T 1000 PLUS INTRODUCTORY GUIDE - Userequip.

108
DATE: 27/11/2011 DOC.MIE92093 REV. 1.35 T 1000 PLUS INTRODUCTORY GUIDE

Transcript of T 1000 PLUS INTRODUCTORY GUIDE - Userequip.

DATE: 27/11/2011

DOC.MIE92093 REV. 1.35

T 1000 PLUS INTRODUCTORY GUIDE

DOC. MIE92093 Rev. 1.34 Page 2 of 108

REVISIONS SUMMARY VISA

N PAGE DATE

1 All 20/10/2008 Issued Lodi

1.28 26 2/02/2010 Added the paragraph

test results handling

Lodi

1.28

-1

17, 37 1/10/2010

Added the TD1000

model and the 15 Hz

Lodi

1.34 11 –13;

41

7/9/2011 Revised to issue 1.34;

removed the serial

I/F

Lodi

1.35 39 27/11/2011 Added the FT1000

connection

Lodi

DOC. MIE92093 Rev. 1.34 Page 3 of 108

SHORT FOREWORD ...................................................................................... 5

INTRODUCTION ............................................................................................. 6

1 TEST SET EXPLANATION ...................................................................... 9

1.1 CONNECTION TO THE RELAY AND POWER-ON ........................................... 9 1.2 TEST CONTROL .......................................................................................... 10 1.3 CURRENT GENERATION ............................................................................. 14 1.4 AC VOLTAGE GENERATION FROM MAIN OUTPUT ..................................... 18 1.5 DC VOLTAGE GENERATION FROM MAIN OUTPUT .................................... 18 1.6 AC VOLTAGE GENERATION FROM THE AUXILIARY OUTPUT..................... 19 1.7 DC VOLTAGE GENERATION FROM THE AUXILIARY OUTPUT .................... 23 1.8 AUXILIARY CONTACTS .............................................................................. 24 1.9 THE TIMER ................................................................................................. 24 1.10 FINDING RELAY THRESHOLDS ................................................................ 26

1.10.1. Introduction ........................................................................ 26 1.10.2. First threshold trip and drop-off ............................... 27 1.10.3. Second threshold trip and drop-off......................... 28

1.11 FINDING RELAY TIMINGS ....................................................................... 28 1.12 TEST RESULTS HANDLING ...................................................................... 30 1.13 BASIC TEST PRINCIPLES ........................................................................ 32

1.13.1. Introduction .......................................................................... 32 1.13.2. Parameter vs. time characteristic ........................... 33 1.13.3. Parameter vs. parameter characteristic .............. 34

1.14 USE OF THE TEST SET AS A MULTIMETER ............................................. 37 1.15 TD1000 MODEL: AC CURRENT GENERATION FROM THE AUXILIARY

OUTPUT .............................................................................................................. 39 1.16 FILTERING HEAVY BURDEN DISTORTION WITH THE FT1000 OPTION 40

2 TEST SET AND POP-UP MENU .......................................................... 42

2.1 THE FRONT PANEL ..................................................................................... 42 2.2 DISPLAY AND CONTROL LIGHTS ............................................................... 44 2.3 THE POP-UP MENU .................................................................................... 45

3 WHAT’S INSIDE? .................................................................................... 61

3.1 PHYSICAL DESCRIPTION ........................................................................... 61 3.2 DETAILED FUNCTION DESCRIPTION ......................................................... 65

3.2.1 Main auxiliary transformer, XTF10330 (3) ............. 65 3.2.2 Auxiliary transformer, XTF10245 (4) ........................ 66 3.2.3 Main output transformer, XTF10345 (5) ................. 68 3.2.4 Main front board PWA11398 (17)................................ 68 3.2.5 MISU T1000 board PWA11402 (26) ........................... 68

DOC. MIE92093 Rev. 1.34 Page 4 of 108

3.2.6 CONV T1000 board PWA11401 (16)........................... 69 3.2.7 INTE ON-OFF T 1000 board PWA11400 (7) ........... 70 3.2.8 MICR T 1000 board PWA41300 (15) .......................... 70 3.2.9 AP-50 AC Voltage board PWA11396 (10) ............... 70 3.2.10 Protection fuses ................................................................... 70 3.2.11 Connectors summary ........................................................ 71

4 THE HELL, IT DOESN’T WORK ......................................................... 73

4.1 INTRODUCTION ......................................................................................... 73 4.2 ERROR MESSAGES .................................................................................... 74 4.3 TROUBLE SHOOTING ................................................................................. 76

4.3.1 Auxiliary supplies .................................................................. 79 4.3.2 No power at power-on ........................................................ 79

4.4 AUXILIARY DC VOLTAGE FAULT ..................................................... 80 4.5 AUXILIARY AC VOLTAGE FAULT ............................................................... 83 4.6 NO OUTPUT FROM THE MAIN CURRENT AND VOLTAGE ........................... 84 4.7 DOES NOT MEASURE THE MAIN CURRENT ............................................... 85 4.8 THE BACKLIGHT OR THE DISPLAY DOES NOT OPERATE .......................... 85 4.9 THE AC VOLTAGE MEASUREMENT IS NOT STABLE .................................. 88 4.10 THE TRIP INPUT IS NOT DETECTED OR TIMING ERROR ........................ 88 4.11 PROBLEMS DURING UPGRADE ................................................................ 90 4.12 THE ENCODER IS BROKEN ...................................................................... 91 4.13 THE THERMAL ALARM DOES NOT DISAPPEAR ........................................ 92 4.14 AN AUXILIARY SWITCH DOES NOT OPERATE......................................... 92 4.15 THE FAULT CANNOT BE FIXED ................................................................ 93 4.16 CALIBRATION .......................................................................................... 96

4.16.1 Introduction .......................................................................... 96 4.16.2 Calibration procedure ..................................................... 96 4.16.3 T 1000 PLUS output calibration ................................ 97 4.16.4 T 1000 PLUS external meas. calibration .............. 99 4.16.5 Phase angle measurement calibration ................ 100

APPENDIX 1 SPARE PARTS LIST ..................................................... 105

APPENDIX 2: MENU SELECTIONS FLUX DIAGRAM ................ 108

DOC. MIE92093 Rev. 1.34 Page 5 of 108

SHORT FOREWORD

Dear T 1000 PLUS user,

I often wondered why the user’s manual is not very much used,

even if it includes valuable information. As me too I am a user of

such manuals, the answer I have given myself is that valuable

information are concealed somewhere in the thick thing, and I do

not have time to waste to find it. So, either the manual is actually

of help, or I ignore it.

This is why I decided to split the T 1000 PLUS manual in three:

specification, with all performance details; application manual,

with instructions about how to use it one its operation is

understood; introductory guide, with the device description and

basic information. The idea is that you may read once the

introductory guide or the specification, while you need to follow

application examples more than once; so, why not to split the

manual in three?

Have a good work with T 1000 PLUS!

Primo Lodi

Q&A Manager

DOC. MIE92093 Rev. 1.34 Page 6 of 108

INTRODUCTION

The single phase relay test set mod. T 1000 PLUS is suited for the

testing and adjustments of the following types of relays; the table

lists also the paragraph that explains the test procedure.

Type of relay IEEE code PARAGRAPH

- Distance* 21 1.12

- Synchronizing 25 1.8

- Over/under-voltage 27 - 59 1.2

- Power, varmetric or wattmetric 32 - 92 1.4

- Under current 37 1.1

- Loss of field 40 1.10

- Reverse phase current 46 1.4

- Instantaneous overcurrent 50 1.1

- Ground fault 50N 1.1

- Timed overcurrent 51 1.1

- Power factor 55 1.4

- Directional overcurrent 67 1.5

- Directional ground fault 67N 1.5

- Automatic reclose 79 1.11

- DC voltage 80 1.3

- Frequency 81 1.6

- Frequency rate of change 81 1.7

- Motor protection 86 1.1

- Differential ** 87 1.1

- Directional voltage 91 1.5

- Tripping relay 94 1.9

- Voltage regulation 1.2

- Thermal 1.1

- Timers 1.9

* For distance relays three T 1000 PLUS are necessary.

** Differential starter circuit

In addition to the above, T 1000 PLUS can test:

. Converters: V; I; φ°; p.f.; W; VAr; f., both 0 to 5 and 4 to 20

mA.

. Energy meters, single phase or three phase.

DOC. MIE92093 Rev. 1.34 Page 7 of 108

The instrument contains three separate generators:

. Main generator, which generates either AC current, AC voltage;

DC voltage;

. Auxiliary a.c voltage generator, that generates an independent,

phase shifting a.c voltage;

. Auxiliary DC voltage generator, that generates the DC voltage

that feeds the relay under test.

All outputs are adjustable and metered at the meantime on the

large, graphic LCD display. With the multi-purpose knob and the

LCD display it is possible to enter the MENU mode that allows

setting many functions, which make T 1000 PLUS a very powerful

testing device, with manual and semi-automatic testing

capabilities, and with the possibility to transfer test results to a

PC via the USB interface. These results can be recorded,

displayed and analyzed by the powerful TDMS software, which

operates with all WINDOWS versions, and allows creating a data

base of all tests in the plant.

The basic T 1000 PLUS function is to generate current and

voltages and to stop generation as the relay trips. Test results are

kept in memory, and can be transferred to a PC at a later time,

along with settings.

The ease of operation has been the first goal of T 1000 PLUS: this

is why the LCD is graphic, and so large. With it, the dialogue in

MENU mode is made easy. Besides, all T 1000 PLUS outputs are

continuously measured, and output values are displayed, with no

extra effort to the operator. Also the show waveform feature can

be of help: any doubt about strange measurements, distortion

and so on can be solved.

This is also why we have added the reduced power feature.

Modern relays have a very low burden. As current output is a low

impedance voltage generator, adjusting low currents and/or

current on low burdens is quite difficult because one has to

operate at the very beginning of the adjustment knob. In this

situation it is possible to connect resistors in series; however, one

must be careful not to exceed the maximum current rating, and

the wiring is more complicated. The solution to this problem is

just to reduce the available power: this is easily performed via

DOC. MIE92093 Rev. 1.34 Page 8 of 108

the multi-function knob. With less power, the maximum voltage is

reduced by a factor of 4.4; the adjustment span on the knob is

increased accordingly.

Additional features are:

. Two meters, current and voltage, with independent inputs, allow

measuring T 1000 PLUS outputs or any other source;

. Two auxiliary contacts, that allow simulating the circuit breaker;

. A set of resistors allows easing output adjustment.

The instrument is housed in a transportable aluminum box, that

is provided with removable cover and handles for ease of

transportation.

NOTE: WINDOWS is a trademark of MICROSOFT inc.

DOC. MIE92093 Rev. 1.34 Page 9 of 108

1 TEST SET EXPLANATION

1.1 CONNECTION TO THE RELAY AND POWER-ON

At first, be sure that the main control knob (6) is turned (rotated)

to the zero position (complete counter-clockwise). The reason is

that the current generator is actually a high current voltage

generator. If the output is connected to the load (typically low

impedance), as soon as the test is started, a very high current

can circulate in the circuit.

Next connect the mains supply cable to the instrument and then

to the supply. THE SUPPLY VOLTAGE MUST BE THE SAME AS

INDICATED ON THE PLATE.

Power-on T 1000 PLUS: a diagnostic sequence controls:

. Key microprocessor board components;

. Auxiliary supply voltages.

If something is wrong, the operator is alerted by a message.

At the end of it, default selections are active; T 1000 PLUS is in

the OFF state.

Perform the first selections, according to the type of relay to be

tested:

. Main output socket, acting on the selector push-button (57).

. Auxiliary AC voltage: range; type of generation; value. NOTE: to

perform the range selection, the output must be ON, and the

value reduced to the minimum.

. Auxiliary DC voltage: range; value. NOTE: to perform the range

selection, the output must be ON, and the value reduced to the

minimum.

. Start and Stop timer inputs.

Connect the relay to be tested to the output sockets that have

the indication light (LED) on.

The following is the list of protections that avoid damaging T 1000

PLUS in case of errors.

DOC. MIE92093 Rev. 1.34 Page 10 of 108

. Fuse on the mains supply.

. Thermal NTC sensor on the main and auxiliary transformers. In

case of over-temperature, an alarm message is displayed.

. Thermal sensors on the SCR that controls current injection, and

of the internal temperature. In case of over-temperature, an

alarm message is displayed.

1.2 TEST CONTROL

The T 1000 PLUS front panel is explained in next paragraph.

T 1000 PLUS generation is controlled by the two keys < (55) and

> (56).

Settings and menu selections are controlled by the multi-function

knob with switch (22): see next paragraph for menu selections

description. At power-on T 1000 PLUS generation is OFF, as

confirmed by LED (50). The ON selection serves for finding relay

thresholds; selections ON+TIME and OFF+TIME serve to measure

relay timing.

The following flow diagram summarises all available test control

selections.

The performance of T 1000 PLUS in Normal Test mode is the

following.

. OFF: main outputs are not generated; Vac aux is generated,

and it can be either the pre-fault value or the fault value,

DOC. MIE92093 Rev. 1.34 Page 11 of 108

according to selections; Vdc aux is generated. In this condition,

any trip of Stop input is ignored.

. ON: timer starts; main outputs are generated; Vac aux has the

fault value; Vdc aux does not change. In this situation any trip at

Stop input is detected; it is possible to verify and memorize the

relay threshold, both trip and reset. As the relay trips, the TRIP

LED (43) turns on for 5 seconds; during 5 seconds, parameters at

trip are displayed; then, the standard measurement is restored.

Test results can be saved according to Save selections.

. From OFF to ON + TIME: main outputs are generated and the

timer starts according to selections; as Stop trips or resets, T

1000 PLUS returns to OFF, the TRIP LED (43) turns on and

parameters at trip are displayed until ON or ON+TIME are

selected. Test results can be saved according to Save selections.

. From ON to OFF + TIME: main outputs are removed the timer

starts according to selections; as STOP is sensed, T 1000 PLUS

returns OFF, the TRIP LED (43) turns on and parameters at trip

are displayed until ON or ON+TIME are selected. Test results can

be saved according to Save selections.

Other test mode selections:

. Relay or Breaker selection. Sometimes the test set is used to

monitor the CB contact instead of the relay contact. In this

situation the delay is longer than the relay delay, as the CB delay

adds to the relay delay.

If the test is time limited, it may occur that, while the relay trips

within the programmed maximum generation time, the CB

contact switches after this time. In such a situation, the time

measurement would be Not tripped.

To solve this issue, there is the selection Relay or Breaker. If

Relay is selected, then the time measurement is stopped as the

maximum fault generation time expires; if Breaker is selected,

the timer keeps on measuring for 100 ms after the decay of the

fault maximum time. The diagram explains the situation.

DOC. MIE92093 Rev. 1.34 Page 12 of 108

. Trip + pulse time: the timer measures the delay and the

duration of the trip impulse.

. Reclose test. It is possible to select via menu the test of a

reclosing scheme. Two selections are available, according to the

type of recloser under test.

