AC BRIDGES - Government Polytechnic Deogarh
-
Upload
khangminh22 -
Category
Documents
-
view
1 -
download
0
Transcript of AC BRIDGES - Government Polytechnic Deogarh
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
AC BRIDGES
2.1 General form of A.C. bridge
AC bridge are similar to D.C. bridge in topology(way of connecting).It consists of four arm
AB,BC,CD and DA .Generally the impedance to be measured is connected between ‘A’ and ‘B’.
A detector is connected between ‘B’ and ’D’. The detector is used as null deflection instrument.
Some of the arms are variable element. By varying these elements, the potential values at ‘B’ and
‘D’ can be made equal. This is called balancing of the bridge.
Fig. 2.1 General form of A.C. bridge
At the balance condition, the current through detector is zero.
4
.
2
.
3
.
1
.
II
II
=
=∴
4
.
3
.
2
.
1
.
I
I
I
I=∴ (2.1)
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
At balance condition,
Voltage drop across ‘AB’=voltage drop across ‘AD’.
2
.
1
.
EE =
(2.2)
Similarly, Voltage drop across ‘BC’=voltage drop across ‘DC’
4
.
3
.
EE =
4
.
4
.
3
.
3
.
ZIZI =∴ (2.3)
From Eqn. (2.2), we have
1
.
2
.
2
.
1
.
Z
Z
I
I=∴ (2.4)
From Eqn. (2.3), we have
3
.
4
.
4
.
3
.
Z
Z
I
I=∴ (2.5)
From equation -2.1, it can be seen that, equation -2.4 and equation-2.5 are equal.
3
.
4
.
1
.
2
.
Z
Z
Z
Z=∴
3
.
2
.
4
.
1
.
ZZZZ =∴
Products of impedances of opposite arms are equal.
33224411 θθθθ ∠∠=∠∠∴ ZZZZ
32324141 θθθθ +∠=+∠⇒ ZZZZ
41 ZZ = 32 ZZ
3241 θθθθ +=+
2
.
2
.
1
.
1
.
ZIZI =∴
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
∗ For balance condition, magnitude on either side must be equal.
∗ Angle on either side must be equal.
Summary
For balance condition,
• 3
.
1
.
II = , 4
.
2
.
II =
• 41 ZZ = 32 ZZ
• 3241 θθθθ +=+
• 2
.
1
.
EE = & 4
.
3
.
EE =
2.2 Types of detector
The following types of instruments are used as detector in A.C. bridge.
• Vibration galvanometer
• Head phones (speaker)
• Tuned amplifier
2.2.1 Vibration galvanometer
Between the point ‘B’ and ‘D’ a vibration galvanometer is connected to indicate the bridge
balance condition. This A.C. galvanometer which works on the principle of resonance. The
A.C. galvanometer shows a dot, if the bridge is unbalanced.
2.2.2 Head phones
Two speakers are connected in parallel in this system. If the bridge is unbalanced, the
speaker produced more sound energy. If the bridge is balanced, the speaker do not produced
any sound energy.
2.2.3 Tuned amplifier
If the bridge is unbalanced the output of tuned amplifier is high. If the bridge is balanced,
output of amplifier is zero.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
2.3 Measurements of inductance
2.3.1 Maxwell’s inductance bridge
The choke for which R1 and L1 have to measure connected between the points ‘A’ and
‘B’. In this method the unknown inductance is measured by comparing it with the standard
inductance.
Fig. 2.2 Maxwell’s inductance bridge
L2 is adjusted, until the detector indicates zero current.
Let R1= unknown resistance
L1= unknown inductance of the choke.
L2= known standard inductance
R1,R2,R4= known resistances.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.3 Phasor diagram of Maxwell’s inductance bridge
At balance condition, 3
.
2
.
4
.
1
.
ZZZZ =
322411 )()( RjXLRRjXLR +=+
322411 )()( RjwLRRjwLR +=+
32324141 RjwLRRRjwLRR +=+
Comparing real part,
3241 RRRR =
4
321
R
RRR =∴ (2.6)
Comparing the imaginary parts,
3241 RwLRwL =
4
321
R
RLL = (2.7)
Q-factor of choke, 324
432
1
1
RRR
RRWL
R
WLQ ==
2
2
R
WLQ = (2.8)
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Advantages
Expression for R1 and L1 are simple.
