CO2 Laser Scanner Therapy Device Service Manual

19
CO2 Laser Scanner Therapy Device Service Manual MODEL: YILIYA- 10600aL BEIJING HONKON TECHNOLOGIES CO., LTD.

Transcript of CO2 Laser Scanner Therapy Device Service Manual

CO2 Laser Scanner Therapy Device

Service ManualMODEL: YILIYA- 10600aL

BEIJING HONKON TECHNOLOGIES CO., LTD.

About us

Beijing HONKON Technologies Co., Ltd. is a professional medical and aesthetic device manufacturercombined with R&D, marketing and services, relating to laser, Intense Pulsed Light (IPL), Oxygen jet,radio frequency, microwave, ultrasonic, and other technical fields, including laser series, IPL series,RF series, Oxygen jet series, E-light series, slimming series, and cosmetics products, and some havegot a certificate of CE Marking and been sold to the United States, Europe, the Middle East, Africaand Southeast Asia, and other regions. Supported by users, we have gradually established a leadingposition in this industry.

With the improving R & D and service system, we provide you at any time with a full range of serviceincluding training and maintenance. Moreover, OEM processing business for merchants is provided.Nearly eight years of development, "HONKON" establish its good reputation and awareness in thedomestic and international industry. Within the next five years, we intend to rapid unconventionaldevelopment, continuous innovation and change, and strive to be a first-class medical and aestheticequipment international manufacturer.

With "Customer-centric, continuing to create value for customers, creating strong brand HONKON" asour principle, we are committed to constant innovation and development with high technology as thefoundation, quality as the force and services as assures for the future prospects to create bestproducts, supply excellent services, so as to accomplish our mission, “creating a strong Chineseinternational brand”.

Copyright Declaration

The manual is for CO2 laser scanner therapy device YILIYA-XS 10600A manufactured by ourcompany (BEIJING HONKON TECHNOLOGIES CO., LTD.). Our products have been continuouslyimproved and updated, and we reserve the right to make changes to the specifications without notice.

Before installation, please read this manual. Operate in accordance with the specification strictly. Weare not responsible for the direct, indirect, intentional, unintentional damage and risk due to improperinstallation, use and maintenance.

The declaration is published by HONKON and the final interpretation right belongs to HONKON.

Attention: Please keep this manual well for future use!

ACCESSORIES LISTPlease check whether the accessories are all available or not according to the list below:

ACCESSORIES NUMBER PHOTOS

transmissionsystem

(with box)

articulatedarm 1

connector 1

hammer 1

calibrated tube 1

handpiece (scanner) 1

cliper 5

power cord(3m) 1

foot switch 1

key 2

eye-shield for patient(with box)

1

protective gogglesfor operator

1

fuse (spare part; 250V 10A) 2

Warnings and Labels

Dangerous voltage (adhered to the rear panel of the system )

Laser symbol: To suggest a strong laser beam (adhered to the rear panel of the device )

Handpiece classification by the level of anti-shock: Type B application part (adhered tothe rear panel of the device)

Emergency stop (beside the emergency button)

device )Attention! Please refer to accompanying documents (adhered to the rear panel of the

Classification of the laser: (adhered to the rear panel of the device )

不可见激光辐射避免眼或皮肤受到直射或散射辐射的照射

4 类激光产品

INVISIBLE LASER RADIATION AVOID

EYEOR SKIN EXPOSURE TO DIRECT

OR SCATTERED RADIATION CLASS 4

LASER PRODUCT

Aperture label: (on the handpiece)

不可见激光窗口INVISIBLE LASER APERTURE

Device information: (adhered to the rear panel of the device )

PRODUCT: CO2 LASER SCANNER THERAPY DEVICEMODEL: YILIYA- 10600ILPOWER SUPPLY: 220V ~,50Hz

POWER INPUT: Max 600VA

ENERGYOUTPUT: 8.3mj-166mj

WAVE LENGTH:10600nmDATE:SERIES NO.:DERECTIVE COMPLIED:93/42/EEC

BEIJINGHONKON TECHNOLOGIES CO., LTD.

www.honkonlaser.net.cn

ADDRESS: No.3 Building, No.11 yard, Kangding Street, BDA,

Beijing, China, 100176

AUTHORIZED REPRESENTATIVE: Renovacio Estetica S.L.

