Comparison of the performance of three ion mobility spectrometers for measurement of biogenic amines
Transcript of Comparison of the performance of three ion mobility spectrometers for measurement of biogenic amines
Comparison of the performance of three ion 1
mobility spectrometers for measurement of 2
biogenic amines 3
Zeev Karpasa*
, Ana V. Guamánb,c
, Antonio Pardob, Santiago Marco
b,c 4
a 3QBD, Arad, Israel, and Chemistry Department, Nuclear research Center, Negev, Beer-5
Sheva 84190, Israel 6
b Departament d’Electrònica, Universitat de Barcelona, Martí i Franqués 1, 08028-7
Barcelona (Spain). 8
c Artificial Olfaction Lab, Institute for Bioengineering of Catalonia, Baldiri i Rexach, 4-8, 9
08028-Barcelona (Spain) 10
A. V. Guaman: [email protected] 11
A. Pardo: [email protected] 12
S. Marco: [email protected] 13
*[email protected], fax: 972-8-6469718; phone: 972-50-6232140 14
Abstract 15
The performance of three different types of ion mobility spectrometer 16
(IMS) devices: GDA2 with a radioactive ion source (Airsense, 17
Germany), UV-IMS with a photo-ionization source (G.A.S. Germany) 18
and VG-Test with a corona discharge source (3QBD, Israel) was 19
studied. The gas-phase ion chemistry in the IMS devices affected the 20
species formed and their measured reduced mobility values. The 21
sensitivity and limit of detection for trimethylamine (TMA), putrescine 22
and cadaverine were compared by continuous monitoring of a stream 23
of air with a given concentration of the analyte and by measurement of 24
headspace vapors of TMA in a sealed vial. Preprocessing of the 25
mobility spectra and the effectiveness of multivariate curve resolution 26
techniques (MCR-LASSO) improved the accuracy of the 27
measurements by correcting baseline effects and adjusting for 28
variations in drift time as well as enhancing the signal to noise ratio 29
and deconvolution of the complex data matrix to their pure 30
components. The limit of detection for measurement of the biogenic 31
amines by the three IMS devices was between 0.1 and 1.2 ppm (for 32
TMA with the VG-Test and GDA, respectively) and between 0.2 and 33
0.7 ppm for putrescine and cadaverine with all three devices. 34
Considering the uncertainty in the LOD determination there is almost 35
no statistically significant difference between the three devices 36
although they differ in their operating temperature, ionization method, 37
drift tube design and dopant chemistry. This finding may have general 38
implications on the achievable performance of classic IMS devices. 39
Keywords: ion mobility spectrometry, comparison of performance, 40
biogenic amines, signal processing, sensitivity, vapor concentration 41
1. Introduction 42
The advent of modern ion mobility spectrometry (IMS) can be traced 43
back to the late 1960's and its appearance as an analytical technique 44
followed shortly after, as described in detail previously [1]. In the four 45
decades that followed, several vendors of analytical equipment have 46
marketed a variety of commercial products, using different ionization 47
sources, drift tube designs, operating temperatures and ranging in size 48
from pocket size to walk-in portals. These focused mainly on 49
applications such as the detection of explosives, narcotics and toxic 50
chemicals where the required response was simply "positive" or 51
"negative" for the presence of the target compound or analyte. 52
Surprisingly, only few direct comparisons of the performance of 53
different instruments are available, although reports of proton-54
attachment versus electron exchange ionization [2] and of high-field 55
ion mobility drift tubes and a FAIMS apparatus [3] were published. 56
Several studies attempted to compare the effect of the ionization 57
method on the performance of IMS instruments. Notable among those 58
are the reports of Borsdorf et al. [4-6] in which the gas phase ion 59
chemistry of isomeric hydrocarbons [4], terpenes [5] and substituted 60
toluene and aniline compounds [6] was compared when a radioactive 61
63Ni, a corona discharge (CD) and a photoionization (PI) ion source 62
was used. In the latter study, where three similar RAID1 (Bruker, 63
Germany) handheld IMS devices that differed only in their ion source, 64
the measured reduced mobility and relative abundance of the product 65
ions differed significantly among these instruments [6]. A few other 66
reports on the comparison of the performance of two types of IMS 67
instruments towards detection of odor signatures of smokeless gun 68
powders [7] and drugs [8,9] were also published. In many cases 69
vendors of IMS devices report the level of detection of their device for 70
a given set of compounds (usually belonging to one of the above 71
applications) allowing consumers to compare the instruments on the 72
basis of the manufacturers' claims [10]. 73
In the last few years, the interest in using IMS techniques has grown 74
considerably for several novel applications as recently reviewed [11], 75
particularly in the biological and medical fields, such as in the areas of 76
monitoring food safety and study of pathological conditions. Biogenic 77
amines are compounds that are present in each living cell and play an 78
important role in regulating the cell functions as described in a mini-79
review [12]. Thus, biogenic amines may serve as important markers 80
for diseases like bacterial vaginosis [13,14] and cancer [15] or food 81
spoilage [16-18]. The high proton affinity of amines in general, and 82
biogenic amines in particular, allows their measurement by stand-alone 83
IMS instruments without need for pre-separation and pre-concentration 84
that may be required for other applications where matrix effects could 85
deleteriously affect the measurements. Three important biogenic 86
amines were tested here: trimethylamine (TMA), putrescine (1,4-87
