A LOW-POWER RECEIVER FOR SIMULTANEOUS ...

87
A LOW-POWER RECEIVER FOR SIMULTANEOUS ELECTROCARDIOGRAM AND RESPIRATORY RATE DETECTION by JIFU LIANG Submitted in partial fulfillment of the requirements For the degree of Master of Science Department of Electrical Engineering and Computer Science CASE WESTERN RESERVE UNIVERSITY August, 2016

Transcript of A LOW-POWER RECEIVER FOR SIMULTANEOUS ...

A LOW-POWER RECEIVER FOR SIMULTANEOUS

ELECTROCARDIOGRAM AND RESPIRATORY RATE DETECTION

by

JIFU LIANG

Submitted in partial fulfillment of the requirements

For the degree of Master of Science

Department of Electrical Engineering and Computer Science

CASE WESTERN RESERVE UNIVERSITY

August, 2016

A Low-Power Receiver for Simultaneous Electrocardiogram and

Respiratory Rate Detection

Case Western Reserve University

Case School of Graduate Studies

We hereby approve the thesis1 of

JIFU LIANG

for the degree of

Master of Science

Dr. Soumyajit Mandal

Committee Chair, Adviser July 1st, 2016Department of Electrical Engineering and Computer Science

Dr. Pedram Mohseni

Committee Member July 1st, 2016Department of Electrical Engineering and Computer Science

Dr. Dominique Durand

Committee Member July 1st, 2016Department of Biomedical Engineering

1We certify that written approval has been obtained for any proprietary material contained therein.

I would like to dedicate this thesis to my advisor, Dr. Soumyajit Mandal,for his endless guidance of my research.

I would also dedicate this thesis to my parents, Yanjun Liang and Jinju Lu,for their support of my study in America.

I would also dedicate this thesis to my girlfriend, Shuyue Li,for her encouragement of my study at CWRU.

Table of Contents

List of Tables vi

List of Figures vii

Acknowledgements xiii

Acknowledgements xiii

Abstract xiv

Abstract xiv

Chapter 1. Introduction 1

Motivation 1

Research Goals 2

Literature Review 3

Structure of the Thesis 13

Chapter 2. First Chip Design 14

Impedance Pneumography 14

Block Diagram 16

Clock Generator Block 18

Controlled Injected Current Block and Mixer 20

Preamplifier Block 21

Wide-linear Range OTA and Filters 24

Second Gain Stage for Amplifying RR signal 29

Chapter 3. Measurement Result of First Chip 32

Programmable Injected Current Block and Mixer 34

iv

ECG Amplifier 35

RR Measurement 40

Chapter 4. Second Chip Design 43

Fully Differential Wide-linear-range OTA 43

Fully Differential Filters 45

Fully Differential Preamplifier Block 47

Digital Block 52

Second Gain Stage for Amplifying RR signal 57

Layout 60

Chapter 5. Conclusions 63

Appendix A. PCB for testing the AFE 66

Appendix. Complete References 68

v

List of Tables

3.1 Comparison table with the state of the art 34

vi

List of Figures

1.1 Conceptual view of a wireless biopotential measurement system

integrated into a chest band. The three patch electrodes are denoted

RA, LA, and REF, respectively. 3

1.2 One of the most frequently used topology of bioelectric amplifier1. 6

1.3 (a) Typology of MOS pseudo resistor. (b) Basic structure of switched

capacitor. 7

1.4 Schematic of ECG amplifier in reference2. 8

1.5 Schematic of ECG amplifier presented by Wen et al.3. 10

1.6 Schematic of ECG amplifier presented by Burke and Gleeson4. 11

2.1 Simplified structure of the respiration circuit when RA and LA

electrodes are used. 15

2.2 Detected respiratory rate signal after filtering with a fourth-order

LPF with cutoff frequency of 2 Hz. A commercial AFE (ADS1298R)

designed by Texas Instruments is used here to measure RR signal

when the volunteer took deep breath. 16

2.3 Simplified block diagram of the integrated front-end receiver. 17

2.4 Typical spectra of the ECG, modulated RR, and baseband RR signals

before the HPF, before the quadrature modulator, and after the

modulator. 19

2.5 Clock generator circuit used to generate 0, 90, 180, and 270 clocks. 19

vii

2.6 Simulation result of the clock generator block. clk1 and clk2 have 90

phase shift. 20

2.7 Programmable current mirrors and double balanced mixer used for

generating I AC . The amplitude of I AC can be programmed over a 1:16

range (4 bits) by the switches φ1-φ4. 21

2.8 Schematic of the low-noise ECG amplifier. 22

2.9 Schematic of the OTA used four times in the preamplifier. 23

2.10 Simulation results of (a) the AC response of the preamplifier. (b) the

output of the preamplifier when an input of 2 mV 100 Hz sinusoid

wave is used. (c) the output of the ECG path when an input of 2 mV

100 Hz sinusoid wave is used. 25

2.11 (a) First-order Gm −C LPF topology. (b) First-order Gm −C HPF

topology. 25

2.12 Basic single-stage five transistor OTA with NMOS input transistors. 26

2.13 The wide-linear-range OTA used several times in the AFE. 27

2.14 Schematic of the BPF and second gain stage used to process each

demodulated RR signal component. The capacitors C1 and C2

are 400 pF off-chip capacitors, while C3 and C4 are 2 pF on-chip

capacitors. 29

2.15 Simulation results of (a) the AC response of the BPF after demodulator.

(b) the I component of RR at the output of mixer if a 1 µV RR signal is

applied at the input of the preamplifier. (c) the final output of RR path

if a 1 µV RR signal is applied at the input of the preamplifier. 31

viii

3.1 Die photograph of the fabricated AFE, which has an active area of

1050 µm × 600 µm. Major blocks are labeled. 32

3.2 The PCB used to test the first chip. 33

3.3 The measured transistor I-V curve used to set the 200 nA off-chip bias

current. This curve was measured by using a Keithley source meter. 34

3.4 Circuit used to test the programmable injected current block and the

mixer. A fixed resistor (100 kΩ) was placed between LA and RA. 35

3.5 (a) Square voltage waveforms across the RA and LA with different

injected currents when a 100 kΩ resistor was used between RA and

LA. (b) Measured positive and negative currents as a function of the

digital code. 36

3.6 (a) Differential frequency response of both the ECG amplifier and the

following LPF. (b) Common-mode frequency response of both the

ECG amplifier and the following LPF without utilizing driven-ground

circuit. 36

3.7 (a) Output noise of the both the ECG amplifier and the following LPF.

(b) Input noise of the both the ECG amplifier and the following LPF. 37

3.8 (a) Output of the ECG path when high-frequency clock is disabled.

The input signal is a 2 mV, 10 Hz sinusoid wave. (b) Output of the ECG

path to the sum of a 1.5 mV, 10 Hz sinusoid (simulating ECG) and a

1.4 mV, 100 Hz square wave (simulating RR carrier). 38

3.9 Detected ECG signal after baseline removal and filtering with a

fourth-order LPF ( fc = 50 Hz). 39

ix

3.10 The DC voltage at the output of the ECG path shifted from 0.3 V to

2.4 V in around 900 seconds. 40

3.11 (a) Body model used to generate ECG-like and RR-like signals. (b)

Simulation result of the modulated RR-like signal when 100 nA 1 kHz

AC currents are fed to LA and RA. (c) The FFT result of the transient

signal generated by the body model. 41

3.12 Input triangle wave used to modulate the impedance between RA and

LA at 0.5 Hz, and measured I and Q RR outputs after filtering with a

fourth-order Butterworth LPf with cutoff frequency of 5 Hz. 42

4.1 (a) Fully differential version of wide-linear-range OTA. (b) Differential-

difference OTA used in CMFB. (c) Fully differential OTA stage. 45

4.2 (a) A single stage of fully differential OTA in the preamplifier. (b) Bode

plot of the fully differential OTA stage. 46

4.3 (a) Fully differential LPF. (b) Fully differential HPF. 47

4.4 The AC simulation results of (a) HPF with cutoff frequency of 1 kHz to

remove ECG signal in RR path. (b) LPF with cutoff frequency of 5 Hz

to remove high-frequency RR carrier signal in RR path. 3 nF off-chip

capacitor is used. 47

4.5 Circuit of fully differential preamplifier. 49

4.6 Modified switched capacitor topology. 51

4.7 Equivalent circuits of the modified switched capacitor topology in

two phases. (a) Equivalent circuit when clk1 is on and clk2 is off. (b)

Equivalent circuit when clk1 is off and clk2 is on. 51

x

4.8 Transient simulation result at the output of the preamplifier when

a 2 mV 100 Hz sinusoid wave simulating ECG signal and a 10 kHz

square wave simulating RR carrier signal are used at the input of the

circuit. 52

4.9 Digital blocks in the AFE. 53

4.10 Frequency response of the first-order LPF by using switched capacitor. 54

4.11 Programmable counter. 55

4.12 Two-phase non-overlapping clocks generator block. 55

4.13 (a) Programmable bus used to set digital inputs. (b) Timing diagram

of SPI. 56

4.14 Simulation results of the circuits generating clocks used in the mixer

for modulation and demodulation. (a) The 38 kHz input square wave

simulating the off-chip clock. (b) The first output clock with one

fourth of the frequency of the input wave. (c) The second output clock

with one fourth of the frequency of the input wave and 90 phase

shift. 57

4.15 Simulation results of the circuits generating clocks used in the

switched capacitor. (a) The first output clock. (b) The second output

clock. (c) 2 ns delay between the two non-overlapping output clocks. 58

4.16 Second gain stage for amplifying RR signal. 58

4.17 Open-loop Bode plot of the op-amp stage. 59

4.18 Closed-loop Bode plot of the op-amp stage. 60

xi

4.19 Transient simulation result at the output of I branch in RR path

(I+− I−). The frequency of RR signal is chosen as 5 Hz to increase the

simulation speed. 60

4.20 Layout of the second chip. 62

A.1 Schematic of the PCB board. 66

A.2 Layout of the PCB board. 67

xii

Acknowledgements

0.1 Acknowledgements

I would especially like to thank my mentor, research advisor, Dr. Soumyajit Mandal, for

his endless support of my research during my M.S. study. I would never accomplish

this project without his guidance. Not only has he taught me much knowledge about IC

design, but also he has assisted me to adapt to the study at CWRU.

Thanks to my parents, Yanjun Liang and Jinju Lu, for their continuously support of

my study in America. Thank you for providing me an opportunity to study abroad. You

will always be my beloved parents.

To my girlfriend, Shuyue Li, who keeps giving me encouragement during my study

at CWRU. Although we are studying at different universities, but your support makes me

a better student and a better man.

I would also dedicate to my M.S. committee members, Dr. Pedram Mohseni and Dr.

Dominique Durand. Thank you for your advice and help on my thesis.

I am very grateful to Yingying Wang, for your enduring help on my chip design.

Thank you for being patient and answering all my questions.

I would also thank many of the students and staffs at CWRU. Thanks to Dr. Steve

Majerus for your instruction in the lab. Thanks to my partner in this project, Shixiong

Li, who helped me design my first chip. Thanks to another partner in this project, Gre-

gory Hessler, who helped me to study the method to transmit data by using RFduino

board. Thanks to my friend, Ali Nikoofard, who helped me test my first chip. Thanks

to my friends, Cheng Chen and Xinyao Tang, for all the happy and tough time we spent

together.

Last but not least, thanks to all the people who helped me finish this thesis.

xiii

Abstract

A Low-Power Receiver for Simultaneous Electrocardiogram and

Respiratory Rate Detection

Abstract

by

JIFU LIANG

0.2 Abstract

Electrocardiogram (ECG) and respiratory rate (RR) signals are useful for doctors to di-

agnose heart and respiratory diseases. This work is to design a low-power analog front-

end (AFE) for simultaneously detecting both ECG and RR signals. In this projectïijN two

chips have been designed on the OnSemi 0.5 µm CMOS process, and the first one has

been tested. For the first chip, both simulation and measurement results prove the func-

tionality of it. The circuit draws 6.2 µA from a 3 V power supply. The gain for the ECG

and RR signals are 40 dB and 70 dB, respectively. To modify the defects found during

the measurement of the first chip, a fully differential AFE with more gain in RR path is

designed. The circuit draws 14 µA from a 3 V power supply, and the simulation results

verify the functionality. The AFE can be further used in a wearable biopotential system

for measuring both ECG and RR signals.

xiv

1

1 Introduction

1.1 Motivation

Heart disease is the leading cause of death in the United States. According to the Amer-

ican Heart Association (AHA), an estimated 85.6 million American adults (more than

one-third of the total population) have one or more types of cardiovascular diseases

(CVDs). Heart disease happens on someone in the United States every 43 seconds,

killing more than 375,000 people a year. Based on the fact that many people are suf-

fering from CVDs, a large amount of money is spent on their healthcare. In 2011, it cost

207.3 billion for direct heart disease, 48.6 billion for hypertension, 33 billion for stroke

and 27.7 for other CVDs5. Consequently, monitoring heart activity has gained consid-

erable attention. It has been known that Electrocardiogram (ECG) shows the heart elec-

trical activity, and it contains much information related to heart rate and rhythm which

can be helpful for doctors to diagnose heart disease. So many different kinds of ECG

monitors have been produced and utilized6.

Respiratory rate (RR) is defined as the number of breaths a person has in one minute.

It is very useful for doctors to diagnose pneumonia and many other respiratory dis-

eases7. High respiratory rate has also been observed on a large proportion of cardiac

arrest patients8,9. Based on the study from Health Grades in 2011, 20% of postoperative

Introduction 2

respiratory failures resulted in death10. However, monitoring RR has not gained much

attention compared to other vital signs11. RR is also not recorded in many hospitals,

even when respiratory condition is the primary problem for patients12,13. One main

reason is that there are not enough reliable RR monitoring systems14.

Wearable health-monitoring systems have drawn much interests in recent years15.

From the study of Trans-European Network Home-Care Management Systems (TEN-

HMS), home telemonitoring systems get better feedback from patients. For home tele-

monitoring patients, it costs 26% fewer days in hospital, leading to a 10% cost savings.

Survival rate is also substantially better for home telemonitoring patients16. Wearable

systems not only have those advantages, but also fulfill the need for monitoring patients

over a long period of time. A distinct example is ECG monitoring. It is desirable that

ECG can be monitored over several weeks or months, and wearable monitoring systems

are more suitable compared to the normal ambulatory systems in this case17. But chal-

lenges for wearable monitoring systems exist, and one main problem is the considera-

tion of power consumption.

1.2 Research Goals

In this project, the goal is to design an integrated low-power analog front-end (AFE) for

simultaneously detecting ECG and RR signals. We assume that there are three electrodes

connecting to the body, including right arm (RA), left arm (LA), and reference electrode

(REF). The impedance pneumography (IP) method18,19, which will be described in the

second chapter in detail, is utilized to detect RR signal. Both ECG and RR are measured

between RA and LA chest electrodes. A driven-ground circuit using the REF electrode is

also utilized to boost common-mode rejection ratio (CMRR).

Introduction 3

This chip could be ultimately used in a wearable biopotential measurement sys-

tem, which can continuously wirelessly monitor heart activity and respiratory condi-

tion. Fig. 1.1 shows the desired chest band, including all the three electrodes, AFE, ADC,

antennas, and batteries. All the raw data is processed by an AFE and then digitized by

an ADC. After that, the digitized signals are buffered and then wirelessly transmitted to

a Simband or a mobile phone. The Simband can be programmed so that users can have

good interface with it, and users can also directly read the data collected from the chest

band. Another advantage of using Simband is that it reduces power consumption by

allowing the chest band to transmit data to a local unit over short distance (i.e., by cre-

ating a so-called body area network20). In this case, it is easy for people to wear this

user-friendly monitoring system for long time.

Figure 1.1. Conceptual view of a wireless biopotential measurement sys-tem integrated into a chest band. The three patch electrodes are denotedRA, LA, and REF, respectively.

