DETERMINING BEST REGRESSIONAL MODEL IN PREDICTING FOETAL
WEIGHT IN PHILIPS ClearVue 350 ULTRASOUND EQUIPMENT
BY
EDEM ANASTASIUS KWAMI BUADY
(10434064)
A DISSERTATION SUBMITTED TO THE, UNIVERSITY OF GHANA, LEGON IN
PARTIAL FULFILLMENT OF THE REQUIREMENT FOR THE AWARD OF MSc
ULTRASONOGRAPHY DEGREE
JULY, 2014
University of Ghana http://ugspace.ug.edu.gh
ii
DECLARATION
I EDEM ANASTASIUS KWAMI BUADY do hereby declare that this thesis which is being
submitted in fulfillment of the requirements for the Master degree of MSc in Ultrasonography is
the result of my own research performed under supervision, and that except where otherwise
other sources are acknowledged and duly referenced, this work has not previously been accepted
in substance for any degree and is not being concurrently submitted in candidature for any
degree. I hereby give permission for the Department of Radiography to seek
dissemination/publication of the dissertation in any appropriate format. Authorship in such
circumstances to be jointly held between me as the first author and the project supervisors as
subsequent authors.
Signed ………………………………….…... Date……………………
EDEM ANASTASIUS KWAMI BUADY
(Student ID No.10434063)
Signed ……………………………………… Date……………………
DR. SAMUEL OPOKU
Signed …………………….…………………. Date……………………..
DR. S. ANIM-SAMPONG
Signed ……………………………………… Date………………………
DR. SAMUEL OPOKU
(Head of Department)
University of Ghana http://ugspace.ug.edu.gh
iii
DEDICATIONS
But for the protection of the Almighty God, this work would have been an impossible task. I
therefore dedicate this write up to the Almighty God for his wisdom, knowledge, understanding,
and guidance.
Consequently, dedication of this work is extended to my ever-loving mother Mrs. Janet Yawa
Komla-Buady and father Anthony Kwaku Buady who even in your sickbeds demonstrated a lot
of support, encouragement and love to me but as nature has it, you have all pass on to glory. May
the Almighty god protect you under his strong wing: Mum and Dad, I miss you all dearly.
University of Ghana http://ugspace.ug.edu.gh
iv
ACKNOWLEDGEMENT
Firstly, I would like to acknowledge the almighty God for his immense grace towards this
dissertation. My immense appreciation goes to the under mentioned personalities who helped in
diverse ways to ensure the success of this work.
Dr. S.Y. Opoku, Mr. K. Bamfo-Quaicoe, Dr. E.K. Ofori, Mr. W.K. Antwi, Mr. Sulley, Mr.
B.O..Botwe, Mr L. Arthur all of Department of Radiographer, School of Biomedical and Allied
Health Sciences for their incalculable intellectual contributions throughout the thesis process. I
would also like to thank Dr. V.K. Hewlett for his time and contribution towards this research
work. Finally, Dr. S. Anim-Sampongis highly acknowledged for his tireless effort in bringing
this dissertation to the successful end.
I also appreciate the cooperation of all the staff of Ussher and Mamprobi Polyclinics‟ ultrasound
and maternity units. To the departments of Radiography and Audiology, School of Allied Health
Sciences, College of Health Sciences of the University of Ghana, I want to render many thanks
for your diverse contribution for this study. In respect of their friendship and encouragement
towards the completion of the thesis, I thank my course mates, Mr. Ampofo, Richard, Seth,
Jonathan (Kamkpee) and Eric Saka-Boateng (Sharpest) for their advice, support and prayers.
And finally to my family and friends, especially my brother Senyo and Nana Akua Adubea your
care and support in physical and financially has been really appreciated. Thanks for your
unceasing love, encouragement and emotional support.
University of Ghana http://ugspace.ug.edu.gh
v
TABLE OF CONTENTS
DECLARATION.............................................................................................................................ii
ACKNOWLEDGEMENT ............................................................................................................. iv
TABLE OF CONTENTS................................................................................................................ v
LIST OF TABLES ......................................................................................................................... ix
LIST OF FIGURES ........................................................................................................................ x
LIST OF ABBREVIATIONS........................................................................................................ xi
ABSTRACT.................................................................................................................................. xii
CHAPTER ONE:INTRODUCTION .......................................................................................... 1
1.1 BACKGROUND.............................................................................................................. 1
1.2 PROBLEM STATEMENT .............................................................................................. 4
1.3 RESEARCH QUESTION/HYPOTHESES ..................................................................... 6
1.4 SIGNIFICANCE OF STUDY.......................................................................................... 6
1.5 AIMS ................................................................................................................................ 7
1.6 OBJECTIVES .................................................................................................................. 7
CHAPTER TWO: LITERATURE REVIEW.............................................................................
8
2.1 INTRODUCTION............................................................................................................ 8
2.2 HISTORICAL REVIEW OF ESTIMATING FOETAL WEIGHT ................................. 8
2.3 TACTILE ASSESSMENT OF FOETAL SIZE............................................................... 9
2.4 FOETAL WEIGHT PREDICTIONS USING ALGORITHM DERIVED FROM
MATERNAL AND PREGNANCY-SPECIFIC CHARACTERISTICS.................................. 10
2.5 OBSTETRIC ULTRASONOGRAPHY IN ESTIMATING FOETAL WEGHT .......... 11
2.5.1 Group 1: Abdominal Circumference Only ............................................................ 12
2.5.2 Group 2: Abdominal Circumference and Femur Length ........................................ 12
University of Ghana http://ugspace.ug.edu.gh
vi
2.5.3 Group 3: Abdominal Circumference and Biparietal Diameter ............................... 13
2.5.4 Group 4: Abdominal Circumference, Head Circumference(±Biparietal Diameter)14
2.5.4 Group 5: Abdominal Circumference, Femur Length, and Biparietal Diameter ..... 14
2.5.6 Group 6: Abdominal Circumference, Femur Length, and Head Circumference.... 15
2.5.7 Group 7: Abdominal, Head Circumferences, Femur Length, Biparietal Diameter 15
2.6 MAGNETIC RESONANCE IMAGING IN ESTIMATING FOETAL WEIGHT ....... 18
2.7 GENDER-SPECIFIC ESTIMATION OF FOETAL WEIGHT..................................... 19
2.8 ETHNIC AND RACIAL DIFFERENCES IN ESTIMATED FOETAL WEIGHT ...... 21
2.9 GENETIC FACTORS.................................................................................................... 23
CHAPTER THREE: METHODOLOGY .................................................................................28
3.1 INTRODUCTION.......................................................................................................... 28
3.2 STUDY DESIGN ........................................................................................................... 28
3.3 STUDY SITES ............................................................................................................... 30
3.4 STUDY POPULATION ................................................................................................ 30
3.6 INCLUSION AND EXCLUSION CRITERIA ............................................................. 32
3.6.1 Inclusion Criteria .................................................................................................... 32
3.6.2 Exclusion Criteria ................................................................................................... 32
3.7 EQUIPMENT/INSTRUMENT CAPABILTY .............................................................. 32
3.7.1 Ultrasound Equipment Suitability for Obstetric Work ........................................... 33
3.8 QUALITY ASURANCE TESTS ON THE EQUIPMENT/INSTRUMENT ................ 34
3.9 ASSESSMENT OF RELIABILITY OF THE MEASUREMENTS .............................. 35
3.10 REGRESION MODELS ................................................................................................ 35
3.10.1 Campbell and Wilkin Approach for EFW (AC) only ............................................. 35
University of Ghana http://ugspace.ug.edu.gh
vii
3.10.2 Hadlock Approaches for EFW Calculations: (AC, FL) only.................................. 35
3.10.3 Hadlock Approaches for EFW Calculations: (AC, BPD) Only.............................. 36
3.10.4 Hadlock Approaches for EFW Calculations: AC, FL and HC Only ...................... 36
3.10.5 Hadlock Approaches for EFW Calculations (AC, BPD, FL) ................................. 36
3.10.6 Hadlock Approaches for EFW Calculations (BPD, HC, AC, FL).......................... 36
3.10.7 Osaka Approach for EFW Calculations (BPD, FTA, FL) ...................................... 37
3.10.8 Tokyo Approach for EFW Calculations (BPD, AD, FL) ....................................... 37
3.10.9 Shephard Approach for EFW Calculations (AC, BPD).......................................... 37
3.10.10 Shinozuka Approach for EFW Calculations (AC, BPD, FL) ..................................38
3.10.11 Shinozuka Approach for EFW Calculations (APTD, TTD, BPD, FL) .................. 38
3.11 PROCEDURE FOR DATA COLLECTION ................................................................. 40
3.12 DATA MANAGEMENT ............................................................................................... 44
3.13 ETHICS .......................................................................................................................... 44
3.13.1 Standard Protocol Adopted for the Study ............................................................... 45
CHAPTER FOUR: RESULTS .....................................................................................................46
4.1 INTRODUCTION.......................................................................................................... 46
4.2 DEMOGRAPHICS ........................................................................................................ 46
4.3 FOETAL BIOMETRICS ............................................................................................... 48
4.4 HYPOTHESIS ............................................................................................................... 49
CHAPTER FIVE: DISCUSSION............................................................................................... 53
5.1 INTRODUCTION........................................................................................................... 53
5.2 RESPONSE RATE ......................................................................................................... 53
5.3 DEMOGRAPHICS ..........................................................................................................53
5.3 FOETAL BIOMETRICS ................................................................................................ 54
University of Ghana http://ugspace.ug.edu.gh
viii
CHAPTER SIX: CONCLUSIONS, RECOMMENDATIONS AND LIMITATIONS.......... 58
6.1 INTRODUCTION........................................................................................................... 58
6.2 CONCLUSIONS ............................................................................................................. 58
6.3 RECOMMENDATIONS ................................................................................................ 59
6.4 LIMITATIONS OF THE STUDY .................................................................................. 60
REFERENCES............................................................................................................................. 61
APPENDICES ............................................................................................................................... 79
APPENDIX I: PARTICIPANTS INFORMATION SHEET......................................................... 79
APPENDIX II: INFORMED CONSENT FORM ......................................................................... 80
APPENDIX III: DATA COLLECTION SHEET .......................................................................... 81
APPENDIX IV: ETHICAL CLEARANCE ...................................................................................82
University of Ghana http://ugspace.ug.edu.gh
ix
LIST OF TABLES
Table 3.1: Technical specifications of Philips ClearVue 350 ultrasound equipment…………...33
Table 3.2: Estimated foetal weight formulae used in the study………………………………....39
Table 4.1: Maternal Demographics……………………………………………………………...47
Table 4.2: Correlation coefficient between maternal demographics and neonates‟ ABWs…......47
Table 4.3: Mean and S.D. of BPD, HC, FL, and AC foetal biometrics………………………....48
Table 4.4: Birth weights at term………….……………………………………………………...48
Table 4.5: Comparison between EFW generated by the regression models to ABWs……….....50
University of Ghana http://ugspace.ug.edu.gh
x
LIST OF FIGURES
Figure3.1: BPD measurement………………………………………………………………….41
Figure 3.2: AC measurement…………………………………………………………………..42
Figure 3.3: HC measurement…………………………………………………………………..43
Figure 3.4: FL measurement…………………………………………………………………...43
Figure 4.1: Scatter plot of ABW against regression models…………………………………...49
Figure 4.2: Box and Whisker distribution of ABW and regression EFWs…………………….52
University of Ghana http://ugspace.ug.edu.gh
xi
LIST OF ABBREVIATIONS
BPD Biparietal diameter
HC Head circumference
AC Abdominal circumference
FL Femur length
AD Abdominal diameter
FTA Foetal trunk area
APTD Anterior posterior trunk diametre
TTD Trans-thoracic diametre
BW Birth weight
ABW Actual birth weight
MRI Magnetic resonance imaging
USG Ultrasonography
EFW Estimated foetal weight
IUEFW Intrauterine estimated foetal weight
OAUTHC Obafemi Awolowo University Teaching Hospital Complex
IQR Interquartile range
PPW Pre-pregnancy weight
University of Ghana http://ugspace.ug.edu.gh
xii
ABSTRACT
Background: Earlier investigations have demonstrated that the accuracy of foetal weight
estimation is significantly higher if several ultrasonic foetal parameters are measured, because
the total body mass depends on the size of foetal head, abdominal circumference and femur
length.
Aim: This study aimed to determine the best ultrasound regression models that use the common
foetal biometric measurements in predicting foetal weight for normal singleton pregnancies in
ClearVue 350 ultrasound equipment, and also assess the effects of few maternal anthropometric
variables on estimated foetal weight in the Greater Accra Metropolis.
Methodology: The study was quantitative, cross-sectional and quasi-experimental in nature and
was carried out prospectively in which convenient sampling method was used. Forty (40)
participants were scanned in the study. Foetal biparietal diameter, head circumference,
abdominal circumference and femur length measurements were measured according to the
established guidelines. Microsoft Excel and IBM SPSS soft-wares were used for data analysis.
Results: The study demonstrated that, all eight regression models showed no statistically
significant difference with (p-value <0.05) at 95% confidence interval of their mean foetal
weight estimates compared to the actual birth weight of the neonates. The mean and standard
deviations were similar across the eight models.
Conclusion: Findings from this study showed that, the best regressional model is Hadlock3
model. The study further revealed that, with 87% of estimated foetal weights which were within
5% of actual birth weights, demonstrate strong correlation between EFWs and ABWs, therefore
(radiologists, sonographers/radiographers)on one side and (gynaecologists, midwives) on the
other side can (to a high degree) rely on estimated foetal weights generated from Philips
University of Ghana http://ugspace.ug.edu.gh
xiii
CleaVue350 ultrasound equipment using the common foetal biometric measurements in a 2D
mode to generate meaningful reports and make important decisions such as mode of delivery
respectively.
University of Ghana http://ugspace.ug.edu.gh
1
CHAPTER ONE
INTRODUCTION
1.1 BACKGROUND Prenatal or antenatal development is the process in which a human embryo or foetus gestates
during pregnancy, from fertilization until birth (Shukla, 2012). The term embryo is used to
describe the developing offspring during the first 8 weeks following conception, and the term
foetus is used from about 2 months of development until birth Shukla (2012).
After fertilization the embryogenesis starts. In humans, when embryogenesis finishes, by the end
of the 10th week of gestational age, the precursors of all the major organs of the body have been
formed and the following period, which is the foetal period, is described both topically on one
hand that is by organ, and strictly chronologically on the other, by a list of major occurrences by
weeks of gestational age Shukla (2012). Rapid growth occurs and the embryo's main external
features begin to take form in a process called differentiation, which produces the varied cell
types (such as blood cells, kidney cells, and nerve cells). At 10th week, embryo measures 30–80
millimeters in length with facial features continuing to develop that is, eyelids are more
developed, head comprises nearly half of the foetus' size, the face is well formed, the limbs are
long and thin, the foetus can make a fist with its fingers and genitals appear well differentiated
(Shukla, 2012). It is these well-developed external features or specific parts of a foetus like the
limbs, head and trunk which are measured and are known as foetal biometry.
Growth rate of foetus is linear up to 37 weeks of gestation, after which it plateaus (Daftary and
Chakravarti 2011). The growth rate of an embryo and infant can be reflected as the weight per
gestational age, and is often given as the weight put in relation to what would be expected by the
University of Ghana http://ugspace.ug.edu.gh
2
gestational age. Neonate within the normal range of weight for that gestational age is known as
appropriate for gestational age (AGA). Those that have large or small range of weights are
known as large for gestational age (LGA) and small for gestational age (SGA) respectively.
According to literature (Oken et al., 2003; NahlaSubhi et al., 2010), birth weight (BW) is a
composite of foetal growth and length of gestation, each of which has different contributions and
different sequel. The absence of the contribution of gestational age to BW is the first step in
understanding the determinants of foetal growth. Sonographic estimated foetal weight (EFW) is
a foetal biometry parameter whose calculation is based on four common foetal biometric
measurements, that is, biparietal diameter (BPD), head circumference (HC), femur length (FL),
abdominal circumference (AC).