In the first operating mode, T1000 PLUS is connected as follows:

Trip command to the STOP input; Reclose command to the START

input. As Reclose is detected, the test set automatically applies

current after the programmable reclaim time delay TD. The test

set measures and stores the relay trip delay, and the delay

between trip leading or falling edge and RECLOSE trailing edge

(see figure 4). The test set waits the close command up to 9999

s.

The sequence is repeated the programmed N times; after this, a

last fault is issued, and the test set verifies that there is no Close

command. On the last fault, the test set waits up to 10 times the

last Close delay before stopping the test.

During this operation, the auxiliary contact A1 can be used to

simulate the CB position: this is mandatory for the test of some

Recloser. The test starts with the CB position Closed; at the end

of the last test, it comes back to Closed.

FAULT

RELAY

CB

MAX TIME

0.1 s

DELAY

DOC. MIE92093 Rev. 1.34 Page 13 of 108

FAULT 1 2

TRIP (STOP) 1 2

RECLOSE 1

(START)

TIME MEAS.

Figure 4: Measure of Delay and Reclose times

The second operating mode refers to pole mounted CB’s. In this

mode, there is only one signal coming from the device under test:

the position of the CB. In this mode, the operation is the

following.

In this test mode the test set is edge triggered; the CB position is

connected to the STOP input, and, from the Closed – Open

position, the test set derives the Start and Stop commands, to

perform the test as above.

Other Fault injection selections:

. Maintained (default):

.. ON mode: fault outputs are generated until OFF is selected.

.. ON+TIME or OFF+TIME: as the STOP input is sensed, T 1000

PLUS returns OFF.

. Momentary: in ON mode, main outputs are generated until the

> push-button is pressed;

OPEN

CLOSED

(START)

(STOP)

D1 R1 D2 R2

TD

DOC. MIE92093 Rev. 1.34 Page 14 of 108

. Timed: in all modes (ON; ON+TIME; OFF+TIME), fault outputs

are generated for the programmed maximum time; after this, T

1000 PLUS returns OFF. Any trip after this time is not sensed.

. External. This mode allows for the synchronization of more T

1000 PLUS: they start generating upon reception of the START

input, that is selected in External mode.

. OFF delay: fault parameters can be maintained for the specified

time after relay trips: this allows simulating the circuit breaker

delay.

Test power selection: it allows reducing the available power; this

increases the adjustment sensitivity for low current tests on low

burden relays.

Save selections:

. No automatic saving.

. Automatic test data saving as relay trips. A pop-up window

confirms the saving and tells the test number.

. Test data can be saved after confirmation. After relay trip,

pressing the multi-function knob the operator can save the test

result.

. Manual test data saving. This selection can be used any time: it

serves if the trip is confirmed by a light and not by a contact.

Auxiliary contacts delays: the switch of the auxiliary contacts can

be timed with respect to test start.

1.3 CURRENT GENERATION

If the following current limits and time duration of main current

outputs are trespassed, the generation is interrupted, and the

operator is warned by an alarm message.

DOC. MIE92093 Rev. 1.34 Page 15 of 108

1) MAXIMUM POWER 300 VA

RANGE A AC

CURR. OUTPUT

A

MAX. POWER

VA

MAX. BURDEN

Ohm

LOAD TIME

s

RECOV. TIME

min

100 30 300 0.33 STEADY -

50 440 0,17 30 min 100

75 600 0.1 600 45

100 750 0.075 60 15

150 900 0.04 3 10

250 1000 0.016 1 5

40 12 300 2 STEADY -

20 450 1.1 30 min 100

30 600 0.66 600 45

40 800 0.5 60 15

60 900 0.25 3 10

80 1000 0.15 1 5

10 5 400 16 STEADY -

7.5 600 10 15 min 45

10 800 8 60 15

15 900 4 5 10

20 1000 2.5 2 5

NOTE: if the current generation lasts less than the maximum

value, the recovery time is proportionally reduced. For instance, if

you tested at 100 A during 6 s, the test set will pause during 1.5

min, or 90 s.

The following table shows the values of the current sunk from the

supply.

RANGE

A AC

CURR.

OUTPUT A

SUPPLY

CURR. @ 230 V

A

SUPPLY

CURR. @ 115 V

A

100 30 1.5 3

50 2.5 5

75 3.7 7.4

100 5 10

150 7.5 15

250 12.3 N.A.

DOC. MIE92093 Rev. 1.34 Page 16 of 108

40 12 1.6 3.2

20 2.6 5.2

30 4 8

40 5.2 10.4

60 7.8 15.6

80 10.5 N.A.

10 5 2.2 4.4

7.5 3.3 6.6

10 4.4 8.8

15 6.6 13.2

20 8.8 N.A.

NOTE: with the supply of 115 V, the current sunk at maximum

current outputs is too high; so, these outputs are not available.

2) MAXIMUM POWER 60 VA

RANGE

A AC

CURRENT

OUTPUT

A

MAXIMUM

POWER

VA

LOAD

TIME

s

RECOVERY

TIME

min

100 30 60 STEADY -

38 10 min 45

53 60 10

70 0.75 2

40 12 60 STEADY -

17 10 min 45

23 60 10

36 1 2

10 5 60 STEADY -

6 10 min 45

7 60 2

10 1,5 2

This generator serves for the test of current, power, directional,

distance relays, where current or current and voltage are

necessary. The procedure is the following.

. At first, be sure that the main control knob (6) is turned

(rotated) to the zero position (complete counter-clockwise).

. Power-on T 1000 PLUS.

DOC. MIE92093 Rev. 1.34 Page 17 of 108

. Select by the push-button (57) the measurement on the desired

output sockets (13), according to the maximum current to be

generated: the LED turns on; the AC voltage value is displayed.

. Connect the relay to be tested to sockets (13). Consider that for

tests of 40 A up it is necessary to connect the relay by a wire

having at least a cross section of 10 sq. mm; for lower currents, a

cross section of 2.5 sq. mm can be used.

. Press ON and adjust the output current to the desired value with

knob (6).

. After you have started the test, if the burden is a short circuit

made of a short cable, you measure at zero knob position a

current that usually is less than 3% of the range. This value does

not influence at all the measurement of the current you are

generating: it is not an error of the measurement

instrument. If the current is a problem, select the 60 VA power,

and/or connect resistors in series.

. There are two more possible problems: the desired current

cannot be reached; the adjustment is difficult because the current

is reached too easily.

.. If it is impossible to reach the desired value, this is because the

burden is too high. Very often the problem comes from

connection wires; so, to perform the test it is necessary either to

shorten them, or to increase the cross section (or both).

.. If the adjustment is reached within 1/5th of the knob rotation,

then it is possible to increase the ease of adjustment by reducing

the test power as follows.

TEST CONTROL > TEST POWER (Power) ESC

.. It is also possible to increase the ease of adjustment by

connecting a resistor of the set in series to the relay. Resistors

are rated 50 W; so, compute the resistance value as follows:

(RESISTANCE) = 50 / (TEST CURRENT)^2

Maximum test current values are resumed here below.

RESISTANCE 0.5 1 22 470 1000 2200

MAX ITEST 10 7 1.5 0.3 0.2 0.15

Note that the test starts and stops as the current passes the zero.

DOC. MIE92093 Rev. 1.34 Page 18 of 108

1.4 AC VOLTAGE GENERATION FROM MAIN OUTPUT

If the current of 3.5 A is exceeded on main AC voltage output,

the generation is interrupted, and the operator is warned by an

alarm message.

This generator serves for the test of synchronism relays, where

two voltages are necessary. The procedure is the following.

. At first, be sure that the main control knob (6) is turned

(rotated) to the zero position (complete counter-clockwise).

. Power-on T 1000 PLUS.

. Select by the push-button (57) the measurement on output

sockets (60): the LED turns on; the AC voltage value is displayed.

. There are two ranges available: 250 V (full power); 57 V

(reduced power). The default at power-on is full power; if 57 V

are enough, for a better adjustment, reduce the voltage as

follows.

TEST CONTROL > TEST POWER (Power) ESC

. Adjust the output voltage to the desired value with knob (6).

. Connect the relay to be tested to sockets (60). Check that the

adjusted voltage does not drop as you connect the relay; else,

this would mean that T 1000 PLUS is overloaded (or that you are

connecting to a live wire). In this situation, remove the cause of

error and connect again.

1.5 DC VOLTAGE GENERATION FROM MAIN OUTPUT

If the current of 3.5 A is exceeded on main DC voltage output,

the generation is interrupted, and the operator is warned by an

alarm message.

This generator serves for the test of timers and all devices that

are driven by a DC voltage. The auxiliary DC voltage generator

cannot be used to this purpose as it is continuously generated: no

time measurement can be performed. To this purpose, act as

follows.

. At first, be sure that the main control knob (6) is turned

(rotated) to the zero position (complete counter-clockwise).

. Power-on T 1000 PLUS.

DOC. MIE92093 Rev. 1.34 Page 19 of 108

. Select by the push-button (57) the measurement on output

sockets (61): the LED turns on; the DC voltage value is

displayed.

. There are two ranges available: 300 V at 300 W continuous (full

power); 68 V at 60 W continuous (reduced power). The default

at power-on is full power; if necessary, reduce the power as

follows.

TEST CONTROL > TEST POWER (Power) ESC

. Adjust the output voltage to the desired value with knob (6).

. Connect the relay to be tested to sockets (61). Check that the

adjusted voltage does not drop as you connect the relay; else,

this would mean that T 1000 PLUS is overloaded (or that you are

connecting to a live wire). In this situation, remove the cause of

error and connect again.

1.6 AC VOLTAGE GENERATION FROM THE AUXILIARY OUTPUT

The auxiliary AC voltage is used to test relays that need voltage

and current at the meantime. In this situation, the voltage is

continuously generated; usually, it is adjusted to the nominal

value, and it is not changed during all tests. It is possible to

phase shift the current with respect to voltage; selections are the

following.

. Power-on T 1000 PLUS: the output voltage is zero. Press the

push-button (70) to enable or disable the output; when enabled,

light (62) is ON, and the AC voltage value is displayed on the

screen.

. There are three ranges available from 45 Hz up: 65; 130 or 260

V AC; the power is 30 VA continuous; 40 VA peak for 1 minute.

For increased power and accuracy, it is better to select the range

that is closest to the value to be generated. The default at

power-on is 65 V; if necessary, select the desired range.

. At lower frequencies, the maximum output and power

decrease. The table summarizes the situation at 15 Hz and 33.33

Hz.

DOC. MIE92093 Rev. 1.34 Page 20 of 108

RANGE @

50 Hz

VMAX @

15 Hz

POWER @

15 Hz

VMAX @

33.3 Hz

POWER @

33.3 Hz

V V VA@

VOUT

V VA@

VOUT

65 25 8 @ 22.5 V 55 35 @ 55 V

130 50 12 @ 45 V 110 35 @ 110

V

260 100 15 @ 90 V 220 35 @ 220

V

DO NOT EXCEED THE MAXIMUM VOLTAGES OF THE ABOVE

TABLE: THE OUTPUT WOULD BE DISTORTED!

. The operating mode is pre-selected as Fault: do not change it.

Do not change also the pre-selected frequency, as Locked to

mains. Last, set the desired current phase angle; however, to

perform this, T 1000 PLUS must be ON, and some current must

circulate. Selections are performed as follows.

AUX VAC/VDC > Aux Vac control > Range > (Range) RET

Phase > Reference:

current > (Phase) ESC

NOTE: to perform the range selection, the output must be ON

(press push-button (70)), and the value reduced to the minimum,

else the test set alerts to reduce it.

This performed, adjust the voltage to the desired value with knob

(20). Eventually, connect the relay to be tested to sockets (62).

Check that the adjusted voltage does not change or the overload

message pops up as you connect the relay; else, this would mean

that T 1000 PLUS is overloaded (or that you are connecting to a

live wire). In this situation, remove the cause of error and

connect again (reset the alarm if it popped up).

Execute the test, modifying the phase angle as necessary.

This output is also used for the test of voltage relays, frequency

relays, synchronism relays; frequency rate of change relays. In

these instances, it is necessary to use the Pre-fault + Fault

selection. This feature allows adjusting two different values: the

DOC. MIE92093 Rev. 1.34 Page 21 of 108

pre-fault voltage, that simulates the situation prior to fault, and

the fault voltage.

The pre-fault voltage adjustment is performed by the control

knob, while knob (20) adjusts the fault value. Voltage output

selection is automatic: pre-fault voltage with test stopped; fault

voltage with test started. The switch from a value to the other

one is performed without falling to zero. The main current or

voltage is generated at the zero crossing; the fault one is

generated at the meantime of main voltage or current. The

selection of the reference is performed automatically, following

the selection of main output measurement. If the main DC

voltage is selected the reference is taken on the main AC voltage.

The Pre-fault + Fault selection is performed as follows.

AUX VAC/VDC > Aux Vac control > Mode > Pre-fault+Fault

> Pre-fault amplitude ESC

Once the nominal voltage is adjusted to the pre-fault value, it is

possible to change the amplitude (V, SYN relays) or frequency (F,

SYN relays), or angle (SYN relays: in this instance, the angle is

referred to V main rather than to I main) or frequency rate of

change (dF relays) of the fault voltage.

The pre-fault (nominal) frequency is always the mains one; the

fault one is adjusted by the control knob, in the range 40 Hz to

500 Hz. Switching from nominal frequency to fault frequency is

performed without altering the output voltage amplitude and

phase. The frequency adjustment is performed as follows.

DOC. MIE92093 Rev. 1.34 Page 22 of 108

AUX VAC/VDC > Aux Vac control > Frequency > Adjust >

(Frequency value) ESC

The angle adjustment is performed as explained above; the

difference is that the adjustment is applied only when T 1000

PLUS is ON; with the instrument OFF, the pre-fault voltage is

normally in phase with the current. It is also possible to phase

shift the pre-fault voltage with respect to the fault voltage. This

parameter is necessary during the test of distance relays, when

phase to phase faults are simulated: as test starts, the auxiliary

voltage changes amplitude and phase with respect to the pre-

fault value.

The pre-fault angle adjustment is performed as follows.

AUX VAC/VDC > Aux Vac control > Mode > Pre-fault+Fault

> Phase (phase) ESC

This angle is referred to the fault voltage; so, for its adjustment is

not necessary to have T 1000 PLUS ON.

Last, it is possible to test frequency rate of change relays, by

setting both the starting frequency, as before, and the frequency

ROC range, with range from ± 0.01 to ± 99.99 Hz/s. The

frequency change stops at 40 or 70 Hz. The frequency ROC

adjustment is performed as follows.

AUX VAC/VDC > Aux Vac control > Frequency > Adjust >

Frequency ROC (ROC) ESC

DOC. MIE92093 Rev. 1.34 Page 23 of 108

As test starts, the frequency goes to the pre-set frequency, and

from that value starts increasing or decreasing with the pre-set

ROC.

NOTE: The auxiliary AC voltage is protected by an electronic

circuit that stops the voltage generation and opens the

connection to outputs socket in case of overload (short circuit

included). In case of intervention, an alarm message is displayed.

Via the control knob the operator can reset the alarm and close

the relay to restore operation.

The auxiliary AC voltage is also protected by a thermo switch that

intervenes in case of over-heating. In case of intervention, an

alarm message is displayed.