1.3 Literature Review

Wet electrodes, dry electrodes and non-contact electrodes are widely used to collect

electro-physiology signals in biopotential systems. The difference between these three

electrodes is discussed in this section. Also, monitoring ECG has gained much attention,

Introduction 4

and there have been many different kinds of ECG monitors in the market6. Several re-

searchers have integrated AFEs with ADCs and digital processors (to extract heart rate,

etc.) into complete systems-on-chips (SoCs)2,21,22, but in this section the literature re-

view is limited to analog front-ends since the focus of this thesis is on the AFE. Some of

the ECG measuring circuits have the active ground circuit but some don’t. The differ-

ence between them is discussed in the following content. In addition, the amplitudes

of physiological signals are in the order of tens of µV to tens of mV and the frequency

varies from DC to a few kHz23. Typically, the frequency of ECG signal is from 1 Hz to

100 Hz and the amplitude of it varies from 0.5 mV to 4 mV24, while the RR signal has a

typical bandwidth of 0.5 Hz. Since the frequencies of the ECG and RR signals are rather

low, filters with very low corner frequency are always needed. It is easy to implement

filters with low cut-off frequencies by using off-chip capacitors and resistors. But either

switched capacitor resistors25 or MOS pseudo resistors1,26 are required for on-chip im-

plementations.

1.3.1 Electrodes

Electrodes are an important part of physiology monitoring systems. And the three kinds

of most common electrodes are wet electrodes, dry electrodes and non-contact elec-

trodes. Wet electrodes like Ag/AgCl electrodes are widely used in bioelectric applica-

tions currently27. The advantages of Ag/AgCl electrodes is obvious that they are simple,

disposable and easy to be used28. And the most important advantage of Ag/AgCl elec-

trodes is the low impedance29. But the electrolytic gel has to be applied, which may

cause allergic reactions28. And they can not be used for long time because of the dehy-

dration30. Also, short circuiting may happen since the gel can smear31. But for the dry

electrodes, these problems do not exist since the gel is not applied. However, since there

Introduction 5

is no adhesion between the dry electrodes and the body, electrodes may shift when the

body moves30. Compared to wet electrodes and dry electrodes, non-contact electrodes

do not have an ohmic connection to the skin. No preparation is needed if non-contact

electrodes are utilized. The non-contact electrodes are insensitive to skin conditions,

so they are very suitable for wearable monitoring systems since they can be embedded

within a comfortable layers of fabric32. But the input impedance of the AFE should be

high so that it can extract the signal from non-contact electrodes29. And any motion of

the non-contact electrodes with respect to the skin may generate artifact.

1.3.2 Amplifiers Using Two Electrodes

It is very appealing that bioelectric amplifiers only uses two electrodes because only

minimal number of contacts are utilized to collect signals. It also saves power since the

active-ground circuit is not utilized. However, without driven-ground circuit, the elec-

tric potential of the human body can not be controlled. It is necessary to eliminate DC

component and very low-frequency signals prior to amplifying biopotentials because of

electrode offset voltages, motion artifacts, and so on. Therefore, AC-coupled amplifiers

are often used to remove DC component and very low-frequency signals. There is one

basic topology as shown in Fig. 1.2 which is commonly used in many biomedical ampli-

fiers with only two electrodes1. The mid-band gain is set by the ratio of capacitors in the

feedback loop. The bandwidth is approximately gm/(AmCL), where gm is the transcon-

ductance of the OTA and Am is the mid-band gain. The low-frequency cutoff is set by

the effective resistance of the pseudo resistor and the capacitor in parallel with it. How-

ever, some circuits use pseudo resistor to work as large resistors1,33–35, while some other

circuits use switched capacitor resistors to behave as large resistors36.

Introduction 6

C1

C1C2

C2

CLVo

V1

V2

Figure 1.2. One of the most frequently used topology of bioelectric amplifier1.

For the MOS pseudo resistor in Fig. 1.2, it behaves as a resistor-like device, which has

a monotonic I-V relationship1,26. To simply understand the device, it contains a pair of

diodes in parallel with different polarity. Therefore, the current increases exponentially

with voltage across them. With negative VGS , the device acts like a diode-connected

PMOS device. With positive VGS , the device acts like a diode-connected bipolar device.

High resistance exists when the voltage across the device is small, so two pseudo resis-

tors are commonly used in series as shown in Fig. 1.2 so that the distortion for large

signals can be reduced.

The switched capacitor is a discrete-time system that can also be used to replace the

large resistor. The basic structure of switched capacitor resistor is shown in Fig. 1.3(b).

Introduction 7

ϕ1 and ϕ2 are nonoverlapping two-phase clocks with the period of T . The average cur-

rent i1 is

i1 = 1

T

∫ T

0i1(t )d t = 1

T

∫ T2

0i1(t )d t (1.1)

i1(t ) = d q1(t )

d t(1.2)

From the equation (1.1) and (1.2), we can get

i1 = 1

T

∫ T2

0

d q1(t )

d td t = C ·VC ( T

2 )−C ·VC (0)

T≈ C ·V1 −C ·V2

T(1.3)

From the continuous time perspective, we also know

i1 = V1 −V2

R(1.4)

where R is the effective resistance of the circuit. By comparing the equation (1.3) and

(1.4), we can get the following conclusion that

R = T

C(1.5)

The structure of switched capacitor can be modified and different equivalent resis-

tance can be obtained, which may be different from the equation (1.5). But all of them

share the same idea that a charge q is transferred from the input to the output in each

clock cycle.

P+ P+ N+

N-well

P-sub

ɸ2ɸ1

CV1

+

V2

+

Vc

+

i1

Figure 1.3. (a) Typology of MOS pseudo resistor. (b) Basic structure ofswitched capacitor.

Introduction 8

Liang et al. present a system of portable ECG monitoring system based on Bluetooth

connectivity to a mobile phone2. In his system, ECG signal is collected by the dry skin

electrodes, and then amplified, filtered, and analog-digital converted. After that, the

digital data is sent to a mobile phone through Bluetooth. In his core monitoring circuit,

a preamplifier and a band-pass filter are designed in his AFE. A commercial microcon-

troller with a built-in ADC is also included. The schematic of preamplifier is shown in

Fig. 1.4.

Vip

Vin

Vt

Vt

C1=20pf

C1=20pf

C2=0.1pf

C2=0.1pf

C3=12pf

C3=1

2p

f

R1=100kΩ

R1=100kΩ

R2=1MΩ

R2=1

Vo

Vcm

Figure 1.4. Schematic of ECG amplifier in reference2.

In this preamplifier, two stages of amplification are utilized. The gain of each stage

is set by the ratio of two capacitors or two resistors in negative feedback. Besides, the

PMOS transistors in the feedback loop work as pseudo resistors. The pseudo resistor

and the capacitor in parallel with it form the high-pass pole. The PMOS transistors are

biased in subthreshold region, and the equivalent resistance is set by Vt . As the author

states, the equivalent resistance changes from 109 Ω to 1014 Ω if Vt changes from 0.8 V to

1.6 V, resulting the pole changing from around 10 mHz to several hundred Hz. Measured

Introduction 9

result shows that the passband is from around 0.01 Hz to 100 Hz, and the input referred

noise is around 2.3 µVr ms . But in this preamplifier, there is a disadvantage that several

resistors are used in the amplifier which produce noise. The thermal noise in a resistor

is described by the following equation.

v2n = 4kT R∆ f (1.6)

where k is Boltzmann’s constant in joules per Kelvin, T is the temperature in Kelvin, and

R is the resistor value in Ohms.

For the amplifiers without driven-ground circuit, there are several advantages. For

example, the circuit takes less area and it may have less power consumption. But the

disadvantage is obvious that the AC common-mode voltage can not be reduced. Al-

though AC coupled amplifier can be used to remove the DC component of the common

mode signal, the power-line interference still needs to be removed. However, for wear-

able monitoring systems, the AC interference is a smaller problem since the short leads

are utilized.

1.3.3 Amplifiers Using Reference Electrodes

Now the amplifiers with driven-ground circuit is discussed. The goal of the driven-

ground circuit is to reduce the common-mode signal by using negative feedback37. Three

electrodes are widely used in the amplifiers with active ground circuit, including two

electrodes measuring electric signals and one electrode working as reference electrode.

The reduction of the common-mode voltage is related to the gain of the driven-ground

circuit. Thus, a relatively high gain is desired in the driven-ground circuit. However, a

high gain feedback loop may lead instability. Also in the driven-ground circuit, buffers

may be needed to drive the body impedance.

Introduction 10

Wen et al. provide another full custom AFE for long-time ECG monitoring3. The

circuit contains the instrumentation amplifier, filters, second amplifying stage, driven-

ground circuit, power management circuit, and leadoff monitoring circuit. The schematic

of the ECG amplifier is shown in Fig. 1.5. The bandwidth varies from 0.5 Hz to 100 Hz

with mid-band gain of 51 dB. The CMRR is 75 dB and input referred noise is 12 µV.

Buffer

Buffer

Buffer

LA

RA

RLOut

R1 R2 R3

R1 R2

R0

R0

Vcm

R3

R4Rr

RL

Cr

Figure 1.5. Schematic of ECG amplifier presented by Wen et al.3.

In this preamplifier, the inputs of two buffers are connected to both RA and LA elec-

trodes. By doing this, it can increase the equivalent input impedance of the amplifier.

For each stage of the instrumentation amplifier, the gain is set by the ratio of resistors

in the feedback loop. Also, the driven-leg circuit is utilized to increase the CMRR. The

common mode voltage Vcm is sensed by the two resistors R0. Sensed Vcm is buffered, in-

verted, amplified, and then fed back to the body. By increasing the gain of the amplifier

in the drive-leg circuit, we can increase the CMRR.

Another ECG amplifier design incorporating driven-ground circuit is presented by

Burke and Gleeson4. Since this amplifier is constructed on a matrix board, there are no

Introduction 11

constraints of resistors and capacitors with large values. The amplifier draws 9 µA from

a 3.3 V power supply. The bandwidth of the amplifier is from 0.05 Hz to 1.9 kHz with the

gain of 43 dB. The CMRR is 55 dB without active ground circuit, and it increases to 88 dB

if driven-ground circuit is used.

VCC

VCC

R1

R1R2

R2

R3

R4

R4

R5R5

R6

R6

R7

R7

R8

R8

R9

R10

R10 R11

R11

R12R13

R14

R15

C1

C1

C2

C2

C3 C4

C4

C5

C6

C7

V1

V2

Vref

Vo

Figure 1.6. Schematic of ECG amplifier presented by Burke and Gleeson4.

The amplifier has three stages. The first stage is AC coupled by using C1, causing

attenuation of DC offset and low-frequency input signals. R1 provides patient protection

by limiting the current coming from the electrodes, and R6 with R3 set up the DC bias

voltage for the op-amps. R2 defines the input impedance of the amplifier to meet CMRR

requirements in the presence of inevitable electrode impedance mismatch. The second

stage is also a differential stage with DC coupled at the input. R9 and C3 set the DC

gain to unity. Since the input capacitance of the op-amp generates a zero, C2 is used

Introduction 12

here to define the zero more reliably. The third stage is a simple DC coupled stage with

resistors in the negative feedback. Besides all the three stages, the reference circuit is

used to increase the CMRR. R7 detects the common-mode voltage. Then, it is inverted,

amplified, and sent to the reference electrode. More information of this amplifier can

be found in reference4.

By using driven-ground circuit, the common-mode voltage including the DC value

and the powerline interference can be greatly reduced. This explains why in this project

the driven-leg circuit is used in AFE.

1.3.4 Summary of Bioelectric Amplifiers

From the overview of the research field, to design a good preamplifier, several conditions

should be satisfied.

The first one is noise. A high-quality amplifier should have very low equivalent input

noise since the amplitude of ECG signal and other bioelectric signals are rather small.

For example, the amplitude of ECG signal varies from 0.5 mV to 4 mV24. To reduce the

noise, several methods like increasing the bias current for the amplifier can be utilized.

The second constraint is common-mode rejection ratio (CMRR). The CMRR should

be good enough to decrease the power-line interference so that the power-line inter-

ference is not able to have effect on the measurement of ECG signal. For a body-worn

device, 20 dB CMRR is needed since there is not a big common-mode voltage38. But for

a device not entirely worn on the body, at least 80 dB CMRR is needed.

The third constraint is low high-pass filter corner frequency to allow low-frequency

signal components (as low as 0.1 Hz for ECG) to be amplified. It is easy to implement

off-chip large capacitors or resistors to get a low cutoff frequency. However, switched

Introduction 13

capacitor or MOS pseudo resistor can also be integrated in the chip if the designer wants

to reduce the size of the monitoring system.

The fourth condition is power consumption. Power consumption should be as low

as possible so that monitoring systems can work for a longer period of time without

changing batteries. This is also good when a bioelectric signal needs to be monitored

for long periods of time to detect transient and infrequent symptoms (such as missing

or irregular heartbeats) that can be used by doctors for diagnosis. Another advantage

of systems with low power consumption is that the weight of the devices is light since

smaller batteries can be used for a given operating lifetime.

1.4 Structure of the Thesis

In this project, two AFE chips have been designed. This thesis is divided into the follow-

ing parts. In Chapter 2, the design of the first chip is introduced in detail. In Chapter

3, the measurement result of the first chip is presented. In Chapter 4, the design of the

second chip is introduced in detail. And Chapter 5 contains the conclusion.

14

2 First Chip Design

In this chapter, the first AFE chip design is presented in detail. ECG signal is collected

directly by two conventional contact electrodes RA and LA. To detect RR, impedance

pneumography (IP) method, which will be described later, is used so that the low-frequency

RR signal is modulated on a high-frequency carrier signal. The low-noise preamplifier

in the AFE amplifies both the ECG and modulated RR signals. After the preamplifier,

LPFs are used to remove modulated RR signal in ECG path. In RR path, HPFs are used to

remove ECG. Then, the mixer is utilized to extract the RR signal, and the following BPF

is used to remove high-frequency carrier and low-frequency noise. Since the amplitude

of RR is small, there is another gain stage for further amplification of the RR signal.

2.1 Impedance Pneumography

Impedance pneumography method is widely used to measure respiratory rate. Fig. 2.1

shows the simplified structure of the respiration circuit when RA and LA electrodes are

used. During respiration, the thorax can be treated as two impedance components: a

constant baseline impedance (Zbod y+2Zel ectr ode ) and a varying impedance (∆Z )18,19,39.

For the relatively constant term, the value of Zbod y is around 500 Ω based on the study

from Grenvik et al18, and Zel ectr ode is related to the kind of electrodes used (around

First Chip Design 15

2 kΩ impedance for a standard patch electrode). For the varying term caused by res-

piration, there are two reasons causing the change of the electrical resistance. Firstly,

the chest expands (contracts) during respiration, causing the length of the conductance

paths increasing (decreasing), which increases (decreases) the impedance of the thorax.

Secondly, the gas volume of the chest increases (decreases), causing the conductivity

decreasing (increasing) because the conductivity of gas is not as good as the one of tis-

sue, which increases (decreases) the impedance of the thorax. Taken together, ∆Z in-

creases when the chest expands and ∆Z decreases while the chest contracts. Typically,

∆Z varies from 0.1Ω to 1Ω19.

Zbody ΔZ

ZElectrode

Body

IAC

ZElectrode

Figure 2.1. Simplified structure of the respiration circuit when RA and LAelectrodes are used.

