Use of silica microspheres having refractive index similar to bacteria for conversion of flow...
Transcript of Use of silica microspheres having refractive index similar to bacteria for conversion of flow...
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 6
Avai lab le a t www.sc iencedi rec t .com
journa l homepage : www.e lsev ie r . com/ loca te /wat res
Use of silica microspheres having refractive indexsimilar to bacteria for conversion of flow cytometricforward light scatter into biovolume
Paola Foladoria,*, Alberto Quarantab, Giuliano Ziglioa
aDepartment of Civil and Environmental Engineering, University of Trento, via Mesiano, 77, 38050 Trento, ItalybDepartment of Materials Engineering and Industrial Technology, University of Trento, via Mesiano, 77, 38050 Trento, Italy
a r t i c l e i n f o
Article history:
Received 3 January 2008
Received in revised form 5 May 2008
Accepted 23 June 2008
Available online 4 July 2008
Keywords:
Bacteria
Biovolume
Flow cytometry
Forward angle light scattering
Rayleigh–Gans theory
Wastewater
* Corresponding author. Tel.: þ39 461 882 66E-mail addresses: [email protected]
0043-1354/$ – see front matter ª 2008 Elsevidoi:10.1016/j.watres.2008.06.026
a b s t r a c t
This research describes an alternative approach for the rapid conversion of flow cytometric
Forward Angle Light Scattering (FALS) into bacterial biovolume. The Rayleigh–Gans theory
was considered for explaining the main parameters affecting FALS intensity: sensitivity
analysis of the model was carried out, taking into account the parameters characteristic
of bacterial cells and characteristics of the flow cytometer. For particles with size in the
typical range of bacteria, the FALS intensity is affected mainly by volume and refractive
index of bacterial cells and is approximately independent of the shape of the cells. The pro-
posed conversion from FALS intensity into bacterial biovolume is based on a calibration
curve determined by using silica microspheres having relative refractive index as far as
possible similar to that of bacteria. The approach was validated for two different flow
cytometers (the first equipped with an arc lamp and the second with a laser) by comparing
the biovolume distribution obtained from FALS conversion with the biovolume measured
conventionally under epifluorescence microscopy. The specific case of bacteria taken
from a WWTP was addressed. Compared to the time-consuming conventional microscopic
approach, the application of FALS for sizing bacterial biovolume could be a very promising
tool being completed in few minutes, simultaneously to the enumeration of bacteria during
the flow cytometric analysis.
ª 2008 Elsevier Ltd. All rights reserved.
1. Introduction offers a great potential, thanks to the possibility of measuring
In biotechnological processes and in environmental samples,
viable bacterial biomass is an important issue involved in
mass balance and in evaluating bacteria dynamics (growth
or decay). Cell concentration, biovolume and the specific car-
bon content or the dry weight of cells need to be known for
the calculation of bacterial biomass (Fry, 1990). In environ-
mental samples, cell sizing with epifluorescence or confocal
microscope is the conventional approach for biovolume
determination. Alternatively, the flow cytometry technique
9; fax: þ39 461 882 672.n.it (P. Foladori), quarantaer Ltd. All rights reserved
simultaneously viable bacterial concentration and their for-
ward angle light scattering which is correlated to the cellular
size (or biovolume). Flow cytometry is a single-cell analysis
and is showing increasing interest in environmental microbi-
ology for its rapidity in quantifying microorganisms. This
technique takes only a few minutes for the analysis of sev-
eral hundred or thousand cells in bacterial suspensions
with accuracy and high precision in the enumeration (Porter
et al., 1997; Steen, 2000). In the case of environmental samples,
flow cytometric analysis requires firstly cell staining with
@ing.unitn.it (A. Quaranta), [email protected] (G. Ziglio)..
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 63758
fluorescent probes in order to discriminate microorganisms
from other non-fluorescent non-biotic particles. For example,
the staining of cellular components by fluorescent dyes allows
the total bacteria number to be identified and their viability,
death or metabolic activity to be discriminated. Then total,
viable or dead bacteria can be rapidly and automatically enu-
merated by flow cytometry on the basis of their fluorescence
emission (Nebe-von-Caron et al., 2000; Ziglio et al., 2002). A
flow cytometer is usually equipped for the simultaneous
acquisition of two or more fluorescent signals and two light
scattering signals for each cell passing through the focus. In
particular, the light scattering acquired by the flow cytometer
in the forward direction – that is for angles of about 1–2� and
indicated as FALS (Forward Angle Light Scattering) – depends
on bacteria size (Salzman et al., 1990).
Empirical equations have often been sought by several
authors for the conversion of FALS signal into bacteria size
(Bouvier et al., 2001; Julia et al., 2000; Davey and Kell, 1996).
For microorganisms, most authors agree that FALS increases
monotonically with cell size and is a non-linear function of
the volume or the diameter of the bacteria (Davey and Kell,
1996; Julia et al., 2000). In some cases the non-linear curve
was fitted using a second-order polynomial curve (Davey
and Kell, 1996; Julia et al., 2000).