Equations area simple
They do not depend on the frequency (as w is cancelled)
R1 and L1 are independent of each other.
Disadvantages
Variable inductor is costly.
Variable inductor is bulky.
2.3.2 Maxwell’s inductance capacitance bridge
Unknown inductance is measured by comparing it with standard capacitance. In this bridge,
balance condition is achieved by varying ‘C4’.
Fig 2.4 Maxwell’s inductance capacitance bridge
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
At balance condition, Z1Z4=Z3Z2 (2.9)
44
44
444 1
1
1||
jwCR
jwCR
jwCRZ
+
×
==
44
4
44
44
11 CjwR
R
CjwR
RZ
+=
+= (2.10)
∴Substituting the value of Z4 from eqn. (2.10) in eqn. (2.9) we get
3244
411
1)( RR
CjwR
RjwLR =
+×+
Fig 2.5 Phasor diagram of Maxwell’s inductance capacitance bridge
)1()( 4432411 CjwRRRRjwLR +=+
3244324141 RRRjwCRRRjwLRR +=+
Comparing real parts,
3241 RRRR =
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
4
321
R
RRR =⇒ (2.11)
Comparing imaginary part,
324441 RRRwCRwL =
3241 RRCL =
(2.12)
Q-factor of choke,
32
4324
1
1
RR
RRRCw
R
WLQ ××==
44RwCQ = (2.13)
Advantages
Equation of L1 and R1 are simple.
They are independent of frequency.
They are independent of each other.
Standard capacitor is much smaller in size than standard inductor.
Disadvantages
Standard variable capacitance is costly.
It can be used for measurements of Q-factor in the ranges of 1 to 10.
It cannot be used for measurements of choke with Q-factors more than 10.
We know that Q =wC4R4
For measuring chokes with higher value of Q-factor, the value of C4 and R4 should be
higher. Higher values of standard resistance are very expensive. Therefore this bridge cannot be
used for higher value of Q-factor measurements.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
2.3.3 Hay’s bridge
Fig 2.6 Hay’s bridge
11111
.
XjIRIE +=
=.
E 3
.
1
.
EE +
4
44
.
44
.
jwC
IRIE +=
333
.
RIE =
4
44
444
11
jwC
CjwR
jwCRZ
+=+=
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.7 Phasor diagram of Hay’s bridge
At balance condition, Z1Z4=Z3Z2
324
4411 )
1)(( RR
jwC
CjwRjwLR =
++
3424411 )1)(( RCjwRCjwRjwLR =++
32444122
11441 RRjwCRCLwjjwLRRjwCR =+++
32411444412
1 )()( RRjwCwLRRwCjRCLwR =++−
Comparing the real term,
04412
1 =− RCLwR
4412
1 RCLwR = (2.14)
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Comparing the imaginary terms,
3241144 RRwCwLRRwC =+
3241144 RRCLRRC =+
1443241 RRCRRCL −= (2.15)
Substituting the value of R1 fro eqn. 2.14 into eqn. 2.15, we have,
4412
443241 RCLwRCRRCL ×−=
24
241
23241 RCLwRRCL −=
3242
42
412
1 )1( RRCRCLwL =+
24
241
2324
11 RCLw
RRCL
+= (2.16)
Substituting the value of L1 in eqn. 2.14 , we have
24
24
2432
24
2
11 RCw
RRRCwR
+= (2.17)
3242
42
24
24
2
24
24
2324
1
1 1
1 RRRCw
RCw
RCw
RRCw
R
wLQ
+×
+
×==
44
1
RwCQ = (2.18)
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Advantages
Fixed capacitor is cheaper than variable capacitor.
This bridge is best suitable for measuring high value of Q-factor.
Disadvantages
Equations of L1and R1 are complicated.
Measurements of R1 and L1 require the value of frequency.
This bridge cannot be used for measuring low Q- factor.
2.3.4 Owen’s bridge
Fig 2.8 Owen’s bridge
11111 XjIRIE +=
I4 leads E4 by 900
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
=.
E 3
.
1
.
EE +
2
2222
.
jwC
IRIE +=
Fig 2.9 Phasor diagram of Owen’s bridge
Balance condition, 3
.
2
.
4
.
1
.
