Pulse wave velocity and digital volume pulse as indirect estimators of blood pressure: pilot study...
-
Upload
independent -
Category
Documents
-
view
2 -
download
0
Transcript of Pulse wave velocity and digital volume pulse as indirect estimators of blood pressure: pilot study...
ORIGINAL PAPER
Pulse Wave Velocity and Digital Volume Pulse as IndirectEstimators of Blood Pressure: Pilot Study on Healthy Volunteers
Juan M. Padilla Æ Enrique J. Berjano ÆJavier Saiz Æ Rafael Rodriguez Æ Lorenzo Facila
� Springer Science+Business Media, LLC 2009
Abstract The purpose of the study was to asses the
potential use of pulse wave velocity (PWV) and digital
volume pulse (DVP) as estimators of systolic (SBP) and
diastolic (DPB) blood pressure. Single and multiple cor-
relation studies were conducted, including biometric
parameters and risk factors. Brachial-ankle PWV (baPWV)
and DVP signals were obtained from a Pulse Trace PWV
and Pulse Trace PCA (pulse contour analysis), respec-
tively. The DVP (obtained by photoplethysmography),
allowed stiffness (SI) and reflection indexes (RI) to be
derived. The first study on 47 healthy volunteers showed
that both SBP and DPB correlated significantly both with
baPWV and SI. Multiple regression models of the baPWV
and the waist-to-hip ratio (WHR) allowed SBP and DBP to
be modeled with r = 0.838 and r = 0.673, respectively. SI
results also employed WHR and modeled SBP and DBP
with r = 0.852 and r = 0.663, respectively. RI did not
correlate either with SBP or DBP. In order to avoid the use
of ultrasound techniques to measure PWV, we then
developed a custom-built system to measure PWV by
photoplethysmography and validated it against the Pulse
Trace. With the same equipment we conducted a second
pilot study with ten healthy volunteers. The best SBP
multiple regression model for SBP achieved r = 0.997 by
considering the heart-finger PWV (hfPWV measured
between R-wave and index finger), WHR and heart rate.
Only WHR was significant in the DBP model. Our findings
suggest that the hfPWV photoplethysmography signal
could be a reliable estimator of approximate SBP and could
be used, for example, to monitor cardiac patients during
physical exercise sessions in cardiac rehabilitation.
Keywords Blood pressure � Digital volume pulse �Pulse wave velocity � Photoplethysmography �Stiffness index � Reflection index
Introduction
Noninvasive monitoring of blood pressure (BP) is usually
performed by traditional sphygmomanometry. However,
this technique is unable to monitor short-term changes, so
that a noninvasive method of measuring beat-to-beat BP
would be extremely valuable. This is especially crucial in
monitoring cardiac patients undergoing treadmill exercise
tests during cardiac rehabilitation. It has been suggested
that rate-pressure product (i.e., heart rate 9 systolic blood
pressure) is the key monitoring parameter during this
procedure (Pierson et al. 2004; Nieminen et al. 2008).
Our objective was therefore to assess the potential use of
certain cardiovascular signals (obtained non invasively)
as estimators of systolic (SBP) and diastolic (DPB) blood
pressure. Since a near linear correlation exists between
transit time TT (measured from the R wave) and BP, this
parameter could be used as a surrogate marker of
J. M. Padilla � E. J. Berjano � J. Saiz � R. Rodriguez
Institute for Research and Innovation on Bioengineering (I3BH),
Universidad Politecnica de Valencia, Valencia, Spain
J. M. Padilla
Instituto Tecnologico de Morelia, Morelia, Michoacan, Mexico
E. J. Berjano (&)
Departamento de Ingenierıa Electronica, Universidad Politecnica
de Valencia, Camino de Vera, 46022 Valencia, Spain
e-mail: [email protected]
L. Facila
Departamento de Cardiologıa, Hospital Provincial de Castellon,
Valencia, Spain
123
Cardiovasc Eng
DOI 10.1007/s10558-009-9080-5
pressure (Geddes et al. 1981; Chen et al. 2000). However,
in order to obtain more general and reproducible results,
we initially chose two cardiovascular signals, the pulse
wave velocity (PWV) and the digital volume pulse
(DVP), obtained from commercially available equipment.
PWV is known to be an indicator of arterial stiffness
(Lehmann 1999) and is largely determined by BP. A
regional estimation of the properties of the arterial wall can
be made by measuring local PWV, for instance in the aorta,
where an invasive recording technique is used to record the
aortic pulse wave velocity (aPWV) (Liu et al. 2004).
Finally, brachial-ankle pulse wave velocity (baPWV) is a
non invasive PWV measurement system which shows a
good correlation with aPWV obtained by invasive record-
ing (Yamashina et al. 2002).