A rational approach for the conversion of FALS into
biovolume based on the Rayleigh–Gans theory was proposed
by Koch et al. (1996). On the basis of this theory, the FALS in-
tensity is proportional to the sixth power of the equivalent
sphere radius (or to the second power of particle volume).
Several authors mention the possibility of evaluating the
bacteria size by comparing FALS signal produced by the cells
to that given by microspheres (beads) of known diameter
(inter alia Koch et al., 1996). In this procedure, as underlined
by Koch et al. (1996), the refractive index of the series of par-
ticles must be the same. Synthetic beads for flow cytometry
applications are usually made of polystyrene or latex. FALS
signal produced by bacteria generally has a lower intensity
than the one produced by synthetic beads of the same size
or volume as the bacterium, due to the much higher refrac-
tive index of polystyrene or latex than cells. This can induce
an underestimation of the actual biovolume of bacteria, and
therefore the use of latex particles as size standards for bio-
logical cells may be problematic (Davey et al., 1993; Robertson
et al., 1998).
This aspect focuses on the importance of comparing FALS of
particles having the same optical characteristics, especially for
the relative refractive index. FALS intensity produced by cells
in starved strains or in natural communities can differ from
exponentially growing cells, as a consequence of the different
refractive indexes due to the different metabolic status of the
cells grown with excess of substrate or under limiting condi-
tions (Bouvier et al., 2001). Refractive index of bacteria may
vary from 1.36–1.40 for bacterial cells growing in minimal me-
dium or in environments with limited substrate (Valkenburg
and Woldringh, 1984; Robertson et al., 1998) to 1.40–1.41 for
bacilli in cultures (Ross, 1957). Considering marine bacteria
and phytoplankton cells growing in natural environments
where nutrients are less abundant, refractive indices have
been estimated to be in the range 1.39–1.40 (Jonasz et al.,
1997; Morel and Ahn, 1990) and in the range 1.39–1.45,
respectively (Twardowski et al., 2001; Stramski et al.,
2001). Refractive index of marine microorganisms such as
Synechococcus and eukaryotes is 1.41–1.43 on average (Green
et al., 2003). The purified organic material from a typical
phytoplankton cell has an average refractive index of
about 1.53 (Twardowski et al., 2001), so that it is not the
organic material but the large proportion of water that
gives bacteria low refractive indices. We are not aware of
any directly measured value for the refractive index of
bacteria present in wastewater and activated sludge in
the literature. It was estimated in this research on the ba-
sis of the agreement between the FALS converted into bio-
volume and the microscopic sizing of bacteria.
The present research describes a new and alternative
approach to rapidly convert FALS intensity measured by
flow cytometry into bacterial biovolume by using silica mi-
crospheres, having optical characteristics – especially the
relative refractive index – as far as possible similar to those
of bacteria.
The Rayleigh–Gans theory was considered for understand-
ing and supporting the conversion from FALS intensity into
bacterial biovolume, according to the model of Koch et al.
(1996). The Rayleigh–Gans law is applicable only if the phase
shift between the waves scattered from different points of
the target is low. This condition is satisfied if: (a) the size of
bacteria is not too large with respect to the incident light
wavelength and (b) the refractive index of bacteria is similar
to that of the surrounding medium: an index of refraction
only 3–6% higher than that of the surrounding medium mini-
mizes phase change and allows the use of Rayleigh–Gans
theory (Robertson et al., 1998; Koch et al., 1996). Both these
hypotheses are usually satisfied for bacteria in environmental
samples and in WWTPs. A sensitivity analysis was applied to
evaluate how the flow cytometer configuration and the phys-
ical properties of bacteria, both involved in the Rayleigh–Gans
model, affect FALS intensity.
The specific case of bacteria taken from a WWTP was
addressed. The approach was applied to bacteria in wastewa-
ter and activated sludge and was validated by comparing the
biovolume distribution obtained from FALS conversion with
the biovolume measured using a conventional approach
based on microscopy.
By means of the proposed calibration procedure, a rapid
conversion of the FALS signal (measured usually in arbi-
trary units) into biovolume (measured in mm3) is possible,
independent of the geometry and set-up of the specific
cytometer.
2. Materials and methods
2.1. Synthetic beads
Non-fluorescent silica microspheres (produced by MicroParti-
cles GmbH, Germany) with different diameters were used to
assess the calibration curve of FALS intensity. The selected
silica beads have a refractive index of 1.42, which is close to
the typical values of environmental bacteria. Six types of
microspheres with diameter between 0.5 and 1 mm were
chosen. Also fluorescent polystyrene beads (produced by
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 6 3759
Spherotech Inc, USA) having diameter ranging from 0.49 to
2.27 mm were used.