1.7 DC VOLTAGE GENERATION FROM THE AUXILIARY OUTPUT

The second auxiliary generator can be used to supply the

auxiliary DC voltage for the relay to be tested. To this purpose,

the voltage is available at sockets (63), after pressing the push-

button (69). Use this generator as follows.

. Power-on T 1000 PLUS: the output voltage is zero. Press the

push-button (69) to enable or disable the output; when enabled,

light (63) is ON, and the DC voltage value is displayed on the

screen.

. There are two ranges available: 130 or 240 V DC; the power is

90 W. For increased power and accuracy, it is better to select the

range that is closest to the value to be generated. The default at

power-on is 130 V; if necessary, select the desired range as

follows.

AUX VAC/VDC > Aux Vdc control > (Range) ESC

NOTE: to perform the range selection, the output must be ON

(press push-button (69)), and the value reduced to the minimum,

else the test set alerts to reduce it.

This performed, adjust the voltage to the desired value.

Eventually, connect the relay to be tested to sockets (63). Check

that the adjusted voltage does not change as you connect the

DOC. MIE92093 Rev. 1.34 Page 24 of 108

relay; else, this would mean that T 1000 PLUS is overloaded (or

that you are connecting to a live wire). In this situation, if it is not

a connection error, it is possible to reduce the voltage until the

voltage does not drop: relays tolerate a wide range of DC supply

voltages.

The auxiliary DC voltage is protected by a current limiter. The

user notices the low voltage and removes the overload. The fuse

protects the case of counter-feed.

1.8 AUXILIARY CONTACTS

- The two auxiliary outputs have make and break contacts, that

close (open) at test start after the programmed delay, and open

(close) as current is cut off after the STOP input is sensed.

Contacts can be used to simulate the circuit breaker state.

- Possibility to delay the auxiliary contacts switch with respect to

test start. Delay range: from 0 to 999.99 s.

- Contacts range: 5 A; 250 V AC; 120 V DC

1.9 THE TIMER

Timer inputs are protected against wrong selections. If the

voltage free input is selected and a voltage is applied less than

250 V ac or 275 V DC, circuits will not be damaged.

Characteristics of Start and Stop inputs:

. Inputs do not have any common point, and are opto-coupled

from the instrument at 1.35 kV AC;

. Inputs connection: two banana sockets per input;

. Type of input: either clean or under voltage; maximum input:

250V a.c or 275 V DC;

. Inputs may be independently selected as Normal Open or

Normal Close or Edge: the latter means that the timer is stopped

by any transition;

. Selections are displayed on the front panel by 10 dedicated

lights;

DOC. MIE92093 Rev. 1.34 Page 25 of 108

. For both inputs, when the input is closed or with voltage an LED

turns on;

. When the relay intervenes the TRIP light turns on;

. Inputs protection: if the voltage free input is selected and a

voltage is applied, no damage occurs, provided that the voltage is

within the specified limits.

. When the input is with voltage, it is possible to select the

voltage threshold. Low threshold is for voltages between 24 and

48 V; high threshold is for voltages above 100 V .

The timer offers a number of possible selections to permit many

different types of testing depending on the selection by the

operator. The following flow shows all possible selections.

The default selection is the following.

. Timer start: internal; as test is started. External start allows

synchronizing more T 1000 PLUS.

. Timer stop: external. Internal stop means that the timer is

stopped as T 1000 PLUS goes OFF.

With this selection, the timer meters the elapsed time between

START and STOP. Alternative selections:

.. Elapsed time plus the duration of STOP input; the selection is

performed as follows:

TEST CONTROL > Test mode > Trip + pulse time ESC

.. Impulse counting:: this mode is foreseen for the test of energy

meters. Maximum input frequency: 10 kHz; voltage threshold can

be set as for tripping. It is possible to select this mode via menu,

and to set the number of impulses; the test set measures the

time corresponding to the set number of complete periods applied

to STOP input after ON and during all generation, and measures

DOC. MIE92093 Rev. 1.34 Page 26 of 108

the corresponding energy (if selected). The selection is performed

as follows:

TIMER START/STOP > STOP > External > Count > (number

of counts) ESC

If the count is selected on START, time measurement will be

performed after the set number of counts has expired: this serves

to pass the start-up of the energy meter.

Input thresholds. When the contact has voltage applied, two

thresholds can be selected. The low setting applies to nominal

voltages of 24 and 48 V; the high setting to 110 V up. The

selection is performed as follows:

TIMER START/STOP > STOP > External > Clean-voltage >

Voltage threshold ESC

Time can be metered as seconds or cycles (second is the default).

The selection is performed as follows:

TIMER START/STOP > Units > s or Cycles ESC

1.10 FINDING RELAY THRESHOLDS

1.10.1. Introduction

There are different types of relay characteristic curves: time

depending or independent; single threshold or multiple

thresholds. For all of them, all the test set can do is to generate

fault parameters, and to verify the corresponding trip time.

Let us consider the following example, that applies to an

overcurrent relay with two time-independent thresholds: I> and

I>>. Of this relay we want to measure the thresholds, both trip

and re-set, and the timings.

DOC. MIE92093 Rev. 1.34 Page 27 of 108

1.10.2. First threshold trip and drop-off

The first session is finding threshold I>, that is the limit between

no trip and trip (with long delays). With any other type of single-

threshold relay, the procedure is the same, provided that the

corresponding parameter is changed.

. Select ON; slowly increase the current.

. As the relay trips, pressing the multi-function knob tripping

values can be saved.

The value in memory is the current as the relay trips. This value

is the threshold one only if the current has increased very little

after the threshold was passed. In order to increase the accuracy,

reduce the speed of variation or use a not-timed relay output.

Next, we find the drop-off threshold for I>.

From the trip current above, slowly decrease the current; as the

relay resets, save test result.

The value in memory is the current as the relay resets. As the

reset timing is usually very short, it is usually fairly accurate.

The TRIP light (43) turns on during 5 s; during this time trip

values are held, and after this time the actual reading is displayed

again.

DOC. MIE92093 Rev. 1.34 Page 28 of 108

1.10.3. Second threshold trip and drop-off

The second session is finding threshold I>>. The problem is that

the test result criterion is no more to find the limit between no

trip and trip; it is instead to find the limit between two different

timings: what we have shown as t>, for currents less than I>>,

and t>> for currents more than I>>. There are many ways to

perform the test; we suggest taking advantage of the Timed

generation option, as follows.

. Start from a current more than I>; select ON+TIME, and check

for time response. Take note of the timing t>. Compute tmax as

80% of t>.

. Set the maximum test duration as tmax: the relay will not trip

until you enter the I>> zone.

. Select ON+TIME, and start the test: the test goes OFF with no

message. Slowly increase the test current and repeat the tests,

until the relay trips within tmax: this is the threshold; pressing

the multi-function knob tripping values can be saved.

NOTE: this procedure applies to any multi-threshold relay,

including distance relays. For these latter relays, as the

Impedance vs. time curve has increasing steps (more delay for

higher fault impedance) the limit will be found setting a fault

voltage (at the given current) and then decreasing it.

Finding the drop-off threshold for I>> is impossible, unless if

there is a separate trip for the instantaneous threshold. In fact,

after the relay trips in I>>, reducing the current keeps the relay

tripped for I>, and the I>> reset threshold cannot be measured.

1.11 FINDING RELAY TIMINGS

After having measured the threshold of the relay, in the manner

stated before, it is possible to measure the intervention time of

the relay. Relevant timings are: trip and drop-off delays; they can

be tested on over-current (voltage, frequency..) relays, or on

under-current (voltage, frequency..) relays.

Available test selections are:

. Over-current relay trip delay: from OFF to ON+TIME;

. Over-current relay drop-off delay: from ON to OFF+TIME;

DOC. MIE92093 Rev. 1.34 Page 29 of 108

. Under-current relay trip delay: from ON to OFF+TIME;

. Under-current relay drop-off delay: from OFF to ON+TIME.

All these tests are performed after having pre-adjusted the

desired test current (or set of parameters).

Let us consider for example the timing test of an over-something

relay, where the trip contact is Normal Open. The test is

performed as follows.

. First thing, select the Maintained fault injection, as follows

. Set the “Automatic save at trip” function, as follows.

. Set the timer with the following selections:

I

t

5A) Overcurrent relay trip delay

5B) Undercurrent relay trip delay

I

t

Stop

Stop

FAULT

FAULT

Trip delay

Trip delay

DOC. MIE92093 Rev. 1.34 Page 30 of 108

. Now, press ON and pre-adjust the first set of fault parameters:

as the relay trips, don’t save test result; go OFF.

. Select ON+TIME: as the relay trips, test goes OFF; pressing the

multi-function knob tripping values can be saved. The TRIP LED

(43) turns on and parameters at trip are displayed until ON or

ON+TIME are selected. Confirm save results pressing the multi-

function knob, and proceed with other test currents, until all

points to be tested are measured.

Now we can measure the drop-off timing.

. First thing, select the NC level for the relay reset contact:

. Now, press ON and pre-adjust the same set of fault parameters.

. Select OFF+TIME: as the relay resets, pressing the multi-

function knob drop-off values can be saved. Confirm save results

pressing the multi-function knob, and proceed.

With different test values, other timings can be measured.

1.12 TEST RESULTS HANDLING

Before performing a test, it is important to program the test

header, that includes the following information:

Plant name;

Operator;

Serial number (of the relay under test);

Model/manufacturer (of the relay under test);

Feeder (protected by the relay).

The header is found selecting Results > Header.

All tests performed after setting the header will be grouped

together: the TDMS software will group them together, and will

allow to show test results with a single result table and diagram.

Once a relay has been tested, it is important to change at least

the relay serial number, so that results of different relays are not

mixed together. If a relay has more than one curve, it is possible

to give different headers, just by adding a suffix letter to the

serial number.

Once selected the Header, the following window is displayed.

DOC. MIE92093 Rev. 1.34 Page 31 of 108

The operation to input the header is the following.

After having entered, if you move the knob it will move

between: arrow up; Plant name; arrow down, Return.

If you press the arrow up or down, you will scroll through:

Plant name, Operator, Serial number, Model/manuf.,

Feeder.

Once you have reached the desired description, select it

and press: the description goes in reverse, and you can

edit it.

The editor is performed as follows. If you turn the knob, it

will reach a number of selections in the bottom, and then

the line with the alphabet letters; at the end of this, there

is a double arrow, that, if pressed, replaces letters with

digits.

The input is performed reaching for a letter with the knob,

and pressing it: the letter is copied into the description.

After the letters, you can select also: /_., .

Commands on the bottom line are: Delete the letter after

the cursor; Delete all the description; move one letter left;

move one letter right; OK, to be pressed at the end of the

editing, before going to another description.

At the end of all editing, press the return arrow.

During a test, results can be memorized selecting Test control >

Save. Four selections are available:

Don’t save;

Automatic at trip;

Confirm at trip;

Manual.

With the selection Automatic at trip, data are immediately saved

as soon as the relay trips: this is to be selected when a series of

ON+TIME tests are performed.

With the selection Confirm at trip, as soon as the relay trips it is

possible to save data pressing MENU and then Yes: this is to be

selected when you are adjusting parameters and you want to be

DOC. MIE92093 Rev. 1.34 Page 32 of 108

sure to save the correct data. Save is performed both in ON mode

and in ON+TIME mode.

With the selection Manual, it is possible to save data at any time,

pressing MENU and then Yes, even if the relay did not trip. As

already explained, a threshold is verified with two tests: with a

value the relay trips, with a slightly different value the relay does

not trip. This is to be selected when you are looking for the no-

trip limit of a threshold.

Once results are saved, it is possible to review or to delete them

with the command Results > Show results. The display shows

the list of the relay serial numbers, followed by the number of

tests performed with the same header.

Pressing on the relay serial number you access a window with:

Plant name, Serial number, operator. You can: return, delete

escape. Pressing Delete, a confirmation message is displayed.

Pressing Yes, all results are deleted; pressing No, you return to

the results list.

If you go to the number of tests and press, tests are shown one

after the other, as they have been recorded. Here you can read

the results, and delete the ones that you want, after

confirmation.

1.13 BASIC TEST PRINCIPLES

1.13.1. Introduction

In the years many different types of protection relays have been

developed:

. With different technologies (electromechanical, solid state,

microprocessor);

. Different ways of setting the same values;

. Other performances (multiple thresholds, conditioning signals).

Testing relays allows a wide range of different modes to get the

same result; so, for the first thing the user must choose the type

of test to be performed.

DOC. MIE92093 Rev. 1.34 Page 33 of 108

The easiest type of test is to check the settings. In this instance it

is sufficient to generate the relevant parameters (voltage,

current, angle) with values close to the settings. With two timing

tests, little below and little above the settings, it is possible to

test that they are correct, within a specified tolerance.

For instance, testing an over-current relay set at 10 A can be

performed with two delay tests: at 9.5 A (-10%) the relay should

not trip; at 10.5 A (+5%) it should trip; also the timing can be

tested.

A more complex and complete test is the control of the complete

characteristic curve of the relay. There are two types of

characteristic curves:

- Parameter vs. time: it displays the trip time as the parameter

changes. Examples are: current vs. time, for an over-current

relay; impedance vs. time for a distance relay. It is understood

that for currents less than the threshold, or for impedances

higher than the general starter, no trip occurs.

- Parameter vs. parameter: they display the trip limit of a

parameter vs. another one; other variables are kept constant. For

example, the current vs. voltage curve of an earth fault relay,

that applies for a given range of phase shifts. Other examples: P

vs. Q curve for a reverse power relay; R vs. X curve for a

distance relay; R vs. X curve for a loss of field relay.

The key difference between these types of curves is that the first

one can be tested with timing tests, while the other one is a

collection of threshold tests, where one of the variables is kept

constant. As a consequence, the first type of curves is easily

tested, with a multiple delay test; the second one requires an

analysis of the shape of the curve.

1.13.2. Parameter vs. time characteristic

As an example of the parameter vs. time curve, we consider the

characteristic curve of a time-dependent overcurrent relay.

DOC. MIE92093 Rev. 1.34 Page 34 of 108

I-t over-current relay curve

The curve means that:

. If the test current is less than I> no trip occurs;

. For current more than I> and less than I>> trip time is a

function of the test current;

. For currents more than I>> the trip time is t>>.

This curve can easily be tested by selecting ON+TIME tests at

increasing test currents. The user can decide how many tests to

perform to reach the desired confidence that test results match

with the nominal curve.

1.13.3. Parameter vs. parameter characteristic

As an example of the parameter vs. parameter curve, let us

consider the characteristic curve of an earth fault directional

relay.

First of all, the curve applies only if the test is executed at given

angles between current and voltage. Then, at these angles, the

curve means that for current and voltages below it there is no

trip, while above it the relay trips. In the curve the trip time is

not shown, yet this is the parameter to be measured in order to

understand if we are below or above the curve. This means that

each point of the curve is a threshold limit between trip and no

trip; the threshold can be found moving one parameter and

t

I

t>>

I> I>>

DOC. MIE92093 Rev. 1.34 Page 35 of 108

keeping the other one constant. However, we must be careful

before executing the test.

