in the bore of the
magnetic but with permanent magnet shimming to reduce the
effects of main B-field inhomogeneity.
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
472 Proceedings of the IEEE | Vol. 96, No. 3, March 2008
flexible body coil. Signal-to-noise ratios for baseline,detector in and on with no shielding, and detector in
and on with Al foil shielding were 7.48, 5.97, and 7.18,
respectively. The small decrease of 4% from baseline was
considered acceptable for operation of the system.
Presence of the Al foil attenuates the X-ray beam by only
1% at 50 keV, the mean energy of our X-ray beam when
imaging at 80 kVp.
We carried out a study of eddy current effects using astandard image-based system-level test. Eddy current
measurements were made using the manufacturer’s tools
(BGrafidy[) and with the assistance of the service
engineer. The X-ray tube had no effect, at least in part
because is it outside the volume in which the gradient coil
has its main effect. When the X-ray detector was in
position below the patient cradle (and well within the
volume affected by the gradient coil), changes in the time-dependent field produced by the gradient coil could be
observed with Grafidy. The effects were of a magnitude
and temporal dependence that was felt could be compen-
sated using the preemphasis filter. We considered having
two eddy-current compensation coefficients, to be used
when the detector is inside or outside of the bore,
respectively. However, before recalibrating the eddy-
current compensation, we examined whether any effectscould be seen in images. Seeing no significant effects, we
decided this complication was not necessary at this time.
In other configurationsVfor example, with stronger
gradients or larger field-of-view detectors, the situation
could be different and dedicated eddy-current corrections
may be necessary.
D. System IntegrationFor maximum utility, the X-ray and MR units in a
hybrid system must be integrated such that switching
between modalities is fast, easy, and efficient; registration
between systems is maintained; and each system can be
used to prescribe the acquisition geometry of the other.
Patient motion between X-ray and MR imaging should also
be minimized, and thus all objects should be MR and X-ray
compatible so that they do not need to be moved in and outof the FOV as the imaging modality is switched. Our
registration approach, and construction and design of an
X-ray compatible RF coil, are described here.
System Co-Registration: The key step in coregistration is
to accurately determine the positions of the X-ray
components in MR coordinatesVthe geometric param-
eters described by the projection matrix. Sixteen fiducialmarkers, provided by the crossing points of lead-cored
fishing line visible in both MR and X-ray images, are
placed in the FOV; detection is carried out semiautomat-
ically in X-ray and MR images [14]. By fitting these MR
and X-ray measurements to the nonlinear function
describing the projection process, the values of the
geometric parameters are determined. Two possible
sources of registration error, fiducial marker measurementnoise and MR gradient nonlinearity, were studied. To
illustrate the use of the calibrated geometric parameters
and to validate their accuracy, we used X-ray images as
Bscouts[ to prescribe MR slices. The calibration process
was implemented as a GUI-based package (Fig. 5) on the
same computer platform as RTHawk, a robust real-time
interactive imaging platform for MR scanner control
[15]–[20], developed at Stanford.
X-Ray Compatible RF Coils: For the design of an X-ray
compatible RF-coil, all coil components that potentially
lie in the FOV of the X-ray image, i.e., in the path of the
X-ray beam, should have minimal X-ray attenuation. To
minimize the attenuation by the loop conductor, we used
aluminum (Al), which has a significantly lower atomic
number ðZ ¼ 13Þ than the commonly used copper (Cu,Z ¼ 29) [21]. Using aluminum foil, we built a four-
element abdominal phased array (PA). The capacitors and
copper lead wires, detuning circuits, and baluns were
placed outside the X-ray FOV, so that in normal use they
would not appear in the X-ray image.
MR image quality of the X-ray compatible PA was
compared to a single-channel receive-only coil (SPGR, TR/
TE/flip/BW¼ 40 ms=5:5 ms=50=31 kHz) and showed ap-proximately a 60% improvement in SNR. Measurements
of the quality factor of the coils resulted in an unloaded/
loaded Q ratio of �270/40 for the individual elements of
the array. For comparison, a conventional abdominal PA
coil of similar design had an unloaded/loaded Q ratio of
185/50. This showed that replacing copper conductors
with aluminum did not noticeably influence the noise
performance of the coils. An X-ray/MR image pair,acquired during a transjugular intrahepatic portosystemic
shunt (TIPS) placement using an X-ray compatible coil, is
shown in Fig. 6. This image can be compared with those
shown in Fig. 7, early X-ray images acquired without the
X-ray compatible coil.