diaminobutane) and cadaverine (1,5-diminopentane). 88
In the present study, the performance of three different types of ion 89
mobility spectrometer (IMS) devices with regard to the measurement 90
of these three biogenic amines was compared. The sensitivity and limit 91
of detection for the three amines were determined by continuous 92
monitoring of a stream of air with a given concentration of the analyte. 93
For the volatile TMA measurement of headspace vapors in a sealed 94
vial to determine the response of the IMS instruments to the vapors 95
emanating from a given quantity of the biogenic amine deposited in the 96
vial was also tested. Advanced signal preprocessing, combined with 97
multivariate curve resolution techniques (MCR-LASSO) was carried 98
out to improve the accuracy of the spectra by extracting pure 99
components from complex data matrix [19,20]. 100
2. Experimental Section 101
2.1 Preparation of the samples and measurement methodology 102
Trimethylamine (purum, 45% in water), putrescine (99%, 1,4-103
diaminobutane), toluene (99.8%) and triethylphosphate (99.8%) were 104
purchased from Sigma-Aldrich and cadaverine (purum >97% 1,5-105
diminopentane) was acquired from Fluka. A stock solution of TMA 106
was prepared by weighing a sample of the analyte and dissolving it in a 107
weighed quantity of distilled water. Further dilute solutions were 108
prepared by mixing a measured volume of the stock solution with 109
distilled water. 110
A sample from each of the three amines was inserted in a permeation 111
tube that was placed in an oven equipped with three independently 112
controlled chambers (Owlstone OVG-4, UK) at the selected 113
temperature. The amount of the sample that emanated from the 114
permeation tube was determined by weighing the tube periodically. 115
However, it should be noted that the reaction of moisture in air with the 116
vapors of the amines may lead to deposition of reaction products on 117
external surfaces of the permeation tube leading to erroneous weight 118
results. Therefore the amount of sample emanating from the tube was 119
calibrated with pure, dry nitrogen. The air flowing through the oven 120
compartment was mixed with a stream of clean air to dilute the 121
concentration of the sample vapors. The rate of permeation depends on 122
the oven temperature so that combining the selected temperature with 123
the dilution factor was used to supply the analyte vapors according to 124
the desired concentration range. Ten different concentration with one 125
replicate were measured and the maximum concentrations (“zero split” 126
in the oven) of TMA (at 70ºC), putrescine (at 90ºC) and cadaverine (at 127
90ºC) in a carrier flow 400 mL min-1
of air were, 11.15, 16.21 and 8.49 128
ppm (by weight), respectively. Measurements were carried out to 129
determine the response of the three IMS instruments to a concentration 130
of analyte vapors in a stream of air. The airstream was introduced by 131
Teflon tubing to the inlet port of each device. The response was 132
measured and a limit of detection to vapors was derived from the 133
calibration curve. 134
For the headspace measurements, 50 µL of 15% KOH were placed in a 135
glass vial and then the selected amount of TMA in solution was added 136
and the vial was immediately sealed with an aluminum cap with a 137
PTFE/silicone septum. At the start of the measurement the septum was 138
pierced simultaneously by two syringe needles: one connected to a 139
stream of carrier (pure air) and the other to a Teflon 1/8" tube attached 140
to the inlet of the IMS. The carrier flow through the headspace vial 141
was 400 mL min-1
for the GDA2 and VG-Test and 100 mL min-1
for 142
the G.A.S. UV instrument. In the VG-Test the analyte vapors were 143
somewhat further diluted by the instrument's own carrier flow of 240 144
mL min-1
. 145
146
2.2 The Ion Mobility Spectrometers 147
The three IMS devices used in the present study were the handheld 148
GDA2 (Airsense, Germany [21]), the portable UV-IMS (G. A. S. 149
Germany [22]) and the desktop VG-Test (3QBD, Israel [23]). These 150
devices differ from one another in several aspects, as summarized in 151
Table 1. The most important features are the type of ion source, dopant 152
ion chemistry and operating temperature. Toluene was used as a 153
dopant in the UV-IMS instrument and the VG-Test contained a 154
permeation tube with triethylphosphate (TEP) as a dopant while the 155
GDA2 did not contain a dopant and thus the ionization processes are 156
based on so called "water chemistry". The drift tube temperature in the 157
GDA2 was 40-45ºC, in the VG-Test was 90ºC while the G.A.S. 158
operated at ambient temperature (about 26ºC). These differences led 159
to some disparities in the types of product ions found from each analyte 160
as described in the following section. 161
162
2.3 Safety considerations 163
According to the MSDS trimethylamine, putrescine, cadaverine and 164
lutidine are quite similar from the viewpoint of safety. They are 165
hazardous in case of skin or eye contact and of inhalation. They are 166
corrosive to skin and eyes on contact and liquid or spray mist may 167
produce tissue damage particularly on mucous membranes of eyes, 168
mouth and respiratory tract. Skin contact may also cause burns. 169
Inhalation of the spray mist may lead to severe irritation of respiratory 170
tract, characterized by coughing, choking, or shortness of breath. They 171
should be handled in a chemical hood. Potassium hydroxide is 172
corrosive and may cause serious burns. It is harmful by ingestion, 173
inhalation and when in contact with skin. If the solid or solution comes 174
into contact with the eyes, serious eye damage may result. 175
176
Table 1. Comparison of the main design and operating parameters of 177
the three IMS devices used in the present study. 178
GDA2 G. A. S.