Inaccuracies with EFW have been reported in several literature. In particular, it has been reported
in the published literature (Colman et al., 2006; Scioscia et al., 2008) of the existence of
considerable differences between the sonographic EFW and the actual birth weight ABW with a
mean error of 7% to 10%, even under ideal sonographic conditions. However, despite its
inaccuracies EFW has been used in clinical decisions for over thirty years (Dudley, 2005; Ben-
Haroush et al., 2004). In the search to improve the ability to accurately predict BW from
sonographic EFW, multiple EFW formulae have been published, but limited data exist on their
comparative accuracies, and considerable variation in the EFWs occurs with different formulae
using the same foetal biometric measurements (Burd et al, 2009).
Again, according to Burd et al., (2009), the large number of published EFW formulae provides
clear evidence that none of them are universally accepted and unfortunately, the error of
University of Ghana http://ugspace.ug.edu.gh
3
sonographic EFW appears to be greatest at the two ends of the weight scale with most EFW
formulae, having low and high BWs are overestimated and underestimated respectively (Baker et
al., 1994).
During the last decade, EFW has been incorporated into the standard routine antepartum
evaluation of high-risk pregnancies and deliveries (Shittu et al., 2007). For instance, management
of diabetic pregnancy, vaginal birth after a previous caesarean section, and intrapartum
management of foetuses with breech presentations will be greatly influenced by EFW
(Sherman et al., 1998; Chauhan et al., 1998). Furthermore, in dealing with anticipated preterm
delivery, prenatal counseling on likelihood of survival, the intervention undertaken to postpone
preterm delivery, optimal route of delivery, or the level of the hospital where delivery should
occur may be based wholly or in part on the estimation of expected BW (Shittu et al., 2007).
Categorization of foetal weight into either small or large for gestational age may lead to timed
obstetric interventions that collectively represent significant departure from routine antenatal care
(Chauhan et al., 1998; Nzeh et al., 2000; Hanretty et al., 1990; Hendrix et al., 2000).
Previous studies have shown that the accuracy of EFW is significantly higher if several
ultrasonic foetal parameters are measured. This is because the total body mass depends on the
size of foetal head, abdominal circumference and femur length (Mladenović-Segedi et al., 2005).
In affirmation of this observation, Chien et al., (2000) validated 4 different formulae and
concluded that ultrasonic EFW at term with all the four formulae was high.
University of Ghana http://ugspace.ug.edu.gh
4
Ultrasound examination and measurements of foetal biometry have become integral to modern
obstetric care and thus provide avenues for dating pregnancies or for assessment of foetal growth
(Leung et al., 2008). The selection of appropriate cross-sectional reference charts is of great
importance to ensuring accurate diagnosis, although some published reference charts are
methodologically flawed (Leung et al., 2008). Some common associated problems include
repeated measurements on the same foetuses, formation of „super normal‟ datasets by
inappropriate exclusion of complicated pregnancies, failure to identify the statistical method of
analysis, and the use of statistical methods which in the opinion of Altman et al., (1994) do not
consider the variability of measurements with gestational age. Consequent to this, appropriate
methodologies have been published (Altman et al., 1994; Royston et al., 1998), and
mathematical equations and foetal biometric charts for various populations using the correct
approaches are now available in the medical literature (Chitty et al., 1994; Kurmanavicius et al.,
1999; Salomon et al., 2006).
1.2 PROBLEM STATEMENT
The influence of ethnicity or race on foetal biometry has been firmly established and widely
published in literature (Yeo et al., 1994; Jacquemyn et al., 2000). Leung et al., (2008) established
that, Hong Kong Chinese have similar reference chart as Singaporeans for all biometric
measurements but differs greatly in biometric measurement with those of UK and France with
respect to femur length (FL).
Several variables including ethnicity are incorporated in developing regression models for global
use in ultrasound foetal biometry. However, these plethora of regression models are mostly
University of Ghana http://ugspace.ug.edu.gh
5
based on Western and European populations which might not be representative of Ghanaian‟s,
and thus create opportunities for deviations and errors in ultrasound foetal biometry using Philips
ClearVue 350 ultrasound equipment.
Presently, there is no specific well documented evidence of regressional models of combined
biometric measurements for EFW in Ghana. Thus the use of these regressional models based on
Western and European populations on Ghana‟s obviously presents a challenge in EFW biometry.
In particular, the absence of knowledge of precise foetal weights presents additional problems of
unnecessary caesarean sections and its associated complications, limit women from their preferred
choice of number of children, and perturb critical interventions in cases of macrosomia and breech
presentation.
Furthermore, anecdotal report has shown that enormous complaints from midwives about
inconsistencies of gestational age with respect to EFW, huge gaps between approximate ABW
and EFW, and the prolonged duration of EFW given at a particular date and onset of labor of an
expectant mother have been reported in Mamprobi Polyclinic. These problems may have arisen
from ultrasound foetal biometry using regression models unrepresentative in Ghana or due to
operator inexperience.
Aside these challenges, the Ghana government through the health agencies (Ministry of Health
and Ghana Health Service) has begun nationwide programme of equipping all polyclinics and
district hospitals with Philips ClearVue 350 ultrasound equipment. It is known that the
regressional model used in this equipment does not provide for variations of race influences in
foetal biometrics. It is therefore imperative to determine regression model based on common
biometric measurements most suitable for the Ghanaian population and minimize the stated
problems above.
University of Ghana http://ugspace.ug.edu.gh
6
1.3 RESEARCH QUESTION/HYPOTHESES
1.3.1 Research Question: What is the best regression model(s) from Philips ClearVue 350
ultrasound equipment, using common biometric measurements in establishing EFW in utero in
Greater Accra Metropolis?
1.3.2 Hypothesis:
Null Hypothesis H 0: There will be no significant difference between EFWs and ABWs.
Alternative Hypothesis H1: There will be significant difference between EFWs and ABWs
1.4 SIGNIFICANCE OF STUDY
Generally, simple and accurate methods of estimating intrauterine foetal weight that can be
easily applied to all pregnancies are important means of reducing prenatal mortality and
morbidity. These are done through early detection of faltering growth and also in managing
breech presentation, macrosomia and intrauterine growth retardation at term by gyaenacologists
and midwives.
To buttress the point above, several investigations have shown that low EFW is associated with
high prenatal mortality and morbidity according to (Mavalankar et al., 1991). Therefore,
importantly, knowing the weight of a foetus and with other examinations like umbilical
Doppler and laboratory tests can help clinicians significantly reduce the occurrences of prenatal
morbidity, mortality and maternal deaths in some cases.
University of Ghana http://ugspace.ug.edu.gh
7
1.5 AIMS
The aim of this research study is to determine the best ultrasound regression model that uses the
common foetal biometric measurements in predicting foetal weight for normal singleton
pregnancies with ClearVue 350 ultrasound equipment, and also assess the effects of some few
maternal anthropometric variables on EFW in the Greater Accra Metropolis.
1.6 OBJECTIVES
To determine EFWs of foetuses.
To determine ABWs of neonates.
To compare EFWs to ABWs.
To obtain records of maternal anthropometrics (height and weight) from the time of
visiting antenatal for the first time and weight at the last antenatal visit.
University of Ghana http://ugspace.ug.edu.gh
8
CHAPTER TWO
LITERATURE REVIEW
2.1 INTRODUCTION Before and after the introduction of ultrasound into obstetric work, some forms of assessing
EFW were carried out globally. Earlier assessments were purely via physical examination of
maternal and foetus. After the advent of ultrasound equipment, others were performed
physically, in combination with ultrasonic measurements when EFW formulae were not
incorporated in ultrasound equipment.
2.2 HISTORICAL REVIEW OF ESTIMATING FOETAL WEIGHT
About four decades ago, Campbell and Wilkin assessed BW prediction of foetuses, using
abdominal circumference only, by ultrasonic means. This measurement showed that:
1. Accuracy of predictions varied with the size of the foetus.
2. The predicted BW and the ABW were not comparable.
For instance, at a predicted weight of 1 kg, 95% of BWs fell within 160 g, while at 2 kg, 3 kg
and 4 kg, the corresponding values were 290 g, 450 g and 590 g respectively. However,
according to Kearney et al., (1978), inclusion of abdominal circumference in determination of
EFW cannot be overemphasized as it presents a higher percentage value for predicting BW.
To date, a lot of mathematical formulae as noted in the literature have been deduced to narrow
the range of foetal weight estimation. Therefore, there is a strong evidence base of using more
University of Ghana http://ugspace.ug.edu.gh
9
than two biometric measurements to obtain accurate EFWs which will lead to the common
methods in estimating foetal weight sonographically.
2.3 TACTILE ASSESSMENT OF FOETAL SIZE
Dare et al., (1990) used the tactile assessment, which is the oldest technique for assessing foetal
weight. This is done via manual assessment of foetal size by obstetricians and midwives
worldwide, i.e. by external palpation of the uterus and foetal parts. This method is extensively
used because it is both convenient and virtually costless. However, it has long been known as a
subjective method that is associated with significant predictive errors (Shittu et al., 2007).
According Bossak et al., (1972), it is both patient- and clinician-dependent for its success (less
accurate for obese gravidas than non-obese and significant inter-observer variation in prediction
of BW even among experienced clinicians).
Maternal self-estimation of foetal weight in multiparous women shows comparable accuracy to
clinical palpation in some studies for predicting abnormally large foetuses (Chauhan et al., 1992;
Chauhan et al., 1998). According to Shittu et al. (2007), various calculations and formulae based
on measuring uterine fundal height above symphysis pubis have been developed. For example,
Ojwang et al., (1984) used the product of symphysio-fundal height and abdominal girth
measurement at various levels in centimeters above the symphysis pubis in obtaining a fairly
acceptable predictive value but with considerable variation from the mean. To further simplify
this method, Dare et al., (1990) in OAUTHC, Ile-Ife, used the product of symphysio-fundal
height and abdominal girth at the level of the umbilicus measured in centimeters and expressed
the result in grams to obtain the EFW at term in-utero. The EFW correlated well with BWs.
The Johnson's formula for EFW weight in vertex presentation defined as:
University of Ghana http://ugspace.ug.edu.gh
10
nFHgEFW 155)( (2.1)
Where FH= fundal height, and
spineischialabovevertex
spineischailbelowvertexn
,12
,11 (2.2)
If a patient weigh more than 91 kg, 1 cm is subtracted from the FH (Shittu et al, 2007).
2.4 FOETAL WEIGHT PREDICTIONS USING ALGORITHM DERIVED
FROM MATERNAL AND PREGNANCY-SPECIFIC CHARACTERISTICS
According to Shittu et al., (2007) again, a new theoretically-defensible equation capable of
predicting BWs prospectively from maternal characteristics was developed. To do this, the
efficacy of 59 scientifically-justifiable terms was evaluated simultaneously, obviating any
confounding co-variation and determining which of the predictions could account for variation in
BW that others could not. Aside maternal race, only six maternal and pregnancy-specific
variables were important in prediction of BW for otherwise normal gravidas. Using these
routinely-recorded variables, an equation based on maternal demographic and pregnancy-related
characteristics alone was developed to help predict BW as shown by Nahum (2003).
)1(81.4(000237.0262.036.9)( pWWHfdgBW Rmmmg (2.3)
Where d= gestational age, fg= fetal gender, Hm= maternal height, Wm= maternal weight measured at
26 weeks, WRm= maternal weight gain rate (Wm/d), p = parity. From the definition of variables,
equation 2.3 can be re-written as
)]1(81.4(000237.0262.036.9[)( pd
WmW mH mf g
dgBW (2.4)
University of Ghana http://ugspace.ug.edu.gh
11
male
unknown
female
f g
,1
,0
,1
(2.5)
As obesity is link to weight gain, Nahum et al., (1995) earlier demonstrated that the effect of
maternal obesity on foetal weight gain at term is small and not clinically significant. The study
also shows that, foetal weight gain was only 0.6g per day greater for obese pregnant women than
for those of normal weight for height and male foetuses gained weight 0.5g per day faster than
females. Therefore, it presupposes that the incorporation of maternal weight gain rate into
Nahum‟s equation is of no significance.
2.5 OBSTETRIC ULTRASONOGRAPHY IN ESTIMATING FOETAL WEGHT
The modern method for assessing foetal weight involves the use of foetal measurement obtained
via ultrasonography and has the advantage of relying on linear and/or planar measurements of in-
utero foetal dimensions that are definable objectively and should be reproducible (Shittu et al.,
2007). However, early expectations that this method might provide an objective standard for
identifying foetuses of abnormal size for gestational age was recently undermined by prospective
studies that showed sonographic estimates of foetal weight to be no better than clinical palpation
predictions based on algorithms using various combinations of foetal parameters, such as AC,
FL, BPD, and HC (Shittu et al., 2007). These common and popular models are classified into
groups depending on the combinations of foetal parameters and are used either singularly or in
combination as shown in literature by (Nahum, 2003; Hadlock et al., 1985; Nzeh et al., 2000;
Campbells, 1975; Comb et al., 1993; Ott et al., 1986 and Deter et al., 1985). The mathematical
equations referred to as Model Reference Equations (MRE) which underpins these algorithms for
calculating EFW are described by Nahum and Stanislaw, (2003) as shown below
University of Ghana http://ugspace.ug.edu.gh
12
2.5.1 Group 1: Abdominal Circumference Only
The unlisted group of equations provide for calculating EFWs using:
1. Campbell and Wilkin (in kg):
200331.0282.0564.4ln ACACEFW (2.5)
2. Hadlock et al.
200275.0)(253.0695.2ln ACACEFW (2.6)
3. Jordaan
32
10 )(000036239.0)(0043.0)(1881.06328.0log ACACACEFW (2.7)
4. Warsof et al.
3
10 )(000019.0)(092.08367.1log ACACEFW (2.8)
5. Higginbottom et al.
3)(0816.0 ACEFW (2.9)
2.5.2 Group 2: Abdominal Circumference and Femur Length
These groups of equations provide for calculation of the EFWs using AC and FL only:
6. Hadlock et al
))((004.0)(01938.0)(05281.0304.1log10 FLACFLACEFW
(2.10)
7. Woo et al.
))((00716.0)(28.0)(08.059.0log10 FLACFLACEFW (2.11)
8. Warsof et al.
))((0027.0)(0036.0)(108.0792.2ln 2 FLACACFLEFW (2.12)
University of Ghana http://ugspace.ug.edu.gh
13
2.5.3 Group 3: Abdominal Circumference and Biparietal Diameter
The EFW are estimated via these groups of equations using AC and BPD only:
9. Vintzileos et al.
)(026.0)(084.0879.1log10 ACBPDFEW (2.13)
10. Warsof et al.
2
10 ))((000111.0)(032.0)(144.0599.1log BPDACACBPDEFW (2.14)
11. Shepard et al.
))((002546.0)(046.0)(166.07492.1log10 BPDACACBPDEFW
(2.15)
12. Jordaan.
))((0015.0)(0950.0)(0377.01683.1log10 BPDACBPDACEFW (2.16)
13. Hadlock et al.
)(16994.0))((000595.0
)(007365.0)(000604.0)(05845.01134.1log 22
10
BPDBPDAC
BPDACACEFW (2.17)
14. Woo et al.
22
10 ))((00008599.0)(00111.0)(16.063.1log ACBPDACBPDEFW (2.18)
University of Ghana http://ugspace.ug.edu.gh
14
15. Hsieh et al.
)(052594.0)(000019782.0
))((00015515.0))((0056541.01315.2log
3
2
10
BPDAC
ACBPDBPDACEFW (2.19)
2.5.4 Group 4: Abdominal Circumference, Head Circumference (±Biparietal Diameter)
EFWs are derived from these groups of equations using AC, HC, and with or without BPD only:
16. Hadlock et al.
))((0002184.0
)(00063.0)(07057.0)(0273.0182.1log 2
10
HCAC
ACACHCEFW (2.20)
17. Jordaan.
))((001599.0)(0824.0)(0488.09119.0log10 HCACACHCEFW
(2.21)