2.2. Samples of wastewater and activated sludge
Samples of wastewater were taken after the primary settler of
a municipal wastewater treatment plant (Trento, Italy). Acti-
vated sludge samples were collected from the oxidation tank
of the same plant, operating with a sludge age of about 12
days and at a specific organic load of 0.15 kg BOD5 kg TSS�1 d�1
(BOD5¼ Biochemical Oxygen Demand at 5 days; TSS¼ Total
Suspended Solids). Grab samples were collected and pro-
cessed in the lab within 1–2 h. Samples of wastewater and
activated sludge were diluted 1:20 and 1:500 v/v, respectively,
in PBS (3 g K2HPO4, 1 g KH2PO4 and 8.5 g NaCl L�1; pH¼ 7.2)
reaching approximately a concentration of 106–107 bacteria
per mL. Before the flow cytometric analysis, activated sludge
samples were pretreated in order to disaggregate flocs and
to obtain a suspension of dispersed bacteria. Pretreatment
was performed using sonication at a transferred specific
energy of 80 kJ L�1. This value of specific energy allows the
disaggregation of bacterial flocs avoiding the death of cells
(Foladori et al., 2007).
2.3. Fluorescent staining of cells
Bacteria in wastewater and activated sludge were stained with
the fluorochrome SYBR-Green I (Molecular Probes, Invitrogen),
adding 10 mL of dye (after 1:30 dilution of the commercial stock
solution) to 1 mL of sample containing about 106–107 bacteria.
An incubation of 15 min at room temperature and in the dark
was ensured for the optimal staining of bacteria by the dye.
SYBR-Green I, labelling nucleic acids inside the cells, has
excitation at a wavelength of 488 nm and emission of green
fluorescence at 525 nm (Ziglio et al., 2002).
2.4. Automated cell sizing under microscope
The size distribution of bacterial cells (previously stained with
SYBR-Green I) was measured under microscope, by using
a Nikon Labophot at 1000� magnification and equipped with
epifluorescence apparatus. For the microscopic examination,
a number of 1500–4000 bacteria in about 50–100 fields on
each slide were observed and recorded using a CCD camera
(Photometrics). An image processing program (LUCIA, Nikon)
extracted the particles from the image background after
green-level thresholding (cells stained with SYBR-Green I
emitted green fluorescence) and binarization. For each bacte-
rial cell the main parameters, such as minimum axis (a) and
maximum axis (b) were automatically measured, after the cal-
ibration of the images with a microscopic micrometer. Finally
the bacterial biovolume (V) was calculated, considering the
bacterial shape equivalent to an ellipsoid of revolution (Davey
et al., 1993):
V ¼ 16
pba2 (1)
The measuring of the cell dimensions was done with an error
in minimum and maximum axes of about 4.4% that corre-
sponds to an error in volume of about 13.2%.
2.5. Flow cytometry
Flow cytometric analyses were performed with two different
instruments: (1) a Bryte-HS flow cytometer (Bio-Rad, Inc.,
Hercules, CA) equipped with an arc lamp and is shown in
Fig. 1; (2) an Apogee-A40 flow cytometer (Apogee Flow Sys-
tems, UK) equipped with an Ar laser (488 nm) and character-
ised by an optical configuration similar to the one indicated
in Fig. 1. Apogee’s A40 flow cytometer is a modern high per-
formance model for counting small particles, such as small
bacteria or large virus, thanks to its greater light scatter
performance.
In both the flow cytometers, single cells are injected
through the flow cell and in front of a focus of a microscope
objective where each particle is hit by an excitation light
beam and the emitted signals are acquired. In the Bryte-HS
flow cytometer the incident light wavelength was selected in
the range 470–490 nm with a bandpass filter from a 150 W
Xe lamp. In the Apogee-A40 flow cytometer the incident light
wavelength is 488 nm produced by an Ar laser.
In both instruments, FALS signal (collected at small solid
angle, equal to 2� in the Bryte-HS), LALS (Large Angle Light
Scattering, collected at large solid angles) signal and 3 or 4
fluorescence signals were collected simultaneously for each
particle passing in front of the focus point. Green fluores-
cence emitted by SYBR-Green I was collected with a band-
pass filter with a bandwidth 515–565 nm in the Bryte-HS
flow cytometer and 515–545 nm in the Apogee-A40 flow
cytometer. During the flow cytometric analysis of bacteria
stained with SYBR-Green I, the threshold (for excluding
electronic noise) was set on the green fluorescence histo-
gram. Green fluorescence was used to gate FALS histogram,
in order to acquire only the FALS produced by nucleic acid-
stained bacteria and to eliminate scattering signals
originated by non-biotic particles or detritus present in
wastewater and activated sludge.
As expected, no difference was observed between the FALS
intensity of bacteria before and after the staining with SYBR-
Green I. This correspondence between the FALS intensity
without and with SYBR-Green I was verified with pure strains
in order to obtain a scattering signal well discriminated from
the instrument background.
FALS intensity of bacteria was recorded by adopting a linear
scale (that allows an easier conversion into biovolume with
respect to the logarithmic scale) mapped onto 256 channels.
A linear amplification factor from 10 to 40 was chosen depend-
ing on the size of the particles analysed.
For the analyses of silica beads the threshold was set only
on the FALS histogram, because beads are not fluorescent.