UV-IMS
VG-Test
Type Handheld Portable Desktop
Ion source 63
Ni
100 MBq
UV
Lamp
10.6 eV
Corona Discharge
Standard
inlet
Gas or
vapors
Gas or
vapors
Swab
Drift Tube
temperature
[ºC]
40-45 Ambient 90
Dopant Water
chemistry
Toluene Triethylphosphate
Standard
flow of
sample [mL
min-1
]
400 100 240 + pump
suction
Drift Gas
flow
[mL min-1
]
200 150 200
Shutter
Grid Type
Bradbury
-Nielson
Bradbury
-Nielson
Tyndall-Powell
Opening
Time
[sec]
200 500 200
Drift
Length [cm]
6.29 12 6.4
Pressure (P) Ambient Ambient Ambient
Inlet Type Membrane Open
System
Open System
Electric
field
[V cm-1
]
289 320 280
179
180
181
2.4 Spectral Processing for the mobility spectra 182
The spectral processing for the mobility spectra used in this work 183
consisted of three main steps. (i) Signal preprocessing. (ii) Multivariate 184
signal analysis using a bilinear decomposition with non-negativite 185
constraints followed by quantitative calibration. (iii) The calculation of 186
the limit of detection. 187
The preprocessing of the mobility spectra (i) is needed to improve the 188
accuracy of the posterior data analysis and includes baseline correction, 189
peak alignment and noise filtering. The baseline from each spectrum 190
was corrected by fitting and subtracting a fourth order polynomial 191
using parts of the spectrum where no peaks were identified. Noise 192
reduction was performed using second order Savitzky-Golay [24] filter 193
with a 15 points sliding window. Finally, misalignment of each 194
spectrum was corrected with shift in x-axis (drift time) taking the 195
position of the reactant ion peaks (RIP) – the water peak for the GDA2 196
and the TEP peak for VG-TEST - as reference. This preprocessing 197
procedure was applied independently spectrum by spectrum. 198
However, a special preprocessing technique was applied to the UV-199
IMS spectra because the signal showed interfering noise from other 200
laboratory equipment. Unfortunately the noise was at quite a low 201
frequency (113 Hz arising from the vent pump of the chemical hood) 202
and in-band, precluding the use of frequency selective filters. Thus, a 203
filtering procedure based on principal component analysis (PCA) was 204
attempted. PCA is a technique widely used in image preprocessing to 205
reduce noise and improve signal to noise ratio [25,26]. This is a good 206
solution in those cases where the noise is approximately orthogonal to 207
the signal. Every measurement consisted of 60 spectra that were all 208
used for building the PCA model. Enough loadings were retained to 209
capture 95% of the total variance. Visual inspection of the loadings 210
was used to identify those dimensions that contained basically additive 211
noise. Subsequent spectra were projected to the subspace orthogonal to 212
the noise. At the end, a new data matrix was constructed using only the 213
loadings related to compound information. Assuming noise was 214
stationary during a single measurement; a dedicated PCA model was 215
used for every measurement. On the other hand, the alignment of 216
spectra was uniquely based on the location of the peak of interest 217
because the reactant ion peak (protonated toluene) in the UV-IMS 218
instrument was very broad so it could not be taken as a reference peak 219
for calibration of the drift time scale. 220
The procedure used for processing the ion mobility spectra (ii) was 221
described in some detail recently [20], so only a brief explanation will 222
be given here. The procedure was used for an optimal physical and 223
chemical interpretation of the system, and the analysis of the IMS 224
dataset was done both qualitatively and quantitatively. Multivariate 225
curve resolution with LASSO (MCR-LASSO) [27] was used to resolve 226
IMS spectra yielding a spectral profile and a concentration profile for 227
each species in the sample. Since the analytes were pure and the 228
concentrations rather low, the number of pure components was set at 229
two. One related to the analyte of interest (protonated monomer) and 230
the other related to the dopant (TEP in the VG-Test, Toluene in the 231
UV-IMS and the water chemistry reactant ion peak in the GDA). The 232
resultant calibration profile was used to build a calibration model. In 233
this context, two types of calibration models were performed. The first 234
model was a univariate model which was applied to the measurements 235
done at constant concentration using the volatile generator, and the 236
other was using polynomial-Partial Least Squares (poly-PLS) [28] in 237
the case of headspace measurement. The use of poly-PLS was applied 238
to the concentration profile (after MCR-LASSO) in which an 239
appreciable change in the evolution during the measurement is 240
observed, and at different concentrations the behaviour of the transient 241
changes. Additionally, both the polynomial order and the number of 242
latent variables were calculated using "leave-one-out" cross-validation. 243
Those models were built before calculating the limit of detection. 244
The calculation of the limit of detection (LOD) was based on the upper 245
and lower boundaries of the calibration model and not on blanks, 246
following IUPAC recommendations [29], 247
where q is the slope of the calibration curve, sy is the variance of the 248
data, c represents the concentration levels, t is the t-statistic value with 249
v degrees of freedom and α (0.05) is the confidence level. The v 250
degrees of freedom are estimated according the n number of calibration 251
points. 252
253
3. Results and Discussion 254
3.1 The effect of preprocessing the spectra 255
The signal to noise ratio (SNR) improved due to the preprocessing as 256
seen in the mobility spectra recorded for the VG-Test, GDA2 and UV-257
IMS in Figures 1a, 1b and 1c, respectively. 258
The three IMS devices were located very close to one another on the 259
laboratory bench but the UV-IMS seemed to be more sensitive to 260
external noise and vibrations than the others. This noise is stationary 261
within a measurement. UV-IMS spectra showed a correlated noise that 262
has been corrected with the preprocessing methodology explained 263
above. Although a good SNR has been obtained, the sensitivity to 264
external noise entails the development of tailored preprocessing 265
techniques. An additional consideration is the lack of a reference peak 266
in the UV-IMS because it does not allow accurate well defined peak 267
location and an appropriate alignment between samples (see Figure 1c). 268
However, adding a calibrant in all measurements with a peak located 269
far from the region of interest could be a good solution for the 270
alignment. 271
On the other hand, the preprocessing methodology proposed above for 272
both VG-Test and GDA can be used for off-line processing and the 273
enhanced SNR will improve the performance of both instruments. 274
(a)
(b)
(c)