18. Jordaan.
)(0058.0)(0079.0)(02904.03231.2log10 BPDHCACEFW (2.22)
2.5.4 Group 5: Abdominal Circumference, Femur Length, and Biparietal Diameter
Using AC, FL and BPD biometrics only, the EFW can be calculated via:
19. Hadlock et al.
)(1623.0
)(0457.0)(0316.0))((0034.0335.1log10
FL
ACBPDFLACEFW
(2.23)
20. Woo et al-7
University of Ghana http://ugspace.ug.edu.gh
15
))((000992.0
)(05.0))((0000764.0)(00111.0)(15.054.1log 22
10
FLAC
FLACBPDACBPDEFW (2.24)
21. Shinozuka et al.
32 )(6230.1))((23966.0 BPDACFLEFW (2.25)
22. Hsieh et al.
2
2
10
))((001745.0
))((00076742.0)(1432.0))((0094962.07193.2log
BPDFL
BPDACFLBPDACEFW (2.26)
2.5.6 Group 6: Abdominal Circumference, Femur Length, and Head Circumference
Using the AC, FL, and HC biometrics, the EFWs can be calculated via:
23. Hadlock et al.
)(158.0
)(0438.0)(0107.0))((00326.0326.1log10
FL
ACHCFLACEFW (2.27)
24. Combs et al.
32 )(03312.0))((23718.0 HCACFLEFW (2.28)
25. Ott et al.
)/(2594.1
))((0008582.0)(05394.0)(04355.00661.2log10
ACFL
HCACACHCEFW (2.29)
2.5.7 Group 7: Abdominal, Head Circumferences, Femur Length, Biparietal Diameter
The EFWs can also be calculated using AC, HC, FL and BPDs via:
26. Hadlock et al.
))((00386.0))((00061.0
)(174.0)(0424.0)(0064.03596.1log10
FLACBPDAC
FLACHCEFW (2.30)
University of Ghana http://ugspace.ug.edu.gh
16
In-utero foetal biometric assessment made by obstetric ultrasonography use best-fit algorithms to
make BW predictions (Benacerraf et al., 1988) and this provides an attractive „objective‟ method
of estimating BW. Willocks et al., (1964) were among the first to report their experience with
ultrasound foetal weight estimation, and commented that clinical estimation of foetal weight is
little more than guess work because of the influence of factors such as abdominal wall thickness,
uterine tension, volume of amniotic fluid, and position of the foetus in utero. It was claimed that
in two thirds of cases, foetal weight could be estimated to within about 2.2kg using ultrasound
(Tlale, 2012).
A large number of foetal weight prediction formulae have since been suggested based on foetal
biometry including head BPD, HC, FL, AC and abdominal diameters (AD) known as common
foetal biometry. In particular, a retrospective study has been done by Shamley and Landon
(1994), to evaluate four published equations by Sherpard et al., (1982), Hadlock et al., (1985),
Rose and McCallum (1987), and Sabbagha et al., (1989), as well as clinical estimation for
accuracy in determining foetal weight in labour. According to the study, Hadlock and Shepard
equations were associated with lower percentages of error 6.1% and 6.2% respectively than the
Sabbagha formula (7.8%; p<0.007). For all four equations, 70-79% of foetal weight predictions
were within 10% of ABW. The Shepard formula, however, has limited application in labour
because head descent obscures the BPD which is essential to the foetal weight calculation. The
study concluded that the use of any of the four standard equations or clinical examinations
provides accurate estimation of foetal weight for patients in labour, even in the presence of
ruptured membranes.
University of Ghana http://ugspace.ug.edu.gh
17
In these comparisons, the percentage of predictions within 10% of the BW for ultrasound
estimation ranged from 39% to 69% (Hendrix et al., 2000; Chauhan et al., 1992; Chauhan et al.,
1992; Barnhard et al., 1996; Sherman et al., 1998). While ultrasound foetal weight estimation may
not appear significantly better than clinical methods, a recent comparison by Peregrine et al.,
(2007) has shown that ultrasound estimations of foetal weight immediately before labour was
more accurate than clinical estimation for low and high BW babies, where clinical estimates are
known to be relatively inaccurate.
A study by Yoni et al., (1996) investigated the effect of oligohydramnios on intrapartum
estimation of foetal weight, and concluded that in term patients, intrapartum sonographic
prediction of BW in the presence of reduced amniotic fluid volume offered no advantage over
EFW obtained by abdominal palpation. However, the presence of oligohydramnios significantly
reduced the accuracy of intrapartum clinical and sonographic foetal weight estimations,
therefore, it was suggested that intrapartum foetal weight estimation be obtained prior to artificial
rupture of membranes (Almonem, 2015)
Although the validity and reproducibility of these formulae have been documented in clinical
practice with a reported systematic error of 10% or less relative to the ABW, it is well
recognized that various foetal factors may influence the accuracy of foetal weight estimations.
However, there are few reports that document the effect of certain maternal characteristics,
specifically maternal size and obesity, on the ability to obtain ultrasonographic foetal biometric
measurements and consequently calculate reliable foetal weights (Field et al., 1995)
University of Ghana http://ugspace.ug.edu.gh
18
Newer technologies may be able to provide better foetal weight estimation, for example, three-
dimensional (3-D) sonography potentially allows superior foetal weight estimation including soft
tissue volume of the foetal thigh, upper arm and abdomen (Schild et al., 2000; Bennini et al.,
2010). Its advantage over conventional two-dimensional (2-D) ultrasonography is that
reproducible circumference and volumetric measurements become feasible by simultaneous
visualization of three orthogonal foetal limb sections. However, the disadvantage is that 3-D
sonography is a more time consuming process, requires technically advanced and expensive
equipment, and special operator training and skills which may also be more difficult to apply
during labour (Tlale, 2012). According to Lindell and Marsal (2009), it seems unreasonable to
abandon the 2-D ultrasound imaging for foetal weight estimation.
2.6 MAGNETIC RESONANCE IMAGING IN ESTIMATING FOETAL WEIGHT
Magnetic resonance imaging (MRI) has recently been used for estimating foetal volume and
weight in diabetic and normal pregnancy using high-resolution MRI machine combined with
semi-automatic segmentation software (Shittu et al., 2007). Its use may be recommended for
clinical situation where accurate estimation is essential. Its strong disadvantage is that it is
expensive even where it is available (Uotila et al., 2000). However, the promising new
development in foetal weight estimation by echo-planer MRI cannot be overemphasized and its
advantages compared with other imaging techniques include the ability to obtain multi-planar
acquisitions and therefore theoretically improved resolution (Shamar et al., 2012)
Although it is feasible to calculate MRI foetal weight in sagittal, coronal, or axial planes, (the
recommended plane of imaging,) slice of thickness, and associated number of acquisitions that
most accurately determine term foetal weight had not yet been established or published at the
University of Ghana http://ugspace.ug.edu.gh
19
time when Hassibi et al. (2004) presented their findings. According to (Tlale, 2012) sought to
determine whether there are differences in foetal weight calculation based on plane of imaging or
thickness by comparing sagittal 5mm, 3mm, and axial 8 mm MRI acquisitions with term BW
and sonography foetal weight calculations. Their goal was to establish an optimal, practical
protocol for foetal MRI in the prediction of foetal weight in the term infant. Calculated weights
from a 90-sec single-shot fast spin- echo sequence MRI acquisition with 8mm thick slices in the
axial plane at term were found to be better than sonographic estimates (Tlale, 2012). The
problem with these methods, according to Loffler (1967) is that, addition to its expensiveness, its
impracticability for obtaining measurements during labor. However as indicated earlier that only
measurements of foetus such as BPD, HC, AC, FL, FTA, AD, APTD, and TTD had stood the
test of time in deducing EFW which is inexpensive. Despite the discrepancies between EFW and
ABW, these equation models based on foetal biometric measurements are extensively being used
global currently.
2.7 GENDER-SPECIFIC ESTIMATION OF FOETAL WEIGHT
It has been previously suggested and also published in the literature that differences exist in the
accuracy of sonographic weight estimation between male and female foetuses, and that these
differences may be the result of gender-specific intrauterine growth patterns (Pang et al., 2003;
Parker et al., 1984; Pineau et al., 2003; Fields et al., 2009; Schild et al., 2004; and Schwarzler et
al., 2004). These also include gender-related differences in body composition and percent of
body fat and in ratios among various biometric indices (Fields et al., 2009; Schwarzler et al.,
2004).
University of Ghana http://ugspace.ug.edu.gh
20
Indeed, Melamed et al., (2011) have recently reported that sonographic weight estimation was
consistently more accurate for male than female foetuses, irrespective of the model used. It may
be reasonable to hypothesize that the use of two distinct gender-specific models, optimized for
male and female foetuses, may overcome this limitation. However, the impact of such gender-
specific models on the accuracy of foetal weight estimation is as yet unclear. Surprisingly, none
of the widely accepted sonographic models for foetal weight estimation (Shepard et al.,1982;
Hadlock et al., 1984; Hadlock et al., 1985; Warsof et al., 1986; Ott et al., 1986; and Combs et al.,
1993) includes foetal gender in the equation, and the results of only one study indicated that the
use of such gender-specific model may result in more accurate estimation than several widely
used models (Schild et al., 2004; Siemer et al.,2008a; Siemer et al., 2008b).
Moreover, it is unknown whether such gender-related model optimization merely reflects a
different set of gender-specific model coefficients or whether these gender-specific models also
differ in the combination of biometric indices incorporated into the model (Melamed et al.,
2012). However, the quest for simple and understandable approach of EFW deduction, other
formulae have emerged recently. For instance, it has been established that gender-specificity has
an effect on EFW, that male foetuses have more weight than their female counterparts for the
same gestational age. Hence, as proposed by Schild et al., (2004), different models should be
applied to different groups of foetuses depending on gender.
To augment Schild‟s work, Siemer et al., (2008) demonstrated that Schild‟s gender-specific
formula had the highest degree of accuracy than those equations widely use in the weight range
between 2,500 and 3,999 g where majority of BWs falls but fare poorly at the extremes with BW
< 2500g and >3999g. In improving on Schild‟s gender-specific model, Melamed et al., (2012)
adjusted Schild‟s formula and hard more accurate result. However, these gender-specific models
University of Ghana http://ugspace.ug.edu.gh
21
were associated with highest accuracy (lowest systematic error) only in the larger interquartile
BW subgroup, while this advantage was not clearly observed in the extremes of BW (below the
first quartile and above the third quartile) as was observed in many studies (Melamed et al.,
2012).
2.8 ETHNIC AND RACIAL DIFFERENCES IN ESTIMATED FOETAL WEIGHT
The effects or impact of racial variations and ethnicity on EFW have been questioned by Richard
et al., (1997) in a study on differing BWs among infants of US-born Blacks, African-born
Blacks, and US-born Whites. In this research, during the past 40 years, epidemiologic research
has elucidated many important associations between the socio-demographic characteristics of
mothers and the BW of infants. For example, the extremes of childbearing age (Kleinman &
Kessel, 1987), cigarette smoking (Fox et al., 1994), inadequate prenatal care (Murray &
Bernfield., 1988), urban poverty (Collins & David, 1990) and black race (David and Collins.,
1991) are well-documented risk factors for low BW. Other obstetrical risk factors account for
part of the racial disparity in BW, but differences persist (Lieberman et al., 1987; Rawlings et
al., 1995; Klebanoff et al., 1989; Sheehan & Gregorio, 1995). Although the incidence of low
BW decreases in both blacks and whites as the number of risk factors declines, the improvement
is faster among whites, resulting in a wider BW gap between blacks and whites among infants of
low-risk women (Kleinman & Kessel, 1987; Collins & David, 1990). This led some investigators
to believe that genetic factors associated with race influence BW. (Naylor & Myrianthopoulos
1967; Little & Sing 1987; Magnus 1984; Hulsey et al., 1991; Goldenberg et al., 1991;
Wildschutt et al., 1991).
University of Ghana http://ugspace.ug.edu.gh
22
In the National Collaborative Prenatal Project, only 1% of the total variance in BW among
18,000 infants was accounted for by socioeconomic variables, leading the authors to conclude
that “race behaves as a real biological variable in its effect on BW and this effect of race is
presumably genetic” (Naylor & Myrianthopoulos, 1967). The assumption that black women
differ genetically from white women in their ability to bear normal or large infants persists in
more recent studies of foetal growth (Hulsey et al., 1991; Amini et al., 1994) one of which, for
example, refers to “genetic factors affecting growth, such as neonatal gender and race” (Amini et
al., 1994). Few data have been published on the BWs of infants born to African-born women in
the United States. Most African Americans trace their origins to Western Africa, where the slave
trade flourished in the 17th and 18th centuries (Oliver & Fage, 1988; Reed 1969). It is estimated
that U.S. blacks derive about three quarters of their genetic heritage from West African ancestors
and the remainder from Europeans (Reed, 1969; Chakraborty, 1991; Adams & Ward, 1973;
Glass, 1953). To the extent that population differences in allele frequency underlie the observed
differences in BW between blacks and whites in the United States, one would expect women of
“pure” West African origin to bear smaller infants than comparable African Americans,
considering the European genetic admixture in the latter (Richard et al., 1997). However, this
assumption appear not to be true according to (Richard et al., 1997).
In summary, African-born black women have infants with a greater mean BW and a different
birth-weight distribution than U.S. - born black women born (Richard et al., 1997). Therefore,
from the literature, there is clear evidence of the effect of race and ethnicity of a particular
population on EFW which clearly suggests that, research must be conducted to find out and
ensure that, the most accurate regression model is used for a particular population in question.
University of Ghana http://ugspace.ug.edu.gh
23
2.9 GENETIC FACTORS Interest in determining the relative contributions of factors that produce BW variation, namely
the maternal and foetal genetic factors and the environment of the foetus have been expressed.
Approximately 40% of total BW variation is due to the genetic contributions from mother and
foetus (approximately half from each), and the other 60% is due to contributions from the foetal
environment (Polani, 1974).
Although both parents' genes affect childhood growth and final adult size, the maternal genes
have the main influence on BW as the classic horse-pony cross-breeding experiments by Walton
and Hammond (1938) demonstrated the important role of the mother that foals of the maternal
horse and paternal pony are significantly larger than foals of the maternal pony and paternal
horse, and foals of each cross are comparable in size to foals of the pure maternal breed which
clearly demonstrated the widely held study of a maternally related constraint on foetal growth.
Similar conclusions of maternal constraint to growth are reached from family studies in humans.
Low and high BWs recur in families with seemingly otherwise normal pregnancies. Sisters of
women with intrauterine growth restriction (IUGR) babies tend to have IUGR babies, a trend
that is not seen in their brothers' babies (Johnstone & Inglis, 1974). Mothers of IUGR infants
were frequently growth restricted at birth (Ounsted & Ounsted, 1966; Simpson et al., 1975).
There is also a greater similarity in BW between maternal half siblings and full siblings than
between paternal half siblings and full siblings Robert and Robert (2012). Although the maternal
phenotypic expression particularly maternal height may affect foetal growth, the evidence for
such an influence is not convincing Robert and Robert (2012). Social deprivation has also been
University of Ghana http://ugspace.ug.edu.gh
24
associated with IUGR, a finding not explained by known physiologic or pathological factors
(Wilcox et al., 199).
The one definite paternal influence on foetal growth and size at birth is the contribution of a Y
chromosome rather than an X chromosome (Robert and Robert, 2012). The male foetus grows
more quickly than the female foetus and weighs approximately 150g to 200g more than the
female at birth (Thomson et al., 1968). There is also a suggestion that paternal size at birth can
influence foetal growth, with BWs potentially increased by 100g to 175g (Klebanoff et al.,
1998). Furthermore, the greater the antigenic dissimilarity between the parents, the larger the
foetus. Whether it is genetically determined or not, women who were small for gestational age
(SGA) at birth have double risk of reduced intrauterine growth in their foetuses (Klebanoff et al.,
1997). In similar fashion, foetuses destined to deliver preterm have a higher incidence of reduced
foetal growth (Bukowski et al., 2001). The role of the genetic constitution of mother or foetus in
these observations is not clear.