Every day, after daily calibration, during the analyses of differ-
ent beads or bacteria, no changes on photomultiplier values
were applied.
2.6. Modelling and theoretical aspects for describingFALS
The FALS intensity produced by bacterial cells was
explained on the basis of the Rayleigh–Gans approximation
of the Mie theory. The validity of using the Rayleigh–Gans
relationship for describing FALS of bacteria in wastewater
Fig. 1 – Optical configuration of the flow cytometer equipped with arc lamp and acquiring two scattering signals (at small
and large angles) and three fluorescences at different wavelength.
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 63760
and activated sludge (but also in most environmental sam-
ples) is justified by two considerations minimizing the
phase shift:
(1) the size of bacteria generally ranges from 0.3 and 1 mm, not
too large with respect to the wavelength (l) of the excita-
tion light applied in the flow cytometer, of 488 nm in the
Apogee-A40 and in the range 470–490 nm in the Bryte-HS;
(2) bacteria are characterised by high water content, which
gives a refractive index similar to the surrounding me-
dium, which is usually water.
In fact in general the equation of Rayleigh–Gans can be
applied to particles with size between 0.3 and 3 mm and with
refractive index (indicated hereafter as n) similar to the one
of the medium (n0) surrounding the particle.
In flow cytometry the intensity of the scattered light
depends on the size (characterised by the radius, r, of a spher-
ical particle having an equivalent volume) and the orientation
of the particles (given by b angle) passing in front of the focus
point. The scattering light intensity (I ) at a certain angle of
observation (q) and distance (R) is (Koch et al., 1996):
I ¼ I0PðqÞ8p4r6n40
R2l4
"ðn=n0Þ2�1
ðn=n0Þ2þ2
#2
vy�1þ cos2q
�(2)
where P(q) for a rotation ellipsoid is:
PðqÞ ¼�3ðsin x� x cos xÞ
x3
�2
(3)
where the x term depends on the two axis (a,b) of the ellipsoid:
x ¼ 4pan0
lsin
�q
2
�"sin2
bþ�
ba
�2
cos2b
#12
(4)
The equations above allow the scattering intensity for any q
value to be calculated.
Summing up the Rayleigh–Gans equation, the FALS
intensity depends on:
- the bacterial properties, such as size or volume (related to r),
elongation (defined by the ratio b/a), orientation (b), and rel-
ative refractive index (n/n0);
- the optical characteristics of the flow cytometer, such as the
intensity of the incident light (I0), its wavelength (l) and the
angle and the distance of observation (q, R, respectively).
Furthermore, from Eq. (2), the scattering intensity is
proportional to the sixth power of r, which means it is propor-
tional to the square of the biovolume.
In order to investigate how the different parameters affect
FALS intensity, a sensitivity analysis of the model was carried
out.
The parameters of the flow cytometer (I0, l, R, q) are
characteristic of each instrument and do not change during
the analysis: therefore they can be assumed constant in the
model. The parameter l was fixed equal to 488 nm for simulat-
ing Apogee-A40, while l was assumed of 480 nm in the Bryte-
HS, as mean value of the range 470–490 nm (the sensitivity
analysis demonstrated that changes in FALS intensity due to
variations in l from 470 to 490 nm were less than 0.2% for par-
ticles as small as bacteria and so this influence is negligible).
The parameter q was assumed equal to 2�.
Among the parameters characteristic of bacterial cells (r, b/
a, n/n0 and b), the radius r (related to the volume) and the
relative refractive index, n/n0, gave the higher sensitivity as
discussed below, whereas the influence of elongation and
cell orientation is quite negligible.
Fig. 2 shows the relationship between the FALS intensity
predicted by the model (expressed in arbitrary units) and the
particle volume for varying values of the refractive index of
the particles and considering spherical particles. Values of
particle volume up to 0.6 mm3, similar to values of bacterial
Fig. 3 – Influence of particle elongation on the FALS
intensity predicted by the model, referring to two different
refractive indexes: (A) particles with n [ 1.38; (B) particles
with n [ 1.42.
Fig. 2 – Relationship between particle volume and FALS
intensity as predicted by the model for different values of
the refractive index of the particles.
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 6 3761
cells, were considered. We selected refractive indexes, n,
typical of microbial cells, in the range 1.38–1.42. Water was
considered as surrounding medium (n0¼ 1.33).
By varying n a considerable difference in the FALS inten-
sity was observed. The higher the n value, the higher the
FALS intensity. Two particles having the same volume, but
different n (1.39 or 1.40), produce FALS intensities that differ
by 37%.
The influence of elongation of the particles (b/a ratio) on
FALS intensity predicted by the model is shown in Fig. 3.
Particles are considered as ellipsoids and elongation ranges
between 1 and 3 (1 in the case of spheres). The elongation
did not significantly affect the FALS intensity in the case of
particles made up of materials with n¼ 1.37–1.42, as indicated
in Fig. 3A and B where the curves coincide.