E. Clinical Operation of the SP-XMR SystemThe X-ray tube and generator are permanently installed
in the MR suite and the equipment room, respectively. Prior
to a clinical procedure, the X-ray detector is placed on a
modified mount under the patient table, and an X-ray
compatible cradle is installed. Switching from X-ray to MR
takes only 60 s and requires the following: a) X-ray system
software is placed in a Bstandby[ state and b) X-ray detectorand display are powered off to prevent RF noise from the
display. Switching fromMR to X-ray takes the same amountof time. Note that during the first group or procedures
carried using SP-XMR guidance, the detector was lowered
by 5 ft during MR scanning, but this is no longer necessary
with the passive shimming described above.
To date, we have completed 56 patient studies under
SP-XMR guidance: 16 TIPS [22], 11 hysterosalpingo-
grams, a brain biopsy [23], a vaginal reconstruction, a cyst
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
Vol. 96, No. 3, March 2008 | Proceedings of the IEEE 473
drainage, a facial vascular malformation sclerotherapy, 11other arteriovenous malformations, a prostate seed
implantation, 3 arthrograms, and 10 cystography studies
looking for renal reflux in infants and young children. For
all studies, several switches back and forth between imagingmodalities were typically required, with X-ray typically
providing device guidance or visualization of small,
contrast-filled structures, while MR was used to visualize
critical soft-tissue structures or to provide three-dimen-sional physiological information.
One particularly illustrative example of the utility of
the XMR system was provided during the creation of a
neovagina for an 18-year-old patient with a severe
congenital anomaly (cloacal atresia) where the caudal
spine was unformed and she was born without a urinary
bladder, urethra, rectum, anus, and vagina. During a dozen
previous surgeries, a loop of small bowel was used to createa neobladder, with a urostomy for drainage, and a
colostomy was constructed to allow evacuation of the
bowel. The patient also had separate halves of a uterus,
neither of which was connected to a cervix or vagina.
Menstrual cycles had started in one half of the uterus,
Fig. 5. (a) System schematic showing the relationship among magnet control, MR console, real-time imaging module, and X-ray/MR calibration
module. The GUI interface for the calibration module is shown in (b), with an MR image of one plane of the X-ray/MR compatible lead-core
fishing line phantom and the projection matrix that defines the relationship between the X-ray system FOV and the MR system FOV.
The RTHawk real-time interface is shown in (c). The software XScout permits definition of an MR plane of interest by drawing a line on
the X-ray projection image and combining the known location of the line in the detector plane with the calibrated projection matrix
values from XCalib. The parameters of the plane of interest are then passed to RTHawk.
Fig. 6. Composite imagegenerated fromimagesacquiredduringaTIPS
procedure. Thewhite arrowhead indicates the trocar and needle being
directed towards the hepatic vein (arrow) during MR guidance of
the puncture. The X-ray image was acquired later in the procedure
during balloon expansion of the stent along the tract of the needle.
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
474 Proceedings of the IEEE | Vol. 96, No. 3, March 2008
causing accumulation of blood in the contained uterine
cavity and attached fallopian tube. The other half
contained no menstruating tissue but developed a large
retained stone. A kidney had also migrated incorrectlyduring development, resting in the bottom of the pelvis.
The challenge, therefore, was to create a connection
between the perineal dimple (vaginal remnant on the skin
surface) and the menstruating uterus in order to allow
normal menstrual flow while avoiding the neobladder,
kidney, bowel, and surrounding blood vessels. Using our
hybrid system, we performed percutaneous transperineal
puncture of the imperforate uterus under MR guidanceusing numerous oblique planes of imaging, verified the
path of the needle relative to sensitive structures by
injecting contrast under X-ray imaging, and then created a
neovagina by progressively balloon-dilating the channel
along the needle tract. Images acquired during the
procedure are shown in Fig. 7. The ability to obtain
three-dimensional MRI images, combined with the speed,
sensitivity, projectional imaging, and spatial resolution ofreal-time X-ray fluoroscopy, allowed successful tract
creation without injury to surrounding structures in thispatient with no surgical option.