Specific maternal genotypic disorders can cause IUGR. An example being phenylketonuria
(Saugstad, 1972). Infants born to homozygously affected mothers almost always have IUGR, but
whether the reason is an abnormal amount of metabolite crossing from mother to foetus or an
inherent problem in the foetus is unknown Robert and Robert (2012). There is a significant
association between IUGR and congenital malformations. Such abnormalities can be caused by
established chromosomal disorders or by dysmorphic syndromes, such as various forms of
dwarfism Robert and Robert (2012). Some of these malformations are the expression of a
specific gene abnormality with a known inheritance pattern, whereas others are only presumed to
be the result of a gene mutation or an adverse environmental influence Robert and Robert (2012).
University of Ghana http://ugspace.ug.edu.gh
25
Although in some reports only 2% to 5% of IUGR infants have a chromosomal abnormality, the
incidence rises to 20% if IUGR and mental retardation are both present (Snijders et al., 1993).
Birth weights in infants with trisomy 13, 18, and 21 are lower than normal (Chen et al., 1972,
Peuschel et al., 1976). The decrease in BW is less pronounced in trisomy 21. The frequency
distribution of BWs in infants with trisomy 21 is shifted to the left of the normal curve after 34
weeks of gestation, resulting in gestational ages 1 to 1.5 weeks less than normal, and BWs and
lengths are less than in control infants from 34 weeks until term (Creasy et al., 2013). This effect
is more marked after 37 weeks of gestation, but BWs are still only approximately 1 standard
deviation from mean weight. BWs in translocation trisomy 21 are comparable to those in
primary trisomy 21. Birth weights of newborns that are mosaic for normal and 21-trisomic cells
are lower than normal but higher than those of21-trisomic infants (Polani, 1974).
Newborns with other autosomal abnormalities, such as deletions (chromosomes 4, 5, 13, and 18)
and ring chromosome structure alterations, also have had impaired foetal growth. Although
abnormalities of the female (X) and male (Y) sex chromosomes are frequently lethal (80% to
95% result in first trimester spontaneous abortions), they could be a cause of IUGR in a newborn
(Kramer, 1987). Infants with XO sex chromosomes have a lower mean BW than control infants
(approximately 85% of normal for gestational age) and are approximately 1.5 cm shorter at birth.
Mosaics of 45 X and 46, XX cells are affected to a lesser degree (Kramer, 1987). Although a
paucity of reports prevents definite conclusions, it appears that the repressive effect on foetal
growth is increased with the addition of X chromosomes, each of which results in a 200- to 300-
g reduction in BW (Kramer, 1987). In general, IUGR is associated with numerous other
dysmorphic syndromes, particularly those causing abnormal brain development. The overall
contribution that chromosomal and other genetic disorders make to human IUGR is estimated to
University of Ghana http://ugspace.ug.edu.gh
26
be 5% to 20% and approximately 25% of foetuses with early-onset foetal growth restriction
could have chromosomal abnormalities, and karyotyping via cordocentesis can be considered
Robert and Robert (2012). A genetic basis should be considered strongly if IUGR is encountered
in association with neurologic impairment or early polyhydramnios.
The two main methods for predicting BW in current sonographic obstetrics are:
(a) clinical techniques based on abdominal palpation of foetal parts, calculations based on
fundal height in combination with some foetal biometric measurements and
(b) sonographic measures of skeletal foetal parts only which are then inserted into regression
equations to derive estimated foetal weight (Hanretty et al., 1990; Hendrix et al., 2000;
Raman et al., 1992).
Although some investigators consider sonographic estimates to be superior to clinical estimates,
others, in comparing both the techniques concurrently, conclude that they confer similar levels of
accuracy (Nzeh et al., 2000; Hanretta et al., 1990; Hendrix et al., 2000; Raman et al., 1992;
Watson et al., 1988; Mehdizadeh et al., 2000; Wilcox et al., 1992;Cecattiet al., 2000; Johar et
al.,1988;Hulsey et al., 1991; Richards et al., 2001;Hadlocket al., 1985, Ebomoyi et al., 1991,
Ojwang et al., 1984).
The available techniques for imaging can be broadly classified under prediction of equations of
BW by ultrasonography (USG) and MRI (Shittu et al., 2007). For imaging purposes, ultrasound
EFW is commonly used worldwide and less expensive. In addition, acquisition of ultrasound
equipment is far affordable compare to other imaging modality such as magnetic resonance
imaging MRI, which is highly expensive. Thus, the aim of the present study was to determine
University of Ghana http://ugspace.ug.edu.gh
27
which of the commonly used sonographic model provide better accuracy or precision of foetal
weight estimation, and to provide a better understanding of the reasons why that/those formulae
should be used in deducing foetal weight estimation.
University of Ghana http://ugspace.ug.edu.gh
28
CHAPTER THREE
METHODOLOGY
3.1 INTRODUCTION Research methods are approaches or techniques employed in conducting research operations
using scientific and systematic ways in answering a research question. This section therefore
presents the methodology employed in conducting this study and describes key aspects such as
the study design, study population and study sites, and experimental procedure.
3.2 STUDY DESIGN
A quantitative study which is cross-sectional and quasi-experimental in nature was
adopted for this study. Numeric figures were produced and analyzed for the study to become
conclusive.
For a study to be quantitative, Babbie (2010) enumerated that, quantitative methods emphasize
objective measurements and the statistical, mathematical, or numerical analysis of data collected
through polls, questionnaires, and surveys, or by manipulating pre-existing statistical data
using computational techniques. Furthermore, quantitative research focuses on gathering
numerical data and generalizing it across groups of people or to explain a particular phenomenon
(Muijs, 2010). This study adopted a research design that required a quantitative documentation
of foetal biometric measurements and neonatal weights.
Quantitative studies are conducted to determine the relationship between an
independent variable and a dependent or outcome variable within a population associated with
an experiment (subjects measured before and after delivery) (Babbie, 2010). As descriptive study
University of Ghana http://ugspace.ug.edu.gh
29
establishes only associations between variables, an experimental study establishes causality
(Babbie, 2010). Therefore, the study ultrasonically measured foetal biometry, compute EFWs
and correlate with the actual BWs.
Quantitative research deals in numbers, logic, and an objective stance and focuses on numeric
and unchanging data and detailed, convergent reasoning rather than divergent reasoning (Brains,
2011) i.e., the generation of a variety of ideas about a research problem in a spontaneous, free-
flowing manner (Brians, 2011). Therefore, this study accordingly, compares EFWs to ABWs in
line with the aim and objectives of this study.
As much as the study was quantitative, with assumption that there was not going be differences
in experimentation and the outcome values, it was cross sectional in nature as well since all the
experimental work was performed on a certain group of people during a particular period and
was carried out prospectively in which convenient sampling method was applied. Convenient
sampling method was used based on the fact that participants in this study must be in their latent
phase during labor which is associated with a lot of pains and discomfort resulting in
participants‟ unwillingness to partake in the study.
Ultrasound scans for EFW were scheduled within thirty minutes on arrival to the delivery ward
after medical histories such as previous antenatal ultrasound and laboratory particulars i.e.
antenatal folders which include their entire previous ultrasound scan and laboratory reports have
been received. The period between the time of arrival and the scheduled time for scanning
became very important due to empirical evidence that, almost 95% of expectant mothers arrived
at the labour-ward in the late latent phase of labour.
University of Ghana http://ugspace.ug.edu.gh
30
Foetal biometrics such as BPD, HC, AC and FL measurements were taken according to the
established guidelines (Loughna et al., 2009). All examinations were performed trans-
abdominally using Philips ClearVue 350 equipment, one day prior to vaginal or normal delivery
and within a maximum of one day prior to induction of labour. Birth weights were measured
using properly calibrated Salter scale (model 914) digital weighing scale of maximum range of
20kg.
3.3 STUDY SITES
The research was conducted at the Mamprobi Polyclinic (M-PC) in the Greater Accra
Region of Ghana. This study site was selected based on the following factors:
1. Out-patients records suggested that patients from all the geo-regional groupings in Ghana
residing in Accra attend this polyclinic for antenatal and delivery services.
2. M-PC has a large antenatal and delivery base.
3. M-PC is equipped with the new Philips ClearVue 350 ultrasound equipment.
3.4 STUDY POPULATION
The study included all pregnant women laboring with singleton pregnancies and had been
admitted for birth deliveries at the study site within the stipulated period of four months of data
collection. An anecdotal report from the study site showed one thousand two hundred and sixty-
seven, (1,267) deliveries per year. Therefore, during the four months of data collection period, an
estimated four hundred and twenty-two, (422) expectant mothers went ultrasound examination
for the purpose of this study and biometric measurements, BPD, HC, AC and FL of foetuses
were taken.
University of Ghana http://ugspace.ug.edu.gh
31
3.5 SAMPLE SIZE
Knowing the population size and using finite population correction factor, a sample size n was
determined via the equation according to courses.wcupa.edu;
)1(Nn
Nnn
o
o (3.1)
Where no = sample size without considering the finite population correction factor
N Population size
Assuming 442N and 100on then Equation (3.1) yields a sample size of
78)1442(100
)442)(100(
)1(Nn
Nnn
o
o (3.2)
Therefore, the actual sample size from finite population size of four hundred and forty-two (442)
was seventy-eighty, (78) participants.
At the study site, estimated numbers of seventy-eight 78 expectant mothers were expected to
undergo the scanning examination during the data collection period. Participants were selected
according to the inclusion and exclusion criteria, taking into account patient‟s right not to partake
in the research, the pain and discomfort level depending on the labour phase (active or latent)
which made few participants to decline in participation and the time duration of completion of
the study.
Based on the reasons given above, seventy-five, 75 participants took part in the study. However
due to, late referrals of some of the participants to tertiary facilities on account of prolong latent
phase of labor, indication of macrosomia and breech presentations, a sample size of 40
participants was arrived at for the study.
University of Ghana http://ugspace.ug.edu.gh
32
3.6 INCLUSION AND EXCLUSION CRITERIA
3.6.1 Inclusion Criteria
Participants were selected taking into account the following:
Expectant mothers with viable singleton pregnancies with cephalic presentations and
admitted for birth delivery at the study sites
Expectant patient‟s presenting with pain and discomfort level on the labour phase (active
or latent).
3.6.2 Exclusion Criteria
Patients presenting the condition below were excluded from the study:
Expectant mothers with polyhydramnios, pre-term labour, ruptured membranes,
abnormal lie and presentation, multiple pregnancies, antepartum haemorrhage,
eclampsia, obvious congenital abnormalities, oligohydramnios, anteriorly-inserted
placenta, and poor visualization of foetal part
Expectant mothers with no contraction and amniotic rapture was excluded from the
study.
3.7 EQUIPMENT/INSTRUMENT CAPABILTY
A Salter scale (model 914) digital weighing scale of maximum range of 20kg and reading up to 6
decimal places was used for the weight measurements of the neonates. Using dry cells as source
of electrical power, makes Salter scale very light in weight and can be moved around with less
effort.
A Philips ClearVue 350 ultrasound imaging equipment was utilized for the study. The technical
specifications of the equipment are listed in Table 3.1.
University of Ghana http://ugspace.ug.edu.gh
33
Table 3.1: Technical specifications of Philips ClearVue 350 ultrasound equipment
Parameters Screen width Control panel height Monitor width Monitor height Depth
Dimensions 51.8cm 83.4cm 43.2cm 138.4cm 58.6cm
Probe: C5-2: Abdominal, obstetrics/gynaecological, pediatric, vascular, and urological
curvilinear transducer
Gray scales: 255 in 2-D, M-mode, and Doppler
Components Scale lines: Up to 1,024 scan lines, transducer and mode dependent
Input signals: Low-level ECG, Four-port transducer receptacles
Electrical parameters: AC 100-240V ±10%
The advantages of this equipment over others in terms obstetric work is based on the following
factors which have been incorporated into the operational function of this equipment.
3.7.1 Ultrasound Equipment Suitability for Obstetric Work
The suitability of the ultrasound equipment for obstetric work in this study was established by
verification of these factors:
1.Amniotic Fluid Index measurement: This factor provides for calculating the amniotic
fluid in each quadrant (Q1+Q2+Q3+Q4) to give the total amniotic fluid index (AFI) value
which is immediately compared to an already provided reference range between 8.1 to
18.0cm (Rutherford et al., 1987).
2. Biophysical profile: An important factor that defines foetal wellbeing and expressed as
Biophysical profile = movement + tone + breathing + amniotic fluid index volume
University of Ghana http://ugspace.ug.edu.gh
34
It ranges from 0 to 2 or „Not Applicable‟ which indicates that the category will not
Contribute to the biophysical profile total (Manning et al., 1990).
3. Selecting of number of foetus in twin, triplets and quadruplets gestations to produce
separate reports on each of the foetuses. In selecting the number of foetuses, typically, the
foetus closest to the cervix is labeled „A‟ and the others are designated sequentially as they
are seen moving up to the fundus as B, C, and D.
4. Colour and power Doppler incorporation. This is done for easy study of arterial and
venous systems of the foetus especial in the foetal brain, umbilical cord and renal vessels.
5. B and M-modes: These mode combinations are used in displaying foetal heart movements
leading to estimation of foetal heart beat and the gradient of the wave produced.
These obstetrics properties of the ultrasound equipment in addition to others like
electrocardiogram, small parts and vascular imaging, and 3-D functions coupled with very clear
quality images produced on a large plasma screen, makes the equipment highly functional.
3.8 QUALITY ASURANCE TESTS ON THE EQUIPMENT/INSTRUMENT
Quality assurance and control was carried out before this study started. The ultrasound
equipment was upgraded by the engineers at the start of the study as biennial routine
maintenance contract to make sure the equipment perform at its maximum operational capacity.
After the upgrading of equipment, an operator manual was used to set its functions to their
optimal level to produce superior image quality for accurate and precise measurements to be
taken.
The Scaler scale was each time set to the zero mark before each neonate was weighed and the
dry cell use to power it was changed at start of work every day.
University of Ghana http://ugspace.ug.edu.gh
35
Additionally, a wheel chair was provided to transfer patients from the ward to the scanning room
due to the distance between these two destinations and patients‟ unstable condition at this period.
Wooden stair was provided for smooth climbing to the scanning bed and very long screen was
used to partition the room to provide privacy to the patients.
3.9 ASSESSMENT OF RELIABILITY OF THE MEASUREMENTS
Before the commencement of the study, pilot survey was done to test the reliability of
measurements on the sonographic images. Twenty patients were used in the pilot assessment of
accurate and precise measuring of foetal biometry. The result shows 98% accuracy after
scanning each participant three times and finds their arithmetic means.
The participants did not pay for participating in both the piloting and main research work.
3.10 REGRESION MODELS
As mentioned in Chapter 2, several regression models can be applied to the calculation of EFW
using ultrasonography. The regression models adopted in this study are presented here.
3.10.1 Campbell and Wilkin Approach for EFW (AC) only
The Campbell and Wilkin (1975) regression model for computing AC only is expressed as
200331.0282.0564.4)(ln ACACkgEFW (3.1)
3.10.2 Hadlock Approaches for EFW Calculations: (AC, FL) only
The Hadlock et al., (1985) approach can be applied for computing AC and FL fetal biometrics via
))((004.0)(1938.0)(05281.0304.1log10 FLACFLACEFW (3.2)
University of Ghana http://ugspace.ug.edu.gh
36
In particular, the system determines 2 standard deviation ranges based on EFW via the expression
)*154.0( EFWEFW (3.3)
3.10.3 Hadlock Approaches for EFW Calculations: (AC, BPD) Only
In respect of EFW calculations using AC and BPD fetal biometrics only, the Hadlock regressions is
represented as
))(1694.0)(000595.0
)(007365.0)(000604.0)(05845.011.1(log 22
10
BPDACBPD
BPDACACEFW (3.4)
With a corresponding standard deviation ranges based on EFW as
)*182.0( EFWEFW (3.5)
3.10.4 Hadlock Approaches for EFW Calculations: AC, FL and HC Only
Based on AC, FL, and HC variables, the Hadlock regression for EFW calculations is expressed as
))(158.0)(0438.0)(0107.0))((00326.0326.1(log10 FLACHCFLACEFW (3.6)
The system determines two standard deviation ranges based on EFW for this calculation as
follows:
)*148.0( EFWEFW (3.7)
3.10.5 Hadlock Approaches for EFW Calculations (AC, BPD, FL)
FLACBPDFLACgEFW 1623.00457.00316.0034.0335.1log)( 10 (3.8)
The EFW formula via Hadlock et al., (1985) using AC (cm), BPD (cm), and FL (cm) is: The
system determines two standard deviation ranges based on EFW for this calculation as follows:
)146.0( EFWEFW (3.9)
3.10.6 Hadlock Approaches for EFW Calculations (BPD, HC, AC, FL)
University of Ghana http://ugspace.ug.edu.gh
37
The formula for the EFW via Hadlock et al., (1985) using AC (cm), BPD (cm), FL (cm), and HC
(cm) is shown below. Normal ranges are broken up by EFW as percent of EFW and a gram offset.