ZZZZ =
2
22
222
11
jwC
RjwC
jwCRZ
+=+=
2
322
411
)1(1)(
jwC
RCjwR
jwCjwLR
×+=×+∴
)1()( 2243112 CjwRCRjwLRC +=+
4322432121 CRCjwRCRCjwLCR +=+
Comparing real terms,
4321 CRCR =
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
2
431
C
CRR =
Comparing imaginary terms,
432221 CRCwRCwL =
4321 CRRL =
Q- factor =43
2432
1
1
CR
CCRwR
R
WL=
22CwRQ =
Advantages
Expression for R1 and L1 are simple.
R1 and L1 are independent of Frequency.
Disadvantages
The Circuits used two capacitors.
Variable capacitor is costly.
Q-factor range is restricted.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
2.3.5 Anderson’s bridge
Fig 2.10 Anderson’s bridge
111111
.
)( XjIrRIE ++=
CEE =3
CC ErIE +=.
4
.
CIII += 42
−−−
=+ EEE 42
−−−
=+ EEE 31
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.11 Phasor diagram of Anderson’s bridge
Step-1 Take I1 as references vector .Draw 111RI in phase with I1
)( 1111 rRR += , 11XI is r⊥ to 1
11RI
111111 XjIRIE +=
Step-2 31 II = , 3E is in phase with 3I , From the circuit ,
CEE =3 , CI leads CE by 900
Step-3 CC ErIE +=4
Step-4 Draw 4I in phase with 4E , By KCL , −−−
+= CIII 42
Step-5 Draw E2 in phase with I2
Step-6 By KVL , −−−
=+ EEE 31 or −−−
=+ EEE 42
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.12 Equivalent delta to star conversion for the loop MON
)(11 4
4
4
47
rRjwC
rjwCR
jwcrR
rRZ
++=
++
×=
)(11
1
4
4
4
4
6rRjwC
R
jwcrR
jwCR
Z++
=++
×=
))(1
()(1
)(4
423
4
41
11
rRjwC
rjwCRRR
rRjwC
RjwLR
+++=
++×+
++
+++=
++
+⇒
)(1
))(1(
)(1
)(
4
4423
4
4111
rRjwC
jwCrRrRjwCRR
rRjwC
RjwLR
344323241411 )( RjwCrRRrRjCwRRRRjwLRR +++=+⇒
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.13 Simplified diagram of Anderson’s bridge
Comparing real term,
32411 RRRR =
32411 )( RRRrR =+
14
321 r
R
RRR −=
Comparing the imaginary term,
4343241 )( RwcrRRrRwCRRwL ++=
rCRRrR
CRRL 34
4
321 )( ++=
++= rRr
R
RCRL )( 4
4
231
Advantages
Variable capacitor is not required.
Inductance can be measured accurately.
R1 and L1 are independent of frequency.
Accuracy is better than other bridges.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Disadvantages
Expression for R1 and L1 are complicated.
This is not in the standard form A.C. bridge.
2.4 Measurement of capacitance and loss angle. (Dissipation factor)
2.4.1 Dissipation factors (D)
A practical capacitor is represented as the series combination of small resistance and
ideal capacitance.
From the vector diagram, it can be seen that the angle between voltage and current is slightly less
than 900. The angle ‘δ ’ is called loss angle.
Fig 2.14 Condensor or capacitor
Fig 2.15 Representation of a practical capacitor
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.16 Vector diagram for a practical capacitor
A dissipation factor is defined as ‘tanδ ’.
wCRX
R
IX
IR
CC
===∴ δtan
wCRD =
QD
1=
1cos
sintan
δδδ
δ ≅==D For small value of ‘δ ’ in radians
≅≅ δD Loss Angle (‘δ ’ must be in radian)
2.4.2 Desauty’s Bridge
C1= Unknown capacitance
At balance condition,
32
41
11R
jwCR
jwC×=×
2
3
1
4
C
R
C
R=
3
241
R
CRC =⇒
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.17 Desauty’s bridge
Fig 2.18 Phasor diagram of Desauty’s bridge
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
2.4.3 Modified desauty’s bridge
Fig 2.19 Modified Desauty’s bridge
Fig 2.20 Phasor diagram of Modified Desauty’s bridge
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
)( 1111 rRR +=
)( 2212 rRR +=
At balance condition, )1
()1
(2
1234
1
11
jwCRRR
jwCR +=+
)2
3123
1
44
11
jwC
RRR
jwC
RRR +=+
Comparing the real term, 1234
11 RRRR =
4
1231
1R
RRR =
4
3221
)(
R
RrRrR
+=+
Comparing imaginary term,
2
3
1
4
wC
R
wC
R=
3
241
R
CRC =
Dissipation factor D=wC1r1
Advantages
r1 and c1 are independent of frequency.