III . THE FUTURE OF TRULYHYBRID XMR
The SP-XMR system, based on a vertically open magnet,
allows the X-ray and MR systems to have overlapping fields
of view with no need to move the patient or equipment
when switching between modalities. This is ideal from
many respects, including integration of the information
from the two techniques. On the other hand, the image
quality performance of both the X-ray and MR systems islimited. The MR performance is limited by the low field
strength and modest gradient performance of the Signa-SP
while the X-ray performance is limited by X-ray tube
output (static anode X-ray tube) and the inability to easily
obtain views other than the AP projection. It is possible to
increase the performance of the X-ray system within this
architecture. Nonetheless, it is doubtful that the X-ray
subsystem could match the capabilities of a high perfor-mance X-ray angiography system. At the opposite extreme,
if the X-ray and MR systems are widely separated so as to
prevent any interaction between them, one would be able
to obtain the best performance from each system at the
expense of integration of the two modalities. The wide
separation would also preclude frequent switching be-
tween modalities, which most likely would lead to limited
use of one or the other.Because of the limitations of commercially available MR
systems based on open magnets, many sites investigating
interventional MRI are turning to short, closed-bore
systems, such as the Siemens Espree (1.2 m long, 75 cm
inner bore diameter) for interventional MRI guidance.
While these systems do not match the full capabilities of the
highest performanceMR systems, their capabilities are quite
good (e.g., allowing functionalMRI, diffusion, and perfusionstudies) while the wider and shorter bore provides much
better patient access than in conventional systems (though
not as good as that in open systems). Placement of an X-ray
fluoroscopy system directly adjacent to a closed-bore MRI
system, while maintaining the shortest possible distance
between the two systems, offers an interesting compromise
between the two extreme architectures described above.
With the Siemens Espree, for example, a separation of only1.1 m is realizable between the imaging volumes of the two
systems, and the mechanical transport system of the MR
scanner can be used to maintain easy registration between
the fluoroscopy and MR systems. This patient/table travel is
on the same order as is currently seen in the cath-angio suite
when a clinician inserts a catheter-based device through
groin access and then follows the catheter into a structure in
the brain.As stated above, one important limitation of our
current X-ray/MR system is X-ray tube output. In fact,
in the minimally invasive abdominal interventions for
Fig. 7. (a)–(c) MR images acquired during passage of the needle from
the perineal dimple into the isolated left-sided uterus. (d) Guidewire
coiled in the lumen of the uterus (in place to maintain tract) and the
injection of radiopaque contrast during retraction of the needle
along the tract in order to verify that the needle did not injure the
neobladder or other structure. (e) Injection of contrast directly into
the neobladder through the stoma; a faint blush from the previous
injection [i.e., shown in (c) during withdrawal of the needle] can be
seen; there is no evidence of connection between bladder and tract.
In this early study, the white arrowheads point to artifacts
originating from the non-X-ray compatible patient cradle; black
arrowheads point to artifacts from the non-X-ray compatible RF coil.
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
Vol. 96, No. 3, March 2008 | Proceedings of the IEEE 475
which the XMR system was used, we could not treat
heavy individuals because the heat loading capacity,
and thus X-ray flux, of the static-anode X-ray tube is
limited. Meeting current X-ray image quality standards
requires a rotating-anode X-ray tube to enable veryshort, high exposure, low-noise imaging. When com-
bined with a small separation between X-ray and MR
systems, the X-ray tube experiences fringe fields of
significant magnitude with a direction that is highly
dependent on position.