I-V directional relay curve

For instance, in order to find the point A it is apparent that the

procedure is:

. Set the voltage;

. Increase the current until the relay trips.

However, in order to find the point B the best procedure is:

. Set the current;

. Increase the voltage until the relay trips.

In fact, if we search the point B threshold keeping constant the

test voltage, a minor error on the voltage would cause a big error

on the current. The same problem would occur testing point A at

constant current.

The test of this relay should be split in two: constant voltage for

the first half; constant current in the second half. In conclusion,

when we perform the test we should consider how does the

variable parameter move on the characteristic plane, and avoid

asymptotic situations.

As a second example of the parameter vs. parameter curve,

consider the following curve: it is the general starter of some type

of distance relays.

V

IImin

Vmin

A

B

DOC. MIE92093 Rev. 1.34 Page 36 of 108

The area to the right is the trip zone; to the left it is the no trip

zone. The test strategy should be the following:

. For points A, B, D: set the test voltage and increase the current

until the relay trips;

. For point C, set the test current and decrease the voltage

until the relay trips

As third example we can study the problem of testing a loss of

field relay, that has the following characteristic.

X

RA

B

CD

V

I

A

I V 0

V M I N

V N

I V N

B

C

D

DOC. MIE92093 Rev. 1.34 Page 37 of 108

The area of intervention is the one within the circle with the

centre in the negative X axis. The first problem is to convert R

and X into the corresponding V, I, angle; then, the current is

set at IN, and only voltage and phase are modified. The test is

performed as follows:

. For point A, set the angle at 270°, start from V=0 and increase

V until the relay trips;

. For point B, set the angle at 270°, start from V higher than

the setting and decrease V until the relay trips;

. For point C, set for V the average of A and B, start from an

angle greater than 270° and reduce it until the relay trips. In a

similar way, for D, start from an angle smaller than 270° and

increase it until the relay trips.

. For other points, you have to decide whether to take the angle

constant and change the voltage or vice versa; you should choose

the variation that is not tangent to the characteristic curve.

1.14 USE OF THE TEST SET AS A MULTIMETER

The test set can be used as a powerful multimeter, as it allows

measuring:

. Voltage, AC and DC;

. Current; AC and DC;

. Frequency;

. Impedance and its components;

. Power (active, reactive, apparent, power factor);

. Phase angle: with respect to the mains or between two inputs.

This feature allows also to check the outputs generated by the

test set itself. The marking CAT II means that these inputs are

suitable for the measurement of voltage or currents connected to

the mains.

It is important not to exceed the maximum ratings, as marked.

The low current input is protected by a fuse; all other inputs are

not protected: if the rating is exceeded, this could cause the

permanent fault of the corresponding input.

. Prior to the measurement, enter the MENU selection, and enable

the display of the parameter you desire.

. For the measurement of voltage or current, just connect to the

corresponding input.

DOC. MIE92093 Rev. 1.34 Page 38 of 108

. The frequency is measured only on the voltage input: connect it

and select the F measurement.

. The measurement of the impedance or of the resistance of a

component fed by an external source is performed as follows.

.. Open the circuit, and connect the component to the

external current measurement;

. Connect the component terminals to the external voltage

measurement: this avoids errors caused by the connection

leads and by the current measurement circuit;

. Select the desired measurement (Z+ φ, or R+X);

. Close the circuit, taking care not to exceed the current

limits. The measurement does not change as soon as current

and voltage are high enough: this confirms the correct

measurement.

. For power measurements of an external source proceed as

follows.

.. Open the circuit, and connect the component to the

external current measurement;

. Connect the component terminals to the external voltage

measurement: this avoids errors caused by the connection

leads and by the current measurement circuit. Take care not

to exchange leads, otherwise the power will be displayed with

a negative sign;

. Select the desired measurement (P, Q or S, p.f.);

. Close the circuit, taking care not to exceed the current

limits: the measured power is displayed.

. For phase angle measurement, first of all select if you want to

measure the angle of your current or voltage with respect to the

mains, or current with respect to voltage. NOTE: if one of the

values is low, the measurement with respect to the mains is more

stable. Then, connect the voltage or current (or both), taking care

of the positive sign, that must be connected to the red socket:

the measured angle is displayed.

The following table summarizes the available measurements.

DOC. MIE92093 Rev. 1.34 Page 39 of 108

N. PARAMETER , AC

INPUTS

DERIVED

FROM

FORMULA UNITS

1 ACTIVE POWER, P Iext, Vext;

φ

P = I*V*cos

(φ)

W

REACTIVE POWER, Q Iext, Vext;

φ

Q =

I*V*sin(φ)

VAr

2 APPARENT POWER, S Iext, Vext S= I*V VA

POWER FACTOR, p.f. φ p.f. = cos(φ) -

3 IMPEDANCE, Z and φ Iext, Vext,

φ

Z = V/I Ohm,

°

4 ACTIVE IMPEDANCE

COMP., R

Iext, Vext;

φ

R = Z*

cos(φ)

Ohm

REACTIVE IMPEDANCE

COMP., X

Iext, Vext;

φ

X = Z*

sin(φ)

Ohm

5 FREQUENCY, F Vext - Hz

6 PASE ANGLE, IE TO V2 Φ, IE-V2;

ref. V2

- °

PASE ANGLE, VE TO V2 Φ, VE-V2;

ref. V2

- °

1.15 TD1000 MODEL: AC CURRENT GENERATION FROM THE

AUXILIARY OUTPUT

The TD1000 model has an additional feature: it allows generating

up to 20 A AC from the auxiliary output.

The selection is performed as follows:

AUX VAC/IAC > AUX VAC/IAC CONTROL > RANGE > 20 A

Take care : this output is not a current source : it is a low

voltage, high current source.

All selections available for the auxiliary voltage outputs apply to

the current output: frequency range, angles, prefault – fault.

When this output is selected, the measurement becomes A rather

than V.

. At lower frequencies, the maximum output and power

decrease. The table summarizes the situation at 15 Hz and 33.33

DOC. MIE92093 Rev. 1.34 Page 40 of 108

Hz: these values are higher than those of the standard T1000

PLUS.

RANGE @

50 Hz

VMAX @

15 Hz

POWER @

15 Hz

VMAX @

33.3 Hz

POWER @

33.3 Hz

V V VA@

VOUT

V VA@

VOUT

3 1.8 20 @ 20 A;

14 @ 10 A

3.2 40 @ 20 A;

25 @ 10 A

65 35 8 @ 35 V 65 45 @ 65 V

130 70 14 @ 70 V 130 35 @ 130

V

260 140 30 @ 140

V

260 35 @ 260

V

1.16 FILTERING HEAVY BURDEN DISTORTION WITH THE FT1000

OPTION

Some very old electromechanical relays may have very inductive

loads, which change in value during the test. As the test set

current output is the ratio of applied voltage divided by the

burden impedance, the inductance change causes current

changes.

The optional FT1000 opposes to such changes, and smoothes

them away. The option includes an inductor, which adds to the

burden, but stabilizes the current.

The connection schematic is the following one (case of 10 A).

First of all, Select the operating frequency at 50 Hz or 60 Hz.

FT1000 RELAY

I1

IN

C 10A

DOC. MIE92093 Rev. 1.34 Page 41 of 108

Next, all you have to do is to connect FT1000 in series to the

T1000 selected output: the FT1000 range shall be the same or

greater than the one of T1000.

DOC. MIE92093 Rev. 1.34 Page 42 of 108

2 TEST SET AND POP-UP MENU

2.1 THE FRONT PANEL

DOC. MIE92093 Rev. 1.34 Page 43 of 108

The following list includes the key components inside T 1000

PLUS; see the schematic on last page.

1) Main supply fuse, rated T10A, incorporated in the supply

socket.

2) Power-on switch.

6) Main outputs adjustment.

8) Earth socket.

13) Output sockets for main current output.

19) Auxiliary DC voltage fuse, F2A.

20) Adjustment of auxiliary AC voltage output.

21) Adjustment of auxiliary DC voltage output (not available on

the model TD1000 – 15 Hz).

22) MENU control knob, with switch.

23) Display.

24) Resistors set.

31) Internal START selection light.

32) Normal Open START selection light.

33) Normal Closed START selection light.

34) Under voltage START selection light.

35) Voltage clean START selection light.

36) Internal STOP selection light.

37) Normal Open STOP selection light.

38) Normal Closed STOP selection light.

39) Under voltage STOP selection light.

40) Voltage clean STOP selection light.

41) START input closed or with voltage light.

42) STOP input closed or with voltage light.

43) TRIPPED condition recognized light.

44) 100 A range selection light.

45) 40 A range selection light.

46) 10 A range selection light.

47) 300 VA (normal power) selection light.

48) 60 VA (reduced power) selection light.

49) ON + TIME light: current is generated and time metered until

STOP is detected.

50) OFF light: no current generation.

51) ON light: current is generated.

DOC. MIE92093 Rev. 1.34 Page 44 of 108

52) OFF + TIME light: current is removed and time metered until

STOP is detected.

53) Main AC voltage selection light.

54) Main DC voltage selection light.

55 and 56) START and STOP push-buttons.

57) Push-button for the selection of main output.

60) Main AC voltage safety sockets.

61) Main DC voltage safety sockets.

62) Auxiliary AC voltage (also current on TD1000) safety sockets

and light: ON = output available.

63) Auxiliary DC voltage safety sockets and light: ON = output

available.

64) Sockets for the connection to the external Start input.

65) Sockets for the connection to the Stop input.

66) Auxiliary outputs Normal Open and Normal Closed contacts.

67) External current meter safety sockets.

68) External voltage meter safety sockets.

69) Auxiliary DC voltage ON-OFF switch.

70) Auxiliary AC voltage ON-OFF switch.

71) USB interface connector.

2.2 DISPLAY AND CONTROL LIGHTS

At power-on, the following control lights turn on (default

situation):

. START: INT (31); NO VOLTAGE (35);

. STOP: NORMAL OPEN (37) and NORMAL CLOSED (38) (EDGE

selection); NO VOLTAGE (40);

. TEST OFF (50);

. MENU SELECTION: 100 A SOCKET (44);

. POWER: 300 VA (47).

These lights change according to program selections.

At power-on and during the standard operation the display shows

the measurements of: main AC current (or main AC voltage or

main DC voltage, according to selection); auxiliary AC voltage;

auxiliary DC voltage; elapsed time. To the left is the area for the

access to the menu selection.

DOC. MIE92093 Rev. 1.34 Page 45 of 108

After power-on, the auxiliary AC voltage is available at sockets

(62) after pressing the push-button (70), and can be adjusted

with the knob (20). Also the auxiliary DC voltage is available at

sockets (63) after pressing the push-button (69), and can be

adjusted with the knob (21).

When the menu is accessed, the measurements are displayed

without the headings. If some further measurement is selected

(external measurements, power..) they are displayed below the

menu, and only while the menu is not accessed.

2.3 THE POP-UP MENU

The following is the list of features that are menu selected. The

menu is operated by means of the control knob marked MENU,

that incorporates a switch. The menu is entered pressing the

knob and selecting the item moving the knob. Once the item has

bee found and programmed, pressing the arrow the menu moves

back of one step, so that other programming can be performed;

else, selecting ESC the menu returns to the main window.

During this operation the display shows output measurements, in

reduced format. After confirmation, menu messages disappear,

and measurements are displayed in the standard format.

Any setting can be saved to and recalled from the memory. Up to

10 settings can be stored and recalled; setting no. 0 is the default

one, and pops up at power-on. Settings are permanently stored

in the memory; new settings can be written to the same address

after confirmation. For normal mode operation it is possible to

recall the standard setting, that cannot be modified.

During the test, test results can be stored in the memory (up to

500 results may be stored). At the end of test, settings and test

results can be transmitted to a PC provided with TDMS. The

software allows saving test results, examining them and so on.

The specification of TDMS is given in a separate document.

When the PC is connected, settings can also be created and

transferred into T 1000 PLUS using TDMS.

DOC. MIE92093 Rev. 1.34 Page 46 of 108

The flux diagram of menu selections can be found in Appendix 1.

LEVEL1 LEVEL 2 LEVEL

3

LEV. 4 FUNCTION

TEST

CONTROL

Test

mode

Normal

(default)

Measures the time

delay from START

(internal, external) to

STOP (internal,

external).

Trip + pulse

time

Measures the time

delay from START

(internal, external) to

STOP (internal,

external), and the

duration of STOP.

Reclose

mode

TD;

No.

reclose

Two delays are

measured: fault to

STOP; STOP to

START (reclose

command). At

START, a new fault is

generated after TD

(0-999.99 s), until

the number of

reclose (max 49) is

reached.

Fault

injection

Maintained

(default)

Generation lasts

indefinitely

Momentary Generation lasts until

the ON button is

pressed

DOC. MIE92093 Rev. 1.34 Page 47 of 108

External Generation starts

upon reception of the

START input: this

allows synchronising

T 1000 PLUS. When

selected, the fault is

injected as soon as

the selected

condition on START

input is met.

Timed Max

time

Generation lasts for

the pre-set time

duration. Max time

999 s.

OFF

delay

T delay The main output OFF

is delayed by the set

amount of time or

cycles.

Output

power

300 VA

(default) – 60

VA

Selection of full (300

VA) or reduced (60

VA) power

Save Don’t save

(default)

Test data are not

saved

Automatic, at

trip

As relay trips data

are saved to the next

memory location

Confirm, at trip As relay trips data

can be saved, after

confirmation

Manual When selected,

generated values are

saved.

Auxiliary

contacts

Timing Sets the contact

timings with respect

to test start

LEVEL

1

LEVEL

2

LEVEL

3

LEVEL

4

FUNCTION

DOC. MIE92093 Rev. 1.34 Page 48 of 108

TIMER

START/

STOP

Start INT (default) Timer starts when ON

or ON+TIME are

activated and outputs

generated.

EXT NO-NC-

EDGE

After ON or ON+TIME,

timer starts on the

external input. External

START input Normally

Open, or Normally

Closed, or Both

(EDGE).

CLEAN-

24 V – 80

V

After ON or ON+TIME,

timer starts on the

external input. External

START input without or

with voltage. If with

voltage, two voltage

thresholds are

available: 24 or 80 V.

COUNT Timer enters the

counting mode; it is

possible to program

the number of

transitions prior to

time measurement.

After ON or ON+TIME,

the test set waits for

these transitions

before measuring the

time.

Stop INT Timer stops when the

current of the main

generator is

interrupted.

EXT

(def.t)

NO-NC-

EDGE

(def.t)

Timer stops when the

STOP input is detected.

External STOP input

Normally Open or

Normally Closed or

Both (EDGE).

DOC. MIE92093 Rev. 1.34 Page 49 of 108

CLEAN-

24 V – 80

V

Timer stops when the

STOP input is detected.

External STOP input

without or with

voltage. If with

voltage, two voltage

thresholds are

available: 24 or 80 V.

COUNT Timer enters the

counting mode; it is

possible to program

the number N of

transitions to be

detected. After ON or

ON+TIME, the time

from the first valid

input to input N+1 is

measured; the

corresponding energy

can be read on the

display.