Figure 1. The raw (left frame) and pre-processed (right frame) mobility 275
spectra of TMA with a concentration of ~5 ppm recorded for the (a) 276
VG-Test (b) GDA (c) UV-IMS. Note that the different reactant ions: 277
TEP, protonated water cluster and toluene, respectively. 278
3.2 The reduced mobility values 279
The reduced mobility values of the ions formed in TMA, putrescine 280
and cadaverine were determined relative to that of 2,4-lutidine [30]. 281
Due to the different operational conditions of the three IMS devices the 282
reduced mobility values appeared to differ significantly in some cases 283
(Table 2). In all cases the major ion produced from each of the 284
biogenic amines was the protonated monomer molecule. The core ions 285
are usually clustered with several water molecules and drift gas 286
molecules and the number of clustered molecules depends on the 287
conditions in the drift tube, mainly on the humidity and operating 288
temperature. Only in the GDA2 were pure proton bridged dimers 289
observed and when both putrescine and cadaverine were present a 290
mixed dimer was also formed. The ratio between the intensity of the 291
dimer and monomer peaks depended on the concentration of the 292
analyte vapors, as expected, but quantitative results were not 293
calculated. In the study of the gas phase ion chemistry of substituted 294
toluene and aniline compounds differences in the reduced mobility 295
values of the product were seen [6]. The formation of proton bridged 296
dimers was observed in the alkyl (methyl and ethyl) aniline derivatives 297
when a 63
Ni and PI ion sources were used but not with the CD source 298
[6]. As noted above, the three RAID1 instruments had similar drift 299
tubes and were operated under similar conditions and the only 300
difference was in the ion source. 301
The reduced mobility value for the putrescine protonated monomer 302
measured with the GDA2 instrument (1.94 cm2V
-1s
-1) differed 303
significantly from the value obtained with UV-IMS (1.99 cm2V
-1s
-1) 304
and the VG-Test device in the present study (2.02 cm2V
-1s
-1) and also 305
from the value of 2.04 cm2V
-1s
-1 reported in a recent work with a drift 306
tube temperature of 232C [31]. However, in studies carried out with 307
the VG-Test at 3QBD, Israel, measuring putrescine or cadaverine 308
vapors emanating from a sample that was placed on a cotton swab, an 309
additional peak for the monomers with reduced mobility values of 1.93 310
and 1.84 cm2V
-1s
-1, respectively, with slightly longer drift times were 311
observed and the relative intensity of the two species changed during 312
the measurement probably due to variations in the operational 313
conditions. Some differences between the three instruments were also 314
observed for the reduced mobility values of protonated monomers of 315
TMA and cadaverine, where TMA was quite close between the VG-316
Test and the UV-IMS and cadaverine was almost similar for the GDA2 317
and UV-IMS. 318
It should be noted that for a given analyte in a given instrument, the 319
drift time and reduced mobility did not change with concentration or 320
with the sample introduction method (constant concentration in an air 321
stream of vapors or headspace vapors). The differences observed in the 322
reduced mobility values were beyond the uncertainties involved in the 323
drift time measurements and mobility scale calibration procedures and 324
are thus indicative of different ion species formed in each IMS due to 325
the differences in operating conditions (ion source, temperature, 326
moisture, reactant ion chemistry and structural features of the drift 327
tube). Thus, discrepancies in reduced mobility values that have been 328
reported for different IMS devices are not necessarily the result of 329
erroneous measurements but rather a natural outcome of variations in 330
ionic species formed under different experimental conditions. This may 331
have major repercussions on the transferability of reduced mobility 332
values between different instruments and even for the same instrument 333
operated under different conditions. 334
335
336
Table 2. The core ions observed in trimethylamine [TMA], putrescine 337
[PUT] and cadaverine [CAD] and their reduced mobility values [cm2 338
V-1
s-1
], calculated relative to 2,4-lutidine [LUT]. All core ions shown 339
here are probably clustered with water and drift gas molecules. 340
Compound
Temperature
Ion Species GDA2
44ºC
G. A. S.
26ºC
VG-Test
90ºC
2,4-Lutidine [LUT]H+ 1.90 1.90 1.90