Digital Volume Pulse (DVP) is an alternative method of
measuring arterial stiffness (Chowienczyk et al. 1999) that
provides valid information on the pressure pulse waveform
(Millasseau et al. 2002). DVP can be recorded in the finger
by photoplethysmography (PPG) using an infrared optical
transducer. It usually exhibits an early systolic peak and a
later peak or point of inflection that occurs a short time
(DTDVP) after the first peak in early diastole (see Fig. 1).
It can provide two indexes: stiffness index (SI), which
relates with large artery stiffness, and the reflection index
(RI), which relates with vascular tone (Millasseau et al.
2002). The DVP waveform is determined mainly by the
characteristics of systemic circulation, including pressure
wave reflection and PWV of pressure in the aorta and large
arteries (Chowienczyk et al. 1999). To date, there have
been no previous studies employing either baPWV or DVP
to estimate BP.
The aim of this work was to asses the potential use of
pulse wave velocity (PWV) and digital volume pulse
(DVP) as estimators of systolic (SBP) and diastolic (DPB)
blood pressure in healthy subjects.
Materials and Methods
This study was divided into two phases. In the first, the
potential use of brachial-ankle PWV (baPWV) and digital
volume pulse (DVP) were studied as estimators of systolic
(SBP) and diastolic (DPB) blood pressure. Both signals
were obtained from commercially available equipment. In
the second phase, we studied the potential use of other
signals related with PWV and obtained by means of custom
built-equipment based on photoplethysmography.
First Phase
The first phase involved 47 healthy volunteers
(33.78 ± 9.04 years). The subjects were examined in the
supine position, with electrocardiogram electrodes placed
on both wrists after a 15 min rest. Blood pressure, baPWV
and DVP were measured three times for averaging. baPWV
was recorded by Pulse Trace PWV (Micro Medical, Kent,
UK). This equipment uses a 4 MHz Doppler probe to
identify the arrival of the arterial pulse. The arterial pulse
waveform was sequentially measured in two locations of
the arterial tree (right brachial and right ankle) and was
timed using the R wave of the ECG (Lead I) as shown in
Fig. 2. The Pulse Trace PWV automatically computed the
transit time (DTPWV) from the onset time difference
between brachial and ankle waveforms. baPWV was then
calculated by dividing the externally measured distance
between brachial-ankle locations (Lba) by the DTPWV:
baPWV ¼ Lba =DTPWV ð1ÞThe distance Lba between the measuring sites (i.e., the
brachial and posterior tibial arteries) was measured with a
tape measure. To reduce the influence of body contours on
the distance, the tape measure was held above the surface of
the body parallel to the plane of the examination table. Five
Fig. 1 a Digital Volume Pulse
(DVP) waveform obtained by
photoplethysmography. It
consists of a systolic peak (a)
and diastolic peak (b) separated
by a transit time delay
(DTDVP). b Arterial system
diagram highlighting the aortic
segment: the systolic peak of
DVP waveform corresponds to
the blood forward wave (Fw)
while the diastolic peak
characterizes the blood reflected
wave(s) (Rw) in the arterial
system
Cardiovasc Eng
123
partial distances were measured, as shown in Fig. 3: (1)
from suprasternal notch to inferior edge of the umbilicus;
(2) from the inferior edge of the umbilicus to the iliac crest;
(3) from to the iliac crest to the medial malleolus in sam-
pling site on the right leg; (4) from suprasternal notch to
head of humerus; and (5) from head of humerus to the
forearm. Lba was then calculated by subtracting the distance
between the suprasternal notch to medial malleolus site on
the posterior tibial artery from the distance between the
suprasternal notch to the forearm site on the brachial artery.
DVP waveform was obtained by using a Pulse Trace
PCA (Micro Medical, Kent, UK) which employs a pho-
toplethysmography (PPG) transducer. This techniques fol-
lows the changes in the reflectance of infrared light
(k = 940 nm) through the finger pulp. The PPG transducer
was placed on the index finger of the right hand. The Pulse
Trace PCA provides two indexes from the DVP: the
Stiffness Index (SI) and Reflection Index (RI). SI (mea-
sured in m/s) is defined as the ratio between subject’s
height (h) and DTDVP (see Fig. 1a):
SI ¼ h =DTDVP ð2Þ
The reflection index (RI) (measured as %) is defined as
the ratio between the amplitudes of the second (b) and first
(a) peaks of the DVP waveform (see Fig. 1a):
RI ¼ b=a ð3Þ
Regarding the risk factors, the waist to hip ratio (WHR)
was calculated as the minimal abdominal circumference
between the xiphoid process and iliac crests (i.e., waist)
divided by the circumference over the femoral heads (i.e.,
hips). The body mass index (BMI) was calculated as the
ratio of body weight (kg) to the square of height (m). BP
was measured using an automatic sphygmomanometer
Omrom M6 (Omrom Healthcare, Milton Keynes, UK),
following the guidelines of the European Society of
Cardiology and Hypertension (2003). Age and current
smoking status were determined by questionnaire.