Timer s (default) Time duration metered

in seconds

cycles Time duration metered

in cycles

DOC. MIE92093 Rev. 1.34 Page 50 of 108

LEVEL

1

LEV.

2

LEV.

3

LEVEL

4

LEVEL 5 FUNCTION

AUX

VAC/

VDC

Aux

Vac control

Range 65 (default) ;

130 ; 260 V.

Mode

Fault (default)

The auxiliary

AC voltage is

adjusted by

the dedicated

knob, and is

available after

pressing the

ON-OFF

pushbutton. If

the auxiliary

voltage should

be applied

along with the

main current

or voltage, go

to next

selection.

DOC. MIE92093 Rev. 1.34 Page 51 of 108

Prefault

+ Fault

Prefault

Amplitud

e

Sets the pre-

fault auxiliary

AC voltage

amplitude.

Entering this

selection in

OFF mode, the

pre-fault

voltage is

generated and

displayed, and

adjusted by

the multi-

function knob.

NOTE: the

fault voltage is

generated

pressing ON or

ON+TIME, and

it is adjusted

as usual by the

knob.

Prefault

Phase

(0..359°)

Sets the pre-

fault auxiliary

voltage phase

with respect to

the fault

voltage; the

angle is

adjusted by

the multi-

function knob.

The pre-set

value is not

metered.

DOC. MIE92093 Rev. 1.34 Page 52 of 108

Prefault

duration

Sets the

duration of the

pre-fault

auxiliary

voltage. When

ON or

ON+TIME are

pressed, the

pre-fault will

be generated

at the mains

frequency for

the selected

duration; then

the fault

voltage is

generated, at

the

programmed

frequency.

Prefault

frequenc

y

The prefault

frequency of

the auxiliary

voltage may be

programmed.

The selected

frequency is

applied when

outputs are

OFF.

Freque

ncy

Locked to mains

(default)

If “Locked”,

the auxiliary

voltage is at

the mains

frequency.

DOC. MIE92093 Rev. 1.34 Page 53 of 108

Adjust

freq

40-

500.00

0

The frequency

of the auxiliary

voltage may be

programmed.

Frequency

changes at test

start; output

voltage does

not change in

amplitude.

Adjust

r.o.c.:

0.01..

9.99

Hz/s

The frequency

ramps at the

programmed

rate of

Change. The

starting

frequency can

be the mains

or the value

set by adjust

freq.

Phase Locked to

mains(default)

With this

selection Vaux

is in phase

with the

mains.

DOC. MIE92093 Rev. 1.34 Page 54 of 108

Adjust phase Vaux

- mains

The fault

auxiliary

voltage can be

phase shifted

with respect to

the mains. The

measured

angle is

displayed. Test

must be ON;

for a correct

angle

measurement,

the auxiliary

voltage must

be more than

20% of the

range. Phase is

adjusted by

the

multifunction

knob.

DOC. MIE92093 Rev. 1.34 Page 55 of 108

Adjust phase Vaux

– I main

The fault

auxiliary

voltage can be

phase shifted

with respect to

the main

current. The

measured

angle is

displayed. Test

must be ON;

for a correct

angle

measurement,

current and

voltage must

be more than

20% of the

range. Phase is

adjusted by

the

multifunction

knob.

DOC. MIE92093 Rev. 1.34 Page 56 of 108

Adjust phase Vaux

– V main

The fault

auxiliary

voltage can be

phase shifted

with respect to

the main

voltage. The

measured

angle is

displayed. Test

must be ON;

for a correct

angle

measurement,

both voltages

must be more

than 20% of

the range.

Phase is

adjusted by

the

multifunction

knob.

Aux

Vdc control

Range 130 V (default)

or 240 V. If

this selection

has to be

changed, it’s

necessary to

adjust the

voltage output

to the

minimum by

the dedicated

knob.

DOC. MIE92093 Rev. 1.34 Page 57 of 108

LEVEL1 LEVEL 2 LEVEL 3 LEVEL 4 FUNCTION

METERS Internal Units of I Normal If selected,

current

values are

displayed in

A.

I/IN I

N

If selected,

displayed

values are

defined as

I/IN, that

can be

defined.

Units of V Normal If selected,

voltage

values are

displayed in

V.

V/VN V

N

If selected,

displayed

values are

defined as

V/VN (phase

voltage),

that can be

defined.

Externa

l I

Enabled AC

(default) -

DC

With

selection AC

the meter

performs

the true rms

measureme

nt; with

selection

DC, the

measureme

nt is

performed

on the

average.

DOC. MIE92093 Rev. 1.34 Page 58 of 108

10A – 20

mA

Selects the

current

input socket

Waveform If selected,

the current

waveform is

displayed

Externa

l V

Enabled AC

(default) -

DC

With

selection AC

the meter

performs

the true rms

measureme

nt; with

selection

DC, the

measureme

nt is

performed

on the

average.

Shunt : 1

– 1000

mOhm

If the

voltage is

coming from

a current

dropping on

a shunt,

specifying

the shunt

value the

current is

displayed;

default 100

mOhm.

Waveform If selected,

the voltage

waveform is

displayed

DOC. MIE92093 Rev. 1.34 Page 59 of 108

LEVEL1 LEVEL 2 LEVEL 3 FUNCTION

METERS

(continue

d)

Other

internal

None (default) No extra

measurement

displayed

Active power P; W

Reactive power Q; VAr

Impedance module Z, Ohm

Impedance argument φ, °

Active impedance

component

R, Ohm

Reactive impedance

component

X, Ohm

Apparent power S; VA

Power factor p.f. = cos(φ V-I)

Active energy (AC) Ea; Wh

Reactive energy (AC) Er; VArh

Other

externa

l

None (default) No extra

measurement

displayed

Active power P; W

Reactive power (AC) Q; VAr

Impedance module Z, Ohm

Impedance argument φ, °

Active impedance

component

R, Ohm

Reactive impedance

component (AC)

X, Ohm

Phase, I (AC) φ, Vmain-Iext;

reference Vaux

Phase, V (AC) φ, Vmain-Vext;

reference Vaux

Apparent power (AC) S; VA

Power factor p.f. = cos(φ V-I)

Frequency of V (AC) f, Hz

Active energy (AC) Ea; Wh

Reactive energy (AC) Er; VArh

DOC. MIE92093 Rev. 1.34 Page 60 of 108

Note: measurements marked AC apply only if both inputs are

selected as alternate current.

LEVEL1 LEVEL 2 LEVEL 3 LEVEL

4

FUNCTION

RESULTS Test header Groups results

Show results Display result

Delete Selected

result(s)

All results

CONFIGU-

RATION

Settings Save to

address

1..10 Saves current

settings to X

Restore

address

1..10 Restores settings

from X

Restore

default

1..10 Restores default

settings

Language UK, FR, SP, PT,

GE, IT

Select the desired

language

Display Speed Slow The displayed

value is refreshed

every 1000 ms

Fast The displayed

value is refreshed

every 300 ms

Hold

mode

Hold

trip

As relay trips,

test data

measured 4

periods before

trip are held.

Hold

min

As relay trips, the

minimum value

within 0.5 s is

held.

Hold

max

As relay trips, the

maximum value

within 0.5 s is

held.

LCD

Contrast

It allows to adjust

the LCD contrast

DOC. MIE92093 Rev. 1.34 Page 61 of 108

3 WHAT’S INSIDE?

3.1 PHYSICAL DESCRIPTION

The test set is contained in an housing, to which it is fixed by

means of nine screws that are located below it: see the following

picture.

After removal, the test set is shown in the following picture.

MOUNTING SCREWS

DOC. MIE92093 Rev. 1.34 Page 62 of 108

The mechanical components are:

. The front plate, on which are mounted: the VARIAC transformer,

all interfacing components and the front board;

. The base plate, on which are mounted all other components;

. Four L-shaped legs on the corners plus a central rod keep

together the two plates.

The following is the list of the most important components

3) Main transformer for the auxiliary voltage outputs, XTF10330.

4) Transformer for the auxiliary supplies, XTF10245.

5) Main outputs transformer, XTF10345.

6) Main outputs adjustment VARIAC.

7) INTE ON-OFF printed circuit board; code YWA11400.

9) Auxiliary DC voltage board; code YWA11395.

10) AP-50 auxiliary AC voltage amplifier; code PWA11396.

11) Back panel board ; code PWA11399.

12) Output current measurement transformer.

14) Auxiliary voltage amplifier output transformer.

15) Microprocessor board code PWA41300.

DOC. MIE92093 Rev. 1.34 Page 63 of 108

16) Converter board code PWA11401.

17) Front panel board; code PWA11398.

24) Resistors set.

26) MISU T 1000 PLUS board code PWA11402.

27) Protection fuses board, code PWA11416.

The arrangement is shown in the figure and pictures below.

The picture above shows the base with the auxiliary DC and AC

modules removed. The following picture shows the front panel

with the front control board. 5

4

12

11

DOC. MIE92093 Rev. 1.34 Page 64 of 108

On the upper corner are mounted the control cards: they are:

MICROT 1000 PLUS, PWA41300; CONVT 1000 PLUS, PWA11401;

INTET 1000 PLUS, PWA11400. they are connected to the back-

panel PWA11399, and to the front panel via three flat cables.

6

17

26

DOC. MIE92093 Rev. 1.34 Page 65 of 108

Their location is shown in the drawing below. Cards are kept in

place by the piece of aluminium sheet on the right, and also by

the threaded bar between the boards. If a board is failing, it is

necessary to remove the two screws as shown, and to un-tighten

the nut of the threaded bar, as shown; in case some component

touches it, it is necessary to remove it.

The small MISUT 1000 PLUS board is mounted on the front panel.

The voltage amplifier AP50, PWA11396, and the DC voltage

module, YWA11395, are mounted on the base plate, and are

connected via cables and flat cables.

3.2 DETAILED FUNCTION DESCRIPTION

The following is the list of the most important boards: for each

board is provided the explanation of the key functions performed.

3.2.1 Main auxiliary transformer, XTF10330 (3)

The main auxiliary transformer (3) has the following secondary

windings. For connections, see the schematic at the end of this

manual.

DOC. MIE92093 Rev. 1.34 Page 66 of 108

VOLTAGE

V

POWER

VA

PURPOSE GOES TO

150 120 Supply of auxiliary DC voltage

module YWA11395 (9)

PWA11395

J800-1; 4

11.5 3 Auxiliary supply for auxiliary

DC voltage module

YWA11395

PWA11395

J800-2, 3

50 60 Low power test supply 2 WAY

CONN.

36 x 2 35 x 2 Power supply of auxiliary AC

voltage board PWA11396 (10)

PWA11396

J1-1; 2; 3

15.2 x 2 10 x 2 General auxiliary ± 15 V

supply for all analog circuits

PWA11399

JP 10

12 10 General auxiliary + 5 V

supply for all digital circuits

PWA11399

JP 8

10 10 General auxiliary + 12 V

unregulated supply (relays, ..)

PWA11399

JP 9

4.5 7 Unregulated + 5 V supply for

the display back-light

PWA11399

JP 7

3.2.2 Auxiliary transformer, XTF10245 (4)

The auxiliary transformer (4) has the following secondary

windings. For connections, see the schematic at the end of this

manual.

VOLTAGE

V

POWER

VA

PURPOSE GOES TO

7 + 15 7 + 5 V and ± 15 V

auxiliary supply for AC

voltage board PWA11396

(10)

PWA11396

J2-1, 2, 3

19 2 START wetting voltage PWA11399

JP 1

19 2 STOP wetting voltage PWA11399

JP 4

4.5 0.4 Mains synchronization

signal

PWA11399

JP 7

7 1 PWA11398

J817-1, 2

8 2 V DC measurement

circuit supply

PWA11398

J815-1, 2

DOC. MIE92093 Rev. 1.34 Page 67 of 108

10 1 Positive voltage for the

buffer of AC voltage

board PWA11396 (10)

PWA11396

J2-4, J1-6

10 1 Negative voltage for the

buffer of AC voltage

board PWA11396 (10)

PWA11396

J1- 4, 5

11.5 x 2 1.5 x 2 I measurement circuit

supply

PWA11398

J812-1, 2,

3

11.5 x 2 1.5 x 2 V measurement circuit

supply

PWA11398

J812-4, 5,

6

The following drawing shows the back panel board (11) and the

JP points, where many wires are soldered.

DOC. MIE92093 Rev. 1.34 Page 68 of 108

3.2.3 Main output transformer, XTF10345 (5)

The main output transformer (5) has the following secondary

windings. For connections, see the schematic at the end of this

manual.

VOLTAGE

V

POWER

VA

PURPOSE

225 +

296

125 Main DC voltage

Main AC voltage (same common)

90+

25+

10

400

300

300

10 A output (60 s maximum)

40 A output (60 s maximum)

100 A output (60 s maximum)

(same common)

3.2.4 Main front board PWA11398 (17)

The board includes many different circuits:

START/STOP contacts filters.

Back-light supply and contrast adjustment.

Push-buttons.

LED’s.

Three isolated AC to DC auxiliary voltage generators for the

following measurement circuits: 600 V AC external; I AC/DC

external; V DC internal.

Control and discharge of the filter capacitor for the main DC

voltage generation.

Reduced power selection relay.

Auxiliary relay.

Transformer and shunt for the auxiliary AC voltage and

current measurement.

At the end of the document are given the mounting layout and

the connectors disposition.

3.2.5 MISU T1000 board PWA11402 (26)

The board includes three isolated circuits for the

measurement of: 600 V external input; I external inputs (25

mA and 10 A); internal auxiliary DC V supply.

DOC. MIE92093 Rev. 1.34 Page 69 of 108

All circuits have: gain selections; high accuracy amplifiers;

high accuracy isolated op amps.

Analog outputs are taken to CONV1000, for the ADC

conversion.

3.2.6 CONV T1000 board PWA11401 (16)

TASK 1: to convert all analog measurement signals (the

above + main I + main V AC + main V DC + auxiliary V AC)

into digital signals. This is performed by two 12-bit, 4

channels each parallel ADC converters.

TASK 2: to generate a squared waveform in phase with the

mains input, and to square the AC current and voltage signals

for the phase meter. Three circuits with amplifiers, limiters,

comparators perform this task.

TASK 3: to generate a sinusoidal waveform, with controlled

frequency and amplitude, that will be passed to the auxiliary

AC voltage amplifier (10). This is made by the local micro DSP

processor that generates the inputs for two DAC’s, that

control the amplitude and the waveform, at 10 kHz.

On the board is located a set of 8 switches; their meaning is

summarized in the following table.

Dip

number

Meaning

1 CONV type: see dip 4

2 T 1000 PLUS at 15 Hz . ON = 15 Hz;

OFF = 40..70 Hz

3 T 1000 PLUS power. ON = more power;

OFF = normal.

4 CONV type: dip sw4 ON , dip sw1 OFF = CONV

rev 2

CONV type: dip sw4 OFF , dip sw1 OFF = CONV

rev 0

CONV type: dip sw4 ON , dip sw1 ON = CONV

rev 1

5 T 1000 PLUS normal / C: OFF normal, ON = C

6 AP50 : OFF normal, ON = 500V (T 1000 PLUS-

E)

7 Main Vac: OFF normal, ON = 500V (T 1000

PLUS-E)

8 Force download : OFF =no , ON = force download

DOC. MIE92093 Rev. 1.34 Page 70 of 108

3.2.7 INTE ON-OFF T 1000 board PWA11400 (7)

TASK 1: to detect the START and STOP inputs closed – open

situation via two isolated constant current generator circuits,

with wet/dry selection and opto-isolated logic output.