ADDRESS: C / Reis Catolics, 62, 46700, Gandia, Valencia,

Spain

CONTACT: +34 – 962871411 Mr. Carles

0197

Tips for safety

Ensure good ventilation to keep stable and reliable work, please place the equipment far

enough from the wall, or other obstacles.

The power cord can not be extruded and placed under other things. Please grip the plasticpart when pulled the power cord out after turned off.

Power can not be overloaded, or it may causes electric shock or even fire.

Cut the power before cleaning, and not to use liquid detergent and chemical spray cleanersto wipe the device. Soft and clean cloth could be used, and focal lens should be wiped withMedical alcohol.

Power supply should comply with that of the device required, before operate please check.

And power cord should be pulled off if the device would not be used for a long time.

Equipment installation and operation environment should be clean, low dust, withtemperature within 10 ℃ -40 ℃, relative humidity higher than 80%, without corrosive particles,conductive dust, etc., and strong magnetic field interference should be avoided

Not to repair the equipment yourself, for open or teardown the equipment may be subject tohigh voltage and caused electric shock and personal injury. For repair, please contact with ourafter-sales service department.

The device can not be used in operation rooms in hospital.

ContentCHAPTER 1 Introduction1.1 Front View1.2 Back View1.3 Top View1.4 Operation Principle and Characteristic1.5 Work Theory1.6 Application1.7 ContraindicationCHAPTER 2 Introduction of Installation, Adjustment and Operation2.1 Unpacking and inspection2.2 Installation of the transmission system2.3 Replace the fuse2.4 Tips for installationCHAPTER 3 Interface OperationCHAPTER 4 Tips for TreatmentCHAPTER 5 Technical InformationCHAPTER 6 MaintenanceCHAPTER 7 Transport and Storage ConditionsCHAPTER 8 Troubleshooting GuideCHAPTER 9 Technical SupportCHAPTER 10 After-Sale Service

CHAPTER 1 Introduction

1.1 Front View

1. articulated arm1 2. signal cord

3. handpiece2(scanner)

3 4. hammer5 5. touch screen

6. emergency stop8 7. key

8. handrail6 9. truckle brake7

9

1.2 Back View 6

1. vent2.power socket

1 3. foot switch jack4. IC card jack(for further use)

2 5. janitor jack6. truckle

3

4

5

6

1.3 Top View

1. screw for fastness2. articulated arm base

1 3. signal line jack4.windpipe jack

2

3 4

1.4 Operation Principle and CharacteristicsModulated by the control circuit, laser light emitted from the laser built-in would be turned to ultrapulse with short duration and high power. Controlled by the signal, the two lens in the scanner wouldbe deflected in some degrees at X and Y directions respectively, then come into being differentpatterns. Through F-θ lens gathering light, spots with very small diameter would be gained (diameter0.25mm).Super-pulsed laser is characterized by the short pulse width, high-energy of single-pulse and microspot size. Instantaneous high energy density can directly gasify target tissue, without bleeding. Inaddition, deep skin collagen can be effectively regenerated with the max energy penetration depthup to 2mm.Statement: Technical information is available for servicing reference if necessary, such as the wiringdiagram.

1.5 Work TheoryThe ultra pulse CO2 laser scanner therapy device scans a highly focused laser beam of high-energyand short pulse rapidly over certain spot sizes randomly. Due to its absorption by water in soft tissue,the fractional CO2 laser incorporating the advanced technologies can immediate ablate discretecolumns of tissue without charring, deposit heat and promote regeneration throughout the underlyinglayers of the skin, eliminating heat conduction to adjacent tissue. Also, random scanning technologyensures every spot have enough time to be cooled to avoid thermal damage. This device can be usedin multiple specialties,such as Oral Surgery, Plastic Surgery, Dermatology.

1.6 Application1. Photic damage2. Scars including atrophic acne scars, surgical scars, traumatic scars, burn scars;3. Pigment diseases, such as speckle, pigmentation, actinic keratosis, sunspots;4. Wrinkles caused by sunlight or early aging wrinkles, on the lower eyelid, forehead, cheekas well as crow's-feet.