Preliminary tests (in order to design the experimental
set-up) showed the marked influence of ambient tempera-
ture on the hemodynamic parameters (blood pressure,
PWV and DVP). This parameter was therefore carefully
controlled during the tests and kept at 21 ± 1�C. The
acquisition of the DVP signal was an automatic process,
i.e., the Pulse Trace PCA (Micro Medical, Kent, UK) only
showed DVP values when the recording was of sufficient
quality, which was achieved after pre-warming the sub-
ject’s finger.
Fig. 2 Transit time (DTPWV) was computed from the onset time
difference between the signals detected in two sites of measurement:
the brachial and ankle arteries. The waveforms were obtained by
means of an ultrasound sensor and are shown as brachial and ankle
waveforms. The R wave of ECG was used by the Pulse Trace PWV
(Micro Medical, Kent, UK) as timing reference to calculate DTPWV
automatically, since only one ultrasound sensor was employed. TS and
TD are systolic and diastolic period, respectively. The figure shows an
actual register obtained in our study
Fig. 3 Partial distances employed to calculate the right brachial–
ankle distance (Lba). Lba was calculated as the sum of the partial
distances 1, 2 and 3, minus the sum of the partial distances 4 and 5
(Lba = 1 ? 2 ? 3 - 4 - 5). The partial distances were (1) from
suprasternal notch to inferior edge of the umbilicus; (2) from the
inferior edge of the umbilicus to the iliac crest; (3) from the iliac crest
to the medial malleolus in the sampling site on the right leg; (4) from
suprasternal notch to head of humerus; and (5) from head of humerus
to the forearm
Cardiovasc Eng
123
Second Phase
In order to avoid the use of ultrasound techniques to
measure PWV, in the second part of the study we designed
a system to measure PWV by photoplethysmography
instead of using a Doppler probe. The equipment (see
Fig. 4) consisted of a signal conditioning module con-
nected to a data acquisition card controlled by LabView
software (National Instruments, Austin, TX, USA). The
module processed an ECG signal (Lead I), and two pho-
toplethysmography signals which registered the changes in
the transmittance of infrared light through the index finger
(index finger pulse wave) and the big toe (big toe pulse
wave). With these three signals, we defined three new
parameters: ftPWV (PWV measured between index finger
and big toe), hfPWV (PWV measured between heart -R
wave of ECG- and index finger), and htPWV (PWV
measured between heart and big toe). In each case PWV
was calculated by dividing the externally measured dis-
tance between each location by the recorded pulse transit
time (PTT). In the case of photoplethysmography signals,
PTT was measured by the onset of each pulse. The Fig. 5
shows typical waveforms obtained from the specially
designed equipment, and the PTT measured for each sig-
nal. These signals provided an additional stiffness index
and an additional reflection index, which were compared to
those obtained from the Pulse Trace PCA (Micro Medical,
Kent, UK) for validation. After validating the system, we
conducted the second phase of the study on ten healthy
volunteers (30.22 ± 7.06 years). The measuring protocol
was similar to the first phase, but included measuring the
new parameters.
Statistical Analysis
All statistical analyses were performed using SPSS 13
(SPSS, Chicago, IL, USA). Data were expressed as
mean ± standard deviation. Pearson0s correlation coeffi-
cients (r) were calculated to compare the relationships
between the variables under study and risk factors. Values
of P \ 0.05 were considered to indicate statistical signifi-
cance. Multiple regression analyses were automatically
conducted by a stepwise regression method (backward
elimination). Once the models had been obtained, they
were tested by means of the Durbin-Watson statistic,
Fisher-Snedecor statistic, variance inflation factor (VIF),
and the condition index.
Results
First Phase Overview
The first phase of the study evaluated the relationships
between cardiovascular parameters, cardiovascular mea-
surements (baPWV, SI and RI), and risk factors. Table 1
shows the basic characteristics of the 47 healthy subjects in
this group. The cardiovascular parameters included SBP,
DBP, mean blood pressure (MBP), pulse pressure (PP) and
heart rate (HR). The values of the variables under study
obtained were: baPWV of 8.4 ± 1.4 m/s, SI of 6.28 ±
0.83 m/s, and RI of 68.53 ± 8.07%. Table 2 shows the
correlation coefficients and significance between baPWV,
SI, RI, cardiovascular parameters and risk factors.