Schematics of two possible implementations, for
guidance of neuro-angiographic interventions and for
guidance of cardiac interventions, are shown in Fig. 8with geometry of closest approach given the somewhat
different imaging requirements. Coronary artery X-ray
imaging requires high output with very short pulses for
optimum visualization of cardiac vessels (typically 9 10 kW
pulses, with a 5 ms pulse width, at 80 kVp). The need for
oblique views of the cardiac vessels places the X-ray tube
farther from the magnet shroud since angulation is
required. The neuroimaging application is less exactingwith respect to X-ray tube output; longer X-ray pulse
widths can be tolerated since motion is small. However,
since most device navigation can be achieved using
standard anteroposterior and lateral views, the X-ray
system can be placed closer to the magnet shroud. Both
applications require high resolution, i.e., good focal spot
distribution, for visualization of small vessels, with a focal
spot size on the order of 0.8 mm � 0.8 mm2. Note that thedistance from the patient table to the X-ray tube is
determined based on government regulations that fix the
minimum source-to-skin distance at greater than 38 cm,
and therefore typical values for the focal spot to center of
rotation of clinical C-arm systems is �66 cm (i.e., for
patient total thickness on the order of 40 cm).
We have shown that a thin-film transistor-based large-
area digital X-ray detector is essentially immune to
external magnetic fields and have verified operation in
magnetic fields as high as 0.5 T. We have also shown that
the electron optics of the X-ray tube are very sensitive toexternal fields, and therefore we orient the motor of the
X-ray tube towards the shroud of the magnet so that it
experiences a somewhat higher field than is present at the
focal spot (separation between motor and electron beam is
�10 cm). An estimate of the magnetic fields that the X-ray
tube components would experience, for the geometries in
Fig. 8, is summarized in Table 1 for a 1.5 T Signa magnet
(GE Medical Systems).
Fig. 8. Two possible implementations of the closed-bore XMR system. The installation shown in (a) permits anteroposterior and lateral
views, while the geometry of (b) with the X-ray system placed slightly farther away from the shroud of the magnet allows angulation
for cardiac applications.
Table 1 Strength of the Magnetic Field at the Focal Spot Location (F.S.)
and at the Motor for the Two Geometries Illustrated in Fig. 8. Br is the Field
Strength in the Radial Direction and Bz is the Field Strength in the Axial
Direction. For the Neurointerventional Applications (Neuro), the Distance
From the Center of the MR Field of View to the X-Ray Tube Components Is
Determined Only by the Length of the Magnet and the Radial Location of
the X-Ray Tube. For the Cardiac Applications (Cardiac and Car. Oblique),
the Need for Angulation Initially Places the X-Ray Tube Farther From the
Magnet Shroud, But Angling the Gantry Then Places the Tube at Smaller R
and Therefore in Higher Magnetic Fields
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
476 Proceedings of the IEEE | Vol. 96, No. 3, March 2008
These geometries have guided our initial investigationsof the operation of a rotating-anode X-ray tube when
placed in external fields of the magnitudes described
above, and the behavior of both a standard induction motor
and the electron optics were characterized [24].
X-Ray Tube Motor in Fringe Field of Closed-Bore Magnet:When an induction motor is operating in the presence of
an external magnetic field, the stator soft iron ringbecomes partially magnetized by the external field, in
addition to the magnetization produced by the stator coils.
These effects might impact the performance of the motor
and hence decrease the frequency of anode rotation (f). Itis important to maintain high f to ensure proper
distribution and dissipation of heat from the focal track
to the body of the anode.
A Machlett DX 69B rotating anode X-ray tube insertwith a 4-in 400 000 heat units anode was used to test the
effect of external magnetic field on the induction motor.
The tube insert was removed from the housing to permit
visual contact with the anode, which allowed measure-
ment of rotational frequency using a calibrated Strobotac
Type 1531 strobe light (GenRad, Condord, MA). The high
voltage and filament cables were disconnected, and the
tube was mounted on a precision-machined aluminumstage with a rotation axis about the central point of the
induction motor soft iron ring in order to provide accurate
angular position control (see Fig. 9 inset). A transformer
circuit was used to drive the motor. Mimicking the usual
startup and run procedure, a 220 V root mean square
(VRMS) ac current was applied to the coils for 2 s to
initiate the rotation. The voltage was then switched to50 VRMS for steady-state operation.