If a high-frequency AC current is injected into a body through the two electrodes,

a varying voltage is then generated across the electrodes. The AC current behaves as

the carrier signal, and it is amplitude-modulated by the low-frequency respiratory sig-

nal. In the following circuit, the low-frequency RR signal is extracted and the carrier

signal is filtered. Fig. 2.2 shows the detected respiratory rate signal after filtering with a

fourth-order LPF with cutoff frequency of 2 Hz. A commercial AFE (ADS1298R) designed

by Texas Instruments, which utilizes impedance pneumography method to measure RR

First Chip Design 16

signal, is used here when the volunteer took deep breath. From Fig. 2.2, it can be es-

timated that the frequency of detected RR signal is around 0.25 Hz. In addition, since

the gain of this AFE is set to 6 and the injected current is around 30 µA, the estimated

varying impedance ∆Z in thorax is around 1.2 Ω. In this design, the impedance pneu-

mography method mentioned above is used to detect RR signal. Another method to

measure RR signal is to extract it directly from ECG signal40. Studies have shown that

ECG signal is modulated because of the respiration. So ECG-derived respiratory activity

measurement is a method to derive RR signal from modulated ECG by using some signal

processing algorithms.

Time (s)4 6 8 10 12 14 16 18 20

Vo

ltag

e (m

V)

0

0.1

0.2

0.3

0.4

0.5

0.6

Figure 2.2. Detected respiratory rate signal after filtering with a fourth-order LPF with cutoff frequency of 2 Hz. A commercial AFE (ADS1298R)designed by Texas Instruments is used here to measure RR signal whenthe volunteer took deep breath.

2.2 Block Diagram

A simplified block diagram of the integrated AFE is shown in Fig. 2.3. The amplitude

of ECG signal varies from 0.5 mV to 4 mV24. For RR, ∆Z typically varies from 0.1 Ω

to 1 Ω19. The AHA recommends that the safe limit current is 10 µA41. In this project,

the maximum injected current is 1.6 µA, which causes the maximum amplitude of RR

First Chip Design 17

signal to be 1.6µA x 1Ω = 1.6µV. Since the quantization step size, i.e., least significant bit

(LSB), of a moderate-resolution (8-10bits) low-power analog-to-digital converter (ADC)

is typically 0.5 - 5 mV, the AFE should provide at least 20 dB gain for ECG signal and 70 dB

gain for RR signal to ensure that output SNR is not affected by the ADC’s quantization

noise.

VCC

VREF

Left Arm (LA)Right Arm (RA)

Left Leg (LL)

ECG signal

RR Q-branch

RR I-branch

Mo

du

lati

on

VCM

10kHz

Clk

Q-Path

I-Path

Chip

Off-chip20kHz Clk

Pre-amp

Figure 2.3. Simplified block diagram of the integrated front-end receiver.

To collect ECG signal, two electrodes (LA and RA) are used to make a differential

voltage measurement. To collect RR signal, a high-frequency AC current of known am-

plitude is injected into the tissue through the two electrodes as mentioned before. The

AC current is the carrier signal which is amplitude-modulated by the RR signal, and the

differential voltage is generated across two electrodes since ∆Z changes during respira-

tion. The modulation frequency of the AC current is set to 10 kHz which is much larger

than the frequency of the ECG signal so that there is little interference between ECG and

modulated RR signals.

First Chip Design 18

After collecting both of the signals, a low-noise preamplifier is used to amplify both

ECG and RR signals. In the amplifier, the active ground circuit is utilized to increase

CMRR, and the third electrode (LL) is used as reference electrode here.

After the preamplifier, in order to get ECG signal, LPFs are needed to remove the

modulated RR signal. To get RR signal, HPFs are utilized first so that the ECG signal is at-

tenuated. A quadrature demodulator is then employed, and both the real and imaginary

components of RR signal are extracted. At the output of the demodulator, band-pass fil-

ters (BPFs) are used to remove the high-frequency carrier signal and the RR signal is

selected. The resulting spectra is shown in Fig. 2.4. Since the amplitude of RR signal is

small and the gain of the preamplifier is not enough, to ensure that the signal can be

digitized without adding significant quantization noise, there is another gain stage for

further amplification of the RR signal.

2.3 Clock Generator Block

In this design, since the RR signal needs to be modulated and demodulated, four clocks

are needed to be fed into the mixers, including 0, 90, 180, and 270 clocks. An easy

method to generate these four clocks is shown in Fig. 2.5 if an off-chip clock is utilized.

D-flip-flop is used to divide the frequency of the input signal by two, and an inverter fol-

lowed by the D-flip-flop is able to generate a clock with 90 phase shift. The simulation

result of the clock generator block is shown in Fig. 2.6. It’s shown that clk1 and clk2 have

90 phase shift, and the frequency of them is half of the frequency of off-chip clock.

Clocks are only needed when user wants to monitor RR signal. The NAND gate can

disable the clocks when user only wants to monitor ECG signal so that clock feedthrough

has no influence on the measurement of the ECG signal.

First Chip Design 19

0-fModulated-RR

-700 Hz

-10 kHz

fModulated-RR

+700 Hz

10 kHz

ECG

f

0-fModulated-RR

-10 kHz

fModulated-RR

10 kHz

ECGf

LO

-fModulated-RR

-20 kHz

fModulated-RR

20 kHz

f

0.65 Hz

LOerror

10 kHz-10 kHz

Signal

Signal

Signal

0.1 Hz

-0.65 Hz

-0.1 Hz

fRR

0.5 Hz

-fRR

-0.5 Hz

LO

LOerror

Figure 2.4. Typical spectra of the ECG, modulated RR, and baseband RRsignals before the HPF, before the quadrature modulator, and after themodulator.

Q

QSET

CLR

S

R

D

clk

Q

QSET

CLR

S

R

D

clk

NAND

1x

1x

1x

3x

3x

9x

9x 9x

9x

clk2_ba

clk1_ba

clk2

clk1clk_in

enable

Figure 2.5. Clock generator circuit used to generate 0, 90, 180, and 270clocks.

First Chip Design 20

Time (ms)0 0.2 0.4 0.6 0.8 1

Vo

ltag

e (V

)

0

1

2

3offchip clock

Time (ms)0 0.2 0.4 0.6 0.8 1

Vo

ltag

e (V

)

0

1

2

3clk1

Time (ms)0 0.2 0.4 0.6 0.8 1

Vo

ltag

e (V

)

0

1

2

3clk2

Figure 2.6. Simulation result of the clock generator block. clk1 and clk2have 90 phase shift.

2.4 Controlled Injected Current Block and Mixer

Since the varying impedance term of thorax ∆Z is hard to predict, the amplitude of AC

current should be able to be controlled. The preamplifier has a limited linear range of

around 7.5 mV and a fixed amount of input-referred noise of 8.1 µVr ms over a band-

width of 300 Hz. Therefore, the SNR will be poor if the modulated RR is small, while the

circuit will saturate and generate inter-modulation distortion between the ECG and RR

signals if the modulated RR signal is too large. To control the injected current in this

design, two programmable binary-weighted 4-bit current mirrors are used to generate

controllable injected current. As Fig. 2.7 shows, the minimum injected current is 100 nA.

If all the switches are closed, the maximum injected current should be 1.6 µA, which is

still smaller than the AHA-recommended safe limit current of 10 µA. Because of the four

switches, the injected current is able to vary from 100 nA to 1.6 µA in steps of 100 nA. In

this case, the user is able to change the injected current if ∆Z changes.

First Chip Design 21

10

0n

A

Inmos-out

VCC

Ipmos-out

RA

LA

Clk

Clk

Clk

ɸ1 ɸ2 ɸ3 ɸ4

ɸ1 ɸ2 ɸ3 ɸ4

8I4I2III

8I4I2III10

0n

A

Figure 2.7. Programmable current mirrors and double balanced mixerused for generating I AC . The amplitude of I AC can be programmed over a1:16 range (4 bits) by the switches φ1-φ4.

Besides the programmable current mirrors, a passive double-balanced mixer is used

here. The output of the two current mirrors are connected to the RF port of the mixer,

and the high frequency clock signal which is set to 10 kHz here is fed to the LO port.

The outputs of the mixer are connected to the RA and LA electrodes. Thus, the AC cur-

rent is injected into the body through the electrodes and the AC current is amplitude-

modulated by the low-frequency RR signal.

2.5 Preamplifier Block

The main preamplifier as shown in Fig 2.8 is a modified version of the circuit described

in42,43. This low-noise amplifier is utilized to amplify both the ECG and RR signals.

Three electrodes are included in the amplifier. RA electrode and LA electrode are con-

nected to the inputs of the amplifier (Vi n+ and Vi n−), and the reference electrode is con-

nected to the output of the common-mode feedback circuit (Vcm f b).

The amplifier is a two-stage instrumentation amplifier. For each stage in the am-

plifier, a single-stage operational transconductance amplifier (OTA) is used, which is

First Chip Design 22

VDD

Vref

Vin+

Vin-

Vcmfb

Vref

Vout

A

A

A

A

A

A

A=

VCMsense

Figure 2.8. Schematic of the low-noise ECG amplifier.

biased in sub-threshold mode to reduce power consumption. The schematic of the OTA

is shown in Fig 2.9. In the OTA, the input differential signals are fed to PMOS devices

since the flicker noise in this process is significantly lower (about 5x) for PMOS devices.

Also, by increasing W/L, the transistors are driven into sub-threshold mode.

For the instrumentation amplifier, the mid-band gain of each stage is set to 10 by

choosing the ratio of capacitors in the feedback loop. Therefore, the total gain of the

amplifier is 100 (40 dB). Compared to some standard instrumentation amplifiers, ca-

pacitors instead of resistors are utilized in the feedback loop to set up the gain. That is

because resistors contribute noise and the matching property of resistors is poor. How-

ever, using capacitors can not provide the DC voltage levels for the OTA. To overcome

this problem, the MOS pseudo resistors described earlier are used so that DC current is

able to flow. These pseudo resistors act as very large resistors for small signals, shown

as the "A" block in Fig 2.8. To reduce the distortion of large signals, two pseudo resistors

First Chip Design 23

are placed in series. Thus, the DC gain is set to 1 by using these pseudo resistors, while

the AC gain is set by the capacitors in parallel with the pseudo resistors.

VCC

OUT

V+ V-

VB

Figure 2.9. Schematic of the OTA used four times in the preamplifier.

Besides, a common-mode feedback path is included in the amplifier to remove common-

mode signal, such as the powerline interference. The reference electrode, usually placed

on the leg, is connected to the output of the driven-ground circuit denoted as Vcm f b in

Fig. 2.8. The common-mode voltage is sensed in the first stage of the instrumentation

amplifier. And the OTA in the common-mode feedback amplifies the difference between

the sensed common-mode voltage and the reference voltage which is set to 0.75 V in this

design. Since the output impedance of the feedback is in series with the impedance of

the reference electrode, it is necessary to reduce the output impedance of the circuit so

that the electrode with an impedance of around 2 kΩ can be driven. Therefore, a super-

buffer circuit is utilized to lower the output impedance.

First Chip Design 24

For the standard source follower, the output impedance is Ro = 1gm+gmb

. However,

for this super-buffer circuit, the negative feedback through the additional PMOS re-

duces the impedance by a factor of about gm2ro1, resulting the output impedance of

Ro = 1gm+gmb

1gm2ro1

. Therefore, a typical electrode with an impedance of around 2 kΩ

can be driven effectively. Another thing that needs to be mentioned is that the open

circuit voltage gains of both standard buffer and super buffer do not have much differ-

ence. The gain of the standard source follower is vovi

= gm ro1+(gm+gmb )ro

, while the gain of the

super-buffer circuit is vovi

= gm1ro1

1+(gm1+gmb1)ro1+ 1gm2ro2

. Thus, if gm2ro2 À 1, both voltage gains

are approximately equal to 1, and the super source follower has little effect on the open

circuit voltage gain. The detail description of the super source follower is in44.

Fig. 2.10 shows the simulation result of the ECG path. The mid-band gain is around

40 dB as shown in Fig. 2.10(a). If a 2 mV 100 Hz sinusoid wave and a 10 kHz carrier signal

are applied, both of them get amplified as shown in Fig. 2.10(b). Fig. 2.10(c) shows the

output of the ECG path, and it is clear to see that the high-frequency carrier is reduced

by the LPFs. From Fig. 2.10(c), it can be estimated that the gain in ECG path is around

40 dB, which is in the agreement with the design value.

2.6 Wide-linear Range OTA and Filters

This AFE is designed to monitor ECG and RR simultaneously. Therefore, LPFs are needed

to remove high-frequency modulated RR signal in ECG path, and both HPFs and LPFs

are needed to remove ECG as well as high-frequency carrier in RR path. First-order

Gm −C LPF and HPF topologies, as shown in Fig. 2.11(a) and (b), are used several times

First Chip Design 25

Frequency (Hz)10-2 100 102 104 106

Gai

n (

dB

)-40

-20

0

20

40

Time (ms)0 2 4 6 8 10 12

Vo

ltag

e (V

)

0.2

0.4

0.6

0.8

ECG Amp output

Time (ms)0 2 4 6 8 10 12

Vo

ltag

e (V

)0.2

0.4

0.6

0.8

ECG output

Figure 2.10. Simulation results of (a) the AC response of the preamplifier.(b) the output of the preamplifier when an input of 2 mV 100 Hz sinusoidwave is used. (c) the output of the ECG path when an input of 2 mV 100 Hzsinusoid wave is used.

in the AFE. The cutoff frequency of the filter shown in Fig. 2.11 is

fcuto f f =1

2πRC= Gm

2πC(2.1)

Vin

CVoutGm

Vin

CVout

Gm

Vref

Figure 2.11. (a) First-order Gm −C LPF topology. (b) First-order Gm −CHPF topology.

First Chip Design 26

VCC

Out

V+ V-

VB

i+ i+

i-iB

iout

Vs

Figure 2.12. Basic single-stage five transistor OTA with NMOS input transistors.

For the OTA in the filters, considering the basic five-transistor OTA in Fig. 2.12 first.

Assuming the input pair of transistors operates in the sub-threshold region, we have

i+ = I0eκs v+−vs

ϕt (2.2)

i− = I0eκs v−−vs

ϕt (2.3)

i++ i− = iB (2.4)

i+− i− = iout (2.5)

κs = 1

1+ γ

2pϕ0+VSB

(2.6)

γ=√

2εsi qNA

Cox(2.7)

where ϕt = kTq is the thermal voltage which has a typical value of 26 mV at the tempera-

ture of 300K; ε is the dielectric constant of Silicon; and NA is the dopant atom density.

From equation (2.2)-(2.5), we can get the conclusion that

iout = iB tanh

(κs(v+− v−)

2ϕt

)(2.8)

First Chip Design 27

and we define the linear range as the slope of iout at the origin, i.e., when v+− v− = 0:

VL = 2ϕt

κs(2.9)

If ks=0.7 and ϕt =26 mV, then the linear range of the basic five-transistor OTA VL is

around 75 mV, which is not sufficient to filter the amplified ECG signal (typical ampli-

tude = 100-200 mV). To overcome this, the OTA used in the AFE is a modified version

of the wide-linear-range OTA (WLR OTA) presented in45, as shown in Fig. 2.13. In this

design, source degeneration and bump linearization techniques are used to increase the

linear range, similar to the circuit in45. However, the circuit here uses gate terminals as

inputs instead of well terminals so that the layout is simplified. In this case, the cost is

the reduced linear range, but it is still sufficient for this application.

VCC

M3

M7

M11

M9

Iout

V+V-

VB

VCC VCC

M1

M4

M2

M5 M6

M8

M10

M12

Figure 2.13. The wide-linear-range OTA used several times in the AFE.

The source degeneration method is a common technique to increase the linear range.

It was first used in vacuum-tube design46, and it was then used in bipolar design as

emitter degeneration47. By using source degeneration, the current flowing a transistor

First Chip Design 28

is converted to a voltage through a resistor, and that voltage is then fed back to the source

of the transistor. In this case, the current is decreased. The linear range increases by a

factor of approximately 2 by using source degeneration method.

The bump linearization is a technique to linearize a hyperbolic tangent function48.