Assessment of baPWV
baPWV was positively and significantly correlated with all
the cardiovascular parameters and presented moderately
high correlation coefficients with SBP (r = 0.57,
P = 0.0001), DBP (r = 0.55, P = 0.0001) and MBP
(r = 0.60, P = 0.0001). It was also positively and signif-
icantly correlated with some risk factors, such as BMI
(r = 0.48, P = 0.001), WHR (r = 0.58, P = 0.0001) and
Fig. 4 Custom-built equipment to measure PWV by means of photoplethysmography. a The equipment consisted of a computer with a data
acquisition card and a signal conditioning module for ECG and photoplethysmography signals. b Inner view of the signal conditioning module
Cardiovasc Eng
123
waist (r = 0.60, P = 0.0001). The lack of correlation
between age and baPWV was unexpected, since baPWV
assesses arterial stiffness and this should correlate with age.
We therefore looked for confusion variables in the ana-
lyzed sample (i.e., variables which distort the relationship
between baPWV and age) and we repeated the correlation
analysis but excluding volunteers who smoked more than
Fig. 5 Typical waveforms
measured by the custom-built
equipment: ECG and two
photoplethysmography signals
at index finger and big toe. The
pulse transit time (PTT)
measured between index finger
and big toe was used to
calculate ftPWV (a), PTT
between heart (wave R of ECG)
and index finger to calculate
hfPWV (b), and PTT between
heart and big toe to calculate
htPWV (c)
Table 1 Basic characteristics of the healthy subjects who took part in
the first phase of the study (n = 47)
Characteristic Mean SD
Cardiovascular parameters SBP (mmHg) 114.37 11.70
DBP (mmHg) 67.93 6.83
MBP (mmHg) 82.86 9.00
PP (mmHg) 46.37 8.06
HR (bpm) 62.00 8.38
Risk factors Age (years) 33.78 9.04
BMI (kg/m2) 24.14 4.12
WHR (cm/cm) 0.86 0.10
Waist (cm) 85.38 13.36
SBP systolic blood pressure, DBP diastolic blood pressure, MBPmean blood pressure, PP pulse pressure, HR heart rate, BMI body
mass index, and WHR waist to hip ratio
Table 2 Correlation coefficients (r) and statistical significance (P)
between cardiovascular parameters, risk factors, brachial-ankle pulse
wave velocity (baPWV), stiffness index (SI) and reflection index (RI)
in the first phase of the study (47 healthy volunteers) (significant
values are highlighted in bold)
baPWV SI RI
r P r P r P
SBP 0.57 0.0001 0.41 0.004 0.07 [0.05
DBP 0.55 0.0001 0.40 0.005 0.16 [0.05
MBP 0.60 0.0001 0.44 0.002 0.13 [0.05
PP 0.35 0.018 0.24 [0.05 -0.04 [0.05
HR 0.36 0.014 0.29 0.046 -0.47 0.001
RI 0.10 [0.05 0.40 0.005
SI 0.55 0.0001 0.40 0.005
Age 0.28* [0.05* 0.56 0.0001 0.13 [0.05
BMI 0.48 0.001 0.19 [0.05 0.16 [0.05
WHR 0.58 0.0001 0.33 0.026 0.14 [0.05
Waist 0.60 0.0001 0.35 0.017 0.22 [0.05
SBP systolic blood pressure, DBP diastolic blood pressure, MBPmean blood pressure, PP pulse pressure, HR heart rate, BMI body
mass index, and WHR waist to hip ratio
* Correlation between age and baPWV became significant (r = 0.37,
P \ 0.05) when 11 smokers and BMI [ 30 kg/m2 volunteers were
excluded from the analysis
Cardiovasc Eng
123
three cigarettes a day and 11 subjects with BMI [ 30 kg/
m2. In this case, we did find a significant relationship
between baPWV and age, with a correlation coefficient
r = 0.37 (P \ 0.05).
In the multiple regression analyses for SBP and DBP,
using baPWV, the best models included baPWV and the
waist to hip ratio (WHR) as independent variables. The
regression models were:
SBP ¼ 25:43 þ 83:10�WHR þ 2:1� baPWV
DBP ¼ 26:33 þ 32:70�WHR þ 1:68� baPWV
with correlation coefficients of r = 0.838 and r = 0.673,
respectively.
Assessment of DVP (SI and RI)
The stiffness index derived from DVP (SI) was positively
and significantly correlated with moderately high correla-
tion coefficients with SBP (r = 0.41, P = 0.004), DBP
(r = 0.40, P = 0.005), MBP (r = 0.44, P = 0.002) and
with a low correlation coefficient with HR (r = 0.29,
P \ 0.05). SI was also positively and significantly corre-
lated with the risk factors WHR (r = 0.33, P \ 0.03),
waist (r = 0.35, P \ 0.02) and age (r = 0.56, P \ 0.001),
but did with BMI. In contrast, RI did not correlate either
with blood pressure parameters (SBP, MBP DBP) or risk
factors (BMI, WHR, waist and age) (see Table 2). RI was
negatively and significantly correlated only with HR
(r = -0.47, P = 0.001). The results also showed that
baPWV was positively and significantly correlated with SI
(r = 0.55, P = 0.001), but not with RI. On the other hand,
SI was positively and significantly correlated with RI
(r = 0.40, P = 0.005).