If FALS intensity would be simulated with a large angle of
observation (for example q higher than 20�), the influence of
elongation would be more evident (data not shown). In
conclusion, we can assume that elongation does not
significantly affect FALS intensity in the case of microbial
cells present in environmental samples (with n around
1.38–1.42).
On the basis of these results and considering particles in
the typical range size of bacteria, it can be concluded that
the FALS intensity is affected mainly by volume and refractive
index of bacterial cells and is approximately independent of
the shape of the cells.
3. Results and discussion
3.1. Calibration of FALS acquired by flow cytometry byusing silica beads as reference
The FALS intensity acquired by flow cytometry is expressed
in arbitrary units, usually applying a scale made up of 256 or
1024 channels. The purpose is to calculate the unknown ab-
solute volume of the bacterial cells starting from their FALS
intensity utilising reference beads having known size and
refractive index equal to bacteria. According to the model
explained above, the spherical shape of these reference
beads is adequate for representing also ellipsoidal bacterial
cells.
On the basis of these considerations, spherical beads made
of silica and not fluorescent were used as reference. Sizes sim-
ilar to bacteria were adopted, with beads volumes between
0.078 and 0.505 mm3. The refractive index of silica is equal to
1.42 (relative refractive index in water, n/n0, is equal to 1.065,
where n0 of water is equal to 1.333). This value is in the range
indicated in Section 1 for microorganisms in low-loaded
environments.
Since silica beads are not fluorescent, during the flow
cytometric analysis only the FALS signal was recorded and
represented in a histogram. The FALS signal produced by the
beads can easily be discriminated from electronic noise by
most flow cytometers on the market in the case of silica beads
Fig. 4 – Peak FALS channels acquired by flow cytometry as
a function of the volume of silica beads (square points) and
modelled FALS intensity predicted by the Rayleigh–Gans
theory: (A) data acquired by Bryte-HS; (B) data acquired by
Apogee-A40. Fig. 5 – Straight line for the conversion of FALS channels
(expressed in arbitrary scale) into particle volume: (A) data
acquired by Bryte-HS; (B) data acquired by Apogee-A40.
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 63762
over 0.7 mm, whereas only some instruments are able to detect
the weak FALS signal produced by silica beads with smaller
diameters (lower than 0.5 mm). Using the Bryte-HS flow cytom-
eter it was quite difficult to distinguish from background
beads with diameter smaller than 0.7 mm on the basis of
FALS signal. The most recently developed Apogee-A40 flow
cytometer allowed to distinguish very clearly silica beads hav-
ing smaller diameters up to 0.4 mm. The low value of n of silica
contributes to cause the weak FALS signal of silica beads; on
contrary the most common polystyrene or latex beads, char-
acterised by higher n values, can be easily detectable also for
small diameters.
For each size of silica beads (six sizes of beads) the channel
corresponding to the peak of the FALS frequency distribution
(indicated hereafter as ‘‘peak FALS channel’’) was recorded
with both Bryte-HS and Apogee-A40 flow cytometers. The
data of peak FALS channels as a function of the correspondent
volume of beads are shown in Fig. 4 (square points) for both
the flow cytometers.
FALS intensity produced by spheres with n of 1.42 and
volume up to 0.6 mm3 was also calculated on the basis of the
Rayleigh–Gans model explained above (and indicated hereaf-
ter as ‘‘modelled FALS intensity’’) and is again shown in
Fig. 4. The modelled FALS intensity was used to evaluate the
congruence of the peak FALS channel obtained experimen-
tally with silica beads by flow cytometry. The good correspon-
dence between the two FALS intensities (Fig. 4) confirms the
effectiveness of the Rayleigh–Gans model in describing the
FALS signal acquired by both the flow cytometers. The perfor-
mance of the two flow cytometers – the Bryte-HS equipped
with arc lamp and Apogee-A40 equipped with laser – resulted
very similar. Therefore, the procedure used for the conversion
of FALS into biovolume is similar in both cases (see paragraph
‘‘Validation of the FALS conversion.’’). The relationship
between peak FALS channels and the second power of the
particle volume is well fitted with a straight line (R2> 0.98)
as indicated in Fig. 5.
This straight line can be used for the immediate evaluation
of biovolume of bacteria after measurement of FALS intensity
with flow cytometry. On the basis of this conversion, each
FALS channel is associated with a specific cellular volume
and the whole scale of FALS expressed in arbitrary units can
be converted in an absolute volume scale (expressing the
particle volume in mm3). From a practical point of view, the
Fig. 6 – Peak FALS channels of polystyrene beads as
a function of the volume (circle points) and modelled FALS
intensity predicted by the Rayleigh–Gans theory.
Fig. 7 – Cytometric analysis of bacteria in activated sludge
by using Bryte-HS: (A) frequency distribution of FALS
intensity; (B) frequency distribution of cellular biovolume
calculated from FALS intensity.
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 6 3763
immediate conversion of FALS intensity into the correspond-
ing biovolume is based on the following equations:
Volume2 ¼ 0:0022 FALS� 0:0478
for the Bryte-HS flow cytometer (5)
Volume2 ¼ 0:0017 FALS� 0:0133
for the Apogee-A40 flow cytometer (6)
where FALS is expressed in channels, as acquired by the used
flow cytometer.