TMA [TMA]H+ 2.22 2.13 2.10
[TMA]2H+ 1.78 Not observed Not observed
Putrescine [PUT]H+ 1.94 1.99 2.02
[PUT]2H+ 1.46 Not observed Not observed
Cadaverine [CAD]H+ 1.81 1.82 1.87
[CAD]2H+ 1.36 Not observed Not observed
Mixed [CAD][PUT]H+ 1.41 Not observed Not observed
341
3.3 Vapor phase measurements: Calibration and limit of detection 342
3.3.1 Vapors in air 343
There are different ways to display the quantitative response of an IMS 344
to changing amounts of the analyte, whether it is in the vapor phase or 345
deposited on a substrate (filter paper, a glass vial or a swab). One could 346
plot the decrease in the reactant ion peak height or peak area, the 347
increase in the product ion, or ions, produced from the analyte or the 348
ratio of the product ion peak to the sum of reactant and product ions. 349
Table 3 summarizes the response of the GDA2, VG-Test and UV-IMS 350
to the concentrations of TMA, putrescine and cadaverine as predicted 351
according to the MCR-LASSO and PLS procedure described above, as 352
a function of the real concentration of the analyte vapors in air. 353
354
Table 3: The dependence of the response of the VG-Test, GDA2 and 355
UV-IMS on the concentration of trimethylamine, putrescine and 356
cadaverine. 357
Sample
Type
Calculation
VG-Test GDA2 G.A.S
RMSECV* R2 RMSECV R
2 RMSECV R
2
TMA Air (ppm) MCR 0.2 0.95 1.5 0.91 1 0.88
TMA
Headspace
(g)
MCR 0.6 0.94 1.1 0.99 1.1 0.92
Putrescine Air (ppm) MCR 0.3 0.96 0.1 0.95 0.7 0.96
Cadaverine Air (ppm) MCR 0.2 0.97 0.09 0.94 0.1 0.97
Root-mean-square error of cross-validation 358
The sensitivity of the VG-Test is quite similar for the three analytes, 359
but the GDA2 and UV-IMS show a different sensitivity towards TMA 360
compared with cadaverine and putrescine. 361
Table 4 shows a summary of the limits of detection obtained for the 362
three IMS instruments for the three biogenic amines that were 363
measured here. The procedure to determine the limit of detection, 364
based on MCR-LASSO, was given in detail previously [19,20] so only 365
a brief description was presented above. 366
It is well known that the ultimate constraint for limit of detection 367
measurements is the noise level (and signal to noise ratio). For this 368
reason noise filtering and signal enhancement (as provided by MCR-369
LASSO) are key elements for improving the LOD performance for 370
different analytical instruments. In the present study, it should be noted 371
that there were differences in the appearance of mobility spectra and 372
response characteristics of the three instruments. The UV-IMS spectra 373
nominally had the best SNR but this was offset by the lower 374
reproducibility of the device for a given concentration of the analyte. 375
The LOD measurements with the GDA suffered from the long 376
stabilization and clearance times between samples that affected the 377
determination of the blank levels. This could be due to the fact that this 378
device had a membrane in the inlet (Table 1). 379
380
381
Table 4. The limit of detection calculated on the basis of MCR-382
LASSO for vapors of trimethylamine, putrescine and cadaverine in air 383
for the GDA2, UV-IMS and VG-Test ion mobility spectrometers. Air 384
density was taken as 1.16 g L-1
at 300K and 1 Bar. Also shown is the 385
limit of detection for TMA in headspace vapors emanating from a 386
sample deposited in a 20 mL headspace vial. 387
Compound Sample type VG-Test GDA2 G.A.S
TMA Vapor in air (ppm)
nmole/L
0.1±0.2
1.5±2.9
1.2±1.5
11.8±14.7
1.0±1.0
8.5±8.5
Headspace vapors (µg)
nmole
0.9±0.6
15.3±8.8
1.9±1.2
32.2±20.3
0.7±1.1
11.9±18.6
Putrescine Vapor in air (ppm)
nmole/L
0.5±0.3
7.3±4.4
0.7±0.1
6.9±1.0
0.7±0.7
5.9±5.9
Cadaverine Vapor in air (ppm)
nmole/L
0.4±0.2
5.8±2.9
0.2±0.1
2.0±1.0
0.4±0.1
3.4±0.8
388
The LOD showed some dependence on the analyte and on the type of 389
IMS device used, but the range between the lowest LOD (0.1 ppm of 390
TMA vapor with the VG-Test) and highest LOD (1.2 ppm of TMA 391
with the GDA) was quite narrow. The LOD for the diamines varied 392
between 0.2 ppm for cadaverine with the GDA to 0.7 ppm for 393
putrescine with the UV-IMS instrument. The LOD for TMA deposited 394
in a headspace vial ranged from 0.7 to 1.9 g for the UV-IMS and 395
GDA, respectively. The fact that the performance of the three devices 396
was quite similar is really surprising considering that they differ so 397
much from each other in their operating temperature, ionization source, 398
dopant chemistry and drift tube design. If one takes into consideration 399