TASK 2: to drive the ON-OFF of the test set. There are two

SCR’s pairs (one per phase) in opposing mounting, mounted

on heat sinks. SCR characteristics: 600 V AC; 20 A steady;

200 A peak.

TASK 3: diagnostic circuits of the main auxiliary supplies.

3.2.8 MICR T 1000 board PWA41300 (15)

The microprocessor board is the same as the one of the DRTS

family, but without the two PLA’s. It is controlling the test set

operation, test and results handling and the communication with

the PC, but not the AC waveform generation, that is controlled by

a dedicated slave microprocessor.

3.2.9 AP-50 AC Voltage board PWA11396 (10)

The amplifier receives the AC signal to be amplified from

CONV1000 board. The signal is first isolated by an high accuracy

opto-isolation op amp; then it goes to the op amp that controls

the loop. The output goes to a level shifter and booster amplifier;

the power output is a class AB amplifier, made of a

complementary pair of power transistors, having the following

ratings: 140 V, 20 A, 250 W; TO-3 metal case.

The output of transistors goes to the primary of transformer (14),

that has three outputs: 65, 130, 260 V. These outputs are taken

to three relays of the same board, that perform the range

selection. The selected secondary goes to the AC voltage output.

Output power is 30 VA at the selected range; maximum currents

are: 460 mA at 65 V; 230 mA at 130 V; 115 mA at 260 V.

An additional circuit stabilizes the power DC supplies.

3.2.10 Protection fuses

DOC. MIE92093 Rev. 1.34 Page 71 of 108

By the side of the boards support is mounted a board with a set

of protection fuses. They protect the following modules:

. Auxiliary DC voltage supply: timed 0.25 A;

. Display backlight: timed 1.6 A;

. Internal supply + 15 V: timed 1 A;

. Internal supply - 15 V: timed 1 A;

. Internal supply + 5 V: timed 0.8 A;

. Internal supply + 12 V: timed 1 A.

The position of fuses is shown in the picture below.

The first one protects the auxiliary supply: if it is blown, there is

no DC voltage output.

The second one protects the LCD backlight: if it is blown, the LCD

display shows up, but the backlight is not there, so the reading is

difficult.

The other ones protect the supplies of electronic components: if

they fail. All the test set cannot operate.

3.2.11 Connectors summary

The following is the summary of all test set connectors: see the

schematic at the end of this manual.

J800: DC Voltage module (9)

PIN 1 2 3 4

VOLTAGE REF. (4) REF. (3) 11,5 V 150 V

TRANSFO (3) (3) (3) (3)

NOTE: REF. (x) is the reference zero of pin (x).

DOC. MIE92093 Rev. 1.34 Page 72 of 108

J812: MAIN BOARD (17)

PIN 1 2 3 4 5 6

VOLTAGE 11,5 V REF. (1-3) 11,5 V 11,5 V REF. (4-6) 11,5 V

TRANSFO (4) (4) (4) (4) (4) (4)

J815: MAIN BOARD (17)

PIN 1 2

VOLTAGE 8 V REF. (1)

TRANSFO (4) -

J817: MAIN BOARD (17)

PIN 1 2

VOLTAGE 7 V REF. (1)

TRANSFO (4) (4)

J820: MAIN BOARD (17)

PIN 1 2

VOLTAGE 4,5 V REF. (1)

TRANSFO (3) (3)

J1: AP-50 (10)

PIN 1 2 3 4 5 6

VOLTAGE 38 V REF. (1-3) 38 V 10 V REF. (4) REF. (J2-4)

TRANSFO (3) (3) (3) (4) (4) (4)

J2: AP-50 (10)

PIN 1 2 3 4

VOLTAGE REF. (2-3) 7 V 15,5 V 10 V

TRANSFO (3) (3) (3) (4)

DOC. MIE92093 Rev. 1.34 Page 73 of 108

4 THE HELL, IT DOESN’T WORK

4.1 INTRODUCTION

Sometimes, when my ears whistle, I wonder if it is because of

some of my customers being angry at us because the test set

doesn’t work: according to Murphy’s law, when it was most

necessary. We at ISA do our best efforts to filter the so-called

infant mortality of electronic components prior to delivery of all

our test sets; and this after extensive testing of prototypes and

pre-production units.

Yet, sometimes faults occur, because everything dies, including

electronic components; so, please, before shooting at us, see if

the following instructions can serve you to fix the problem. If not,

e-mail us the problem, not forgetting to mention the unit’s

serial number: our business is to minimize your downtime. My

e-mail address is:

[email protected]

Last, our experience is that our test sets withstand very heavy

duty cycles for long wiles, if correctly used; most problems arise

because of:

. Lack of grounding in the mains supply;

. Voltage or current neutral is connected to earth;

. Severe spikes on the mains (spikes are not always so kind to

respect standard limits);

. Transit, with the associated drops and thermal cycling;

. Errors, such as the connection to live wires, or generating

current on a short circuit with the handle completely turned to the

maximum current.

Please mention in your e-mail how did the fault occur: this serves

us for our continuous improvement program.

DOC. MIE92093 Rev. 1.34 Page 74 of 108

4.2 ERROR MESSAGES

The test set performs a number of tests at power-on and during

the generation. The following table lists all the messages, and the

corrective action.

A) TESTS AT POWER-ON (and runtime)

ERROR MESSAGE CONSEQUENCE CORRECTIVE

ACTION

+ 15 V supply failure If the error is not

confirmed it is

possible to continue;

else, the test cannot

proceed

(NOTE 1)

Try some power

on – off; if

persists, three

steps:

. Check the

fuses;

. Try to replace

the CONV board;

. Return the

test set.

- 15 V supply failure If the error is not

confirmed it is

possible to continue;

else, the test cannot

proceed

(NOTE 1)

Try some power

on – off; if

persists, three

steps:

. Check the

fuses;

. Try to replace

the CONV

board;

. Return the

test set.

NVM access If the error is not

confirmed it is

possible to continue;

else, the test cannot

proceed

(NOTE 1)

Fault in the

MICR board:

replace it

DOC. MIE92093 Rev. 1.34 Page 75 of 108

B) RUNTIME TESTS

ERROR

MESSAGE

CONSEQUENCE CORRECTIVE

ACTION

60 VA IAC supply

time

The test lasted

too long; tests

are blocked for

the specified

duration

Wait until the

message disappears

300 VA IAC

supply time

The test lasted

too long; tests

are blocked for

the specified

duration

Wait until the

message disappears

IAC while VDC

supplied

Generation is

stopped

The test set is

sensing a burden

both on main IAC

and VAC or VDC,

what is forbidden.

Remove one of the

burdens

MAIN VAC The main AC

voltage is

overloaded;

generation is

stopped

Reduce voltage or

burden of main AC

voltage

MAIN VDC The main DC

voltage is

overloaded;

generation is

stopped

Reduce voltage or

burden of main DC

voltage

AUXILIARY VAC The auxiliary AC

voltage is

overloaded;

generation is

stopped

Reduce voltage or

burden of the

auxiliary voltage

ERROR

OVERLOAD

RECOVERY xxx s

If you try to

restart before

the above

message

Wait until the

message disappears

DOC. MIE92093 Rev. 1.34 Page 76 of 108

disappears, the

test set informs

you about the

residual time

OVERTEMP

POWER

TRANSFORMER

Generation is

stopped

(NOTE 1)

The main

transformer

temperature is too

high because of

heavy loads or very

long test duration:

wait until it cools

down. If it does not

disappear, see the

text.

OVERTEMP

SCR’S HEAT

Generation is

stopped

(NOTE 1)

The temperature of

the SCR that drives

the test is too high

because of heavy

loads or very long

test duration: wait

until it cools down

OVERTEMP T

1000 PLUS

ENGINE

Generation is

stopped

(NOTE 1)

The test set

temperature is too

high because of

heavy loads or very

long test duration:

wait until it cools

down

NOTE 1: Sometimes the alarm is false, i.e. the fault is in the

diagnostic circuits (they also can fail), and the test set is OK. In

this case, you can proceed by pressing the buttons <, > and the

control wheel at power on. TAKE CARE, because this operation

skips ALL faults, and the test set is no more providing alarms in

case of too long test duration during high current generation:

MAKE THESE TESTS AS SHORT AS POSSIBLE!

4.3 TROUBLE SHOOTING

DOC. MIE92093 Rev. 1.34 Page 77 of 108

The following table summarizes main faults, cause of fault and

correction. When in the correction you find “see text”, please

proceed with the following paragraphs.

SYMPTOM POSSIBLE CAUSE CORRECTION

Problems after

transport

Heavy shock Open T 1000 PLUS

and check for

loose boards or

connections

At power-on does

not turn on

Mains supply fuse

open

Other causes

See text

The variable

transformer gets

very hot as soon

as the test set is

powered-on

The INTE board is

faulty

Replace INTE

After the general

over-temperature

alarm message it

does not turn on

The power

transformer was

over-heated

Wait at least 30

minutes before

trying again: it

needs time to cool

down.

The auxiliary DC

voltage is not

generated

The protection fuse

was broken by a

contact with a live

wire.

One protection fuse

is blown

Replace fuse (19).

Replace fuse on

the fuses board.

Replace fuses on

the module (see

text below)

The auxiliary

contact does not

close

The contact was

over-loaded and

permanently

damaged

Send back for

repair.

It is impossible to

measure on the 20

mA range

The input was

overloaded; the

PTC protection is

not restored

Wait 15 minutes

before trying

again.

No generation of I

main

Damaged INTE ON-

OFF

See text

No measurement

of I main

Broken connection See text

DOC. MIE92093 Rev. 1.34 Page 78 of 108

Measurements are

completely false

The microprocessor

has lost its

correction factors

Go to the

calibration

procedure and

repeat it.

The display is dark Protection fuse

blown

See text

Unstable auxiliary

AC voltage

measurement.

Fault of AC voltage

generation circuit

See text

The TRIP input is

not detected

Damaged INTE ON-

OFF

See text

Over-load alarm

on the auxiliary AC

voltage cannot be

restored; AC

voltage cannot be

generated.

Fault of AC voltage

amplifier.

Contact your agent

for support,

detailing:

. test set number;

. how did it occur

A knob

(adjustment,

multi-function

wheel) does not

turn

Friction against the

front panel

. Remove the

plastic (blue)

protection on the

front of the knob:

you gain access to

a nut;

. You can un-

tighten the nut:

the knob is free;

. Slightly withdraw

the knob,

interpose one or

two sheets of

paper between

knob and front

panel, and tighten

back the knob.

Removing the

paper, the knob

will be free of

turning.

DOC. MIE92093 Rev. 1.34 Page 79 of 108

4.3.1 Auxiliary supplies

On the edge of the CONV board are mounted a number of ring-

shaped test points. They carry the key auxiliary supplies, which

operate the test set. The arrangement of the test points is the

following.

If any of these voltages is missing, the test set cannot operate: it

is necessary to find out if it is a problem of blown fuse, or repair

the auxiliary voltage circuit. If the main auxiliary transformer (3)

is damaged, it can be replaced with the following procedure.

The replacement transformer comes cabled with all crimped

connectors, in order to ease the replacement. There are eight

remaining wires that are to be soldered directly to the small fuse

board. The two transformers are perfectly identical: wires come

out from the same position and have the same colours. The

replacement procedure is he following.

- Unscrew the faulty transformer.

- Place the new transformer in the same position.

- Unsolder the first wire from the fuse board.

- Solder in the same place the new wire fro the new transformer,

having the same colour.

- Proceed until all wires have been soldered.

4.3.2 No power at power-on

The mains fuse is located in the power supply plug, in the small

drawer. Replace it.

At power-on, almost no power (50 W) should be drawn from the

mains. If the power is more, the most likely faulty component is

the auxiliary DC voltage circuit; next, it is the auxiliary AC voltage

circuit.

To ascertain this, it is necessary to remove:

DOC. MIE92093 Rev. 1.34 Page 80 of 108

. three connectors from the DC voltage generator: two bigger,

white, one on the front, one on the rear, plus one flat cable;

. Five connectors from the AC voltage amplifier: four white,

mounted on top of the board, plus a flat cable.

After this, power-on, and verify that the display operates, and

that the LED’s on the front panel turn on with the default values.

If so, re-insert the modules one after the other, in order to detect

which one is faulty.

In case of DC voltage generator, see here below what else to do;

in case of the AC voltage amplifier, it should be replaced.

4.4 AUXILIARY DC VOLTAGE FAULT

If the output voltage cannot be adjusted, then the fault can be

located in the adjustment potentiometer. In this instance, the

behaviour is the following:

- Select the 130 V range: the output is fixed at about 10 V;

- Select the 240 V range: the output is fixed at about 20 V.

The potentiometer is easily replaced as follows:

. Remove the test set from the container;

. The potentiometer is soldered to the control board by two wires:

unsolder them;

. Remove the plastic cap from the knob: the nut blocks the knob

to the potentiometer. Un-tighten the knob and remove it: you get

access to the nut that holds the potentiometer to the front panel.

. Unscrew it, replace the potentiometer, screw it back, solder

wires, mount the knob so that it does not touch the front panel.

Take care not to exchange the four-wire connector from

the transformer with the four-wire connector to the AC

voltage amplifier: see the picture below.

DOC. MIE92093 Rev. 1.34 Page 81 of 108

If the potentiometer was not the problem and fuse (19) is OK, it

is possible that the fuse on the fuses board is blown, or that one

of the fuses mounted on the board has blown up. In this instance,

proceed as follows.

. Open the test set, as explained above, and locate the fuse on

the fuses board. Check it and replace if necessary.

. If OK, locate fuses on the auxiliary DC voltage board (see

picture below).

. Check fuses and replace them if necessary. They are 20 mm

long; F1 is F3.15A; F2 is F2A.

. If a fuse is blown, replace it, power-on and check if the DC

voltage is available. If it is OK, the fuse was weak; if it is not,

there is a fault on the board, that needs to be replaced.

4 wires with different colours: they come from

the transformer

4 wires with the same colour

4 wires with the same colour

4 wires with different colours: they come from the transformer

DOC. MIE92093 Rev. 1.34 Page 82 of 108

The last test is to remove the connector J800 coming from the big

round power transformer, and located on the rear of the module.

The connector has 4 pins: the voltage between pins 1 and 4

should be 155 V AC nominal; between pins 2 and 3 should be

11.5 V AC nominal. If this is correct, the module is faulty; else,

the power transformer is faulty.

The module can be removed by un-screwing the two screws

below it, and it can be withdrawn from its mounting position

without removing other cables.

When setting back the board, if the short cable with 4-way

connectors has been removed on both sides, take care because it

is not symmetrical: you have to mount it back so that the ferrite

that protects two wires is towards the outside of the test set

(bottom side).