)(174.0)(0424.0
))((00061.0)(0064.0))((00386.03596.1log10
FLAC
ACBPDHCFLACEFW
(3.10)
3.10.7 Osaka Approach for EFW Calculations (BPD, FTA, FL)
The EFW via Osaka University (Nobuaki et al., 1990) using BPD (range: 3.1 t0 10.0cm), fetal
trunk abdominal area (range: 20 to 180.0cm2), and FL (range: 1.0 to 8.0cm) is:
30994.6))((50665.3)(25647.1)( FLFTABPDgEFW (3.11)
3.10.8 Tokyo Approach for EFW Calculations (BPD, AD, FL)
The EFW formula via Tokyo University (Mizuno et al., 1989), using BPD (range: 3.1 to 10.0cm),
AD (AP, range: 5.0 to 15.0cm), AD (transverse: range: 5.0 to 15.0cm), and FL (range: 1.0 to
8.0cm) is:
FLADtrvADapBPDgEFW 42.3)(07.1)( 3 (3.12)
Where ap and trv are antero-posterior and transverse.
3.10.9 ShephardApproach for EFW Calculations (AC, BPD)
The Shephard (1982) formula for EFW using AC, (range: 15.0 to 40.0cm), and BPD (range: 3.1 to
10.0cm) is:
)002646.0046.0166.07492.1(101000)( BPDACACBPDgEFW (3.13)
The system determines one standard deviation range based on EFW for this calculation according
to the following:
ggEFW
ggEFW
ggEFW
566500,3
405500,3500,2
218500,2
(3.14)
University of Ghana http://ugspace.ug.edu.gh
38
3.10.10 Shinozuka Approach for EFW Calculations (AC, BPD, FL)
The formula for EFW via (Shinozuka et al., (1996), using AC (cm) (range: none), BPD (cm)
(range: none), and FL (cm) (range: none) is:
)30.0(07.1()( 23 FLACBPDgEFW (3.15)
3.10.11 Shinozuka Approach for EFW Calculations (APTD, TTD, BPD, FL)
The Shinouzuka (1989) formula for EFW using AP trunk diameter (cm) (rang: none), transverse
trunk diameter (cm) (range: none), BPD (cm) (range: none), and FL (cm) (range: none) is:
FLTTDAPTDBPDgEFW 42.3)07.1()( 3 (3.16)
Therefore from the immediate groups of equations above, the equations used in this study are
presented in Table 3.2.
University of Ghana http://ugspace.ug.edu.gh
39
Table 3.2: Estimated fetal weight formulae used in the study
Study Year Formulae
Campbell
1975
200331.0282.0564.4)(ln ACACkgEFW
Hadlock1
1984
))((004.0)(1938.0)(05281.0304.1log10 FLACFLACEFW
Hadlock6⃰
1985
)(1694.0)(000595.0
)(007365.0)(000604.0)(05845.011.1log
22
10BPDACBPD
BPDACACEFW
Hadlock3
1985
)(158.0
)(0438.0)(017.0)(00326.0326.1log10
FL
ACHCFLACEFW
Hadlock7
1985
FL
ACBPDFLACgEFW
1623.0
0457.00316.0034.0335.1log)( 10
Hadlock4
1985 )(174.0)(0424.0
))((00061.0)(0064.0))((00386.03596.1log10
FLAC
ACBPDHCFLACEFW
Shephard 1982 )002646.0046.0166.07492.1(101000)( ACBPDACBPDgEFW
Shinozuka2 1989 )30.0()07.1()( 23 FLACBPDgEFW
University of Ghana http://ugspace.ug.edu.gh
40
3.11 PROCEDURE FOR DATA COLLECTION
Expectant mother made to lie on the scanning bed in a supine position with the abdomen
exposed, a coupling gel is applied on it and a probe used to search for foetal parts like head,
abdomen and femur length and various measurements were made.
The BPD measurements were done for each foetus by obtaining a longitudinal section of the
foetus using small sliding movements of the transducer on each side of the foetal spine and
tracing the head along the spine. On locating the head, the transducer was moved to the vertex. In
this manner, the probe was slided to the lateral side of the foetal head facing anteriorly which at
this point looks ovoid in shape.
At this position, the probe was adjusted until strong midline echo with the cavum-septum
pellucida is demonstrated. When the section was not the required ovoid shape, minor rotational
adjustments were made until the shape was correct. Slight up and down movements of the
transducer on the maternal abdomen located the correct section, which was then frozen. The
BPD was measured by placing the horizontal component of the on-screen calipers on the outer
aspects of the echoes from the foetal skull. This was done at right angles to the midline and at
widest diameter of the ovoid head shape (Figure 3.1).
University of Ghana http://ugspace.ug.edu.gh
41
Figure 3.1: BPD measurement Source: Field data (2014)
Foetal AC was measured by scanning the foetus in a longitudinal plane until the foetal aorta was
visualized in its course, through the foetal chest and abdomen. The probe was rotated through
90° at the level of the lower end of the foetal ribs, with the rips showing posterior acoustic
shadowing. A minor adjustment was then made by moving the transducer up or down the foetal
body through maternal abdomen until the section was obtained. The section appeared almost a
perfect circle, containing only a short section of the umbilical vein in the anterior third of the
foetal abdomen.
University of Ghana http://ugspace.ug.edu.gh
42
Figure 3.2: AC measurement Source: Field data (2014)
For HC measurements, ellipsoidal tracing was made by drawing a line at the outer perimeter
around the exact image used in measuring BPD (Figure 3.3). The FL measurement was obtained
by scanning the foetus in longitudinal plane and the spine traced to its tapering end. The probe
was then rotated through small angles to locate a single long bone lying almost horizontal on the
monitor, producing a clean posterior acoustic shadow (Figure 3.4).
University of Ghana http://ugspace.ug.edu.gh
43
Figure 3.3: HC measurement Source: Field data (2014)
Figure 3.4: FL measurement Source: Field data (2014)
University of Ghana http://ugspace.ug.edu.gh
44
Maternal height and weights were obtained from participants antenatal records which are
retrospective in nature. The participants first visit weight is referred to as initial weight and the
last weight before onset of labour is final weight.
3.12 DATA MANAGEMENT
All procedures were carried out with strict adherence to standard protocols. Cross checking of
data entries were made to ensure accurate data capture for accurate statistical analysis adhered to.
Participants were given codes for identification. Data collecting sheets were store in safe cabinet
and electronic data is stored in a password protected database. The outcome of the study will be
disseminated to the Department of Radiography, School of Biomedical and Allied Health
Science, College of Health Sciences, University of Ghana. Articles of this study will be
published in Scientific Journals. The study reports shall be presented in various seminars and
conferences.
3.13 STATISTICAL ANALYSIS
The data was entered using Microsoft Excel software package and was exported onto SPATA
software package for all data analysis. Linear regression and Box and Whisker were used to
show the relationship between the various EFWs and the ABWs and correlation coefficient was
used to show the relationship between maternal heights, various weights and ABWs.
3.14 ETHICS
The study conformed to the guidelines for human experimentation in the Helsinki Declaration
“Ethical Principles for Medical Research Involving Human Subjects .Approval was obtained
University of Ghana http://ugspace.ug.edu.gh
45
from Ethics and Protocol Review Committee of the School of Biomedical and Allied Health
Sciences. Informed consent was obtained before the study after adequate explanation. Informed
written consents were obtained from the workers prior to scanning (Appendix III). Objectives
and examination procedure of the study were made available for the expectant mothers
(Appendix VI). Confidentiality of participants was maintained in accordance with the conditions
of ethical approval from the above-named ethics committee and privacy was ensured by keeping
the identity of the participants anonymous (Appendix V).
3.14.1 Standard Protocol Adopted for the Study
To prevent the re-occurrence of medical abuse of human subjects, the study was carried out with
strict adherence to the following protocol guidelines:
1. Voluntary participation was read to each participant, explanation given and their consent
obtained.
2. The researcher showed respect for participants and treated each of them as autonomous
agents.
3. Participants were given the right to decline participation in the research at any material
stage.
4. Their privacy was respected and their well-being protected.
5. Participants‟ integrity was safe guarded.
6. Participants were protected from mental, physical, and emotional harm.
University of Ghana http://ugspace.ug.edu.gh
46
CHAPTER FOUR
RESULTS
4.1 INTRODUCTION This Chapter presents the results of the study conducted to determine the best formula for
estimating foetal weight. The results are presented in accordance with stated objectives of the
study and included key aspects such as demographics characteristic, biometric measurements
obtained ultrasonically, computation of EFW by the chosen regression equations, weighing of
neonates and comparison of EFW and ABWs.
Seventy-five, 75 expectant mothers reporting for normal delivery at the study site initially
consented to participation following adherence to study protocols and ethics requirements out of
calculated sample size of seventy-eight, 78. However, only forty, 40 participants were used in the
study due to late referral of patients on suspicion of macrosomia, prolonged latent phase of
labour and breech presentations. Therefore, 96% response rate was achieved for the study.
4.2 DEMOGRAPHICS
The age and body mass index (BMI) demographics of the participants are shown in Table 4.1.
The mean age of the population was 26.93±6.06 years. The initial median BMI of the
participants at their first visit in the first trimester was 23.31and an inter-quartile range (IQR) of
5.15, and 27.80(IQR=5.33) at third trimester.
University of Ghana http://ugspace.ug.edu.gh
47
Table 4.1: Maternal Demographics (N= 40)
Maternal age (years)
Minimum Maximum Mean ± s.d
17 38 26.93 ± 6.06
BMI (kg/m2)
Median Inter-quartile range
Initial BMI 23.31 5.15
Final BMI 27.80 5.33
Statistical analysis was performed to establish correlation between maternal weight, height, initial
and final BMI, and neonates‟ weights (ABWs). The results are presented in Table 4.2.
Table 4.2: Correlation coefficient between maternal demographics and neonates’ ABWs
Demographic variables Correlation coefficient(R2) p-value
Initial maternal weight (kg) vrs. ABW (kg) 0.0251 0.8777
Final maternal weight (kg) vrs. ABW (kg) 0.1785 0.2705
Initial maternal BMI vrs. ABW (kg) 0.0502 0.7586
Final Maternal BMI vrs. ABW (kg) 0.2043 0.2061
Maternal height (cm) vrs. ABW (kg) -0.0856 0.5996
Source: Field data (2014)
The range of computed correlated coefficients {weight: R2 = 0.0251-0.1787; BMI: R
2= 0.0502-
0.2043; height: R2=-0.0856} and p-values {0.2061-0.8777> 0.05} within 95% confidence
University of Ghana http://ugspace.ug.edu.gh
48
intervals established significant differences between maternal anthropometrics and neonates
ABWs.
4.3 FETAL BIOMETRICS
Table 4.3 below shows the results of the sonographically determined fetal biometric data.
Table 4.3: Mean and standard deviation of BPD, HC, FL, and AC (N=40)
Fetal biometric measurements Lower limit Upper limit Mean ± s.d
BPD 8.99 9.28 9.13 ± 0.45
AC 34.91 36.42 35.67 ± 2.37
HC 31.94 33.16 32.56 ± 1.91
FL 7.44 7.71 7.58 ± 0.43
Source: Field data (2014)
Fetal weights determined via the eight regression models (Campbell & Wilkin 1975, Hadlock 2, 7,
4, 3, 6, 1985, Shephard, 1982, Shinozuka 2 1996) were compared with ABWs of 40 neonates. For
normal delivery at term, 35, (87.5%) newborns had BWs between 2.5kg and 3.9kg (Table 4.4)
Table 4.4: Birth weights at term
Birth weight at term (kg) Number Percent
< 2.5 2 5.0
2.5 – 3.9 35 87.5
> 3.9 3 7.5
Total 40 100.0
Source: Field data (2014)
The mean ABW ranged from 3.0kg –3.3kg with a mean of 3.2 ± 0.46kg.
University of Ghana http://ugspace.ug.edu.gh
49
Act
ual
bir
th w
eigh
t (k
g)
4.4 HYPOTHESIS
As hypothesized that there will be no significant difference between EFWs and ABWs, all the
eight equations or regression models present p-values<0.05 significance level. Therefore, the
study failed to reject the null hypothesis.
5
4.5
4
3.5
3
2.5
2
1.5 2 2.5 3 3.5 4 4.5
Regression model weights eqn_1 eqn_2 eqn_3 eqn_4 eqn_5 eqn_6 eqn_7 eqn_8
Linear (eqn_1) Linear (eqn_5)
Linear (eqn_2) Linear (eqn_6)
Linear (eqn_3) Linear (eqn_7)
Linear (eqn_4) Linear (eqn_8)
Figure 4.1: Scatter plot of ABW against regressional models Source: Field data (2014).
A scatter plot in Figure 4.1 showed a linear trend among all the 8 regression models with outliers
and their statistical measures as shown in table 4.5 below.
University of Ghana http://ugspace.ug.edu.gh
50
Table 4.5: Comparison between EFW generated by the regression models to ABWs
ABW ®ression models EFWs p-value
Lower limit Upper limit Mean± s.d
ABW (N = 40) 3.03 3.33 3.18 ± 0.46
Eqn_1: Campbell 1975 3.44 3.69 3.57 ± 0.07
Difference -0.52 -0.26 -0.39 ± 0.06 0.0001
Eqn_2: Hadlock 1 1885 3.03 3.33 3.57 ± 0.07
Difference -0.52 -0.26 -0.39 ± 0.07 0.0001
Eqn_3: Hadlock 6 1885 3.03 3.33 3.35 ± 0.05
Difference -0.31 -0.30 -0.17 ± 0.07 0.0200
Eqn_4: Hadlock 3 1885 3.53 3.57 3.55 ± 0.07
Difference -0.50 -0.23 -0.37 ± 0.07 0.0001
Eqn_5: Hadlock 7 1885 3.54 3.58 3.56 ± 0.07
Difference -0.52 -0.25 -0.38 ± 0.07 0.0001
Eqn_6: Hadlock 4 1885 3.53 3.57 3.55 ± 0.11
Difference -0.50 -0.24 -0.37 ± 0.07 0.0001
Eqn_7: Shephard 1982 3.36 3.73 3.55 ± 0.09
Difference -0.52 -0.22 -0.37 ± 0.07 0.0001
Eqn_8: Shinozuka 2 1996 3.55 3.59 3.57 ± 0.08
Difference -0.52 -0.25 -0.39 ± 0.07 0.0001
Source: Field data (2014)
University of Ghana http://ugspace.ug.edu.gh
51
With exception of the Hadlock31985 regression model, the results showed that the absolute
mean difference between the ABWs and the EFW sin all the regressions models were
statistically significant [p-value=0.0001<0.05, with a 95% CI, -0.52 to -0.23-0.26]. The absolute
mean difference between the ABW and Hadlock31985 EFW was also statistically significant [p-
value = 0.02 < 0.05 with a 95% CI of -0.31 to -0.30], showing no difference between the mean
ABW and the mean EFW.
In Table 4.5 Hadlock3 i.e. equation 4 shows a narrow interval of limits around the estimated mean
and the smallest standard deviation.