the uncertainty in the LOD determination there appears to be almost no 400
statistically significant difference in the sensitivity between these three 401
instruments. This finding may have general implications as to the 402
possible limit of detection that can be achieved with classic IMS drift 403
tubes (without pre-concentration or separation). 404
3.3.2 Headspace vapors from deposited sample 405
The response of the three IMS devices to TMA vapors in a headspace 406
vial was similar in principle though different in detail. First a rapid 407
increase within a few seconds in the intensity of the analyte signal, 408
concomitant with the decrease in the reactant ion peak, and then a 409
gradual decrease in the analyte signal and increase in the reactant ion 410
peak as the headspace vapors begin to clear out. This is clearly seen in 411
Figure 2 that depicts the analyte (TMA) and reactant ion peaks in the 412
VG-Test and GDA2 as a function of time. These changes are due to the 413
fact that the flow of the carrier through the headspace vial dilutes the 414
analyte vapors and the rate of dilution depends on the flow and volume 415
of the vial. 416
Theoretically, with perfect mixing and transport, a carrier flow of 400 417
mL min-1
(6.67 mL s-1
) with a 20 mL vial volume should dilute the 418
analyte vapors as shown in the top and middle of Figure 3 for the VG-419
Test and GDA2, respectively. The bottom trace of Figure 3 was 420
recorded for the UV-IMS at a flow of 100 mL min-1
. The graph was 421
adjusted to allow for the 5 seconds delay until the maximum signal 422
intensity of the analyte is obtained. Evidently, the clearing time is 423
longer due to imperfect mixing and transport of analyte vapors from 424
the sample vial into the drift tube. 425
426
427
Figure 2. Top frame: The signal intensity of the analyte (TMA) and reactant ion (TEP) 428
peaks in a headspace vial as a function of time. Bottom frame: The signal intensity of 429
the analyte (TMA) and reactant ion (RIP) peaks in a headspace vial as a function of 430
time. 431
432
433
434
435
436
437
438
439
440
441
442
443
444
445
446
447
448
449
450
451
452
453
454
Figure 3: The theoretical (with perfect mixing) dilution of the TMA 455
headspace analyte vapors for a carrier flow of 400 mL min-1
(6.67 mL 456
s-1
) and a 20 mL vial volume for the VG-Test device (top) and the 457
GDA (middle). The response of the UV-IMS and theoretical dilution 458
for a flow rate of 100 mL min-1
is also shown (bottom trace). For the 459
actual measurement the graph (based on Figure 2) was adjusted to 460
allow for the first 5 seconds until the maximum signal intensity of the 461
analyte is obtained. 462
Figure 4 shows the response of the VG-Test to different amounts of 463
TMA that were placed in a headspace vial as a function of time for the 464
VG-Test (top) and the GDA (bottom). The plots of the ratio of the 465
analyte peak [(TMA/(TMA+TEP)] in the VG-Test and 466
[(TMA/(TMA+RIP)] in the GDA2 show that the ratio and clearance 467
time increases with the amount of TMA deposited in the vial, as 468
expected. The calibration curve and limit of detection calculated 469
according to the MCR-LASSO procedure are shown in Tables 3 and 4. 470
In this case the limit of detection for TMA headspace vapors with the 471
GDA2 is slightly inferior to the LOD calculated for the VG-Test and 472
UV-IMS but all values are in the range of 0.7-1.9 µg (6.9-21.6 nmole) 473
of TMA deposited in the headspace vial. 474
Measurement of headspace vapors of putrescine produced a different 475
response in all three IMS devices. The putrescine signal was not 476
observed initially, even when the headspace vial was heated in the 477
homemade oven to 100ºC for several minutes, after about five minutes 478
the intensity of the putrescine peak started to increase and continued to 479
do so long after, theoretically, all the vapors in the headspace vial 480
should have been transported to the IMS. The intensity continued to 481
increase even after the vial was removed, indicating that vapors 482
remained in glass vial, the transfer line and the instrument itself. This 483
phenomenon made it quite impossible to create a calibration plot for 484
putrescine similar to that made for TMA headspace vapors. 485
486
487
488
489
490
491
492
493
494
495
Figure 4: The response to different amounts of TMA (in µg) that were 496
placed in a headspace vial as a function of time for the VG-Test (top) 497
and the GDA (bottom). Note that the ratio [(TMA/(TMA+TEP)] (top) 498
