BERITA ACARA KALIBRASI MESIN HEMODIALISA
RUMAH SAKIT MODEL
NAMA TEKNISI NOMOR SERI
TGL KALIBRASI JAM KERJA MINUTE
A. SYSTEM SENSOR ELETRONIK & WATER WAYTOUCH SCREEN HASIL OK NO
SENSOR ADAPTER A-B HASIL OK NOSENSOR HANSEN DIALYSER HASIL OK NO
SENSOR WETER EMPTY ALARM HASIL OK NOSENSOR BLOOD LEVEL HASIL OK NO
SENSOR AIR BUBBLE HASIL OK NOBLOOD LEAK (Value ± 1600) NILAI TERUKUR HASIL OK NO
B. WETER INLET PRESSUREPRE PUMP 1,2 - 1,4 Kg/cm2 NILAI TERUKUR HASIL OK NO
POST PUMP 1,3 - 1,5 Kg/cm2 NILAI TERUKUR HASIL OK NOC. ARTERIAL & VENOUS PRESSURE
ARTERIAL PRESSURE 0 ± 2 mmHg NILAI TERUKUR HASIL OK NOVENOUS PRESSURE 0 ± 50 mmHg NILAI TERUKUR HASIL OK NO
D. CONDUCTIVITYCONDUCTIVITY 13,5 - 14,3 ms/Cm NILAI TERUKUR HASIL OK NO
E. TEMPERATURTEMPERATUR 33,5 - 40 ˚C NILAI TERUKUR HASIL OK NO
F. HEPARIN TESTSTROKE 0,49 – 0,54 ml/H NILAI TERUKUR HASIL OK NO
G. TEST & SIMULASIRUN 1200 cc/H NILAI TERUKUR HASIL OK NO
CATATAN HASIL KALIBRASI
DAPAT DIGUNAKAN PERLU DILAKUKAN PERBAIKAN
TEKNISI KA. TEKNISI RUMAH SAKIT
( ) ( ) ( )