In a bump differential pair, there is a central arm including two transistors in series. The

current flowing through these two bump transistors is a bump-shaped function of the

differential voltage. The current in the outer arm (I) is the usual hyperbolic tangent func-

tion of the differential voltages (V). However, the bump transistors steal current near the

origin of the I-V curve. If ω is the ratio of the W/L of the bump transistors and the W/L

of the transistors in the outer arm, we can get the following equation

iout = si nhx

β+ coshx(2.10)

where

β= 1+ ω

2(2.11)

x = κ(v+− v−)

ϕt(2.12)

The optimal value of β is 2, in the following sense. If we Taylor expand the equation

(2.10) at β=2, it can be found that no cubic distortion term exists.

si nhx

2+ coshx= x

3− x5

540+ x7

4536+· · · (2.13)

As a comparison, the cubic distortion exists when β=0.

t anhx

2= x

2− x3

24+ x5

240+· · · (2.14)

The bump linearization increases the linear range by a factor of around 1.5, and it

does not contribute noise. Therefore, by using both the source degeneration and bump

linearization techniques, the linear range is increased by a factor of 2×1.5 = 3, setting

First Chip Design 29

the VL to around 270 mV, which is enough to filter out the amplified ECG and RR signals.

In the ECG path, two first-order Gm −C LPFs with cutoff frequency of 1 kHz are used to

reduce the high-frequency modulated RR signal in ECG path. In RR path, four first-order

HPFs with cutoff frequency of 700 Hz are first used to reduce the ECG signal before the

demodulator.

2.7 Second Gain Stage for Amplifying RR signal

After the demodulator, both the I and Q components of RR signal are extracted. But

the amplitude of the RR signal is still small, as analyzed below. Considering the varying

impedance term of the thorax ∆Z is 1 Ω19 and the injected AC current is the maximum

current of 1.6 µA. The preamplifier can provide 40 dB gain, however there is 4 dB conver-

sion loss because of the mixer. Thus, after the demodulator, the amplitude of RR signal

is around 100 µV, which is hard to directly be digitized with a moderate-resolution ADC,

or displayed on an oscilloscope or other monitoring devices. To overcome this problem,

another gain stage is used and the circuit is shown in Fig. 2.14.

I+

I-

Ibia

s1

Ibia

s2

RR

I-branchGm1

Gm

2

Vref

LPF (0.65 Hz) HPF (0.1 Hz) Second gain stage

C1

C2

C3

C4

Vref

Figure 2.14. Schematic of the BPF and second gain stage used to processeach demodulated RR signal component. The capacitors C1 and C2 are400 pF off-chip capacitors, while C3 and C4 are 2 pF on-chip capacitors.

First Chip Design 30

The signals coming from the output of the demodulator, including both the I+ (Q+)

and I− (Q−) components, are passed through pseudo-differential BPFs. The upper side-

band, the high-frequency carrier, DC offset, and low-frequency noise are removed. In

the BPF, two first order Gm −C LPFs, in which the WLR OTA described earlier is used,

are utilized in each path of I+ (Q+) and I− (Q−). The -3 dB frequency of the LPFs is set

to 0.65 Hz, which contains the fundamental frequency of RR signal (a typical bandwidth

of 0.5 Hz). Also, since the corner frequency of the LPF is relatively low, 400 pF off-chip

capacitors are used here while the bias current of OTAs is kept at a reasonable value

(1 nA). The HPF is a simple CR circuit, in which adaptive element is used to behave as

a big resistor. The -3 dB frequency of the HPF is set to 0.1 Hz. Also, by using this HPF,

the DC voltage of the input of the OTA can be set to the reference voltage. However,

the disadvantage of using a pseudo resistor in the filter is that its effective resistance

is process-dependent because of threshold voltage variations, which makes the cutoff

frequency also process-dependent.

The circuit following the BPF is the second gain stage for amplifying the RR signal.

The OTAs used here are still the WLR OTA described earlier. This stage also performs the

differential to single-ended conversion. The relationship between the output signal and

the input signal is described by the following equation.

Vout

(Vi n1 −Vi n2)= Gm1

Gm2= Ibi as1

Ibi as2(2.15)

From the equation (2.15), it is easy to set the gain of this stage by choosing the ratio

of two bias currents. Here a current ratio of 50 (34 dB) is set. By doing this, the total gain

for RR signal is around 40−4+34 = 70 dB. And the -4 dB term comes from the loss of the

mixer.

First Chip Design 31

Fig. 2.15 shows the simulation results of RR path. The AC response of the second

RR gain stage is presented in Fig. 2.15(a). If a 1 µV RR signal is applied at the input

of the preamplifier, the output signals of the demodulator in I branch are presented in

Fig. 2.15(b). The final outputs of RR path, including both the I and Q components, are

presented in Fig. 2.15(c). From Fig. 2.15(c), it can be estimated that the gain in RR path

is around 72 dB, which is in good agreement with the theoretical design value.

Freq (Hz)10-3 10-2 10-1 100 101 102

Vo

ltag

e (d

B)

-40

-30

-20

-10

0

10

LPF (0.65 Hz)HPF (0.1 Hz)

Time (ms)0 0.2 0.4 0.6 0.8 1

Vo

ltag

e (V

)

0.7

0.75

0.8

0.85 I+I-

Time (sec)0 0.5 1 1.5 2

Vo

ltag

e (m

V)

-3

-2

-1

0

1

2

3

I branch outputQ branch output

Figure 2.15. Simulation results of (a) the AC response of the BPF after de-modulator. (b) the I component of RR at the output of mixer if a 1 µV RRsignal is applied at the input of the preamplifier. (c) the final output of RRpath if a 1 µV RR signal is applied at the input of the preamplifier.

32

3 Measurement Result of First Chip

In this chapter, the measurement of the first chip is discussed. The chip was fab-

ricated in the OnSemi 0.5 µm CMOS process through the MOSIS Educational Program

(MEP). The die photograph of the AFE is shown in Fig. 3.1, which includes the active area

of 1050 µm × 600 µm.

Q-path RR Amp and LPF I-path RR Amp and LPF

CL

K d

ivid

er RR HPF

LPF

Amp

EC

G p

art

Current

mirrorAFE

Figure 3.1. Die photograph of the fabricated AFE, which has an activearea of 1050 µm × 600 µm. Major blocks are labeled.

To test the chip, a printed circuit board (PCB) was designed to set up the measure-

ment environment. All off-chip capacitors for the filters were soldered on the board. Sig-

nal generators were used to provide a sinusoid wave (simulating ECG) and a square wave

(simulating high-frequency RR carrier). These signals were fed into the chip through the

Measurement Result of First Chip 33

Figure 3.2. The PCB used to test the first chip.

BNC connectors on the PCB. For off-chip bias currents, Keithley source meters were

used to source the desired current and measure the corresponding voltages. Then, po-

tentiometers were used to provide the voltages corresponding to the desired currents. In

this case, it is then only necessary to provide voltages instead of off-chip bias currents.

Specifically, by using source meters, the current could be swept so that an I-V curve of

the on-chip bias transistor could be obtained. It was easy to find the corresponding volt-

ages of the desired off-chip bias current by using the measured I-V curve. The I-V curve

used to set a 200 nA off-chip bias current is shown in Fig. 3.3. More information about

the schematic and the layout of the PCB board can be found in the appendix.

The AFE drew 6.2 µA from a single 3 V power supply. Table 3.1 compares our re-

ceiver with other reported ECG AFEs. It can be concluded that our AFE has lower power

consumption, and our AFE can monitor both ECG and RR signals while others can only

monitor ECG. For the future use of this AFE, it is easy to power this AFE by using two

1.5 V alkaline batteries or one 3 V lithium battery.

Measurement Result of First Chip 34

Current (uA)0 0.2 0.4 0.6 0.8 1.0

Vo

ltag

e (V

)

0.8

1

1.2

1.4

1.6

I-V curve used to set 200 nA bias current

Figure 3.3. The measured transistor I-V curve used to set the 200 nA off-chip bias current. This curve was measured by using a Keithley sourcemeter.

Table 3.1. Comparison table with the state of the art

Parameters This work 20133 201249 20112

TechnologyCMOS CMOS CMOS CMOS0.5 µm 0.18 µm 0.35 µm 0.35 µm

Power Supply 3V 2.9-5.5V 2.0-3.5V 3.3VCurrent 6.2µA 190µA 170µA 725µA

Gain (ECG) 40 dB 51 dB 40 dB 66.5 dB

Note: This work monitors both ECG and RR signals, while the circuits in other reference onlymeasure ECG.

3.1 Programmable Injected Current Block and Mixer

The first part tested was the programmable injected current block and the mixer. To test

them, a fixed resistor (100 kΩ) was placed between LA and RA as shown in Fig. 3.4. By

placing a fixed resistor here, there should be a square input voltage waveform across the

RA and LA. That is because the injected current goes through the fixed resistor from RA

to LA when Clk was high and from LA to RA when Clk was low.

The measurement result of the programmable current mirrors and mixer is shown in

Fig. 3.5. There are 4 bits in the programmable injected current block, so 16 square wave-

forms are included in Fig. 3.5(a) by changing the programmable bits. To further analyze

Measurement Result of First Chip 35

10

0n

A

Inmos-out

VCC

Ipmos-out

RA

LA

Clk

Clk

Clk

ɸ1 ɸ2 ɸ3 ɸ4

ɸ1 ɸ2 ɸ3 ɸ4

8I4I2III

8I4I2III10

0n

A

100k

Figure 3.4. Circuit used to test the programmable injected current blockand the mixer. A fixed resistor (100 kΩ) was placed between LA and RA.

the function of this block, the actual amount of injected current from the programmable

current mirrors were calculated. Fig. 3.5(b) shows the calculated positive and negative

currents from current mirrors as a function of the four-bits code. The actual currents

are larger than the theoretical values, and this is probably because the I-V curves used

to set the off-chip bias currents were not calibrated well. The injected current is depen-

dent on the off-chip bias current, therefore the actual injected current is different from

the design value when the off-chip bias current is not the theoretical value. In addition,

there is an offset |IP − IN | which has an average value of 65 nA, which is relatively small

(less than 1 LSB = 100 nA).

3.2 ECG Amplifier

The ECG amplifier is the most important part in the AFE. And the measurement result

is presented here.

The differential frequency response of the ECG signal path (consisting of ECG am-

plifier and LPF) is shown in Fig. 3.6(a). The measured upper cutoff frequency is around

Measurement Result of First Chip 36

Time (sec)0 0.02 0.04 0.06 0.08 0.1

Vo

ltag

e (V

)

-0.4

-0.2

0

0.2

0.4

Digital code1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Cu

rren

t (μ

A)

0

0.5

1

1.5

2

2.5

3PositiveNegativeOffset

Figure 3.5. (a) Square voltage waveforms across the RA and LA with dif-ferent injected currents when a 100 kΩ resistor was used between RA andLA. (b) Measured positive and negative currents as a function of the digi-tal code.

150 Hz, however, the theoretical value is 1 kHz. The reason for this is probably a differ-

ent off-chip bias current was used so that the cutoff frequency of the LPF changed. Also,

the mid-band gain is around 39.5 dB, which is in good agreement with the design value

of 40 dB. Fig. 3.6(b) is the common-mode frequency response of the ECG path without

utilizing driven-ground circuit. The average gain of common-mode signal in the band of

100 Hz is around 0.8 dB, resulting in the CMRR in ECG path of 39.5 dB - 0.8 dB = 38.7 dB

without driven-ground circuit.

Frequency (Hz)100 101 102 103 104

Gai

n (

dB

)

-20

-10

0

10

20

30

40

Frequency (Hz)100 101 102 103

Gai

n (

dB

)

-40

-30

-20

-10

0

10

Figure 3.6. (a) Differential frequency response of both the ECG amplifierand the following LPF. (b) Common-mode frequency response of both theECG amplifier and the following LPF without utilizing driven-ground cir-cuit.

Measurement Result of First Chip 37

The input and output noise measurement results of the both the ECG amplifier and

the following LPFs is shown in Fig. 3.7(a) and (b). From the result, it is easy to con-

clude that the 1/ f corner frequency is around 25 Hz. The thermal noise PSD is around

254 nV/Hz1/2, which is slightly larger than the simulation result of 162 nV/Hz1/2. The

minimum detectable signal is around 38 µV in the band of 1-150 Hz. Since the upper

cutoff frequency of the LPF is around 150 Hz, the thermal noise is flat in the range of

25 Hz to 150 Hz but not flat when frequency is higher than 150 Hz.

Freq (Hz)100 101 102 103

No

ise

(V2 /H

z)

10-12

10-10

10-8

10-6

10-4

Output noise

Freq (Hz)100 101 102 103

No

ise

(V/s

qrt

(Hz)

)

10-7

10-6

10-5

10-4

Input noise

Figure 3.7. (a) Output noise of the both the ECG amplifier and the follow-ing LPF. (b) Input noise of the both the ECG amplifier and the followingLPF.

Fig. 3.8 proves that the preamplifier can amplify both the ECG and high-frequency

carrier signals as expected. As described in the second chapter, the high-frequency clock

can be disabled when user only wants to monitor ECG signal. Fig. 3.8(a) shows the out-

put of the ECG path when clock was disabled. The input signal, a 2 mV, 10 Hz sinusoid

wave generated from a signal generator, which simulated ECG signal, was amplified by

around 40 dB as expected.

Fig. 3.8(b) shows the output of the ECG path when input signal was the sum of a

1.5 mV, 10 Hz sinusoid (simulating ECG) and a 1.4 mV, 100 Hz square wave (simulating

Measurement Result of First Chip 38

RR carrier). Since the upper cutoff frequency of the ECG LPF is around 150 Hz as de-

scribed earlier, a relatively low-frequency clock (100 Hz) was used here so that the ECG

LPF did not have effect on it. It is easy to see that both the ECG and the RR carrier signals

got amplified by around 40 dB in good agreement with simulations.

Time (s)0 0.2 0.4 0.6 0.8 1

Vo

ltag

e (V

)

-0.2

-0.1

0

0.1

0.2ECG output

Time (s)0 0.05 0.1 0.15 0.2

Vo

ltag

e (V

)

-0.2

0

0.2

output signal

Figure 3.8. (a) Output of the ECG path when high-frequency clock is dis-abled. The input signal is a 2 mV, 10 Hz sinusoid wave. (b) Output of theECG path to the sum of a 1.5 mV, 10 Hz sinusoid (simulating ECG) and a1.4 mV, 100 Hz square wave (simulating RR carrier).

Fig. 3.9 shows a real ECG signal detected from human body by using two standard

patch electrodes which were placed on the right and left upper arms. The data pre-

sented has been filtered with a fourth-order Butterworth LPF with cutoff frequency of

50 Hz. The frequency of R wave in the output ECG signal is around 1.3 Hz which is in

good agreement with the frequency of the heart rate. But the amplitude of the output

signal is smaller than what we expected. One possible reason is that the placement of

the electrodes may not be optimal for measuring ECG signal.

Although the measurement results above have proven that the preamplifier can am-

plify both ECG and RR carrier signals as expected, there is a major problem that we

found during the experiment. The DC voltage at the output of the ECG path shifted

as time went by as shown in Fig. 3.10. Theoretically, the DC voltage at the output of the

preamplifier should be equal to the reference voltage (0.75 V) which was provided from

Measurement Result of First Chip 39

0 1 2 3 4 50

5

Time (s)0 1 2 3 4 5

Vo

ltag

e (m

V)

0

3

6

9

12

15

ECG signal

Figure 3.9. Detected ECG signal after baseline removal and filtering witha fourth-order LPF ( fc = 50 Hz).

a external power supply. But every time the chip was turned on, the DC voltage at the

output of the ECG path started to increase from around 0.3 V and finally saturated at

around 2.4 V after approximately 900 seconds. To make the AFE work as expected, a

suitable range for this DC voltage was concluded during the experiment, in which 40 dB

gain could be obtained from this preamplifier. And this DC voltage range is from around

0.6 V to 1.5 V. In this case, the preamplifier could only work well for about 2 minutes.

The reason causing this problem is the leakage current in the adaptive element, which is

a well-known problem with pseudo resistors50. In this AFE, NMOS transistors are used

in the pseudo resistors. The body terminal of NMOS transistor has to be connected to

ground because of the process of this chip. In this case, if we assume the source is the

high-impedance node, connecting bulk to ground creates a non-zero reverse-bias volt-

age across the source to body junction diode, resulting in the leakage current in a pseudo

resistor. Because of this leakage current, the DC voltage starts to change slowly when the

leakage current flowing through the adaptive element. It is possible to avoid this prob-

lem by using PMOS transistors instead of NMOS transistors to work as pseudo resistors

because the body terminal of a PMOS transistor can be connected to the drain so that

Measurement Result of First Chip 40

all diodes are zero biased. Another way to avoid this problem is to use switched capaci-

tor resistors instead of pseudo resistors, but the cost is that the complexity of the circuit

increases.