and [(TMA/(TMA+RIP)] (bottom) and the clearance time increase with 499
the amount of TMA deposited in the vial. 500
501
502
4. Conclusions 503
The variation in the reduced mobility values determined with the three 504
IMS instruments is indicative of the fact that there are different ion 505
species formed in each IMS due to the differences in operating 506
conditions (ion source, temperature, moisture level, reactant ion 507
chemistry and structural features of the drift tube). This could be 508
attributed to variation in the degree of clustering that is affected by the 509
different operating temperatures and the humidity in the three devices 510
and to the nature and characteristics of the core ion. As mentioned 511
above, this may affect the transferability of reduced mobility values 512
between different instruments (especially when operated at low drift 513
tube temperatures and uncontrolled humidity) and even for the same 514
instrument operated under different conditions. A case in point is the 515
study where different product ions were found in three similar IMS 516
devices, operated with identical parameters except the type of ion 517
source used [6]. Quantitative measurements to determine the limit of 518
detection were not reported in that study [6]. The practical conclusion 519
is that while a database of reduced mobility values can serve as a 520
guideline for identification of ion species, the analyte should be 521
measured and calibrated under the same operational conditions with the 522
same IMS instrument that will serve as an analytical tool. 523
The difference between the theoretical dilution of the analyte vapors 524
and the measured signal intensity in the headspace experiment shows 525
that the clearing time is longer, due to imperfect mixing and delayed 526
transport of analyte vapors from the sample vial into the drift tube. The 527
measurement of headspace vapors of putrescine showed that transport 528
of vapors from the heated headspace vial was much less efficient than 529
the transport of the volatile TMA vapors. This was established by the 530
fact that the putrescine peak increased slowly after a considerable delay 531
and continued to do so long after all the vapors in the headspace vial 532
should have been transported to the IMS theoretically. This 533
phenomenon made it quite impossible to create a calibration plot for 534
putrescine similar to that made for TMA headspace vapors. 535
Building calibration curves with raw spectra or noise filtered spectra 536
can lead to very different results. Correct preprocessing of spectra and 537
applying the appropriate multivariate signal processing is essential 538
before establishing an appropriate comparison, since this study shows 539
that different instruments have different noise levels. Those differences 540
may be partially due to different conditions regarding grounding, 541
shielding and cabling and obviously different electronic filtering and 542
amplification signal chains in the instrument electronics as well 543
external mechanical interfering noise. We believe that proper digital 544
filtering can recover the inherent noise limits of the different IMS 545
technologies. Surprisingly, the three devices showed quite similar 546
limits of detection for the three analytes although they differ so much 547
in their operating temperature, ionization source, dopant chemistry and 548
drift tube design. Considering the uncertainty in the LOD 549
determinations there appears to be almost no statistically significant 550
difference between these three instruments. This finding may have 551
general implications as to the possible limit of detection that can be 552
References 556
1. G.A. Eiceman, Z. Karpas, Ion mobility spectrometry – Second Edition, 557
Chapter 1, CRC Press, Boca-Raton, FL, 2005. 558
2. R. R. Kunz, W. F. Dinatale, P. Becotte-Haigh, P. Int. J. Mass 559
Spectrom. 226 (2003) 379–395. 560
3. L.A. Viehland, R. Guevremont, R.W. Purves, D.A. Barnett, Int. J. 561
Mass Spectrom. 197 (2000) 123–130. 562
4. H. Borsdorf, J.A. Stone, G.A. Eiceman, Int. J. Mass Spectrom. 246 563
(2005) 19-28. 564
5. H. Borsdorf, M. Rudolph, Int. J. Mass Spectrom. 208 (2001) 67-72. 565
6. H. Borsdorf, K. Neitsch, G.A. Eiceman, J.A. Stone, Talanta 78 (2009) 566
1464-1475. 567
7. M. Joshi, Y. Delgado, P. Guerra, H. Lai, J.R. Almirall, Forens. Sci. Int. 568
188 (2009) 112-118. 569
8. C.W. Su, K. Babcock, S. Rigdon, Int. J. Ion Mobil. Spectrom. 1 570
(1998)15-27. 571