The Box and Whisker plot illustrating distinctly the relationship between the ABWs and EFWs
generated by the regression models and the associated distribution of the z-scores (median,
minimum, maximum, first quartile, third quartile and inter-quartile ranges) is shown in Figure
4.2 below.
University of Ghana http://ugspace.ug.edu.gh
Figure 4.2: Box and Whisker distribution of ABW and EFWs Source: Field data (2014)
52
University of Ghana http://ugspace.ug.edu.gh
53
CHAPTER FIVE
DISCUSSION
5.1 INTRODUCTION This Chapter discusses the findings of the study in relationship with reviewed literature. The
findings were presented in relationship to the stated objectives of the study under various themes.
5.2 RESPONSE RATE
A response rate of 75, (96%) out of 78, (100%) was achieved for this study. However, only 40,
53% out of 75 expectant mothers were used in the study. The low final figure of 40 was attributed
to late referral of 35, (47%) patients to tertiary (Specialist) centres on suspicion of macrosomia,
prolong latent phase of labour and breech presentations. Follow up on these participants by
telephone yielded no result in one week of hospitalization. This scenario even present higher
percentage comparable to study by Fernandez et al., (2009) in which the study reported that,
about 15% of participants were lost to follow up after 6weeks post hospitalization.
5.3 DEMOGRAPHICS
The ages of the 40 expectant mothers ranged from 17-38 years with a mean of 26.93± 6.1 years.
An initial and final median BMI of 23.31 and 27.80kg/m2and their corresponding IQR of 5.15
and 5.33 were also obtained respectively. The mean age of the population is representative of the
reproductive age group in Ghana and comparable to a mean age of 27.01±5.40years reported by
Kumarasiri et al. (2013) in an EFW study conducted on singleton pregnancies in Sri Lanka.
However the BMI and IQR in this study are in sharp contrast with the work of Kumarasiri et al.
(2013), which reported a median BMI of 20kg/m² and an IQR of 2.2.Additionally, statistical
University of Ghana http://ugspace.ug.edu.gh
54
significance was established between all groups of maternal anthropometric and ABW [p-
values>0.05] i.e. range of computed correlated coefficientR²-values and p-values (> 0.05) within
95% confidence intervals established significant differences between maternal anthropometrics
(Weight, Height, BMI) and neonates ABWs as shown in Table
4.2.
On the contrary, a study by Ojha & Malla (2007) indicated that low maternal anthropometric
(weight, height and BMI) are some risk factors that contribute to low BW, as confirmed by
Jananthan et al., (2009), that a strong correlation existed between maternal anthropometric and
ABW. Additionally, in WHO publication (Backstrand, 1995) of the Mexico Nutrition CRSP
research project in six Leishmann villages in the rural Solis Valley of central Mexico during
1984-1986 also established that weight and BMI showed the highest association with BW while
the effect of maternal height on ABW was insignificant. The contrast of the findings of this study
with the literature might be due to large sample size of those previous studies. However, findings
from this study are consistent with (Nahum et al., 1995).
5.3 FOETAL BIOMETRICS
All the 40 uncomplicated singleton pregnancies seen during the study period were between 35
0
and 406
weeks of gestation. Foetal weights determined via the 8 regression models were
compared with ABWs of 40 neonates. The mean duration from ultrasound scan to delivery was 1
day (range 0-2). For normal delivery at term, most of the newborns (n=35, 87.5%) had BWs in
the range of 2.5 and 3.9kg, while 7.5% (n=3) weighed more than 3.9kg. Only 5% (n=2) weighed
less than 2.5kg. According to WHO, low and high BW limits are defined as BW less than 2.5 kg
and greater than 3.9kgrespectively. Comparatively, the results obtained in this study are
consistent with the WHO limits.
University of Ghana http://ugspace.ug.edu.gh
55
According to Nahum et al., (1995) in Gerard et al., (2002), the daily rate of foetal growth is
13.0gfor boys and 12.4g for girls respectively. Hence for term deliveries of 270 days, these
figures translate into 3.51kg for boys and 3.35kg for girls. The EFWs obtained in this study are
in agreement with Nahum et al., (1985) and Gerard et al., (2002) as the mean EFW of all the
models is 3.46±0.65 and ABW ranged from 3.0kg –3.3kg with a mean of 3.2 ± 0.46kg.
From figure 4.2, the findings in this study confirmed many studies which concluded that most
EFW models generally overestimated foetal weight. Heer et al., (2008) established that, among
all known influencing factors, only a time interval of more than 7 days between EFW and
delivery had a negative impact on foetal weight estimation. Therefore, the principle of foetal
weight compensation from scan time to delivery time was therefore ignored in this study as it
presents no clinical significance of affecting neonates ABWs.
Luria et al., (2004) concluded that, there is a very good correlation between ultrasound EFW and
ABW. However, limits of agreement are fairly wide, thus ultrasound estimates of BW turn to
overestimate the neonatal weight by an average of 52g.Differences in anthropometry during
pregnancy appear to be related to BW. It was found that the smallest infants were those whose
mothers had a relatively lower pre-pregnancy weight (PPW) and early pregnancy weight, and
who subsequently failed to gain much weight by the third trimester (Backstrand, 1995) was also
established by this study.
Arguably, no previous study has been done in Ghana to assess the accuracy of using established
regression models in predicting EFW. This present study showed that the regression models
established no significant differences in predicting ABWs. The means and standard deviations
University of Ghana http://ugspace.ug.edu.gh
56
within 95% CI of the 4 foetal biometrics (BPD, HC, FL and AC) were very close, indicating the
precision of the theoretical models in determining ABWs in this study. However, inconsistent
with Kearney et al., (1978), who reported that inclusion of AC in estimating BW is the only
biometric measurement that can positively influence accuracy of EFW as it present the highest
standard deviation value of ± 2.37 around its mean compared standard deviation value of ± 0.43
for (FL) shown in Table-3.3.The disagreement might stem from the fact that, foetal abdominal
circumference in other populations differs greatly from that of Ghanaian one or differences in
accurate measurements.
A linear trend was noticed among all the 8 regression models with outliers. This means that,
individual point estimates as well as differences in estimates between any of the regression
models was a precise measure of the ABW. Per the earlier hypothesis, that there will be no
statistically significant difference between the ABWs and any of the 8 regression models. This
study indeed demonstrated that all the regression models showed no statistically significant
difference at (p-value =0.0001) in their mean EFWs compared to the neonates ABWs. This study
demonstrated that 87% of EFWs are within 5% of ABWs agreed with studies by Harlev et al.,
(2006), Shittu et al., (2007) and Atalie et al., (2006) which reported that about 75% of EFWs
were within 10% of ABWs. The 87% of EFW within 5% of ABWs compared to 75% of EFWs
within 10% of ABWs indicated how superior i.e. accuracy and precision of measurements in this
study as compared to others.
The 8 regression models adopted in this work are typically based on western populations and
hence does not account for racial and ethnic differences which present diversities in
anthropometrics. Further to this, regression analysis was used in the comparisons.
University of Ghana http://ugspace.ug.edu.gh
57
On the contrary, Altman and Bland (1983) questioned the appropriateness of using regression
analysis in such comparisons, suggesting that many factors influenced the accuracy of EFWs.
Kumarasiri et al., (2013) also concluded that operator experience played an important role in
accurate estimations of foetal weight among others. In this study, the mean BWs at term among
the Ghanaian population (3.53± 0.05kg - 3.57± 0.08kg) was comparable to the reference range of
3.08 ± 0.4kgderived from the global reference range based on WHO surveys of mean BW at
term (Mikolajczyk, 2011) who also used regression analysis.
In conclusion, all the 8 regression models predicted high accuracies of estimating foetal weights
within 5% of ABWs. With the exception of the Hadlock6(1985) regression model which had p-
value of 0.02, the results showed that the absolute mean difference between the ABWs and the
EFW in all the regression models were statistically significant with p-values = 0.0001. This is
suggestive that all the regressions models can be apply to Ghanaian population. However,
measuring BPD in labour period becomes less accurate as the head distends into the pubic bone
which turns to masked the foetal head from complete visualization and accurately measurement
coupled with narrow interval of limits around the estimated mean and the smallest standard
deviation, this study projects Hadlock3 i.e. equation 4 as regressional model which does not
depend on BPD for EFW, to be the best regressional model to be use in Ghanaian population at
term for estimating foetal weight.
.
University of Ghana http://ugspace.ug.edu.gh
58
CHAPTER SIX
CONCLUSIONS, RECOMMENDATIONS AND LIMITATIONS
6.1 INTRODUCTION
The study on the accuracy of ultrasound regression models for predicting foetal biometrics for
normal singleton pregnancies has been conducted. It was aimed at determining the best
ultrasound regression models using common foetal biometric measurements in EFW predictions
using the ClearVue 350 ultrasound equipment. The study included assessing records of maternal
anthropometrics (height and weight) from the time of first antenatal visit and weight at the last
antenatal visit, accurate measurements of foetal biometrics (BPD, HC, FL, and AC) as input data
into 8 established regression models, weight measurements of neonates within 1 hour of
delivery, comparisons of EFWs generated by the regression models to ABWs.
A summary of the present findings, conclusions, recommendations and limitations arising from
this research study are presented in this Chapter.
6.2 CONCLUSIONS
Comparisons of EFWs and ABWs have been investigated in various populations like that of (US,
Europe, Latin Americas and Africa) and by different methods. The findings of the study showed
that:
all the 8 regression models with the exception of the Hadlock7 which had (p-value =
0.02), established p-values = 0.0001 < 0.05 level of significance. Therefore, the best
University of Ghana http://ugspace.ug.edu.gh
59
ultrasound regressional model in predicting foetal weight in Philips ClearVue 350
ultrasound equipment is/are all the 8 models as there p-values <0.05 significant level.
the regression models predicted 87% of all EFWs in 5%of ABWs within 95%
confidence interval demonstrating high degree of accuracy and consistency.
Per these findings, gynaecologists, radiologists, sonographers/radiographers and midwives can
rely on EFWs produced from the Philips CleaVue ultrasound equipment using common foetal
biometric measurements in a 2D mode.
Prenatal care is essential for preventing foetal and/or maternal mortality and other complications.
However, timely interventions for patients at risk are required to prevent these complications. As
a result, if the method used for identification of patients at risk lacks sensitivity and specificity,
the trial of any intervention will not yield good results. It is therefore a challenge to estimate
actual foetal and BWs using ultrasound scans. Despite the limitations of ultrasound EFWs, the
modality remains the key basis for decisions amongst clinicians in many countries including
Ghana.
6.3 RECOMMENDATIONS
As noted from the study, it was only conducted in a microcosm population. Therefore,
geographical zones and country wide study should be done for better generalization.
Radiographers, sonographers, radiologists, gynaecologists, and midwives (professionally
trained and qualified to scan) should be trained every six months on emerging scan
procedures and produce accurate images and EFWs to enable good clinical decisions to
be taken.
University of Ghana http://ugspace.ug.edu.gh
60
The findings from this study have clearly shown that, with 87% of EFWs were within 5%
of ABW. Hence, radiographers, sonographers, gynecologists, radiologists, and midwives
can, to a high degree of accuracy rely on EFWs generated from the Philips ClearVue
ultrasound equipment using the common foetal biometric measurements in a 2D mode.
Caution must however be exercised especially in suspected large for date foetuses as all
models appear to slightly overestimated ABW.
6.3 LIMITATIONS OF THE STUDY
The major limitation as far as this study is concerned was the large number of referrals
from the study site to tertiary facilities after scanning. This has greatly affected the
sample size of this study.
Midwives in the labour-wards, most often, did not call the researcher when patients came
for delivery.
The scanning site was far from the labour-ward making it difficult for patients to walk in.
University of Ghana http://ugspace.ug.edu.gh
61
REFERENCES
Adams, J ., Ward, R.H. (1973). Admixture studies and the detection of selection. Science;
180:1137-1143CrossRef | Web of Science | Medlineage birth across generations.
American Journal of Obstetrics & Gynecology. 176:521
Airede, A.I. (Ed) (1995). Birth weight of Nigerian newborn infants. Areview: West Africa
Journal of Medicine: 14(2):116-120.
Almonen, E.A.A.H.A. (2015). Accuracy of ultrasound estimation of foetal weight in normal
vagina deliveries at Almuglad Hospital West Kordafan State- Sudan.
Altman, D.G., Bland, J.M. (1983). Measurement in medicine: the analysis of method comparison
studies. The Statistician: 32: 307-317.
Altman, D.G., Chitty, L.S. (1994). Charts of foetal size Methodology. British Journal of
Obstetrics & Gynaecology. 101: 29–34.
Amini, S.B., Catalano, P.M., Hirsch, V., Mann, L.I. (1994). An analysis of birth weight by
gestational age using a computerized perinatal data base. American Library of medicine,
National institute of Health Obstetrics &Gynecology; 83:342-352
Atalie, C., Dushyant, M., John H., Jeremy T. (2006). Reliability of ultrasound estimation of
foetal weight in term singleton pregnancies. The New Zealand Medical Journal;
119(1241).
Babbie, Earl, R. (2010). The Practice of Social Research. 12th ed. Belmont, CA: Wadsworth
Cengage,
Backstrand, J.R. (1995). Maternal anthropometry as a risk predictor of pregnancy
outcome:Bulletin of the World Health Organization, Vol. 73, No. SPEISS, Annual
University of Ghana http://ugspace.ug.edu.gh
62
Baker, P.N., Johnson, I.R., Gowland, P.A. (1994). Foetal weight estimation by echo-planar
magnetic resonance imaging. Lancet. 343:644–645.
Barlow, P. (1973). The influence of inactive chromosomes on human development. Anomalous
sex chromosome complements and the phenotype. American Library of medicine,
National institute of Health Humans; 17(2):105-36.
Baum, J.D., Gussman, D., Stone, P. (2002). Clinical and patient estimation of foetal weight vs.
Ultrasound estimation. American Library of medicine, National institute of Health
Journal of Reproduction Medicine.47:194–8.
Benacerraf, B.R., Gelman, R., Frigoletto, F.D. (1988). Sonographically estimated foetal weights:
accuracy and limitation. America Journal of Obstetric & Gynecoogyl.159:1118-21.
Ben-Haroush, A., Yogev, Y. Bar, J. (2004). Accuracy of sonographically estimated foetal weight
in 840 women with different pregnancy complications prior to induction of labour.
Ultrasound Obstetrics & Gynecology. 23:172–176.
Bennini, J.R., Marussi, E, F., Barini R., Faro, C., Peralta, C.F. (2010). Birth-weight prediction by
two-and three-dimensional ultrasound imaging. Ultrasound Obstetrics &Gynecology;
35:426-33.
Bland, J.M., Altman, D.G. (1986). Statistical methods for assessing agreement between two
methods of clinical measurement. Lancet.i: 307–310.
Bossak, W, S., Spellacy, W. N. (1972). Accuracy of estimating foetal weight by abdominal
palpation. Journal of Reproductive Medicine.9:58–60.
Brians, Craig L., et al. (2011). Empirical Political Analysis: Quantitative and Qualitative
Research Methods. 8th ed. Boston, MA: Longman,
University of Ghana http://ugspace.ug.edu.gh
63
Bukowski, R ., Gahn, D ., Denning, J ., S a a d e , G . (2001). Impairment of g r o w t h in foetuses
destined to deliver preterm. American Journal Obstetrics & Gynecology. 185(2):463-7
Burd, I, Srinivas S, Paré E., Dharan V.,Wang, E. (2009).Is Sonographic Assessment of Foetal
Weight Influenced by Formula Selection? American Library of medicine, National
institute of Health Journal of Ultrasound Medicine vol. 28 no. 8 1019-1024.
Campbells& Wilkin D. (1975). Ultrasonic measurement of foetal abdominal circumference in
the estimation of foetal weight. British Journal of Obstetrics & Gynaecology.82:689–97.