The mounted tube was advanced axially into the fringe
field of a full body, unshielded 1.5T GE Signa CV/I MRI
system (S3 magnet), providing a relatively uniform field
of known direction throughout the insert volume,
depending on the distance of the insert from the magnet.
We assumed that B is parallel to the MRI central axis and
uniform throughout the induction motor volume andverified that the gradient in Bz throughout the motor
volume did not exceed 4% at any point and that Br always
remained below 4% of Bz along the center line of the
magnet. We evaluated the performance of the induction
motor for fields up to 0.24 T and angles between B and
the axis of rotation of the motor ranging from 0 to 90� inincrements of 15�.
The results are summarized in Fig. 9 and show that atransverse field (a field at 90� to the anode axis of rotation)has a significant effect on motor operation with rotation
speed falling below 3000 rpm for fields greater than 0.03 T.
Electron Optics in Fringe Field of Closed-Bore Magnet:When an X-ray tube is placed in a magnetic field of
arbitrary direction, the electron trajectories are the joint
result of the electric field (from the electrodes and thefocusing cup) and the magnetic field. One direct method to
study these trajectories is to solve the equations of motion
of the electrons. The problem may be simplified to a
classical one of an electron, with an effective initial velocity
v0, moving in a combination of electric and magnetic fields
E and B, both being uniform and static. The regime of
interest for this investigation was the presence of a smallBwith arbitrary direction relative to E. The motion of theelectrons may be viewed as a combination of 1) linear
uniformly accelerated motion in the direction of the
magnetic field with an initial velocity due to the electric
field component in that direction, 2) a drift at a constant
velocity, and 3) a rotation about B at a constant speed.
We verified our analytical solution using both the FE
model described above and for fields up to 150 Gauss (8 A)
using a Helmholtz coil pair (inner radius 21 cm, 380 turnsof magnet wire, dc power supply: 20 A maximum current,
50 V, PowerTen, Elgar Electronics Corp., San Diego, CA)
to generate a known field at the location of the focal spot of
our static anode X-ray tube. The resulting deflection has a
measured slope of �0.25 mm/A, which agrees well with
the perturbation theory calculation of 0.28 mm/A and can
be converted to deflection per tesla using the Helmholtz-
coil field-current coefficient of 1.870� 103 T/A to get aslope of 150 mm/T or�2 mm per 0.0150 T [25]. Using this
value, we calculated the expected deflection of the e-beam
in the rotating-anode X-ray tube due to a Br in the range of
0.015 to 0.1000 T. For the largest Br of 0.1022 T, the total
deflection could be as high as 15 mm, which could easily
drive the focal spot off the anode track in a rotating anode
tube with similar electron optic geometry.
Fig. 9. The effect of an externally applied magnetic field on the
induction motor of a rotating anode X-ray tube. The speed of rotation
was measured for a range of field magnitudes and orientations. The
measured data in the inset show the results for low magnetic fields at
large angles. The inset photograph shows the precision-machined
aluminum stand used tomaintain a known angle between the external
field and the axis of rotation of the motor.
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
Vol. 96, No. 3, March 2008 | Proceedings of the IEEE 477
We have built a nonmagnetic rotating-anode X-ray tubeinsert with minimal magnetic components (Rytech Inc.,
Toronto, ON, Canada) including removal of the iron core
between the stator windings, and replacement of all Ni and
magnetic stainless steel with nonmagnetic stainless steel
(see Fig. 10). The main goal of this effort is to allow
verification of our FE model of e-beam deflection for a
geometry where the field at the focal spot is determined onlyby the external field and not disturbed by other magneticmaterial. Preliminary results are shown in Fig. 11, where
good agreement between the FE model and measured
deflection is seen for a small field of 195 G in a directionequivalent to Br. These preliminary investigations point to
the need for active deflection coils with feedback to control
the magnetic field at the location of the focal spot.
SummaryVClosed-Bore XMR System: These preliminary
results illustrate the challenges to the tube design yet to be
met before full integration between a closed-bore MR
system and an X-ray fluoroscopy system can be achieved.The two main challenges to be addressed are X-ray tube
motor operation and efficiency and focal spot deflection.