F1 F 3.15 A

F2 F 2 A

DOC. MIE92093 Rev. 1.34 Page 83 of 108

4.5 AUXILIARY AC VOLTAGE FAULT

Possible modes for AC voltage fault are:

. Over-load message at power-on, that cannot be re-set;

. Over-load message even with a minimum load;

. No AC voltage output.

In such instances, the most likely cause of fault is the amplifier

(10), PWA11396. You should open the test set, disconnect the

five connectors, remove the amplifier by unscrewing the four

screws below it, replace with the new one and verify that the test

set is operational. In order to be able to remove the amplifier,

first remove the DC voltage module (two screws), so that you can

take it away moving it to the left.

When fitting back the connectors, cables are short or with

different pin numbers, so that they cannot be exchanged, unless

for two 6-way connectors towards the bottom of the test set. The

correct location is sketched below (see the picture above)

If, after the replacement of the amplifier, the problem is still

there, it is possible to go to a further investigation as follows.

. Open the test set, and look at the AP-50 amplifier from the

edge, and below it, at the location of the connector that receives

the flat cable: see the picture below.

DOC. MIE92093 Rev. 1.34 Page 84 of 108

. Verify the signals as summarized in the following table.

PIN 1 2 3 4 5 6 7

SIGNAL Ext meas 80 V

0 Vin 0 Vin 0-2 V AC

Range sel

H = 260 V;

L = 130

V

+ 12 V Ext meas 10 V

PIN 8 9 10 11 12 13 14

SIGNAL 0 Range sel H =

260/130 V;

L = 65 V

- 15 V

+ 15 V 0 H =close

L = open relay

+ 5 V

The AC signal between pins 2 and 4 is coming from the CONV

board: it is 2 V AC when the output is at maximum.

4.6 NO OUTPUT FROM THE MAIN CURRENT AND VOLTAGE

This is the case when you can start the test, but no output is

generated from the main current and voltage outputs. In this

situation the fault is located in the INTE ON-OFF board (7), that is

located in the boards pack. The replacement is very fast: just

PIN 2

PIN 1

DOC. MIE92093 Rev. 1.34 Page 85 of 108

follow the instructions above for the opening of T 1000 PLUS and

board replacement.

4.7 DOES NOT MEASURE THE MAIN CURRENT

Remove T 1000 PLUS from the container. Just below the output

sockets there are two yellow wires going to the metering current

transformer. These yellow wires are made of a single enameled

wire coming from the secondary of the measurement

transformer: it is possible that they broke, and so you loose the

measurement. The fixing is easy: just solder the wires, taking

care to peel the enamel so that it can be soldered.

If this is not the cause of the problem, further tests:

. On the main board, close to the point where the transformer is

soldered, is located the resistor R71, rated 1 Ohm: verify that it is

correctly soldered.

. Generate 1 A on the 10 A output, and measure the voltage

across the 1 Ohm resistor: you should measure 50 mV

(Transformer ratio 200//1; 10 turns make the ratio 20//1).

. The voltage drop across the shunt is taken to the CONV board

via a 12-way flat cable, of the type FLEXSTRIP, located in the

center of the board. These connectors are rather hard to fit: press

it very well on both sides.

. Last, it is possible that there is a fault on the CONV board where

the measurement is performed: if this is the case, the board has

to be replaced. Prior to this, verify that the voltages on the CONV

board test points (see 3.4.2 above) are correct.

4.8 THE BACKLIGHT OR THE DISPLAY DOES NOT OPERATE

1. The display backlight does not turn on.

The backlight is supplied by a DC to DC converter that is mounted

on the main board, that is mounted below the front panel. If the

backlight does not turn on, the converter is likely to be faulty.

The converter replacement is performed as follows.

. Remove T 1000 PLUS from the container.

. Go to the left side: the sketch below shows the main

components.

. The converter is mounted as shown in the design below.

DOC. MIE92093 Rev. 1.34 Page 86 of 108

. First of all, verify that the two-way, AMP type white connector is

correctly fit, and that wires are not loose.

. Next, try to re-solder the four connecting pins of the converter

to the main board: we once had a case of poor soldering.

. If this is not enough, it is necessary to replace the converter. To

this purpose, we suggest to remove the faulty one by cutting pins

as short as possible, so that cleaning holes is easier; then, solder

the new converter.

2. The display does not operate.

In this situation, verify the following.

. The three FLATSTRIP cables from the control board back panel

to the main board could be not completely fit in, as they are

rather hard: check it.

DOC. MIE92093 Rev. 1.34 Page 87 of 108

. Verify the voltages on the test points of the CONV board, as

explained above. If some voltage is missing, then the auxiliary

transformer should be replaced.

. The display is connected to the main front board via a number

of pins, located in the center of the display: verify that there is no

dirt causing a short circuit between them.

. If everything is OK, there is also the possibility that the MICR

board is faulty: replace it.

. If changing the MICR board the fault is still there, it is necessary

to return the test set to ISA for the replacement of the display,

that involves a difficult labor.

DOC. MIE92093 Rev. 1.34 Page 88 of 108

4.9 THE AC VOLTAGE MEASUREMENT IS NOT STABLE

First of all, check that there is no unplugged connector or broken

wire. Next, it is necessary to understand if the problem is in the

measurement or if the output is actually unstable. To this

purpose, connect the output to an AC voltage meter and verify

the reading. If it is unstable as well, the problem is in the

amplifier; otherwise, it is in the measurement circuit.

In the first instance, amplifier, we have to verify if the problem

comes from the adjustment potentiometer. To this purpose,

proceed as follows:

. Go to the menu and select PREFAULT + FAULT;

. In the PREFAULT mode, the output voltage is adjusted by the

multi-function knob, and not by the potentiometer: please verify

the output. If it is stable, the problem is located in the

adjustment knob. If it is unstable, it can be located on the AP/50

amplifier, or on the CONV board. The problem when replacing

CONV is that, as it hosts the measurement circuits, after

replacing it, it is necessary to repeat the calibration. It is up to

you if you want to receive the replacement boards or to deliver

the test set to us.

In the second instance, the fault is in the measurement, we have

to replace the CONV board, with the problem explained above.

4.10 THE TRIP INPUT IS NOT DETECTED OR TIMING ERROR

There is a possibility that the problem is coming from the

selections performed on the input. In fact, we can select the

contact as free of voltage or with voltage (LED's on the front

panel show the active selection); besides, if it is with voltage, we

can select two thresholds: low for voltages above 24 V; high for

voltages above 110 V. In case of doubt, please select the default

settings: both inputs are selected without voltage. Note that if

you select the inputs as voltage free, and apply 12 to 24 V, the

detection becomes uncertain.

When the voltage clean trip input is selected, a voltage of about

20 V DC can be measured on START or STOP inputs; when the

DOC. MIE92093 Rev. 1.34 Page 89 of 108

contact closes, a current of about 10 mA flows though it and

detects the contact status. If with this selection the closed contact

is not detected, and 20 V are not applied, the fault is located in

the INTE ON-OFF board (7), that is located in the boards pack.

The replacement is very fast: just follow the instructions above

for the opening of T 1000 PLUS and board replacement.

If the replacement of the INTE ON-OFF board is not solving the

problem, check the following.

. There are two connectors on the front board: they carry the

START and STOP inputs. Check that they are OK.

. These wires are soldered to the back-panel board, on test points

marked START and STOP respectively. Test that they are

correctly soldered by putting a jumper into the sockets and

verifying that the wires are short-circuited.

STOP

START P

DOC. MIE92093 Rev. 1.34 Page 90 of 108

. Last, on the test points shown above as STOP SUPPLY and

START SUPPLY are soldered the wires that supply the START and

STOP circuits for the voltage free selection, and that come from

the auxiliary transformer located below the boards: if a voltage is

missing, the transformer should be replaced.

If there is a doubt about the timing metered by T 1000 PLUS,

please proceed as follows.

. First of all, go to the CONFIG selection; settings, and select

RESTORE DEFAULT: this avoids that some exotic setting is the

cause of the problem.

. Next, connect a jumper to the STOP input. Start the ON+TIME

test: the timer should display 0 to 2 ms.

. Now, go to the TEST CONTROL selection, and to the AUXILIARY

CONTACT. Set the timing of 10 ms. Connect the C, NO contacts of

the auxiliary output to STOP inputs. Start ON-+TIME: it should

display 10 to 12 ms. You can test more with different timings of

the auxiliary contact.

4.11 PROBLEMS DURING UPGRADE

If during the Upgrade operation the power went off, and the test

set is no more operational, to recover the situation proceed as

follows.

DOC. MIE92093 Rev. 1.34 Page 91 of 108

- Remove (or unlock a little) the metal bar that fixes the two

aluminium columns.

- Remove the two connectors placed on the INTE board.

- Remove the INTE board, and then the pack MICRO + CONV.

- Open the pack MICRO + CONV.

- Change the position of the mini-dip switch N. 8 on the CONV

board (it is OFF; it must be set ON).

- Insert again all the boards (and the INTE connectors too).

- Turn on the test set.

- At this point , the instrument should be connect to your PC.

Execute the firmware download.

Verify that the new firmware version appears; if yes, you have

to turn off and set the dipswitch N. 8 to OFF (as it was initially).

4.12 THE ENCODER IS BROKEN

In this instance, it is necessary to have some skill and patience.

The procedure is the following.

. During the time, we have used three different type of encoder

mounting; they are:

1. The front-end printed circuit board does not allow to

access the encoder;

2. The front-end printed circuit board allows to access the

encoder, which does not carry a small PCB;

3. The front-end printed circuit board allows to access the

encoder, which carries a small PCB.

Before proceeding, shoot a picture at the encoder, so that we can

understand your situation. If you recognize the case 1, the test

set should be returned us.

. You have in your hands the new encoder: it must replace the

faulty one. This means that you have to dismount the old one,

unsolder the five wires going to it, solder them to the new one,

mount it.

. It is necessary to gain access to the faulty encoder. If you look

at it, it is surrounded by other components: no access. It is

necessary to remove the DC voltage generator board, and also to

remove the variable transformer: remove the control knob, and

dismount it removing the three screws below the knob.

. Now you have room enough to operate. Remove the encoder:

the five wires going to it are long enough to allow you keeping it

in your hand.

DOC. MIE92093 Rev. 1.34 Page 92 of 108

. Unsolder the first wire, and solder it to the new encoder, in the

same position: with this procedure you are sure not to exchange

wires.

. At the end of soldering, mount back the new encoder, power-on

and check that it is correctly operational. After this, complete the

mounting of the test set.

4.13 THE THERMAL ALARM DOES NOT DISAPPEAR

The thermal alarm is given by either of two thermal switches put

in series: one is located on the main transformer; the other one is

located on the AP-50 amplifier. One of them is faulty.

In order to understand which one is faulty, proceed as follows:

. Dismount the AP-50 amplifier (four screws from below), and lift

it enough to reach for the thermal sensor: it ends with two wires,

that go to a connector.

. Disconnect the connector, and perform a temporary short circuit

on the mating connector.

. Power-on. If the error signal disappears, the fault is on the AP-

50 sensor (unless AP-50 is actually hot); else, the fault is on the

sensor mounted on the power transformer (5).

. On the Main Output Transformer there is a thermistor with two

pink wires leaving it, screwed below the holding disk. Its value at

rest is 10 kOhm. For a temporary fix, replace it with a 10 kOhm

resistor until the replacement arrives.

4.14 AN AUXILIARY SWITCH DOES NOT OPERATE

Auxiliary switches are protected by fuses located on the front

board, inside the test set, and connected in series to the C

contact of the relay. If one of the two is not operating, very likely

the fuse is opened. The intervention procedure is the following.

. Open the test set.

. Protection fuses are located by the side of the auxiliary relay:

see the picture.

DOC. MIE92093 Rev. 1.34 Page 93 of 108

. The fuse to the right protects the auxiliary contact no. 1.

Replace the fuse and verify again.

. If the fuse is not the problem, it is possible, even if quite

difficult, to unsolder the relay and to replace it.

4.15 THE FAULT CANNOT BE FIXED

If the fault is too hard to be fixed, you have to deliver T 1000

PLUS back to your agent. We have encountered problems caused

by a poor packing of instruments that have been delivered us for

calibration or repair. In order to avoid such inconveniences,

please apply the following procedure.

First of all, compile the following form, and attach it to the

instrument. Please do not forget to compile it.

With the instrument should come the mains supply cable. The

user’s manual originally delivered with the test set is not

necessary.

Cover the instrument itself with a polyester film, in order to

protect it against dust and foam.

The instrument should be protected by anti-shock foam having a

minimum thickness of 5 cm ON ALL SIDES.

Use a new carton box as a container. On the box apply the UP

and the FRAGILE labels.

In the box the instrument will be placed horizontal or standing;

not upside down.

If the set is heavier than 20 kg it is better to use also a pallet:

this ensures that the box will not be packed upside down.

Last but not least, do not declare an high value for customs: this

expedites clearance of the good and lowers fees.

DOC. MIE92093 Rev. 1.34 Page 94 of 108

INSTRUMENT RETURN FORM

DATE ________________________________________

AGENT _______________________________________

COUNTRY _____________________________________

TYPE OF INSTRUMENT __________________________

SERIAL NO. ___________________________________

INSTRUMENT RETURNED FOR:

CALIBRATION ____

REPAIR ____

In case of repair, please specify the following.

DATE OF FAULT _________________________________

REPORTED BY E-MAIL, PHONE ______________________

COMPANY ______________________________________

USER’S REFERENCE _______________________________

FAULT DESCRIPTION

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

DOC. MIE92093 Rev. 1.34 Page 95 of 108

HOW DID IT OCCUR

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

LOCAL ANALYSIS OR ATTEMPTS TO REPAIR

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

RECOMMANDATIONS AND NOTES

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

_______________________________________________

DOC. MIE92093 Rev. 1.34 Page 96 of 108

4.16 CALIBRATION

4.16.1 Introduction

T 1000 PLUS does not need to be calibrated, as metering circuits

employ high stability components.

It is suggested to check the unit every 2 years, by comparing T

1000 PLUS measurements to external meters. Tests should be

performed with an high accuracy multi-meter, that should

guarantee a maximum AC measurement error of 0.1%, both for

voltage and current. Besides, as such multi-meters do not have

current ranges greater than 2 A, for the test of the high current

ranges it is necessary a class 0.1 measurement Current

Transformer. The adoption of lower-class instruments may cause

false interventions, that introduce errors into the test set. . Last,

as internal generators follow the mains, it should be stable

enough to allow for a stable generation.

At the end of the test, if deviations are not acceptable it is

possible to enter the calibration mode, as explained in the

followings.

4.16.2 Calibration procedure

The calibration mode is accessed by pressing push-buttons < and

> at the meantime, and powering on T 1000 PLUS. After a

confirmation message, the following window is opened.

The calibration is divided in two: measurements and phase

angles.

For the measurements, after the selection, the following window

is opened.

DOC. MIE92093 Rev. 1.34 Page 97 of 108

4.16.3 T 1000 PLUS output calibration

Calibrations are divided between external measurements and

internal generators. Starting from an internal generator, for

instance Main IAC, the following window is opened.