The intercept and the slope of the straight line (Eq. (5) or (6))
are coefficients depending on the specific characteristics of
the instrument used and therefore they could change for
different flow cytometers. Furthermore, the FALS intensity
produced by the same particles could change daily depending
on the setting of each instrument. As a consequence, Eq. (5) or
(6) may change and therefore the equation has to be verified
periodically in order to obtain accurate volume estimations.
The flow cytometric analysis of 4–6 different sizes of silica
beads and the calculation of Eq. (5) or (6) requires about 20–
30 min. Therefore, the calibration curve can be done easily
every day.
3.2. Inaccurate calibration of FALS by using polystyreneor latex beads as reference
As previously stated, inaccurate results would be expected, by
using beads made of fluorescent latex or polystyrene (widely
applied for daily calibration of cytometers) to simulate FALS
produced by bacterial cells. In fact, two particles having the
same shape and volume but significantly different refractive
indexes produce significantly different FALS intensities.
Polystyrene microspheres having diameter of 0.49, 0.87,
1.84, 2.27 mm and n of 1.585 were chosen. The FALS intensity
of beads was very strong, producing histograms well discrim-
inated from the electronic noise of the instrument. For each
size of beads, the channel corresponding to the peak of the
FALS frequency distribution was recorded (‘‘peak FALS chan-
nel’’) by using Bryte-HS flow cytometer and data are indicated
in Fig. 6 (circle points). Polystyrene beads suspended in water,
despite having size similar to bacteria, do not fall in the Ray-
leigh–Gans domain, as demonstrated by the large difference
from the modelled FALS intensity indicated in Fig. 6 as a con-
tinuous line.
3.3. Validation of the FALS conversion on bacteriapresent in wastewater and activated sludge
Bacterial cells present in wastewater and activated sludge
need to be stained with fluorescent probes, such as SYBR-
Green I, in order to be discriminated from other organic and
inorganic non-biotic particles, which can represent more
than 70% of the total solid concentration and could produce
the same FALS as bacteria. The scattering due to other non-
biotic particles was excluded from the analysis by gating
FALS on the bacteria population only. Before flow cytometric
Fig. 8 – Comparison between the bacterial biovolume
distribution measured under microscope and by Bryte-HS
flow cytometer (after conversion of FALS intensity) in the
case of (A) activated sludge and (B) wastewater.
Fig. 9 – Comparison between the bacterial biovolume
distribution measured under microscope and by Apogee-
A40 flow cytometer (after conversion of FALS intensity) in
the case of (A) activated sludge and (B) wastewater.
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 63764
analysis, aggregates were dispersed by sonication in order to
obtain a suspension composed mainly of single cells.
The FALS distribution obtained for activated sludge
bacteria by using Bryte-HS flow cytometer is shown in
Fig. 7A. The typical arbitrary scale from 0 to 256 channels is
indicated on the horizontal axis. Only 200 channels were indi-
cated, because very few cells fell in the range 201–256 chan-
nels. To evaluate bacterial biovolume, FALS intensity was
converted on the basis of Eq. (5). The applicability of this con-
version is warranted by the following considerations:
- the elongation (b/a ratio) of bacteria in wastewater and
activated sludge was in the range between 1.5 and 2.5,
within the maximum frequency for ratios of 1.7. This evalu-
ation was done under epifluorescence microscopy after
fluorescent staining of cells with SYBR-Green I. For these
values of b/a ratio and bacterial biovolume lower than
1.0 mm3, FALS intensity is not significantly affected by elon-
gation, as demonstrated in Fig. 3 according to the model.
Therefore, the use of the calibration curve obtained with
spherical beads is correct;
- the influence of cell orientation (b) on FALS signal is
negligible, as evaluated by the model;
- the refractive index of silica beads should be similar to that
of bacteria.
As a consequence the FALS signal produced by bacteria is
related exclusively to their size or volume (related to r).
The frequency of biovolume distribution (expressed in
mm3), converted from the FALS intensity, is shown in Fig. 7B.
According to Eq. (5), channel 22 is the minimum FALS
channel that can be used, in order to avoid negative biovo-
lumes. Consequently, some (very small) particles emitting
FALS intensity lower than 22 are excluded from the conver-
sion. This minor fraction of small cells does not significantly
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 6 3765
affect the sizing of the entire bacterial population or the
subsequent calculation of cellular biomass.
The bacterial biovolume distribution calculated convert-
ing the FALS signal acquired by flow cytometry was com-
pared to the biovolume of the same bacteria directly
measured under epifluorescence microscopy using an auto-
mated image analysis (Fig. 8). The size classes used in flow
cytometry are smaller and consequently much more numer-
ous than in microscopy (about 180 against 20–30 classes).
The biovolume distribution obtained according to the two in-
dependent experimental approaches show a good similarity
both for bacteria in activated sludge (Fig. 8A) and wastewater
(Fig. 8B); these results are referred to the application of the
Bryte-HS flow cytometer.