Time (s)100 200 300 400 500 600 700 800 900 1000

Vo

ltag

e (V

)

0

0.5

1

1.5

2

2.5

Output of ECG path

Figure 3.10. The DC voltage at the output of the ECG path shifted from0.3 V to 2.4 V in around 900 seconds.

3.3 RR Measurement

To test the functionality of the RR path, a body model as shown in Fig. 3.11(a) was used

to generate both the ECG-like and RR-like signals at the same time. In the body model,

the source V1 is a sinusoid wave behaved as ECG signal. By choosing the values of resis-

tors and capacitors, the ECG-like signal can be fed to RA and LA. The source V2 is a 1 Hz

triangle wave which is able to turn the discrete FETs on or off, so that the impedance

between RA and LA is modulated by around 1%. Fig. 3.11(b) shows the transient sim-

ulation result of this body model. It is able to find the modulated RR-like signal when

a 100 nA 1 kHz AC current was used. Fig. 3.11(c) shows that the body model is able to

generate the modulated RR signal (999 Hz and 1001 Hz). This body model was soldered

on the PCB board.

Measurement Result of First Chip 41

LA

RA

33

μF

33

μF

1.8 k

1.8 k

1.8 k

1.8 k

0.56 k

0.56 k

5.6 k

5.6 k

VrefV1 V2

Time (s)0 1 2 3 4 5

Vo

ltag

e (V

)

×10-3

-1.5

-1

-0.5

0

0.5

1

1.5

2

Frequency (Hz)997 998 999 1000 1001 1002 1003

20lo

g(a

mp

litu

de/

1Vo

lt)

(dB

)

-160

-140

-120

-100

-80

Figure 3.11. (a) Body model used to generate ECG-like and RR-like sig-nals. (b) Simulation result of the modulated RR-like signal when 100 nA1 kHz AC currents are fed to LA and RA. (c) The FFT result of the transientsignal generated by the body model.

Fig. 3.12 shows the measured I and Q components of RR signal after filtering with a

fourth-order Butterworth LPF with cutoff frequency of 5 Hz. The 0.5 Hz ramp wave was

used to control the FETs on or off. The results prove the functionality of the RR path with

a total gain of around 72 dB.

Measurement Result of First Chip 42

Time (s)0 2 4 6 8 10 12

Vo

ltag

e (V

)

-0.04-0.03-0.02-0.01

00.010.020.030.04

I branch outputQ branch output

Time (s)0 2 4 6 8 10 12

Vo

ltag

e (V

)

0

2

4

6Input ramp wave

Figure 3.12. Input triangle wave used to modulate the impedance be-tween RA and LA at 0.5 Hz, and measured I and Q RR outputs after fil-tering with a fourth-order Butterworth LPf with cutoff frequency of 5 Hz.

43

4 Second Chip Design

From the measurement results of the first chip, several updates are needed. For the

second chip, a fully differential AFE is designed to improve CMRR and reduce distor-

tion. Also, more gain in RR path is needed since the amplitude of RR is small. Besides, it

is necessary to get rid of the DC voltage shifting problem caused by the leakage current

in the adaptive element. In addition, the driven-ground circuit needs to be updated be-

cause it did not work well during the experiment. Finally, four-terminal sensing method

is utilized to detect respiratory rate. AC current is injected into body through a pair of

electrodes while the signal is sensed by another pair of electrodes, resulting more accu-

rate measurement of body impedance. The simulation result proves the functionality of

the second AFE, which draws 14 µA from a 3 V power supply. The gain for ECG signal is

40 dB (46 dB if the preamplifier works in high gain mode) while the gain for RR signal is

82 dB (88 dB if the preamplifier works in high gain mode).

4.1 Fully Differential Wide-linear-range OTA

The second chip is a fully differential AFE. Therefore, a fully differential WLR OTA is

needed. Fig. 4.1(a) is the fully differential version of the wide-linear-range OTA. For this

fully differential OTA, a common-mode feedback (CMFB) must be used. In the feedback,

Second Chip Design 44

the common-mode voltage is sensed and then compared with a suitable reference volt-

age. Then the circuit feeds back the correcting signal so that the output common-mode

voltage is set to a desired value. Normally, two resistors are used to detect the common-

mode voltage, and an OTA is used to compare the sensed common-mode voltage with a

reference. However, in our chip, a differential-difference operational transconductance

amplifier (DDOTA)51, as shown in Fig. 4.1(b), is used to detect the common-mode volt-

age and compare it with a reference voltage. No passive component is used to detect

the common-mode voltage here, which ensures that the CMFB circuit has high input

impedance and does not load the main OTA (which would reduce its differential voltage

gain). The stage of the fully differential OTA with the common-mode feedback is shown

in Fig. 4.1(c). For the load capacitors C1 and C2, the CMFB half circuit only sees C1 as

the load capacitor while the differential half circuit sees C1+2C2 as the load capacitor52.

Thus, the values of C1 and C2 can be adjusted to independently set the bandwidths of

the differential and common-mode paths. We connect the output of the DDOTA (V f b)

to the bias voltage node (VB ) so the negative common-mode feedback is generated. The

fully differential OTA stage in the preamplifier is shown in Fig. 4.2(a), and the gain is set

by the ratio of capacitors in the feedback loop. To set up the input DC voltage, a modified

switched capacitor resistors topology is used, which will be described in detail later.

To check the stability of this single stage of fully differential OTA in the preamplifier

shown in Fig. 4.2(a), the Bode plot is plotted as shown in Fig. 4.2(b). From the Bode plot,

it is clear to see that the phase margin is around 80 and the gain margin is around 12 dB,

which proves the stability of the fully differential OTA.

Second Chip Design 45

VCC

M3

M7

M11

M9

out+

V+V-

Vbias

VCC VCC

M1

M4

M2

M5 M6

M8

M10

M12

out-

VB

Vpp

Vfb

Vpn Vnp Vnn

VCC VCC

Ib

Vref

C1

C1

C2

Figure 4.1. (a) Fully differential version of wide-linear-range OTA. (b)Differential-difference OTA used in CMFB. (c) Fully differential OTA stage.

4.2 Fully Differential Filters

In order to design fully differential filters, the fully differential OTA presented above is

used. In ECG path, the LPFs are needed to remove high-frequency modulated RR signal.

In RR path, HPFs are needed to remove ECG signal before the demodulator. After the

demodulator, LPFs with off-chip capacitors are needed to remove high-frequency RR

carrier signal.

Second Chip Design 46

Vin1

Vin2

1pf

1pf

0.1pf

0.1pf

Sw

itch

cap

12

Sw

itch

cap

12

0.1pf

0.1pf

0.1pf

Vout1

Vout2

Vref

Vref

Frequency (Hz)10-3 10-2 10-1 100 101 102 103 104 105 106 107

Ph

ase

(deg

ree)

-200

-100

0

100Frequency (Hz)

10-3 10-2 10-1 100 101 102 103 104 105 106 107

Gai

n (

dB

)

-40

-20

0

20

Figure 4.2. (a) A single stage of fully differential OTA in the preamplifier.(b) Bode plot of the fully differential OTA stage.

Fig. 4.3 shows the filters by using the fully differential OTA. The LPF shown in Fig. 4.3(a)

is the modified version of the second RR gain stage in the first chip. Fig. 4.3(b) is the stan-

dard fully differential Gm −C HPF. The time constant of the fully differential filters is CGm

(if Gm1 is equal to Gm2 in the LPF). Fig. 4.4(a) shows the AC simulation result of the HPF

with cutoff frequency of 1 kHz to remove ECG signal in RR path. Fig. 4.4(b) shows the AC

simulation result of the LPF with cutoff frequency of 5 Hz to remove high-frequency RR

carrier signal in RR path. To get this low cutoff frequency, 3 nF off-chip capacitor is used

here.

Second Chip Design 47

Ibia

s1

Ibia

s2

Gm1

Gm

2

C

C

Ibia

s

C

C

Figure 4.3. (a) Fully differential LPF. (b) Fully differential HPF.

Frequency (Hz)10-2 100 102 104 106

Gai

n (

dB

)

-120

-100

-80

-60

-40

-20

0

Frequency (Hz)10-4 10-2 100 102 104 106 108

Gai

n (

dB

)

-150

-100

-50

0

Figure 4.4. The AC simulation results of (a) HPF with cutoff frequency of1 kHz to remove ECG signal in RR path. (b) LPF with cutoff frequency of5 Hz to remove high-frequency RR carrier signal in RR path. 3 nF off-chipcapacitor is used.

4.3 Fully Differential Preamplifier Block

By using the fully differential OTA described above, a fully differential preamplifier is de-

signed as shown in Fig. 4.5. There are two stages in the preamplifier. In each stage, the

Second Chip Design 48

OTA is AC coupled. The gain is set by the ratio of capacitors in the feedback loop. The

gain of the first stage is set to 10 while the gain of the second stage is programmable.

By turning the switches on and off, the gain can be set to 10 or 20 in the second stage.

In addition, the driven-ground circuit is updated. Two OTAs in unity feedback work as

buffers to sense the common-mode voltage. Then, the sensed common-mode voltage

will be compared to a reference voltage, and the difference of them is amplified by two

gain stages. The first gain stage in the driven-ground circuit includes two OTAs, with

the one in unity feedback acting as a buffered resistor. The gain is set by the ratio of

two bias currents. In order to ensure that Gm2 can work in linear range, Ibi as1 is a pro-

grammable bias current so that the gain of this stage can be controlled. In this design,

the programmable gain varies from 1 to 8. The second gain stage includes an op-amp

with resistors in the feedback loop. The gain is set by the ratio of two resistors, which is

10 (20 dB) in this case. Another reason that an op-amp is used here is that the output

impedance of it is relatively high to drive the electrode impedance of around 2 kΩ.

To set up the DC voltage in the preamplifier, MOS pseudo resistors are generally used

but the DC voltage shifting problem exists. Therefore, in order to create a large resistor

to generate a low-frequency pole, switched capacitor instead of pseudo resistor is used

here. Instead of the standard switched capacitor as described in the first chapter, a novel

structure of switched capacitor shown in Fig. 4.6, which is described in53, is utilized to

provide DC voltage for the amplifier. The idea to design this modified switched capaci-

tor topology is the same as the standard switched capacitor. It works by moving charges

into and out of capacitors when switches are on and off. The equivalent circuits in two

phases are presented in Fig. 4.7, and the effective capacitances are 0.5C and 2.5C, re-

spectively.

Second Chip Design 49

Vin1

Vin2

1pf

1pf

0.1pf

0.1pfS

witch

cap

12

Sw

itch

cap

12

1pf

1pf

0.1pf

0.1pf

Sw

itch

ca

p1

2S

witch

ca

p1

2

0.1pf

0.1pf

0.1pf

Vout1

Vout2

0.1pf

0.1pf

0.1pf

Vref Vref

Vref Vref

Ibia

s1

Ibia

s2

Gm1

Gm

2

Vref

Vref

15pf

Vcmfb

1pf

1pf

200k20k

Vref

Figure 4.5. Circuit of fully differential preamplifier.

When clk2 is high, the effective capacitance is 0.5C as shown in Fig. 4.7(a), and the

total charge q0 is

q0 = (V1 −V2)×0.5C (4.1)

When clk2 becomes low and clk1 becomes high, the effective capacitance is 2.5C as

shown in Fig. 4.7(b). Since the total charge in the switched capacitor keeps the same, the

voltage Vclk1 becomes

Vclk1 =(V1 −V2)×0.5C

2.5C(4.2)

Second Chip Design 50

Thus, some charge transfers from the two capacitors in parallel to the two capacitors

in series, and the amount of it is

q1 =Vclk1 ×0.5C (4.3)

From equation (4.2) and (4.3), it can be concluded that

q1 = (V1 −V2)×0.5C

2.5C×0.5C = (V1 −V2)×0.1C (4.4)

The term q1 here is the charge that is transferred during one cycle. From the con-

tinuous time perspective, we also know that the rate of transfer of charge per unit time

is

I = q × f (4.5)

where f is the transfer rate.

Taking equation (4.4) and (4.5) together, we can get

I = (V1 −V2)×0.1C × f (4.6)

Therefore, for this modified switched capacitor topology as shown in Fig. 4.6, the

equivalent resistance is

Req = (V1 −V2)

I= 10

f ×C(4.7)

This equivalent resistance is 10 times larger than the standard switched capacitor

topology when the same frequency and capacitor value are used. There are several rea-

sons that this modified topology is needed. Since the frequencies of ECG and RR are low,

a low cutoff frequency of around 0.1 Hz is needed, resulting in a big resistor of around

200 GΩ. By using switched capacitor, reducing the clock frequency or reducing the on-

chip capacitor value can generate a larger resistor, as shown in equation (4.7). However,

the minimum reliable on-chip capacitor value is around 50 fF. To get a cutoff frequency

Second Chip Design 51

of 0.1 Hz, the equivalent resistance should be around 200 GΩ if a big on-chip capacitor

of 10 pF is used. In this case, for the standard switched capacitor topology, if the ca-

pacitor in the switched capacitor is 50 fF, the clock frequency should be around 100 Hz,

which is in the band of ECG signal. Therefore, this modified switched capacitor topology

is needed. By using this new structure, we can set the capacitor value as 50 fF and the

clock frequency as 1000 Hz, which is not in the band of ECG signal anymore.

clk1clk2

C

C

C

C

clk2

clk2

clk2

clk2

clk1

clk1

clk1

V1 V2

Figure 4.6. Modified switched capacitor topology.

C CV1 V2

C C C

C

V1 V2Vclk1

Figure 4.7. Equivalent circuits of the modified switched capacitor topol-ogy in two phases. (a) Equivalent circuit when clk1 is on and clk2 is off.(b) Equivalent circuit when clk1 is off and clk2 is on.

Fig. 4.8 shows the simulation result at the output of the preamplifier (Vout1 −Vout2)

when a 2 mV 100 Hz sinusoid wave simulating ECG signal and a 10 kHz square wave sim-

ulating RR carrier signal are used at the input of the circuit. It can be seen that both the

ECG-like signal and the high-frequency carrier signal can be amplified by the preampli-

fier.

Second Chip Design 52

Time (ms)0 2 4 6 8 10 12

Vo

ltag

e (V

)

-0.6

-0.4

-0.2

0

0.2output of preamplifier

Figure 4.8. Transient simulation result at the output of the preamplifierwhen a 2 mV 100 Hz sinusoid wave simulating ECG signal and a 10 kHzsquare wave simulating RR carrier signal are used at the input of the cir-cuit.

4.4 Digital Block

Since the switched capacitor is used in the AFE, some digital circuits are needed to gen-

erate the clocks for switched capacitor. Fig. 4.9 shows the digital blocks in the AFE. The

input frequency is chosen as 38 kHz. To generate clocks for the mixer to extract I com-

ponent and Q component of RR signal, a D flip-flop and the same clock generator block

described in the second chapter are used in the AFE. In this case, clock generator block

has four outputs, and clk1 and clk2 have 90 phase shift. Since there is a D flip-flop in

this branch, the frequency of clk1 and clk2 is 38 kHz/4=9.5 kHz, which is the frequency

of RR carrier signal. Also, the enable function is still available in this block. If the user

only wants to monitor ECG, the high-frequency RR carrier can be disabled.