In the multiple regression analyses for SBP and DBP,
using SI and RI, the best models included SI and the waist
to hip ratio (WHR) as independent variables. The regres-
sion models were:
SBP ¼ 17:84 þ 72:23�WHR þ 2:7� SI
DBP ¼ 19:17 þ 42:03�WHR þ 2:1� SI
with correlation coefficients of r = 0.852 and r = 0.663,
respectively.
Second Phase
For the second phase of the study, we designed special
equipment to measure PWV by photoplethysmography.
We conducted a validation test against the Pulse Trace
PCA and PWV by measuring the parameters SI and transit
time (TT), respectively. The correlation coefficients
between the measurements from both systems were 0.88
(P = 0.004) for SI, and 0.744 (P = 0.034) for TT. Table 3
shows the basic characteristics of the 10 healthy volunteers
who took part in the second part of the study. Table 4
shows the values of the signals obtained in the study
measured by Pulse Trace PCA and PWV (Micro Medical,
Kent, UK) and by the custom-built equipment.
In the multiple regression analyses for SBP, using the
variables measured by the custom-built equipment, the best
model included hfPWV, the parameter waist to hip ratio
(WHR), and heart rate (HR) as independent variables. The
regression model was:
SBP ¼ 26:93 þ 51:49�WHR þ 15:82� hfPWV
� 0:375� HR
with a correlation coefficient r = 0.997. In contrast, the
best model for DBP only included the waist to hip ratio
(WHR) as an independent variable:
DBP ¼ 36:89 þ 38:90�WHR
with a correlation coefficient r = 0.758.
Discussion
Assessment of the Brachial-Ankle Pulse Wave Velocity
(baPWV)
The baPWV value obtained in our study (8.40 ± 1.42 m/s)
was lower than those previously reported: 11.08 ± 1.87 m/
s (Shiotani et al. 2005), 11.15 ± 1.04 m/s (Naidu et al.
2005), 11.04 ± 1.26 m/s (Liu et al. 2005), and
16.23 ± 0.93 m/s (Nakamura et al. 2003). In these studies,
baPWV was measured by techniques completely different
to ours: while Shiotani et al. (2005), Liu et al. (2005), and
Nakamura et al. (2003) used a volume-plethysmographic
system, and Naidu et al. (2005) employed a custom-made
Table 3 Basic characteristics of the healthy volunteers considered in
the second phase of the study (n = 10)
Characteristic Mean SD
Cardiovascular parameters SBP (mmHg) 111.25 12.45
DBP (mmHg) 69.37 6.80
MBP (mmHg) 78.87 16.59
PP (mmHg) 41.80 8.02
HR (bpm) 62.00 8.38
Risk factors Age (yrs) 30.22 7.06
BMI (kg/m2) 23.05 4.08
WHR (cm/cm) 0.83 0.12
Waist (cm) 81.67 16.58
SBP systolic blood pressure, DBP diastolic blood pressure, MBP,
mean blood pressure, PP pulse pressure, HR heart rate, BMI body
mass index, and WHR waist to hip ratio
Cardiovasc Eng
123
device based on the oscillometric method. The discrepancy
between our absolute baPWV value and those from the
previous studies could be due to the different measuring
techniques.
In the relationship between baPWV and the cardiovas-
cular parameters (see Table 2), we observed a moderately
high correlation between baPWV and BP (r = 0.57 with
SBP, and r = 0.55 with DBP). These values are in general
inside the range of correlation coefficients previously
reported for MBP by Shiotani et al. (2005) (r = 0.65,
P \ 0.001), and for SBP by Nakamura et al. (2003)
(r = 0.643, P \ 0.0001) and Liu et al. (2005) (r = 0.32,
P \ 0.05).