1.7 Contraindication1. Has other skin remodeling operations, such as Dermabrasion;2. Has a skin disease or inflammation of the skin3. Smoking, diabetes, physical scars;

4. Mental disorder;5. Has serious systemic disease;6. Use of retinoic acid drugs recently;7. Do not want to accept the risk or prevent solarization after the operation;

CHAPTER 2 Introduction of Installation, Adjustment and OperationCaution: This device should be operated only by professionals and trained persons.If operate the device not by the method given there would be dangerous radiation!

2.1 Unpacking and inspectionOpen the packaging box, please check to ensure that the device is not seriously damaged, circuitcables are well connected and accessories are available (see accessories list).

2.2 Installation of the transmission systemTransmission system has been calibrated by the professionals before the device leaving the factory.Unauthorized persons should not calibrate the transmission system with the calibrate tube.Articulated armInsert the articulated arm directly into the articulated arm base at the top of the main device, thentighten the screw at the side of the pipe to fix up the articulated arm, as shown in figure 1 and 2;

screw for fasten

HammerFig.1 Fig.2

Connect the two parts of the heavy hammer with the stud and screw through the hole in the top part ofthe articulated arm, as shown in figure 3 to 5;

Fig.3 Fig.4 Fig.5Caution:a.When installing the heavy hammer, please check firstly to ensure the articulated arm be fixed ina safe station to prevent damage.b. Be sure to carry the heavy hammer carefully to prevent damage to people or the device.

Scanning handpieceTightening the focal lens clockwise to scanner,as shown in figure 6 to 7, then cover the windpipe onthe stud with hole in it, as shown in figure 8 to 9.

Fig.6 Fig.7 Fig.8 Fig.9Connect the end part of the articulated arm and the scanning handpiece with the connector, as shownin figure 10 to 13.

Fig.10 Fig.11 Fig.12 Fig.13Signal cord and windpipePlug signal cord into the signal socket on the top of the device and tighten the screws on both sides.Then insert the windpipe into the air vent beside, as shown in figure 14 to 17.

Fig.14 Fig.15 Fig.16 Fig.17Foot switchWith the notch on the top, plug the foot switch cable into the foot switch socket on the back part of thedevice, then tighten it clockwise, as shown in figure 18 to 19.

JanitorFig.18 Fig.19

The connecter of the janitor is the same with that of the foot switch, and for the connection, pleaserefer to the foot switch, as shown in figure 18 to 19.Power cordPlug the two terminals of the power cord into the power socket on the back part of the instrument and

the power input socket connected with the network power, then put down the anti-off clip to fix up thepower cord, as shown in figure 20 to 22.

Fig.20 Fig.21 Fig.22Caution: Three-hole power outlet with ground wire must be used, and the voltage must match that ofthe device.Turning on the deviceMake sure the emergency stop switch off, then insert the switch key and rotate it clockwise (thedirection indicated by the arrow) by 90° to turn on the device.

Caution: Do not replace the components of the device optionally, otherwise it is as your own risk.

2.3 Replace the fuseYou can replace the fuse only with those of 250V/10A.

Fig.23 Fig.24 Fig.25First of all, pull out the fuse base with tools (the screwdriver is available), replace the fuse with newones, then push the base into the filter. As shown in figure 23 to 25.

2.4 Tips for installationa. The device should be placed stably in a drafty location with room temperature of 10 ℃ -40 ℃,relative humidity no higher than 80%, and not to be in the direct sunlight, closing to the radiator, suchas the ventilation pipe, and in somewhere there would be mechanical vibration;b. For the device is an optical equipment of high sensitivity, the installation and operationenvironment should be low-dust, and has a good air quality (non-corrosive particles, conductive dust,etc.); Otherwise, the dust would irreversibly damage the optical components inside the device.c. The device should be installed and placed carefully to prevent falling and bump, the optical devicesin which are precise and fragile.d. Make sure that all the cables be connected in place and stably;e. There should not be strong electromagnetic interference when operating the device, if necessary,with the use of filter;f. Do not to replace the components of the device optionally for the internal structure is complicated.In order to ensure the normal use, please contact with the manufacturers or agents to buy or replacecomponents needed.

g. Please check before operating the device whether the power supply to meet the specificationsrequired. ( as detailed in the technical parameters ).