The other branch of the digital circuit is used to generate the clocks for switched

capacitor. A programmable frequency divider block and a nonoverlapping clock gen-

erator block are included. For the programmable frequency divider block as shown in

Fig. 4.11. Five D flip-flops are placed in series so that they behave like a counter. The

Q of each D flip-flop connects to the input of a XNOR gate, while the other input of the

Second Chip Design 53

Q

QSET

CLR

S

R

D

clk

Programmable

freq divider

clock

generator

nonoverlapping

clk_in

enableclk1

clk1_ba

clk2

clk2_ba

clk1_sc

clk2_sc

Figure 4.9. Digital blocks in the AFE.

XNOR gate is the programmable bit. The programmable bits are set to the desired fre-

quency divider ratio. The reason we need to use the programmable frequency divider

is that the switched capacitor resistor is a discrete-time system. The switched capaci-

tor resistor along with any load capacitance presented at its output acts as an RC LPF.

The frequency response is periodic, and the peaks exist at integer multiples of the clock

frequency fSC . The frequency of signals should be midway between these peaks so the

maximum attenuation can be obtained from the filter, as shown in Fig. 4.10. Therefore,

we need to set the frequency of RR carrier ( fRR ) signal to be an odd multiple of fSC /2,

i.e., fRR = (2n +1) fSC /2, where n = 0,1,2 · ··. In this case, it is easy for users to adjust the

frequency divider ratio by using the programmable frequency divider circuit.

The way this block works is that the counter counts one by one until it is reset when it

counts the value of programmable bits. For example, if the programmable bits are 11000,

which is 24 in decimal, the counter counts starting from 1. When the counter counts 24,

all the Q outputs of each D flip-flop are equal to the corresponding programmable bits,

resulting the following NAND gate generating a high output. This output triggers all the

D flip-flops to reset. The problem of this block is that the delays in each path are not the

same. Therefore, another D flip-flop clocked with the inverse of the input clock is used

Second Chip Design 54

Chapter 9 – Section 1 (5/2/04) Page 9.1-20

CMOS Analog Circuit Design © P.E. Allen - 2004

Example 9.1-5 - Continued

Frequency Response of the First-order, Switched Capacitor, Low Pass Circuit:

0

0.2

0.4

0.6

0.8

1

0 0.2 0.4 0.6 0.8 1

Mag

nitu

de

0.707

|H(jω)|

|Hoo(ejωT)|

ω = 1/τ1

ω/ωc

-100

-50

0

50

100

0 0.2 0.4 0.6 0.8 1

Phas

e Sh

ift (

Deg

rees

)

ω/ωc

Arg[Hoo(ejωT)]

Arg[H(jω)]

ω = 1/τ1

Fig. 9.1-12

Better results would be obtained if fc > 20kHz.

Chapter 9 – Section 1 (5/2/04) Page 9.1-21

CMOS Analog Circuit Design © P.E. Allen - 2004

SUMMARY• Resistance emulation is the replacement of continuous time resistors with switched

capacitor approximations

- Parallel switched capacitor resistor emulation

- Series switched capacitor resistor emulation

- Series-parallel switched capacitor resistor emulation

- Bilinear switched capacitor resistor emulation• Time constant accuracy of switched capacitor circuits is proportional to the

capacitance ratio and the clock frequency• Analysis of switched capacitor circuits includes the following steps:

1.) Analyze the circuit in the time-domain during a selected phase period.2.) The resulting equations are based on q = Cv.3.) Analyze the following phase period carrying over the initial conditions from the

previous analysis.4.) Identify the time-domain equation that relates the desired voltage variables.5.) Convert this equation to the z-domain.6.) Solve for the desired z-domain transfer function.

7.) Replace z by ejωT and examine the frequency response.

Figure 4.10. Frequency response of the first-order LPF by using switchedcapacitor.

following the output of the AND gate. This additional D flip-flop removes all the glitches

at the output of the AND gate, and the output of this additional D flip-flop is used to

reset the five D flip-flops in the counter. This reset signal does not have 50% duty cycle.

Thus, another D flip-flop is used to divide the frequency by two, generating a 50% duty

cycle clock.

There are five programmable bits so the maximum frequency divider ratio is 32. The

frequency of the output clock is described as

fout = fi n

r ati o ×2(4.8)

The clocks of switched capacitor should be two-phase non-overlapping clocks, so

the non-overlapping clocks generator circuit is needed. The circuit of this block is shown

in Fig. 4.11. Several inverters in series are used so that a delay is generated. By using the

NOR gates and these inverters with feedback, two output clocks are non-overlapping

and the delay between them comes from the delay generated by the inverters in series.

Since many digital inputs are needed for the digital circuit, in order to save some

pins of the chip, the programmable three-wire serial bus circuit is used, as shown in

Second Chip Design 55

Q

QSET

CLR

S

R

D

clkQ

QSET

CLR

S

R

D

clkQ

QSET

CLR

S

R

D

clkQ

QSET

CLR

S

R

D

clkQ

QSET

CLR

S

R

D

clk

ratio5

ratio4

ratio3

ratio2

ratio1

Q

QSET

CLR

S

R

D

clk

Q

QSET

CLR

S

R

D

clk

1x

AND

clk_in

clk_outreset

Figure 4.11. Programmable counter.

1x 1x 1x1x 1x 1x 1x 3x 9x

1x 1x 1x 1x 3x 9x

clk1_sc

clk2_sc

Figure 4.12. Two-phase non-overlapping clocks generator block.

Fig. 4.13(a). Microcontroller can be used off-chip and some code can be written to trans-

mit all the digital inputs by using Serial Peripheral Interface (SPI). The D flip-flops work

as shift registers, and the latches are used to save the bits. When the enable signal is high,

the latches retain their previous values to ensure that incorrect values are not passed

into the circuit while the shift register is being programmed. The enable signal goes low

after shift register programming is complete, thus the circuit loads the correct data val-

ues into the latches. The timing diagram for enable, clk, and data signals are shown in

Fig. 4.13(b).

To test the functionality of all the digital circuit in the AFE, the circuit shown in

Fig. 4.9 has been simulated. The following simulation results prove the functionality

of all the digital blocks, as shown in Fig. 4.14 and Fig. 4.15. The input clock is a 38 kHz

Second Chip Design 56

Q

QSET

CLR

S

R

D

clk

DATA

Q

en

Q

QSET

CLR

S

R

D

clk

DATA

Q

en

Q

QSET

CLR

S

R

D

clk

DATA

Q

en

Q

QSET

CLR

S

R

D

clk

DATA

Q

en

clk clk clk clk

DATA

enable enable enable enable

bit[1] bit[2] bit[n] bit[n+1]

enable

clk

data

Figure 4.13. (a) Programmable bus used to set digital inputs. (b) Timingdiagram of SPI.

square wave simulating the off-chip clock, as shown in Fig. 4.14(a). The Fig. 4.14(b) and

Fig. 4.14(c) are the outputs of two clocks with 90 phase shift, which are further used in

the mixer for modulation and demodulation. It is clear to see that the frequencies of the

two clocks in the Fig. 4.14(b) and Fig. 4.14(c) are same and one fourth of the frequency

of the input clock shown in Fig. 4.14(a).

The other branch of the digital circuit shown in Fig. 4.9 is used to generate non-

overlapping clocks for switched capacitor. The simulation results are in Fig. 4.15. Fig. 4.15(a)

and Fig. 4.15(b) show the outputs of two non-overlapping clocks. In this simulation, the

programmable frequency divider ratio is set to 19, therefore the frequency of the clocks

in Fig. 4.15(a) and Fig. 4.15(b) is 1 kHz, which is in agreement with the theoretical value.

If these two clocks are zoomed in, it is clear to see that they are non-overlapping, as

shown in Fig. 4.15(c). In the non-overlapping clock generator block as shown in Fig. 4.12,

Second Chip Design 57

8 inverters are used in series in each branch, resulting a 2 ns delay between the two out-

put clocks for switched capacitor, as shown in Fig. 4.15(c).

Time (s) ×10-30 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1

Vo

ltag

e

0

2

4off-chip clk

Time (s) ×10-30 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1

Vo

ltag

e (V

)

0

2

4clk1

Time (s) ×10-30 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1

Vo

ltag

e (V

)

0

2

4clk2

Figure 4.14. Simulation results of the circuits generating clocks used inthe mixer for modulation and demodulation. (a) The 38 kHz input squarewave simulating the off-chip clock. (b) The first output clock with onefourth of the frequency of the input wave. (c) The second output clockwith one fourth of the frequency of the input wave and 90 phase shift.

4.5 Second Gain Stage for Amplifying RR signal

As we described earlier, the amplitude of RR signal is really small. Therefore, another

gain stage for amplifying RR signal is needed. The circuit shown in Fig. 4.16 contains

three parts, including a HPF and two gain stages.

For the HPF, the modified switched capacitor described earlier is used. The theo-

retical equivalent resistance of the switched capacitor is 200 GΩ, resulting in the cutoff

frequency of this HPF around 0.08 Hz. In order to amplify the extracted RR signal, there

are two gain stages in this block. The first gain stage includes an op-amp. The gain is set

Second Chip Design 58

Time (s) ×10-30 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

Vo

ltag

e (V

)

0

2

4

SC clk1

Time (s) ×10-30 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

Vo

ltag

e (V

)

0

2

4SC clk2

Time (us)494.01 494.011 494.012 494.013 494.014 494.015 494.016 494.017 494.018 494.019 494.02

Vo

ltag

e (V

)

0

2

4

SC clk1SC clk2

Figure 4.15. Simulation results of the circuits generating clocks used inthe switched capacitor. (a) The first output clock. (b) The second outputclock. (c) 2 ns delay between the two non-overlapping output clocks.

Ibia

s1

Ibia

s2

Gm1

Gm

2

Sw

itch

cap

12

Sw

itch

cap

12

Sw

itch

ca

p1

2S

witch

cap

12

10pF

10pF

2pF

2pF

0.1pF

0.1pFVref

Vref

Vref

Iout+

Iout-

Iin+

Iin-

HPF (0.08 Hz) Gain stage 1 Gain stage 2

Figure 4.16. Second gain stage for amplifying RR signal.

by the ratio of the capacitors in the feedback loop, which is set to 20 in the design. To

set up the input DC voltages, the same switched capacitors are used. The second gain

stage is the fully differential version of the second RR gain stage used in the first AFE,

Second Chip Design 59

as described in the second chapter. The gain of this stage is set by the ratio of two bias

currents, which is set to 10 in the design. In this case, the total gain of this block utilized

to further amplify the RR signal is 20×10 = 200 (46 dB).

In order to test the stability of the op-amp stage, the open-loop Bode plot is plotted

as shown in Fig. 4.17. The phase margin is 65 and the gain margin is 18 dB, which is

sufficient for this application. Fig. 4.18 shows the simulated closed-loop Bode plot of the

op-amp stage shown in Fig. 4.16. The phase margin is around 50 and the gain margin is

around 5 dB. Fig. 4.19 shows the transient simulation result at the output of I branch in

RR path (I+− I−). In this simulation, the body model used in Section 3.3 is utilized here

to generate modulated RR signal. The frequency of RR signal is set to 5 Hz to increase

the simulation speed, which is modulated on a 10 kHz carrier signal. At the output of the

I branch in RR path as shown in Fig. 4.19, it can be seen that the high-frequency carrier

signal is removed by the filters in RR path and the 5 Hz RR signal is extracted.

Frequency (Hz)10-2 100 102 104 106 108

Gai

n (

dB

)

-50

0

50

100

Frequency (Hz)10-2 100 102 104 106 108

Ph

ase

(deg

ree)

-400

-300

-200

-100

0

Figure 4.17. Open-loop Bode plot of the op-amp stage.

Second Chip Design 60

Frequency (Hz)10-2 100 102 104 106 108

Gai

n (

dB

)-10

0

10

20

30

Frequency (Hz)10-2 100 102 104 106 108

Ph

ase

(deg

ree)

-200

-100

0

100

Figure 4.18. Closed-loop Bode plot of the op-amp stage.

Time (s)0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1

Vo

ltag

e (V

)

-0.1

0

0.1

0.2

0.3

0.4

0.5I component of RR signal

Figure 4.19. Transient simulation result at the output of I branch in RRpath (I+− I−). The frequency of RR signal is chosen as 5 Hz to increasethe simulation speed.

4.6 Layout

Good layout leads to good performance. In the design, several methods are utilized to

make layout better.

For critical components like differential pairs and current mirrors, they should be

laid out carefully. The first technique is common-centroid layout. A common-centroid

array should cancel systematic mismatches due to gradients. Also, making devices larger

is another way to improve matching, while the trade-off is the increased layout area.

Second Chip Design 61

It is also useful to use small transistors to create many fingers so that transistors with

large W/L can be achieved. Metal instead of poly should be used to interconnect gates if

possible. In addition, guard rings are often utilized to reduce substrate coupling noise.

NMOS rings should be tied to ground while PMOS rings should be connected to the

power supply. It is also very important to use as many contacts as possible to increase

the conductance between two layers.

For passive components, poly-poly capacitors with the typical capacitance density

of around 900 aF/µm2 and high resistance (Hi-Res) resistors with typical resistivity of

around 1192 Ω/ä are widely used in this AFE. To have better layout, both of them have

common-centroid layouts to improve matching. Guard rings are also used to protect

capacitors and resistors.

Fig. 4.20 shows the final layout of the whole AFE. The active area is around 1050 µm

× 1050 µm, and the total area is around 1500 µm × 1500 µm.

Second Chip Design 62

Figure 4.20. Layout of the second chip.

63

5 Conclusions

This thesis has demonstrated two AFEs for simultaneous measurement of ECG and

RR signals. The first chip has been fabricated on the OnSemi 0.5 µm CMOS process and

it has been tested. Both simulation and measurement results prove the functionality of

it. One conference paper emerged based on the results of the first chip54. The second

AFE has been designed to correct the defects in the first chip, and it will be submitted

for fabrication on the same process. In this chapter, we will summarize the results of the

main blocks in the AFEs and discuss the future work.

ECG Amplifier: Two different preamplifiers have been designed in each AFE. In the

first chip, the preamplifier used is a modified version of the amplifier presented in42,43.

Since the frequencies of bioelectrical signals are low (typical ECG bandwidth of 1 Hz to

100 Hz and RR bandwidth of 0.5 Hz)24, the MOS pseudo resistors are used to generate

a low cutoff frequency. The driven-ground circuit is also utilized to boost the CMRR.

Measured results show that the 1/ f corner frequency is around 25 Hz and the thermal

noise PSD is around 254 nV/Hz1/2, which are in good agreement with simulations. The

minimum detectable signal is around 38 µV in the band of 1-150 Hz. The mid-band gain

is 40 dB as expected.

Conclusions 64

However, certain design defects were found during the measurement. The DC volt-

age at the output of the amplifier shifted. This is because the leakage current exists in

the pseudo resistor. To overcome this problem, another preamplifier is designed in the

second AFE. For the second amplifier, switched capacitor is used to replace the MOS

pseudo resistor so that the leakage current in the pseudo resistor does not have effect on

the circuit. Another advantage of using switched capacitor resistors is that the effective

resistance can be easily controlled by changing the clock frequency fed to switched ca-

pacitor. In this case, the high-pass cutoff frequency can be controlled during operation.

It is possible to control the effective resistance of pseudo resistor by changing the gate

bias voltage23,55, but this method is much less straightforward. Also in this preampli-

fier, the second gain stage has programmable gain. Besides, the driven-ground circuit is

updated since we found the original driven-ground circuit did not work well. Source fol-

lowers and on-chip resistors are used to detect the common-mode voltage. Two OTAs,

with the one in unity feedback acting as a buffered resistor, are used to compare the

sensed common-mode voltage with a reference voltage and amplify the difference be-

tween them. An op-amp in unity feedback is then used to act as a buffer so that the

driven-ground circuit is able to drive the electrode impedance.