The first challenge will possibly require redesign of the X-ray
tube motor to provide alternative mechanical approaches to
maintain anode rotation at the desired speed. A recent
publication [26] outlines several approaches that could be
considered, including springs and piezoceramic motors. The
second challenge can likely be met using a combination ofmagnetic shielding and active deflection coils. We measured
the field inside and outside of a 3-mm-thick box of mu metal
and found that a field reduction of �0.05 T can be achieved
in fields up to �0.1 T; new formulations of such metals are
quoted as having screening factors of five to seven for static
fields, with a saturation field of 0.8 T (MagnoShield
COMPOUND panels, Aaronia AG, DE-54597 Euscheid,
Germany). The impact of motor options, shielding, andactive coil solutions on the field homogeneity of the magnet
must, of course, be considered, and some compromise in
X-ray tube operation may be necessary.
IV. CONCLUSIONS
The hybrid X-ray/MR system developed using the Signa SP
open bore magnet has proven the utility of closeintegration between the two systems, facilitating fast and
Fig. 10. (a) An X-ray tube built to our specifications (Rytech Inc.) with minimal magnetic material including (b) an iron-free stator
and (c) a nickel-free focusing cup. The tube is shown during heating and outgassing in (d).
Fig. 11. Our first experiment with the nonmagnetic X-ray tube to
determine deflection of the focal spot when placed in a magnetic field
that is oriented perpendicular to the anode-cathode direction or in
the Br direction (see Fig. 8). Our finite-element simulation (same scale
as experimental data) shows good agreement with the simulations,
with both indicating a deflection of approximately 5 mm.
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
478 Proceedings of the IEEE | Vol. 96, No. 3, March 2008
repeated switching between the two imaging modalitiesduring an interventional procedure. We have shown that,
in principle, TFT-based flat-panel digital detectors are
immune to magnetic field effects and that static-anode
X-ray tubes operate well in magnetic fields if alignment
between the magnetic and electric field is maintained.
The homogeneity of the magnetic field can be maintained
in spite of the proximity of the X-ray detector to the MR
field of view, and RF noise and eddy currents do notsignificantly degrade MR image quality. We have also
carried out preliminary experiments to evaluate the
performance of a rotating-anode X-ray tube in the fringe
fields of a closed-bore 1.5 T magnet. Some modification
of X-ray tube design will likely be required for closest
integration of X-ray fluoroscopy with a closed-bore MR
system, although relatively simple solutions may be
possible. Our experience has shown that interventionalphysicians need, and are demanding, better image
guidance than is currently available. XMR systems are,
we believe, practical systems that can achieve the goal of
better guidance. While some technical challenges remain,
there are several possible approaches to overcoming these
challenges and making the dream of a practical, dual-
modality image guidance system a reality. h
Acknowledgment
The authors would like to thank S. Williams,
P. Neysmith, H. Easton, N. R. Bennett, and G. DeCrescenzo
for technical assistance. They also greatly appreciate the
technical and surgical assistance provided by C. Cooper and
G. Johnson during patient procedures.
REFERENCES
[1] T. J. Vogl, J. O. Balzer, M. G. Mack, G. Bett,and A. Oppelt, BHybrid MR interventionalimaging system: Combined MR andangiography suites with single interactivetable. Feasibility study in vascular livertumor procedures,[ Eur. Radiol., vol. 12,pp. 1394–400, 2002.
[2] M. W. Wilson, N. Fidelman, O. M. Weber,A. J. Martin, R. L. Gordon, J. M. LaBerge,R. K. Kerlan, Jr., K. A. Wolanske, andM. Saeed, BExperimental renal arteryembolization in a combined MR imaging/angiographic unit,[ J. Vasc. Interv. Radiol.,vol. 14, pp. 1169–1175, 2003.
[3] K. S. Rhode, D. L. Hill, P. J. Edwardsa,J. Hipwell, D. Rueckert, G. Sanchez-Ortiz,S. Hegde, V. Rahunathan, and R. Razavi,BRegistration and tracking to integrate X-rayand MR images in an XMR facility,[ IEEETrans. Med. Imag., vol. 22, pp. 1369–1378,2003.