They are independent of each other.
Source need not be pure sine wave.
2.4.4 Schering bridge
41111 XjIrIE −=
C2 = C4= Standard capacitor (Internal resistance=0)
C4= Variable capacitance.
C1= Unknown capacitance.
r1= Unknown series equivalent resistance of the capacitor.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
R3=R4= Known resistor.
Fig 2.21 Schering bridge
1
11
111
11
jwC
rjwC
jwCrZ
+=+=
44
4
44
44
411
1
RjwC
R
jwCR
jwCR
Z+
=+
×=
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.22 Phasor diagram of Schering bridge
At balance condition, 3
.
2
.
4
.
1
.
ZZZZ =
2
3
44
4
1
11
1
1
jwC
R
RjwC
R
jwC
rjwC=
+×
+
)1()1( 44132411 rjwCCRCRrjwC +=+
134413241122 CRRjwCCRCRrjwCCR +=+
Comparing the real part,
3
241
R
CRC =∴
Comparing the imaginary part,
14342411 CRRwCCRrwC =
2
341
C
RCr =
Dissipation factor of capacitor,
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
2
34
3
2411
C
RC
R
CRwrwCD ××==
44RwCD =∴
Advantages
In this type of bridge, the value of capacitance can be measured accurately.
It can measure capacitance value over a wide range.
It can measure dissipation factor accurately.
Disadvantages
It requires two capacitors.
Variable standard capacitor is costly.
2.5 Measurements of frequency
2.5.1 Wein’s bridge
Wein’s bridge is popularly used for measurements of frequency of frequency. In this bridge, the
value of all parameters are known. The source whose frequency has to measure is connected as
shown in the figure.
1
11
111
11
jwC
rjwC
jwCrZ
+=+=
22
22
1 RjwC
RZ
+=
At balance condition, 3
.
2
.
4
.
1
.
ZZZZ =
322
24
1
11
1
1R
RjwC
RR
jwC
rjwC×
+=×
+
13242211 )1)(1( jwCRRRRjwCrjwC ×=++
[ ]4
3212121
211221
R
RRjwCRrCCwrjwCRjwC =−++
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.23 Wein’s bridge
Fig 2.24 Phasor diagram of Wein’s bridge
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Comparing real term,
01 21212 =− RrCCw
121212 =RrCCw
2121
2 1
RrCCw =
2121
1
RrCCw = ,
21212
1
RrCCf
Π=
NOTE
The above bridge can be used for measurements of capacitance. In such case, r1 and C1 are
unknown and frequency is known. By equating real terms, we will get R1 and C1. Similarly by
equating imaginary term, we will get another equation in terms of r1 and C1. It is only used for
measurements of Audio frequency.
A.F=20 HZ to 20 KHZ
R.F=>> 20 KHZ
Comparing imaginary term,
4
3211122
R
RRwCrwCRwC =+
4
3211122
R
RRCrCRC =+ …………………………………..(2.19)
21221
1
RrCwC =
Substituting in eqn. (2.19), we have
14
32
2122
122 C
R
RR
RrCw
rRC =+
Multiplying 32
4
RR
R in both sides, we have
132
4
222
32
422
1C
RR
R
RCwRR
RRC =×+×
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
3222
24
3
421
RRCw
R
R
RCC +=
122112 =RCrCw
+
==
3222
24
3
4222
212221
11
RRCw
R
R
RCRCw
CRCwr
+
=
32
4
3
422
22
1
RR
R
R
RRCw
+
=∴
22
22
2
4
31
1
1
RRCw
R
Rr
+=∴
)1
(
1
22
22
24
31
RRCw
R
Rr
2.5.2 High Voltage Schering Bridge
Fig 2.25 High Voltage Schering bridge
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
(1) The high voltage supply is obtained from a transformer usually at 50 HZ.