The correlations found between baPWV and risk factors
in the healthy volunteers in our study were partially in
disagreement with previous studies. We observed a mod-
erate correlation between baPWV and BMI (r = 0.48),
which is in agreement with Liu et al. (2005) (r = 0.46,
P \ 0.05), but not with Shiotani et al. (2005), who did not
find any correlation. We also found a significant correlation
with WHR (r = 0.58), which had not been reported pre-
viously. Initially, we did not find a significant correlation
between baPWV and age. However, since previous studies
on healthy volunteers had found high correlations between
baPWV and age (Nakamura et al. 2003; Liu et al. 2005),
we repeated the correlation analysis but excluding 11
smokers and BMI [ 30 kg/m2 volunteers. In this case, we
did find a significant relationship between baPWV and age,
with a correlation coefficient r = 0.34. Since our popula-
tion was composed of 70% males, our results are in good
agreement with those published by Im et al. (2007), who
found r = 0.30 in men and r = 0.11 in women. Our results
also confirm that smoker and BMI [ 30 kg/m2 status could
possibly be confusion variables in this type of analysis, as
has previously been suggested (McVeigh et al. 1997; Yufu
et al. 2007).
baPWV as an Estimator of Blood Pressure
To our knowledge, this is the first study where the baPWV
signal was employed to estimate SBP and DBP. However,
our findings are comparable to those obtained in similar
studies. For instance, Millasseau et al. (2002) reported a
regression model to infer the carotid-femoral PWV
(cfPWV) by using age and mean blood pressure (MBP), and
found a correlation coefficient of r = 0.71 (P = 0.0001).
Asmar et al. (1995) also reported a regression model with
healthy volunteers in which cfPWV was explained by age
and SBP (r = 0.685, P = 0.001). On the other hand, Kubo
et al. (2002) reported different models to explain baPWV by
using age and MBP in cardiac patients. Our results are
partially in agreement with those obtained by Millasseau
et al. (2002) and Kubo et al. (2002) since the highest cor-
relation coefficient found in our study was also for MBP
(r = 0.60) (see Table 2). However, our interest is focused
on models that explain SBP and DBP by using other car-
diovascular parameters, and in this respect, our findings are
hence new.
Assessment of the Digital Volume Pulse (DVP)
The SI value found in our study (6.28 ± 0.83 m/s) was
lower than those previously reported by Millasseau et al.
(2002) (8.4 m/s) and by Alty et al. (2007) (9.4 m/s), both
on healthy subjects. Since age is known to affect SI, this
disagreement could be due to age differences between our
subjects (33.78 ± 9.04 years) and those studied by
Millasseau et al. (2002) (47 ± 13.8 years) and Alty et al.
(2007) (50 ± 13.6 years).
We observed a moderate correlation between SI and
blood pressure (r = 0.41 for SBP, r = 0.40 for DBP and
r = 0.44 for MBP), which is in agreement with that found
by Millasseau et al. (2003) (r = 0.32 for SBP, r = 0.48 for
DBP and r = 0.45 for MBP) in a group of healthy subjects.
Concerning the relationship between SI and risk factors
in healthy volunteers (see Table 2), our results showed a
moderately high correlation between SI and age (r = 0.56),
which was similar to that previously reported by Millas-
seau et al. (2003) (r = 0.63). We also observed that SI
presented a low correlation with WHR (r = 0.33), and was
slightly higher with waist measurement (r = 0.35). Finally,
we did not find any correlation between SI and BMI. These
relationships had not been studied previously.
Mean RI value found by us in healthy volunteers was
68.53 ± 8.07%, which is in agreement with that reported
by Chowienczyk et al. (1999), also in healthy volunteers
(60.00 ± 5.50%). In our study RI did not correlate either
Table 4 Values of the signals obtained in the second phase of the
study on 10 healthy volunteers measured by both Pulse Trace PWV
(Micro Medical, Kent, UK) and custom-built equipment
Measuring equipment Parameter Mean SD
Pulse trace PCA and PWV
(Micro Medical, Kent, UK)
baPWV (m/s) 7.91 0.89
SI (m/s) 6.13 0.76
RI (%) 67.73 10.52
Custom-built equipment ftPWV (m/s) 6.94 1.20
hfPWV (m/s) 4.12 0.49
htPWV (m/s) 4.97 0.59
SI* (m/s) 6.15 1.20
RI* (%) 67.32 9.06
BaPWV brachial-ankle pulse wave velocity, SI stiffness index, RIreflection index (these two parameters measured from Pulse Trace
PWV of Micro Medical, Kent, UK), ftPWV index finger-big toe pulse
wave velocity, hfPWV heart-index finger pulse wave velocity, htPWVheart- big toe pulse wave velocity, SI* stiffness index; RI* reflection
index (both parameters measured from the custom-built equipment)
Cardiovasc Eng
123
with SBP, DBP, MBP or with PP (see Table 2), which is
also in agreement with Millasseau et al. (2003). We also
found a significant negative correlation between RI and HR
(r = -0.473, P = 0.001), which has also been previously
reported by Chowienczyk et al. (1999) and Millasseau et al.
(2003).