[4] J. Kucharczyk, W. A. Hall, W. C. Broaddus,G. T. Gillies, and C. L. Truwit, BCost-efficacyof MR-guided neurointerventions,[Neuroimag. Clin. North Amer., vol. 11,pp. 767–772, 2001, xii.
[5] R. M. Chu, R. P. Tummala, J. Kucharczyk,C. L. Truwit, and R. E. Maxwell, BMinimallyinvasive procedures. Interventional MRimage-guided functional neurosurgery,[Neuroimag. Clin. North Amer., vol. 11,pp. 715–725, 2001.
[6] K. S. Rhode, M. Sermesant, S. Hegde,I. S. Gerardo, D. Rueckert, R. Razavi, andD. L. Hill, BXMR guided cardiacelectrophysiology study and radio frequencyablation,[ in Proc. SPIE Med. Imag. 2004:Physiol., Function Structure Med. Images, 2004,vol. 5369, pp. 10–21.
[7] K. S. Rhode, P. Lambaise, S. Hegde,M. Sermesant, D. L. Hill, R. Razavi, andJ. S. Gill, BReal-time XMR guidance forcardiac electrophysiology procedures,[ inProc. 5th Interventional MRI Symp., Boston,MA, 2004.
[8] S. Hegde, M. E. Miquel, K. S. Rhode,V. Muthurangu, D. L. Hill, S. A. Qureshi, andR. Razavi, BTowards safer cardiacintervention: A novel approach that combinesX-ray and magnetic resonance imaging forguidance of stent implantation,[ in Proc. 5th
Interventional MRI Symp., Boston, MA, 2004,pp. 82–83.
[9] A. J. Dick, V. K. Raman, A. N. Raval,M. A. Guttman, R. B. Thompson, C. Ozturk,D. C. Peters, A. M. Stine, V. J. Wright,W. H. Schenke, and R. J. Lederman, BInvasivehuman magnetic resonance imaging:Feasibility during revascularization in acombined XMR suite,[ Cath. Cardio. Interv.,vol. 64, pp. 265–274, 2005.
[10] R. Fahrig, Z. Wen, A. Ganguly,G. DeCrescenzo, J. A. Rowlands,G. M. Stevens, R. F. Saunders, andN. J. Pelc, BPerformance of a static-anode/flat-panel X-ray fluoroscopy system in adiagnostic strength magnetic field: A trulyhybrid X-ray/MR imaging system,[Med. Phys., vol. 32, pp. 1775–1784, 2005.
[11] J. A. Rowlands and J. Yorkston, BFlat paneldetectors for digital radiography,[ inHandbook of Medical Imaging, Vol. 1:Physics and Psychophysics, vol. 1, J. Beutel,H. L. Kundel, and R. L. Van Metter, Eds.Bellingham, WA: SPIE, 2000, pp. 223–328.
[12] J. T. Dobbins, 3rd, D. L. Ergun, L. Rutz,D. A. Hinshaw, H. Blume, and D. C. Clark,BDQE(f) of four generations of computedradiography acquisition devices,[ Med. Phys.,vol. 22, pp. 1581–1593, 1995.
[13] Z. Wen, R. Fahrig, and N. J. Pelc, BX-ray tubein parallel magnetic fields,[ in Proc. SPIE,2003, vol. 5030, pp. 972–979.
[14] H. Yu, R. Fahrig, and N. J. Pelc,BCo-registration of X-ray and MR fields ofview in a hybrid XMR system,[ J. Magn. Res.Imag., vol. 22, pp. 291–301, 2005.
[15] J. M. Santos, G. A. Wright, and J. M. Pauly,BFlexible real-time magnetic resonanceimaging framework,[ in Proc. 26th Annu.Int. Conf. IEEE EMBS, 2004, p. 1048.
[16] J. M. Santos, M. McConnell, G. C. Scott,M. Hyon, and J. M. Pauly, BMulti-coilreal-time interventional system,[ in Proc. Int.Soc. Magn. Res. Med. 11th Annu. Meeting 2003,2003, p. 1197.