2.6 Wagner earthing device:
Fig 2.26 Wagner Earthing device
Wagner earthing consists of ‘R’ and ‘C’ in series. The stray capacitance at node ‘B’ and ‘D’ are
CB, CD respectively. These Stray capacitances produced error in the measurements of ‘L’ and
‘C’. These error will predominant at high frequency. The error due to this capacitance can be
eliminated using wagner earthing arm.
Close the change over switch to the position (1) and obtained balanced. Now change the
switch to position (2) and obtained balance. This process has to repeat until balance is achieved
in both the position. In this condition the potential difference across each capacitor is zero.
Current drawn by this is zero. Therefore they do not have any effect on the measurements.
What are the sources of error in the bridge measurements?
Error due to stray capacitance and inductance.
Due to external field.
Leakage error: poor insulation between various parts of bridge can produced this error.
Eddy current error.
Frequency error.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Waveform error (due to harmonics)
Residual error: small inductance and small capacitance of the resistor produce this error.
Precaution
The load inductance is eliminated by twisting the connecting the connecting lead.
In the case of capacitive bridge, the connecting lead are kept apart.(d
rAC
∈∈=
0Q )
In the case of inductive bridge, the various arm are magnetically screen.
In the case of capacitive bridge, the various arm are electro statically screen to reduced
the stray capacitance between various arm.
To avoid the problem of spike, an inter bridge transformer is used in between the source
and bridge.
The stray capacitance between the ends of detector to the ground, cause difficulty in
balancing as well as error in measurements. To avoid this problem, we use wagner
earthing device.
2.7 Ballastic galvanometer
This is a sophisticated instrument. This works on the principle of PMMC meter. The only
difference is the type of suspension is used for this meter. Lamp and glass scale method is used
to obtain the deflection. A small mirror is attached to the moving system. Phosphorous bronze
wire is used for suspension.
When the D.C. voltage is applied to the terminals of moving coil, current flows through it. When
a current carrying coil kept in the magnetic field, produced by permanent magnet, it experiences
a force. The coil deflects and mirror deflects. The light spot on the glass scale also move. This
deflection is proportional to the current through the coil.
t
Qi = , ∫== idtitQ
Q∝θ , deflection ∝ Charge
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.27 Ballastic galvanometer
2.8 Measurements of flux and flux density (Method of reversal)
D.C. voltage is applied to the electromagnet through a variable resistance R1 and a reversing
switch. The voltage applied to the toroid can be reversed by changing the switch from position 2
to position ‘1’. Let the switch be in position ‘2’ initially. A constant current flows through the
toroid and a constant flux is established in the core of the magnet.
A search coil of few turns is provided on the toroid. The B.G. is connected to the search
coil through a current limiting resistance. When it is required to measure the flux, the switch is
changed from position ‘2’ to position ‘1’. Hence the flux reduced to zero and it starts increasing
in the reverse direction. The flux goes from + φ to - φ , in time ‘t’ second. An emf is induced in
the search coil, science the flux changes with time. This emf circulates a current through R2 and
B.G. The meter deflects. The switch is normally closed. It is opened when it is required to take
the reading.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
2.8.1 Plotting the BH curve
The curve drawn with the current on the X-axis and the flux on the Y-axis, is called
magnetization characteristics. The shape of B-H curve is similar to shape of magnetization
characteristics. The residual magnetism present in the specimen can be removed as follows.
Fig 2.28 BH curve
Fig 2.29 Magnetization characteristics
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Close the switch ‘S2’ to protect the galvanometer, from high current. Change the switch
S1 from position ‘1’ to ‘2’ and vice versa for several times.
To start with the resistance ‘R1’ is kept at maximum resistance position. For a particular value of
current, the deflection of B.G. is noted. This process is repeated for various value of current. For
each deflection flux can be calculated.( A
Bφ
= )
Magnetic field intensity value for various current can be calculated.().The B-H curve can be
plotted by using the value of ‘B’ and ‘H’.
2.8.2 Measurements of iron loss:
Let RP= pressure coil resistance
RS = resistance of coil S1
E= voltage reading= Voltage induced in S2
I= current in the pressure coil
VP= Voltage applied to wattmeter pressure coil.
W= reading of wattmeter corresponding voltage V
W1= reading of wattmeter corresponding voltage E
PEW
VW
→
→
1
V
WEW
V
E
W
W ×=⇒= 1
1
W1=Total loss=Iron loss+ Cupper loss.
The above circuit is similar to no load test of transformer.