Cecatti, J.G, Machado, M.R., dos Santos, F.F., Marussi, E.F. (2000). Curve of normal foetal
weight values estimated by ultrasound according to gestational age. Cad Saude Publica
Brazil.; 16:1083–90. A
Chakraborty, R., Kamboh, M.I., Ferrell, R.E. (1991). `Unique' alleles in admixed populations: a
strategy for determining `hereditary' population differences of disease frequencies.
American Library of medicine, National institute of Health: Ethnicity & Disease; 1:245-
256. CrossRef | Web of Science | Medline
Chauhan, S.P., Lutton, P.M., Bailey, K.J., Guerrieri, J.P., Morrison, J.C. (1992). Intrapartum
clinical, sonographic, and parous patients‟ estimates of newborn birth weight. British
Journal Obstetrics & Gynecology. 79:956–8.
Chauhan, S.P., Hendrix, N.W., Magann, E.F., Morrison, J.C., Jenney, S.P., Devoe, L.D (1998).
Limitations of clinical and sonographic estimates of birth weight. American Journal
Obstetrics & Gynecology. 91:72–7.
Chen ATL, Chan Y-K, Falek A: The effects of chromosome abnormalities on birth weight in
man: II. Autosomal defects. Hum Hered 1972; 22:209. Publication retrieved from
http://www.researchgate.net/publication/232420247 on May 2014.
University of Ghana http://ugspace.ug.edu.gh
64
Chien, P.F., Owen, P., Khan, K.S. (2000). Validity of ultrasound estimation of foetal weight.
American Journal of Obstetrics & Gynecology. 95(6 Pt 1):856-60.
Chitty LS, Altman, D.G., Henderson, A., Campbell, S. (1994). Charts of foetal size: 4. Femur
length. British Journal of Obstetrics & Gynaecology.101: 132–135.
Chitty, L.S, Altman, D.G., Henderson, A., Campbell, S. (1994). Charts of foetal size: 2. Head
measurements. British Journal Obstetrics & Gynaecology.101: 35–43.
Chitty, L.S., Altman, D.G., Henderson, A., Campbell‟s. (1994). Charts of foetal size: 3.
Abdominal measurements. British Journal of Obstetrics & Gynaecology .101: 125–131.
Collins, J.W. Jr., David, R.J. (1990). The differential effect of traditional risk factors on infant
birthweight among blacks and whites in Chicago. America Journal of Public
Health.80:679-681
Colman,A., Maharaj.D., Hutton, J., Tuohy,J. (2006). Reliability of ultrasound estimation of
foetal weight in term singleton pregnancies. New Zealand Medical Journal; 119: U2146.
Combs, C.A., Jaekle, R.K., Rosenn, B., Pope, M., Miodovnik, Siddiqi, T.A. (1993). Sonographic
estimation of foetal weight based on a model of foetal volume. American Journal
Obstetrics& Gynecology. 82:365–70.
Courses.wcupa.edu; Publication retrieved from www. Courses.wcupa.edu on 17th March 2014
Creasy, R.K., Resnik R., Iam J.D. (2013). Creasy and Resnik‟s maternal foetal medicine:
Principles and practice.
Daftary, S., Chakravarti, S. (2011). Manual of Obstetrics, 3rd Edition. Elsevier. pp. 1-16. ISBN
Dare F.O., Ademowore, A.S., Ifaturoti, O.O., Nganwuchu, A. (1988). The value of
symphysiofundal height /abdominal girth measurement in predicting foetal weight.
International Journal of Gynaecology & Obstetrics.31:243–8.
University of Ghana http://ugspace.ug.edu.gh
65
David, R.J., Collins, J.W. Jr. (1991). Bad outcomes in black babies: race or racism? American
Library of medicine, National institute of Health: Ethnicity & Diseases; 1:236-244.
Deter, R.L., Hadlock, F.P. (Ed) (1985). Use of ultrasound in the detection of macrosomia: a
review. In American Library of medicine, National institute of Health: Journal of
Clinical Ultrasound. (pp 13-519).
Diejomaoh, F.M.E. (Ed) (1988). Perinatal and maternal mortality. Textbook of obstetrics and
gynaecology for medical student: (pp. 304–14). Lagos: University Services Educational
Publishers.
Dudley, N.J. (2005). A systematic review of the ultrasound estimation of foetal weight.
American Library of medicine, National institute of Health: Ultrasound Obstetrics &
Gynecology; 25:80–89.
Ebomoyi, E., Adetoro, O.O., Wickremasinghe, A.R. (1991). Birth weight and sociological
factors in Ilorin, Nigeria. Journal Bio Social Science. 23:417–23. US National Library of
Medicine, National Institute of Health.
Elliot, K., Knight, J, (Ed.). Size at Birth, Amsterdam: Associated Scientific Publishers; excluding
physiologic and pathologic effects. BJOG. International Journal of Obstetrics &
gynaecology; 102:918.
Fernandez, R.S, Davidson P., Griffiths R., Juergens, C., Stafford B., Salamonson Y. (2009). A
Pilot randomized controlled trial comparing a health-related lifestyle self-management
Intervention with standard cardiac rehabilitation following an acute cardiac event:
Implications for a larger clinical trial. Australian Critical Care; 22, 17-27.
University of Ghana http://ugspace.ug.edu.gh
66
Fields, D.A., Krishnan, S., Wisniewski, A.B. (2009). Sex differences in body composition early
in life. Gender Medicine 6: 369–375. US National Library of Medicine, National Institute
of Health.
Field, N., Piper, J., Langer, O., (1995). Effect of obesity in the accuracy of foetal weight
estimation.
Fox,S.H., Koepsell, T.D., Daling, J.R. (1994). Birth weight and smoking during pregnancy --
effect modification by maternal age. American Journal of Epidemiology. 139:1008-1015.
Glass, B., Li, C.C. (1953). The dynamics of racial intermixture -- an analysis based on the
American Negro. American Journal of Humuman Genetics; 5:1-20 Web of
Science | Medline
Goldenberg, R.L., Cliver, S.P., Cutter, GR. (1991). Black-white differences in newborn
anthropometric measurements. American Journal of Obstetrics Gynecology. 78:782-788.
Hadlock, F.P., Harrist, R.B., Carpenter, R.J., Deter, R.L., Park, S.K. (1984). Sonographic
estimation of foetal weight. The value of femur length in addition to head and abdomen
measurements. American Journal of Radiology 150: 535–540.
Hadlock, F.P., Harrist, R.B., Sharman, R.S., Deter, R.L., Park, S.K. (1985). Estimation of foetal
weight with the use of head, body, and femur measurements–a prospective study.
American Journal of Obstetrics & Gynecology. 151: 333–337.
Hanretty, K.P., Neilson, J.P., Fleming, E.E. (1990). Re-evaluation of clinical estimation of foetal
weight: a comparison with ultrasound. American Library of medicine, National institute
of Health: Journal of Obstetrics &Gynaecology.10:199–201.
Harlev A., Walfisch A., Bar-David, J., Hershkovitz R., Friger, M., Hallak, M. (2006). Maternal
estimation of foetal weight as a complementary method of foetal weight assessment: a
University of Ghana http://ugspace.ug.edu.gh
67
prospective clinical trial. American Library of medicine, National institute of Health
Journal of Reproductive Medicine; 51(7):515-20.
Hassibi, S., Farhataziz, N., Zaretsky, M., McIntire, D., Twickler, D.M. (2004). Optimization of
foetal weight estimates using MRI: comparison of acquisitions. American Journal
Roentgenology; 183:487-92?
Heer, I.M., Kümper C., NadinVögtle N., Müller-Egloff S., Dugas M., Strauss, A.
(2008). Analysis of Factors Influencing the Ultrasonic Foetal Weight
Estimation. Foetal Diagnosis and Therapy 23:204–210 DOI:
10.1159/000116742.
Hendrix,N.W., Grady CS, Chauhan, S.P. (2000). Clinical versus sonographic estimates of birth
weight in term of parturients. A randomized clinical trial. American Library of medicine,
National institute of Health; Journal of Reproductive Medicine 45:317–22.
Hoffman, H.J., Cliver, S.P., Brumfield, C.G. (1993). Foetal biparietal diameter, head
circumference, abdominal circumference and femur length. A comparison by race and
sex. American Journal of Reproductive Medicine: 38:201–206.
Hulsey, T.C., Levkoff, A.H., Alexander, G.R. (1991). Birth weights of infants of black and white
mothers without pregnancy complications. American Journal Obstetrics & Gynecology;
164:1299-1302
Jacquemyn, Y., Sys, S.U., Verdonk, P. (2000). Foetal biometry in different ethnic groups.
Science Direct: Early Human Development: 57: 1–13.
Jananthan, R., Wijesinghe, D.G.N.G., Sivananthawerl, T. (2009). Maternal Anthropometry as a
Predictor of Birth Weight. Tropical Agricultural Research Vol. 21(1): 89 - 98
University of Ghana http://ugspace.ug.edu.gh
68
Johar, R., Rayburn, W., Weir, D., Eggert, L. (1988). Birth weight in term infants. A 50-year
perspective. American Library of medicine, National institute of Health: Journal of
Reproductive Medicine; .33:813–6.
Johnstone, F., Inglis, L. (1974). Familial trends in low birth weight. British Medical Journal.
3:659.
Kearney, K., Vigneron, N., Frischman, P., Johnson, J.W. (1978) Foetal weight estimation by
ultrasonic measurement of abdominal circumference. American Journal of Obstetrics &
Gynecology. 51(2):156-62.
Klebanoff, M.A., Shino, P.H., Berendes, H.W., Rhoads, G.G. (1989). Facts and artifacts about
anemia and preterm delivery. American Library of medicine, National institute of Health
Journal of 262(4):511-515
Klebanoff, M.A., Schulsinger, C., Mednick, B.R., Secher, N.J.(1997). Preterm and small-for-
gestational-age birth across generations. American; 176(3):521-6.
Klebanoff, M.A., Mednick, B.R., Schulsinger, C., Secher, N.J., Shiono, P.H. (1998). Father‟s
effect on infant birth weight. American Journal of Obstetrics and Gynecology; 178 1022–
1026.
Kleinman, J.C., Kessel, S.S. (1987). Racial differences in low birth weight: trends and risk
factors. New England Journal of Medicine; .317:749-75
Kramer, M.S. (1987). Determinants of low birth weight: methodological assessment and meta-
analysis. Bull World Health Organ. 65(5):663-737.
Kumaral, D.M.A., Perera, H. (2009). Evaluation of six commonly used formulae for sonographic
estimation of foetal weight in a Sri Lankan population. Sri Lankan Journal of Obstetrics
and Gynaecology. 3'L: 20-33
University of Ghana http://ugspace.ug.edu.gh
69
Kumarasiri, S., Wanigasekara R., Wahalawatta L., Jayasinghe L., Padeniya T., Dias T. (2013).
Accuracy of ultrasound estimated foetal weight formulae to predict actual birth weight
after 34 weeks: prospective validation study. Ceylon; 58(3):116-21.
Kurmanavicius, J, Wright, E.M., Royston, P., Zimmermann, R., Huch, R., Huch, A., Wisser, J.
(1999). Foetal ultrasound biometry: 2. Abdomen and femur length reference values.
British Journal Obstetrics &Gynaecology.106: 136–143. Lancet 1:809.
Kurmanavicius, J., Wright, E.M., Royston, P., Wisser, J., Huch, R., Huch, A., Zimmermann, R.
(1999). Foetal ultrasound biometry: 1. Head reference values. British Journal of
Obstetrics & Gynaecology .106: 126–135.
Leung, T.N., Pang, M.W., Daljit, S.S., Leung, T.Y., Poon, C.F., Wong, S.M., Lau, T.K. (2008).
Foetal biometry in ethnic Chinese: Biparietal diameter, Head circumference, Abdominal
circumference and Femur length. American Library of medicine, National institute of
Health Journal of Ultrasound Obstetrics &Gynecology, 31: 321–327.
Lieberman, E., Ryan, K.J., Monson, R.R., Schoenbaum, S.C. (1987). Risk factors accounting for
racial differences in the rate of premature birth. New England Journal of Medicine
31:743-748
Lindell, G., Marsal, K. (2009). Sonographic foetal weight estimation in prolonged pregnancy:
Comparative study o f two-and three-dimensional methods. American Library o f
medicine, National institute of Health Journal of Ultrasound Obstetrics &Gynecology;
33:295-300.
Little, R.E., Sing, C.F. (1987). Genetic and environmental influences on human birth weight.
American Journal Human Genetics .40:512-26.
University of Ghana http://ugspace.ug.edu.gh
70
Loeffler, F.E. (1967). Clinical foetal weight prediction. Journal of Obstetrics &Gynaecology
British Commonwealth; 74:675-7.
Loughna, P., Chitty, L., Evans, T., Chudleigh, T. (2009). Foetal size and dating: Charts
recommended for clinical obstetric practice. Ultrasound, 17 (3) 160 - 166.
Luria, M., Rabinowitz, R., Shen, O. (2004). Can the sonographer predict the accuracy of
sonographic estimation of birth weight? 14th World Congress on Ultrasound in
Obstetrics & Gynecology. P14.6624: 269–372
Magnus, P. (1984). Further evidence for a significant effect of foetal genes on variation in birth
weight. Clinical Genetics.26:289-296.
Manning, F.A., Harman, C.R., Morrison, I., Menticoglou, S.M., Lange, I.R., Johnson, J.M.
(1990). Foetalassessments based on foetal biophysical profile scoring. American Journal
of Obstetrical Gynecology.Vol.162, No.3:703-709,
Mavalankar, D.V., Trived, C.R., Gray, R.H. (1991). Levels and risk factors for prenatal mortality
in Ahmadabad, India. Bull World Health Organ. 69:35–42
Mehdizadeh, A., Alaghehbandan, R., Horsan, H. (2000). Comparison of clinical versus
ultrasound estimation of foetal weight. . American Journal Perinatology. 17:233–6.
Melamed,N., Ben-Haroush, A., Meizner,I., Mashiach, R., Glezerman, M., Yogev, Y. (2011)
Accuracy of sonographic weight estimation as a function of foetal sex. American Library
of medicine, National institute of Health: Journal of Ultrasound Obstetrics
& Gynecology: 38: 67–73.
Melamed, N., Yogev, Y., Ben-Haroush., A., Meizner, I., Mashiach, R., Glezerman, M. (2012).
Does use of a sex-specific model improve the accuracy of sonographic weight
University of Ghana http://ugspace.ug.edu.gh
71
estimation? American Library of medicine, National institute of Health Journal of
Ultrasound Obstetrics &Gynecology, 39: 549–557.
Mikolajczyk, R.T., Zhang, J., Betran, A.P., Souza, J.P., Mori, R., Gülmezoglu, A.M., Merialdi
M. A. (2011). Global reference for foetal-weight and birth weight percentiles. Lancet;
377: 1855-61.
Mladenović-Segedi, L., Segedi, P. (2005). Accuracy of ultrasonic foetal weight estimation using
head and abdominal circumference and femur length. Medicinski Pregled: American
Library of medicine, National institute of Health Journal of 58(11-12):548-52.
Muijs, Daniel. (2010). Doing Quantitative Research in Education with SPSS. 2nd ed.
London: SAGE Publications,
Murray, J.L., Bernfield, M. (1988). The differential effect of prenatal care on the incidence of
low birth weight among blacks and whites in a prepaid health care plan. New England
Journal of Medicine; 319:1385-1391.
NahlaSubhi, A., Adel, T.A. (2010). Foetal weight normograms for singleton pregnancies in a
Jordanian population: Annals of Saudi Medicine. 30(2): 134–140.
Nahum, G.G., Stanislaw, H., Huffaker, B.J. (1995). Foetal weight gain at term: linear with
minimal dependence on maternal obesity. American Journal Obstetrics &
Gynecology.172 (5):1387-94.
Nahum, G.G. Estimation of foetal weight. Retrieved from Medicine/WebMD Obstetrics &
Gynecology on April 2014
Nahum, G.G., Stanislaw, H., Durham, N.C., Turlock, C. (2003). Ultrasonographic prediction of
term birth weight: How accurate is it? American Journal of Obstetrics and Gynecology;
188(2):566–574.