9. S.S. Choi, et al. Bull. Kor. Chem. Soc. 31 (2010) 2382-2385. 572
10. K. Cottingham, Anal. Chem, (2003) (Oct. 1), 435A-439A. 573
11. S. Armenta, M. Alcala, M. Blanco, Anal. Chim. Acta 703 (2011) 114-574
123. 575
12. K. Igarashi, K. Kashiwagi, Biochem. Biophys. Res. Comm. 271 (2000) 576
559-564. 577
13. J.D. Sobel, Z. Karpas, A. Lorber, Eur. J. Obstet. Gynecol. Reprod. 578
Biol. 163 (2012) 81–84. 579
14. Z. Karpas, W. Chaim, R. Gdalevsky, B. Tilman, A. Lorber, Anal Chim 580
Acta 474 (2002) 115–123 581
15. U. Bachrach, Amino Acids 26 (2004) 307–309. 582
16. G.M. Bota, P.B. Harrington, Talanta 68 (2006) 629–635. 583
17. Z. Karpas, B. Tilman, R. Gdalevsky, A. Lorber, Anal Chim Acta 463 584
(2002) 155–163. 585
18. C. Y. Chong, F. Abu Bakar, A. R. Russly, B. Jamilah, N. A. Mahyudin, 586
Intern. Food Res. J. 18 (2011) 867-876. 587
19. Z. Karpas, A.V. Guaman, D. Calvo, A. Pardo, S. Marco, Talanta 93 (2012) 588
200-205. 589
20. V. Pomareda, A.V. Guamán, M. Mohammadnejad, D. Calvo, A. Pardo, 590
S. Marco, Chemeometrics Intel. Lab. Syst. (2012), 591
DOI: 10.1016/j.chemolab.2012.06.002. 592
21. www.airsense.com/en/products/gda-2/ (accessed October 25, 2012) 593
22. www.gas-dortmund.de/Products/UV-IMS/1.336.html (accessed 594
October 25, 2012) 595
23. www.3qbd.com/English/ (accessed October 25, 2012) 596
24. A. Savitzsky, M.J.E. Golay, Anal. Chem. 36 (1964) 1627–1639. 597
25. P. Razifar, H.H. Muhammed, F.Engbrant, P. Svensson,J.Olsson,E. 598
Bengtsson, B Langstrom, M. Bergstrom, The Open Neuroimaging J. 3 599
(2009) 1-16. 600
26. M. Statheropoulos, A. Pappa, P. Karamertzanis, H.L.C Meuzelaar, 601
Anal. Chim. Acta, 401(1999) 35-43 602
27. V. Pomareda, D. Calvo, A. Pardo, S. Marco, Chemometrics Intel. Lab. 603
Syst. 104 (2010) 318-332. 604
28. S. Wold, N. Kettaneh-Wold, B. Skageberg, Chemometrics Intel. Lab. 605
Syst., 7 (1989) 53-65. 606
29. J. Mocak, A. M. Bond, S. Mitchell, S. Scollary, Pure & Appl. Chem, 607
69 (1997) 297-328. 608
30. G.A. Eiceman, E.B. Nazarov, J.A. Stone, Anal. Chim. Acta 493 (2003) 609
185-194. 610
31. S. Armenta, M. Blanco, Analyst (2012) doi:10.1039/C2AN35965K. 611
612
613
Legend of figures 614
Figure 1. The raw (left frame) and pre-processed (right frame) mobility 615
spectra of TMA with a concentration of ~5 ppm recorded for the (a) 616
VG-Test (b) GDA (c) UV-IMS. Note that the different reactant ions: 617
TEP, protonated water cluster and toluene, respectively. 618
Figure 2. Top frame: The signal intensity of the analyte (TMA) and 619
reactant ion (TEP) peaks in a headspace vial as a function of time. 620
Bottom frame: The signal intensity of the analyte (TMA) and reactant 621
ion (RIP) peaks in a headspace vial as a function of time. 622
Figure 3: The theoretical (with perfect mixing) dilution of the TMA 623
headspace analyte vapors for a carrier flow of 400 mL min-1
(6.67 mL 624
s-1
) and a 20 mL vial volume for the VG-Test device (top) and the 625
GDA (middle). The response of the UV-IMS and theoretical dilution 626
for a flow rate of 100 mL min-1
is also shown (bottom trace). For the 627
actual measurement the graph (based on Figure 2) was adjusted to 628
allow for the first 5 seconds until the maximum signal intensity of the 629
analyte is obtained. 630
Figure 4: The response to different amounts of TMA (in µg) that were 631
placed in a headspace vial as a function of time for the VG-Test (top) 632
and the GDA (bottom). Note that the ratio [(TMA/(TMA+TEP)] (top) 633
and [(TMA/(TMA+RIP)] (bottom) and the clearance time increase with 634
the amount of TMA deposited in the vial. 635
636
List of Tables 637
Table 1. Comparison of the main design and operating parameters of 638
the three IMS devices used in the present study. 639
Table 2. The core ions observed in trimethylamine [TMA], putrescine 640
[PUT] and cadaverine [CAD] and their reduced mobility values [cm2 641
V-1
s-1
], calculated relative to 2,4-lutidine [LUT]. All core ions shown 642
here are probably clustered with water and drift gas molecules. 643
Table 3: The dependence of the response of the VG-Test, GDA2 and 644
UV-IMS on the concentration of trimethylamine, putrescine and 645
cadaverine. 646
Table 4. The limit of detection calculated on the basis of MCR-647
LASSO for vapors of trimethylamine, putrescine and cadaverine in air 648
for the GDA2, UV-IMS and VG-Test ion mobility spectrometers. Air 649
density was taken as 1.16 g L-1
at 300K and 1 Bar. Also shown is the 650
limit of detection for TMA in headspace vapors emanating from a 651
sample deposited in a 20 mL headspace vial. 652
653