CHAPTER 3 Interface OperationTurn on the device and wait for about 60 seconds to get into the user interface. As the followingpicture:

30 秒激光器预热延时计数器

Buttons:

Test: Before treatment, run the device on the black test-paper to confirm it works well.

Standby mode: to set parameters and change work model.

Press , after three seconds, the machine enter into Ready to treatment; Only pulse

energy can be adjusted; when the foot switch is pressed, laser is emitting;

Continue work mode

Ultro-pluse laser work mode

Fractional laser output work mode:

Red arm light indicated.

Increase energy, step 2mj.

Reduce energy, step 2mj.

Increase the Area、Density、Proportional and Pulse Interval by treatment.

Reduce the Area、Density、Proportional and Pulse Interval by treatment.

In fractional laser work model, could be chosed

freely by treatment.

l 1.0 x 1.0 cm2

l 2.0 x 2.0 cm2

l 3.0 x 3.0 cm2

l 6 x 6 point / cm2

l 12 x 12 point/ cm2

l 24 x 24 point / cm2

graphics only 6x6point / cm2 and 12x12piont / cm2 are optional.

In fraction laser work mode, adjust the by and , in order

to get different proper AREA and DENSITY :

Proportion6 x 6

point/cm2

12 x 12point/cm2

24 x 24point/cm2

1/1 (100%) √ √ √ √ √ √ √ √

1/2 (50%) √ √ √ √ √ √ √

1/4 (25%) √ √ √ √ √ √

1/8 (12.5%) √ √ √ √ √

l For example, area 1.0x1.0cm2,density 6x6point/ cm2,if the Proportion is 1/2, then thefinal area 0.5x0.5 cm2 with density 6x6point/ cm2

Only available in work mode and work mode,Adjustable range 1ms~250ms ,Step 1ms

and Work mode, Adjustable range 2 mj ~ 200 mj,Step 2 mj.

Work mode , Adjustable range 1 w ~ 30 w , Step 1 w.

&

Only available in wrok mode

l laser output singly ,only when press foot switch down, if it finishs oneoutput will stop automatically.

l laser output continually ,when press foot switch down until released byoperater. The laser output will be continue in 1s (seconds) interval.

The inner machine temperature indicatedl When the inner temperature up to 75 ℃, the machine will trun to Standby

automatically and off work. please trun off machine till to moral WrokTemperature ≤ 45℃.l

Wrong Operation Indicated

l Indicate the operator presses foot switch dwon when the machine in standby,please check and release it.

l Indicate the control system cann’t transfer signal to control board, please check

and the cable behind the screen if they are connect well. Or contact

our engineer to help.

l When the machine is in working situation, change parameter are not

available. Please press turn to . Then change parameter as necessary.

l Indicated the parameter cann’t be changed in current work mode.

l Indicated the parameter are not available in current work mode.

CHAPTER 4 Tips for Treatment

a. Please check before the operation to confirm the health status of the patient. Communicate withthe patient and take photographs to establish the treatment files;

b. Clean the skin. Be sure to remove the makeup firstly if the operation area be part of the face;

c.Wear protective glasses to avoid probable damage to eyes caused by strong and direct light;

d. Start to operate. First of all, select an area below ears or arms to do a short test to observe thereactions with a low energy, at the same time ask the feeling of the patient, and then adjust theenergy to an appropriate one accordingly.

e. Cool with the ice after the operation if necessary.

Note: As a result of lots of factors such as health condition, skin type, lifestyle and so on, the finaltreatment effect varies from person to person. In general, the result can be maintained for a longertime as long as the patient cooperate actively and attend a reasonable diet, prevent sun light andhave enough rest.