The three current outputs are displayed; for each output are

shown the measurement ranges that the test set will

automatically select, according to the input value. There are two

columns of squared dots for each range: the left one, marked of,

refers to the off-set calibration; the right one, marked m, refers

to the measurement calibration. Squares are white when the

calibration has not yet been performed, then they become black.

The operating mode is the following:

. Select the range to be calibrated.

. Set to zero the adjustment knob.

. With no output, ad just the off-set: it will be performed with an

open circuit for current outputs, and with a short-circuit with

voltage outputs.

Following the example of main IAC output, the following is the

window for the calibration of the 1 A range.

DOC. MIE92093 Rev. 1.34 Page 98 of 108

. The first thing is the off-set calibration: to this purpose, remove

the connection to the meter, then select RESET: as you click on

it, the square dot to the left side turns white:

. Now, press SET: the square dot turns black, and the calibration

is done.

. On the upper part of the screen it is displayed the nominal

current to be used for the calibration of the selected range: in our

instance, it is 1 A. To the right, there are two adjustment

parameters: COARSE and FINE, and then the measurement. The

CARSE and FINE values are just calibration parameters, with no

direct meaning for the measurement.

. Connect the meter to the output to be calibrated; then, press >

to generate the output. The external meter measures the output;

T 1000 PLUS displays its measurement. Adjust the current until

the external meter displays a value within 5% of the nominal

calibration value. NOTE: on order to ease the calibration, you can

use the power resistors to put them in series to the output.

Go to the COARSE adjustment, and modify it (reduce it to

increase the measurement) until the T 1000 PLUS measurement

is the minimum less than the external one; then, go to the FINE

adjustment (increase it to increase the measurement) in order to

get the best match with the external meter (the difference should

be less than 0.2% of the reading).

. Press the return icon to confirm. Go to the next range: T 1000

PLUS goes OFF.

. In our instance, next range is 19.99; the calibration value is 11

A. Repeat the above procedure, and then do the same for all

ranges of the other outputs, 40 A and 100 A.

DOC. MIE92093 Rev. 1.34 Page 99 of 108

. Next is the calibration of the AC voltage output: the window is

the following.

.

. Here there are four ranges to calibrate. The procedure is the

same as above, but the off-set is calibrated by putting a short

circuit on the output.

. The main DC voltage output is similar to the above. The

auxiliary AC voltage output calibration opens the following screen.

. Also here, the calibration is performed with the output short-

circuited. The auxiliary DC voltage output window is the following.

. In this instance, DON’T MODIFY THE OFFSET CALIBRATION: the

minimum value for the DC voltage is not zero. Just calibrate the

measurement.

4.16.4 T 1000 PLUS external meas. calibration

DOC. MIE92093 Rev. 1.34 Page 100 of 108

Now you can proceed with the external measurements

calibration. For this operation, IT IS NECESSARY TO HAVE AN

EXTERNAL SOURCE, SUCH AS A DRTS TEST SET: T1000 ittself

cannot be used to source the inputs.

. The window for the external IAC measurement is the following

one.

. NOTE: the first two calibrations will be performed on the 25 mA

input; the other two on the 10 A input. Also here, first calibrate

the offset (open circuit, and then the gain, inputting a current

close to the one displayed.

. Calibrations are divided in two: AC calibration and DC

calibration. The following is the widow for the 600 V input.

4.16.5 Phase angle measurement calibration

. Once these calibrations are finished, you can calibrate phase

angle measurements. To this purpose it is necessary to use a

phase meter with a measurement error of 0.2° maximum. The

phase angle measurement calibration is performed taking into

account the selected phase reference. The first window that

opens is the following one.

DOC. MIE92093 Rev. 1.34 Page 101 of 108

1) Angle Auxiliary AC Voltage (reference) – Main AC

current.

Once the selection is performed, the following window is opened.

The calibration has to be performed for each voltage and current

range: nine calibrations in all. Connect the auxiliary AC voltage

output to input 1 of the reference phase meter, and the main

current output to input 2 of the reference phase meter. Take

care to connect the T1000 black sockets to the negative

inputs of the phase meter.

Press the key >: T1000 goes ON. Adjust at 57 V the auxiliary AC

voltage output, and at 5 A the main current output. Go to the first

angle to be adjusted, and press the button: the squared dot will

turn white, and the phase meter to the right displays the internal

measurement.

Rotate the knob until the internal phase meter measurement is

equal to the external one : pressing the knob, the correction

factor is saved, and the square dot turns black.

The adjustment will be repeated on all main current outputs, with

current values of 10 A for the 40 A sockets, and 20 A for the 100

DOC. MIE92093 Rev. 1.34 Page 102 of 108

A sockets: the auxiliary AC voltage will be the same for all

ranges.

2) Angle Auxiliary AC Voltage (reference) – Main Voltage.

Once the selection is performed, the following window is opened.

Connect the auxiliary AC voltage output to input 1 of the

reference phase meter, and the main AC voltage output to input

2 of the reference phase meter. Take care to connect the

T1000 black sockets to the negative inputs of the phase

meter.

Both voltage outputs are adjusted at 57 V. on the 62.5 V range,

and the main voltage output, The adjustment procedure is the

same described above.

3) Angle Auxiliary AC Voltage (reference) – External

Current.

Once the selection is performed, the following window is opened.

Connect the auxiliary AC voltage output to input 1 of the

reference phase meter, and the external AC current input to

input 2 of the reference phase meter. Take care to connect the

T1000 black sockets to the negative inputs of the phase

meter.

The external AC current can be generated by the main AC current

generator; it is advisable to connect in series a resistor of the

DOC. MIE92093 Rev. 1.34 Page 103 of 108

resistor set. Press the key >, and adjust 1 A for the first

calibration, and 5 A for the second one. The procedure is the

same as above.

4) Angle Auxiliary AC Voltage (reference) – External

Voltage.

Once the selection is performed, the following window is opened.

Connect the auxiliary AC voltage output to input 1 of the

reference phase meter, and the external AC current input to

input 2 of the reference phase meter. Take care to connect the

T1000 black sockets to the negative inputs of the phase

meter.

For this adjustment, you can use the Auxiliary voltage output as

the external voltage: they will surely be in phase. However, as

the smallest External voltage range is 20 V, it is necessary to

reduce the 65 V Auxiliary voltage output to a smaller value: this

is performed taking advantage of the auxiliary resistors set.

Connect the Auxiliary voltage output to the 2200 Ohm, and the

Auxiliary voltage zero to 100 Ohm; the External voltage input is

taken from the partition, as shown in the following figure.

DOC. MIE92093 Rev. 1.34 Page 104 of 108

With this connection, the voltage to VEXT is about 10 V when the

input is 65 V: this is to be used for the angle calibration of the 20

V External input range. For other ranges, the partitioning is not

necessary.

The program alerts if a scale is not calibrated. At the end of the

session, all parameters are saved in the non volatile FLASH

EPROM.

common

black

DOC. MIE92093 Rev. 1.34 Page 105 of 108

APPENDIX 1 SPARE PARTS LIST

The following is the suggested list of spare parts for a duration of

five years and for up to five test sets.

The reference is made to the list of components.

1) N. 10 Fuses in the supply socket (1), rated T10A, code

XFU22101.

2) N. 10 Fuses of the auxiliary DC voltage (19), rated F2A, code

XFU23109.

3) N. 1 DC to DC converter for the display (23) backlight, code

XCA10348.

4) N. 1 Encoder with switch, for MENU control knob (22), code

XCM10160.

5) N. 1 Printed circuit board INTE ON-OFF (7), code PWA11400.

6) N. 1 Auxiliary DC voltage supply module (9), code YWA11395.

7) N. 1 Auxiliary AC voltage amplifier AP-50 (10), code

PWA11396.

J819- 3

RO SSO

BLU 0V

100A

0.5 QUADRI

CONNETTORE VOLANTE

NERO 0V

M ARRO NE

XTF20404

1 3

4

2

5

6

0.5 QUADRI L= cm

VCC

BI ANCO

12 PIN

1

G I ALLO 0V

ARANCI O NE 0V

BLU 0V

0.25 QUADRI

F

BLU 4. 5V( 11 )

BI ANCO

BI ANCO 0V

BLU

0

VISTA LATO FILI

TP14

I NTE O N- O FF PWA11400

J804- 1

J804- 2J808

VI O LA

XTF10245

1

4

2

32.5 QUADRI

MODIFICA 1

NERO

0.5 QUADRI L= cm

0.5 QUADRI L=20 cm

P1

SERI ALE RS232

594837261

M

BI ANCO

0.25 QUADRI

( 17 )

G I ALLO 11. 5V

M ARRO NE

PICCOLO

12 PIN

( 18 )

YWA11395 VCC

J806

J818- 1

J818- 2

J818- 3

J818- 4

J800- 1

J800- 2

J800- 3

J800- 4

BI ANCO 296V

BLU

J818

0.5 QUADRI

M

BLU 8V

22R

0.5 QUADRI L=20 cm

RO SSO 0V

0.25 QUADRI

( 8 )

4

BLU

( 9 )

Fr iday, Febr uary 14, 2003

10093 3

CABLAG G I O T1000

A2

1 1

Tit le

Size Docum ent Num ber Rev

Dat e: Sheet of

Designer Check and Approval

0.5 QUADRI L=20 cm

12 cm 0.5 QUADRI

CONNETTORE DI XTF20404

0.5 QUADRI

F

G I ALLO 225V

0.5 QUADRI L=20 cm

BLU 0V

NERO 10V

0.25 QUADRI

( 10 )

0

NERO 0V

NERO 0V

NTC 10K

1 2

PRESA + FI LTRO

T

F

N

0.5 QUADRI L=20 cm

TP13

1

12

10 GIRI

4

VERDE

BI ANCO

XTF10330

1

4

2

3

VARI AC

0.5 QUADRI L=20 cm

VI O LA

TERRA

0.25 QUADRI

9 POLI

( 14 )

( 12 )

PWA11398 BP APPO G G I O

TP17

TP18

TP19

TP20

TP9

TP10

TP11

TP12

TP13

TP14

TP3

TP6

TP5

TP4

TP2

TP15

TP16

J809- 1

J809- 2

J809- 3

J809- 4

J810- 1

J810- 2

J810- 3

J810- 4

J821

J818- 1

J818- 2

J818- 3

J818- 4

J819- 2

J819- 3

J813- 1

J813- 2

J813- 3

TP7TP8

J801

J802

J803

J812- 1

J812- 2

J812- 3

J812- 4

J812- 5

J812- 6

J816- 1

J816- 2

J816- 3

J816- 4

J816- 5

J816- 6

J817- 1

J817- 2

J815- 1

J815- 2

J820- 1

J820- 2

J811- 1

J811- 2

J808- 1

J808- 2

J819- 5

J819- 4

J819- 1

J819- 6

470R

CONNETTORE VOLANTE

0.5 QUADRI L=20 cm

( 19 )

DI SPLAY

-

-

VARI STO R275V J819- 3

BI NCO

G I ALLO 11. 5V

RO SSO

VERDE

0.5 QUADRI

0.5 QUADRI L=20 cm

30 cm

5uH

0.25 QUADRI

( 1 )

( 13 )

PWA11396 AP50

J5

J2- 1

J2- 2

J2- 3

J2- 4

J4- 1

J4- 2

J4- 3

J4- 4

J4- 5

J4- 6

J6

J1- 1

J1- 2

J1- 3

J1- 4

J1- 5

J1- 6

40A

VCC

( 20 )

0.5 QUADRI L=8cm

M ARRO NE 90V

0.5 QUADRI L=20 cm

12

CO NV T1000

J805

J802

VERDE 0V

RO SA 0V

12

1

FUSIBILI

230V = T5A 250V

110V = T10A

250VM ARRO NE 38V

G RI G I O 0V

1

M ARRO NE 19V

BLU

SW1TERM O SWI TCH

12

VI O LA 0V

1 QUADRO FASTON

VCA

100K M ULTI G I RO

13

2

VAC

( 21 )

0.5 QUADRI L=8cm

BLU 0V

BLU 10V

RO SSO

RO SSO

1KSOPRA

1

M ARRO NE 0V

4 GIRI

12 GIRI

1R

110V

VI O LA

RO SSO

0.25 QUADRI

12 POLI

( 22 )

( 2 )

G I ALLO 11. 5V

100K M ULTI G I RO

13

2

1234

BLU 10V

XTF10345

1

5

6

4

7

8

9

10

2

3

1112

13

14

M ARRO NE 38V

12

4

F

RO SSO 0V

J6

123456

PASSANTE

TP10

VERDE

1 QUADRO FASTON

110V

J4 ( AP50 )

VCC

( 23 )

( 3 )

BLU

T2XTF10245

1

5

6

4

78

910

1112

1314

15

18

19

2

3

24

2526

2728

29

3031

32

33

AC

VERDE 25V

6K8

M ARRO NE

1

M

TP11

BI ANCO

1

1234

110V

BI ANCO 11. 5V

VERDE 15. 2V

VERDE 15. 2V

0.25 QUADRI

( 4 )

NERO

28 cm

M

J816 ( BP )

TP9

ARANCI O NE 0V

RO SA 4. 5V

XTF20034

G I ALLO

VERDE 10V

FUSI BI LE T2A

10A

1.5 QUADRI

BI ANCO 0V

4.5cm

0.25 QUADRI

0.5 QUADRI

NERO 15. 5V

SOPRA

( 7 )

1

( 24 )

F

2.5 QUADRI

1 QUADRO

G I ALLO

G I ALLO 50V

FUSI BI LE RI PRI STI NABI LE 5A

0.5 QUADRI

14 PIN

100R

( 5 )

MEDIO

( 25 )

RO SSO 0V

RO SSO 12V

12

RO SSO

MONTATO

SU AP50

TP12

1.5 QUADRI

RO SSO 7V

0.5 QUADRI L= cm

NERO 150V

16 PIN

PWA11399 BP SEG NALI

TSM - 1

TSM - 2

J801

J802

J803

J805

J806

19Vac-1 JP1

19Vac-2 JP1

19Vac-1 JP4

19Vac- JP4

4. 5Vac- 1

4. 5Vac- 2

17Vac-1 JP2

17Vac-2 JP117Vac-1 JP5

17Vac-2 JP5

TSA-1

TSA-2

15. 2Vac

15. 2Vac

0

12Vac-1

12Vac-2

10Vac-1

10Vac-2

START-1

START-2

STO P-1

STO P-2

J838

J831

J835

0.25 QUADRI

( 6 )

0

I NTERRUTTO RE LUM I NO SO

( 15 )

RO SSO

VI O LA

NERO

XTF10330

1

5

6

4

7

89

1011

12

1314

15

1617

1819

2021

22

2

3

23

24

12 20 cm 0.5 QUADRI

G I ALLO 11. 5V

BI ANCO

0

9 POLI L = 55

1 QUADRO

0.5 QUADRI L= cm

ENCO DER

-

-

-

-

-

16 PIN

VI O LA 19V

0R5

1.5 QUADRI

AC

0.25 QUADRI

( 16 )

M I CRO T1000

J801

J802

RO SSO 7V

XTF10345

1

4

2

3

GROSSO

2K2

1 QUADRO

APPENDIX 2: MENU SELECTIONS FLUX DIAGRAM