Second Gain Stage for RR:

In the first AFE, the circuit uses two WLR OTAs, with the one in unity feedback be-

having as a resistor. The gain is set by the ratio of two bias currents, which is set to 50

(34 dB) in the chip. However, it is not good to set the ratio of bias currents too high since

the OTA in unity feedback may be saturated. The second chip is a fully differential AFE,

and a fully differential version of the original RR gain stage is used. The gain is set to

10 (20 dB) which is more reasonable. To provide more gain in RR path, the circuit also

Conclusions 65

uses an op-amp to amplify the RR signal. The gain of the op-amp stage is set by the ratio

of capacitors in the feedback loop, which is set to 20 (26 dB) in the AFE, resulting the

total RR gain of 40 dB+20 dB+26 dB-4 dB=82 dB. Here the -4 dB term comes from the

conversion loss of the mixer.

Future Work: The ultimate goal of this thesis is to use the AFE designed in a wearable

biopotential measurement system, which can continuously monitor high-quality ECG

and respiratory condition. The chest band with three electrodes collects the signals. The

AFE designed can amplify the sensed signals and an ADC is also needed to digitize the

amplified outputs. Digitized signals shall be buffered locally and transmitted to a wrist

band or a personal computer through low-power Bluetooth.

The biopotential measurement system can be further used in several applications.

One application is to use the system to discriminate between the non-epileptic seizures

and epileptic seizures56. This technique has been demonstrated on a limited number

of patients at University Hospitals (UH) Case Medical Center (CMC) to identify psy-

chogenic non-epileptic seizures (PNES).

Appendix 66

Appendix A

PCB for testing the AFE

This appendix presents the schematic and layout of the PCB board used to test the

first AFE.

1

1

2

2

3

3

4

4

D D

C C

B B

A A

Title

Number RevisionSize

A4

Date: 2016/6/7 Sheet ofFile: C:\Users\..\chiptest1.SchDoc Drawn By:

B1

51K101-400A4

C9CAP 1uF 6.3V 0805(2012)

C10CAP 100nF 6.3V 0805(2012)

GND

C11CAP 1uF 6.3V 0805(2012)

C12CAP 100nF 6.3V 0805(2012)

GND

C1

CAP 1nF 6.3V 0805(2012)

C2

CAP 1nF 6.3V 0805(2012)

C3

CAP 1nF 6.3V 0805(2012)

C4

CAP 1nF 6.3V 0805(2012)

C5

CAP 1nF 6.3V 0805(2012)

C6

CAP 1nF 6.3V 0805(2012)

C7

CAP 1nF 6.3V 0805(2012)

C8

CAP 1nF 6.3V 0805(2012)

GND

GND

GND

C16CAP 100nF 6.3V 0805(2012)

GND

VDD APVDD

APVDD

VDDVDD

5

4

3

2

1

13

12

11

10 17

16

9

614

715

8

D-Sub

212612043

SCLK1

CS2

GND 3

AIN0 4

AIN1 5

AIN2 6

AIN3 7

VDD8

DOUT/DRDY9 DIN10

U1

ADS1118IDGSTGND

APVDD

Iout

IoutQout

Qout

ECGout

ECGout

SC1SC2SC3SC4

GND GND

C18CAP 100nF 6.3V 0805(2012)

GND

APVDD

0.75V

0.75V

B3

51K101-400A4

GND

clk

clk

PADFOL

PADFOL

C20CAP 100nF 6.3V 0805(2012)

GND

APVDD

C13

CAP 100nF 6.3V 0805(2012)C14

CAP 100nF 6.3V 0805(2012)

C24

CAP 100nF 6.3V 0805(2012)

C15

CAP 100nF 6.3V 0805(2012)

GND

GND GND

GND

C17CAP 100nF 6.3V 0805(2012)

C19CAP 100nF 6.3V 0805(2012)

C21CAP 100nF 6.3V 0805(2012)

GND GND GND

LL

RALA

123

P5

Header 3

J1

Socket

C22CAP 100nF 6.3V 0805(2012)

GND

APVDD

C23CAP 100nF 6.3V 0805(2012)

GND

123

P6

Header 3

J2

Socket200nA

200nA

5nA

5nA

123

P1

Header 3GND

APVDDSC1

123

P2

Header 3GND

APVDDSC2

123

P3

Header 3GND

APVDDSC3

123

P4

Header 3GND

APVDDSC4

123

P7

Header 3

123

P8

Header 3

123

P9

Header 3

LL

LA

RA12

P11

Header 2

12

P12

Header 2

123

P10

Header 3GND

APVDDCLK_en

CLK_en

J6

SocketJ7

Socket

J8Socket

GND

J9Socket

GND

SC11

SC22

SC33

SC44

55

66

77

88

99

1010

1111

1212

1313

1414

APGND15

APVDD16

PADFOL17

VDDD18

GNDD19

CLK_en20 0.75v 21clk 22VDDA 23Qout 24Gnda 25cap1 26cap2 27cap3 28cap4 29cap5 30cap6 31cap7 32cap8 335nA 34Iout 35ECGout 36RA 37LA 38LL 39200nA 40CH1

chip1

1234

P13

Header 4

B2

51K101-400A4

VDDin APVDDin

J11Socket

J10Socket

1

32

Pot3pot11

32

Pot2pot11

32

Pot1pot1 1

32

Pot4pot1

GNDAPVDD

21R2

21

R5

21

R7

21

R8

21

R621

R421

R3

21R1

D1

G 2

S3

Tran1

2n7000

D1 G 2S3Tran2

2n7000

12

c31cap33u

12

c32cap33u

J21

Socket

J22

SocketJ20Socket

LLB

LLB

RAB

RAB

LAB

LAB

PIB101

PIB10MH

COB1PIB201

PIB20MH

COB2PIB301

PIB30MH

COB3

PIC101 PIC102

COC1 PIC201 PIC202

COC2 PIC301 PIC302

COC3 PIC401 PIC402

COC4PIC501 PIC502

COC5PIC601 PIC602

COC6 PIC701 PIC702

COC7 PIC801 PIC802

COC8

PIC901

PIC902COC9

PIC1001

PIC1002COC10

PIC1101

PIC1102COC11

PIC1201

PIC1202COC12

PIC1301PIC1302

COC13

PIC1401PIC1402

COC14

PIC1501PIC1502

COC15

PIC1601

PIC1602COC16

PIC1701

PIC1702COC17 PIC1801

PIC1802COC18

PIC1901

PIC1902COC19 PIC2001

PIC2002COC20

PIC2101

PIC2102COC21 PIC2201

PIC2202COC22

PIC2301

PIC2302COC23

PIC2401PIC2402

COC24

PIc3101

PIc3102COc31

PIc3201

PIc3202COc32

PICH101

PICH102

PICH103

PICH104

PICH105

PICH106

PICH107

PICH108

PICH109

PICH1010

PICH1011

PICH1012

PICH1013

PICH1014

PICH1015

PICH1016

PICH1017

PICH1018

PICH1019

PICH1020 PICH1021

PICH1022

PICH1023

PICH1024

PICH1025

PICH1026

PICH1027

PICH1028

PICH1029

PICH1030

PICH1031

PICH1032

PICH1033

PICH1034

PICH1035

PICH1036

PICH1037

PICH1038

PICH1039

PICH1040

COCH1

PID0Sub01

PID0Sub02

PID0Sub03

PID0Sub04

PID0Sub05

PID0Sub06

PID0Sub07

PID0Sub08

PID0Sub09

PID0Sub010

PID0Sub011

PID0Sub012

PID0Sub013

PID0Sub014

PID0Sub015

PID0Sub016

PID0Sub017

COD0Sub

PIJ101

COJ1

PIJ201

COJ2

PIJ601COJ6

PIJ701

COJ7

PIJ801COJ8

PIJ901COJ9

PIJ1001COJ10

PIJ1101COJ11

PIJ2001COJ20

PIJ2101

COJ21

PIJ2201

COJ22

PIP101

PIP102

PIP103

COP1

PIP201

PIP202

PIP203

COP2

PIP301

PIP302

PIP303

COP3

PIP401

PIP402

PIP403

COP4

PIP501

PIP502

PIP503

COP5

PIP601

PIP602

PIP603

COP6

PIP701

PIP702

PIP703

COP7

PIP801

PIP802

PIP803

COP8

PIP901

PIP902

PIP903

COP9

PIP1001

PIP1002

PIP1003

COP10

PIP1101

PIP1102

COP11PIP1201

PIP1202

COP12

PIP1301

PIP1302

PIP1303

PIP1304

COP13

PIPot101

PIPot102

PIPot103 COPot1

PIPot201

PIPot202

PIPot203 COPot2

PIPot301

PIPot302

PIPot303 COPot3

PIPot401

PIPot402

PIPot403 COPot4

PIR101 PIR102

COR1PIR201 PIR202

COR2

PIR301 PIR302

COR3PIR401 PIR402

COR4

PIR501

PIR502COR5

PIR601

PIR602COR6

PIR701

PIR702COR7

PIR801

PIR802COR8

PITran101

PITran102

PITran103

COTran1

PITran201

PITran202

PITran203

COTran2

PIU101

PIU102

PIU103

PIU104

PIU105

PIU106

PIU107

PIU108

PIU109

PIU1010

COU1

PIC1702

PICH1021

PIPot101

NL0075V

PICH1034

PIP602NL5nA

PICH1040

PIP502

NL200nA

PIC1102 PIC1202

PIC1301

PIC1602 PIC1802 PIC2002 PIC2202

PIC2401

PICH1016

PIP103 PIP203

PIP303 PIP403

PIP1003

PIP1202

PIPot103 PIPot203 PIPot303 PIPot403

PIU108NLAPVDD

PIB201 PIP1201NLAPVDDin

PIB301

PICH1022NLclk

PICH1020

PIP1002

NLCLK0en

PICH1036PIJ601

PIU106NLECGout

PIB10MH PIB20MH PIB30MH

PIC102

PIC202

PIC302

PIC402

PIC502

PIC602

PIC702

PIC802

PIC901 PIC1001 PIC1101 PIC1201

PIC1302

PIC1402 PIC1501

PIC1601PIC1701

PIC1801PIC1901

PIC2001PIC2101

PIC2201PIC2301

PIC2402

PICH1015

PICH1019

PICH1025

PIJ901 PIJ1001 PIJ1101

PIP101 PIP201

PIP301 PIP401

PIP1001

PIPot102 PIPot202 PIPot302 PIPot402

PIU103

PIU107

PICH1035 PIJ701

PIU105NLIout

PICH1038

PIP802NLLA

PIP801

PIR402NLLAB

PICH1039

PIP902NLLL

PIP901

PIR701

PIR802

PITran103

PITran203

NLLLB

PIC101PICH1026

PIC201PICH1027

PIC301PICH1028

PIC401PICH1029

PIC501PICH1030

PIC601PICH1031

PIC701PICH1032

PIC801PICH1033

PIC2102

PIP503

PIPot301 PIC2302

PIP603

PIPot401

PIc3101PIJ2101

PIc3102

PIR101

PIc3201

PIR301

PIc3202PIJ2201

PICH105

PICH106

PICH107

PICH108

PICH109

PICH1010

PICH1011

PICH1012

PICH1013

PICH1014

PID0Sub01

PID0Sub02

PID0Sub03

PID0Sub04

PID0Sub05

PID0Sub06

PID0Sub07

PID0Sub08

PID0Sub09PIP703

PID0Sub010

PIP803

PID0Sub011

PIP903

PID0Sub012

PID0Sub013

PID0Sub014

PID0Sub015

PID0Sub016

PID0Sub017

PIJ101 PIP501 PIJ201 PIP601

PIJ2001

PITran102

PITran202

PIP1301 PIU102

PIP1302 PIU109

PIP1303 PIU1010

PIP1304 PIU101

PIR102 PIR201PIR502

PIR302 PIR401PIR601

PIR501

PIR702 PITran101

PIR602

PIR801 PITran201

PIC1902

PICH1017

PIPot201

NLPADFOLPICH1024

PIJ801

PIU104NLQout

PICH1037

PIP702NLRA

PIP701

PIR202NLRAB

PICH101PIP102NLSC1

PICH102

PIP202NLSC2

PICH103

PIP302

NLSC3

PICH104

PIP402

NLSC4

PIC902 PIC1002

PIC1401 PIC1502PICH1018 PICH1023

PIP1102

NLVDD

PIB101 PIP1101NLVDDin

Figure A.1. Schematic of the PCB board.