University of Ghana http://ugspace.ug.edu.gh
72
Naylor, A.F., Myrianthopoulos, N.C. (1967). The relation of ethnic and selected socio-economic
factors to human birth-weight. Annals of Human Genetics: 31:71-83
Nesbitt-Hawes, E.M., Tetstall, E., Gee,K., Welsh, A.C.(2014).Ultrasound (in)accuracy: it's in
the formulae not in the technique - assessment of accuracy of abdominal circumference
measurement in term pregnancies. Australasian Journal Ultrasound Medicine.17 (1):
Nzeh, D.A., Oyawoye, O., Adetoro, O.O. (2000). Ultrasound estimation of birth weight in late
pregnancy among African women. West African Journal of Ultrasound. 1:9–14.
Ojha, N., and Malla, D.S. (2007). Low birth weight at term: Relationship with maternal
anthropometry. Journal of Nepal Medical Association. 2007 Apr-Jun; 46(166):52-6
Ojwang, S., Ouko, B, C. (1984). Prediction of foetal weight in utero by fundal height/girth
measurements. Journal Obstetrics Gynecology East Central Africa. 3:111.
Oken, E., Kleinman, K.P., Rich-Edwards J., Gillman, M.W. (2003). A nearly continuous
measure of birth weight for gestational age using a United States national reference.
Biomed Central Pediatrics. 3:55–65.
Oliver, R., Fage, J.D. A short history of Africa. 6th ed. New York: Facts on File, 1988.
Ott, W.J., Doyle, S., Flamm, S., Wittman, J. (1986). Accurate ultrasonic estimation of foetal
weight. Prospective analysis of new ultrasonic formulae. American Journal of
Perinatology. 3:307–10.
Ounsted, M., Ounsted, C. (1966). Maternal regulations of intrauterine growth. Nature; 187:777.
Pang, M.W., Leung, T.N., Sahota, D.S., Lau, T.K., Chang, A.M. (2003). Customizing foetal
biometric charts. American Library of medicine, National institute of Health Journal of
Ultrasound Obstetrics &Gynecology; 22: 271–276.
University of Ghana http://ugspace.ug.edu.gh
73
Parker, A.J., Davies, P., Mayho, A.M, Newton, Jr. (1984). The ultrasound estimation of sex-
related variations of intrauterine growth. American Journal of Obstetrics & Gynecology.
149: 665–669.
Peregrine, E., O‟brien, P., Jauniaux, E. (2007). Clinical and ultrasound estimation of birth weight
priortoinduction of labor at term. American Library of medicine, National institute of
Health Journal: Ultrasound Obstetrics &Gynecology; 29: 304 – 309
Peuschel, S.M., Rothman, K.J., Ogilvy, J.D. (1976). Birth weight of children with Down's
Syndrome. American Journal of Mental Deficiency. 80(4):442-5
Pineau, J.C., Guihard-Costa, A.M., Droulle, P. (2003). Only two-phase models, computed
independently for males and females are appropriate to describe foetal head growth.
Foetal Diagnosis Thererap.18: 207–216.
Polani, P.E. (1974). Chromosomal and other genetic influences on birth weight variation.
In:Elliot, K., Knight, J., Ed:Size at birth(1992),Amsterdam. Associated Scientific
Publisher.
Raman, S., Urquhart, R., Yusof, M..Clinical versus ultrasound estimation of foetal weight.
Australian New Zealand Journal of Obstetrics Gynecology. 32:196–9.
Rawlings, J.S., Rawlings, V.B., Read, J.A. (1995). Prevalence of low birth weight and preterm
delivery in relation to the interval between pregnancies among white and black women.
New England Journal of Medicine; 332:69-74. Recommended for clinical obstetric
practice. Ultrasound. Volume 17, Issue 3, pp. 160-166
Reed, T.E. (1969). Caucasian genes in American Negroes. Science 1969; 165:762-768
CrossRef | Web of Science | Medline
University of Ghana http://ugspace.ug.edu.gh
74
Resnik, R., Creasy, R.K. (2012). Itrauterine growth restriction. Creasy and Resnik‟s maternal
foetal medicine.34:637
Richard, J., David, M.D., James W., Collins, Jr., M.D., (1997). Differing Birth Weight among
Infants of U.S.-Born Blacks, African-Born Blacks, and U.S.-Born Whites: New England
Journal of Medicine; 337:1209-1214.
Richards, M., Hardy, R., Kuh, D., Wadsworth, M.E. (2001). Birth weight and cognitive function
in the British 1946 birth cohort. Longitudinal population based study. British Medical
Journal; 322:199–203.
Robert, R., & Robert, K. C. (2010). Intrauterine Growth Restriction. Maternal-foetal Medicine.
34:1-2.
Rose and McCallum (1987). A simplified method for estimating foetal weight using ultrasound
measurements. American Library of medicine, National institute of Health Journal:
Obstetrics & Gynecology.64 (4):671-5
Royston, P., Wright, E.M. (1998). How to construct „normal ranges‟ for variables. American
Library of medicine, National institute of Health Journal: Ultrasound
Obstetrics& Gynecology. 11: 30–38.
Rutherford, S.E., Smith, C.V., Phelan, J.P., Kawakami, K., Ahn M.O. (1987). Four-quandrant
assessment of amniotic fluid volume. Inter-observer variation. American Library of
medicine, National institute of Health Journal: Journal of Reproductive Medicine, 32
(8):567-589.
Sabbagha, R.E., Minoque, J., Tamura, R.K., Hunqerford, S.A. (1989). Estimation of birth weight
by use of ultrasonographic formulae targeted to large-, appropriate-, and small-for
Gestational-age foetuses. American Journal of Obstetrics & Gynecology.160 (4):854-60
Salomon, L.J, Duyme, M., Crequat, J., Brodaty, G., Talmant, C., Fries N., Althuser, M. (2006).
University of Ghana http://ugspace.ug.edu.gh
75
French foetal biometry: Reference equations and comparison with other charts. American
Library of medicine, National institute of Health Journal: Journal of Obstetrics
Gynecology; 2006:28: 193–198.
Saugstad, L.F. (1972) Birth weights in children with phenylketonuria and in their siblings.
Lancet, 299 (7755): 809-813
Schild, R.L., Fell, K., Fimmers, R., Gembruch, U., Hansmann, M. (2004). A new formula for
calculating weight in the foetus of < or = 1600 g. American Library of medicine, National
institute of Health: Ultrasound Journal Obstetrics & Gynecology 24: 775–780.
Schild, R.L., Sachs, C., Fimmers, R., Gembruch U., Hansmann, M. (2004). Sex-specific foetal
weight prediction by ultrasound. American Library of medicine, National institute of
Health Journal: Ultrasound Obstetrics & Gynecology; 23: 30–35.
Schild, R.L., Fimmers, R., Hansmann, M. (2000). Foetal weight estimation by three -
dimensional ultrasound. American Library of medicine, National institute of Health
Journal: Ultrasound Obstetrics & Gynecology; 16:445-52
Schwarzler, P., Bland, J.M., Holden, D., Campbell, S., Ville, Y. (2004). Sex-specific antenatal
reference growth charts for uncomplicated singleton pregnancies at 15–40 weeks of
gestation. American Library of medicine, National institute of Health Journal:
Ultrasound Obstetrics & Gynecology; 23: 23–29.
Scioscia, M., Vimercati, A., Ceci, O., Vicino, M., Selvaggi, L.E. (2008). Estimation of birth
weight by two-dimensional ultrasonography: A critical appraisal of its accuracy.
American Library of medicine, National institute of Health Journal: Obstetrics &
Gynecology. 111:57–65.
Shamley, K.T.., Landon, M.B. (1994). Accuracy and modifying factors for ultrasonographic
University of Ghana http://ugspace.ug.edu.gh
76
determination of foetal weight at term. American Library of medicine, National institute
of Health Journal: Obstetrics & Gynecology. 84(6):926-30.
Sheehan, T.J., Gregorio, D.I. (1995). Low birth weight in relation to the interval between
pregnancies. New England Journal of Medicine: 333:386-387.
Shepard, M.J., Richards, V.A, Berkowitz, R.L, Warsof, S.L., Hobbins. J.C. (1982). An evaluation
of two equations for predicting foetal weight by ultrasound. American Journal of
Obstetrics & Gynecology; 142:47-54.
Sherman, D.J., Arieli, S., Tovbin, J., Siegel, G., Caspi, E., Bukovsky, I.A. (1998). A comparison
of clinical and ultrasound estimation of foetal weight. American Library of medicine,
National institute of Health Journal: Obstetrics and Gynecology.91:212–7.
Shittu, A.S, Kuti, O., Orji, E.O., Makinde, O.N., Solomon,O., Ogunniyi, S.O., Ayoola, O.O.,
Sule, S.S. (2007).Clinical versus Sonographic Estimation of Weight in Southwest
Nigeria.Journal of Health Population and Nutrition. ; 25(1): 14–23.
Shukla, N. (2014). Foetal development and foetal physiology. Retrieved May 2014 from
http://www.slideshare. net /nidhi shukla77736/foetal-development-42841132.
Siemer, J., Hilbert, A., Wolf, T., Hart, N., Muller, A., Schild, R.L. (2008). Gender- specific
weight estimation of foetuses between 2,501 and 3,999 g–new regression formulae.
Foetal Diagnosis and Therapy 24: 304-309.
Siemer, J., Wolf, T., Hart, N., Schrauder, M., Meurer, B., Goecker, T., Beckmann, M.W., Schild,
R.L. (2008). Increased accuracy of foetal weight estimation with a gender-specific weight
formula. Foetal Diagnosis and Therapy. 24: 321–326.
University of Ghana http://ugspace.ug.edu.gh
77
Simpson, J.W., Lawless, R.W., Mitchell, A.C. (1975): Responsibility of the obstetrician to the
foetus: 2. Influence of prepregnancy weight and pregnancy weight gain on birth weight.
American Library of medicine, National institute of Health Journal; Obstetrics &
Gynecology; 45:481.
Snijders, R.J.M., Sherrod, C., Gosden, C.M. (1993). Foetal growth retardation: Associated
malformations and chromosomal abnormalities. American; 168 (2):547–555.
Tamme, G., Matthias W., Beckmann, Ralf, L., Schild. (2008). Increased accuracy of foetal
weight estimation with a gender-specific weight formula: Departments of Obstetrics and
Gynaecology, University Hospitals, Mannheim, Erlangen, Germany: Foetal Diagnosis
and Therapy.24:321–326.
Thomson, A.M., Billewicz, W.Z., Hytten, F.E. (1968). The assessment of foetal growth. Journal
75(9):903-16.
Tlale, K.J. (2012). Validation of a simple clinical formula for predicting birth weight in women
who are in labour at term. 2011. Journal of Ultrasound 17: 161-7.
Uotila, J., Dastidar, P., Hannone, T., Ryymin, R., Punonen, R., Laasonan, E. (2000). Magnetic
resonance imaging compared to ultrasonography in foetal weight and volume estimation
in diabetic and normal pregnancy. Acta Obstetrics & Gynaecology Scandinavian journal:
79:255–9.
Walton, A., Hammond, J. (1938): The maternal effects on growth and conformation in the Shire
horse-Shetland pony crosses: Proceedings of the Royal society of London; Biological
Sciences 125:311
University of Ghana http://ugspace.ug.edu.gh
78
Warsof, S.L., Wolf, P., Coulehan, J., Queenan, J.T. (1986). Comparison of foetal weight
estimation formulae with and without head measurements. American Library of medicine,
National institute of Health Journal: Obstetrics &Gynecology. 67:569–73.
Watson, W.J., Soisson, A.P., Harlass, F.E. (1988). Estimated weight of the term foetus. Accuracy
of ultrasound vs. clinical examination. Journal of Reproductive Medicine. 33:369–71.
Wilcox, A.J., Skjaerven, R. (1992). Birth weight and perinatal mortality: the effect of gestational
age. American Journal of Public Health. 82:378–82.
Wilcox, M.A., Smith, S.J., Johnson, I.R., Maynard, P.V., Chilvers, C.E. (1995). The effect of
social deprivation on birth weight, excluding physiological and pathological effects.
British Journal Obstetrics & Gynecology. 102(11):918-24.
Wildschutt, H.I., Lumey, L.H., Lunt, P.W. (1991). Is preterm delivery genetically
determined? Journal of Paediatric Perinatal Epidemiology: 5 (4):363–372.
Willocks, J., Donald I., Duggan, T. C., Dayn. (1964). cephalometric by ultrasound. British.
Journal Obstetrics & Guynecology; 11: 71.
Yeo, G.S., Chan, W.B., Lun, K.C., Lai, F.M. (1994). Racial differences in foetal morphometryin
Singapore. Annals Academy of Medicine Singapore 23: 371–376.
Yoni, B. (1996). Accuracy of intrapartum estimates of foetal weight – effect of oligohydramnios.
Journal of Reproductive Medicine; 41:907-10
University of Ghana http://ugspace.ug.edu.gh
79
APPENDICES
APPENDIX I
PARTICIPANTS INFORMATION SHEET
DETERMINING BEST REGERESSIONAL MODEL IN PREDICTING FOETAL
WEIGHT IN PHILIPS ClearVue 350 ULTRASOUND EQUIPMENT
Purpose of the Study
You are being asked to participate in this study in order to test for the best regression model
(Equation) for predicting birth weight (BW).
Study Procedure and General Information.
If you agree to be part of this study, ultrasound scanning will be performed on your abdomen to
measure the head, abdomen, and thigh bone of your foetus.
Risk to the Patient
Patient lying on the examination bed for longer period can cause unnecessary discomfort due to
the pressure exerted on both the aorta and inferior vena cava. Therefore, the examination must be
done quickly with precision.
Additionally, although there is no risk associated with the use of ultrasound waves, however the
principle of ALARA (AS LOW AS REASONABLE ACHIEVABLE) with respect to ultrasound
dose will be strictly adhere to.
Patient’s Right
Any patient has the right to partake or withdraw at any time during study. All participants will be
fairly treated equally with no patient paying for the examination throughout the period of the
study. Any ethical violation by the investigator can be reported to the SAHS Ethical and protocol
review committee, College of Health Sciences, University of Ghana.
All participants shall be identified by codes. The following under listed names who are members
of the Dissertation Advisory Team (DAT) can be contacted for any concerns you have with
respect to this study:
University of Ghana http://ugspace.ug.edu.gh
80
APPENDIX II INFORMED
CONSENT FORM
Department of Radiography, School of Biomedical and Allied Health Sciences, College of
Health Sciences, University of Ghana
Title of Project: Determining Best Regeressional Model in Predicting Foetal Weight in Philips
Clearvue 350 Ultrasound Equipment
Purpose of Research: In Partial Fulfilment of MSc Degree
Name of Researcher: EDEM ANASTASIUS KWAMI BUADY
Name of Supervisors: DR. SAMUEL OPOKU, DR. ANIM-SAMPONG CONFIDENTIALITY:
All data will be anonymously and confidentially handle.
I………………………………………………………………………. agree to take part in the
above research and I consent to my participation. I understand that my participation is entirely
voluntary and that there is no personal benefit that I will derive by participating in this study
If you have any concern about this study and wish to contact someone independently, you may
contact the supervisors below. SIGNATURE…………………………………..
DR. SAMUEL OPOKU DR. ANIM-SAMPONG
Dept. of Radiography Dept. of Audiology
SBAHS SBAHS
University of Ghana University of Ghana
P.O. Box KB 143 P.O. Box KB 143
Korle-Bu Korle-Bu
Tel: 0246909286/0207934650 Tel: 0207774000
University of Ghana http://ugspace.ug.edu.gh
81
APPENDIX III
DATA COLLECTION SHEET
SECTION A: MATERNAL DEMOGRAPHIC INFORMATION
A. Code number………………………...
B. Age……………………………..years.
C. Height…………………….............cm
D. Weight………………………….....kg
E. Weight Difference………………...kg
SECTION B: FOETAL BIOMETRIC MEASUREMENTS
BPD…………………………………………..cm
AC………………………...………………….cm
HC……………………………………………cm
FL…………………………………………….cm