Finally, even though the relationship between baPWV
and SI had not been studied previously, the correlation we
found (r = 0.55, P = 0.0001) was similar to that found by
Millasseau et al. (2002) between carotid-femoral PWV
(cfPWV) and SI (r [ 0.65, P \ 0.01). The relationship
between PWV and arterial stiffness has been previously
discussed (Hamilton et al. 2007). In addition, by using the
Moens–Kortweg formula, it is know that PWV is propor-
tional to the square root of the elastic modulus of the
arterial wall. However, it has been also pointed out that the
square root relationship also means that a change in PWV
is not a particularly sensitive measure of change in physical
arterial properties (Hughes et al. 2004).
DVP as an Estimator of Blood Pressure
Only the SI parameter derived from DVP was found to be
useful in the multiple regression models. Once more, and
although there are no previous studies with which to
compare our SBP and DBP models, we should point out
that Millasseau et al. (2002) reported a model with a high
correlation coefficient (r = 0.69, P = 0.0001) in which SI
was explained both by age and by MBP. This is in close
agreement with our experimental results, in which the
highest correlation coefficients were between SI and age,
and between SI and MBP.
hfPWV as an Estimator of Blood Pressure
First of all, we found a high correlation coefficient
(r = 0.88) between the SI measured by the custom-built
equipment and that measured by the Pulse Trace PCA
(Micro Medical, Kent, UK), which suggests that both the
transmittance photoplethysmography technique (in the
custom-built equipment) and the reflectance technique
(Pulse Trace PCA) offer similar results. Moreover, the
mean ftPWV value measured in our study (6.94 ± 1.20 m/
s) was similar to those reported in healthy volunteers by
Tsai et al. (2005) (6.39 ± 0.93 m/s) and Chen et al. (2004)
(6.49 ± 0.92 m/s).
Secondly, in the multiple regression analyses for SBP
the best model included hfPWV, waist to hip ratio (WHR),
and heart rate (HR) as independent variables. The high
correlation coefficient (r = 0.997) given by this model
suggests that hfPWV could be a good estimator of the beat-
to-beat SBP. hfPWV could therefore be useful, for
instance, for monitoring cardiac patients during physical
exercise in cardiac rehabilitation. Unfortunately, since
DBP was only explained by the WHR, this means that we
did not find any signal that could potentially be used to
estimate DBP. Finally, future work should be conducted
with subjects undergoing physical exercise to assess the
reliability of our current results.
Conclusions
The findings of this study suggest that:
(1) Both the baPWV and SI signals (measured from
DVP) are good estimators of SBP and DBP. In
contrast, RI (measured from DVP) does not signifi-
cantly correlate with any blood pressure parameter.
(2) Estimation of SBP and DBP in multiple regression
analysis is improved by including the waist to hip
ratio (WHR). In this case, similar correlation coeffi-
cients are obtained from baPWV and SI.
(3) Results are improved when new variables measured
by the custom-built equipment are considered in the
multiple regression analyses. The best model to
explain the SBP (r = 0.997) includes the hfPWV
(PWV measured by photoplethysmography between
R-wave of ECG and index finger), along with the
waist to hip ratio (WHR) and heart rate (HR) as
independent variables. DBP could not be explained
by any signal from the custom-built equipment.
(4) These results suggest that hfPWV could be a good
estimator of the beat-to-beat SBP, which could be
very useful for monitoring cardiac patients during
physical exercise in cardiac rehabilitation.
Acknowledgments The translation of this paper was funded by the
Universidad Politecnica de Valencia, Spain. This study was partially
supported by Merce V. Electromedicina (Valencia, Spain) and by
‘‘IMIDT-Programa de Investigacion y Desarrollo Tecnologico’’ of the
‘‘Instituto de la Mediana y Pequena Industria de la Comunidad Val-
enciana (IMPIVA)’’.
References
Alty SR, Angarita-Jaimes N, Millasseau SC, Chowienczyk PJ.
Predicting arterial stiffness from the digital volume pulse
waveform. IEEE Trans Biomed Eng. 2007;54:2268–75.
Asmar R, Benetos A, Topouchian J, Laurent P, Pannier B, Brisac AM,
et al. Assessment of arterial distensibility by automatic pulse
wave velocity measurement. Validation and clinical application
studies. Hypertension. 1995;26:485–90.
Chen W, Kobayashi T, Ichikawa S, Takeuchi Y, Togawa T.
Continuous estimation of systolic blood pressure using the pulse
arrival time and intermittent calibration. Med Biol Eng Comput.
2000;38:569–74.
Chen YK, Wu HT, Chi CK, Tsai WC, Chen JY, Wang MC. A new dual
channel pulse wave velocity measurement system. Bioinformatics
Cardiovasc Eng
123
and Bioengineering, 2004. BIBE 2004. Proceedings. Fourth IEEE
symposium on volume, 19–21 May 2004: 17–21.