[17] J. M. Santos, G. C. Scott, and J. M. Pauly,BReal-time imaging and active device trackingfor catheter-based MRI,[ in Proc. Int. Soc.Magn. Res. Med. 9th Annu. Meeting 2001, 2001,p. 2169.
[18] J. M. Santos, G. A. Wright, P. Yang, andJ. M. Pauly, BAdaptive architecture forreal-time imaging systems,[ in Proc. Int. Soc.Magn. Res. Med. 10th Annu. Meeting 2002,2002, p. 468.
[19] P. C. Yang, J. M. Santos, P. K. Nguyen,G. C. Scott, J. Engvall, M. V. McConnell,G. A. Wright, D. G. Nishimura, J. M. Pauly,and B. S. Hu, BDynamic real-time architecturein magnetic resonance coronaryangiographyVA prospective clinical trial,[J Cardio. Magn. Res., vol. 6, pp. 885–894,2004.
[20] J. M. Santos, C. H. Cunningham, M. Lustig,B. A. Hargreaves, B. Hu, D. G. Nishimura, andJ. Pauly, BSingle breath-hold whole-heartMRA using variable-density spirals at 3T,[Magn. Res. Med., vol. 55, pp. 371–379, 2006.
[21] V. Rieke, A. Ganguly, B. L. Daniel, G. Scott,J. M. Pauly, R. Fahrig, N. J. Pelc, and K. Butts,BX-ray compatible radiofrequency coil formagnetic resonance imaging,[ Magn. Res.Med., vol. 53, pp. 1409–1414, 2005.
[22] S. T. Kee, A. Ganguly, B. L. Daniel, Z. Wen,K. Butts, A. Shimikawa, N. J. Pelc, R. Fahrig,and M. D. Dake, BMR-guided transjugularintrahepatic portosystemic shunt creationwith use of a hybrid radiography/MR system,[J. Vasc. Interv. Radiol., vol. 16, pp. 227–234,2005.
[23] R. Fahrig, G. Heit, Z. Wen, B. L. Daniel,K. Butts, and N. J. Pelc, BFirst use of atruly-hybrid X-ray/MR imaging system forguidance of brain biopsy,[ Acta Neurochir(Wien), vol. 145, pp. 995–997, 2003,discussion 997.
[24] L. Brzozowski, A. Ganguly, M. Pop, Z. Wen,R. Bennett, R. Fahrig, and J. A. Rowlands,BCompatibility of interventional X-ray andmagnetic resonance imaging: Feasibility of aclosed bore XMR (CBXMR) system,[ Med.Phys., vol. 33, pp. 3033–3045, 2006.
[25] Z. Wen, R. Fahrig, S. Conolly, and N. J. Pelc,BInvestigation of electron trajectories of anX-ray tube in magnetic fields of MRscanners,[ Med Phys, vol. 34, pp. 2048–2058,2007.
[26] W. F. Block and J. S. Price, BMR/X-rayscanner having rotatable anode,[ U.S. Patent6 973 162 B2, 2005.
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
Vol. 96, No. 3, March 2008 | Proceedings of the IEEE 479
ABOUT THE AUTHORS
Rebecca Fahrig, photograph and biography not available at the time of
publication.
Arundhuti Ganguly, photograph and biography not available at the time
of publication.
Prasheel Lillaney, photograph and biography not available at the time
of publication.
John Bracken, photograph and biography not available at the time of
publication.
John A. Rowlands, photograph and biography not available at the time
of publication.
Zhifei Wen, photograph and biography not available at the time of
publication.
Huanzhou Yu, photograph and biography not available at the time of
publication.
Viola Rieke, photograph and biography not available at the time of
publication.
Juan M. Santos, photograph and biography not available at the time of
publication.
Kim Butts Pauly, photograph and biography not available at the time of
publication.
Daniel Y. Sze, photograph and biography not available at the time of
publication.
Joan K. Frisoli, photograph and biography not available at the time of
publication.
Bruce L. Daniel, photograph and biography not available at the time of
publication.
Norbert J. Pelc, photograph and biography not available at the time of
publication.
Fahrig et al. : Hybrid X-Ray/MR System for Interventional Guidance
480 Proceedings of the IEEE | Vol. 96, No. 3, March 2008