CHAPTER 5 Technical InformationPower Supply AC 220V 50Hz

Transmission System 7-joint articulated armWave Lengths 10600nmMode Structure TEM00Spot Diameters 150-500umSpot Pattern circle / triangle / square / rectangle / rhombus /lineSpot Number(Square)

6×6、12×12、24×24

Spot Size 1×1 / 2×2 / 3×3Aiming Beam Wavelengths 635 nm

Aiming Beam Power ≤5mW

Output Mode Ultra pulseScanning Mode Random scanningPulse Energy 10mj~200mj step 2mjPulse Interval 1ms~100ms step 1msWorking Mode Continuous operation with intermittent loading

Cooling Air&Water

CHAPTER 6 Maintenancea. After turning on the device, it will take one minute to come into the interface, then press the simmerkey.b. Put the handpiece on the bracket after every operation avoiding falling off to be damaged.c. Do not use flammable things, such as alcohol, to clean and disinfect the lens of the handpiece andthe operation area.d. Do not clean the surface of the machine using gasoline, alcohol, thinner materials or otherchemical agents, as it will pill off the coat of the device.e. Please note that not to make it dusty when cleaning the device or the operation room, avoidingmaking the optical system fail.f. Please remove the key switch and turn off the emergency stop when the device would not be usedfor a long time.g. Do not take down the device without permission, for the internal structure is complex. When it goeswrong, please contact with the service department of the company.h. In order to ensure the normal use of the equipment, please contact with the manufacturers oragents when accessories are needed to be replaced.Caution:a. Customers are only permitted to do some daily maintenance or deal with some simple failuresunder the direction of professionals of our company, and other servicing should be done by technicalstaffs authorized by our Company.b. Make sure to turn off the device and pull out the power line before doing maintenance, or it ishazardous to both bodies and device.

CHAPTER 7 Transport and Storage Conditions1. Transport and storage environment requires: temperature within -40℃ to 55℃, relative humidity nohigher than 80% and good air quality;2. Before transportation, the scanner, hammer, articulated arm should be removed, avoiding damage.2. For it is precise equipment, strong vibration and shocks should be avoided and no upside downwhen remove and transport to prevent damage to the internal system and electronic devices;3. Do not remove the equipment in dusty environment when there is no dust-proof packaging;4. Towing, suspension and clip are avoided in transport, and holding the bottom is needed.

CHAPTER 8 Troubleshooting Guide

Failure Failure Analysis Solution

No power

Power cord failed Replace a new oneEmergency button beenpressed

Turn to up it

Key switch failed Replace a new oneFuse burned Replace new onesTouch buttons failed Turn off the device, touch the

screen by hand, then get intothe orientation interface after asound. Point the sign "+" on thetop left corner of the screen andthen the bottom right corner oneto complete the reorientation.

Have the feeling ofelectric shock

Dry weather Not to wear chemical fiberclothing

Footswitch nofunction

Cables were not well-connected Reconnect the cablesFootswitch failed Contact with the company

CHAPTER 9 Technical Support

Promise"The customer is always right and continuing to be provided with first-class service" hasbeen the principle of the after-sales service of our company, and an independenttechnical service department has been set up, having a professional team for customersto resolve technical and clinical problems at any time. Experts from some famoushospitals and skin-care specialists have been specially retained to solve technicalproblems and provide more scientific technology more valuable at any time forcustomers. Further more, large scale national meetings are regularly carried out todeliver advanced cosmetic and medical technology, information of world-class productsto customers.

How to servicea. After purchase, we will send technical instructors to provide training service for free;b. Regular following service by telephone is available at any time to help you solve thetechnical problems encountered;c. In addition to telephone contact, you can also visit our website to download relevanttechnical information at any time.

CHAPTER 10 After-Sale Service

We ensure that our products work well normally. We supply warranty service for free within one yearfrom the date of purchase and life-long maintenance service is also available (except specialagreement)

Free warrantya. If failure is identified by the company as a result of materials or process defects on thepremise of correct use and maintenance within one year, free service is available.b. Free service does not include the scanner, foot switch and articulated arm, and the mail costs arealso not included.c. Users applying for free service are required to hold a "warranty card" or a valid shoppingcertification;d. Free repair service (damages caused by human factors are except) within threemonths from the date of the repair service and life-long maintenance service are still supplied forthe replacement of parts;e. About special provisions, carry out according to the contract strictly.

Beyond free warrantya. In the free warranty period, damages as a result of improper installation, usage, maintenance orcustody, such as severe impact, vibration, drop and wrong operation;b. Damages caused by improper repair by unauthorized departments by the company or consumerself-removal.c. No warranty card or valid certification to prove;d. The warranty card is incomplete or altered;e. Damages caused by irresistible factors, such as fires, earthquakes and other natural disasters;f. Beyond the warranty period;