Chowienczyk PJ, Kelly RP, MacCallum H, Millasseau SC, Andersson
TL, Gosling RG, et al. Photoplethysmographic assessment of
pulse wave reflection: blunted response to endothelium-depen-
dent beta2-adrenergic vasodilation in type II diabetes mellitus. J
Am Coll Cardiol. 1999;34:2007–14.
European Society of Cardiology and Hypertension. European Society
of Hypertension-European Society of Cardiology guidelines for
the management of arterial hypertension. J Hypertens.
2003;21:1011–53.
Geddes LA, Voelz MH, Babbs CF, Bourland JD, Tacker WA. Pulse
transit time as an indicator of arterial blood pressure. Psycho-
physiology. 1981;18:71–4.
Hamilton PK, Lockhart CJ, Quinn CE, McVeigh GE. Arterial
stiffness: clinical relevance, measurement and treatment. Clin
Sci (Lond). 2007;113:157–70.
Hughes SM, Dixon LJ, McVeigh GE. Arterial stiffness and pulse
wave velocity: problems with terminology. Circulation.
2004;109:e3. (author reply e3).
Im JA, Lee JW, Shim JY, Lee HR, Lee DC. Association between
brachial-ankle pulse wave velocity and cardiovascular risk
factors in healthy adolescents. J Pediatr. 2007;150:247–51.
Kubo T, Miyata M, Minagoe S, Setoyama S, Maruyama I, Tei C. A
simple oscillometric technique for determining new indices of
arterial distensibility. Hypertens Res. 2002;25:351–8.
Lehmann ED. Clinical value of aortic pulse-wave velocity measure-
ment. Lancet. 1999;354:528–9.
Liu PY, Tsai WC, Lin CC, Hsu CH, Haung YY, Chen JH. Invasive
measurements of pulse wave velocity correlate with the degree
of aortic valve calcification and severity associated with matrix
metalloproteinases in elderly patients with aortic valve stenosis.
Clin Sci (Lond). 2004;107:415–22.
Liu H, Yambe T, Zhang X, Saijo Y, Shiraishi Y, Sekine K, et al.
Milyagin VA and Nitta S Comparison of brachial-ankle pulse
wave velocity in Japanese and Russians. Tohoku J Exp Med.
2005;207:263–70.
McVeigh GE, Morgan DJ, Finkelstein SM, Lemay LA, Cohn JN.
Vascular abnormalities associated with long-term cigarette
smoking identified by arterial waveform analysis. Am J Med.
1997;102:227–31.
Millasseau SC, Kelly RP, Ritter JM, Chowienczyk PJ. Determination
of age-related increases in large artery stiffness by digital pulse
contour analysis. Clin Sci (Lond). 2002;103:371–7.
Millasseau SC, Kelly RP, Ritter JM, Chowienczyk PJ. The vascular
impact of aging and vasoactive drugs: comparison of two digital
volume pulse measurements. Am J Hypertens. 2003;16:467–72.
Naidu MU, Reddy BM, Yashmaina S, Patnaik AN, Rani PU. Validity
and reproducibility of arterial pulse wave velocity measurement
using new device with oscillometric technique: a pilot study.
Biomed Eng. 2005;4:49. (Online Aug 23, 2005).
Nakamura U, Iwase M, Nohara S, Kanai H, Ichikawa K, Iida M.
Usefulness of brachial-ankle pulse wave velocity measurement:
correlation with abdominal aortic calcification. Hypertens Res.
2003;26:163–7.
Nieminen T, Leino J, Maanoja J, Nikus K, Viik J, Lehtimaki T, et al.
The prognostic value of haemodynamic parameters in the
recovery phase of an exercise test. The finnish cardiovascular
study. J Hum Hypertens. 2008;22:537–43.
Pierson LM, Miller LE, Herbert WG. Predicting exercise training
outcome from cardiac rehabilitation. J Cardiopulm Rehabil.
2004;24:113–8.
Shiotani A, Motoyama M, Matsuda T, Miyanishi T. Brachial-ankle
pulse wave velocity in Japanese university students. Intern Med.
2005;44:696–701.
Tsai WC, Chen JY, Wang MC, Wu HT, Chi CK, Chen YK, et al.
Association of risk factors with increased pulse wave velocity
detected by a novel method using dual-channel photoplethys-
mography. Am J Hypertens. 2005;18:1118–20.
Yamashina A, Tomiyama H, Takeda K, Tsuda H, Arai T, Hirose K,
et al. Validity, reproducibility, and clinical significance of
noninvasive brachial-ankle pulse wave velocity measurement.
Hypertens Res. 2002;25:359–64.
Yufu K, Takahashi N, Hara M, Saikawa T, Yoshimatsu H.
Measurement of the brachial-ankle pulse wave velocity and
flow-mediated dilatation in young, healthy smokers. Hypertens
Res. 2007;30:607–12.
Cardiovasc Eng
123