In the case of no load test the reading of wattmeter is approximately equal to iron loss. Iron loss
depends on the emf induced in the winding. Science emf is directly proportional to flux. The
voltage applied to the pressure coil is V. The corresponding of wattmeter is ‘W’. The iron loss
corresponding E is V
WEE = . The reading of the wattmeter includes the losses in the pressure
coil and copper loss of the winding S1. These loses have to be subtracted to get the actual iron
loss.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
2.9 Galvanometers
D-Arsonval Galvanometer
Vibration Galvanometer
Ballistic C
2.9.1 D-arsonval galvanometer (d.c. galvanometer)
Fig 2.30 D-Arsonval Galvanometer
Galvanometer is a special type of ammeter used for measuring µ A or mA. This is a
sophisticated instruments. This works on the principle of PMMC meter. The only difference is
the type of suspension used for this meter. It uses a sophisticated suspension called taut
suspension, so that moving system has negligible weight.
Lamp and glass scale method is used to obtain the deflection. A small mirror is attached
to the moving system. Phosphors bronze is used for suspension.
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
When D.C. voltage is applied to the terminal of moving coil, current flows through it.
When current carrying coil is kept in the magnetic field produced by P.M. , it experiences a
force. The light spot on the glass scale also move. This deflection is proportional to the current
through the coil. This instrument can be used only with D.C. like PMMC meter.
The deflecting Torque,
TD=BINA
TD=GI, Where G=BAN
TC=KSθ =Sθ
At balance, TC=TD ⇒ Sθ =GI
S
GI=∴θ
Where G= Displacements constant of Galvanometer
S=Spring constant
2.9.2 Vibration Galvanometer (A.C. Galvanometer )
The construction of this galvanometer is similar to the PMMC instrument except for the moving
system. The moving coil is suspended using two ivory bridge pieces. The tension of the system
can be varied by rotating the screw provided at the top suspension. The natural frequency can be
varied by varying the tension wire of the screw or varying the distance between ivory bridge
piece.
When A.C. current is passed through coil an alternating torque or vibration is produced. This
vibration is maximum if the natural frequency of moving system coincide with supply frequency.
Vibration is maximum, science resonance takes place. When the coil is vibrating , the mirror
oscillates and the dot moves back and front. This appears as a line on the glass scale. Vibration
galvanometer is used for null deflection of a dot appears on the scale. If the bridge is unbalanced,
a line appears on the scale
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.31 Vibration Galvanometer
Example 2.2-In a low- Voltage Schering bridge designed for the measurement of
permittivity, the branch ‘ab’ consists of two electrodes between which the specimen under
test may be inserted, arm ‘bc’ is a non-reactive resistor R3 in parallel with a standard
capacitor C3, arm CD is a non-reactive resistor R4 in parallel with a standard capacitor C4,
arm ‘da’ is a standard air capacitor of capacitance C2. Without the specimen between the
electrode, balance is obtained with following values , C3=C4=120 pF, C2=150 pF,
R3=R4=5000Ω.With the specimen inserted, these values become C3=200 pF,C4=1000
pF,C2=900 pF and R3=R4=5000Ω. In such test w=5000 rad/sec. Find the relative
permittivity of the specimen?
Sol: Relative permittivity( rε ) =mediumair with measured ecapacitanc
mediumgiven with measured ecapacitanc
2020
CLASS NOTES ON ELECTRICAL MEASUREMENTS & INSTRUMENTATION
Fig 2.32 Schering bridge
)(3
421
R
RCC =
Let capacitance value C0, when without specimen dielectric.
Let the capacitance value CS when with the specimen dielectric.
pFR
RCC 150
5000
5000150)(
3
420 =×==
pFR
RCCS 900
5000
5000900)(
3
42 =×==
6150
900
0
===C
CSrε
Example 2.3- A specimen of iron stamping weighting 10 kg and having a area of 16.8 cm2 is
tested by an episten square. Each of the two winding S1 and S2 have 515 turns. A.C. voltage
of 50 HZ frequency is given to the primary. The current in the primary is 0.35 A. A
voltmeter connected to S2 indicates 250 V. Resistance of S1 and S2 each equal to 40 Ω.
Resistance of pressure coil is 80 kΩ. Calculate maximum flux density in the specimen and
iron loss/kg if the wattmeter indicates 80 watt?
2020