Appendix 67

PAB10MH

PAB101

COB1

PAB20MH

PAB201

COB2PAB30MH

PAB301

COB3

PAC102PAC101 COC1

PAC202PAC201 COC2PAC302PAC301

COC3

PAC402PAC401 COC4

PAC502PAC501 COC5

PAC602PAC601 COC6

PAC702PAC701 COC7

PAC802PAC801 COC8

PAC902PAC901

COC9PAC1002PAC1001

COC10

PAC1102PAC1101

COC11

PAC1202PAC1201COC12

PAC1302PAC1301

COC13

PAC1402PAC1401

COC14

PAC1502 PAC1501

COC15

PAC1602PAC1601COC16

PAC1702PAC1701COC17

PAC1802PAC1801

COC18

PAC1902PAC1901

COC19

PAC2002PAC2001

COC20

PAC2102

PAC2101COC21

PAC2202PAC2201 COC22

PAC2302PAC2301

COC23

PAC2402PAC2401

COC24

PAc3101

PAc3102

COc31PAc3201

PAc3202COc32

PACH1021

PACH1022

PACH1023PACH1024

PACH1025PACH1026

PACH1027

PACH1028

PACH1029

PACH1030

PACH1031

PACH1032

PACH1033

PACH1034

PACH1035

PACH1036

PACH1037

PACH1038

PACH1039

PACH1040

PACH1020

PACH1019PACH1018PACH1017

PACH1016PACH1015PACH1014

PACH1013

PACH1012

PACH1011

PACH1010

PACH109

PACH108

PACH107

PACH106

PACH105

PACH104

PACH103

PACH102

PACH101

COCH1

PAD0Sub09

PAD0Sub014

PAD0Sub015

PAD0Sub01

PAD0Sub08

PAD0Sub07

PAD0Sub02

PAD0Sub03PAD0Sub04

PAD0Sub05

PAD0Sub06PAD0Sub013PAD0Sub012

PAD0Sub011

PAD0Sub010

PAD0Sub016

PAD0Sub017

COD0Sub

PAJ101COJ1

PAJ201COJ2

PAJ601

COJ6

PAJ701

COJ7

PAJ801

COJ8

PAJ901 COJ9

PAJ1001

COJ10

PAJ1101 COJ11

PAJ2001 COJ20

PAJ2101

COJ21

PAJ2201COJ22

PAP103PAP102PAP101COP1

PAP203PAP202PAP201COP2

PAP303PAP302PAP301COP3

PAP403PAP402PAP401COP4

PAP503PAP502PAP501COP5

PAP603 PAP602 PAP601COP6

PAP703PAP702PAP701

COP7PAP803PAP802PAP801

COP8

PAP903PAP902PAP901

COP9

PAP1003PAP1002

PAP1001

COP10

PAP1102PAP1101

COP11

PAP1202PAP1201

COP12PAP1301

PAP1302PAP1303

PAP1304

COP13

PAPot103PAPot101PAPot102COPot1

PAPot203PAPot201PAPot202

COPot2

PAPot303PAPot301

PAPot302

COPot3

PAPot403PAPot401PAPot402

COPot4

PAR101

PAR102

COR1

PAR201

PAR202 COR2

PAR301

PAR302

COR3

PAR401

PAR402COR4 PAR501 PAR502 COR5PAR601 PAR602COR6

PAR701 PAR702

COR7

PAR801 PAR802

COR8

PATran101PATran102PATran103

COTran1

PATran201 PATran202 PATran203

COTran2

PAU101PAU102

PAU103PAU104PAU105

PAU1010PAU109

PAU108PAU107PAU106

COU1

PAC1702

PACH1021

PAPot101

PACH1034

PAP602

PACH1040

PAP502

PAC1102

PAC1202

PAC1301

PAC1602

PAC1802

PAC2002

PAC2202

PAC2401

PACH1016

PAP103

PAP203

PAP303

PAP403

PAP1003PAP1202

PAPot103

PAPot203

PAPot303

PAPot403

PAU108

PAB201PAP1201

PAB301

PACH1022

PACH1020

PAP1002

PACH1036

PAJ601

PAU106

PAB10MH

PAB20MH

PAB30MH

PAC102

PAC202

PAC302

PAC402

PAC502

PAC602

PAC702

PAC802

PAC901PAC1001

PAC1101

PAC1201

PAC1302

PAC1402

PAC1501

PAC1601

PAC1701

PAC1801

PAC1901

PAC2001

PAC2101

PAC2201

PAC2301

PAC2402

PACH1015

PACH1019

PACH1025

PAJ901

PAJ1001

PAJ1101

PAP101

PAP201

PAP301

PAP401

PAP1001 PAPot102

PAPot202

PAPot302

PAPot402

PAU103PAU107

PACH1035

PAJ701

PAU105

PACH1038

PAP802PAP801

PAR402

PACH1039

PAP902PAP901

PAR701PAR802

PATran103PATran203

PAC101

PACH1026

PAC201

PACH1027PAC301

PACH1028PAC401

PACH1029 PAC501

PACH1030PAC601

PACH1031PAC701PACH1032

PAC801PACH1033

PAC2102

PAP503 PAPot301

PAC2302PAP603

PAPot401

PAc3101

PAJ2101

PAc3102

PAR101

PAc3201

PAR301

PAc3202 PAJ2201

PAD0Sub09

PAP703

PAD0Sub010

PAP803

PAD0Sub011

PAP903PAJ101

PAP501

PAJ201

PAP601

PAJ2001

PATran102PATran202

PAP1301

PAU102

PAP1302

PAU109

PAP1303

PAU1010

PAP1304PAU101

PAR102

PAR201PAR502

PAR302

PAR401 PAR601 PAR501

PAR702

PATran101

PAR602

PAR801

PATran201

PAC1902

PACH1017

PAPot201

PACH1024

PAJ801

PAU104

PACH1037

PAP702PAP701

PAR202

PACH101

PAP102

PACH102PAP202PACH103

PAP302 PACH104

PAP402

PAC902PAC1002

PAC1401 PAC1502PACH1018 PACH1023

PAP1102

PAB101

PAP1101

Figure A.2. Layout of the PCB board.

Bibliography 68

Complete References

[1] Reid R Harrison and Cameron Charles. A low-power low-noise CMOS amplifier forneural recording applications. Solid-State Circuits, IEEE Journal of, 38(6):958–965,2003.

[2] Liang Kai, Xu Zhang, Yuan Wang, Huang Suibiao, Guan Ning, Pang Wangyong,Li Bin, and Hongda Chen. A system of portable ECG monitoring based on Blue-tooth mobile phone. In IT in Medicine and Education (ITME), 2011 InternationalSymposium on, volume 2, pages 309–312. IEEE, 2011.

[3] Meiying Wen, Yayu Cheng, and Ye Li. A full custom analog front-end for long-timeECG monitoring. In Engineering in Medicine and Biology Society (EMBC), 201335th Annual International Conference of the IEEE, pages 3234–3237. IEEE, 2013.

[4] Martin J Burke and Denis T Gleeson. A micropower dry-electrode ECG preamplifier.Biomedical Engineering, IEEE Transactions on, 47(2):155–162, 2000.

[5] Dariush Mozaffarian, Emelia J Benjamin, Alan S Go, Donna K Arnett, Michael JBlaha, Mary Cushman, Sandeep R Das, Sarah de Ferranti, Jean-Pierre Després,Heather J Fullerton, et al. Heart disease and stroke statistics-2016 update a reportfrom the American Heart Association. Circulation, pages CIR–0000000000000350,2015.

[6] M Ayat, K Assaleh, and H Al-Nashash. Prototype of a standalone fetal ECG monitor.In Industrial Electronics & Applications (ISIEA), 2010 IEEE Symposium on, pages617–622. IEEE, 2010.

[7] World Health Organization et al. Pocket book of hospital care for children:guidelines for the management of common illnesses with limited resources.Geneva: World Health Organization, 2005.

[8] Timothy J Hodgetts, Gary Kenward, Ioannis G Vlachonikolis, Susan Payne, andNicolas Castle. The identification of risk factors for cardiac arrest and formulationof activation criteria to alert a medical emergency team. Resuscitation, 54(2):125–131, 2002.

[9] Jerry P Nolan, Jasmeet Soar, David A Zideman, Dominique Biarent, Leo L Bossaert,Charles Deakin, Rudolph W Koster, Jonathan Wyllie, Bernd Böttiger, et al. Euro-pean resuscitation council guidelines for resuscitation 2010 section 1. executivesummary. Resuscitation, 81(10):1219–1276, 2010.

Bibliography 69

[10] Jason J Liu, Ming-Chun Huang, Wenyao Xu, Xiaoyi Zhang, Luke Stevens, NabilAlshurafa, and Majid Sarrafzadeh. Breathsens: A continuous on-bed respiratorymonitoring system with torso localization using an unobtrusive pressure sensingarray. Biomedical and Health Informatics, IEEE Journal of, 19(5):1682–1688, 2015.

[11] Michelle A Cretikos, Rinaldo Bellomo, Ken Hillman, Jack Chen, Simon Finfer, andArthas Flabouris. Respiratory rate: the neglected vital sign. Medical Journal ofAustralia, 188(11):657, 2008.

[12] Jackie McBride, Debbie Knight, Jo Piper, and Gary B Smith. Long-term effect ofintroducing an early warning score on respiratory rate charting on general wards.Resuscitation, 65(1):41–44, 2005.

[13] Jacqueline Hogan. Why don’t nurses monitor the respiratory rates of patients?British Journal of Nursing, 15(9), 2006.

[14] Michelle Cretikos, Jack Chen, Ken Hillman, Rinaldo Bellomo, Simon Finfer, ArthasFlabouris, Merit Study Investigators, et al. The objective medical emergency teamactivation criteria: a case–control study. Resuscitation, 73(1):62–72, 2007.

[15] Paolo Bonato. Wearable sensors and systems. Engineering in Medicine and BiologyMagazine, IEEE, 29(3):25–36, 2010.

[16] Philips Med. Syst. North Amer. Corp. TEN-HMS study demonstrates clinical andfinancial efficacy of home telemonitoring. Sep. 2003.

[17] Paolo Bonato et al. Wearable sensors/systems and their impact on biomedical en-gineering. IEEE Engineering in Medicine and Biology Magazine, 22(3):18–20, 2003.

[18] Ake Grenvik, Stanley Ballou, Edward McGinley, J Eugene Millen, Wils L Cooley, andPeter Safar. Impedance pneumography: Comparison between chest impedancechanges and respiratory volumes in II healthy volunteers. Chest Journal, 62(4):439–443, 1972.

[19] Sameer P Kelkar, Niranjan D Khambete, and Sujata S Agashe. Development ofmovement artefacts free breathing monitor. J. Instrum. Soc. India, 38:34–43, 2004.

[20] Min Chen, Sergio Gonzalez, Athanasios Vasilakos, Huasong Cao, and Victor CMLeung. Body area networks: A survey. Mobile networks and applications, 16(2):171–193, 2011.

[21] Mahmood Khayatzadeh, Xiaoyang Zhang, Jun Tan, Wen-Sin Liew, and Yong Lian.A 0.7 V 17.4 µW 3-lead wireless ECG SoC. Biomedical Circuits and Systems, IEEETransactions on, 7(5):583–592, 2013.

Bibliography 70

[22] Hyejung Kim, Sunyoung Kim, Nick Van Helleputte, Antonio Artes, Mario Konijnen-burg, Jos Huisken, Chris Van Hoof, and Refet Firat Yazicioglu. A configurable andlow-power mixed signal SoC for portable ECG monitoring applications. BiomedicalCircuits and Systems, IEEE Transactions on, 8(2):257–267, 2014.

[23] Xiaodan Zou, Xiaoyuan Xu, Libin Yao, and Yong Lian. A 1 V 450 nW fully integratedprogrammable biomedical sensor interface chip. Solid-State Circuits, IEEE Journalof, 44(4):1067–1077, 2009.

[24] Gari D Clifford and Matt B Oefinger. ECG acquisition, storage, transmission, andrepresentation.

[25] Bedrich J Hosticka, Robert W Brodersen, and Paul R Gray. MOS sampled data recur-sive filters using switched capacitor integrators. Solid-State Circuits, IEEE Journalof, 12(6):600–608, 1977.

[26] Tobi Delbruck and Carver A Mead. Adaptive photoreceptor with wide dy-namic range. In Circuits and Systems, 1994. ISCAS’94., 1994 IEEE InternationalSymposium on, volume 4, pages 339–342. IEEE.

[27] Nika Zolfaghari, Shahini Sirikantharajah, Mohsen Shafeie, and Kristiina M ValterMcConville. Adaptive filtering technique and comparison of PS25015A dry elec-trodes and two different Ag/AgCl wet electrodes for wearable ECG applications.Sensors & Transducers, 184(1):84, 2015.

[28] Yu Wang, WeiHua Pei, Kai Guo, Qiang Gui, XiaoQian Li, HongDa Chen, and Jian-Hong Yang. Dry electrode for the measurement of biopotential signals. ScienceChina Information Sciences, 54(11):2435–2442, 2011.

[29] F Refet, Chris Van Hoof, Robert Puers, et al. Biopotential readout circuits forportable acquisition systems. Springer Science & Business Media, 2008.

[30] Ebrahim Nemati, M Jamal Deen, and Tapas Mondal. A wireless wearable ECGsensor for long-term applications. Communications Magazine, IEEE, 50(1):36–43,2012.

[31] A Searle and L Kirkup. A direct comparison of wet, dry and insulating bioelectricrecording electrodes. Physiological measurement, 21(2):271, 2000.

[32] Yu M Chi and Gert Cauwenberghs. Wireless non-contact EEG/ECG electrodesfor body sensor networks. In Body Sensor Networks (BSN), 2010 InternationalConference on, pages 297–301. IEEE, 2010.

Bibliography 71

[33] Woradorn Wattanapanitch, Michale Fee, and Rahul Sarpeshkar. An energy-efficientmicropower neural recording amplifier. Biomedical Circuits and Systems, IEEETransactions on, 1(2):136–147, 2007.

[34] Honglei Wu and Yong Ping Xu. A 1V 2.3µW biomedical signal acquisition IC. In 2006IEEE International Solid State Circuits Conference-Digest of Technical Papers.

[35] Roy H Olsson, Derek L Buhl, Anton M Sirota, Gyorgy Buzsaki, and Kensall DWise. Band-tunable and multiplexed integrated circuits for simultaneous record-ing and stimulation with microelectrode arrays. Biomedical Engineering, IEEETransactions on, 52(7):1303–1311, 2005.

[36] Tim Denison, Kelly Consoer, Wesley Santa, Al-Thaddeus Avestruz, John Cooley, andAndy Kelly. A 2µW 100 nV/sqrtHz chopper-stabilized instrumentation amplifier forchronic measurement of neural field potentials. Solid-State Circuits, IEEE Journalof, 42(12):2934–2945, 2007.

[37] Bruce B Winter and John G Webster. Driven-right-leg circuit design. BiomedicalEngineering, IEEE Transactions on, (1):62–66, 1983.

[38] AC Metting van Rijn. The Modelling of Biopotential Recordings and its Implicationsfor Instrumentation Design. PhD thesis, TU Delft, Delft University of Technology,1993.

[39] David Prutchi and Michael Norris. Design and development of medical electronicinstrumentation: a practical perspective of the design, construction, and test ofmedical devices. John Wiley & Sons, 2005.

[40] Shishir Dash, Kirk H Shelley, David G Silverman, and Ki H Chon. Estimation of res-piratory rate from ECG, photoplethysmogram, and piezoelectric pulse transducersignals: a comparative study of time–frequency methods. IEEE Transactions onBiomedical Engineering, 57(5):1099–1107, 2010.

[41] Michael M Laks, Robert Arzbaecher, James J Bailey, David B Geselowitz, and Alan SBerson. Recommendations for safe current limits for electrocardiographs a state-ment for healthcare professionals from the committee on electrocardiography,American Heart Association. Circulation, 93(4):837–839, 1996.

[42] Leon Fay, Vinith Misra, and Rahul Sarpeshkar. A micropower electrocardiogramamplifier. Biomedical Circuits and Systems, IEEE Transactions on, 3(5):312–320,2009.

Bibliography 72

[43] Lorenzo Turicchia, Bruno Do Valle, Jose L Bohorquez, William R Sanchez, VinithMisra, Leon Fay, Maziar Tavakoli, and Rahul Sarpeshkar. Ultralow-power elec-tronics for cardiac monitoring. Circuits and Systems I: Regular Papers, IEEETransactions on, 57(9):2279–2290, 2010.

[44] Mr DK Shedge, Mr DA Itole, Mr MP Gajare, and PW Wani. Analysis and design ofCMOS source followers and super source follower. 2013.

[45] Rahul Sarpeshkar, Richard F Lyon, and Carver Mead. A low-power wide-linear-range transconductance amplifier. Analog Integrated Circuits and SignalProcessing, 13(1-2):123–151, 1997.

[46] Robert W Landee. Electronic designers. 1957.

[47] Paul R Gray, Paul Hurst, Robert G Meyer, and Stephen Lewis. Analysis and design ofanalog integrated circuits. Wiley, 2001.

[48] Tobi Delbruck. Bump circuits for computing similarity and dissimilarity of ana-log voltages. In Neural Networks, 1991., IJCNN-91-Seattle International JointConference on, volume 1, pages 475–479. IEEE, 1991.

[49] Single-Lead, Heart Rate Monitor Front End, 2012.

[50] Yi Chen, Anirban Basu, Lei Liu, Xiaodan Zou, Ramamoorthy Rajkumar, Gavin Stew-art Dawe, and Minkyu Je. A digitally assisted, signal folding neural recording ampli-fier. Biomedical Circuits and Systems, IEEE Transactions on, 8(4):528–542, 2014.

[51] Eduard Säckinger and Walter Guggenbuhl. A versatile building block: the CMOSdifferential difference amplifier. Solid-State Circuits, IEEE Journal of, 22(2):287–294, 1987.

[52] Scott K Arfin, Soumyajit Mandal, and Rahul Sarpeshkar. Dynamic-range analysisand maximization of micropower Gm-C bandpass filters by adaptive biasing. InCircuits and Systems, 2009. ISCAS 2009. IEEE International Symposium on, pages2954–2957. IEEE, 2009.

[53] Naveen Verma, Ali Shoeb, Jose Bohorquez, Joel Dawson, John Guttag, and Anan-tha P Chandrakasan. A micro-power EEG acquisition SoC with integrated featureextraction processor for a chronic seizure detection system. 2010.

[54] Jifu Liang, Shixiong Li, Ali Nikoofard, and Soumyajit Mandal. A low-power receiverfor simultaneous electrocardiogram and respiration rate detection. Poster sessionpresented at IEEE International Symposium on Circuits and Systems, Montreal,Canada, May 2016.

Bibliography 73

[55] Xiaoyuan Xu, Xiaodan Zou, Libin Yao, and Yong Lian. A 1 V 450 nW fully integratedbiomedical sensor interface system. In VLSI Circuits, 2008 IEEE Symposium on,pages 78–79. IEEE, 2008.

[56] Sándor Beniczky, Isa Conradsen, Mihai Moldovan, Poul Jennum, Martin Fabricius,Krisztina Benedek, Noémi Andersen, Helle Hjalgrim, and Peter Wolf. Automateddifferentiation between epileptic and nonepileptic convulsive seizures. Annals ofneurology, 77(2):348–351, 2015.