Analogous procedure for the conversion of FALS intensity
was applied in the case of Apogee-A40 flow cytometer. The
obtained bacterial biovolume distributions are indicated in
Fig. 9A and B for bacteria in activated sludge and wastewater,
respectively. The samples shown in Fig. 9 are different from
those presented in Fig. 8 being characterised by bacteria
with smaller size. A good agreement was observed between
the biovolume measured under microscope and that calcu-
lated from FALS intensity acquired by the Apogee-A40
laser-based flow cytometer. Although the refractive index of
bacteria in wastewater and activated sludge is not known,
this good agreement between cytometric and microscopic
sizing confirms that the refractive index of bacteria in waste-
water and activated sludge is around 1.42. However, because
populations of cells are generally heterogeneous in regard to
their refractive indices, the value of 1.42 has to be assumed as
an average value.
The biovolume frequency distribution allows some
important parameters to be calculated: (1) total cellular bio-
volume present in a bacterial suspension (Vtot), by multiply-
ing each volume class by the number of cells falling in this
class; (2) mean cellular volume, by dividing Vtot by the total
number of cells; (3) peak volume, corresponding to the
most frequently occurring volume of bacterial cells; (4)
other frequency indicators such as median or percentiles.
Furthermore, the cellular biovolume can be converted into
bacterial biomass expressed as dry weight (for example
referring to VSS, Volatile Suspended Solids, or COD, Chem-
ical Oxygen Demand). VSS and COD are typical parameters
used in mass balance in WWTPs.
It can be observed from Fig. 8 that bacteria present in
wastewater are slightly greater (peak at 0.20 mm3) than the bio-
volume of cells in activated sludge (peak at 0.16 mm3), probably
due to the longer sludge age of bacteria in activated sludge
(about 12 days) with respect to raw wastewater coming from
the sewerage. The mean cellular biovolume (excluding
aggregates greater than 0.6 mm3) evaluated from the data
shown in Figs. 8 and 9 ranged from 0.14–0.32 mm3 for wastewa-
ter to 0.12–0.25 mm3 for activated sludge. Only particles with
biovolume smaller than 0.6 mm3 (corresponding to an equiva-
lent sphere diameter of 1.05 mm) could be considered as single
cells, while particles greater than 0.6 mm3 are attributed either
to large-size cells, such as eukaryotic organisms, or to
bacterial aggregates not dispersed during pretreatment by
sonication. In wastewater and activated sludge, the number
of particles having volume greater than 0.6 mm3 was less
than 12% with respect to the total number of cells counted
by flow cytometry.
Similar results were founded by other authors. Vollertsen
et al. (2001) referred mean and median values of 0.22 mm3 and
0.29 mm3, respectively, in raw wastewater, by sizing bacteria
under microscope. In activated sludge samples, Frølund
et al. (1996) measured a mean biovolume equal to
0.25� 0.24 mm3, while Jahn and Nielsen (1998) found cellular
biovolumes in the range of 0.13–0.37 mm3 for sewerage
biofilm.
4. Conclusions
The procedure for the rapid conversion of FALS acquired
with two models of flow cytometers (Bryte-HS and Apo-
gee-A40) into bacterial biovolume can be described as
follows:
(1) selecting 4–6 types of silica beads having known size and
relative refractive index similar to that of bacteria;
(2) building a calibration curve between the silica beads
volume and the correspondent FALS intensity (choosing
adequate PMT, gain and number of FALS channel);
(3) analysing the bacterial suspension with flow cytometry,
measuring the bacterial concentration and the FALS inten-
sity (applying the same PMT, gain and number of FALS
channels of point 2);
(4) converting the FALS intensity into bacterial biovolume
(each FALS channel can be immediately converted
into a correspondent volume) and finally plotting the
biovolume distribution for the bacteria present in the
sample.
The proposed procedure was validated in the specific case
of bacteria present in wastewater and activated sludge,
obtaining a good correspondence between the FALS conver-
sion and the sizing under microscope.
The main parameter affecting the FALS conversion is the
bacterial refractive index: for example by assuming a variation
of n from 1.40 to 1.42, an underestimation of 22% for biovo-
lume could occur.
Although the refractive index of bacteria in wastewater
and activated sludge is not known, the good agreement
between cytometric and microscopic size distributions con-
firms that the value of 1.42 may be a good estimation of the
refractive index of these bacteria.
Acknowledgements
The Authors are grateful to the reviewers for their precious
suggestions. The authors thank L. Bruni of the Biological and
Chemical Laboratories of the Autonomous Province of Trento
for her appreciated collaboration during the analytical stages.
S. Tamburini, M. Ferrai and F. Ierace of the Department of Civil
and Environmental Engineering of the University of Trento are
also gratefully acknowledged for their help in data analysis
and processing.
w a t e r r e s e a r c h 4 2 ( 2 0 0 8 ) 3 7 5 7 – 3 7 6 63766
r e f e r e n c e s
Bouvier, T., Troussellier, M., Anzil, A., Courties, C., Servais, P.,2001. Using light scatter signal to estimate bacterial biovolumeby flow cytometry. Cytometry 44, 188–194.
Davey, H.M., Davey, C.L., Kell, D.B., 1993. On the determination ofthe size of microbial cells using flow cytometry. In: Lloyd, D.(Ed.), Flow Cytometry in Microbiology. Springer-Verlag,London, pp. 49–65.
Davey, H.M., Kell, D.B., 1996. Flow cytometry and cell sortingof heterogeneous microbial populations: the importanceof single-cell analyses. Microbiological Reviews 60,641–696.
Foladori, P., Bruni, L., Andreottola, G., Ziglio, G., 2007. Effectsof sonication on bacteria viability in wastewater treatmentplants evaluated by flow cytometry – fecal indicators,wastewater and activated sludge. Water Research 41,235–243.
Frølund, B., Palmgren, R., Keiding, K., Nielsen, P.H., 1996.Extraction of extracellular polymers from activated sludgeusing a cation exchange resin. Water Research 30, 1749–1758.
Fry, J.C., 1990. Direct methods and biomass estimation. Methodsin Microbiology 22, 41–85.
Green, R.E., Sosik, H.M., Olson, R.J., DuRand, M.D., 2003. Flowcytometric determination of size and complex refractive indexfor marine particles: comparison with independent and bulkestimates. Applied Optics 42 (3), 526–541.
Jahn, A., Nielsen, P., 1998. Cell biomass and exopolymer compositionin sewer biofilms. Water Science and Technology 37, 17–24.
Jonasz, M., Fournier, G., Stramski, D., 1997. Photometricimmersion refractometry: a method for determining therefractive index of marine microbial particles from beamattenuation. Applied Optics 36 (18), 4214–4225.
Julia, O., Comas, J., Vives-Rego, J., 2000. Second-order functionsare the simplest correlations between flow cytometric lightscatter and bacterial diameter. Journal of MicrobiologicalMethods 40, 57–61.
Koch, A.L., Robertson, B.R., Button, D.K., 1996. Deduction of thecell volume and mass from forward scatter intensity ofbacteria analysed by flow cytometry. Journal ofMicrobiological Methods 27, 49–61.
Morel, A., Ahn, Y.H., 1990. Optical efficiency factors of free-livingmarine bacteria: influence of bacterioplankton upon the
optical properties and particulate organic carbon in oceanicwaters. Journal of Marine Research 48, 145–175.
Nebe-von-Caron, G., Stephens, P.J., Hewitt, C.J., Powell, J.R.,Badley, R.A., 2000. Analysis of bacterial function by multi-colour fluorescence flow cytometry and single cell sorting.Journal of Microbiological Methods 42, 97–114.
Porter, J., Deere, D., Hardman, M., Edwards, C., Pickup, R., 1997. Gowith the flow – use of flow cytometry in environmentalmicrobiology. FEMS Microbiology Ecology 24, 93–101.
Robertson, B.R., Button, D.K., Koch, A.L., 1998. Determination ofthe biomasses of small bacteria at low concentrations ina mixture of species with forward light scatter measurementsby flow cytometry. Applied and Environmental Microbiology64, 3900–3909.
Ross, K.F.A., 1957. The size of living bacteria. Quarterly Journal ofMicroscopical Science 98 (4), 435–454.
Salzman, G.C., Singham, S.B., Johnston, R.G., Bohren, C.F., 1990.Light scattering and cytometry. In: Melamed, M.R.,Mullaney, P.F., Mendelsohn, M.L. (Eds.), Flow Cytometry andSorting. John Wiley & Sons, Inc, New York, pp. 81–108.
Steen, H.B., 2000. Flow cytometry of bacteria: glimpses from thepast with a view to the future. Journal of MicrobiologicalMethods 42, 65–74.
Stramski, D., Bricaud, A., Morel, A., 2001. Modeling the inherentoptical properties of the ocean based on the detailedcomposition of the planktonic community. Applied Optics 40,2929–2945.
Twardowski, M.S., Boss, E., Macdonald, J.B., Pegau, W., Barnard, A.H., Zaneveld, J.R.V., 2001. A model for estimating bulkrefractive index from the optical backscattering ratio and theimplications for understanding particle composition in case Iand case II waters. Journal Of Geophysical Research 106 (C7),14129–14142.
Valkenburg, J.A.C., Woldringh, C.L., 1984. Phase separationbetween nucleoid and cytoplasm in Escherichia coli as definedby immersive refractometry. Journal of Bacteriology 160,1151–1157.
Vollertsen, J., Jahn, A., Nielsen, J.L., Hvitved-Jacobsen, T.,Nielsen, P.H., 2001. Comparison of methods for determinationof microbial biomass in wastewater. Water Research 35,1649–1658.
Ziglio, G., Andreottola, G., Barbesti, S., Boschetti, G., Bruni, L.,Foladori, P., Villa, R., 2002. Assessment of activated sludgeviability with flow cytometry. Water Research 36, 460–468.